510(k) Data Aggregation

    K Number
    K191010

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-10-10

    (177 days)

    Product Code
    Regulation Number
    868.5450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Palladium High Flow Therapy System is intended to deliver warmed and humidified high-flow respiratory gases to spontaneously breathing adult and pediatric patients weighing at least 3.5 kg within hospital, sub-acute and homecare settings.

    Device Description

    The Palladium High Flow Therapy System is designed to heat and humidify respiratory gases and deliver high flow respiratory therapy via a small-bore nasal cannula to spontaneously breathing pediatric and adult patients within hospital, sub-acute and homecare settings.

    Palladium is portable, self-contained, AC driven or battery powered with high flow therapy driven by an internal blower room air source and supplied with external oxygen or other gas sources, which is the same as the predicate Vapotherm Flowrest® device.

    Palladium utilizes similar semi-permeable small-bore tubing technology for humidification and integrity of a closed, single-use water path disposable as the reference Vapotherm Precision Flow® device.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the "Palladium High Flow Therapy System". It details the device's indications for use, its comparison to predicate devices, and non-clinical performance testing. The document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a standalone study with specific acceptance criteria and detailed quantitative results for the Palladium system's performance.

    Based on the provided text, a table of acceptance criteria and reported device performance, and details about a specific study proving the device meets these criteria, is not fully available in the requested format. The document describes several non-clinical tests (biocompatibility, electrical safety, software, and bench/performance testing) and states that the "results demonstrated that the device performance was met after conditioning and was substantially equivalent to the predicate device." However, it does not provide the specific numeric acceptance criteria or the measured performance values for the Palladium device in a clear, quantifiable manner for each test.

    Here's a breakdown of what can be extracted based on your requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes general categories of performance testing and states that the outcomes met expectations or demonstrated substantial equivalence. Specific measurable acceptance criteria values and the actual results of the Palladium device (e.g., "Oxygen accuracy: ±X%" and "Measured: Y%") are not provided.

    Test CategoryAcceptance Criteria (as described generally)Reported Device Performance (as described generally)
    BiocompatibilityCompliance with ISO 10993-1, ISO 18562-1, -2, -3, -4, ISO 10993-5, ISO 10993-10; appropriate biocompatibility profile for "externally communicating, tissue contacting, permanent duration of use" materials."Testing of the patient-contacting parts of the Palladium High Flow Therapy System demonstrates an appropriate biocompatibility profile for the device."
    Electrical Safety and EMCCompliance with IEC 60601-1:2005 Ed.3+A1:2012, IEC 60601-1-2: 2014 Ed.4, and AIM 7351731:2017 (RFID immunity). Specifics include basic safety, essential performance, emissions, and immunity."The testing demonstrated the appropriate electrical safety and electromagnetic compatibility (EMC) profile for the device."
    Software V&VCompliance with FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" classified as "moderate" level of concern."Software verification and validation testing were conducted and documentation was provided as recommended... The software for this device was considered as a 'moderate' level of concern."
    Bench/Performance TestingOxygen accuracy: Not explicitly stated but implied to meet predicate performance.Temperature accuracy: Expected to be ±2 °C (from comparison table).Humidification system output: Compliance with ISO 80601-2-74:2017.Flow rate accuracy: Not explicitly stated but implied to meet predicate performance.Patient contacting surface temperature: Not explicitly stated but implied to meet safety standards.Continuous use (Use Lifespan): Not explicitly stated but implied to demonstrate safe operation over time.Operating environment: Not explicitly stated but implied to meet specified environmental conditions.Nurse call compatibility: Not explicitly stated but implied to work with nurse call systems."The results demonstrated that the device performance was met after conditioning and was substantially equivalent to the predicate device."Regarding temperature accuracy, the comparison table states "Temperature Accuracy: SAME - K103316 ± 2 °C" for the subject device, indicating this was the acceptance criteria and presumably met. For other performance metrics, only a general statement of meeting performance and substantial equivalence is given.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified for any of the non-clinical tests.
    • Data Provenance: The tests are non-clinical (bench testing, lab-based safety, and performance evaluations), therefore no patient data or geographical provenance is applicable in the context of the supplied text. The company (Vapotherm, Inc.) is based in Exeter, NH, USA. The testing would have been conducted in a controlled lab environment.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not applicable. The documentation refers to non-clinical (bench) testing, software verification, and compliance with established standards. There is no mention of "ground truth" established by experts in the context of clinical expert review.

    4. Adjudication method for the test set:

    • This information is not applicable. Adjudication methods are typically used in clinical studies involving interpretation of medical data by multiple human readers, which is not described here.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC study was conducted or mentioned in this document. This is a 510(k) submission for a respiratory gas humidifier, not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This is not applicable as the device is a physical medical device (humidifier), not an algorithm or AI system. Its performance is evaluated through engineering and physiological testing, not AI-centric standalone performance studies.

    7. The type of ground truth used:

    • For the non-clinical performance and safety tests, the "ground truth" would be the objective measurements against pre-defined engineering specifications and internationally recognized standards (e.g., ISO, IEC). For example, temperature measurements against a calibrated reference, flow rate measurements, or chemical analysis for biocompatibility tests. It's based on technical standards and specifications, not expert consensus, pathology, or outcomes data in the usual clinical sense.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/Machine Learning device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable for the same reason as above.
    Ask a Question

    Ask a specific question about this device

    K Number
    DEN170001

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-04-10

    (462 days)

    Product Code
    Regulation Number
    868.5454
    Type
    Direct
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Precision Flow® HVNI is intended for use to add warm moisture to breathing gases from an external source for administration to a neonate/infant, pediatric and adult patients in the hospital and subacute institutions settings. It adds heat and moisture to a blended medical air/ oxygen mixture and assures the integrity of the precise air/oxygen mixture via an integral oxygen analyzer. The flow rates may be from 1 to 40 liters per minute via nasal cannula.

    Precision Flow® HVNI provides high velocity nasal insufflation (HVNI) with simultaneous oxygen delivery to augment breathing of spontaneously breathing patients suffering from respiratory distress and/or hypoxemia in the hospital setting. Precision Flow® HVNI is not intended to provide total ventilatory requirements of the patient and not for use during field transport.

    Device Description

    The Precision Flow® HVNI system delivers high flow rates of heated and humidified blended breathing gas through high flow nasal cannulas. The Precision Flow® HVNI system can connect to air and oxygen source. The Precision Flow® HVNI may be operated with limited performance at gas inlet pressures as low as 4 psi (28 kPa). For the full specified range of gas flows and oxygen percentages, both gas inlet pressures must be at minimum 40 psi (276 kPa).The main unit contains an integrated blender that delivers the targeted gas mixture to the disposable patient circuit (DPC). The Disposable Patient Circuit contains:

    • Water path: tubing from sterile water supply to vapor transfer cartridge
    • . Vapor transfer (humidification) cartridge: 2 types; low flow (1-8 lpm) and high flow (5-40 lpm)
    • Delivery Tube: triple lumen tubing
    • . Nasal Cannula

    The device automatically senses cartridge type. The available set temperature range is 33 °C to 39 °C. The device also contains a backup battery to provide power only for 15 minutes.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study information for the Precision Flow® HVNI, based on the provided text:

    Acceptance Criteria and Device Performance

    The acceptance criteria for the Precision Flow® HVNI are primarily non-clinical, focusing on safety, performance standards, software validation, and labeling requirements. The clinical studies establish safety and effectiveness by demonstrating non-inferiority or comparable outcomes to existing treatments.

    Acceptance Criteria CategoryAcceptance CriteriaReported Device Performance/Evidence
    Biocompatibility/MaterialsPatient-contacting components must be demonstrated to be biocompatible (Special Control 1). Tests include Cytotoxicity, Sensitization, Intracutaneous Reactivity, Extractables and Leachables, Volatile Organic Compounds (VOC), and Particulate Matter (PM2.5).All biocompatibility tests passed, demonstrating the biocompatibility of the device, in accordance with ISO 10993-1.
    Shelf Life/Reprocessing/SterilityCleaning instructions for reusable components must be validated (Special Control 3). Acceptance criteria for cleaning validation: no visible soil, protein level <6.4 micrograms/cm², TOC <12 micrograms/cm². Low-level disinfection validated by ≥ 6 log reduction of specified test organisms.Cleaning validation study met acceptance criteria: no visible soil, protein level <6.4 micrograms/cm², and TOC <12 micrograms/cm². Low-level disinfection demonstrated ≥ 6 log reduction of E. coli, S. aureus, K. pneumoniae, and P. aeruginosa.
    EMC/Electrical/Mechanical/Thermal SafetyElectrical safety, thermal safety, mechanical safety, electromagnetic compatibility, and radiofrequency identification (RFID) testing must be performed (Special Control 4). Standards: IEC 60601-1-2:2014, AAMI / ANSI ES60601-1:2005/(R) 2012 and A1:2012, AIM Standard 7351731 (RFID reference).Device passed tests in accordance with IEC 60601-1-2:2014, AAMI / ANSI ES60601-1:2005/(R) 2012 and A1:2012. Tested for RFID exposure using AIM Standard 7351731 as a reference.
    SoftwareSoftware verification, validation, and hazard analysis must be performed (Special Control 5). Adequate software documentation consistent with a "Major" level of software concern (FDA Guidance, May 11, 2005).Adequate software documentation provided. Software validation and verification testing demonstrated that the device met its design, implementation, and cybersecurity requirements.
    Non-clinical Performance TestingDevice performs as intended under anticipated conditions for use (Special Control 2). Specific tests: a. Alarm testing b. Continuous use thermal stability testing c. Humidity output testing d. Blender performance (FiO2 blending accuracy) Additional tests: Battery performance, Cannula pressure drop (blocked tube alarm not initiated). Standards: AAMI / ANSI / IEC 60601-1-8:2006 & A1:2012, ISO 8185:2007, ISO 80601-2-74.Blender performance: Verified set FiO2 at various flowrates (21%-100% FiO2) with high/low flow cartridges. Thermal Stability: Output temperature remained within specifications for 15 days across full range of settings, and another 15 days at max temp (43°C) and max flow (8/40 LPM). Battery performance: No change in set flow and FiO2 observed. Nasal cannula performance: Acceptable pressure drop at applicable flows, blocked tube alarm not initiated. Alarm testing, continuous use thermal stability, humidity output, and blender performance were implied to be successful as per "Additional testing".
    Clinical Effectiveness (Adults)Non-inferiority to NIPPV in avoidance of failure (including intubation/mechanical ventilation) within 72 hours of initiation (Primary Endpoints for Doshi et al. study).Doshi et al. study: Intubation within first 72 hours: 7% (HVNI) vs. 13% (NIPPV). Demonstrated non-inferiority (Wald p<0.001) for intubation rates. While arm failures (crossover) were higher in HVNI (26% vs 17%), the intubation rate was lower, suggesting clinical usefulness. Secondary findings: HVNI favored for pCO2 normalization at 240 min (p=0.02), physician perception (respiratory response, comfort/tolerance, simplicity of use), and patient perception of dyspnea was similar.
    Clinical Effectiveness (Neonates)Safety and effectiveness comparable to CPAP/nCPAP/NIPPV for respiratory distress/hypoxemia, without increasing adverse outcomes (pneumothorax, infection, ROP surgery, extubation failure, BPD, air leaks, IVH, NEC, sepsis). Non-inferiority for intubation rates within 72 hours of therapy initiation (Lavizzari et al. study).McQueen et al.: Compared to VON database (CPAP predominant), HFT did not increase pneumothorax (4.8% vs 4.4%) or nosocomial infection (14.8% vs 15.1%). Lower ROP surgery (2.2% vs 3.2%). Supported effectiveness and safety. Collins et al.: Non-inferior extubation failure rates for HFT (22%) vs nCPAP (34%). Similar clinical outcomes (BPD), trend for shorter supplemental oxygen in HFT. Improved nasal trauma score with HFT. Lavizzari et al.: Non-inferiority for primary endpoint (mechanical ventilation within 72 hrs). Failure rates: HFT (10.8%) vs nCPAP/BiPAP (9.5%). Lower bound of 95% CI (-6.0%) below zero met non-inferiority criteria. Similar secondary outcomes (duration of support, surfactant, air leaks, BPD). Kugelman et al.: Comparable failure to intubate rates for HFT (34.2%) vs NIPPV (31.6%). No clinically meaningful differences in secondary outcomes (air leak, nasal trauma, BPD, IVH, NEC, sepsis).
    LabelingLabeling must meet 21 CFR 801.109, include specific information (FiO2 ranges, flowrates, continuous monitoring warning, condensation warning, alarm description/function, not a CPAP device warning, aspiration risk warning) (Special Control 6).Labeling meets 21 CFR 801.109 and includes all specified warnings and descriptions as per Special Control 6.
    Adverse EventsNo device-related serious adverse events.In the Doshi et al. adult clinical study, there were no adverse events related to the use of the device. All adverse events were related to the underlying patient condition. The neonate studies also did not raise new safety concerns.

    Study Information

    Due to the nature of this particular device (high flow humidified oxygen delivery device) and the provided text, the "studies" are clinical trials and published literature, rather than an AI/ML-specific validation study with test sets, ground truth experts, or training sets in the typical sense. The regulatory submission leverages existing clinical evidence.

    For the Adult Clinical Study (Doshi et al.):

    1. Sample size used for the test set and the data provenance:

      • Sample Size: 204 adult patients (104 in HVNI arm, 100 in NIPPV arm).
      • Data Provenance: Multi-center, prospective, randomized clinical trial conducted at 5 sites. Country of origin not explicitly stated but implies US context for FDA submission.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This is a clinical trial where clinical outcomes (e.g., intubation, failure to tolerate device, physiological parameters like O2 saturation, PaO2, PCO2, relief of respiratory distress, deteriorating medical status) were the ground truth. These outcomes were assessed by treating clinicians (physicians, nurses, respiratory therapists) at the study sites. The "experts" are the medical professionals directly involved in patient care and decision-making within the clinical trial setting. Specific qualifications beyond "physician" are not detailed in the text.
    3. Adjudication method for the test set:

      • Not explicitly stated as a formal adjudication panel in the typical AI/ML context. Treatment failure was defined by clear clinical criteria (need for intubation, device intolerance, failure to oxygenate/ventilate, failure to relieve distress, deteriorating medical status). These assessments are inherent to clinical practice and were likely made by the treating clinicians in real-time, within the trial's protocol.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, this was not an MRMC study comparing human readers with and without AI assistance. It was a comparative effectiveness study between two medical devices/therapies (HVNI vs. NIPPV).
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • No, this device is a medical device for patient breathing support; it is not an AI algorithm performing a diagnostic task. Its performance is measured by its clinical impact on patients.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Outcomes Data: Primary ground truth was clinical outcomes data, specifically "treatment failure rate, defined as the need for intubation, and arm failure rate, defined as the decision for crossover to the alternate therapy, within 72 hours of initiation of assigned therapy." Other outcomes included vital signs, blood gas analysis (e.g., pCO2), physician perception, and adverse events.
    7. The sample size for the training set:

      • Not applicable in the context of this device being a medical hardware system, not an AI/ML algorithm requiring a specific training set of patient data.
    8. How the ground truth for the training set was established:

      • Not applicable for the same reason as above.

    For the Neonate Patient Population (McQueen et al., Collins et al., Lavizzari et al., Kugelman et al.):

    1. Sample size used for the test set and the data provenance:

      • McQueen et al.: Retrospective cohort study. HFT group: Data from five centers, no specific n given for test set, but stated "used Vapotherm HVNI as their primary and predominant mode." Compared to Vermont Oxford Network (VON) database: 176,599 very low birth weight baby admissions (control/comparison data).
      • Collins et al.: 132 low birth weight infants (< 2500g, <32 weeks gestation). Randomized (67 HFT, 65 nCPAP). Data provenance: Randomized Controlled Trial.
      • Lavizzari et al.: Randomized trial, 298 infants (158 HFT, 158 nCPAP/BiPAP). Data provenance: Prospective, randomized clinical non-inferiority trial.
      • Kugelman et al.: 76 infants (<35 week gestational age, >1000g). Randomized (38 HFT, 38 NIPPV). Data provenance: Randomized Pilot Study.
      • Data Provenance for all neonate studies: Published literature, likely international based on author names and journal types, but not explicitly stated for each study.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Similar to the adult study, ground truth involved clinical outcomes assessed by treating medical professionals (neonatologists, nurses, etc.) within the context of their respective clinical trials or retrospective data collection. Specific expert qualifications beyond clinical roles are not detailed within the provided text.
    3. Adjudication method for the test set:

      • Not explicitly stated as a formal adjudication panel. Outcomes were generally defined by protocol-driven clinical events (e.g., extubation failure, need for intubation, physiological parameters).
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No. These are clinical studies comparing therapies, not AI assistance for human readers.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • No.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • Outcomes Data: Primary ground truths included extubation failure rates, rates of intubation/mechanical ventilation, occurrences of pneumothorax, nosocomial infection, ROP surgery, bronchopulmonary dysplasia (BPD), air leaks, IVH, NEC, sepsis, duration of respiratory support, need for surfactant, time to full feeds, length of stay, mortality, and nasal trauma scores.
    7. The sample size for the training set:

      • Not applicable.
    8. How the ground truth for the training set was established:

      • Not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K111640

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-10-07

    (116 days)

    Product Code
    Regulation Number
    868.5450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Precision Flow® - Heliox is intended to warm and add moisture to breathing gases from an external source for administration to neonate/infant, pediatric and adult patients in the hospital and sub-acute institutions. It adds heat and moisture to a blended medical heliox (79% helium. 21% oxygen)/oxygen mixture and assures the integrity of the precise heliox (79% helium, 21% oxygen)/oxygen mixture via an integral oxygen analyzer. The flow rates may be from 1 to 40 liters per minute via nasal cannula.

    Device Description

    The Precision Flow® - Heliox consists of two parts: The main unit which contains all the electrical and electronic components including the electronic blender and flow controllers. All the sensors are located in the main unit. The main unit has no water pathways and the gas pathway contains only dry gas at room temperature, and consequently does not need internal cleaning or disinfection. The disposable components comprising the disposable water module, vapor transfer cartridge and heated delivery tube. Conditions in the circulating water and gas streams are sensed remotely via the interface between the main unit and the disposable module.

    AI/ML Overview

    The Vapotherm Precision Flow® - Heliox is a medical device intended to warm and humidify breathing gases, specifically a heliox/oxygen mixture, for neonate/infant, pediatric, and adult patients in hospital and sub-acute settings.

    Here's a breakdown of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document lists various standards that the Precision Flow® - Heliox met, with some exceptions. It also includes functional and design verification tests. The "Reported Device Performance" for these tests is generally stated as: "In all instances, the Precision Flow® - Heliox functioned as intended and the results observed were as expected." Specific quantitative performance data is limited within this summary but is generally implied by compliance with the referenced standards.

    Acceptance CriteriaReported Device Performance
    I. Compliance with Standards:"In all instances, the Precision Flow® - Heliox functioned as intended and the results observed were as expected." (Implied compliance with the intent of the standard, with specific exceptions noted)
    ISO 62304: Medical Device Software - Software Life Cycle ProcessMet
    ISO 8185:2007: Respiratory tract humidifiers for medical useMet
    ISO 10993-5:2009: Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicityMet
    ISO 10993-10:2010: Tests for Irritation and SensitizationMet
    ISO 11195:1995: Gas mixers for medical use - Stand-alone gas mixersMet
    ISO 14971:2007: Medical devices - Application of risk management to medical devicesMet
    ANSI/AAMI/ISO 15223-1:2007/A1:2008: Medical Devices - Symbols to be used with Medical Device Labels, Labeling and information to be supplied - Part 1: General RequirementsMet
    ANSI/AAMI ES60601-1:2005: Medical electrical equipment – Part 1: General requirements for basic safety and essential performanceMet
    ANSI/AAMI/IEC 60601-1-2:2007 3rd edition: Medical Electrical Equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic Compatibility - Requirements and testsMet (with exceptions: Section 5: Device cannot produce hazardous radiation; Section 6: Anesthetic mixtures not applicable)
    EN60601-1-4:1999: Medical Electrical Equipment: Part 1: General Requirements for Safety 4. Collateral Standard: Programmable Electrical Medical SystemsMet
    IEC 60601-1-8 2006-10: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systemsMet
    IEC 60529 IPX1: Drip ProofMet
    AAMI TIR 32:2004: Software Risk ManagementMet
    AAMI TIR 36:2007: Validation of SoftwareMet
    ISTA -1A:2001: Procedure 1A: Packaged-Products weighing 150 LB (68 KG) or LessMet
    EPA/625/R-96/010b (VOC) Method TO-15 (1999): Compendium Method TO-15 Determination of Volatile Organic Compounds (VOCs) in Air collected in specially-Prepared canisters and analyzed by chromatography/mass spectrometry (GC/MS)Met
    NIOSH Method 0500 (1994): NIOSH manual of Analytical methods (NMNM) fourth edition August 1994 (Particulate Matter Testing)Met
    II. Design Verification Plan (Required Tests):"In all instances, the Precision Flow® - Heliox functioned as intended and the results observed were as expected."
    Functional Performance TestMet
    Alarms & Fault Conditions TestMet
    Patient Connector Attachment TestMet
    Water Connector Attachment TestMet
    Extended Life TestMet
    Membrane Switch and Adjustment Dial TestMet
    Software Verification TestMet
    Fluid & Gas Pathways Leak TestMet
    Cartridge Insertion TestMet
    Door TestMet
    IPX TestMet
    Environmental Temperature/Humidity Extremes TestMet
    III. Performance Characteristics (Implied by Comparison for Substantial Equivalence):
    Oxygen % Range21 to 100% (Equivalent to predicates)
    FiO2 Accuracy2 to 40 lpm: +/-2%; 1 to 2 lpm: +/-5% (Equivalent to predicates)
    Max Flow Range40 lpm (Equivalent to predicates)
    Alarm Sound Level (Medium Priority)47 dB @ 1m (Equivalent to predicate)
    Alarm Sound Level (Low Priority)45 dB @ 1m (Equivalent to predicate)

    2. Sample Size Used for the Test Set and Data Provenance

    The 510(k) summary does not provide specific sample sizes for particular performance tests or explicitly mention a "test set" in the context of clinical or image-based studies. The performance data section broadly states: "In all instances, the Precision Flow® - Heliox functioned as intended and the results observed were as expected." This implies that the design verification tests listed were conducted on the device.

    • Sample Size for Test Set: Not explicitly stated for each test, but implied to be sufficient for demonstrating compliance with cited standards and internal design verification plans.
    • Data Provenance (Country of Origin, Retrospective/Prospective): Not specified. The tests are likely laboratory or bench testing, not clinical studies involving patient data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • This information is not applicable to this type of device and study. The testing for the Precision Flow® - Heliox involves engineering and performance validation against objective technical standards (e.g., flow rates, temperature, alarm function, electrical safety), not subjective interpretation requiring expert consensus.

    4. Adjudication Method for the Test Set

    • This information is not applicable as the evaluation relies on objective measurements against engineering specifications and regulatory standards, not subjective interpretations requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of AI vs. Without AI Assistance

    • No, an MRMC comparative effectiveness study was not done. This device is a medical gas humidifier/blender, not an AI-powered diagnostic or decision-support tool. It does not involve human readers interpreting cases with or without AI assistance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    • No, this is not applicable. The device is hardware with integrated software for control, not a standalone algorithm. Its performance is inherent to its physical operation and control systems.

    7. The Type of Ground Truth Used

    • The "ground truth" for this device's performance is established by objective engineering specifications, technical standards (e.g., ISO, ANSI/AAMI, IEC), and manufacturer-defined design requirements. This includes accuracy of gas blending, flow rates, temperature control, alarm thresholds, electrical safety, and biocompatibility.

    8. The Sample Size for the Training Set

    • This is not applicable in the context of a traditional "training set" for machine learning or AI models. The device's software is likely developed using standard software engineering practices and verified accordingly, rather than trained on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    • This is not applicable for the reasons stated above. The "ground truth" for the device's operational parameters is derived from engineering design, relevant medical standards, and intended physiological effects of humidified gas delivery.
    Ask a Question

    Ask a specific question about this device

    K Number
    K103316

    Validate with FDA (Live)

    Device Name
    FLOWREST
    Manufacturer
    Date Cleared
    2011-02-08

    (90 days)

    Product Code
    Regulation Number
    868.5450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Flowrest® delivers warmed humidified high-flow breathing gases. The Flowrest® is intended for treating spontaneously breathing patients who require warmed and humidified high-flow respiratory gases within the homecare, subacute and hospital settings.

    Device Description

    The Flowrest® is an integrated non-invasive high flow device incorporating a blower, an air flow heater, a heated humidifier, and a dual lumen breathing circuit and cannula. Filtered room air is drawn in through a blower, where it passes a heater. The heated air is split between the inner lumen and outer lumen of the dual lumen breathing circuit. The heated air in the inner lumen passes through a heated humidifier and is delivered to the patient through a cannula. The heated air in the outer lumen passes through the entire length of the breathing circuit and serves to insulate the humidified patient air, thus minimizing condensation and rainout.

    The Flowrest® is comprised of two functional, yet integrated components. One is a motorized fan assembly that provides the physician prescribed high flows of heated breathing and insulating gases between 15 and 35 liters per minute. The fan speed is directly related to flow rate and is controlled by software. The blower assembly output connects directly to a humidification chamber at the front of the device.

    The second component of the Flowrest® is a heated humidifier. The water is contained in the humidification chamber positioned on a heater plate at the front of the unit. The chamber connects directly to the blower assembly. Air flow from the device passes through the heated humidification chamber, is warmed and humidified and after passing through the breathing circuit, is delivered to the patient through a cannula. The second flow of warm air through the outside lumen of the delivery tube bypasses the humidification chamber through the top of the humidifier chamber. Ambient temperature is monitored in order to reduce humidified patient air condensation. Temperature controls are controlled by software.

    AI/ML Overview

    The provided document is a 510(k) summary for the Vapotherm Flowrest®, a respiratory gas humidifier. It details the device's intended use, technological characteristics, and a comparison to a predicate device to establish substantial equivalence.

    However, the document does not contain the kind of information typically found in studies proving an AI/Machine Learning device's performance against acceptance criteria. Specifically, it lacks:

    • A table of acceptance criteria for AI performance metrics (e.g., sensitivity, specificity, AUC).
    • Details about a test set (sample size, provenance).
    • Information on expert ground truth establishment (number of experts, qualifications, adjudication).
    • MRMC studies, standalone AI performance, or training set details.
    • The type of ground truth used (pathology, outcomes data, etc.)

    The document focuses on the engineering and safety performance of a medical device (a humidifier) against established medical device standards (e.g., ISO, IEC), and its substantial equivalence to a predicate device.

    Therefore, I cannot fulfill your request for AI-specific performance criteria and study details based on the provided text. The "Performance Data" section in the document refers to compliance with safety and design standards for a gas humidifier, not the performance of an AI algorithm.

    If this was an AI/ML device, the information would typically look like this (hypothetically, what you are asking for):

    1. A table of acceptance criteria and the reported device performance (Hypothetical for an AI device):

    MetricAcceptance CriteriaReported Device Performance (Hypothetical)
    Sensitivity≥ 90% for detecting Condition X92.5%
    Specificity≥ 80% for ruling out Condition X85.1%
    Area Under ROC Curve≥ 0.900.93
    Inference Time< 2 seconds per image1.5 seconds

    2. Sample size used for the test set and the data provenance (Hypothetical):

    • Test Set Sample Size: 5,000 cases (e.g., chest X-rays, retinal images).
    • Data Provenance: Retrospective data collected from multiple hospitals across the United States, United Kingdom, and Germany.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (Hypothetical):

    • Number of Experts: 3 board-certified radiologists/pathologists/ophthalmologists.
    • Qualifications: All experts had at least 10 years of post-residency experience in their respective specialties, with sub-specialty expertise relevant to the target condition.

    4. Adjudication method (Hypothetical):

    • Adjudication Method: 2+1 (two experts independently read each case; in case of disagreement, a third, senior expert provided a decisive read). Or, alternatively, consensus reading among all three experts.

    5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done (Hypothetical):

    • MRMC Study Done: Yes.
    • Effect Size of Human Reader Improvement: The MRMC study showed that human readers aided by the AI significantly improved their diagnostic accuracy (e.g., AUC increased from 0.82 to 0.91, a statistically significant gain of 0.09) and reduced their reading time by 20% compared to reading without AI assistance for detecting Condition X.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done (Hypothetical):

    • Standalone Performance: Yes, standalone performance was evaluated on a held-out test set (as described in point 2) and met the pre-defined acceptance criteria (as in point 1).

    7. The type of ground truth used (Hypothetical):

    • Ground Truth Type: Expert consensus (as established by the process in point 3 and 4), validated against, wherever possible, pathology reports or long-term clinical outcomes data.

    8. The sample size for the training set (Hypothetical):

    • Training Set Sample Size: 50,000 cases.

    9. How the ground truth for the training set was established (Hypothetical):

    • Training Set Ground Truth: Established by a combination of clinical reports, a single expert read (for efficiency with high volume), and automated labeling where robustly feasible, with a subset verified by multi-expert consensus. The training data was regularly reviewed and curated to ensure label accuracy.

    In summary, the provided document K103316 does not describe a study related to AI/ML device performance. It is for a traditional medical device (a humidifier) and focuses on compliance with established safety and performance standards for that type of device, demonstrating substantial equivalence to a predicate.

    Ask a Question

    Ask a specific question about this device

    K Number
    K072845

    Validate with FDA (Live)

    Device Name
    PRECISION FLOW
    Manufacturer
    Date Cleared
    2008-07-17

    (287 days)

    Product Code
    Regulation Number
    868.5450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Precision Flow™ is intended for use to add warm moisture to breathing gases from an external source for administration to a neonate/infant, pediatric and adult patients in the hospital, sub-acute institutions, and home settings. It adds heat and moisture to a blended medical air/oxygen mixture and assures the integrity of the precise air/oxygen mixture via an integral oxygen analyzer. The flow rates may be from 1 to 40 liters per minute via nasal cannula.

    Device Description

    The Precision Flow™ consists of two parts: The main unit which contains all the electrical and electronic components including the electronic blender and flow controllers. All the sensors are located in the main unit. The main unit has no water pathways and the gas pathway contains only dry gas at room temperature, and consequently does not need internal cleaning or disinfection. The disposable components comprising the disposable water module, vapor transfer cartridge and heated delivery tube. Conditions in the circulating water and gas streams are sensed remotely via the interface between the main unit and the disposable module.

    AI/ML Overview

    This 510(k) summary describes a medical device, the Vapotherm Precision Flow™, which is a respiratory gas humidifier. It is not an AI/ML powered device, therefore, many of the requested categories related to AI performance, ground truth, and expert evaluation are not applicable.

    Here's an analysis of the provided text based on your request, highlighting the non-applicability of AI/ML specific criteria:

    Acceptance Criteria and Device Performance (Non-AI/ML Device)

    The Vapotherm Precision Flow™ is cleared based on demonstrating substantial equivalence to predicate devices, primarily through non-clinical performance data (testing) and a comparison of its technological characteristics to existing devices. The "acceptance criteria" here are implicitly that the device performs its intended functions (humidifying and blending medical gases) safely and effectively, and that these performance characteristics are comparable to legally marketed predicate devices without raising new safety or effectiveness concerns.

    Since this is not an AI/ML device, performance metrics like sensitivity, specificity, accuracy, or AUC are not applicable or reported in this 510(k) summary. Instead, the performance is demonstrated through various engineering, safety, and biocompatibility tests.

    Acceptance Criteria (Implied)Reported Device Performance
    Biocompatibility: No adverse biological reactions.Biocompatibility Tests (LLNA, Intracutaneous Reactivity, MEM Elution): Performed, implying acceptable results (otherwise it wouldn't be marketed).
    Emissions: No harmful volatile organic compounds or particulates.Volatile Organic Compounds (VOC), Particulate Matter tests: Performed to recognized standards, implying acceptable results.
    Electrical Safety: Meets international safety standards.EMC Test, Emissions Test, Safety Inspection Test, IEC 60601-1, IEC 60601-1-4, IEC 60601-1-8: Performed, ensuring compliance with general requirements for safety, programmable electrical medical systems, and alarm systems.
    Microbiological Safety: No detectable bacteria after extended use.Bio Burden, 30 Day Comparative Use Testing: "No detectible bacteria in any water condensation samples from the Precision Flow® devices either initially (3 days) or after 30 days of operation."
    Software Functionality: Verified and validated.Ximedica TRP 1097, TRP 1071 (Black Box), Unit Test Case Listing - 3VAP1004/TRP 1092 (White Box): Software verified and validated in accordance with applicable FDA guidance.
    Thermal Stability: Maintains temperature regulation.Thermal Stability: Performed, indicating proper temperature control.
    Blender Accuracy: Maintains precise air/oxygen mixture.Blender Comparison Performance: Performed, indicating comparable performance to predicate blenders, assuring integrity of the precise air/oxygen mixture via an integral oxygen analyzer.
    Substantial Equivalence: Comparable to predicate devices.The device has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. Performance data demonstrate it is as safe and effective as the predicate devices, with minor differences raising no new issues.

    Non-Applicable (N/A) for AI/ML Specific Questions:

    Given that K072845 describes a humidifier, not an AI/ML-powered device, the following categories are not applicable. The submission process for such a device focuses on engineering specifications, safety testing, and performance validation against established standards and predicate devices, rather than statistical performance against a ground truth dataset.

    2. Sample size used for the test set and the data provenance: N/A (Not an AI/ML device, no "test set" in the context of an algorithm's classification/prediction performance). Testing involved physical hardware, software, and biological compatibility, not a data-driven test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: N/A (Not an AI/ML device, no ground truth established by experts for algorithmic performance).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: N/A (Not an AI/ML device, no adjudication of algorithm outputs).

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: N/A (Not an AI/ML device. No human-in-the-loop AI assistance involved).

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: N/A (Not an AI/ML device, no standalone algorithm performance to evaluate).

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): N/A (Not an AI/ML device). The "ground truth" for this device relates to physical measurements meeting specifications (e.g., oxygen percentage, temperature, bacterial counts being zero).

    8. The sample size for the training set: N/A (Not an AI/ML device, no "training set"). Software verification and validation are described, but this refers to traditional software engineering processes, not machine learning model training.

    9. How the ground truth for the training set was established: N/A (Not an AI/ML device, no "training set" or ground truth in the ML context).

    Ask a Question

    Ask a specific question about this device

    K Number
    K070056

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-04-13

    (98 days)

    Product Code
    Regulation Number
    868.5450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vapotherm 2000i and 2000h are designed to add moisture to and to warm breathing gases for administration to patients, including neonates/infants, and adults. The flow rates may be from 1 to 40 liters per minute via nasal cannula. Environments for use -- Home, Hospital, Sub-acute Institutions.

    Device Description

    The Vapotherm 2000i device is a modification to cleared Vapotherm 2000i device. The base device of the modified Vapotherm 2000i device has the same technological characteristic as the predicate Vapotherm 2000i device. The software contained in the modified device has not been modified as a result of the corrections being effected and remains identical to the software resident in the predicate device.

    AI/ML Overview

    The provided 510(k) summary for Vapotherm's 2000i Respiratory Gas Humidifier is for a Special 510(k) Corrective Action Being Effected. This means the submission is for a modification to an already cleared device, and the focus is on demonstrating that the modifications do not raise new issues of safety or effectiveness.

    Therefore, the document does not contain specific acceptance criteria, study details, sample sizes, expert qualifications, or ground truth establishment methods in the way a traditional de novo or full 510(k) submission for a novel device or significant change might.

    Here's an analysis based on the provided text, addressing your points where possible and noting where information is absent:


    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Not explicit in document"Vapotherm has conducted extensive testing to verify and validate the changes to the previously marketed Vapotherm 2000i device."
    • Explanation: The document states that the software has not been modified and remains identical to the predicate device. The modifications are implied to be non-software related and do not raise new issues of safety or effectiveness. The "extensive testing" would likely have focused on verifying that the changes did not negatively impact the device's original performance specifications (e.g., humidification output, temperature control, safety alarms), rather than establishing new performance criteria.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size for Test Set: Not mentioned.

    • Data Provenance: Not mentioned.

    • Explanation: Given this is a Special 510(k) for corrective action on an existing device, the testing would likely be internal verification and validation testing, not a clinical study on a patient population.


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Number of Experts: Not applicable.

    • Qualifications of Experts: Not applicable.

    • Explanation: As this is a medical device humidifier, not an AI or diagnostic device, the concept of "ground truth" derived from expert consensus for a test set is not relevant in this context. Performance would be measured against engineering specifications and regulatory standards for humidification.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication Method: Not applicable.

    • Explanation: Again, this is for a hardware modification to a humidifier, not a diagnostic or AI device where adjudication of expert opinions would typically be required for ground truth.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No.

    • Effect Size: Not applicable.

    • Explanation: This type of study is relevant for AI-powered diagnostic tools. It is not applicable to a respiratory gas humidifier.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance Test: Not applicable in the context of an algorithm.

    • Explanation: The device is a physical humidifier. While it has embedded software, the performance assessment would be of the device's physical functions (e.g., humidification, temperature control, flow rates), not a standalone algorithm's diagnostic or predictive capabilities. The document explicitly states the software "has not been modified" and "remains identical to the software resident in the predicate device."


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Type of Ground Truth: Not explicitly stated, but for a humidifier, "ground truth" would be established through engineering specifications and quantifiable physical measurements (e.g., hygrometry, thermometry, flow meters) against established performance standards for respiratory gas humidifiers.

    8. The sample size for the training set

    • Sample Size for Training Set: Not applicable.

    • Explanation: The term "training set" is typically used for machine learning models. This device is a traditional medical device, not an AI model, so there is no training set in that sense.


    9. How the ground truth for the training set was established

    • Ground Truth Establishment for Training Set: Not applicable.

    • Explanation: Similar to point 8, this pertains to AI/ML development and is not relevant for this device.


    Summary of Device and Approval:

    • Device Name: Vapotherm™ 2000i and 2000h Respiratory Gas Humidifier
    • Submission Type: Special 510(k) Corrective Action Being Effected
    • Purpose of Submission: Modification to a previously cleared Vapotherm 2000i device (K042245).
    • Key Statement for Performance: "Vapotherm has conducted extensive testing to verify and validate the changes to the previously marketed Vapotherm 2000i device."
    • Reason for Approval: The modified device has the same intended use and indications for use, similar principles of operation, and similar technological characteristics as the previously cleared predicate, and the modifications "do not raise any new issues of safety or effectiveness." The software was explicitly not modified.
    • Regulatory Outcome: Found substantially equivalent to the predicate device (K042245).
    Ask a Question

    Ask a specific question about this device

    K Number
    K042245

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-08-30

    (11 days)

    Product Code
    Regulation Number
    868.5450
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vapotherm™ 2000h and 2000i are designed to add moisture to and to warm breathing gases for administration to patients, including neonates/infant, pediatrics, and adults. The flow rates may be from 1 to 40 liters per minute via nasal cannula. Environments of use - Home, Hospital, Sub-acute Institutions

    Device Description

    The Vapotherm 2000i and Vapotherm 2000h are identical and share the concept of humidification by transpiration of water vapor across a membrane by the use of a low or high flow cartridge with membrane bundles. The difference in the cartridges is only the number of membrane bundles included, fewer in the low flow. Both units and cartridges produce a highly humidified air (relative humidity >95%), virtually free of droplets, at body temperature or above at flow rates from 1 to 40 lpm via a nasal cannula. The water content at 41℃ is 40-50 mg/liter, about fourfold higher than can be achieved by humidification at room temperature. The unique combination of high flow and high vapor-phase humidity allow an unusually wide range of clinical applications. Applications previously considered impractical because of limited patient tolerance for high nasal flow can now be routine because of the comfort provided by warmth and high humidity.

    AI/ML Overview

    The provided text primarily consists of regulatory summaries and approval letters for the Vapotherm 2000h and 2000i devices. It details the device's function, indications for use, and a comparison to predicate devices. However, it does not contain information about acceptance criteria, specific studies conducted to demonstrate performance against those criteria, sample sizes, expert involvement, or any statistical evaluation of device performance.

    Therefore, I cannot directly answer your request based on the provided input. The document focuses on regulatory equivalence rather than an empirical study demonstrating performance against specific acceptance criteria.

    To elaborate, the document states:

    • Device Description: The Vapotherm 2000i and 2000h produce "a highly humidified air (relative humidity >95%), virtually free of droplets, at body temperature or above at flow rates from 1 to 40 lpm via a nasal cannula. The water content at 41℃ is 40-50 mg/liter..."
    • Comparison to Predicate Devices: It highlights "Same" or "No changes" for several attributes when comparing to predicate devices (K000401, K013486, K040862).

    These descriptions are technical specifications and comparisons, not acceptance criteria for a validation study. The FDA approval K042245 is based on substantial equivalence to predicate devices, meaning it largely relies on the predicate devices having already met safety and effectiveness standards, rather than requiring extensive new clinical trial data for this specific iteration.

    Ask a Question

    Ask a specific question about this device

    K Number
    K013486

    Validate with FDA (Live)

    Device Name
    VAPOTHERM 2000H
    Manufacturer
    Date Cleared
    2003-05-30

    (588 days)

    Product Code
    Regulation Number
    868.5450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vapotherm 2000h (5000) can mix air with oxygen from an external source and adds heat and moisture to the gas. The Vapotherm 2000h (5000) is for use in home, hospital or sub-acute institutional settings.

    Device Description

    The Vapotherm 2000i, predicate, and the Vapotherm 2000h (5000) are identical and share the concept of humidification by transpiration of water vapor across a membrane. Both produce a high flow of highly humidified air (relative humidity >95%), virtually free of droplets, at body temperature or above. The water content at 41°C is 40-50 mg/liter, about fourfold higher than can be achieved by humidification at room temperature. The unique combination of high flow and high vapor-phase humidity allow an unusually wide range of clinical applications. Applications previously considered impractical because of limited patient tolerance for high nasal flow can now be routine because of the comfort provided by warmth and high humidity.

    The Vapotherm 2000h (5000) is an unmodified Vapotherm 2000i but with an internal air compressor and external oxygen source, wall or cylinder oxygen. There is a single inlet fitting. like the 2000i for connection to an external oxygen source, if desired.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Vapotherm 2000h (5000), a humidifier for respiratory gas. This submission focuses on demonstrating substantial equivalence to a predicate device (Vapotherm 2000i) rather than presenting a study with specific acceptance criteria and performance data in the context of AI/machine learning.

    Therefore, many of the requested categories (e.g., sample size for test set, number of experts, adjudication method, MRMC study, standalone performance, training set details) are not applicable to this type of regulatory submission and the information provided. This is a medical device, not an AI/ML-based diagnostic system, and thus does not involve "ground truth" derived from expert consensus on images or pathology in the way an AI study would.

    Here’s the information that can be extracted and a clear explanation for the N/A categories.

    1. Table of Acceptance Criteria and Reported Device Performance

    For this 510(k), the "acceptance criteria" is demonstrating substantial equivalence to the predicate device (Vapotherm 2000i). The "reported device performance" is essentially a comparison of attributes to show similarity.

    AttributeAcceptance Criteria (Predicate: Vapotherm 2000i)Reported Device Performance (Vapotherm 2000h (5000))
    Indications for useTo add moisture to and to warm breathing gases for administration to a patientSame (Adds "with an air or air/oxygen mixture" for patent population)
    Environments of useHome, Hospital, Sub-acute Institutions, not specifiedThe same
    Patient PopulationFor use with any patient utilizing supplemental oxygen in which humidification would be beneficialAdd - “ with an air or air/oxygen mixture” Otherwise the same.
    ContraindicationsNoneThe same
    Dimensions11" x 5.5" x 4.5"12" x 15" x 10"
    Weight< 6 lbs without water25 lbs. without water
    Power max.250 VA (warm-up), 80 VA (continuous)The same (for Vapotherm), Compressor - 49 watts
    Input power115 VAC, 60 HzThe same
    Gas Source Pressure4-50 psi, wall source or cylinderInternal compressor - 8 psi
    Gas SourceWall source or CylinderInternal AC compressor or wall or cylinder oxygen
    Gas fittingsOneOne
    Method to regulate flowFlow meter at wall or cylinderFlow meter on device for compressor, plus flow meter at oxygen source.
    Capable of variable FiO₂No – set at wall if practitioner desiresNo – set at wall if practitioner desires
    Monitoring of FiO₂NoneInstructs to verify with an in-line oxygen analyzer
    Membrane typeBasic membrane type humidifier, hollow fiber cartridgeThe same
    Method of humidifyingWarmed water circulated over outside of hollow fibers, air pumped through center. Fibers permeable to water, water transpires as vapor. Water temp software-controlled.The same
    Water typeCommon tap waterThe same
    Compressor - Meets UL 544Yes (De Vilbiss K963349)Yes (Proposed Thomas)
    Compressor - PortableYes (De Vilbiss K963349)Yes (Proposed Thomas)
    Compressor - Max. pressure90 psi (De Vilbiss K963349)8.3 psi (Proposed Thomas)
    Compressor - Max liter flow30 lpm (De Vilbiss K963349)25 lpm in this application (Proposed Thomas)

    Conclusion of Substantial Equivalence: The document states that the Vapotherm 2000h (5000) is "substantially equivalent" to the predicate device because:

    • They have substantially equivalent intended uses.
    • They have the same environments for use.
    • They are similar in design.
    • They employ the same technology.
    • "There are no significant differences that affect the safety or effectiveness of the intended device as compared to the predicate devices."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • N/A: This document pertains to a 510(k) submission for a non-AI medical device (a humidifier). The submission relies on a comparison of device specifications and an assessment of technical and functional equivalence to a predicate device, as opposed to performance studies on a "test set" in the context of diagnostic accuracy or AI/ML. No clinical or performance data from a "test set" is presented in this summary in the way it would be for an AI device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • N/A: Not applicable for a non-AI device demonstrating substantial equivalence. There is no concept of "ground truth" established by experts for this type of submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • N/A: Not applicable for this type of device and submission.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • N/A: Not applicable. This is not an AI-assisted diagnostic device, and no MRMC study was performed or is relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • N/A: Not applicable. This is not an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • N/A: Not applicable. The "ground truth" concept as defined for AI/diagnostic studies is not relevant here. The device's performance is gauged by meeting physical and functional specifications compared to a predicate.

    8. The sample size for the training set

    • N/A: Not applicable. This document does not describe an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    • N/A: Not applicable. This document does not describe an AI/ML device that requires a training set.
    Ask a Question

    Ask a specific question about this device

    K Number
    K000401

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2000-08-24

    (199 days)

    Product Code
    Regulation Number
    868.5450
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1