Search Filters

Search Results

Found 1134 results

510(k) Data Aggregation

    K Number
    DEN250007

    Validate with FDA (Live)

    Device Name
    Delivery Date AI
    Manufacturer
    Date Cleared
    2026-02-11

    (331 days)

    Product Code
    Regulation Number
    N/A
    Type
    Direct
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K251423

    Validate with FDA (Live)

    Date Cleared
    2026-02-06

    (274 days)

    Product Code
    Regulation Number
    884.5160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Spectra S1 Pro and Spectra S2 Pro are single-user, powered breast pumps intended to express and collect milk from the breasts of lactating women.

    Device Description

    The Spectra S1 Pro and Spectra S2 Pro are single-user, powered breast pumps intended to express and collect milk from the breasts of lactating women. Pumping can be performed on one breast (single pumping) or both breasts (double pumping) at the same time. The Spectra S1 Pro and Spectra S2 Pro are comprised of the following components: pump unit, adapter, wide breast shield, back-flow protector, tubing, silicone valve, bottle cap, bottle screw, nipple, airtight cap, and PP bottle. The device is provided non-sterile.

    The Spectra S1 Pro and Spectra S2 Pro support a single pumping mode in which only one breast is expressed and a dual pumping mode in which both breasts are expressed. The user can switch between Massage Mode and Expression Mode using designated buttons and adjust the vacuum level and cycle speed within each mode.

    Spectra S1 Pro and Spectra S2 Pro includes the following features:

    • Expression can be performed on one breast only (single mode), on both sides simultaneously (dual mode).
    • Expression mode: 1-15 level (50-270 mmHg), cycle speeds (38, 42, 46, 50, 52, 58)
    • Massage mode: 1-5 level (50-180 mmHg), cycle speeds (70, 80, 90, 100, 105)

    The Spectra S1 Pro operates using an AC adapter (100-240V AC, 50/60Hz, DC 12V) or a rechargeable Li-Polymer battery (11.1V, 2000mAh). The Spectra S2 Pro operates using an AC adapter (100-240V AC, 50/60Hz, DC 12V) only.

    When the backflow protector is assembled between the pump unit and the breast shield, the silicone membrane inside the backflow protector creates a physical barrier by preventing air and milk from flowing back into the pump unit.

    All milk contacting components are compliant with 21 CFR 177.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252087

    Validate with FDA (Live)

    Date Cleared
    2026-02-03

    (216 days)

    Product Code
    Regulation Number
    882.4560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Navigation Module of the Cortium® System is intended to be used during the preparation and placement of Cortium® Universal OCT Spinal Fixation System screws during spinal surgery, to assist the surgeon in precisely locating anatomical structures. The Navigation Module of the Cortium® System is designed for use with the Medtronic StealthStation System, which is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as a vertebra, can be identified relative to a CT or MRI based model, fluoroscopy images, or digitized landmarks for the anatomy. Instruments of the Navigation Module of the Cortium® System are not intended for navigation of occipital screws.

    Device Description

    The Navigation Module of the Cortium® System contains non-sterile, reusable, manual surgical instruments including drills, taps, and screwdrivers. The instruments are manufactured from stainless steels which conform to ASTM F899. The instruments are designed for use with the Medtronic StealthStation surgical navigation system for implantation of the Cortium® Universal OCT Spinal Fixation System components. Instruments of the Navigation Module of the Cortium® System were tested for compatibility using StealthStation S8 software version 2.1.0.

    Instruments of the Navigation Module of the Cortium® System are not intended for navigation of occipital screws.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251354

    Validate with FDA (Live)

    Device Name
    UV Smart D60
    Date Cleared
    2026-01-29

    (274 days)

    Product Code
    Regulation Number
    880.6511
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UV Smart D60 is indicated for use in a healthcare environment to achieve a high level disinfection (HLD) of external surfaces of Transesophageal Echocardiogram (TEE) probes that do not contain lumens and that do not contain indentations or channels that are deeper than their widths.

    Device Description

    The UV Smart D60 provides chemical-free high level disinfection using Ultraviolet-C (UV-C) light within a fully enclosed disinfection chamber. The system features eight UV-C lamps, UV-C-reflective chamber walls, and UV-C-transmitting instrument holders designed to suspend the intended load in an optimal position. The fixed disinfection cycle time of 120 seconds is the only critical determinant of disinfection efficacy. The disinfection efficacy is verified during each cycle by two independent threshold sensors—one for UV-C dose and one for power consumption.

    The UV Smart D60 is intended to be used in healthcare facilities by trained personnel. Prior to disinfection, each soiled instrument must be cleaned at the bedside and manually cleaned in accordance with the original equipment manufacturer (OEM) instructions for use. Under normal operating conditions, the UV Smart D60 has an expected service life of ten years, with preventive maintenance and lamp replacement required annually, as specified in the UV Smart D60's Instructions for Use.

    To ensure that the intended load is both physically compatible with the disinfection chamber and capable of receiving the required HLD UV-C dose (Minimum Effective Dose or MED) on its surfaces, the intended load is approved for use with the UV Smart D60. Once approved, the intended load is released for use in the UV Smart D60's traceability system. The UV Smart D60 will not initiate a disinfection cycle unless a registered instrument is recognized, ensuring that only approved instruments can be processed.

    Each disinfection cycle accommodates a single instrument, which is suspended entirely within the D60's disinfection chamber—ensuring that all surfaces, including handles and attached cables, receive the MED for validated high level disinfection.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K253964

    Validate with FDA (Live)

    Date Cleared
    2026-01-09

    (29 days)

    Product Code
    Regulation Number
    870.1120
    Age Range
    0 - 0.08
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Unimed disposable blood pressure cuff (Models U1681S-C51N to U1685S-C51N) is an accessory used in conjunction with noninvasive blood pressure measurement systems. It is non-sterile and for single-patient use. The cuff is available in neonatal sizes (3-15 cm limb circumference). The cuff is not designed, sold, or intended for use except as indicated.

    Device Description

    The Unimed Disposable Neonatal NIBP Cuff (Model U1681S-C51N) is a single-patient-use accessory for non-invasive blood pressure (NIBP) measurement. The cuff is designed for neonatal limb circumferences ranging from 3 cm to 6 cm. The cuff is intended to be used with compatible blood pressure monitors employing oscillometry. It is non-sterile, and is disposed of after single-patient use to reduce cross-contamination risks.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K253281

    Validate with FDA (Live)

    Device Name
    UpDoc
    Manufacturer
    Date Cleared
    2025-12-23

    (85 days)

    Product Code
    Regulation Number
    868.1890
    Age Range
    18 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    UpDoc is a software as a medical device (SaMD) intended to provide medication management for patients aged 18 years and older who have been diagnosed with type 2 diabetes.

    UpDoc provides patients with insulin treatment plan instructions based on a healthcare provider (HCP)-specified treatment plan.

    UpDoc contains two user-interactive software components:

    Patient User Interface (UpDoc mobile application): Intended for use by patients with type 2 diabetes as an aid in optimizing insulin management. Patients use the mobile application to log blood glucose, meal, symptom, and medication adherence data, and receive treatment plan instructions. Data may be entered manually or reported via voice or text-based interactions. The application may also receive blood glucose data via a Bluetooth-enabled glucometer or continuous glucose monitor.

    HCP User Interface (UpDoc web portal): Intended for use by trained healthcare providers to configure and manage the patient-specific insulin treatment plan. This includes insulin dosing instructions (type, starting and maximum doses, adjustment algorithm, and blood glucose targets) and safety protocols to address non-emergency hypoglycemia, hyperglycemia, and related symptoms.

    Insulin instructions are computed in UpDoc's cloud-based application based on the HCP-defined treatment parameters.

    Device Description

    UpDoc is a software as a medical device (SaMD) designed to assist patients aged 18 years and older with insulin management for type 2 diabetes. Healthcare providers (HCPs) set an individualized treatment plan for their patients that includes monitoring and insulin titration instructions. UpDoc engages with patients to help them follow their designated treatment plan and supports HCPs in monitoring reported health data, medication adherence, and treatment progress.

    UpDoc is composed of three modular software components: a provider-facing web portal (UpDoc Provider Portal), a patient mobile application (UpDoc Patient App), and a cloud-based application consisting of a Conversation Service (UpDoc Agent) and a Clinical Service. These components work together to support safe and effective provider-directed insulin therapy.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251011

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-12-19

    (262 days)

    Product Code
    Regulation Number
    884.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Chiavaye personal moisturizer is a lubricant for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. Chiavaye personal moisturizer is not compatible with natural rubber latex, polyurethane, or polyisoprene condoms.

    Device Description

    Chiavaye Personal Moisturizer is a clear to slightly yellow and odorless oil-based personal lubricant that is not compatible with condoms made of natural rubber latex, polyisoprene, or polyurethane. This device is a non-sterile personal lubricant for penile and/or vaginal application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication.

    This device is sold as an over-the-counter (OTC) product in 30 mL and 100 mL sizes provided in black 'MIRON' glass, with a Polypropylene (PP) protective cap. The pump cap is made of Polypropylene (actuator), Nitrile (seal), and Polyethylene (tube).

    The device is composed of Almond Oil, Coconut Oil, Grape Seed Oil, Sunflower Oil, Evening Primrose Oil, and Vitamin E.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K252235

    Validate with FDA (Live)

    Device Name
    PVAD IQ Software
    Manufacturer
    Date Cleared
    2025-12-18

    (154 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    18 - 120
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PVAD IQ software is intended for non-invasive analysis of ultrasound images to detect and measure structures from cardiac ultrasound of patients 18 years old and above, with a Percutaneous Ventricular Assist Device (PVAD). Such use is typically utilized for clinical decision support by a qualified physician.

    Device Description

    PVAD IQ is a Software as a Medical Device (SaMD) solution designed to support clinicians in the positioning of Percutaneous Ventricular Assist devices (PVADs) through ultrasound image-based assessment. Percutaneous Ventricular Assist device is a temporary device used to provide hemodynamic support for patients experiencing cardiogenic shock or undergoing high-risk percutaneous coronary interventions (PCI).

    The PVAD IQ software is a machine learning model (MLM) based software, that operates on ultrasound clips (as the system input) and provides two outputs with regards to PVAD patients:

    1. Landmark identification and measurement - provides the two landmarks position detection, and computation of the mean distance between the two landmarks- the aortic annulus and the PVAD inlet.

    2. Acceptability classification, which is a binary classification of ultrasound clips that are "acceptable" or "non-acceptable", in terms of visibility of the two landmarks. A clip is defined as acceptable when both landmarks are simultaneously visible in a manner suitable for quantitative imaging.

    The User Interface (UI) enables the user to review or hide the mean distance measurement, annotate desired images, and add manual measurement, while keeping the raw data for further review as needed.

    The software output is shown on the screen either as the mean distance measurement, or as a notification related to non-acceptable clips.

    AI/ML Overview

    The PVAD IQ Software, a machine learning model (MLM) based software, provides two primary outputs for patients with Percutaneous Ventricular Assist Devices (PVADs): landmark identification and measurement (specifically, the distance between the aortic annulus and the PVAD inlet) and acceptability classification of ultrasound clips.

    1. Acceptance Criteria and Reported Device Performance

    The study established pre-specified acceptance criteria for the PVAD IQ software's performance, which it met.

    Acceptance CriteriaThresholdReported Device Performance
    Distance Measurement (MAE)Below 0.5 cm0.42 cm (95% CI: 0.38–0.47 cm)
    Acceptability Classification (Cohen's Kappa)Above 0.60.71 (95% CI: 0.66–0.75)
    Landmark Detection (AUC) - PVAD InletAbove 0.80.92 (0.9–0.94)
    Landmark Detection (AUC) - Aortic AnnulusAbove 0.80.98 (0.95, 1)
    Landmark Position (MAE) - PVAD InletBelow 0.5 cm0.44 cm (0.41–0.48 cm)
    Landmark Position (MAE) - Aortic AnnulusBelow 0.5 cm0.31 cm (0.3–0.33 cm)

    2. Sample Size and Data Provenance for Test Set

    • Sample Size: 963 clips
    • Number of Patients: 186 patients
    • Data Provenance: Geographically distinct test datasets. While specific countries are not mentioned, the ground truth annotations were provided by US (United States) board certified cardiac sonographers. The timing (retrospective or prospective) is not specified, but the data was used for evaluating a previously trained model, which often implies a retrospective application to a held-out test set.

    3. Number and Qualifications of Experts for Ground Truth (Test Set)

    • Number of Experts: Not explicitly stated as a specific number of individual experts. The document refers to "US (United States) board certified cardiac sonographers."
    • Qualifications of Experts: "US (United States) board certified cardiac sonographers experienced in PVAD/Impella® echocardiographic imaging."

    4. Adjudication Method for Test Set

    The adjudication method is not explicitly stated in the provided document. It only mentions that ground truth annotations were "provided by US (United States) board certified cardiac sonographers." It does not specify if multiple sonographers reviewed each case, how disagreements were resolved, or if a consensus mechanism (like 2+1 or 3+1) was used.

    5. MRMC Comparative Effectiveness Study

    An MRMC (Multi-Reader Multi-Case) comparative effectiveness study comparing AI assistance with unassisted human readers was not mentioned in the provided document. The study focused on the standalone performance of the PVAD IQ software.

    6. Standalone Performance Study

    Yes, a standalone (algorithm only without human-in-the-loop performance) study was conducted. The reported performance metrics (MAE, Cohen's Kappa, AUC) directly assess the algorithm's performance against the established ground truth.

    7. Type of Ground Truth Used

    The ground truth used was expert consensus/annotations. Specifically, "Ground truth annotations for the distance between the aortic annulus and the PVAD inlet were provided by US (United States) board certified cardiac sonographers experienced in PVAD/Impella® echocardiographic imaging." This implies human experts manually defining the "correct" measurements and classifications.

    8. Sample Size for the Training Set

    The sample size for the training set is not provided in this document. The document states that the PVAD IQ software is "trained with clinical data" but does not specify the volume or characteristics of this training data.

    9. How Ground Truth for Training Set Was Established

    The method for establishing ground truth for the training set is not explicitly detailed in this document. It broadly states that the software uses "non-adaptive machine learning algorithms trained with clinical data" and "refining annotations" is part of model retraining (under PCCP). While it can be inferred that ground truth for training data would also involve expert annotations, similar to the test set, the specific process, number of experts, or their qualifications for the training data are not provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K251416

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-12-17

    (224 days)

    Product Code
    Regulation Number
    892.2100
    Age Range
    18 - 91
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UltraSight Guidance is intended to assist medical professionals (not including expert sonographers) in acquiring cardiac ultrasound images. UltraSight Guidance is an accessory to compatible general-purpose diagnostic ultrasound systems. UltraSight Guidance is indicated for use in two-dimensional transthoracic echocardiography (2D-TTE) for adult patients, specifically in the acquisition of the following standard views: Parasternal Long-Axis (PLAX), Parasternal Short-Axis at the Aortic Valve (PSAX-AV), Parasternal Short-Axis at the Mitral Valve (PSAX-MV), Parasternal Short Axis at the Papillary Muscle (PSAX-PM), Apical 4-Chamber (AP4), Apical 5-Chamber (AP5), Apical 2-Chamber (AP2), Apical 3-Chamber (AP3), Subcostal 4-Chamber (SubC4), and Subcostal Inferior Vena Cava (SC-IVC).

    Device Description

    UltraSight Guidance is a software application based on machine learning that uses artificial intelligence to provide dynamic real-time guidance on the position and/or orientation of the transducer to help non-expert users acquire diagnostic quality tomographic views of the heart. The system provides guidance for ten standard cardiac views.

    Main features:
    • Quality Bar: The system displays an image quality bar that is continuously updated while the user scans the subject, while attempting to find the maximal quality. The quality bar is a score for image diagnosability. It represents the classification between high and low quality images, where high quality images are defined as grade 3 or more based on American College of Emergency Physicians (ACEP) guidelines (Rachel B. Liu et al., "Emergency Ultrasound Standard Reporting Guidelines", 2018, American College of Emergency Physicians)
    • Probe Guidance: The probe guidance feature provides graphic on-screen instructions for the user to emulate how a sonographer would manipulate the transducer to acquire the target cardiac view. The five possible guidance cues are rotation, tilt, rock, and slides in the lateral-medial and up/down directions (with respect to the subjects' head).

    The guidance user interface (UI) is composed of a 3D probe display that shows orientation guidance cues (rotations, tilts and rocks) and a cross that shows slide guidance cues (slides in x and y directions). The users infer the maneuver they should perform from viewing the 3D probe display and the slides cross. Supporting text messages appear on the screen regarding guidance cues.

    This device is a modification of a previously marketed device, where the main modification introduced is the addition of a pre-processing step done prior to feeding images as an input to the device. This modification is aimed to enable future compatibility with potentially additional ultrasound probes that meet pre-requisites as part of a pre-determined change control plan.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the UltraSight Guidance device, based on the provided FDA 510(k) clearance letter:

    UltraSight Guidance: Acceptance Criteria and Study Details

    The UltraSight Guidance device focuses on two main features: the "Quality Bar" and "Probe Guidance." The performance testing for each is described below.

    1. Table of Acceptance Criteria and Reported Device Performance

    FeatureAcceptance CriteriaReported Device Performance
    Quality BarMean Area Under the ROC Curve (AUC) > 0.8Mean AUC was within the pre-defined acceptance criteria of AUC > 0.8 with 95% CI showing good classification performance relative to the success criteria. (p. 10)
    Mean False Positive Rate (FPR) < 0.2The mean false positive rate (FPR) met the acceptance criteria of FPR < 0.2 with 95% CI showing good classification performance. (p. 10)
    Probe GuidanceMean Area Under the ROC Curve (AUC) > 0.8 (for each guidance cue)The mean AUC was within the acceptance criteria of AUC > 0.8 with 95% CI showing good classification performance. (p. 13) (Note: This is stated for each guidance cue, implying individual performance met this criterion, though not explicitly broken down per cue in the summary).

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size:
      • Quality Bar: 134 patients, comprising 26,362 ultrasound clips.
      • Probe Guidance: 134 patients, comprising 2.4 million ultrasound frames.
    • Data Provenance: The test dataset was collected from clinical sites geographically distinct from those used for the development dataset.
      • Countries of Origin: 111 patients from the US and 23 patients from Israel.
      • Retrospective/Prospective: Not explicitly stated as retrospective or prospective, but the description of "collected from clinical sites" and "comprising [number] ultrasound clips/frames" suggests it was likely retrospective collection of existing data or data collected for the study.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Quality Bar: Two expert cardiologists initially annotated the clips for diagnosability.
    • Probe Guidance: Ground truth was established by expert sonographers and/or cardiologists.
    • Qualifications of Experts: The document explicitly states "expert cardiologists" and "expert sonographers and/or cardiologists," implying recognized professionals in their respective fields, but does not provide specific details on years of experience or board certifications.

    4. Adjudication Method for the Test Set

    • Quality Bar: 2+1 Adjudication. Initially, two expert cardiologists provided annotations. In case of disagreement, a third cardiologist provided additional annotation, and the final label was determined by majority vote among the three experts. (p. 9-10)
    • Probe Guidance: Not explicitly detailed. The statement "Ground truth was established by expert sonographers and/or cardiologists" (p. 13) does not specify an adjudication method for potential disagreements among multiple experts, or if a single expert reviewed each case.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    • No MRMC comparative effectiveness study was specifically described for the initial clearance in the provided text. The testing focuses on the standalone performance of the algorithm (AI model) for the Quality Bar and Probe Guidance features.
    • It does mention a "Clinical utility test" as a potential future testing method as part of the PCCP for "Expanded list of cardiac views": "Clinical utility test will evaluate the ability of non-expert users to obtain diagnostic-quality cardiac images using UltraSight software for the newly added cardiac views." (p. 12). This might involve comparing performance with and without the device, but it is for future modifications, not the current clearance, and details are scarce.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    • Yes, standalone performance (algorithm only) was done for both the Quality Bar and Probe Guidance features. The studies describe the classification performance of the algorithm against expert-established ground truth without human users interacting with the device for the testing itself. For example, for the Quality Bar, the device's classification of "diagnosable/non-diagnosable" was compared to expert labels. For Probe Guidance, the algorithm's prediction of guidance cues was compared to expert-defined optimal probe positions.

    7. The Type of Ground Truth Used

    • Quality Bar: Expert Consensus (diagnosability label). The ground truth ("diagnosable / non-diagnosable") was established by two expert cardiologists, with a third cardiologist resolving disagreements through majority vote. (p. 9-10)
    • Probe Guidance: Expert-established optimal probe position/orientation. Ground truth was established by "expert sonographers and/or cardiologists using the recorded probe position during ultrasound acquisition." This defined the "required rotation, tilt, rock and slides adjustments" against which the algorithm's guidance was compared. (p. 13)

    8. The Sample Size for the Training Set

    • The document states that the test dataset was "collected from clinical sites geographically distinct from those used for the development dataset" (p. 9). However, the specific sample size of the training (development) set is not provided in the clearance letter summary.

    9. How the Ground Truth for the Training Set Was Established

    • Similar to the training set sample size, the method for establishing ground truth for the training (development) set is not explicitly detailed in the provided text. It is generally assumed that similar expert annotation processes would have been followed for the training data, but the document does not specify this.
    Ask a Question

    Ask a specific question about this device

    K Number
    K253719

    Validate with FDA (Live)

    Date Cleared
    2025-12-17

    (23 days)

    Product Code
    Regulation Number
    884.5470
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Unscented Menstrual Tampons are inserted into the vagina to absorb menstrual discharge.

    Device Description

    The proposed unscented menstrual tampons are cylindrical devices designed to absorb menstrual blood during a women's period. They consist of a tampon, including an absorbent pledget ("absorbent core") completely surrounded by an overwrap ("security veil") and a removal string ("withdrawal cord"), and an applicator (except for the digital style). The tampon design is cylindrical, bullet-like shape. The applicator has a smooth, rounded tip for ease of insertion. According to different composition materials and design styles, the unscented menstrual tampons of this submission are divided into 3 categories (Unscented menstrual long applicator tampon, Unscented menstrual Cardboard Applicator Tampon, Unscented menstrual Digital Tampons) and 7 sub-categories (Cotton + cardboard tube, Organic cotton + cardboard tube, Cotton (digital style), Organic cotton(digital style), viscose (digital style), Cotton + plastic tube and Viscose + cardboard tube). For applicator-style tampons, they feature a built-in applicator made of plastic or cardboard that helps the users insert the tampons into the vagina by pushing the tampon out of the applicator. The digital style tampons consist of a plain tampon designed to be inserted manually using the fingers.

    The tampons are provided in 4 absorbencies: Light(L) (≤6g), Regular(R) (6-9g), Super (S)(9-12g) and Super Plus (SP)(12-15g). For Cotton + cardboard tube and Organic cotton + cardboard tube, L and SP are applicable for this application. For Cotton (digital style), Organic cotton (digital style) and Viscose (digital style), only L is applicable. For Cotton + plastic tube and Viscose + cardboard tube, all those four absorbencies are applicable for this application.

    Each device is individually wrapped and then packaged in sealed multi-unit containers for retail sale. All tampons and applicators are provided non-sterile and for single use only.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    Page 1 of 114