510(k) Data Aggregation

    K Number
    K200695

    Validate with FDA (Live)

    Device Name
    MATRx plus
    Date Cleared
    2020-08-24

    (160 days)

    Product Code
    Regulation Number
    868.2375
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MATRx plus is indicated for use by a lay person in a home and hospital use under the direction of a Health Care Professional (HCP).

    MATRx plus records the following data: patient respiratory nasal airflow, snoring, blood oxygen saturation, pulse, respiratory effort and body position during sleep.

    MATRx plus uses these recordings to produce a report for the HCP that may aid in the diagnosis and assessment of sleep disordered breathing for adult patients.

    The MATRx plus device may also be used with an automated mandibular positioner that uses feedback control to record changes in the patient's respiratory status related to repositioning of the mandible during an overnight study.

    MATRx plus uses these recordings to produce a report for the HCP that can be used to prospectively identify patients with mild to moderate obstructive sleep apnea who may be suitable for oral appliance therapy and to recommend a target mandibular position.

    The use of the device does not replace the need for follow-up testing to determine the initial and ongoing effectiveness of the therapy as recommended by clinical practice guidelines.

    Device Description

    MATRx plus is a ventilatory effort recorder and closed loop auto-titration device for oral appliances. MATRx plus is a 5-channel battery-powered respiratory pressure sensor and oximetry system. MATRx plus provides recordings of respiratory pressure, snoring, respiratory effort, pulse rate, oxygen saturation, and body position during sleep.

    The device is worn on the patient's abdomen attached to a reusable effort belt and all relevant respiratory information during sleep is collected via nasal cannula, pulse oximetry module, and respiratory effort sensor. The disposable plastic nasal cannula is connected to the MATRx plus recorder and fixed at the patient's nose. The cannula is dual lumen and functions to individually sample the pressure from each naris. The oximetry sensor is fixed at the patient's finger and connects directly to the device. The Recorder receives input from the sensors and wirelessly transmits the data to a bedside Tablet during the study. The Tablet is set up through a connection to the web portal by the healthcare professional (HCP) prior to deploying the device to the patient for home use. The Tablet runs the MATRx plus application and records and stores the data. The MATRx plus application on the Tablet is also used to guide the user through the set up and conduct of the study through a stepwise user interface. At the end of a recording, the Tablet advises the patient if sufficient data for analysis were recorded during the night.

    During an auto-titration study, the patient wears temporary Titration Trays connected to a mandibular positioner. During the study, the device moves the mandible in response to respiratory events in order to determine if the patient is a suitable candidate for oral appliance therapy.

    After recording, the MATRx plus must be returned to the HCP. The data are automatically uploaded to the secure Portal where they can be accessed and downloaded to the Data Viewer. The Data Viewer can generate a report with the recorded and analyzed data (respiratory pressure, respiratory effort, pulse rate, oxygen saturation) to aid in diagnosis (when used as a ventilatory effort recorder) or to assist the HCP in the clinical management of oral appliance therapy for patients with obstructive sleep apnea.

    The device is intended to be used on adult patients, upon referral from their healthcare provider.

    The device is not to be used as an apnea monitor or in a life supporting or life sustaining situation and must not be used in the vicinity of an MRI device.

    AI/ML Overview

    The MATRx plus device underwent non-clinical and clinical testing to demonstrate its substantial equivalence to its predicate device (K191925).

    Here's an breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Single Arm Study (Prediction of Outcome for Mild/Moderate Sleep Apnea Patients)
    Sensitivity (LCL) > 0.60 (against established clinical endpoints for responder/non-responder prediction)Sensitivity = 97.1% (LCL = 84.7%)
    Target Accuracy (LCL) > 0.60Target Accuracy = 90.9% (LCL = 75.7%)
    Cross-over Study (Non-inferiority in Predictive Accuracy compared to Predicate Device)
    Predictive accuracy non-inferiority margin: 0.125 (compared to predicate device)Predictive accuracy demonstrated to be non-inferior to the predicate device with a negligible mean target accuracy difference of 0.17 mm. ("meaning the difference was less than 0.125, but not explicitly stated what percentage it was exactly")
    Mean target accuracy difference < 0.125 mm (compared to predicate device for recommended target mandibular position)Mean target accuracy difference = 0.17 mm

    2. Sample Size and Data Provenance for Test Set:

    The provided document does not explicitly state the exact sample size for the test sets in either the single-arm or cross-over study. It mentions the "mild/moderate patient population" for the single-arm study.
    The provenance of the data is not specified in terms of country of origin. The studies are described as prospective.

    3. Number of Experts and Qualifications for Ground Truth:

    This information is not provided in the document.

    4. Adjudication Method for Test Set:

    This information is not provided in the document.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    The document does not describe a multi-reader multi-case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance. Instead, it describes a "cross over study" to directly compare the subject device's performance to the predicate device, focusing on predictive accuracy and target protrusive position.

    6. Standalone Performance:

    Yes, a standalone performance study was done. The "single arm study" and "cross over study" evaluate the device's ability to predict outcomes and recommend mandibular positions, which are standalone functions of the algorithm. The device, an "Auto Titration Device For Oral Appliances," appears to operate without direct human intervention in its primary functions during the study.

    7. Type of Ground Truth Used:

    For the single-arm study, the ground truth was "previously established clinical endpoints used to clear the original MATRx plus device" for determining "responder/non-responder" status in the mild/moderate patient population. For the cross-over study, the ground truth for comparison was the performance of the predicate MATRx plus device (K191925) in terms of predictive accuracy and target protrusive position.

    8. Sample Size for Training Set:

    This information is not provided in the document. The document describes testing the subject device, suggesting the training would have occurred prior to these validation studies.

    9. How Ground Truth for Training Set Was Established:

    This information is not provided in the document.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191925

    Validate with FDA (Live)

    Device Name
    MATRx Plus
    Date Cleared
    2019-10-29

    (103 days)

    Product Code
    Regulation Number
    868.2375
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MATRx plus is indicated for use by a lay person in the home and hospital use under the direction of a Health Care Professional (HCP).

    MATRx plus records the following data: patient respiratory nasal airflow, snoring, blood oxygen saturation, pulse, respiratory effort and body position during sleep.

    MATRx plus uses these recordings to produce a report for the HCP that may aid in the diagnosis and assessment of sleep disordered breathing for adult patients.

    The MATRx plus device may also be used with an automated mandibular positioner that uses feedback control to record changes in the patient's respiratory status related to repositioning of the mandible during an overnight study.

    MATRx plus uses these recordings to produce a report for that can be used to prospectively identify patients with mild to moderate obstructive sleep apnea who may be suitable for therapy with an oral appliance and to recommend a target mandibular position.

    The use of the device does not replace the need for follow-up testing to determine the initial and ongoing effectiveness of the therapy as recommended by clinical practice guidelines.

    Device Description

    MATRx plus is a ventilatory effort recorder and closed loop auto-titration device for oral appliances. MATRx plus is a 5-channel battery-powered respiratory pressure sensor and oximetry system. MATRx plus provides recordings of respiratory pressure, snoring, respiratory effort, pulse rate, oxygen saturation, and body position during sleep.

    The device is worn on the patient's abdomen attached to a reusable effort belt and all relevant respiratory information during sleep is collected via nasal cannula, pulse oximetry module, and respiratory effort sensor. The disposable plastic nasal cannula is connected to the MATRx plus recorder and fixed at the patient's nose. The cannula is dual lumen and functions to individually sample the pressure from each naris. The oximetry sensor is fixed at the patient's finger and connects directly to the device. The Recorder receives input from the sensors and wirelessly transmits the data to a bedside Tablet during the study. The Tablet is set up through a connection to the web portal by the healthcare professional (HCP) prior to deploying the device to the patient for home use. The Tablet runs the MATRx plus application and records and stores the data. The MATRx plus application on the Tablet is also used to guide the user through the set up and conduct of the study through a stepwise user interface. At the end of a recording, the Tablet advises the patient if sufficient data for analysis were recorded during the night.

    During an auto-titration study, the patient wears temporary Titration Trays connected to a mandibular positioner. During the study, the device moves the mandible in response to respiratory events in order to determine if the patient is a suitable candidate for oral appliance therapy.

    After recording, the MATRx plus must be returned to the HCP. The data are automatically uploaded to the secure Portal where they can be accessed and downloaded to the Data Viewer. The Data Viewer can generate a report with the recorded and analyzed data (respiratory pressure, respiratory effort, pulse rate, oxygen saturation) to aid in diagnosis (when used as a ventilatory effort recorder) or to assist the HCP in the clinical management of oral appliance therapy for patients with obstructive sleep apnea.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the MATRx Plus device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA clearance for the MATRx Plus (K191925) is based on demonstrating substantial equivalence to its predicate devices (K181996 and DEN170090), rather than explicit numerical acceptance criteria for clinical performance. The focus is on showing that the minor changes to the new MATRx Plus device do not alter its safety or effectiveness compared to the predicate MATRx Plus devices. Therefore, the "acceptance criteria" are implied by the performance of the predicate device and the demonstration that the subject device meets the same performance standards.

    Feature/Performance AspectAcceptance Criterion (Implied by Predicate)Reported Device Performance (Subject Device)Discussion of Differences/Equivalence
    Intended UseIdentical to predicate deviceIdentical to predicate deviceIdentical
    Patient PopulationAdults (Identical to predicate)Adults (Identical to predicate)Identical
    Environment of UseDeployed from clinics/hospitals, unsupervised in home/clinic, analyzed from physician's office (Identical to predicate)Deployed from clinics/hospitals, unsupervised in home/clinic, analyzed from physician's office (Identical to predicate)Identical
    Outcome1) Data to assist in diagnosis of sleep disordered breathing; 2) Identifying patients with mild to moderate sleep apnea for whom an oral appliance is a suitable therapy (Identical to predicate).1) Data to assist in diagnosis of sleep disordered breathing; 2) Identifying patients with mild to moderate sleep apnea for whom an oral appliance is a suitable therapy (Identical to predicate).Identical
    Oximeter SensorMasimo SET 2040, 1 Hz sampling frequency (Predicate performance)Masimo SET 2040D, 1 Hz sampling frequency (Functionally identical)Functionally identical; oximeters have same technology and performance standards. Substantially equivalent.
    Airflow Signal AlertPredicate: Airflow signal absent for 90s alertSubject Device: Airflow signal absent for 120s alertNo changes to airflow collection or analysis. Change to airflow alert in auto-titration study mode has no effect on data collected or test outcome. Substantially equivalent.
    Mandibular Positioner Actuator MaterialAcetal resin or anodized stainless steel (Predicate)Acetal resin, anodized stainless steel, or anodized aluminum (Subject Device)Identical aside from addition of anodized aluminum actuator. Mandibular positioner function is not affected by change to actuator material. Substantially equivalent.
    Titration Trays Use DurationPredicate: 24 hoursSubject Device: 6 nights (maximum duration allowed by tablet software)The Titration Trays themselves are unchanged from the predicate. Additional biocompatibility and performance testing (in reference device K190051) demonstrated suitability for longer use. This change increases allowable test duration without changing tray design, thus substantially equivalent.
    Titration Trays Fitting PersonnelPredicate: Qualified dentist or dental professionalSubject Device: Trained healthcare professionalThe change to fitting of Titration Trays is supported by reference devices (e.g., Zyppah K182312) which are OTC oral appliances fitted by untrained individuals. Since a non-HCP fitting for an oral appliance has been cleared as safe and effective, the subject device's trays can similarly be fitted by a non-healthcare professional. Substantially equivalent.
    Recorder Dimensions/Weight6.30 x 7.90 x 6.60 cm, 230 g (Predicate)11.0 x 8.0 x 3.5 cm, 166 g (Subject Device)Subject device is smaller and lighter than predicate. Substantially equivalent. (Implies performance maintained)
    Mandibular Positioner Dimensions/Weight5.2 x 5.7 x 2.8 cm, 80 g (Predicate)5.2 x 5.5 x 2.6 cm, 80 g (Subject Device)Subject device is slightly smaller than predicate and the same weight. New design has been tested to ensure it meets the same performance criteria as predicate device. Substantially equivalent.
    Battery Amperage5V - 2A medical grade wall power supply (Predicate)5V - 1A medical grade wall power supply (Subject Device)Decrease in amperage leads to slightly increased charging time (5 h vs 4 h). Device is usable at 85% charge. Deemed negligible. Substantially equivalent.
    Study Setup and ManagementSetup on device software and transferred to Device via internet (Predicate)Setup on device software and transferred to Device via internet. HCP may start new study directly on Tablet. HCP may opt to set up a study containing both study types (basic sleep study preceding or following auto-titration study).HCP may continue to set up study in web Portal and load on Tablet or can alternatively start the study directly on the Tablet. No new functionality introduced (just change of function location). Addition of option for HCP to set up multiple studies for one patient removes the need for the device to be returned to the HCP between studies. Study types remain the same. Addition of risk that is mitigated by Tablet software. Substantially equivalent.
    Reprocessing Instructions(Original instructions)Simplified instructions for reprocessingNo changes to reprocessing requirements. New instructions have been validated for cleaning and low-level disinfection. Substantially equivalent.
    BiocompatibilityAll body contacting components previously clearedAll body contacting components previously cleared. Titration Trays cleared for use up to 29 nights (in K190051) – allowed test duration changed from 3 to 6 nights.Change to auto-titration test duration supported by biocompatibility testing on file for reference device (K190051). Substantially equivalent.
    Other (Respiratory Effort, Position, Snoring, Data Collection, Processor, Indicators, Recording Time, Internal Memory, Battery Cover, Data Display, Data Reporting/Analysis - Indices Generated)Identical to predicateIdentical to predicateIdentical
    Safety TestingTested to IEC 60601-1, IEC 80601-2Tested to IEC 60601-1, IEC 80601-2Testing performed as necessary for subject device to demonstrate updated standards as predicate devices. Substantially equivalent.

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states: "Clinical performance testing was not required for the changes made to the subject device." This means there was no new clinical test set used for this specific 510(k) submission (K191925) to demonstrate the device meets acceptance criteria. The clearance relies on the existing clinical data and performance of the predicate devices (K181996 and DEN170090).

    Therefore, details like sample size, data provenance (country of origin, retrospective/prospective) for a new clinical test set are not provided for this submission. The original predicate devices would have undergone such studies, but that information is not detailed here.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Since no new clinical performance testing was conducted for this submission, there's no mention of experts establishing ground truth for a new test set. For the predicate devices, it's assumed appropriate medical experts (likely sleep specialists/physicians) would have been involved in establishing ground truth for their respective clinical studies.

    4. Adjudication Method for the Test Set

    Not applicable, as no new clinical performance testing was performed.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done

    Not applicable, as no new clinical performance testing was performed, and the device is a diagnostic/therapeutic aid, not an AI interpretation system requiring MRMC for reader improvement.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    The regulatory submission focuses on the safety and performance changes of a medical device hardware and software update, not on a standalone algorithm's performance in a traditional AI sense. The device does perform automated data processing (e.g., auto-scoring of events, calculating indices), which represents an "algorithm only" component. The performance of these algorithmic aspects is implicitly validated through the equivalence to the predicate devices and their prior clearance. For example, the document states: "autoscored 4% desaturation and 4% or 3% apnea-hypopnea events," indicating an algorithm operating on the collected data.

    7. The Type of Ground Truth Used

    For the predicate devices, the ground truth for diagnosing sleep-disordered breathing and assessing oral appliance efficacy would have been established through methods standard in sleep medicine, likely including:

    • Expert Consensus: Interpretation of polysomnography (PSG) data by board-certified sleep physicians.
    • Outcomes Data: Demonstrating clinical benefit (e.g., reduction in AHI, improvement in ODI) with the oral appliance therapy guided by the device.

    For this specific 510(k) (K191925), the ground truth for demonstrating equivalence rests on the technical performance of the modified components meeting the standards and identical functional outcomes of the previously cleared predicate. For example, performance bench testing was completed to demonstrate the mandibular positioner and oximeter meet the same performance criteria as the predicate.

    8. The Sample Size for the Training Set

    Not applicable, as this submission did not involve the development or re-training of a new AI model with a distinct training set. The device is an updated version of existing hardware and software, with the core analytical algorithms presumably carried over from the predicate.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable for the same reasons as #8. For the original development of the algorithms in the predicate device, ground truth would have been established through expert-labeled sleep study data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190051

    Validate with FDA (Live)

    Device Name
    TD Clip
    Date Cleared
    2019-06-07

    (147 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TD Clip is intended to be used with MATRx or MATRx plus Titration Trays to form a mandibular repositioning device intended to reduce or alleviate snoring and mild to moderate obstructive sleep apnea (OSA) in patients 18 years and older. The device is intended to be fitted with assistance from a healthcare professional and used during sleep for a total of less than 30 nights.

    Device Description

    The TD Clip is a temporary oral appliance. The patient uses the temporary oral appliance to alleviate mild to moderate obstructive sleep apnea and/or snoring while waiting for a custom oral appliance to be manufactured. The dentist fabricates temporary titration trays consisting of upper and lower dental trays. The trays are custom fit to the patient using polyvinyl siloxane impression material. The healthcare provider may use the clip assembly to fasten the titration trays at the target therapeutic position to form the TD Clip temporary oral appliance. Patients may use the temporary oral appliance as treatment while waiting for their custom oral appliance. Adjustment of the position requires the setting of a new TD Clip by the healthcare professional.

    AI/ML Overview

    The acceptance criteria and study proving the device meets them are described below for the TD Clip.

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance Criteria (Implied)Reported Device Performance
    Intraoral Material PerformanceNo presence or increased risk of breakage that could lead to unacceptable harm over the maximal study length (30 nights of use).Evaluated under simulated aging, including environmental conditions and representative worst-case loading schemes. The device met the implied acceptance criteria. (Specific quantitative results or detailed findings are not provided in this summary, but the conclusion is that the device is safe and effective).
    TD Clip PerformanceMaintained the set protrusive positioning over 30 nights of use.Evaluated under repeated use and exposed to normal sleep forces over 30 nights of use, including static and cyclic loading under worst-case use. The device met the implied acceptance criteria. (Specific quantitative results or detailed findings are not provided in this summary, but the conclusion is that the device is safe and effective).
    Tray Removal Simulation100% of users were able to remove the Titration Trays from their mouth quickly and without causing injury to themselves or the Titration Trays.Evaluated to ensure users could remove the trays quickly and without injury. The device met the implied acceptance criteria. (Specific quantitative results or detailed findings are not provided in this summary, but the conclusion is that the device is safe and effective).
    Biocompatibility TestingMaterials in contact with mucosal membrane adhere to ISO 10993-1 and associated FDA guidance.All materials in contact with the mucosal membrane were assessed and tested in accordance with ISO 10993-1 and the associated FDA guidance. The device met the implied acceptance criteria.
    Risk ManagementNo unacceptable risks identified.During risk management and performance testing, no new risks were identified. The risk management concluded that the TD Clip had no unacceptable risks.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not specified for any of the performance tests. The document only mentions "mechanical and environmental use conditions" and "repeated use and exposed to normal sleep forces over 30 nights of use".
    • Data Provenance: The tests described are "Bench testing" and "simulated mechanical and environmental use conditions," indicating these were laboratory/engineering tests rather than human clinical trials. Thus, there is no country of origin or retrospective/prospective distinction for data provenance in the traditional sense; the data is from an engineering test setting.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    • Not applicable. The ground truth for these performance tests would be based on engineering specifications and direct measurements against those specifications, rather than expert interpretation of medical data. The evaluation was for mechanical and material performance.

    4. Adjudication Method for the Test Set:

    • Not applicable. Adjudication methods like 2+1 or 3+1 are used for clinical studies with human evaluators to establish ground truth for ambiguous cases. These were bench tests against predetermined engineering criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size:

    • No MRMC comparative effectiveness study was done. Clinical testing was explicitly stated as "not performed as the device has no technological differences from the predicate."

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. The TD Clip is a physical medical device (an intraoral appliance) and not an algorithm or AI software. Therefore, the concept of "standalone performance" for an algorithm doesn't apply here. The performance tests focused on the physical integrity and function of the device itself.

    7. The Type of Ground Truth Used:

    • The ground truth for the performance testing was based on engineering specifications and predefined performance limits (e.g., no breakage, maintenance of protrusive positioning, quick and injury-free removal, adherence to biocompatibility standards). This is a form of objective, empirical measurement and adherence to established material and mechanical standards.

    8. The Sample Size for the Training Set:

    • Not applicable. As a physical medical device undergoing bench testing, there is no "training set" in the context of machine learning or AI.

    9. How the Ground Truth for the Training Set Was Established:

    • Not applicable, as there is no training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K181996

    Validate with FDA (Live)

    Device Name
    MATRx Plus
    Date Cleared
    2018-11-21

    (118 days)

    Product Code
    Regulation Number
    868.2375
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is indicated for use by the Health Care Professional (HCP), where it may aid in the diagnosis and assessment of sleep disordered breathing for adult patients.

    MATRx plus records the following data: patient respiratory nasal airflow. snoring, blood oxygen saturation, pulse. respiratory effort and body position during sleep.

    The device uses these recordings to produce a report for the HCP that may aid in the diagnosis of sleep disordered breathing. The device is intended for home and hospital use under the direction of an HCP.

    Device Description

    The MATRx plus device is a ventilatory effort recorder. The MATRx plus recorder is a 5-channel battery-powered respiratory pressure sensor and oximetry system. MATRx plus provides recordings of respiratory pressure, respiratory effort, pulse rate, oxygen saturation, snoring, and body position during sleep. The physician-prescribed device is worn on the patient's abdomen attached to a reusable effort belt, and all relevant respiratory information during sleep is collected via nasal cannula, pulse oximeter, and respiratory effort sensor. The disposable plastic nasal cannula is connected to the MATRx plus recorder and positioned at the patient's nose. The cannula is dual lumen and functions to individually sample the pressure from each naris. The oximetry sensor is fixed at the patient's finger and connects directly to the device. The Recorder receives input from the sensors and wirelessly transmits the data to a bedside Tablet during the study.

    The Tablet is set up through a connection to the web portal by the Health Care Professional (HCP) prior to deploying the device to the patient for use in the home. The Tablet runs the MATRx plus application. The Tablet records and stores the data. The MATRx plus application on the Tablet is also used to guide the user through the set up and conduct of the study through a stepwise user interface. At the end of a recording the Tablet advises the patient if sufficient data for analysis were recorded during the night.

    After recording, the MATRx plus must be returned to the HCP. The data are automatically uploaded to the secure Portal where they can be accessed and downloaded to the Data Viewer. The Data Viewer can generate a report with the recorded and analyzed data (respiratory pressure, respiratory effort, pulse rate, oxygen saturation) to aid in diagnosis.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the MATRx Plus device, a ventilatory effort recorder. The submission focuses on demonstrating substantial equivalence to a legally marketed predicate device (K163665 MATRx Plus) and includes modifications to the scoring algorithms.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided information:

    Acceptance Criteria and Device Performance

    The document doesn't explicitly list "acceptance criteria" in a table format with specific thresholds. Instead, it demonstrates performance through clinical validation and by showing the new features (3% oxygen desaturation index, 3% and 4% apnea-hypopnea index, preceding peak baseline option, and improved event scoring algorithms) produce "adequate results" and that the device has "comparable sensitivity and accuracy" to a cleared predicate.

    However, we can infer the implicit acceptance criteria from the clinical validation study's goal: to demonstrate that the autoscoring algorithm for the detection of apneas and hypopneas was accurate, and that its sensitivity and accuracy are comparable to previously FDA cleared devices.

    Inferred Acceptance Criteria & Reported Performance:

    Acceptance Criteria (Inferred)Reported Device Performance
    Accuracy of Apnea and Hypopnea Detection: The autoscoring algorithm for detecting apneas and hypopneas should be accurate."The autoscoring algorithm for the detection of apneas and hypopneas was accurate." (This is a qualitative statement of meeting the goal rather than a specific quantitative metric like a percentage, sensitivity, or specificity).
    Comparability to Cleared Predicates (Sensitivity & Accuracy): The device's apnea and hypopnea detection algorithms should have comparable sensitivity and accuracy to previously FDA cleared ventilatory effort recorder devices (e.g., Embletta MPR Sleep Data Recording System K122516)."The results of this testing demonstrate that the MATRx plus subject's apnea and hyponea detection algorithms have comparable sensitivity and accuracy with previously FDA cleared ventilatory effort recorder devices such as the Embletta MPR Sleep Data Recording System (K122516)." (Again, a qualitative statement affirming comparability rather than providing specific comparative metrics).
    Additional Features Performance: The new features (ODI 3% autoscoring, preceding peak baseline, event rounding, reduced erroneous scoring) perform as intended."All features that were changed from the predicate devices are covered by additional verification testing to ensure the intended performance and was found to produce adequate results. Performance bench testing was performed to validate the addition of the ODI3% autoscoring method, the preceding peak baseline, event rounding, and the improvements to reduce erroneous scoring."

    Note: The document provides very high-level summaries of the testing and the results. Specific quantitative metrics (e.g., exact sensitivity/specificity values, or statistical results demonstrating "comparable" performance) are not detailed within this 510(k) summary.


    Study Details:

    1. Sample Size Used for the Test Set and Data Provenance:

      • Sample Size: 27 individuals
      • Data Provenance: Subjects had "confirmed or suspected sleep apnea." The document does not specify the country of origin of the data or whether the study was retrospective or prospective. It is implied to be prospective clinical validation.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

      • The document states that the performance was compared to "the gold standard, polysomnography." It does not specify the number of experts (e.g., sleep specialists, polysomnography technologists) involved in interpreting the polysomnography or establishing the ground truth, nor their specific qualifications. It is generally understood that polysomnography interpretation is performed by qualified sleep physicians or technologists.
    3. Adjudication Method for the Test Set:

      • The document does not describe any specific adjudication method for the ground truth of the test set beyond stating that polysomnography was the "gold standard." There is no mention of 2+1, 3+1, or other multi-reader adjudication processes.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

      • No. The study described is a comparison of the device's autoscoring algorithm against a polysomnography gold standard, not a comparative effectiveness study involving human readers with and without AI assistance. Therefore, there is no effect size reported for human readers' improvement with AI vs. without AI.
    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

      • Yes. The clinical validation directly assesses the performance of the "autoscoring algorithm for the detection of apneas and hypopneas" by comparing its output to polysomnography. This is a standalone assessment of the algorithm's performance.
    6. The Type of Ground Truth Used:

      • Expert Consensus / Physiological Gold Standard: The ground truth used was polysomnography (PSG), which is considered the clinical gold standard for diagnosing sleep disordered breathing. While PSG involves expert interpretation, the document frames PSG itself as the "gold standard" rather than raw pathological findings or purely outcome data.
    7. The Sample Size for the Training Set:

      • The document does not provide any information about the sample size (or any details) of the training set used for the algorithm's development. The focus of this 510(k) summary is on the validation of the modified algorithm.
    8. How the Ground Truth for the Training Set Was Established:

      • This information is not provided in the document. As mentioned, the document focuses on the validation of the device's modified algorithms.
    Ask a Question

    Ask a specific question about this device

    K Number
    DEN170090

    Validate with FDA (Live)

    Device Name
    MATRx plus
    Date Cleared
    2018-08-23

    (245 days)

    Product Code
    Regulation Number
    872.5571
    Type
    Direct
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MATRx plus device may also be used with an automated mandibular positioner that uses feedback control to record changes in the patient's respiratory status related to repositioning of the mandible during an overnight study.

    MATRx plus uses these recordings to produce a report for the HCP that can be used to prospectively identify patients with mild to moderate obstructive sleep apnea who may be suitable for therapy with an oral appliance and to recommend a target mandibular position.

    The use of the device does not replace the need for follow-up testing to determine the initial and ongoing effectiveness of the therapy as recommended by clinical practice guidelines.

    Device Description

    This De Novo request is for an expanded indication (new intended use) for use with an automated mandibular positioner that uses feedback control to record changes in the patient's respiratory status related to repositioning of the mandible during an overnight study.

    The device is an automated temporary oral appliance that manipulates the mandible through multiple protrusion levels during an overnight sleep study in the home. The device uses the same interface as the MATRx plus Home Sleep Apnea Test, also from Zephyr Sleep Technologies, cleared under product code MNR, regulation 21 CFR 868.2375, cleared in K163665. These components include the recorder, tablet, oximeter, effector belt, and nasal cannula. The MATRx plus device for a titration study includes a mandibular positioner to determine an optimal titration position based on inputs of airflow and an oxygen desaturation index (ODI) of less than 10 events per hour. The device analyses the collected information and generates a report to assist the Healthcare Provider in the clinical management of oral appliance therapy for patients with mild to moderate obstructive sleep apnea. From this report, patients may be informed on the future efficaciousness of oral appliance (OA) therapy.

    Specifically, the clinical management may be assisted by this type of oral appliance assessment study by prospectively identifying patients that are expected to achieve therapeutic success with oral appliance therapy. The oral appliance assessment study also includes the provision of a recommended target protrusive setting for the mandible through use of a legally marketed intraoral appliance (regulated under 21 CFR 872.5570, product code LQZ). The target protrusive setting is the amount of mandibular protrusion where the patient is expected to achieve efficacious therapy with the intraoral appliance and is used by the HCP to more effectively complete the titration process.

    The use of the MATRx plus device for an oral appliance assessment study does not replace the clinical titration process or the need for follow-up testing to determine the initial and ongoing therapy as recommended by clinical practice guidelines.

    The MATRx plus test identifies patients as suitable for intraoral therapy if a mandible protrusion level can be found where the patient has an oxygen desaturation index (ODI) of less than 10 events per hour with a 4% desaturation criterion.

    AI/ML Overview

    Here's a structured breakdown of the acceptance criteria and the study that proves the MATRx plus device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Clinical Endpoints

    Acceptance Criteria CategorySpecific Acceptance Criteria (Effectiveness, Safety, Usability)Reported Device Performance (as stated in the document)
    Effectiveness1. Predictive Accuracy (Primary Endpoint): Sensitivity and specificity statistically significant to reject the null hypothesis that sensitivity ≤0.6.Sensitivity: 94.2% (95% CI: 0.86 – 1.00)
    Specificity: 75.0% (95% CI: 0.40 – 1.00)
    2. Target Protrusive Position (Secondary Endpoint): Statistically significant to reject the null hypothesis that target accuracy ≤0.6.Identification of Efficacious Target Protrusion: 86.3% of participants were correctly predicted as responders and successfully provided with an effective target protrusion.
    SafetyNo significant safety concerns reported by coordinator, dentist, or participant.No significant safety concerns reported. Adverse events were temporary discomfort (in teeth, joint, jaw, minor gum irritations, temporary bite changes, soreness in teeth) that resolved within a few hours.
    Only minor discomfort reported on post-study questionnaires.Temporary discomfort reported (resolved within hours).
    No software failure resulting in injury.No software failure reported that resulted or could result in injury.
    No movement beyond full retrusion and maximum protrusion set values.No movement beyond full retrusion and maximum protrusion set values reported.
    UsabilityAccurate prediction for OA therapy and effective titration level demonstrated by actual use testing (setup, placement, use, removal by participant in home).Human factors assessment conducted as part of the clinical trial showed participants and providers evaluated for comprehension-based tasks. The study supported usability.

    Bench Performance Testing

    Acceptance Criteria CategorySpecific Acceptance CriteriaReported Device Performance (as stated in the document)
    Mandibular Positioner, Accuracy Data(i) Device will be able to send a position feedback signal to the controller that reports a new position within 0.5mm of the actual physical position of the trays/mandible; and the accuracy will be maintained under expected clinical loads. (ii) The mandibular positioner will be able to physically endure the use life of the device under normal operating load without failure and maintain the accuracy. Simulated use life testing was conducted based on days of single patient use, movements per patient use, and approximated number of insertion and removal cycles per patient use.The data provided showed each test was passed according to its pre-specified acceptance criteria.
    Mandibular Positioner, Force Limit TestMandibular positioner will be incapable of applying a force that could induce serious injury under a worst-case fault scenario of 70 Newtons.The data provided showed each test was passed according to its pre-specified acceptance criteria.
    Device Force Limitation Simulation(i) Demonstrate the ability of users to voluntarily stall the mandibular positioner in a fault position (i.e., full stroke speed). (ii) Demonstrate that without voluntary action to resist the force, the trays dislodge from their seated position on the teeth prior to injury or maximum force (i.e., 70N).The data provided showed each test was passed according to its pre-specified acceptance criteria.
    Impression Material BreakageDemonstrate the ability of the impression material used to form the temporary oral appliance to maintain integrity over the use life of the temporary oral appliance.The data provided showed each test was passed according to its pre-specified acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set:
      • Clinical Study: The document provides "49" for "Successful Identification of Responders" and "2" for "Failure to Identify Responders" in the "Predictive Success Category," and "3" for "Successful Identification of Responders" and "6" for "Failure to Identify Responders" in the "Predictive Failure Category." This totals 49 + 2 + 3 + 6 = 60 participants in the mild to moderate OSA population for the effectiveness endpoints.
      • Human Factors: 15 patients were recruited for each study.
    • Data Provenance: The MATRx plus is by Zephyr Sleep Technology, located in Calgary, Alberta, Canada. The clinical study was conducted in a clinical setting (dental professional's office for deployment, home environment for use). The study is prospective, as it involved recruiting participants, conducting the MATRx plus study, and then following up with custom oral appliance therapy.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts:
      • Effectiveness Ground Truth (Oral Appliance Outcome): While not explicitly stated how many individual experts reviewed each case, the process involved a "Sleep Health professional" reviewing baseline HSAT data and a "licensed dentist" conducting intraoral exams and subjective interviews post-study. The final determination of "successful treatment with an oral appliance" (ODI < 10 with custom appliance) and "unsuccessful treatment" (ODI ≥ 10) was based on "automated analysis of a full night home sleep test," which implies an objective measurement, potentially overseen or interpreted by these professionals. For the "Identification of Efficacious Target Protrusion," if the initial target wasn't successful (ODI < 10), the oral appliance was advanced and retested, indicating involvement of a clinical expert.
      • Human Factors: Not explicitly stated, but the process involved "Participants and providers were evaluated for comprehension-based tasks."
    • Qualifications of Experts:
      • Sleep Health Professional: Reviewed baseline HSAT data. (Specific qualifications like "radiologist with 10 years of experience" are not provided, only the role title).
      • Licensed Dentist: Evaluated for adequate dentition, fitted titration trays, measured protrusion values, conducted intraoral exams, and performed subjective interviews post-study. (Specific qualifications like years of experience are not provided).

    4. Adjudication Method for the Test Set

    The document does not explicitly describe a formal "adjudication method" (like 2+1 or 3+1 consensus) for establishing the ground truth of the test set results. The primary effectiveness endpoint, successful/unsuccessful treatment with an oral appliance, was defined retrospectively by an "automated analysis of a full night home sleep test" using the ODI < 10 criterion. This suggests an objective, rule-based approach rather than a multi-expert consensus for the final outcome. However, "All baseline data were reviewed by a Sleep Health professional," and the follow-up for target protrusion involved clinical discretion ("adjusted accordingly based on clinical expertise of the dentist"), indicating expert involvement in data interpretation and treatment adjustments.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study evaluated the standalone performance of the MATRx plus device in predicting OA therapy success and recommending a target mandibular position. It did not compare human readers' performance with and without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, a standalone study assessing the algorithm's performance was conducted. The "Predictive Accuracy" effectiveness endpoint explicitly states: "The MATRx plus predictive accuracy endpoint was assigned if the sensitivity and specificity was found to be statistically significant..." and "The predication was done using a trained binary classifier." This refers to the algorithm's ability to identify responders and recommend a target protrusion on its own, based on the collected physiological data. While human factors and clinical oversight were part of the overall process, the core predictive power evaluated was that of the device's algorithm.

    7. Type of Ground Truth Used

    The primary ground truth for the effectiveness endpoint ("Successful treatment with an oral appliance") was outcomes data: achieving an Oxygen Desaturation Index (ODI) of less than 10 events per hour with a custom oral appliance, determined by automated analysis of a full night home sleep test.
    For the target protrusion accuracy, the ground truth was the clinically determined effective protrusion (the position where ODI < 10 was achieved with the custom OA, potentially after clinician-guided titration).

    8. Sample Size for the Training Set

    The document does not provide the sample size for the training set. It only describes the pivotal clinical trial (Phase II) which appears to be the test set for evaluating the device's performance. It mentions "The predication was done using a trained binary classifier," implying a training phase occurred, but details about the data used for training are absent from this text.

    9. How the Ground Truth for the Training Set Was Established

    The document does not describe how the ground truth for the training set was established, as it does not detail the training process or the data used for it. It only mentions that the prediction was made using a "trained binary classifier," implying ground truth data was used for its training.

    Ask a Question

    Ask a specific question about this device

    K Number
    K163665

    Validate with FDA (Live)

    Device Name
    MATRx plus
    Date Cleared
    2017-05-05

    (129 days)

    Product Code
    Regulation Number
    868.2375
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is indicated for use by the Health Care Professional (HCP), where it may aid in the diagnosis and assessment of sleep disordered breathing for adult patients.

    MATRx plus records the following data: patient respiratory nasal airflow, snoring, blood oxygen saturation, pulse, respiratory effort and body position during sleep.

    The device uses these recordings to produce a report for the HCP that may aid in the diagnosis of sleep disordered breathing. The device is intended for home and hospital use under the direction of a HCP.

    Device Description

    The MATRx plus is a new ventilatory effort recorder device. The MATRx plus recorder is a 5-channel battery-powered respiratory pressure sensor and oximetry system. MATRx plus provides recordings of respiratory pressure, respiratory effort, pulse rate, oxygen saturation, snoring and body position during sleep. The physician prescribed device is worn on the patient's abdomen attached to a reusable effort belt and all relevant respiratory information during sleep is collected via nasal cannula, pulse oximetry module and respiratory effort sensor. The disposable plastic nasal cannula is connected to the MATRx plus recorder and fixed at the patient's nose. The cannula is dual lumen and functions to individually sample the pressure from each naris. The oximetry sensor is fixed at the patient's finger and connects directly to the device. The Recorder receives input from the sensors and wirelessly transmits the data to a bedside Tablet during the study.

    The Tablet is set up through a connection to the web portal by the Health Care Professional (HCP) prior to deploying the device to the patient for use in the home. The Tablet runs the MATRx plus application according to the selected study type, records and stores the data. The MATRx plus application on the Tablet is also used to guide the user through the set up and conduct of the study through a stepwise user interface. At the end of a recording the Tablet advises the patient if sufficient data for analysis were recorded during the night.

    After recording, the MATRx plus must be returned to the HCP. The data are automatically uploaded to the secure Portal where they can be accessed and downloaded to the Data Viewer. The Data Viewer can generate a report with the recorded and analyzed data (respiratory pressure, respiratory effort, pulse rate, oxygen saturation) to aid in diagnosis.

    AI/ML Overview

    The Zephyr Sleep Technologies MATRx plus device, a ventilatory effort recorder for aiding in the diagnosis and assessment of sleep-disordered breathing in adult patients, underwent bench testing and performance comparisons to demonstrate substantial equivalence to its predicate device, the ApneaLink Air (K143272).

    Here's an analysis of the provided information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative "acceptance criteria" for clinical performance metrics (like sensitivity, specificity for diagnosing sleep-disordered breathing). Instead, it focuses on demonstrating technological equivalence to the predicate device and satisfactory performance in bench testing.

    Therefore, the table below reflects the comparison of technological characteristics between the MATRx plus and its predicate device, ApneaLink Air, as presented in the submission. The acceptance criterion is implicit: the MATRx plus should be "substantially equivalent" in its performance and features to the predicate device.

    Feature / CriterionAcceptance Criteria (Implicit: Substantially Equivalent to ApneaLink Air K143272)Reported Device Performance (MATRx plus)Comments on Equivalence
    Intended UseAid in diagnosis of sleep disordered breathing for adult patients. Records nasal airflow, snoring, SpO2, pulse, resp effort, body position.Aid in diagnosis and assessment of sleep disordered breathing for adult patients. Records nasal airflow, snoring, SpO2, pulse, resp effort, body position.Substantially Equivalent (MATRx plus includes "assessment" in addition to "diagnosis").
    Environment of UseDeployed from Clinics, Hospitals; Used unsupervised in Home or Clinic; Analyzed from physician's office.Deployed from Clinics, Hospitals; Used unsupervised in Home or Clinic; Analyzed from physician's office.Equivalent.
    OutcomeData to aid in the diagnosis of sleep disordered breathing.Data to aid in the diagnosis of sleep disordered breathing.Equivalent.
    ContraindicationsNot for use near MRI.Not for use near MRI; Not an apnea monitor or for life-supporting situations; Not for use by persons under 18.Substantially Equivalent (MATRx plus has additional safety contraindications).
    Oximeter SensorThird-party sensor (Nonin Xpod 3012 LP), 1 Hz sampling.Third-party sensor (Masimo SET 2040), 1 Hz sampling.Substantially Equivalent.
    Airflow SensorSingle channel nasal cannula, 100 Hz sampling for pressure, airflow, snoring.Dual channel nasal cannula, 350 Hz sampling for pressure, airflow, snoring.Substantially Equivalent - Dual channel can reduce artifact.
    Respiratory Effort SensorPneumatic principle, 10 Hz sampling.Inductance principle (third-party belt Sleep Sense, K042253), 25 Hz sampling.Substantially Equivalent - Both are commonly used and sensitive enough.
    Position Sensor3D axis accelerometer.3D axis accelerometer.Substantially Equivalent.
    Snoring DetectionNasal airflow fluctuation envelope (10-60 Hz), user adjustable threshold.Nasal airflow fluctuation envelope (10-70 Hz), set threshold.Substantially Equivalent.
    Recorder DimensionsApprox. 2.4" x 4" x 1.2" (predicate).Approx. 2.48" x 3.11" x 0.83" (MATRx plus).Substantially Equivalent - Compared to reference device (K021176) at 2.5"x4.5"x1.2".
    Recorder Weight66 g (predicate).230 g (MATRx plus).Substantially Equivalent - Compared to reference device (K021176) at 234g.
    Data CollectionSensors direct to recorder, stored on recorder, uploaded via USB post-study.Sensors direct to recorder, wirelessly transferred (Bluetooth) to tablet, stored on tablet, wirelessly uploaded to web portal post-study.Substantially Equivalent - Bluetooth connectivity comparable to Nox T3 (K082113). Allows earlier data review.
    ProcessorARM Cortex M3 based LPC1853.ARM Cortex M4 based, STM32 F4, 168 MHz; Tablet minimum requirements: 1.33 GHz, 1 GB RAM.Substantially Equivalent - MATRx plus includes a tablet for data storage and user interaction.
    IndicatorsVia LED: Test Complete, Respiratory Flow.Via Tablet UI: Test Started, Test Paused, Test Complete, Respiratory Flow, Oximetry Connection, Effort Sensor Connection, Battery Charge.Substantially Equivalent - Additional feedback provided by MATRx plus (via tablet).
    Recording Time4 x 12 hours.6 x 8 hours.The total number of study hours in the MATRx plus is set by the software.
    Internal Memory48 hours.48 hours (reported); > 100 hours (comment).Substantially Equivalent.
    Patient Set-upOn device software, transferred via USB.On device software, transferred via internet.Substantially Equivalent - Removal of USB port to increase cybersecurity.
    Data Storage and AccessStored and accessible via local database. Copies downloaded locally.Stored and accessible via manufacturer's secure internet-accessible server. Copies downloaded locally.Substantially Equivalent.
    Data Analysis (Indices)AHI, RI, Apnea Index (various), Hypopnea Index, flow limited breaths, Cheyne-Stokes, ODI, various saturations, various pulse rates.ODI, Average Saturation, minimum saturation, maximum saturation, min pulse rate, max pulse rate, average pulse rate. (Does NOT calculate AHI, RI, Apnea Index, Hypopnea Index, flow limited breaths, Cheyne-Stokes).Substantially Equivalent - Raw data available for manual calculation of uncalculated parameters.
    BiocompatibilityCleared under K131932.All body contacting components previously cleared (Nasal Cannula K151506; Oximeter sensor K051212, K090662, K101896).Substantially Equivalent.
    Safety TestingIEC 60601-1-11: 2010; IEC 60601-1-2:2007; IEC 60601-1:2005.(Implied equivalent; states "Design and system verification testing included... mechanical and environmental testing according to IEC 60601-1-11:2010, electrical safety testing according to IEC 60601-1:2005+1 and EMC testing according to IEC 60601-1-2:2007 has been performed.") Rechargeable battery tested to IEC 62133:2012.Equivalent.

    2. Sample size used for the test set and the data provenance

    The document mentions validation of the "Autoscoring algorithm for ODI" (Oxygen Desaturation Index) by comparing its output with the Snore SAT monitor (K002159).

    • Test Set Sample Size: 179 patients.
    • Data Provenance: Not explicitly stated (e.g., country of origin). It's mentioned as "data from 179 patients" when comparing the autoscoring algorithm. The study appears to be retrospective, using existing data for comparison.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document states: "Autoscoring algorithm for ODI was compared with the Snore SAT monitor output (K002159) on data from 179 patients."

    • This implies that the "ground truth" for the ODI autoscoring comparison was the output of the Snore SAT monitor (K002159).
    • The document does not specify the number of human experts, their qualifications, or their role in establishing this ground truth for the test set. It relies on the predicate device's (or reference device's) established output as the comparative standard.

    4. Adjudication method for the test set

    Not applicable. The ground truth for the ODI autoscoring was based on the output of a previously cleared device (Snore SAT monitor), not on human consensus or adjudication in this specific context.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No MRMC comparative effectiveness study is mentioned for the MATRx plus. The submission focuses on demonstrating substantial equivalence to a predicate device through technological comparison and bench testing, not on assessing human reader improvement with AI assistance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Yes, a standalone performance assessment was done for the Autoscoring algorithm for ODI.

    • "Autoscoring algorithm for ODI was compared with the Snore SAT monitor output (K002159) on data from 179 patients."
    • This suggests the MATRx plus's algorithm-driven ODI calculation was directly compared to the output of a reference device without human interpretation as part of this specific comparison.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the ODI autoscoring algorithm validation, the ground truth was the output of a legally marketed predicate/reference device, the Snore SAT monitor (K002159). This is a form of "reference device data" ground truth, not pathology, expert consensus, or outcomes data in this specific study.

    8. The sample size for the training set

    The document does not provide information on the sample size used for training any algorithms within the MATRx plus device. The focus of this 510(k) summary is on demonstrating equivalence, not on detailing algorithm development and training.

    9. How the ground truth for the training set was established

    Not provided in the document. As no information on the training set is included, the method for establishing its ground truth is also absent.

    Ask a Question

    Ask a specific question about this device

    K Number
    K103704

    Validate with FDA (Live)

    Date Cleared
    2012-03-23

    (459 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Overnight use of a temporary oral appliance to determine in which patients with obstructive sleep apnea mandibular advancement by an oral appliance will reduce obstruction of airflow and thereby to identify patients suitable for oral appliance therapy. The device is also used to recommend a target mandibular position that will reduce obstruction of airflow in patients determined to be suitable for oral appliance therapy.

    Device Description

    The RCMP appliance is a device used in the clinical setting to identify patients with obstructive sleep apnea that will be successfully treated with an oral appliance. A temporary appliance fits to the patient's teeth and incrementally and reversibly advances the patient's mandible anteriorly with respect to the maxilla while the patient is under full polysomnographic monitoring. The repositioning of the mandible pulls the tongue forward and increases the patient's airway space. Such obstruction can be a causative factor in obstructive sleep apnea and snoring. The RCMP is used in the clinical setting by a technician to advance the mandible until the polysomnographic data indicates removal of the obstruction of airflow in the airway. The device is used to recommend a target mandibular position that will reduce obstruction of airflow in patients determined to be suitable for oral appliance therapy.

    The RCMP appliance consists of upper and lower disposable dental titration trays that are used to hold onto the teeth by means of a quick-set retention material. The trays attach to a mandibular positioner that retracts and extends a rod to adjust the tray positioning by means of a small force linear actuator. By extending the linear actuator rod attached to the upper tray, the RCMP pushes on the upper teeth and pulls on the lower teeth, thus displacing the jaw. RCMP software interacts with the polysomnography to allow the technician to control fine adjustments of the relative position of the mandible.

    AI/ML Overview

    Here's an analysis of the provided text to extract information about the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly state quantitative acceptance criteria for the device beyond demonstrating that it is "at least as safe and effective as the predicate device at selecting candidates for oral appliance therapy." It highlights performance characteristics and differences from the predicate without setting specific numerical targets.

    However, based on the qualitative descriptions and the overall goal of the study, we can infer some implied performance objectives:

    Acceptance Criterion (Implied)Reported Device Performance
    Safety: Device operation does not cause patient harm.The validation study provides evidence that the RCMP test, when implemented in a polysomnographic setting by a trained sleep technician, is safe.
    Patient Comfort/Procedure Tolerability: Procedure is comfortable.The dental titration procedure and approach was shown to be comfortable for the patient.
    Technician/Expert User-Friendliness: Easy for users to operate.The dental titration procedure and approach was shown to be user friendly to both the technician and the sleep expert.
    Non-Disruptive during Sleep Study: Device does not wake patient.The study included a detailed arousal analysis that demonstrated that the RCMP sleep titration does not wake the patient and therefore the titration can be completed during REM sleep.
    Effectiveness in Candidate Selection: Accurately identifies suitable patients for oral appliance therapy.The study demonstrated that using prospectively determined criteria, the RCMP test can accurately select favorable candidates for oral appliance therapy. The study also demonstrated that the RCMP test can recommend a therapeutically effective target protrusive distance for the therapy. The clinical study demonstrated that the RCMP is at least as safe and effective as the predicate device at selecting candidates for oral appliance therapy.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: 66 patients.
    • Data Provenance: The study was a prospective clinical study. Patients were recruited from a sleep center and dentist.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and their Qualifications

    The document mentions that:

    • The RCMP is used "by a trained sleep technician".
    • "The final assessment and recommendation for treatment by an appliance is completed by a physician who reviews the positioning data and associated cardio-respiratory reaction."
    • The procedure was "user friendly to both the technician and the sleep expert."

    While it confirms the involvement of a trained sleep technician for device operation and a physician (sleep expert) for final assessment and recommendation, it does not specify the exact number of physicians/sleep experts used to establish the ground truth (i.e., assess the suitability for oral appliance therapy and optimal mandibular position for each patient in the test set). It implies at least one physician. The qualifications are broadly described as "physician" and "sleep expert."

    4. Adjudication Method for the Test Set

    The document does not explicitly describe an adjudication method like 2+1 or 3+1 for establishing ground truth from multiple experts. It states that "the final assessment and recommendation for treatment by an appliance is completed by a physician who reviews the positioning data and associated cardio-respiratory reaction." This suggests a single physician's review is part of the process for determining the "therapeutically effective target protrusive distance" and identifying suitable patients, rather than a multi-expert adjudication of the ground truth itself based on labels provided by independent readers.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its Effect Size

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study comparing human readers with and without AI assistance was not done. The study focused on demonstrating the safety and effectiveness of the RCMP device itself, particularly its ability to identify suitable candidates and recommend an effective mandibular position, not on improving human reader performance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was done

    This device (RCMP) is not an AI algorithm but a physical, remotely controlled mechanical device. Its operation is explicitly stated to involve a "trained technician" and "physician" in a clinical setting. Therefore, a standalone (algorithm only) performance assessment is not applicable and was not performed. The device requires human interaction for its use and interpretation of its outputs.

    7. The Type of Ground Truth Used

    The ground truth used was based on polysomnographic data reviewed by a physician/sleep expert. Specifically, the device is used "in the clinical setting under full polysomnographic cardio-respiratory monitoring" to identify patients and then a "physician... reviews the positioning data and associated cardio-respiratory reaction" to make the final assessment and recommendation for treatment. This combines objective physiological measurements (polysomnography) with expert clinical judgment.

    8. The Sample Size for the Training Set

    The document does not mention a training set or its sample size. The clinical study described is a 66-patient validation study. Given that the RCMP is a device with mechanical and software components rather than a machine learning model, a distinct "training set" in the context of AI/ML is not relevant here. The device's design, software, and operational parameters would have been developed and tested through bench and engineering processes, not via a machine learning training dataset.

    9. How the Ground Truth for the Training Set was Established

    As no training set (in the ML sense) is mentioned, this question is not applicable. The device's functionality and efficacy were validated through the clinical study and bench testing, as opposed to being "trained" on a dataset.