Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K232625

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-06-07

    (283 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Steriking® Pouch for Robotic Instruments serve as an enclosure for Intuitive Endowrist Robotic Instruments during steam sterilization that maintains the sterility of the enclosed medical devices until they are used.

    Pouches are intended for use as packaging material of Intuitive Endowrist Robotic Instruments for steam sterilization in health care establishments. The products are for single use only. The pouches allow sterilization, mamtain sterility and enable aseptic presentation of packed medical device.

    The recommended sterilization cycles are as follows: Pre-vacuum steam at 132ºC for 4 minutes; Drying time of 20 minutes Pre-vacuum steam at 135℃ for 3 minutes: Drying time of 16 minutes

    The maximum load for the Steriking® Pouch for Robotic Intuitive Endowrist Robotic Instrument or other medical devices with a combined weight of metal and plastics of 2.6 pounds or less.

    Device Description

    Steriking® Pouch for Robotic Instruments consist of a paper backing (Bleached wood pulp, grammage 100 g/m2) with transparent plastic film laminate front (2 sheets of laminated plastic) with a total grammage of 55 g/m2, 1 sheet of oriented polyester 12 microns thick, 1 sheet of coextruded polypropylene 40 microns thick. The plastic laminate is triple heat sealed to the backing paper.

    Steriking® Pouch for Robotic Instruments consists of pouch sizes that can be heat sealed, and pouch sizes that are selfsealed.

    For heat-sealed sizes: The open end of the pouch is to be heat sealed once a device is inserted. Heat sealing parameters to provide a sterile barrier are 165°C-200°C (329°F - 392°F).

    Steriking® Pouch for Robotic Instruments have Dimensional heat-seal configurations (2 sizes 200mm x 800mm, 250mm x 900mm).

    Steriking® Pouch for Robotic Instruments with heat-seal maintains the sterility of the enclosed devices for up to 1 year post Steam sterilization and before sterilization has a maximum shelf life of 5 years from the date of manufacture.

    For self-sealed sizes: The open end of the pouch is to be self sealed once a device is inserted. Self-sealable pouches are featured with adhesive strip allowing tight, impermeable closing flap is pre-folded to facilitate the closure. When closing the self-seal pouch the paper flap shall be folded line. Flap should be pressed firmly against the laminate from the center working outwards to ensure a good, even seal.

    Steriking® Pouch for Robotic Instruments have Dimensions (2 sizes 200mm x 800mm x 800mm x 875mm).

    Steriking® Pouch for Robotic Instruments with self-seal maintains the sterility of the enclosed devices for up to 1 year post Steam sterilization and before sterilization has a maximum shelf life of 3 years from the date of manufacture.

    AI/ML Overview

    The provided text describes the acceptance criteria and the results of performance tests for the "Steriking® Pouch for Robotic Instruments". However, it does not include information about any AI-powered function, a "study that proves the device meets the acceptance criteria" in the context of an algorithm's performance, or details regarding ground truth establishment, expert involvement, or sample sizes related to AI/algorithm development.

    The document is a 510(k) Premarket Notification from the FDA for a medical device (sterilization pouch), which is a physical product, not a software or AI device. Therefore, the questions about AI, algorithms, ground truth, expert consensus, and training/test set sizes are not applicable to this document.

    Here's a summary of the available information regarding acceptance criteria and performance, as it relates to a physical medical device.

    1. Table of acceptance criteria and the reported device performance:

    ElementStandard ReferenceAcceptance CriteriaResults
    Sterilant PenetrationAAMI TIR 12:2020Demonstrates that a minimum of 1.0x10^6 Geobacilus stearothermophilus spores were killed in a half-cycle (6-log reduction)Pass
    Drying TimeISO 17665-1: 2006/2016Test samples do not exceed a 3% weight gainPass
    Material CompatibilityEN 868-5: 2009 4.2.2.1, Annex Bno objectionsPass
    EN 868-5: 2009 4.2.2.1, Annex Cno pinholesNo result listed, but "Pass" can be inferred from conclusion.
    ASTM D 882:2012, procedure A>= 20 N/15 mmNo result listed, but "Pass" can be inferred from conclusion.
    ISO 1924-3> 66 N/15 mmNo result listed, but "Pass" can be inferred from conclusion.
    > 33 N/15 mmNo result listed, but "Pass" can be inferred from conclusion.
    Internal Method152 μm +/- 10%No result listed, but "Pass" can be inferred from conclusion.
    ISO 1974> 550 NNo result listed, but "Pass" can be inferred from conclusion.
    DIN 53363> 10 NNo result listed, but "Pass" can be inferred from conclusion.
    ISO 2758> 230 kPaNo result listed, but "Pass" can be inferred from conclusion.
    ISO 5636-3>=3.9 - 5.7 μm/Pa•sNo result listed, but "Pass" can be inferred from conclusion.
    Puncture ResistanceISO 14477Internal protocol, Accepted by SponsorPass
    Package IntegrityASTM F1929 - Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye PenetrationNo leaks detected after dye migrationPass
    Microbial Barrier PropertiesDIN EN ISO -11607-1:2017-10 Section 5.1.6. a) microbial barrier; DIN 953-6, section 4.8.6.Evaluated as "sufficiently germ-proof"Pass
    Shelf Life Prior to SterilizationASTM F1929-15No leaks detected after dye migrationNo result listed, but "Pass" can be inferred from conclusion.
    EN 868-5 Packaging for terminally sterilized medical devices — Part 5>= 1.5 N/15 mmPass
    ISO 5636-3 Paper and board — Determination of air permeance>=3.4 μm/Pa •sNo result listed, but "Pass" can be inferred from conclusion.
    ASTM F1980 - 16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical DevicesExecuted as per standard, No Acceptance CriteriaNo result listed, likely because it's a guide, not a pass/fail test.
    Shelf Life Post sterilizationASTM F1929-15No leaks detected after dye migrationPass
    ASTM F88/F88M-15: Standard Test Method for Seal Strength of Flexible Barrier Materials>= 1.5 N/15 mmNo result listed, but "Pass" can be inferred from conclusion.
    BiocompatibilityISO 10993-5, Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicityNot greater than Grade 2 reactivity (mildly reactive)Pass
    Evidence that printing ink on paper material does not leach into the interior of the pouchDIN EN 6465 Evaluation gradesPass

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
    The document does not specify the sample sizes used for each test. The data provenance is not explicitly stated, but the submission is from Wipak Oy, Nastola, Finland, suggesting the testing was conducted or overseen by this company. The tests are for device performance, not a dataset in the context of an algorithm.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
    Not applicable. This is a physical medical device; the performance tests are based on established international and national standards for sterilization packaging, not on expert interpretation or ground truth derived from clinical data in the way an AI algorithm would be.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    Not applicable. This is for a physical medical device, not an algorithm where adjudication of results would be relevant.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This is a physical medical device, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
    Not applicable. The "ground truth" for this device's performance is objective measurement against established industry standards and regulations for sterilization, package integrity, microbial barriers, and biocompatibility.

    8. The sample size for the training set:
    Not applicable. This is a physical medical device, not an AI algorithm requiring a training set.

    9. How the ground truth for the training set was established:
    Not applicable. This is a physical medical device, not an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231996

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-09-07

    (64 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Steriking® LT-Blueline Pouches with Tyvek® are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERIS® V-Pro® Sterilizer Systems.

    The device is intended to allow sterilization of enclosed devices and to maintain sterility of the enclosed devices until used up to 1 year post sterilization. The products are for single use only.

    Steriking® LT-Blueline Pouches with Tyvek are intended for use in the following STERIS® V-Pro® Sterilization Cycles: Lumen Cycle Non Lumen Cycle Flexible Cycle

    Device lumen dimensions:

    Flexible Cycle: 1 lumen max and 1 mm min ID x 1050 mm max length for all pouch sizes.

    Lumen Cycle: 1 lumen max for all pouch sizes.

    1 mm min ID for pouch sizes: 130 x 380 mm, 130 x 270 mm, 90 x 250 mm, 90 x 200 mm 2 mm min ID for pouch sizes: 300 x 450 mm, 250 x 400 mm, 190 x 330 mm 125 mm max length for pouch sizes: 130 x 380 mm, 130 x 250 mm, 90 x 250 mm, 90 x 200 mm 250 mm max length for pouch sizes: 300 x 450 mm, 250 x 400 mm, 190 x 330 mm

    Max Weights: 1.234 pounds for pouch sizes below (Flexible, Lumen, and Non-Lumen Cycles) 300 x 450 mm 250 x 400 mm 190 x 330 mm

    .122 pounds for pouch sizes below (Flexible and Lumen Cycles) 130 x 380 mm 130 x 270 mm 90 x 250 mm 90 x 200 mm

    .204 pounds for pouch sizes below (Non-Lumen Cycle) 130 x 380 mm 130 x 270 mm 90 x 250 mm 90 x 200 mm

    Pouch Sizes: 300 x 450 mm 250 x 400 mm

    Device Description

    Steriking® LT-Blueline Pouches with Tyvek® are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERIS® V-Pro® Sterilizer Systems.

    The device is intended to allow sterilization of enclosed devices and to maintain sterility of the enclosed devices until used up to 1 year post sterilization. The products are for single use only.

    Steriking® LT-Blueline Pouches with Tyvek are intended for use in the following STERIS® V-Pro® Sterilization Cycles: Lumen Cycle Non Lumen Cycle Flexible Cycle

    Device lumen dimensions:

    Flexible Cycle: 1 lumen max and 1 mm min ID x 1050 mm max length for all pouch sizes.

    Lumen Cycle: 1 lumen max for all pouch sizes.

    1 mm min ID for pouch sizes: 130 x 380 mm, 130 x 270 mm, 90 x 250 mm, 90 x 200 mm 2 mm min ID for pouch sizes: 300 x 450 mm, 250 x 400 mm, 190 x 330 mm 125 mm max length for pouch sizes: 130 x 380 mm, 130 x 250 mm, 90 x 250 mm, 90 x 200 mm 250 mm max length for pouch sizes: 300 x 450 mm, 250 x 400 mm, 190 x 330 mm

    Max Weights: 1.234 pounds for pouch sizes below (Flexible, Lumen, and Non-Lumen Cycles) 300 x 450 mm 250 x 400 mm 190 x 330 mm

    .122 pounds for pouch sizes below (Flexible and Lumen Cycles) 130 x 380 mm 130 x 270 mm 90 x 250 mm 90 x 200 mm

    .204 pounds for pouch sizes below (Non-Lumen Cycle) 130 x 380 mm 130 x 270 mm 90 x 250 mm 90 x 200 mm

    Pouch Sizes: 300 x 450 mm 250 x 400 mm

    AI/ML Overview

    This document is an FDA 510(k) clearance letter for a medical device: "Steriking® LT-Blueline Pouches with Tyvek®". It confirms that the device is substantially equivalent to legally marketed predicate devices.

    However, the provided text does not contain any information about acceptance criteria or a study proving that an AI/ML device meets those criteria. The document is a regulatory approval for a physical medical device (sterilization pouches), not an AI/ML diagnostic or therapeutic device.

    Therefore, I cannot answer your specific questions regarding acceptance criteria and performance studies for an AI/ML device based on the input provided. The requested information (e.g., sample size for test set, number of experts for ground truth, MRMC studies, standalone performance, training set details) is relevant for AI/ML device submissions, but this document pertains to a physical sterilization product.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231999

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-09-07

    (64 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Steriking® LT-Blueline Pouches with Tyvek® are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERRAD® Sterilizer Systems. The device is intended to allow sterilization of enclosed devices and to maintain sterility of the enclosed devices until used up to 1 years post sterilization. The products are for single use only. The Steriking® LT-Blueline Pouches with Tyvek are intended for use in the following STERRAD® Sterilization Cycles:
    NX Standard
    NX Advanced
    100NX Duo
    100NX Flex
    100NX Standard

    Max weights:
    NX Standard Cycle:
    For pouch sizes 90 x 200 mm, 90 x 250 mm, 130 x 270 mm, 130 x 380 mm, .094 lbs. For pouch sizes 190 x 330 mm, 250 x 400 mm, 300 x 450 mm, 1.194 pounds.

    NX Advanced Cycle:
    For pouch sizes 90 x 200 mm, 90 x 250 mm, 130 x 270 mm, 130 x 380 mm, .096 lbs. For pouch sizes 190 x 330 mm, 250 x 400 mm, 300 x 450 mm, 1.224 pounds.

    100NX Duo Cycle:
    For pouch sizes 90 x 200 mm, 90 x 250 mm, 130 x 270 mm, 130 x 380 mm, .116 lbs. For pouch sizes 190 x 330 mm, 250 x 400 mm, 300 x 450 mm, 1.222 pounds.

    100NX Flex Cycle:
    For pouch sizes 90 x 200 mm, 90 x 250 mm, 130 x 270 mm, 130 x 380 mm, .202 lbs. For pouch sizes 190 x 330 mm, 250 x 400 mm, 300 x 450 mm, 1.220 pounds.

    100NX Standard Cycle:
    For pouch sizes 90 x 200 mm, 90 x 250 mm, 130 x 270 mm, 130 x 380 mm, .202 lbs. For pouch sizes 190 x 330 mm, 250 x 400 mm, 300 x 450 mm, 4.478 pounds.

    Max Count of lumens:
    1 lumen max for All cycles

    Max ID of lumens:
    NX Standard Cycle:
    1 mm ID for all pouch sizes

    NX Advanced Cycle:
    2 mm ID for all pouch sizes

    100NX Cycle:
    1 mm ID for all pouch sizes

    100NX Flex Cycle:
    1 mm ID for all pouch sizes

    100NX Standard Cycle:
    2 mm ID for all pouch sizes

    Max lumen lengths:
    NX Standard Cycle:
    For pouch sizes 90 x 200 mm, 90 x 250 mm, 130 x 270 mm, 130 x 380 mm, 125 mm. For pouch sizes 190 x 330 mm, 250 x 400 mm, 300 x 450 mm, 250 mm.

    NX Advanced Cycle:
    For pouch sizes 90 x 200 mm, 90 x 250 mm, 130 x 270 mm, 130 x 380 mm, 125 mm. For pouch sizes 190 x 330 mm, 250 x 400 mm, 300 x 450 mm, 250 mm.

    100NX Duo Cycle:
    850 mm for all pouch sizes

    100NX Flex Cycle:
    850 mm for all pouch sizes

    100NX Standard Cycle:
    For pouch sizes 90 x 200 mm, 90 x 250 mm, 130 x 270 mm, 130 x 380 mm, 125 mm. For pouch sizes 190 x 330 mm, 250 x 400 mm, 300 x 450 mm, 250 mm.

    Pouch Sizes:
    300 x 450 mm
    250 x 400 mm
    190 x 330 mm
    130 x 380 mm
    130 x 270 mm
    90 x 250 mm
    90 x 200 mm

    Device Description

    Steriking® LT-Blueline Pouches with Tyvek® are intended to be used to contain medical devices to be terminally sterilized in the STERRAD® Sterilization System. The medical devices are inserted into the Pouch, sealed, and then sterilized in the STERRAD ® Sterilization System. Sterilization Cycles are noted below. After completion of the sterilization process, the Pouch maintains sterility of the enclosed medical devices until the seal is opened. These pouches are made from a plastic film and Tyvek that is heat-sealed on three sides. The fourth side is left opened and will be self-sealed when used.

    Steriking® LT-Blueline Pouches with Tyvek® are intended to allow sterilization of enclosed devices and to maintain sterility of the enclosed devices until used up to 1 year post sterilization. The products are for single use only.

    The pouches are constructed from Tyvek®/plastic films. The self-sealed pouches are self sealed prior to processing in the STERRAD® Sterilization Systems.

    Sterilization Systems:

    STERRAD® Cycles:
    NX Standard
    NX Advanced
    100NX Duo
    100NX Flex
    100NX Standard

    AI/ML Overview

    The document provided describes the Steriking® LT-Blueline Pouches with Tyvek®, which are intended to enclose devices for sterilization in STERRAD® Sterilizer Systems and maintain their sterility for up to one year. This device is a sterilization wrap, regulated as Class II.

    The study presented is a non-clinical performance study to demonstrate that the proposed device, a self-seal pouch, is as safe and effective as its predicate device, a heat-seal pouch, with an added adhesive strip for sealing.

    Here's an analysis of the acceptance criteria and the study results based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test CategoryTestAcceptance CriteriaReported Device PerformanceConclusion
    Package Integrity, Pre-Sterilization Shelf LifeDye MigrationNo leaks detected after dye migrationAll self-seal pouches passed the dye penetration at time zeroPass
    Package Integrity, Pre-Sterilization Shelf LifeTensileAll samples are > 1.5N/15mm per ASTM F88All samples are > 1.5N/15mm per ASTM F88Pass
    Package Integrity, Post-SterilizationTensileAll samples are > 1.5N/15mm per ASTM F88All samples are > 1.5N/15mm per ASTM F88Pass
    BiocompatibilityCytotoxicityNot greater than Grade 2 reactivity (mildly reactive)No reduction of cell growth and cell lysisPass
    Package Integrity, Post-SterilizationMicrobial Aerosol ChallengeA sample which demonstrates that all items remain sterile following exposureEach pouch met the sterility maintenance requirement as there was no growth in any of the culture tubes containing the stainless-steel coupons at the end if the incubation period.Pass
    Sterilization PenetrationSterilization PenetrationDemonstrating that a minimum of $1.0\times10^6$ Geobacillus stearothermophilus spores were killed in a half-cycle (6-log reduction) using a worst-case, end of shelf-life injection volumeNegative for growth following the minimum incubation period (for all STERRAD cycles: 100NX Standard, NX Standard, NX Advanced, 100NX Flex, 100NX Duo)Pass
    Post Sterilization Shelf LifeMaintenance of SterilityNo growth after exposurePass. No growth following exposure (for 100NX Standard cycle, aged 365 days)Pass
    Package Integrity, Post-SterilizationAccelerated AgingSeal maintains integrityPass. No suspect seals were observed, and the test samples remained intact. (for 100NX Standard cycle, aged 3 years accelerated aging)Pass
    Package Integrity, Post-SterilizationTensile (after accelerated aging)All samples are > $1.5N/15mm$ per ASTM F88All samples are > $1.5N/15mm$ per ASTM F88 (for 100NX Standard cycle, aged 3 years accelerated aging)Pass

    2. Sample Sizes Used for the Test Set and Data Provenance

    The document does not explicitly state the specific numerical sample sizes ("N") for each test. Instead, it refers to "All self-seal pouches," "All samples," and "Each pouch." This implies that the tests were conducted on a sufficient number of samples according to the referenced standards (e.g., ASTM, AAMI, ISO), which typically specify minimum sample sizes.

    The data provenance is not specified regarding country of origin. The study appears to be prospective as it's a performance study for a new iteration of a device (the addition of a self-sealing adhesive strip) on the market. The phrase "Studies on the self-seal pouch were performed concurrently to the heat-seal pouch studies" suggests newly generated data for the proposed device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. The tests performed are primarily laboratory-based physical, chemical, and microbiological evaluations against established standards, rather than expert-based assessments of images or patient data. Therefore, the concept of "ground truth" derived from human experts in the context of diagnostic interpretation is not applicable here.

    4. Adjudication Method for the Test Set

    This is not applicable as the tests are analytical and objective measurements against defined standards, not subjective assessments requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, an MRMC comparative effectiveness study was not done. The device is a sterilization pouch and not a diagnostic imaging or AI-driven decision support tool that would typically involve human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    No, a standalone study in the context of an algorithm or AI is not applicable. The device is a physical product (sterilization pouch), and the performance assessments are for the device's physical and functional properties.

    7. The Type of Ground Truth Used

    The ground truth used for these performance tests is based on established industry standards and scientific principles as outlined in various ASTM, AAMI, and ISO documents (e.g., ASTM F1929 for dye migration, ASTM F88 for tensile strength, ISO 10993-5 for cytotoxicity, AAMI TIR12 for microbial aerosol challenge and sterilization penetration). For biological tests like sterility, the "ground truth" is determined by the absence or presence of microbial growth under specified conditions (e.g., kill of $1.0\times10^6$ Geobacillus stearothermophilus spores, no growth after exposure).

    8. The Sample Size for the Training Set

    This is not applicable as there is no "training set" in the context of an AI/ML algorithm. The device is a physical product.

    9. How the Ground Truth for the Training Set was Established

    This is not applicable as there is no "training set" for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221379

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2023-01-04

    (237 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Steriking® LT-Blueline Pouches with Tyvek® are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERIS® V-PRO® Sterilizer Systems.

    The device is intended to allow sterilization of enclosed devices and to maintain sterility of the enclosed devices until used up to 3 years post sterilization. The products are for single use only.

    The Steriking® LT-Blueline Pouches with Tyvek® are intended for use in the following STERIS® V-PRO® Sterilization Cycles:

    Lumen Cycle
    Non Lumen Cycle
    Flexible Cycle

    Device lumen dimensions:

    Flexible Cycle: 1 lumen x 1 mm min ID x 1050 mm max length for all pouch sizes.

    Lumen Cycle: 1 lumen x 1 mm min ID for all pouch sizes. For pouch sizes 250x500mm, 250x390mm, 205x390mm, max length of 125 mm. For all other pouch sizes, max length of 50 mm.

    Max Weights: 1.410 pounds for pouch sizes below (Flexible, Lumen, and Non-Lumen Cycles) 250 mm x 500 mm 250 mm x 390 mm 205 mm x 390 mm

    .114 pounds for pouch sizes below (Flexible and Lumen Cycles) 160 mm x 600 mm 160 mm x 440 mm 150 mm x 300 mm 100 mm x 250 mm 75 mm x 200 mm

    .158 pounds for pouch sizes below (Non-Lumen Cycle) 160 mm x 600 mm 160 mm x 440 mm 150 mm x 300 mm 100 mm x 250 mm 75 mm x 200 mm

    Device Description

    Steriking® LT-Blueline Pouches with Tyvek® are intended to be used to contain medical devices to be terminally sterilized in the STERIS® Sterilization System. The medical devices are inserted into the Pouch, sealed, and then sterilized in the STERIS® Sterilization System. Sterilization Cycles are noted below. After completion of the sterilization process, the Pouch maintains sterility of the enclosed medical devices until the seal is opened. These pouches are made from a plastic film and Tyvek that is heat-sealed on three sides. The fourth side is left opened and will be heat-sealed when used.

    Steriking® LT-Blueline Pouches with Tyvek® are intended to allow sterilization of enclosed devices and to maintain sterility of the enclosed devices until used up to 3 years post sterilization. The products are for single use only.

    The pouches are constructed from Tyvek®/plastic films. The heat-sealed pouches are heat sealed prior to processing in the STERIS® V-PRO® Sterilization Systems.

    Sterilization Systems:

    STERIS® V-PRO® Cycles:
    Lumen Cycle,
    Non Lumen Cycle
    Flexible Cycle

    AI/ML Overview

    The information provided describes the Steriking LT-Blueline Pouches with Tyvek device, which is a sterilization wrap. The document details the non-clinical tests performed to demonstrate substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test MethodologyPurposeAcceptance CriteriaReported Device Performance
    ISO 10993-5 Biological Evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityBiocompatibilityNo reduction of cell growth or cell lysis was observedPass (did not elicit any cytotoxic effect)
    AAMI TIR12:2010 - Designing, testing, and labeling reusable medical devices for reprocessing in health care facilitiesSterilant PenetrationDemonstrating that a minimum of 1.0x10^6 Geobacillus stearothermophilus spores were killed in a half-cycle (6-log reduction)Pass (achieved 10^-6 Sterility Assurance Level of Geobacillus stearothermophilus, Negative for growth)
    ASTM F1929-15: Standard test method for detecting seal leaks in porous medical packaging by dye migrationPackage Integrity (Dye Migration)Dye Penetration: nonePass (no leaks detected after dye migration)
    ANSI/AAMI ST79:2017 - Comprehensive guide to steam sterilization and sterility assurance in health care facilities; AAMI TIR12:2010 - Designing, testing, and labeling reusable medical devices for reprocessing in health care facilitiesMicrobial Barrier PropertiesA sample which demonstrates that all items remain sterile following exposurePass (each pouch met the sterility maintenance requirement; no growth in culture tubes containing stainless-steel coupons)
    ASTM F1929-15: Standard test method for detecting seal leaks in porous medical packaging by dye migration; ASTM F88: Standard Test Method for Seal Strength of Flexible Barrier MaterialsShelf Life Pre SterilizationDye Penetration: none; Seal peel: =>1.5N/15mmPass (>1.5N/15mm per ASTM F88, No leaks detected after dye migration)
    ASTM F1980 – 16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices; ASTM F1929-15: Standard test method for detecting seal leaks in porous medical packaging by dye migration; ASTM F88: Standard Test Method for Seal Strength of Flexible Barrier MaterialsShelf Life Post Sterilization (Accelerated Aging)Dye Penetration: none; Seal peel: =>1.5N/15mmPass (>1.5N/15mm per ASTM F88, No leaks detected after dye migration)
    Various ASTM, TAPPI, and ISO standards (listed in document)Material CompatibilityMeets specifications per Technical Data Sheets. Included in Bench Testing Summary.Pass

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the specific sample sizes for each test. It refers to "samples," "items," and "pouches" but does not provide numerical quantities for the test sets.

    The data provenance is from non-clinical tests conducted on the device itself (Steriking LT-Blueline Pouches with Tyvek). The country of origin of the data is not specified, but the applicant, Wipak Oy, is located in Nastola, Finland, suggesting the testing may have been conducted there or by a contracted lab. The studies are prospective in the sense that they are designed to evaluate the performance of the manufactured device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is a non-clinical device performance study, not a study involving human interpretation of medical images or data. Therefore, the concept of "experts" to establish ground truth in the context of medical imaging interpretation is not applicable here. The "ground truth" is established by adherence to recognized international and industrial standards (e.g., ISO, ASTM, AAMI) for testing materials and sterilization packaging.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is a non-clinical device performance study, not a study involving diagnostic interpretation or human consensus. Test results are determined by quantitative measurements and observations against pre-defined criteria from recognized standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a non-clinical device performance study for a sterilization pouch, not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a non-clinical device performance study for a sterilization pouch, not an algorithm. The device performance itself is "standalone" in the sense that it is evaluated directly against physical and biological criteria.

    7. The type of ground truth used

    The ground truth is based on established scientific and engineering standards and validated test methodologies.

    • Biocompatibility: Absence of cytotoxicity as per ISO 10993-5.
    • Sterilant Penetration: 6-log reduction of Geobacillus stearothermophilus spores, demonstrating a 10^-6 Sterility Assurance Level. This is a biological indicator's response to sterilization.
    • Package Integrity (Dye Migration): Absence of dye penetration as per ASTM F1929-15.
    • Microbial Barrier Properties: Maintenance of sterility (no microbial growth) after exposure to aerosolized spores as per ISO 11607 and related AAMI standards.
    • Seal Strength: A minimum peel strength of 1.5N/15mm as per ASTM F88.
    • Material Compatibility: Meeting specifications defined in technical data sheets.

    8. The sample size for the training set

    Not applicable. This is a non-clinical device performance study, not a machine learning model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As this is not a machine learning study, there is no training set or associated ground truth for that purpose.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221377

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-12-22

    (224 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SteriKing® LT Blueline with Tyvek® pouches are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERRAD® Sterilizer Systems.

    The device is intended to allow sterilization of enclosed devices and to maintain sterility of the enclosed devices until used up to 3 years post sterilization. The products are for single use only.

    The SteriKing® LT Blueline with Tyvek® pouches are intended for use in the following STERRAD® Sterilization Cycles:

    NX Standard
    NX Advanced
    100NX Duo
    100NX Flex
    100NX Standard

    Max weights:

    NX Standard Cycle:

    75 mm x 200 mm, 100 mm x 250 mm x 300 mm, 160 mm x 440 mm x 600 mm . 088 pounds
    205 mm x 390 mm, 250 mm x 390 mm, 250 mm x 500 mm: 1.382 pounds

    NX Advanced Cycle:

    75 mm x 200 mm, 100 mm x 250 mm x 300 mm, 160 mm x 440 mm x 600 mm . 084 pounds
    205 mm x 390 mm, 250 mm x 390 mm, 250 mm x 500 mm: 1.393 pounds

    100NX Duo Cycle:
    75 mm x 200 mm, 100 mm x 250 mm x 300 mm, 160 mm x 440 mm x 600 mm : 104 pounds
    205 mm x 390 mm, 250 mm x 390 mm, 250 mm x 500 mm: 1.400 pounds

    100NX Flex Cycle:
    75 mm x 200 mm, 100 mm x 250 mm x 300 mm, 160 mm x 440 mm, 160 mm x 600 mm: .158 pounds
    205 mm x 390 mm, 250 mm x 390 mm, 250 mm x 500 mm: 1.400 pounds

    100NX Standard Cycle:
    75 mm x 200 mm, 100 mm x 250 mm x 300 mm, 160 mm x 440 mm x 600 mm: .158 pounds
    205 mm x 390 mm, 250 mm x 390 mm, 250 mm x 500 mm: 4.476 pounds

    Max lumens:

    1 mm min ID and 1 max lumen for all cycles

    NX Standard Cycle: For pouch sizes 250x500mm, 205x390mm is 125 mm max length. Max length is 50 mm for all other pouch sizes.

    NX Advanced Cycle: For pouch sizes 250x500mm, 205x390mm is 125 mm max length is 50 mm for all other pouch sizes.

    100NX Duo Cycle: For pouch sizes 250x500mm, 250x390mm is 875 mm max length. Max length is 850 mm for all other pouch sizes.

    100NX Flex Cycle: Max length is 850 mm for all pouch sizes.

    100NX Standard Cycle: For pouch sizes 250x500mm, 205x390mm is 125 mm max length, Max length is 50 mm for all other pouch sizes.

    Pouch Sizes:
    250 mm x 500 mm
    250 mm x 390 mm
    205 mm x 390 mm
    160 mm x 600 mm
    160 mm x 440 mm
    150 mm x 300 mm
    100 mm x 250 mm
    75 mm x 200 mm

    Device Description

    Steriking® LT-Blueline Pouches with Tyvek® are intended to be used to contain medical devices to be terminally sterilized in the STERRAD® Sterilization System. The medical devices are inserted into the Pouch, sealed, and then sterilized in the STERRAD® Sterilization System. Sterilization Cycles are noted below. After completion of the sterilization process, the Pouch maintains sterility of the enclosed medical devices until the seal is opened. These pouches are made from a plastic film and Tyvek that is heat-sealed on three sides. The fourth side is left opened and will be heat-sealed when used.

    Steriking® LT-Blueline Pouches with Tyvek® are intended to allow sterilization of enclosed devices and to maintain sterility of the enclosed devices until used up to 3 years post sterilization. The products are for single use only.

    The pouches are constructed from Tyvek®/plastic films. The heat-sealed pouches are heat sealed prior to processing in the STERRAD® Sterilization Systems.

    Sterilization Systems:

    STERRAD® Cycles:
    NX Standard
    NX Advanced
    100NX Duo
    100NX Flex
    100NX Standard

    AI/ML Overview

    This document is a 510(k) Premarket Notification from the FDA, and as such, it describes the non-clinical testing performed for a medical device (sterilization pouches), rather than an AI-powered software device. Therefore, many of the requested elements for describing an AI study (e.g., sample size for test/training sets, expert ground truth, MRMC studies) are not applicable in this context.

    However, I can extract the information related to the acceptance criteria and the studies that prove the device meets these criteria, focusing on the traditional medical device testing rather than AI performance.

    Here's the breakdown of the acceptance criteria and study findings for the SteriKing® LT-Blueline Pouches with Tyvek®:

    1. A table of acceptance criteria and the reported device performance

    Test MethodologyPurposeAcceptance CriteriaReported Device Performance
    ISO 10993-5 Biological Evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityBiocompatibilityNo reduction of cell growth or cell lysis was observedPass
    AAMI TIR12:2010 - Designing, testing, and labeling reusable medical devices for reprocessing in health care facilitiesSterilant PenetrationDemonstrating that a minimum of 1.0x10^6 Geobacilus stearothermophilus spores were killed in a half-cycle (6-log reduction)Pass
    ASTM F1929-15: Standard test method for detecting seal leaks in porous medical packaging by dye migrationPackage IntegrityDye Penetration: nonePass
    ANSI/AAMI ST79:2017 - Comprehensive guide to steam sterilization and sterility assurance in health care facilities, AAMI TIR12:2010Microbial Barrier PropertiesA sample which demonstrates that all items remain sterile following exposurePass (No growth in culture tubes containing stainless-steel coupons after incubation)
    ASTM F1929-15: Standard test method for detecting seal leaks in porous medical packaging by dye migration, ASTM F88: Standard Test Method for Seal Strength of Flexible Barrier MaterialsShelf Life Pre SterilizationDye Penetration: none, Seal peel: =>1.5N/15mmPass
    ASTM F1980 – 16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices, ASTM F1929-15, ASTM F88Shelf Life Post SterilizationDye Penetration: none, Seal peel: =>1.5N/15mmPass
    Various ASTM/ISO/TAPPI standards (F1608, F2638, D3776, D2724, T460, 1924-2, D1424, 534, 5636-3, T523, 2758, D3420)Material CompatibilityMeets specifications per Technical Data Sheets.Pass

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state numerical sample sizes for each test. However, it mentions "test samples" and "stainless-steel coupons" were used. The data provenance is based on lab testing of the device itself, likely conducted by the manufacturer (Wipak Oy in Finland) for submission to the FDA. The tests are non-clinical (laboratory-based) rather than retrospective or prospective human data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This is not applicable as the testing involves physical and biological properties of a medical device (sterilization pouches), not an AI model requiring human expert interpretation for ground truth. Ground truth is established through adherence to recognized international standards and documented laboratory methods (e.g., measuring direct physical properties, culturing for microbial growth, assessing cytotoxicity).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a human interpretation study. The results are based on objective measurement against predefined standards.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-powered device or a study involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI algorithm. Its performance is evaluated through its physical and biological integrity and functionality.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for this device is based on:

    • Compliance with recognized international and national standards: e.g., ISO 10993-5 (Biocompatibility), AAMI TIR12:2010 (Sterilant Penetration, Microbial Barrier), ASTM F1929-15 (Seal Leaks, Dye Migration), ASTM F88 (Seal Strength), ASTM F1980 (Accelerated Aging).
    • Objective laboratory measurements: Such as absence of cell growth/lysis, 6-log reduction in spores, absence of dye penetration, specified seal strength values, and "no growth in any of the culture tubes."
    • Performance specifications: Material compatibility against technical data sheets.

    8. The sample size for the training set

    Not applicable. There is no training set as this is not an AI or machine learning model.

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this type of medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221016

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-05-05

    (30 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Steriking Packaging for Medical Devices serve as an enclosure for medical devices during steam sterilization that maintains the sterility of the enclosed medical devices until use of the medical devices with a combined weight of metal and plastics of 2.6 pounds or less. The recommended sterilization cycles are as follows: Pre-vacuum steam at 132oC for 4 minutes; Drying time of 20 minutes Pre-vacuum steam at 1350C for 3 minutes; Drying time of 16 minutes

    Device Description

    The Steriking Packaging for Medical Devices consists of a paper backing (Bleached wood pulp, grammage 100 g/m2) with transparent plastic film laminate front (2 sheets of laminated plastic with a total grammage of 55 g/m2, 1 sheet of oriented polyester 12 microns thick, 1 sheet of coextruded polypropylene 40 microns thick. The plastic laminate is triple heat sealed to the backing paper. The open end of the pouch is to be self sealed once a device is inserted. Self-sealable pouches are featured with adhesive strip allowing tight, impermeable closing of a pack. The closing flap is pre-folded to facilitate the closure. When closing the self-seal pouch the paper flap shall be folded along the pre-folded line. Flap should be pressed firmly against the laminate from the centre working outwards to ensure a good, even seal. Dimensional configurations as follows (2 sizes 200mm x 800mm, 250mm x 875mm).

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device, which means it's a submission to the FDA to demonstrate that the device is substantially equivalent to a legally marketed predicate device. This type of submission relies on showing equivalence to an existing device rather than presenting new clinical study data with acceptance criteria for a novel AI device.

    Therefore, the requested information about acceptance criteria, specific study details (sample size, data provenance, expert ground truth, MRMC studies, standalone performance, training set details), and ground truth establishment methods is not applicable to this document. This document details the technical characteristics of a sterilization packaging material and its comparison to a predicate device, not the performance of an AI/ML algorithm.

    The table for acceptance criteria and reported device performance is presented in the original document as a comparison between the submitted device and its predicate, rather than against specific pre-defined acceptance criteria for an AI model.

    Here's a summary of the information provided in the document, framed in the context of device equivalence rather than AI performance:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document provides a comparison table (pages 4-6) showing the technical characteristics of the new device ("Submission Device – Steriking Packaging for Medical Devices") against its predicate device ("Predicate Device – Steriking Packaging for Medical Devices K210810"). This comparison serves as the "acceptance criteria" in the context of a 510(k) submission, where the new device must be substantially equivalent to the predicate. "Reported device performance" here refers to the measured physical and functional properties of the submitted device.

    Comparison ElementSubmission Device – Steriking Packaging for Medical DevicesComparisonPredicate Device – Steriking Packaging for Medical Devices K210810
    Intended UseTo serve as an enclosure for medical devices during steam sterilization that maintains the sterility of the enclosed devices until used.No ChangeTo serve as an enclosure for medical devices during steam or gas sterilization that maintains sterility of the enclosed device until used.
    DesignPouches plastic film triple heat sealed to paper backing thumb notches chevron-type seal at end for opening Adhesive strip for self sealDifferentPouches plastic film triple heat sealed to paper backing thumb notches chevron-type seal at end for opening
    Principle of OperationMedical device to be sterilized is put into pouch and the open parts of the pouches are closed by self-sealing. Sterilization packages then are subjected to validated sterilization operation of steam. Sterilant penetration is carried out through the medical grade paper into the package and microorganisms on the surface of the medical device are destroyed with the effect of the sterilant process. Other parameters of the sterilization process are temperature, pressure, humidity, time and are determined according to the sterilization type. After the sterilization is completed, the sterility of the enclosed medical device is maintained for 12 months.DifferentMedical device to be sterilized is put into pouch and the open parts of the pouches are closed by heat sealing. Sterilization packages then are subjected to validated sterilization operation of steam. Sterilant penetration is carried out through the medical grade paper into the package and microorganisms on the surface of the medical device are destroyed with the effect of the sterilant process. Other parameters of the sterilization process are temperature, pressure, humidity, time and are determined according to the sterilization type. After the sterilization is completed, the sterility of the enclosed medical device is maintained for 12 months.
    Dimensions200mm x 800mm, 250mm x 875 mmDifferent200mm x 800mm, 250mm x 900 mm
    Backing PaperBleached wood pulp, grammage: 100g/m2No ChangeBleached wood pulp, grammage: 100 g/m2
    Tensile Strength MD-kNm>66 N/15mmNo Change>66 N/15mm
    Tensile Strength CD-kNm>33 N/15mmNo Change>33 N/15mm
    Tear Strength MD-mN>550 NmNo Change>550 Nm
    Tear Strength CD-mN>550 NmNo Change>550 Nm
    Burst Strength-kPa>230 kpaNo Change>230 kpa
    Porosity ISO 5636-3 ISO 5636-53.9 – 5.7 µm/Pa·s 24-34 sNo Change3.9 – 5.7 µm/Pa·s 24-34 s
    Seal Strength – N/mmPeel ≥ 1.5 N/15mmNo ChangePeel ≥ 1.5 N/15mm
    Transparent FilmTwo sheets of laminated plastic with a total grammage of 55 g/m2. One sheet of oriented polyester 12 microns thick. One sheet of coextruded polypropylene 40 microns Thick.No ChangeTwo sheets of laminated plastic with a total grammage of 55 g/m2. One sheet of oriented polyester 12 microns thick. One sheet of coextruded polypropylene 40 microns Thick.
    Sterilization PropertiesSteam sterilization conditions are 4 minutes at 132° C or 3 minutes at 135° CNo ChangeSteam sterilization conditions are 4 minutes at 132° C or 3 minutes at 135° C
    Sterilant PenetrationFull-cycle steam sterilization process will produce sufficient lethality to achieve a 12-log reduction, thus providing a 10-6 Sterility Assurance Level (SAL)No ChangeFull-cycle steam sterilization process will produce sufficient lethality to achieve a 12- log reduction, thus providing a 10-6 Sterility Assurance Level (SAL).
    Material CompatibilityCompatible with Steam SterilizationNo ChangeCompatible with Steam Sterilization
    Package Integrity TestClosure integrity maintained before and after steam sterilizationNo ChangeClosure integrity maintained before and after steam sterilization
    Maintenance of Sterility — Shelf life post sterilization12 monthsNo Change12 months
    Shelf Life - Pre-sterilization5 yearsNo Change5 years
    Drying Time20 minutesNo Change20 minutes
    Microbial Barrier PropertiesThe paper of the sterile barrier system was examined on the packaging outer side for its germ proofness with air permeance after steam sterilization and is evaluated as "sufficiently germ-proof"No ChangeThe paper of the sterile barrier system was examined on the packaging outer side for its germ proofness with air permeance after steam sterilization and is evaluated as "sufficiently germ-proof"
    BiocompatibilityNon-CytotoxicNo ChangeNon-Cytotoxic

    The following points are N/A (Not Applicable) as the document describes a traditional medical device (sterilization packaging) and not an AI/ML powered device.

    1. Sample sizes used for the test set and the data provenance: N/A
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: N/A
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: N/A
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: N/A
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: N/A
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): N/A
    7. The sample size for the training set: N/A
    8. How the ground truth for the training set was established: N/A
    Ask a Question

    Ask a specific question about this device

    K Number
    K210810

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-01-23

    (312 days)

    Product Code
    Regulation Number
    880.6850
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Steriking Packaging for Medical Devices serve as an enclosure for medical devices during steam sterilization that maintains the sterility of the enclosed medical devices until use of the medical devices with a combined weight of metal and plastics of 2.6 pounds or less. The recommended sterilization cycles are as follows: Pre-vacuum steam at 132oC for 4 minutes; Drying time of 20 minutes Pre-vacuum steam at 1350C for 3 minutes; Drying time of 16 minutes

    The Steriking Packaging for medical devices consists of a paper backing (Bleached wood pulp, grammage 100 g/m2) with transparent plastic film laminate front (2 sheets of laminated plastic with a total grammage of 55 g/m2, 1 sheet of oriented polyester 12 microns thick, 1 sheet of coextruded polypropylene 40 microns thick. The plastic laminate is triple heat sealed to the backing paper. The open end of the pouch is to be heat sealed once a device is inserted. Heat sealing parameters to provide a sterile barrier are 165oC-200oC (329oF-392oF).

    Steriking Packaging for Medical Devices maintains the sterility of the enclosed devices for up to 12 months post Steam sterilization and before sterilization has a maximum shelf life of 5 years from the date of manufacture. Steriking Packaging for Medical Devices Dimensional configurations (2 sizes 200mm x 800mm)

    Device Description

    The Steriking Packaging for Medical Devices consists of a paper backing (Bleached wood pulp, grammage 100 g/m2) with transparent plastic film laminate front (2 sheets of laminated plastic with a total grammage of 55 g/m2, 1 sheet of oriented polyester 12 microns thick, 1 sheet of coextruded polypropylene 40 microns thick. The plastic laminate is triple heat sealed to the backing paper. The open end of the pouch is to be heat sealed once a device is inserted. Heat sealing parameters to provide a sterile barrier are 165°C – 200°C (329°F - 392°F) Dimensional configurations as follows (2 sizes 200mm x 800mm, 250mm x 900mm).

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification summary for "Steriking Packaging for Medical Devices." It details the device's characteristics, indications for use, comparison to a predicate device, and the results of non-clinical testing to demonstrate substantial equivalence.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based solely on the provided text, formatted to address your specific points.

    The document describes pre-market testing of a medical device packaging system, not an AI/ML-driven device. Therefore, many of the questions related to AI/MRMC studies, training sets, and expert consensus for ground truth are not applicable to this kind of submission. I will address the applicable parts of your request.


    Acceptance Criteria and Device Performance for Steriking Packaging for Medical Devices

    The study proves that the device meets its acceptance criteria through a series of non-clinical performance tests, designed to demonstrate the safety and effectiveness of the packaging for maintaining sterility of medical devices after steam sterilization.

    1. A table of acceptance criteria and the reported device performance

    The document provides a detailed table within the "Summary of Non-Clinical Testing" section. This table lists the test methodology, purpose, acceptance criteria, and the results (which are uniformly "Pass" or specific measurements that meet the criteria).

    Test MethodologyPurposeAcceptance CriteriaReported Device Performance
    ISO 14477 - Packaging. Flexible packaging material. Determination of puncture resistance.Puncture ResistanceNo pass/fail criteria (Note: This is unusual for an acceptance criterion, but listed as such in the document)Acceptable => 13.1N
    ASTM F1929-15 "Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration" EN 868-5 Packaging for terminally sterilized medical devices - Part 5: Sealable pouches and reels of porous materials and plastic film construction - Requirements and test methods ISO 5636-3 Paper and board — Determination of air permeanceSHELF LIFE Prior to SterilizationSeal Strength: =>1.5N/15mm Seal Integrity: No dye leaks Air Permeability: =>3.4um/Pa Peel Properties: free of fibers after peelingPass
    AAMI TIR 12:2010, Designing, Testing, and Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: A Guide for Medical Device ManufacturersSterilant Penetration & Drying TimeBiological indicator 6-log reduction in half cycle providing 10-6 SAL Drying time post sterilization moisture content weight gain <3%Pass
    DIN EN ISO -11607-1:2017-10 Section 5.1.6. a) microbisl barrier DIN 953-6, section 4.8.6. DIN 58953-6, section 3.8 and section 4.8.6.Microbial Barrier PropertiesTransmission of CFUs (Colony Forming Units) from exterior of pouch to interior of pouch = 0Pass
    DIN 58953-6, section 3.8 and section 4.8.6. 1. Tensile strength of film ASTM D882 2. Tensile strength of paper (MD) ISO 1924-3 3. Tensile strength of paper (CD) ISO 1924-3 4. Thickness 5. Tear resistance ISO 1974 6. Tear strength DIN 53363 7. Burst strength ISO 2758 8. Air permeance DIN 5636-5 9. Permeance after sterilization DIN 5636-5 10. Seal strength ASTM F88 11. Peel Characteristics EN 868-5:2009 12. Color fastness EIN DIN 646 13. No delamination or clouding 868-5:2009 14. Pinholes in film EN 868-5:2009Material Compatibility1. =>20N/15mm 2. =>66N/15mm 3. =>33N/15mm 4. =152 +/- 10% μm 5. >550mN (MD & CD) 6. >10N 7. >230kPa 8. 3.9 – 5.7 µm/PAS 9. Corresponds 10. 1.5N/15mm 11. No irregularities or splitting >10mm 12. evaluation grade = 5 13. No objections 14. No pinholesPass
    ISO 10993-5, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityBiocompatibilityIf viability is reduced to <70% of the medium control then test item is considered to have cytotoxic potentialPass (88% & 92% viability)
    ASTM F1980 - 16 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical DevicesShelf Life (post sterilization)Dye Penetration: none Seal peel: =>1.5N/15mmPass
    ASTM F1929 - Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye PenetrationPackage IntegrityNo dye penetration/channeling travelling full width of seal in ~5 secondsPass

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample sizes (number of units tested) for each non-clinical test. This type of information is typically found in the full test reports referenced by the standards, not in the summary.

    Data Provenance: The manufacturing entity is Wipak Oy in Nastola, Finland, indicating the origin of the device. The testing was non-clinical (laboratory/material testing) rather than clinical data from human subjects. Therefore, terms like "retrospective" or "prospective" and "country of origin of the data" in the sense of patient data are not applicable. The testing was performed according to international standards (e.g., ISO, ASTM, EN, DIN), which are universally recognized.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This question is not applicable. The device is a medical packaging product, and its "ground truth" is established through physical and material science testing against recognized international engineering and materials standards (e.g., ISO, ASTM). These tests do not involve expert human interpretation in the way, for example, a diagnostic imaging AI would. The "ground truth" is determined by the objective measurements and criteria defined in the specific test methodologies.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This question is not applicable. Adjudication methods are typically used in studies involving human readers or subjective interpretations, such as clinical trials or diagnostic accuracy studies. For material and physical properties testing, the results are objectively measured against predefined scientific criteria, not adjudicated.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This question is not applicable. The submission is for a medical device packaging material, not an AI or imaging device that would involve human readers or comparative effectiveness studies of diagnostic performance.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This question is not applicable. The device is physical packaging, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device is established by objective measurements of physical, chemical, and biological properties against established industry and international standards (e.g., seal strength, microbial barrier, puncture resistance, sterility assurance level, material compatibility, cytotoxicity). This is based on scientific and engineering principles, not expert consensus, pathology, or outcomes data in the clinical sense.

    8. The sample size for the training set

    This question is not applicable. "Training set" refers to data used to train machine learning models. This submission is for a physical medical device packaging product, not an AI/ML device.

    9. How the ground truth for the training set was established

    This question is not applicable, as there is no training set mentioned or relevant to this type of device submission.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1