Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K994324

    Validate with FDA (Live)

    Date Cleared
    2000-02-23

    (63 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACUVUE Contact Lens (spherical) is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or non-aphakic persons with non-diseased eyes who may have 1.00 D or less of astigmatism.

    The ACUVUE BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic, aphakic or non-aphakic persons with non-diseased eyes who may have 0.75 D or less of astigmatism.

    The ACUVUE TORIC Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic or non-aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D or less of astigmatism.

    The ACUVUE TORIC BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic aphakic or non-phakic persons with non-diseased eyes who may have 10.00 D of astigmatism or less.

    ACUVUE Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

    Eye care practitioners may prescribe the lens for either single-use disposable wear or for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system only.

    Device Description

    The device description does not change from that cleared under K991134.

    AI/ML Overview

    The provided 510(k) submission (K994324) for the ACUVUE (etafilcon A) soft contact lenses is a labeling modification submission, not a submission for a new device requiring a comprehensive performance study against acceptance criteria.

    The purpose of this 510(k) is to revise the wear schedule in the labeling to include a statement about the reduced risk of giant papillary conjunctivitis (GPC) with frequent replacement.

    Therefore, many of the typical elements of a performance study required for novel device clearance are not present or applicable in this document.

    Here's an analysis based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    This 510(k) submission does not present a table of acceptance criteria or new device performance data in the way a new device would. The core claim is a labeling modification supported by existing clinical literature, not new performance metrics for the device itself.

    The "acceptance criteria" in this context would be the FDA's acceptance of the scientific literature supporting the labeling change.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set (for the labeling claim): The submission states that the labeling modification is supported by "clinical studies published in the July 1999 CLAO Journal (see Attachment A)." Attachment A is not provided in the document. Therefore, the sample size and data provenance of these specific studies are not detailed in this 510(k) submission.
    • Data Provenance: The document does not specify the country of origin of the data mentioned in the CLAO Journal. It states the studies were "published," implying they are retrospective in terms of this 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the 510(k) submission. As the evidence relies on published clinical studies, the details about expert involvement in establishing ground truth would typically be found within those external publications, not in this regulatory submission.

    4. Adjudication Method for the Test Set

    This information is not provided in the 510(k) submission for the same reasons as above.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done, nor would it be relevant for a contact lens labeling modification based on GPC risk. This type of study is more common for diagnostic imaging or AI-assisted diagnostic devices.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    N/A. This device is a contact lens; there is no algorithm or human-in-the-loop performance component.

    7. The Type of Ground Truth Used

    The "ground truth" for the labeling modification regarding reduced GPC risk would be derived from the clinical outcomes data reported in the published studies cited (July 1999 CLAO Journal). This would include observations and diagnoses of GPC in patients using contact lenses with different replacement schedules.

    8. The Sample Size for the Training Set

    N/A. This device is a contact lens. There is no training set in the context of machine learning or algorithms. The "training data" for general contact lens safety and efficacy would have been the extensive clinical trials conducted for the original device clearance (K991134).

    9. How the Ground Truth for the Training Set Was Established

    N/A. See point 8. The ground truth for the original ACUVUE lens (K991134) would have been established through standard clinical trial methodologies, including patient examinations, objective measurements, and subjective patient feedback, reviewed and interpreted by qualified clinicians, to demonstrate the lens's safety and effectiveness for its intended use.

    Ask a Question

    Ask a specific question about this device

    K Number
    K994318

    Validate with FDA (Live)

    Date Cleared
    2000-02-23

    (63 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VISTAKON Contact Lens (spherical) is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or non-aphakic persons with non-diseased eves who may have 1.00 D or less of astigmatism.

    The VISTAKON BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic, aphakic or non-aphakic persons with nondiseased eyes who may have 0.75 D or less of astigmatism.

    The VISTAKON TORIC Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic or non-aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D or less of astigmatism.

    The VISTAKON TORIC BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic aphakic or non-phakic persons with non-diseased eyes who may have 10.00 D of astigmatism or less.

    VISTAKON Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

    Eye care practitioners may prescribe the lens for either single-use disposable wear or for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system only.

    Device Description

    The device description does not change from that cleared under K991163.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the VISTAKON contact lenses:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly state specific quantitative acceptance criteria or detailed device performance metrics in the way typically seen for new medical device clearances. This 510(k) submission is for a labeling modification to an already cleared device (K991163), not for a new device or a significant change in its design or performance.

    The "acceptance criteria" here are implicitly related to demonstrating that the revised labeling (specifically, the claim about reduced risk of giant papillary conjunctivitis) is supported by existing clinical evidence.

    However, based on the information, we can infer the "performance" goal as demonstrating the established safety and efficacy for the original device, as the current submission relies on that prior clearance.

    Acceptance Criterion (Inferred)Reported Device Performance
    Safety and Efficacy for intended use (refractive error correction, UV protection)Established by prior clearance K991163 (details not provided in this document)
    Support for revised labeling claim regarding reduced risk of giant papillary conjunctivitis (GPC) with specific wear schedulesSupported by clinical studies published in the July 1999 CLAO Journal. (Specific findings not detailed here, but the FDA's clearance implies the claim was adequately supported).

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated in this document. The submission references "clinical studies published in the July 1999 CLAO Journal" (Attachment A) as the basis for the labeling modification. The sample size would be found within those external publications, but is not provided here.
    • Data Provenance: The document does not specify the country of origin of the data. It refers to "clinical studies published in the July 1999 CLAO Journal," which could be from various locations. The studies are assumed to be prospective given they are clinical studies, likely trials, investigating the impact of wear schedules.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the submission. The "ground truth" for the claims about GPC risk reduction would have been established by the methods described in the referenced clinical studies (e.g., clinical examination, diagnosis by ophthalmologists/optometrists). The number and qualifications of the experts involved in those studies are not detailed in this 510(k).

    4. Adjudication Method for the Test Set

    This information is not provided. As the submission refers to external clinical studies, the adjudication methods (if any, for outcomes like GPC diagnosis) would be detailed within those publications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done for this submission. MRMC studies are typically used to evaluate the impact of AI-assisted diagnostic devices on human reader performance. This submission concerns a contact lens, not a diagnostic AI system.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No, a standalone performance evaluation was not done. This is not an AI/algorithm-based device. The "performance" being evaluated is related to the clinical outcomes associated with the wear schedule of the contact lens.

    7. The Type of Ground Truth Used

    The ground truth used for the claims about the reduction of giant papillary conjunctivitis (GPC) would have been clinical diagnosis by eye care professionals. This would involve direct observation and examination of the patient's eyes for signs of GPC as per established diagnostic criteria. It is a form of "expert consensus" or "clinical diagnosis" based on observable medical conditions.

    8. The Sample Size for the Training Set

    This information is not applicable as this is not an AI/machine learning device that requires a training set. The clinical studies supporting the claims would have involved a patient cohort, but it's not a "training set" in the context of an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable as there is no training set for an AI/ML algorithm. For the clinical studies, the "ground truth" for GPC diagnosis would have been established by the clinical assessments and diagnoses of the investigators or study clinicians.

    Ask a Question

    Ask a specific question about this device

    K Number
    K991163

    Validate with FDA (Live)

    Date Cleared
    1999-06-08

    (62 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VISTAKON Contact Lens (spherical) is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or non-aphakic persons with non-diseased eyes who may have 1.00 D or less of astigmatism.

    The VISTAKON BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic, aphakic or non-aphakic persons with non-diseased eyes who may have 0.75 D or less of astigmatism.

    The VISTAKON TORIC Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic or non-aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D or less of astigmatism.

    The VISTAKON TORIC BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic aphakic or non-phakic persons with non-diseased eyes who may have 10.00D of astigmatism or less.

    VISTAKON UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

    Eye care practitioners may prescribe the lens for either single-use disposable wear or for frequent/ planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system only.

    Device Description

    The device description does not change from that cleared under K983912.

    AI/ML Overview

    The provided 510(k) submission (K991163) for the VISTAKON (lenefilcon A) soft (hydrophilic) contact lenses does not describe a study involving acceptance criteria and device performance in the typical sense of a new medical device efficacy study. Instead, this submission is a labeling modification for an already cleared device (K983912).

    Therefore, the following information is largely not applicable or directly extractable from the provided text, as the focus is on the equivalence of the device and its existing data to support new indications, rather than a de novo study demonstrating meeting new acceptance criteria.

    Here's an analysis based on the provided document:


    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not present a table of acceptance criteria or new device performance data. The core of this 510(k) is that the device material and manufacturing process are unchanged from a previously cleared device (K983912). The only new "performance" mentioned relates to UV blocking, which was also established in a prior submission or existing data.

    • UV Blocking:
      • UVA: 95%
      • UVB: 99.9%
      • Acceptance Criteria for UV Blocking: Not explicitly stated in this document, but implied to be met based on the reported values. Generally, contact lenses classified as UV-blocking must meet certain FDA-defined percentages for UVA and UVB absorption.

    2. Sample Sizes Used for the Test Set and Data Provenance

    • Test Set Description: Not applicable, as no new clinical or non-clinical studies were conducted specifically for this 510(k) to establish performance against new acceptance criteria.
    • Data Provenance: The submission explicitly states: "There is no change in lens material, the manufacturing process, nor the parameters and properties, therefore, the clinical data previously submitted in K983912 supports the clinical safety of the subject device." This means any performance data supporting the indications would stem from studies related to K983912. The provenance for those original studies is not detailed in this document.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not applicable, as no new test set was created for a study mentioned in this 510(k). The clinical safety is supported by prior data.

    4. Adjudication Method for the Test Set

    • Not applicable, as no new test set was created for a study mentioned in this 510(k).

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC study was done. This device is a contact lens, and the 510(k) is for an expanded indication statement, not for an AI diagnostic or interpretive device. Therefore, a human reader improvement effect size is not relevant.

    6. Standalone Performance Study (Algorithm Only)

    • No standalone performance study was done. This device is a contact lens. The concept of an "algorithm only" performance is not applicable.

    7. Type of Ground Truth Used

    • For the original submission (K983912) that established the safety and efficacy of the lens material, the "ground truth" would have likely been based on clinical outcomes data from human subjects wearing the contact lenses (e.g., visual acuity, comfort, adverse events, fit characteristics, etc.) as evaluated by optometrists or ophthalmologists.
    • For the UV blocking claim, the ground truth would be based on laboratory measurements of UV transmission through the lens material.

    8. Sample Size for the Training Set

    • Not applicable. This 510(k) is not for a machine learning or AI device that requires a training set. The clinical data from K983912 would have involved a cohort of human subjects. The details of that sample size are not in this document.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable, as there is no "training set" in the context of this 510(k). For the original clinical data from K983912, the ground truth would have been established by clinical investigators (eye care professionals) assessing patient outcomes and device performance against defined endpoints in clinical trials.
    Ask a Question

    Ask a specific question about this device

    K Number
    K991134

    Validate with FDA (Live)

    Date Cleared
    1999-05-06

    (34 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACUVUE Contact Lens (spherical) is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or nonaphakic persons with non-diseased eyes who may have 1.00 D or less of astigmatism.

    The ACUVUE BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic, aphakic or non-aphakic persons with non-diseased eyes who may have 0.75 D or less of astigmatism.

    The ACUVUE TORIC Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic or non-aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D or less of astigmatism.

    The ACUVUE TORIC BIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic aphakic or non-phakic persons with non-diseased eyes who may have 10.00D of astigmatism or less.

    ACUVUE UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.

    Eye care practitioners may prescribe the lens for either single-use disposable wear or for frequent/ planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system only.

    Device Description

    The device description does not change from that cleared under K962804.

    AI/ML Overview

    The provided text describes a 510(k) submission for ACUVUE (etafilcon A) soft (hydrophilic) contact lenses. This submission is for a labeling modification to add an additional Indications Statement. It is not an AI/ML device, and therefore, many of the requested criteria regarding AI/ML device performance and studies are not applicable.

    Here's an analysis of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not provide specific acceptance criteria or reported device performance for an AI/ML device. Instead, it refers to the existing clearance of the same lens material (etafilcon A) and its associated performance as sufficient due to the nature of the 510(k) being a labeling modification.

    Acceptance Criteria (Not explicitly stated for AI/ML)Reported Device Performance (Historical for etafilcon A contact lenses)
    UV Blocking (UVA)82%
    UV Blocking (UVB)97%
    Clinical SafetySupported by previously submitted data (N18-033/S31 and K962804)
    Non-Clinical SafetySupported by previously submitted data (N18-033/S31 and K962804)

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable as this is not an AI/ML device requiring a test set for algorithm performance evaluation. The submission relies on existing clinical and non-clinical data for the lens material itself from previous submissions (N18-033/S31 and K962804). The provenance of that historical data is not detailed in this document.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable as this is not an AI/ML device using a test set with ground truth established by experts.

    4. Adjudication Method for the Test Set

    Not applicable as this is not an AI/ML device using a test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study was not done. This is not an AI/ML device, so such a study to evaluate human reader improvement with AI assistance is not relevant.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    No, a standalone performance study was not done. This is not an AI/ML device.

    7. The Type of Ground Truth Used

    The concept of "ground truth" as typically applied to AI/ML device performance (e.g., expert consensus, pathology, outcomes data) is not directly applicable here. The demonstration of safety and effectiveness for a contact lens material (etafilcon A) generally relies on:

    • Clinical safety and efficacy data: Data from human trials assessing visual acuity, comfort, ocular health, and adverse events.
    • Non-clinical data: Laboratory testing for material properties, microbiology, toxicology, chemistry, shelf-life, and leachability.

    The submission states that this data was sufficient for the predicate devices and supports the current device due to no changes in material or manufacturing process.

    8. The Sample Size for the Training Set

    Not applicable as this is not an AI/ML device requiring a training set.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable as this is not an AI/ML device requiring a training set with established ground truth.

    Conclusion:

    This 510(k) submission is for a labeling modification for an existing contact lens product. The "device" in question is a contact lens, not an AI/ML diagnostic or therapeutic system. Therefore, most of the requested information pertaining to AI/ML device acceptance criteria, studies, and ground truth establishment is not relevant to this specific document. The submission relies on the previously established safety and effectiveness of the etafilcon A material clear in prior 510(k)s and PMAs.

    Ask a Question

    Ask a specific question about this device

    K Number
    K983912

    Validate with FDA (Live)

    Date Cleared
    1999-01-26

    (84 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    1 - 5
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VISTAKON Contact Lens is indicated for daily wear for the correction of refractive ametropia (myopia and hyperopia) in aphakic or non-aphakic persons with non-diseased eyes who may have 1.00 D of astigmatism or less.
    The VISTAKON MULTIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic aphakic or non-aphakic persons with non-diseased eyes who may have 0.75 D of astigmatism or less.
    The VISTAKON TORIC Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic or non-aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D of astigmatism or less.
    The VISTAKON TORIC MULTIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic aphakic or non-aphakic persons with non-aphakic persons with non-diseased eyes who may have 10.00 D of astigmatism or less.
    Eye care practitioners may prescribe the lens either for single-use disposable wear or frequent/planned replacement wear with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system only.

    Device Description

    The VISTAKON (lenefilcon A) Soft (hydrophilic) Contact Lens is available as a spherical lens, a multifocal lens, a toric lens and a toric multifocal lens.
    The lens material (lenefilcon A) is a copolymer of 2-hydroxyethyl 2-methyl-2 propenoate and 2-propenic acid, 2-methyl-2,3-dihydroxypropyl ester cross-linked with poly (oxy-1,2-ethanediyl) (4) bis (2-methyl-2-propenoate). The VISTAKON Contact Lens Visibility Tint with UV Blocker is tinted blue using Reactive Blue Dye #4 to make the lens more visible for handling. A benzotriazole UV-absorbing monomer is used to block UV radiation. The transmittance characteristics are less than 1% in the UVB range of 280nm to 315nm and less than 10% in the UVA range of 316nm to 380nm. The VISTAKON Contact Lens is a hemispherical or hemitoric shell.

    AI/ML Overview

    The VISTAKON (lenefilcon A) Soft Contact Lens is a contact lens with various designs (spherical, multifocal, toric, and toric multifocal) intended for daily wear to correct refractive ametropia. The study for this device did not involve an AI component, nor did it involve the evaluation of expert performance with or without AI assistance. Instead, it was a clinical and non-clinical study to demonstrate substantial equivalence to a predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance

    ParameterAcceptance Criteria / Predicate Device Performance (NewVues® (vifilcon A))Reported Device Performance (VISTAKON (lenefilcon A))
    Material Properties:
    Water Content, %55 (labeled) / 56 (measured)55 (labeled) / 55 (measured)
    Refractive Index @ 20°C1.415 (labeled) / #1.41 (calculated)1.41 (labeled) / 1.41 (measured)
    Dk, edge corrected16 (labeled) / 18 (measured)16 (labeled) / 17 (measured)
    Dk, non-edge corrected16 (labeled) / 22 (measured)21 (labeled) / 22 (measured)
    Color (if tinted)BlueBlue
    Physical Dimensions:
    Base Curve, mm8.5 (labeled) / 8.5 (measured)8.84 (labeled) / 8.86 (measured)
    Diameter, mm14.0 (labeled) / 14.3 (measured)14.05 (labeled) / 14.10 (measured)
    Power, D-1.00 (labeled) / -1.02 (measured)-0.74 (labeled) / -0.79 (measured)
    Toxicology:No evidence of toxicity/irritation and meeting ISO requirementsPassed Cytotoxicity, Ocular Irritation (Rabbit), Acute Systemic Toxicity (Mouse)
    Microbiology:Minimum SAL of 10-6 for sterilization; demonstrated disinfection efficacy with care productsSterilization process validated to minimum SAL of 10-6; shelf-life stability data support sterility
    Clinical Equivalence:Substantial clinical equivalence to predicate deviceFound to be at least substantially clinically equivalent to the predicate device
    UV Blocker Transmittance:N/A (implicit that it blocks UV)Less than 1% in UVB (280-315nm), less than 10% in UVA (316-380nm)

    Notes on Acceptance Criteria:

    • The acceptance criteria for the material properties and physical dimensions are implied by demonstrating substantial equivalence to the predicate device. The performance data for the VISTAKON lens is compared directly to the predicate device's data.
    • For non-clinical studies (toxicology, chemistry, microbiology), the acceptance criteria are based on established standards and previous demonstrations of compatibility (e.g., ISO studies, validated sterilization processes).

    2. Sample Size for Test Set and Data Provenance

    • Sample Size for Test Set: At least 50 patients participated in the clinical trial. The ratio of subject device to predicate device was 2:1.
    • Data Provenance: Not explicitly stated, but clinical trials for US FDA submissions are typically conducted in the US. The study evaluated "Daily Wear contact lens materials with a new USAN name according to the Premarket Notification (510(k)) Guidance Document for Daily Wear Contact Lenses (May 1994)," which implies US-centric regulatory guidelines. The study was prospective as it involved a clinical trial.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The submission does not specify the number or qualifications of experts (e.g., ophthalmologists, optometrists) involved in establishing the ground truth or evaluating patient outcomes in the clinical trial. However, clinical trials for contact lenses typically involve qualified eye care professionals for examinations and assessments.

    4. Adjudication Method for the Test Set

    The submission does not explicitly describe an adjudication method for the clinical study's evaluations (e.g., 2+1, 3+1). Clinical trials generally involve standardized protocols for examinations and reporting adverse events, which implicitly serve as a form of "ground truth" establishment by the clinical investigators.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No such study was performed or is applicable to this device. This submission is for a medical device (contact lens), not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a contact lens and does not involve an algorithm or AI.

    7. The Type of Ground Truth Used

    For the clinical study:

    • The "ground truth" was established through direct clinical observation and measurement of various parameters in patients wearing the contact lenses. These parameters included:
      • Adverse reactions
      • Symptoms
      • Problems and complaints reported by patients
      • Slit lamp evaluations (objective assessment of ocular health)
      • Visual acuity (objective measurement of vision correction)
      • Lens wear time
      • Fit assessment
      • Discontinuations from the study
    • For non-clinical studies (chemistry, toxicology, microbiology), the ground truth was based on established scientific protocols, validated methods, and regulatory standards (e.g., ISO studies, validated sterilization processes).

    8. The Sample Size for the Training Set

    Not applicable. This device is a contact lens, not an AI model requiring a training set. The "training" for the device would be its manufacturing process and design, refined through R&D.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI model. The development of the VISTAKON contact lens material and design would have involved extensive R&D, material characterization, and design optimization based on scientific principles and previous knowledge of contact lens technology.

    Ask a Question

    Ask a specific question about this device

    K Number
    K964612

    Validate with FDA (Live)

    Date Cleared
    1997-01-30

    (73 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VISTAKON Contact Lens (spherical) is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and presbyopia) in aphakic or non-aphakic persons with non-diseased eyes who may have 1.00 D of astigmatism or less.
    The VISTAKON MULTIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic aphakic or non-aphakic persons with non-diseased eyes who may have 0.75 D of astigmatism or less.
    The VISTAKON TORIC Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic or non-aphakic persons with non-diseased eyes that are hyperopic or myopic and may have 10.00 D of astigmatism or less.
    Eye care practitioners may prescribe the lens for single-use disposable wear or for frequent/planned replacement wear, with cleaning disinfection and scheduled replacement (see WEARING SCHEDULE). When prescribed for frequent/planned replacement, the lens may be disinfected using a chemical disinfection system only.

    Device Description

    The VISTAKON (genfilcon A) Soft (hydrophilic) Contact Lens is available as a spherical lens, a spherical multifocal lens and an astigmatic (toric) lens. The lens material (genfilcon A) is a copolymer of 2-hydroxyethyl methacrylate and methacrylic acid cross-linked with tetraethylene glycol dimethacrylate and ethylene glycol dimethacrylate. The VISTAKON Contact Lens with visibility tint is tinted blue using Reactive Blue Dye #4 to make the lens more visible for handling. The VISTAKON Contact Lens is a hemispherical shell.

    AI/ML Overview

    This document describes the Vistakon (genfilcon A) Contact Lens. The provided text, however, focuses on showing substantial equivalence to a predicate device and does not define explicit acceptance criteria or provide a detailed study report with performance metrics against those criteria. Instead, it outlines the non-clinical and clinical studies conducted to demonstrate safety and effectiveness.

    Here's an analysis of the provided information based on the request, with the understanding that specific numerical "acceptance criteria" against which a device's performance is measured are not explicitly stated in this 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    As mentioned, explicit, quantitative "acceptance criteria" are not provided in the typical sense (e.g., "sensitivity must be >X%"). Instead, the studies aim to demonstrate that the device is safe and effective for its intended use and substantially clinically equivalent to the predicate device. The performance is implicitly assessed by showing that it does not pose new or greater risks, and offers comparable benefits to the predicate.

    Criteria/CategoryAcceptance Criteria (Implicit)Reported Device Performance
    Material Properties (Chemistry)Properties consistent with FDA Group I lenses; compatible with approved lens care products; acceptable shelf-life stability; no significant difference in leachable materials compared to predicate.Material property data reflected properties of Group I lenses. Lens care product manufacturers previously showed compatibility with Group I lenses. Shelf-life stability based on approved protocols (initial claim after 6 months real-time data). No significant differences in leachable materials found between subject and predicate device.
    ToxicologyExtracts not considered ocular irritants; extracts not considered systemically toxic; lens not cytotoxic.Ocular Irritation study indicated extracts would not be considered ocular irritants. USP Systemic Toxicity study indicated extracts would not be considered systemically toxic. Cytotoxicity study indicated the lens is not cytotoxic.
    Microbiology/SterilizationSterilization process validated to deliver a minimum SAL of 10^-6; compatible with approved disinfection efficacy of lens care products; sterility maintained throughout shelf-life.Moist heat sterilization validated to deliver a minimum SAL of 10^-6. Lens care product manufacturers established disinfection efficacy in lens groups. Shelf-life stability data supports lens sterility.
    Clinical Safety & Effectiveness (Daily Wear)Safe and effective for intended use; substantially clinically equivalent to the predicate device (Optima FW). No new or increased adverse reactions, symptoms, problems, or complaints; comparable slit lamp evaluations, visual acuity, lens wear time, and fit assessment; comparable discontinuation rates.The VISTAKON (genfilcon A) trial contact lens was found to be safe and effective for its intended use and at least substantially clinically equivalent to the predicate device. Parameters measured included adverse reactions, symptoms, problems and complaints, slit lamp evaluations, visual acuity, lens wear time, fit assessment, and discontinuations. (Specific comparative metrics not provided in this summary, but the conclusion states equivalence).
    Overall Legal Standing (510(k) Purpose)Safety and effectiveness supported; substantially equivalent to the predicate device."The data presented in this Premarket Notification support the safety and effectiveness of the subject device... The subject device has been shown to be substantially equivalent to the predicate device."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: "at least 50 patients with a 2:1 ratio of subject device to predicate device".
      • This means 33 patients for the subject device (VISTAKON) and 17 patients for the predicate device (Optima FW), totaling 50 patients.
    • Data Provenance: Not explicitly stated regarding country of origin. The study was a "clinical trial for daily wear contact lens materials with a new USAN name according to the Premarket Notification (510(k)) Guidance Document for Daily Wear Contact Lenses (May 1994)," which implies it was conducted to meet US regulatory requirements. It was a prospective clinical trial.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • The document does not describe the establishment of a "ground truth" in the way it would for an AI-powered diagnostic device (e.g., expert consensus on image labels). Instead, patient outcomes and clinical assessments were the primary endpoints.
    • The clinical trial involved "eye care practitioners" who evaluated parameters such as slit lamp evaluations, visual acuity, and fit assessment. Their specific qualifications (e.g., "ophthalmologist with X years of experience") are not detailed in this summary.

    4. Adjudication Method for the Test Set

    • The document does not mention an adjudication method (such as 2+1 or 3+1 consensus) for the clinical trial data. Clinical trials typically rely on standardized protocols and trained clinical investigators to collect and assess data, not necessarily an "adjudication" panel for ground truth in the AI sense.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    • No, an MRMC comparative effectiveness study, as typically described for AI systems, was not performed. This product is a contact lens, not an AI diagnostic system. The clinical study compares the subject device directly to a predicate device in human patients, assessing endpoints related to safety and performance. There is no concept of "human readers improving with AI vs without AI assistance" in this context.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • This is not applicable as the submission is for a physical medical device (contact lens), not an algorithm or AI system.

    7. The Type of Ground Truth Used

    • The "ground truth"
      • For non-clinical studies: Laboratory measurements, chemical analyses, toxicology tests, and microbiological validation following established scientific protocols and GLP regulations.
      • For clinical studies: Patient-reported outcomes (symptoms, complaints), objective clinical measurements (slit lamp evaluations, visual acuity), and assessments by qualified eye care practitioners (fit assessment, adverse reactions). The overall "ground truth" for the clinical trial was the observed safety and effectiveness in patients compared to the predicate device.

    8. The Sample Size for the Training Set

    • This is not applicable, as this is a contact lens submission and does not involve AI or machine learning models that require a "training set."

    9. How the Ground Truth for the Training Set Was Established

    • This is not applicable, as there is no training set for an AI model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K962804

    Validate with FDA (Live)

    Date Cleared
    1996-10-01

    (75 days)

    Product Code
    Regulation Number
    886.5925
    Age Range
    All
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ACUVUE Contact Lens (spherical) is indicated for daily wear for the correction of refractive ametropia (myopia, hyperopia and presbyopia) in aphakic or non-aphakic persons with non-diseased eyes who may have 1.00 D of astigmatism or less.

    The ACUVUE MULTIFOCAL Contact Lens is indicated for daily wear for the correction of distance and near vision in presbyopic, aphakic or nonaphakic persons with non-diseased eyes who may have 0.75 D of astigmatism or less.

    The ACUVUE Toric Contact Lens is indicated for daily wear for the correction of visual acuity in aphakic or non-aphakic persons with nondiseased eyes that are hyperopic or myopic and may have 10.00 D of astigmatism or less.

    Eye care practitioners may prescribe the lens for either single-use disposable wear or frequent/ planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfection system only.

    Device Description

    The ACUVUE (etafilcon A) Soft (hydrophilic) Contact Lens is available as a spherical lens, spherical multifocal lens and an astigmatic (toric) lens. The lens material (etafilcon A) is a copolymer of 2-hydroxyethyl methacrylate and methacrylic acid cross-linked with 1,1,1-trimethylol propane trimethacrylate and ethylene glycol dimethacrylate. The ACUVUE Contact Lens with visibility tint is tinted blue using Reactive Blue Dye #4 to make the lens more visible for handling. A benzotriazole UV absorbing monomer is used to block UV radiation. The average transmittance characteristics are less than 5 % in the UVB range of 280 to 315 nm and less than 30 % in the UVA range of 316 to 380 nm.

    AI/ML Overview

    The provided text describes a 510(k) submission for ACUVUE (etafilcon A) Contact Lenses, focusing on the substantial equivalence to a predicate device rather than a new medical device requiring extensive acceptance criteria and a standalone study for performance. Here's an analysis based on the provided information:

    This submission is for a contact lens, which is a medical device, but the context indicates it's an update to an existing product (ACUVUE with UV blocker) seeking substantial equivalence to a predicate device. Therefore, the "acceptance criteria" and "study" are geared towards demonstrating this equivalence rather than establishing completely new performance benchmarks for a novel AI or diagnostic device.

    Here's an attempt to extract and format the information as requested, acknowledging the limitations of applying these categories to a contact lens 510(k) submission:


    Acceptance Criteria and Device Performance for ACUVUE (etafilcon A) Contact Lens (with UV blocker)

    The primary "acceptance criteria" for this 510(k) submission are based on demonstrating substantial equivalence to the predicate device, ACUVUE (etafilcon A) Contact Lens clear and with visibility tint. Performance is largely assessed through a comparative analysis of physical and chemical characteristics.

    1. Table of Acceptance Criteria and Reported Device Performance

    CharacteristicAcceptance Criterion (Predicate Device Label Claim)Reported Device Performance (Subject Device Measured Avg)Comparison/Outcome
    % Water Content58%60%Close to claim, slightly higher
    Refractive Index @ 20° C1.40 ± 0.011.40Meets claim
    Dk (Fatt method, non-edge corrected)28 + 5.6 x 10⁻¹¹ cm²/sec (ml O₂/ml*mmHg)26 x 10⁻¹¹ cm²/sec (ml O₂/ml*mmHg)Within the range of the predicate, slightly lower than mean but within acceptable variance (implied by 5.6 variability)
    Colorlight bluelight blueMatches predicate
    % T @ 593 nm85% minimumConformsMeets minimum
    % T @ 280 - 315 nmavg < 5%2.0%Meets criterion (significantly less than 5%)
    % T @ 316 - 380 nmavg < 30%15.3%Meets criterion (significantly less than 30%)
    Base Curve Radius, mm8.8 mm8.87 mmVery close to claim
    Diameter, mm14.0 mm14.05 mmVery close to claim
    Power, Diopters-1.00 D-0.97 DVery close to claim

    Note: The "acceptance criteria" here are the established specifications or performance of the predicate device, or general industry standards for contact lenses. The subject device demonstrates performance within these established ranges or meets the specific new UV blocking claims.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not explicitly stated for each characteristic measurement. The values "% Water Content", "Refractive Index", "Dk", "Base Curve Radius", "Diameter", and "Power" are reported as "Measured (avg)", suggesting measurements were taken on a sample of devices.
    • Data Provenance: Not explicitly stated (e.g., country of origin). The studies appear to be internal or contracted by Vistakon, Johnson and Johnson Vision Products, Inc. The toxicology studies were conducted by a contract laboratory under Good Laboratory Practice Regulations.
    • Retrospective or Prospective: The physical and chemical characterization data would be considered prospective in the sense that new samples of the subject device were manufactured and tested for these properties.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • N/A. For this type of device (contact lens) and submission (510(k) for substantial equivalence), "experts" are not typically used to establish a "ground truth" for the test set in the way they would be for an AI diagnostic device (e.g., radiologists for image interpretation). The "ground truth" is based on objective physical, chemical, and biological measurements and established safety standards.

    4. Adjudication method for the test set

    • N/A. Adjudication methods (like 2+1, 3+1) are common in studies involving subjective interpretation (e.g., image analysis by multiple readers). Here, the testing involves objective measurements of material properties, toxicology, and sterility, which do not require interpretive adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • N/A. This is not an AI-assisted device. The submission pertains to contact lenses, not diagnostic or interpretive software.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • N/A. This is not an algorithm or software device. Performance is inherent to the physical product.

    7. The type of ground truth used

    For the various aspects of the device, the "ground truth" is established by:

    • Physical and Chemical Measurements: Objective measurements using validated laboratory methods (e.g., for water content, refractive index, Dk, transmittance, dimensions, power).
    • Toxicology Studies: In vivo and in vitro laboratory assays (e.g., ocular irritation, systemic toxicity, cytotoxicity, oral gavage toxicity, Ames test) with outcomes interpreted against pre-defined safety thresholds and regulatory standards.
    • Microbiology Studies: Validation of sterilization processes to achieve a specified Sterility Assurance Level (SAL) and compatibility with approved lens care products.
    • Material Property Data: Comparison of chemical compositions and properties to established standards and the predicate device.
    • Leachable Studies: Laboratory analysis to detect leachable monomers or additives against detection limits.

    8. The sample size for the training set

    • N/A. This is not a machine learning or AI device that requires a "training set." The product's properties are determined by its manufacturing process and material composition, not by learning from a dataset.

    9. How the ground truth for the training set was established

    • N/A. As there is no training set, this question is not applicable.

    Summary of Non-Clinical Studies Mentioned (Acting as the "Study" for Acceptance Criteria Demonstration):

    The document primarily relies on non-clinical studies to demonstrate safety and substantial equivalence:

    • Toxicology Studies:
      • Ocular Irritation (on molded etafilcon A contact lenses)
      • USP Systemic Toxicity (on molded etafilcon A contact lenses)
      • Cytotoxicity (on molded etafilcon A contact lenses)
      • 4-week Oral Gavage Toxicity Study in Rats (on neat UV absorbing additive)
      • Ames Salmonella/Microsome Reverse Mutation Assay (on neat UV absorbing additive)
      • Conducted by a contract laboratory under Good Laboratory Practice Regulations.
    • Microbiology Studies:
      • Validation of moist heat sterilization process (minimum SAL of 10⁻⁶).
      • Shelf-life stability data to support sterility claims.
    • Chemistry Studies:
      • Material property data comparison between current and modified materials (showing no significant differences except for UV transmittance).
      • Compatibility with Group IV lens care products (relying on prior demonstrations).
      • Leachable materials studies from finished lenses (indicating no leachable monomers or additives at detection levels).
    • Shelf-life Studies:
      • Protocols approved under N18-033, with initial shelf-life declared based on data, and a minimum of six months of real-time data collected.

    Conclusion Drawn from Studies:
    The studies support the safety and effectiveness of the device and its substantial equivalence to the predicate device, especially regarding the addition of a UV blocker. Clinical studies were explicitly not required for this Premarket notification because the USAN name and process are the same as the predicate device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1