510(k) Data Aggregation

    K Number
    K200618

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-12-07

    (273 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HydroSens A (hioxifilcon A) Spherical Hydrogel Soft Contact Lens are daily disposable soft contact lens indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

    Device Description

    The HydroSens A (hioxifilcon A) Hydrogel Soft Contact Lens is made of hydrogel material, hioxifilcon A, with UV blocker available as spherical lens. The composition of the lens is 43% hioxifilcon A and 57% water. A light blue color tinted with "reactive Blue19" listed in 21 CFR Part 73.3121 is for handling visibility purpose. The UV transmission (the thinnest lens measured by spectrophotometry as stated in ISO 18369) is less than 50% in the UVA range of 315 - 380 nm and less than 5% in the range of UVB range of 280-315 nm. Lenses are supplied sterile in sealed blister packs containing MPC polymer (2-(Methacryloyloxy) ethyl 2-(trimethylammonio) ethyl phosphate-n-butyl methacrylate copolymer) with isotonic buffered saline solution.

    AI/ML Overview

    The provided text is a 510(k) Summary for the HydroSens A (hioxifilcon A) Spherical Hydrogel Soft Contact Lens. It outlines the device's characteristics and its comparison to predicate devices to demonstrate substantial equivalence, but it explicitly states that no clinical study was conducted to prove the device meets acceptance criteria. Therefore, most of the requested information regarding a study, sample sizes, experts, adjudication, MRMC, and standalone performance cannot be provided as they are not applicable to this submission.

    However, based on the provided "Technological characteristic" and "Performance Data" sections, we can infer some aspects of what would typically be considered "acceptance criteria" for a contact lens and how its performance was assessed non-clinically.

    Here's a breakdown of the requested information, adapted to what is available in the text:

    1. Table of acceptance criteria and the reported device performance:

    Since no clinical study was performed with explicit acceptance criteria and corresponding performance metrics from that study, I will extract relevant technical specifications of the HydroSens A lens from the "Technological characteristic" and "Performance Data" sections, along with performance from non-clinical physiochemical studies. These values represent the device's intended performance.

    Criterion (Inferred from Technological Characteristics & Performance Data)Acceptance Criteria (Implied / Design Specification)Reported Device Performance (from Non-Clinical Tests)
    Dimensional Characteristics:
    Diameter Range13.0 to 15.0 mm, ±0.2mmStated as design specification
    Base Curve8.0 to 9.2 mm, ±0.2mmStated as design specification
    Center Thickness (-3.00D)0.08 mm, ± [0.010 mm + 0.10 tc]Stated as design specification
    Power Range (forFL≤ 10.00 D)
    Power Range (for 10.00 D <FL≤ 20.00 D)
    Material Properties:
    Water Content57%57%
    Light Transmittance95%95%
    Refractive Index (hydrated)N/A (compared to 1.404 for predicates)1.402
    Oxygen Permeability (Dk, 35°C)N/A (compared to 18x10-11 for predicates)22x10-11
    Specific Gravity (hydrated)N/A (compared to 1.136 for predicates)1.1300
    ModulusN/A (compared to 0.34 Mpa for predicate)0.38 Mpa
    Tensile StrengthN/A (compared to 36 gf/mm² for predicate)35.0 gf/mm²
    Elongation at BreakN/A (compared to 128% for predicate)108%
    ToughnessN/A (compared to 0.215 J/m3 for predicate)0.210 J/m3
    UV Protection:
    UV-A Transmission (315-380 nm)Less than 50%Less than 50%
    UV-B Transmission (280-315 nm)Less than 5%Less than 5%
    Sterilization & Biocompatibility:
    SterilitySterile in sealed blister packsDemonstrated through testing against ISO17665-1/2
    BiocompatibilityNon-toxic and biocompatibleDemonstrated through testing against ISO 10993 series

    2. Sample size used for the test set and the data provenance:

    • Sample Size for Test Set: Not applicable. No clinical study was performed. Non-clinical tests would have used various sample sizes per test (e.g., a batch of lenses for physical property testing, materials for biocompatibility), but specific numbers are not provided.
    • Data Provenance: Not applicable for a clinical test set. Non-clinical tests were conducted by the manufacturer (Visco Vision Inc., Taiwan).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. No clinical study for which experts would establish ground truth was conducted.

    4. Adjudication method for the test set:

    • Not applicable. No clinical study with adjudication was performed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. This is a contact lens, not an AI-powered diagnostic device, and no clinical study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is a contact lens, not an algorithm.

    7. The type of ground truth used:

    • Not applicable for a clinical study. For non-clinical performance, the "ground truth" is based on established scientific and engineering standards and validated test methods (e.g., ISO 18369 for ophthalmic optics, ISO 10993 for biocompatibility) that define the acceptable range for physical, chemical, and biological properties of contact lenses.

    8. The sample size for the training set:

    • Not applicable. No machine learning algorithm requiring a training set was used.

    9. How the ground truth for the training set was established:

    • Not applicable. No machine learning algorithm requiring a training set was used.

    In summary: The submission for the HydroSens A contact lens relies on demonstration of substantial equivalence to legally marketed predicate devices through comparison of indications for use, technological characteristics, and non-clinical performance testing (biocompatibility, sterilization, stability, and physiochemical properties) to established standards, rather than a clinical study with acceptance criteria for patient outcomes.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Daily Breeze B (olifilcon B) Spherical Silicone Hydrogel Soft Contact Lens is indicated as daily wear and single use soft contact lens for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with nondiseased eyes who exhibit refractive astigmatism of 1.00 D or less here the astigmatism does not interfere with visual acuity.

    The Daily Breeze B (olifilcon B) Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear and single use for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic persons with nondiseased eyes and whose powers are from -12.00 to +8.00 D and astigmatism corrections are from -0.75 to -2.25 D.

    The Daily Breeze B (olifilcon B) Multifocal Silicone Hydrogel Soft Contact Lenses are indicated as daily wear and single use for the correction of ametropia (myperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -12.00 to +8.00 D with add powers from +0.25 to +2.75 D. The lenses may be worn by persons who exhibit astigmatism of 1.00 D or less where the astigmatism does not interfere with visual acuity.

    Evecare practitioners prescribe the lens for single-use disposable wear.

    Device Description

    The Daily Breeze B (olifilcon B) Silicone Hydrogel Soft Contact Lens is made of hydrogel material, olifilcon B, with UV blocker available as spherical lens, toric lens and multifocal lens. The composition of the lens is 53% olifilcon B, which is a copolymer of 1-vinyl-2-pyrrolidinone (NVP) and Siloxane macromer, and 47% water. A light blue color tinted with "reactive Blue19" listed in 21 CFR Part 73.3121is for handling visibility purpose. A benzotriazole UV absorbing monomer is used to block UV radiation. The UV transmission (the thinnest lens measured by spectrophotometry as stated in ISO 18369) is less than 50% in the UVA range of 315 - 380 nm and less than 5% in the range of UVB range of 280-315 nm.

    Lenses are supplied sterile in sealed blister packs containing isotonic buffered saline solution. The compatibility and package integrity of the blister pack packaging system has been demonstrated and successfully used for other marketed lens products, and packaged lenses are effectively steam sterilized in a validated autoclave. Blister pack containers are labeled with the lens parameters, lot number and product expiration date. The expiration date has been established through stability studies that have assessed the chemical stability of the lens and package integrity (ability to maintain sterility). Shelf-life studies are ongoing to establish and extend the labeled expiration date.

    AI/ML Overview

    The provided text is a 510(k) summary for a contact lens device, not an AI/ML medical device. Therefore, it does not contain the information requested regarding acceptance criteria and study details for an AI/ML device.

    Specifically, the document states:
    "This 510(k) submission does not utilize clinical study for establishing substantial equivalence therefore this section does not apply." (Page 7, Section 11. Clinical study)

    The information requested regarding AI/ML device performance, such as sample sizes for test and training sets, expert qualifications, ground truth establishment, MRMC studies, and standalone performance, is not present in this document.

    The document focuses on the physicochemical properties, manufacturing process, and comparison to a predicate contact lens device to demonstrate substantial equivalence, which is typical for a contact lens submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191397

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-10-02

    (131 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OxySoft (olifilcon C) Spherical Silicone Hydrogel soft contact lenses are daily wear single use soft contact lens indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with nondiseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

    The OxySoft (olifileon C) Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters.

    The OxySoft (olifilcon C) Multifocal Silicone Hydrogel Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with nondiseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.

    Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be discarded upon removal. Therefore, no cleaning or disinfecting is required.

    Device Description

    The OxySoft (olifilcon C) Silicone Hydrogel Soft Contact Lens is made of silicone hydrogel material with UV blocker available as spherical lens, toric lens and multifocal lens. The composition of the lens is 46% olifilcon C and 54 % water. A light blue color tinted with "reactive Blue19" is for handling visibility purpose. A benzotriazole UV absorbing monomer is used to block UV radiation. The transmittance characteristics are 1.9% (< 5%) in the UVB range of 280-315nm and 14.7% (< 50%) in the UVA range of 316-380nm. It is supplied in a sterile state packaged in a buffered saline solution.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a soft contact lens called OxySoft (olifilcon C). It does not describe an AI medical device. Therefore, a response matching the format of an AI device's acceptance criteria and study cannot be generated. However, I can summarize the information provided regarding the clinical study conducted for the contact lens.

    Here's a summary of the clinical study for the OxySoft (olifilcon C) silicone hydrogel soft contact lens:

    Clinical Study for OxySoft (olifilcon C) Silicone Hydrogel Soft Contact Lens

    1. Acceptance Criteria and Reported Device Performance (as inferred from the summary):

    The document does not explicitly state quantitative acceptance criteria for device performance in the same way an AI device would have specific metrics (e.g., sensitivity, specificity). Instead, the study aimed to demonstrate safety and effectiveness by comparison to a predicate device, focusing on:

    MetricDesired Outcome (Inferred Acceptance)Reported Performance
    SafetyNo significant adverse events or findings requiring treatment.- No adverse event (AE) occurred in the study.- No symptom or syndrome requiring treatment for all subjects during the study period.- No significant slit lamp findings (i.e., Grade 3 or Grade 4) were found.
    Effectiveness (Visual Acuity)Visual acuity comparable to the predicate device and correctable to 0.1 LogMAR or better.- The mean visual acuity (VA) for both lenses (OxySoft and predicate Si-Hy) was similar at each visit.- Visual acuity could be corrected to 0.1 LogMAR or better for all subjects.- None of the subjects had VA decreased more than 2 lines.
    Wear TimeSafe for daily wear up to 9 hours a day for 30 days.Found safe for daily disposable single use, worn up to 9 hours a day during 30 days follow-up.

    2. Sample Size and Data Provenance:

    • Test Set Sample Size: 36 subjects were enrolled.
      • Trial group (OxySoft): 24 subjects
      • Control group (Si-Hy predicate): 12 subjects
    • Data Provenance: The document states, "All of them are Asia people." This suggests the data was collected in an Asian population, likely Taiwan, given the manufacturer's location. The study was prospective.

    3. Number of Experts and Qualifications for Ground Truth:

    • This information is not provided in the document. For contact lens studies, "ground truth" usually refers to clinical measurements (visual acuity, slit lamp evaluations by optometrists/ophthalmologists) rather than expert consensus on images. The clinical evaluations were likely performed by the study investigators or their staff.

    4. Adjudication Method for the Test Set:

    • This information is not provided in the document. Clinical trials typically have protocols for handling discrepancies in measurements or adverse events, but a specific "adjudication method" in the context of expert review for ground truth (as relevant to AI) is not applicable here.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    • No, an MRMC study was not done. This type of study model is generally used for diagnostic devices that involve human interpretation of images, often with AI assistance. This document describes a clinical study for a physical medical device (contact lens).

    6. Standalone (Algorithm Only) Performance:

    • No, this is not applicable. The device is a physical contact lens, not an algorithm.

    7. Type of Ground Truth Used:

    • The "ground truth" for this study was established through a combination of clinical measurements and observations by healthcare professionals. This included:
      • Visual acuities
      • Slit lamp findings (e.g., for corneal health, lens fit)
      • Subjective symptoms, problems, and complaints (patient-reported outcomes)
      • Adverse reactions

    8. Sample Size for the Training Set:

    • This is not applicable. The study describes a clinical trial for a physical medical device, not an AI model that requires a "training set."

    9. How Ground Truth for the Training Set Was Established:

    • This is not applicable as there is no AI training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K183670

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-09-13

    (260 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vexillum Zephyr (olifilcon A) with Tangible Polymers Silicone Hydrogel Soft Contact Lenses are indicated as daily wear for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

    Eye care practitioners may prescribe the lens for either single-use daily disposable wear, or for frequent/ planned replacement wear, with cleaning, disinfection, and scheduled replacement. When prescribed for frequent replacement, the lens may be disinfected using a chemical disinfection system only. The Vexillum Zephyr (olifilcon A) with Tangible Polymers Silicone Hydrogel Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

    The OxyPure A (olifilcon A) Silicone Hydrogel Soft Contact Lenses are indicated as daily wear for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

    Eye care practitioners may prescribe the single-use daily disposable wear, or for frequent/ planned replacement wear, with cleaning, disinfection, and scheduled replacement. When prescribed for frequent replacement, the lens may be disinfected using a chemical disinfection system only. The OxyPure A (olifilcon A) Silicone Hydrogel Soft (Hydrophilic) Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

    Device Description

    The Vexillum Zephyr (olifilcon A) with Tangible Polymers Silicone Hydrogel Soft Contact Lens is made of silicone hydrogel material, olifilcon A, with UV blocker available as spherical lens. The composition of the lens is 53% olifilcon A and 47% water. A light blue color tinted with "reactive Blue19" listed in 21 CFR Part 73.3121is for handling visibility purpose. A benzotriazole UV absorbing monomer is used to block UV radiation. The UV transmission (the thinnest lens measured by spectrophotometry as stated in ISO 18369) is less than 50% in the UVA range of 316 - 380 mm and less than 5% in the range of UVB range of 280-315 nm.

    Lenses are supplied sterile in sealed blister packs containing Tangible Polymers (coating on the lens surface during sterilization process) with isotonic buffered saline solution.

    The OxyPure A (olifilcon A) Silicone Hydrogel Soft Contact Lens is made of silicone hydrogel material, olifilcon A, with UV blocker available as spherical lens. The composition of the lens is 53% olifilcon A and 47% water. A light blue color tinted with "reactive Blue19" listed in 21 CFR Part 73.3121is for handling visibility purpose. A benzotriazole UV absorbing monomer is used to block UV radiation. The UV transmission (the thinnest lens measured by spectrophotometry as stated in ISO 18369) is less than 50% in the UVA range of 316 - 380 nm and less than 5% in the range of UVB range of 280-315 nm.

    OxyPure A (olifilcon A) Silicone Hydrogel Soft Contact Lenses are supplied sterile in sealed blister packs containing Sodium Hyaluronate and Sodium Alginate with isotonic buffered saline solution.

    AI/ML Overview

    This document is a 510(k) summary for contact lenses, which focuses on demonstrating substantial equivalence to a predicate device rather than proving performance against specific acceptance criteria for an AI device. Therefore, much of the requested information regarding AI device testing (e.g., sample sizes for test/training sets, ground truth establishment methods, expert qualifications, MRMC studies) is not applicable to this document.

    However, I can extract information related to the device attributes and the comparisons made to establish equivalence, which can be seen as meeting "acceptance criteria" for chemical/physical properties rather than AI performance.

    Here's a breakdown of what can be inferred and what is explicitly stated:

    Acceptance Criteria and Device Performance (Analogous for Contact Lenses)

    In the context of contact lenses, "acceptance criteria" are not thresholds for AI performance, but rather specifications and physical/chemical properties that demonstrate the new devices (Vexillum Zephyr and OxyPure A) are substantially equivalent to the predicate device (OxyPure Color). The table below summarizes these comparisons.

    CategoryAcceptance Criteria (Predicate Device K171447: OxyPure Color)Reported Device Performance (New Devices: Vexillum Zephyr & OxyPure A)Result of Comparison
    ApplicantVisco Vision IncVisco Vision IncSame
    ClassificationClass IIClass IISame
    Regulation Number886.5925886.5925Same
    Product CodeLPL, MVNLPL, MVNSame
    Intended UseMyopia, Hyperopia, astigmatism, PresbyopiaMyopia, Hyperopia, astigmatism, PresbyopiaSame
    Replacement ScheduleDaily WearDaily WearSame
    USAN Nameolifilcon Aolifilcon ASame
    FDA Category (Group)Group 5C (Nonionic, Water < 50 wt %)Group 5C (Nonionic, Water < 50 wt %)Same
    Manufacturing MethodCast MoldedCast MoldedSame
    Lens DesignSphericalSphericalSame
    Water Content47%47%Same
    Light Transmittance94%94%Same
    Refractive Index1.410 (hydrated)1.410 (hydrated)Same
    Oxygen Permeability (DK, 35°C)150 (Fatt method)150 (Fatt method)Same
    Diameter Range13.0 to 15.0 mm13.0 to 15.0 mmSame
    Power Range-20.00D to +20.00D-20.00D to +20.00DSame
    Center Thickness0.08mm @ -3.00D (Varies with Power)0.08mm @ -3.00D (Varies with Power)Same
    Base Curve8.0 mm to 9.2 mm8.0 mm to 9.2 mmSame
    Blue handling tintNoReactive Blue19 (for both new devices)Different
    Color Additives in DyeIron oxide, Titanium dioxide, Phthalocyanine green, Carbazole violetReactive Blue19 (for both new devices)Different
    Leachabilityno leachable monomers and addictive residuesno leachable monomers and addictive residuesSame
    PackagingBlister PackBlister PackSame
    Packaging SolutionSterile isotonic borate buffered salineVexillum Zephyr: sterile isotonic borate buffered saline with Tangible polymersOxyPure A: sterile isotonic borate buffered salineDifferent
    Coating on the lens surfaceNoVexillum Zephyr: Yes, coating with Tangible Polymer during sterilization processOxyPure A: NoDifferent
    Sterilization MethodSteamSteamSame
    Shelf Life5 years5 yearsSame
    Sterility of DeviceSAL= 10^-6SAL= 10^-6Same
    Tensile Strength (Mpa)0.78 ± 0.20.80 ± 0.2Different
    Modulus (Mpa)0.60 ± 0.10.60 ± 0.1Same
    Elongation at break (%)195 ± 20250 ± 50Different
    Toughness (J/m3)0.87 ± 0.051.10 ± 0.05Different
    Compliance Standard (Biocompatibility)ISO10993-1, ISO10993-5, ISO10993-10, ISO10993-11ISO10993-1, ISO10993-5, ISO10993-10, ISO10993-11Same
    Compliance Standard (Sterilization & Shelf Life)ISO 17665-1, ISO11737-1, ISO 11987ISO 17665-1, ISO11737-1, ISO 11987Same
    Compliance Standard (Performance)ISO18369-3, ISO18369-4, ASTM D792-13ISO18369-3, ISO18369-4, ASTM D792-13Same

    Study Proving Device Meets Criteria:

    The study proving the devices meet the "acceptance criteria" (i.e., are substantially equivalent) is based on non-clinical bench testing and comparison of technical characteristics to the predicate device.

    1. Sample Size and Data Provenance:

      • Test Set Sample Size: Not explicitly stated as a "test set" in the context of AI. The "test" consists of various physical, chemical, and mechanical tests on the contact lenses. The specific number of lenses tested for each parameter (e.g., tensile strength, oxygen permeability) is not provided in this summary but is implied to be sufficient for standard quality control and regulatory submission.
      • Data Provenance: The document does not specify the country of origin for the testing data, but the applicant's address is Taiwan. The studies are non-clinical bench tests, implying they are laboratory-based, not patient-based data. They are presumably prospective in the sense that they were specifically conducted for this submission.
    2. Number of Experts and Qualifications for Ground Truth:

      • Not Applicable. This document is not evaluating AI performance or subjective diagnostic interpretations. Ground truth, in this context, refers to the objective measurements of the lens properties themselves, assessed through established laboratory procedures according to recognized ISO and ASTM standards. No human expert "ground truth" establishment in the diagnostic sense is mentioned.
    3. Adjudication Method for Test Set:

      • Not Applicable. As there are no human-in-the-loop diagnostic interpretations requiring adjudication, this concept does not apply.
    4. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

      • No. The document explicitly states: "This 510(k) submission does not utilize clinical study for establishing substantial equivalence therefore this section does not apply." This confirms that no clinical studies, including MRMC studies, were conducted.
    5. Standalone Performance (Algorithm Only):

      • Not Applicable. This is not an AI device.
    6. Type of Ground Truth Used:

      • For the physical and chemical properties, the "ground truth" is established through objective laboratory measurements following recognized international standards (ISO, ASTM). Examples include:
        • Spectrophotometry for UV transmission.
        • Fatt method for oxygen permeability.
        • Standardized tests for tensile strength, modulus, elongation, and toughness.
        • Biocompatibility testing as per ISO 10993 series.
        • Sterility testing.
      • For the overall determination of substantial equivalence, the "ground truth" is the established safety and effectiveness of the predicate device (OxyPure Color Silicone Hydrogel Soft Contact Lenses, K171447). The new devices are compared to this established benchmark.
    7. Sample Size for Training Set & How Ground Truth for Training Set was Established:

      • Not Applicable. There is no training set as this is not an AI/machine learning device. The entire process relies on direct measurement and comparison to a known good (predicate) device, and adherence to established industry standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K181349

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-09-19

    (121 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vexillum (olifilcon B) with Tangible Polymers Spherical Silicone Hydrogel Soft Contact Lenses are daily wear single use soft contact lenses indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

    The Vexillum (olifilcon B) with Tangible Polymers Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear single use soft contact lenses for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters.

    The Vexillum (olifilcon B) with Tangible Polymers Multifocal Silicone Hydrogel Soft Contact lenses are indicated as daily wear single use soft contact lenses for the correction of ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.

    Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be discarded upon removal. Therefore, no cleaning or disinfecting is required.

    The Daily Breeze (olifilcon B) Spherical Silicone Hydrogel Soft Contact Lenses are daily wear single use soft contact lenses indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with nondiseased eyes who exhibit refractive astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.

    The Daily Breeze (olifilcon B) Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic persons with non-diseased eyes and whose powers are from -20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters.

    The Daily Breeze (olifilcon B) Multifocal Silicone Hydrogel Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) and emmetropia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.

    Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be discarded upon removal. Therefore, no cleaning or disinfecting is required.

    Device Description

    The Vexillum (olifilcon B) with Tangible Polymers Silicone Hydrogel Soft Contact Lens is made of silicone hydrogel material, olifilcon B, with UV blocker available as spherical lens, toric lens and multifocal lens. The composition of the lens is 53% olifilcon B and 47% water. A light blue color tinted with "reactive Blue19" listed in 21 CFR Part 73.3121is for handling visibility purpose. A benzotriazole UV absorbing monomer is used to block UV radiation. The UV transmission (the thinnest lens measured by spectrophotometry as stated in ISO 18369) is less than 50% in the UVA range of 316 - 380 nm and less than 5% in the range of UVB range of 280-315 nm. Lenses are supplied sterile in sealed blister packs containing Tangible Polymers (coating on the lens surface during sterilization process) with isotonic buffered saline solution.

    The Daily Breeze (olifilcon B) Silicone Hydrogel Soft Contact Lens is made of silicone hydrogel material, olifilcon B, with UV blocker available as spherical lens, toric lens and multifocal lens. The composition of the lens is 53% olifilcon B and 47% water. A light blue color tinted with "reactive Blue19" listed in 21 CFR Part 73.3121is for handling visibility purpose. A benzotriazole UV absorbing monomer is used to block UV radiation. The UV transmission (the thinnest lens measured by spectrophotometry as stated in ISO 18369) is less than 50% in the UVA range of 316 - 380 mm and less than 5% in the range of UVB range of 280-315 nm. Lenses are supplied sterile in sealed blister packs containing Sodium Hyaluronate and Sodium Alginate with isotonic buffered saline solution.

    AI/ML Overview

    The provided document discusses the 510(k) premarket notification for two soft contact lenses: Vexillum (olifilcon B) with Tangible Polymers and Daily Breeze (olifilcon B). It asserts substantial equivalence to a predicate device (Si-Hy (olifilcon B) Silicone Hydrogel Soft Contact Lenses, K160344).

    It is important to note that this document does not describe a study involving an AI/Machine Learning device. It pertains to contact lenses and their physical/chemical properties. Therefore, many of the requested criteria (such as sample size for test set, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable to this type of medical device submission.

    The "acceptance criteria" here relate to demonstrating substantial equivalence to a predicate device through non-clinical bench testing and comparisons of material, design, and performance characteristics.

    Here's an attempt to answer the applicable parts of your request based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present "acceptance criteria" in the format of a typical AI/ML study (e.g., target specificity and sensitivity values). Instead, it demonstrates "substantial equivalence" by comparing various parameters of the new devices (Vexillum and Daily Breeze) against a legally marketed predicate device (Si-Hy, K160344). The acceptance criterion, implicitly, is that the new devices are "substantially equivalent" to the predicate in terms of their intended use, technological characteristics, and performance characteristics, and that any differences do not raise new questions of safety or effectiveness.

    Here's a table summarizing the comparison parameters and their reported performance/characteristics, which effectively serve as the "acceptance criteria" for substantial equivalence in this context:

    CategoryAcceptance Criteria (Predicate: Si-Hy (olifilcon B) K160344)Reported Device Performance (Vexillum)Reported Device Performance (Daily Breeze)Result of Comparison
    ApplicantVisco Vision IncVisco Vision IncVisco Vision IncSame
    ClassificationClass IIClass IIClass IISame
    Regulation number886.5925886.5925886.5925Same
    Product codeLPL, MVNLPL, MVNLPL, MVNSame
    Intended useMyopia, Hyperopia, astigmatism, PresbyopiaMyopia, Hyperopia, astigmatism, PresbyopiaMyopia, Hyperopia, astigmatism, PresbyopiaSame
    Replacement ScheduleDaily Disposable (Single use)Daily Disposable (Single use)Daily Disposable (Single use)Same
    USAN Nameolifilcon Bolifilcon Bolifilcon BSame
    FDA Category (Group)Group 5C (Nonionic, Water < 50 wt %)Group 5C (Nonionic, Water < 50 wt %)Group 5C (Nonionic, Water < 50 wt %)Same
    Manufacturing MethodCast MoldedCast MoldedCast MoldedSame
    Lens DesignSpherical, toric or multifocalSpherical, toric or multifocalSpherical, toric or multifocalSame
    Water Content47%47%47%Same
    Light Transmittance94%94%94%Same
    Refractive Index1.410 (hydrated)1.410 (hydrated)1.410 (hydrated)Same
    Oxygen Permeability (DK, 35°C)120 (Fatt method)120 (Fatt method)120 (Fatt method)Same
    Diameter Range13.0 to 15.0 mm13.0 to 15.0 mm13.0 to 15.0 mmSame
    Power Range-20.00D ~ +20.00D-20.00D ~ +20.00D-20.00D ~ +20.00DSame
    Center Thickness0.08mm @ -3.00D (Varies with Power)0.08mm @ -3.00D (Varies with Power)0.08mm @ -3.00D (Varies with Power)Same
    Base Curve8.0 mm to 9.2 mm8.0 mm to 9.2 mm8.0 mm to 9.2 mmSame
    Blue handling tintReactive Blue19Reactive Blue19Reactive Blue19Same
    Leachabilityno leachable monomers and addictive residuesno leachable monomers and addictive residuesno leachable monomers and addictive residuesSame
    PackagingBlister PackBlister PackBlister PackSame
    Packaging Solutionsterile isotonic borate buffered salinesterile isotonic borate buffered saline with Tagible polymerssterile isotonic borate buffered salineDifferent (for Vexillum)
    Coating on the lens surfaceNoYes, coating with Tangible Polymer during the sterilization processNoDifferent (for Vexillum)
    Sterilization methodSteamSteamSteamSame
    Shelf Life5 years5 years5 yearsSame
    Sterility of DeviceSAL= 10^-6SAL= 10^-6SAL= 10^-6Same
    Tensile strength (Mpa)0.80±0.20.80±0.20.80±0.2Same
    Modulus (Mpa)0.50 ± 0.10.50 ± 0.10.50 ± 0.1Same
    Elongation at break (%)200± 20%200± 20%200± 20%Same
    Toughness (J/m3)0.90±0.050.90±0.050.90±0.05Same
    BiocompatibilityISO10993-1, -5, -10, -11ISO10993-1, -5, -10, -11ISO10993-1, -5, -10, -11Same
    Sterilization & Shelf lifeISO 17665-1, ISO11737-1, ISO 11987ISO 17665-1, ISO11737-1, ISO 11987ISO 17665-1, ISO11737-1, ISO 11987Same
    Performance (Measurement Methods)ISO18369-3, ISO18369-4, ASTM D792-13ISO18369-3, ISO18369-4, ASTM D792-13ISO18369-3, ISO18369-4, ASTM D792-13Same

    The crucial acceptance here is based on the conclusion that while differences exist in "Packaging Solution" and "Coating on the lens surface" for Vexillum, these differences "do not raise any new questions of safety or effectiveness."

    Study Proving Device Meets Acceptance Criteria:

    The study used to prove the device meets these "acceptance criteria" is a non-clinical bench testing and comparative study against a predicate device.

    1. Sample size used for the test set and the data provenance:

      • The document does not specify exact sample sizes for each bench test. However, it indicates that "bench tests were performed in accordance with following FDA Recognized Consensus Standards." These standards (e.g., ISO and ASTM) dictate the required sample sizes for specific tests.
      • Data Provenance: The company is Visco Vision Inc., located in Taoyuan City, TAIWAN. The studies would have been conducted by them or their contracted labs. This is a premarket submission, so it's prospective in the sense of demonstrating performance for a new device, but the tests themselves follow established methodologies.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable. This is a physical device (contact lens), not an AI/ML diagnostic system requiring expert interpretation of medical images. "Ground truth" is established through standardized laboratory measurement techniques (e.g., spectrophotometry for UV transmission, tensile strength testing, chemical analysis for leachables). The "experts" are the lab technicians and scientists performing these tests according to recognized standards.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not Applicable. As above, this is for physical/chemical testing, not for subjective interpretation of data requiring adjudication. Compliance with a standard means the results fall within defined tolerances.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not Applicable. This is not an AI/ML device, nor does it involve human readers interpreting data.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Not Applicable. This is not an AI/ML device. The "standalone" performance here refers to the physical and chemical properties of the contact lens itself as measured in a lab setting (e.g., water content, oxygen permeability).
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for the performance of these contact lenses is established through standardized physical, chemical, and biological test methods (bench tests) as defined by international standards (e.g., ISO, ASTM). These methods are designed to objectively measure properties like oxygen permeability, water content, tensile strength, sterility, and biocompatibility. The predicate device's performance also serves as a "ground truth" benchmark for comparison.
    7. The sample size for the training set:

      • Not Applicable. This isn't an AI/ML device, so there is no "training set." The development of the contact lens material (olifilcon B) and manufacturing processes would involve research and development, but not in the sense of "training data" for a machine learning model.
    8. How the ground truth for the training set was established:

      • Not Applicable. As there is no training set for an AI/ML model, this question does not apply.
    Ask a Question

    Ask a specific question about this device

    K Number
    K180322

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-04-20

    (74 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Si-Hy (olifilcon B) Color Spherical Silicone Hydrogel soft contact lenses are daily wear single use soft contact lens indicated for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with nondiseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

    The Si-Hy (olifileon B) Color Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic persons with non-diseased eyes and whose powers are from -20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters.

    The Si-Hy (olifilcon B) Color Multifocal Silicone Hydrogel Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) and emmetropia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.

    The lens may also be indicated to enhance or alter the apparent color of the eye. Eye care practitioners may prescribe the lens for daily wear (disposable use) single use. The lenses are to be discarded upon removal. Therefore, no cleaning or disinfecting is required.

    Device Description

    The Si-Hy (olifilcon B) Color Silicone Hydrogel Soft (Hydrophilic) Contact Lenses are avaialbe as spherical lens, toric lens and multifocal lens with cosmetic tint in an annular pattern by providing a clear optic zone. They are marked in different color by containing a combination of the following color additives: Iron oxide, [Phthalocyaninato (2-)] copper, Titanium dioxide, Phthalocyanine green, Carbazole violet. The lens material is composed by 53% olifilcon B and 47% water. A benzotriazole UV absorbing monomer is used to block UV radiation. The transmittance characteristics are 2.2% (< 5%) in the UVB range of 280-315nm and 8.7% ( < 50%) in the UVA range of 316-380nm. All colored additives used including ultraviolet absorbing additive (benzophenone based) are listed in 21CFR 73 subpart Dand 74 sbupart D.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device, specifically a soft contact lens. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device rather than proving clinical effectiveness through a new clinical trial with specific acceptance criteria.

    Therefore, the information requested in your prompt (acceptance criteria, study design for proving acceptance, sample sizes, expert involvement, MRMC studies, standalone performance, ground truth, and training set details) is not applicable in this context as this submission is for a contact lens which does not utilize AI. The FDA letter and the 510(k) summary explicitly state that clinical data is not required because the new device uses the same lens material as a previously cleared device and the same color additives as another previously cleared device.

    The review primarily focuses on:

    • Substantial Equivalence: Demonstrating that the new device has the same intended use, similar technological characteristics, and is as safe and effective as a legally marketed predicate device.
    • Nonclinical Testing: Performance of physicochemical studies (e.g., according to ISO 18369) and toxicology studies to ensure the material and lens properties meet established specifications and are biocompatible.

    Here's how to interpret the provided information in the context of your request, highlighting why many aspects are not present:

    1. A table of acceptance criteria and the reported device performance:

    • Acceptance Criteria: Not overtly stated as distinct "acceptance criteria" in the sense of a clinical trial endpoint. Instead, the "acceptance" is suitability for substantial equivalence, based on demonstrating similar properties and performance to the predicate devices.
      • Physicochemical Properties: The reported performance is that the "physical, optical and chemical properties of the lens are within established specifications for the lenses" (Section 1.9.1).
      • Toxicology: "toxicology studies report shows that the lenses are non-toxic and biocompatibility result is acceptable in ocular environment" (Section 1.9.2).
    • Device Performance: The comparison table (Section 1.8) shows various characteristics of the new device and the predicate devices. The "performance" is implicitly demonstrated by these characteristics being similar or equivalent to the predicates. For example:
      • Water Content: 47% (new device) vs. 47% (predicate K160344) and 42% (predicate K162317)
      • Oxygen Permeability (DK, 35°C): 120 (new device) vs. 120 (predicate K160344) and 10.89 (predicate K162317)
      • Light Transmittance: 94% (new device) vs. 94% (predicate K160344) and 97% (predicate K162317)

    2. Sample size used for the test set and the data provenance:

    • Not Applicable. No human clinical test set was used given the nature of the submission. The "test sets" would be the batches of lenses manufactured for physicochemical and toxicology testing. No information on the number of lenses tested is provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. No human-interpreted "ground truth" was established. The "ground truth" for the non-clinical tests is based on objective measurements and established scientific methods (e.g., ISO standards, toxicology protocols).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not Applicable. No human adjudication was involved.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This is not an AI device, and no MRMC study was conducted.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the non-clinical tests (physicochemical, toxicology), the "ground truth" is derived from objective laboratory measurements and adherence to established international standards (e.g., ISO 18369) and biocompatibility testing protocols.

    8. The sample size for the training set:

    • Not Applicable. There is no training set for an AI/ML model for this device.

    9. How the ground truth for the training set was established:

    • Not Applicable. There is no training set for an AI/ML model for this device.

    In summary, the provided document is a regulatory submission for a contact lens, not an AI/ML-powered medical device. Therefore, the detailed criteria for AI/ML model validation are not present and not relevant to this specific product's clearance. The "proof" of meeting "acceptance criteria" here is the demonstration of substantial equivalence to predicate devices through shared material, similar intended use, and satisfactory non-clinical laboratory testing.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173958

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-04-03

    (96 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OxyAqua (olifilcon D) Spherical Silicone Hydrogel soft (hydrophilic) contact lenses are daily wear single use soft contact lens indicated for the corrective ametropia (myopia and hyperopia) in phakic or aphakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

    The OxyAqua (olifilcon D) Toric Silicone Hydrogel Soft (hydrophilic) Contact Lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters.

    The OxyAqua (olifilcon D) Multifocal Silicone Hydrogel Soft (hydrophilic) Contact lenses are indicated as daily wear for the correction of ametropia (myopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.

    Eye care practitioners may prescribe the lens for daily wear disposable use (single use). The lenses are to be discarded upon removal. Therefore, no cleaning or disinfecting is required.

    Device Description

    The OxyAqua (olifilcon D) Silicone Hydrogel Soft Contact Lens is made of silicone hydrogel material with UV blocker available as spherical lens, toric lens and multifocal lens. The composition of the lens is 42% olifilcon D and 58 % water. A light blue color tinted with "reactive Blue19" is for handling visibility purpose. A benzotriazole UV absorbing monomer is used to block UV radiation. The transmittance characteristics are 2.4% (< 5%) in the UVB range of 280-315nm and 13.7% (< 50%) in the UVA range of 316-380nm. It is supplied in a sterile state packaged in a buffered saline solution.

    AI/ML Overview

    The document describes the premarket notification for a new soft contact lens, the "OxyAqua (olifilcon D) silicone hydrogel soft contact lens," seeking substantial equivalence to a predicate device. Here's an analysis of the acceptance criteria and supporting study:

    The provided document does not explicitly present a table of acceptance criteria with corresponding device performance metrics in a way that is typically found in a clinical trial report for software. However, the regulatory submission for the OxyAqua contact lenses implicitly relies on demonstrating that the device is "as safe, as effective, and performs as well as the predicate device."

    Here's an interpretation of the implied acceptance criteria and reported device performance based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implicit)Reported Device Performance (OxyAqua - Olifilcon D)
    Safety
    Biocompatibility (non-toxic in ocular environment)Toxicology studies report shows that the lenses are non-toxic and biocompatibility result is acceptable in ocular environment.
    Absence of new or increased adverse reactionsClinical study found no significant adverse reactions reported or Grade 3 or Grade 4 slit lamp findings. No significant differences in slit-lamp findings were found between groups (OxyAqua vs. Predicate).
    Acceptable Slit Lamp FindingsNo significant slit lamp findings (i.e. Grade 3 or Grade 4) were noted. No significant differences in slit-lamp findings were found between groups.
    Effectiveness/Performance
    Visual Acuity correction (for ametropia/presbyopia)Mean VA for both lenses (OxyAqua and Predicate) was similar at each visit. Visual acuity could be corrected to 0.1 (log MAR) or better in the clinical study.
    Lens Wearing Time (acceptable daily wear)Clinical study included lens wearing time as a parameter, and participants wore the lenses on a daily wear basis, demonstrating acceptable wearing time (though specific duration not explicitly stated as a result).
    Physiochemical properties within established specificationsPhysiochemical studies were conducted according to ISO 18369. The physical, optical, and chemical properties of the lens are within established specifications for the lenses.
    Substantial Equivalence to Predicate
    Overall similarity "as safe" and "as effective" as predicate"Comparison to the predicate device for chemical composition, physical and optical properties, it shows that 'OxyAqua silicone hydrogel soft contact lens' is as safe, as effective and performs as well as the predicate device."
    Matching Indications for UseThe indications for use for OxyAqua (Spherical, Toric, Multifocal) are clearly defined for correction of ametropia, astigmatism, and presbyopia in non-diseased eyes, for daily wear and single use, aligning with the type of conditions typically addressed by soft contact lenses and, by implication, those addressed by the predicate.
    Similar Technological Characteristics (despite material change)Despite material difference (Olifilcon D vs B, 58% vs 47% water content, Dk 85 vs 120), other characteristics like diameter, base curve, center thickness, powers, refractive index, light transmittance, handling tint, and manufacturing method are presented as similar or within acceptable ranges for substantial equivalence. The predicate device shared the same manufacturer and product code (MVN), implying a similar regulatory and safety profile as a daily disposable soft contact lens.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: "At least 30 evaluable subjects were participated."
    • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective/prospective). However, the description of "a three-month clinical study" strongly suggests it was a prospective study conducted to specifically support this submission. The manufacturer is based in Taiwan.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is generally not applicable for contact lens clinical trials in the same way it is for AI/computer vision devices. Ground truth for contact lenses is established through direct clinical observations and measurements by eye care professionals (e.g., ophthalmologists, optometrists) during the study visits. The document does not specify the number or qualifications of these clinical professionals, but it can be inferred that appropriately qualified clinicians would have performed the evaluations as part of a "clinical study."

    4. Adjudication Method for the Test Set

    Not specified. Clinical studies for contact lenses typically involve direct observation and measurement by qualified eye care practitioners. There is no mention of a separate adjudication process by experts, as common in AI device evaluations where multiple readers interpret images.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, an MRMC study was not done. MRMC studies are typically performed to evaluate the diagnostic performance of a new system (often AI-based) by comparing reader interpretations with and without the system's aid. This study compared two contact lenses (OxyAqua vs. an existing predicate) in a clinical trial setting.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable. The device is a physical medical device (contact lens), not an AI algorithm. Its performance is inherent to the lens itself when worn by a human.

    7. The Type of Ground Truth Used

    The ground truth was established through clinical observations and measurements (e.g., visual acuity, slit-lamp findings, adverse event reporting) by clinicians during a prospective clinical study directly on human subjects. This is considered direct clinical evidence.

    8. The Sample Size for the Training Set

    This question is not applicable. The device is a manufactured physical product (contact lens), not a machine learning model that requires a training set. The "study" refers to the clinical trial conducted for regulatory submission, not machine learning model development.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable for the same reason as above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171447

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2017-12-05

    (203 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The OxyPure Color Spherical Silicone Hydrogel Soft Contact Lenses are indicated as daily wear single use soft contact lens for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

    Eyecare practitioners may prescribe the lens for either single-use disposable wear, or for frequent/planned replacement wear, with cleaning, disinfection, and scheduled replacement. When prescribed for frequent replacement, the lens may be disinfected using a chemical disinfection system only.

    Device Description

    The OxyPure Color Silicone Hydrogel Soft Contact Lenses is a daily wear lens in spherical design. It may be prescribed as daily wear. It is a copolymer of 1-vinyl-2-pyrrolidinone (NVP) and Siloxane macromer. The water content is 47%. A benzotriazole UV absorbing monomer is used to block UV radiation. OxyPure Color Silicone Hydrogel Soft Contact Lenses contain a pigmented area that will alter or enhance the appearance of the natural iris. The lens is colored with one or more of the following color additives: iron oxides, titanium dioxide, phythalocyanine green and carbazole violet. Lenses are supplied sterile in sealed blister packs containing isotonic buffered saline solution. The compatibility and package integrity of the blister pack packaging system has been demonstrated and successfully used for other marketed lens products, and packaged lenses are effectively steam sterilized in a validated autoclave. Blister pack containers are labeled with the lens parameters, lot number and product expiration date. The expiration date has been established through stability studies that have assessed the chemical stability of the lens and package integrity (ability to maintain sterility). Shelf-life studies are ongoing to establish and extend the labeled expiration date.

    AI/ML Overview

    This document is a 510(k) premarket notification for the OxyPure Color Silicone Hydrogel Soft Contact Lenses. As such, it focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study with acceptance criteria for the new device. Therefore, much of the requested information regarding acceptance criteria, study sample sizes, ground truth establishment, expert involvement, and MRMC studies is not directly applicable or available in this specific document.

    Here's a breakdown of the available information related to acceptance criteria and performance:

    1. Table of Acceptance Criteria and Reported Device Performance

    Instead of clearly defined acceptance criteria and reported device performance for a new study, this document provides a comparison table between the new device (OxyPure Color Silicone Hydrogel Soft Contact Lenses) and its predicate device (VISCO SOFT CONTACT LENS K141348). The "Result of Comparison" column essentially indicates whether the new device meets the performance characteristics of the predicate (implied acceptance) or if there's a difference.

    CategoryOxyPure Color Silicone Hydrogel Soft Contact Lenses (New device)VISCO SOFT CONTACT LENS (K141348)Result of Comparison
    Physical/Material Properties
    USAN NameOlifilcon AOlifilcon ASame
    FDA Category (Group)Group 1 Non-ionic, Low waterGroup 1 Non-ionic, Low waterSame
    Water Content47%47%Same
    Light Transmittance>94%>94%Same
    Refractive Index1.410 (hydrated)1.410 (hydrated)Same
    Oxygen Permeability (DK, 35°C)150 (Fatt method)150 (Fatt method)Same
    Leachabilityno leachable monomers and addictive residuesno leachable monomers and addictive residuesSame
    Mechanical Properties
    Tensile strength (Mpa)0.78±0.20.5±0.2Different
    Modulus (Mpa)0.60 ± 0.10.60±0.1Same
    Elongation at break (%)195 ± 20%100± 20%Different
    Toughness (J/m3)0.87±0.050.25±0.05Different
    Design/Manufacturing
    Manufacturing MethodCast MoldedCast MoldedSame
    Tint processPad printingNo Pad PrintingDifferent
    Print ZoneCorresponds to the irisN/ADifferent
    Location of printingembedded between HEMA (Sandwich Process)embedded between HEMA (Sandwich Process)Same
    Lens DesignSphericalSphericalSame
    Diameter Range13 to 15 mm13 to 15 mmSame
    Power Range- 20.00D to +20.00D in 0.25 steps- 20.00D to +20.00D in 0.25 stepsSame
    Center Thickness0.08mm @ -3.00D (Varies with Power)0.08mm @ -3.00D (Varies with Power)Same
    Base Curve8.0 mm to 9.2 mm8.0 mm to 9.2 mmSame
    Blue handling tintNoYes, reactive Blue19Different
    Color Additives in dyeIron oxide, Titanium dioxide, Phthalocyanine green, Carbazole violetReactive Blue19Different
    Packaging/Sterilization
    Single UseNo, Frequent/ Planned ReplacementNo, Frequent/ Planned ReplacementSame
    PackagingBlister PackBlister PackSame
    Packaging Solutionsterile isotonic Borate buffered salinesterile isotonic phosphate buffered salineSame
    Sterilization methodSteamSteamSame
    Shelf Life5 years5 yearsSame
    Sterility of DeviceSAL= 10-6SAL= 10-6Same

    The document states: "The differences between those two devices are mechanical properties and color additives, but other reference devices have same technical characteristic as new devices to support the substantial equivalence." This implies that while the values might be numerically different for some mechanical properties and the color additives differ, these differences are not considered to raise new questions of safety or effectiveness because they align with established characteristics of other marketed devices. The acceptance criteria essentially appear to be "comparable to the predicate device or other legally marketed devices, with no new questions of safety or effectiveness raised by differences."

    2. Sample size used for the test set and the data provenance

    The document explicitly states: "This 510(k) submission does not utilize clinical study for establishing substantial equivalence therefore this section does not apply." This indicates that there was no specific "test set" of patients or data collected for a clinical study with this device. The data provenance would be from laboratory bench testing (non-clinical tests) and comparison to the predicate device's characteristics, not a clinical trial with patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable, as no clinical study with a test set requiring expert ground truth was conducted.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable, as no clinical study with a test set requiring adjudication was conducted.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a contact lens, not an AI-powered diagnostic tool, and no MRMC comparative effectiveness study was performed.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a contact lens, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical tests, the "ground truth" refers to established scientific and engineering standards and validated test methods. These include:

    • ISO18369-3: Ophthalmic optics - Contact lenses - Part 3: Measurement Methods
    • ISO18369-4: Ophthalmic optics Contact lenses Part 4: Physicochemical properties of contact lens materials
    • ISO18369-2: Ophthalmic optics - Contact lenses - Part 2: Tolerances
    • ISO11986: Ophthalmic optics -- Contact lenses and contact lens care products -- Guidelines for etermination of preservation uptake and release
    • ISO11987: Ophthalmic optics -- Contact lenses and contact lens care products --Determination of physical compatibility of contact lens care products with contact lenses
    • ISO17665-2: Sterilization Of Health Care Products - Moist Heat - Part 2: Guidance on the Application of ISO 17665-1. (Sterility)
    • ISO17665-1: Sterilization of health care products -- Moist heat -- Part 1: Requirements for the development, validation and routine control of a sterilization process for medical devices
    • ISO11980: Ophthalmic optics -- Contact lenses and contact lens care products -- Guidance for clinical investigations (though a clinical study was not used for this submission, compliance with this standard for potential future studies is implied).
    • ASTM D792-13: Standard Test Methods For Density And Specific Gravity (Relative Density) Of Plastics By Displacement. (Materials)
    • ISO10993-1, -5, -10, -11: For Biocompatibility.

    The "ground truth" for the device's properties is established through these standardized measurement methods, with the results compared to the known and acceptable properties of the predicate device and other legally marketed devices.

    8. The sample size for the training set

    Not applicable, as no AI/algorithm training set was used.

    9. How the ground truth for the training set was established

    Not applicable, as no AI/algorithm training set was used.

    Ask a Question

    Ask a specific question about this device

    K Number
    K160245

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2016-10-24

    (266 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VISCO (olifilcon A) Toric Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters.

    The VISCO (olifilcon A) Multifocal Soft Contact lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.

    Eye care practitioners may prescribe the lens for either single-use daily disposable wear, or for frequent/planned replacement wear, with cleaning, disinfection, and scheduled replacement. When prescribed for frequent replacement, the lens may be disinfected using a chemical disinfection system only.

    Device Description

    The VISCO (olifilcon A) toric/multifocal soft contact lenses are manufactured in toric or multifocal configurations with UV blocker. The lens material, olifilcon A, is a silicon hydrogel with water content 47%. The VISCO Soft Contact Lens is light blue tinted with "reactive Blue19" for handling visibility purpose. A benzotriazole UV absorbing monomer is used to block UV radiation. The transmittance characteristics are less than 5% in the UVB range of 280-315nm and less than 50% in the UVA range of 316-380mm. The lens is supplied in a sterile state, packaged in a buffered saline solution containing Sodium Hyaluronate and Sodium Alginate

    AI/ML Overview

    The provided document is a 510(k) Premarket Notification for a soft contact lens (VISCO (olifilcon A) Soft Contact Lens). It explicitly states that clinical studies were NOT required for this submission. Therefore, the document does not contain the information requested regarding acceptance criteria, study details, sample sizes, expert ground truth establishment, adjudication methods, MRMC studies, or standalone algorithm performance.

    The basis for the substantial equivalence claim relies on physicochemical studies and the equivalence of the technical characteristics, formulation, manufacturing, and sterilization processes to previously cleared predicate devices.

    Specifically, the document states:

    • "1.10 Clinical Studies: The technical characteristic, formulation, manufacturing, and sterilization processes of the subject device are equivalent to the VISCO (olifilcon A) soft contact lens currently available (K141348). Therefore, the clinical data is not required." (Page 9)
    • "1.9 Nonclinical Tests Performed: Physiochemical studies were conducted according to ISO 18369 First edition 2006-08-15, Ophthalmic optics - Contact lenses (Ophthalmic). The physical, optical and chemical properties of the lens are within established specifications for the lenses. Toxicology studies show that the lenses are non-toxic and biocompatibility result is acceptable in ocular environment." (Page 9)

    In summary, as this is a 510(k) submission for a contact lens cleared based on substantial equivalence to existing devices through non-clinical testing and manufacturing data, the detailed information about an AI/algorithm performance study (which would typically involve ground truth, expert adjudication, MRMC, etc.) is not applicable and not present in this document.

    Therefore, I cannot provide the requested table and study details.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Si-Hy Spherical Silicone Hydrogel Soft Contact Lenses are indicated as daily wear single use soft contact lens for the correction of refractive ametropia (myopia and hyperopia) in phakic persons with non-diseased eyes who exhibit refractive astigmatism of 1.00D or less where the astigmatism does not interfere with visual acuity.

    The Si-Hy Toric Silicone Hydrogel Soft Contact Lenses are indicated as daily wear for the correction of ametropia (myopia and hyperopia) with astigmatism in aphakic and non-aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters and astigmatic corrections are from -0.75 to -2.25 diopters.

    The Si-Hy Multifocal Silicone Hydrogel Soft Contact lenses are indicated as daily wear for the correction of anetropia (myopia and hyperopia) and emmetropia with presbyopia in aphakic persons with non-diseased eyes and whose powers are from -20.00 to +20.00 diopters with add powers from +0.75 to +2.75 diopters. The lenses may be worn by persons who exhibit astigmatism of 1.00 diopters or less where the astigmatism does not interfere with visual acuity.

    Eye care practitioners may prescribe the lens for daily wear (disposable use) single use. The lenses are to be discarded upon removal. Therefore, no cleaning or disinfecting is required.

    Device Description

    The Si-Hy (olifilcon B) Silicone Hydrogel Soft Contact Lens is made of silicone hydrogel material with UV blocker available as spherical lens, toric lens and multifocal lens. The composition of the lens is 53% olifilcon B and 47% water. A light blue color tinted with "reactive Blue19" is for handling visibility purpose. A benzotriazole UV absorbing monomer is used to block UV radiation. The transmittance characteristics are about 2.2% (<5%) in the UVB range of 280-315nm and 8.7% (<50%) in the UVA range of 316-380nm. It is supplied in a sterile state packaged in a buffered saline solution.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a soft contact lens device. It describes the device, its intended use, and comparative information with predicate devices, including nonclinical and clinical study summaries.

    Here's an analysis of the acceptance criteria and study details based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly list "acceptance criteria" in the format of a table with specific thresholds for device performance. Instead, it states that "the physical, optical and chemical properties of the lens are within established specifications for the lenses" and that "the clinical study provided data to establish substantial equivalence with the predicate, control lens." The reported performance is primarily compared to the predicate device to demonstrate substantial equivalence, rather than meeting predefined numerical acceptance criteria for specific metrics.

    However, we can infer some performance metrics from the device characteristics and the clinical study summary.

    Performance MetricAcceptance Criteria (Inferred from Predicate/Study Purpose)Reported Device Performance
    Physical/Optical/Chemical PropertiesWithin established specifications and comparable to predicate devices.Si-Hy (olifilcon B) lens characteristics are detailed (e.g., Water Content: 47%, Oxygen Permeability (DK): 120, Refractive Index: 1.410, Light Transmittance: 94%). The document states that these properties "are within established specifications for the lenses" and are "as safe, as effective and performs as well as the predicate device" in comparison.
    Material CompositionSubstantially equivalent to predicate lenses (Silicone Hydrogel).Olifilcon B (53% olifilcon B and 47% water) - Silicone Hydrogel. Compared to olifilcon A and narafilcon A.
    Biocompatibility/ToxicityNon-toxic and acceptable in ocular environment."Toxicology studies report shows that the lenses are non-toxic and biocompatibility result is acceptable in ocular environment."
    Visual Acuity (VA)Mean VA similar to predicate and correctable to 0.1 (log MAR) or better."The mean VA for both lenses [device and predicate] was similar at each visit. The visual acuity could be corrected to 0.1 (log MAR) or better."
    Slit Lamp FindingsNo significant findings (Grade 3 or Grade 4) and no significant differences between groups."No significant slit lamp findings (i.e. Grade 3 or Grade 4) were noted. No significant differences in slit-lamp findings were found between groups."
    Adverse ReactionsNot explicitly quantified as an acceptance criterion, but implies a safe profile comparable to the predicate.Measured, but specific rates are not provided in this summary. The overall conclusion is that the device is "as safe" as the predicate.
    Symptoms, Problems, ComplaintsNot explicitly quantified, but taken into account for safety and effectiveness.Measured, but specific details are not provided in this summary.
    Lens Wearing TimeNot explicitly quantified, but taken into account for safety and effectiveness.Measured, but specific details are not provided in this summary.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: "At least 50 evaluable subjects were participated."
    • Data Provenance: The document does not explicitly state the country of origin or whether the study was retrospective or prospective. However, clinical studies conducted for regulatory submission for a new device are almost always prospective. The manufacturer is based in Taiwan, so the study was likely conducted there or in a region convenient for the manufacturer.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document. The clinical study describes "parameters measured," implying a quantitative assessment by clinical investigators, but it does not specify the number or qualifications of experts involved in establishing a "ground truth" in the way one might for diagnostic accuracy studies (e.g., radiologist consensus). For contact lens clinical trials, "ground truth" is typically established by measurements and observations made by trained optometrists or ophthalmologists during eye examinations.

    4. Adjudication Method for the Test Set

    This information is not provided in the document. Adjudication methods (like 2+1 or 3+1 consensus) are common in studies where expert interpretation is subjective and needs consensus, such as pathology or imaging reads. For contact lens clinical evaluations, direct measurements and clinical findings by investigators typically serve as the primary data, and formal adjudication by multiple independent experts might not be a standard practice unless there are highly subjective or ambiguous outcomes.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study involving AI assistance was not done. This device is a contact lens, not an AI-powered diagnostic tool. The clinical study performed was a direct comparison of the new contact lens ("Si-Hy (olifilcon B)") against a legally marketed predicate contact lens ("Vistakon (narafilcon A) Contact Lens").

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm performance study was not done. As mentioned, this is a physical medical device (contact lens), not an algorithm.

    7. The Type of Ground Truth Used

    For the clinical study, the "ground truth" was established through:

    • Clinical Measurements and Observations: This includes objective measurements (e.g., visual acuities) and subjective observations/assessments by investigators (e.g., slit lamp findings, adverse reactions, symptoms, complaints).
    • Comparison to a Predicate Device: The study's primary goal was to demonstrate substantial equivalence to a legally marketed control contact lens for safety and effectiveness. The "ground truth" for what constitutes an acceptable contact lens performance is implicitly tied to that of the predicate device.

    8. The Sample Size for the Training Set

    This information is not applicable/not provided. The concept of a "training set" typically applies to machine learning models. For a physical medical device like a contact lens, there isn't a "training set" in the computational sense. The device underwent manufacturing and quality control processes, and then was tested in nonclinical and clinical studies.

    9. How the Ground Truth for the Training Set Was Established

    This question is not applicable as there is no "training set" for this device in the context of AI/machine learning. For the manufacturing and design of the contact lens, the "ground truth" for its properties and performance would have been established through extensive material science research, engineering specifications, and validated in vitro testing to meet performance standards.