Found 2 results

510(k) Data Aggregation

    K Number
    K053656

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-07-14

    (196 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    PivitAB™ is intended for the repair of hernias and other abdominal fascial deficiencies requiring the addition of a reinforcing or bridging material to obtain the desired surgical result. The resorbable polymer coating on the mesh contains the antimicrobial agents, rifampin and minocycline to help provide protection from microbial colonization of the device during surgical implantation. Examples of applications where PivitAB™ may be used include, but are not limited to: inguinal, femoral, umbilical, abdominal, incisional and intramuscular hernias and muscle flap reinforcement.

    Device Description

    PivitAB™ is dual component (resorbable and non-resorbable), sterile prosthesis designed for the reconstruction of soft tissue deficiencies. PivitAB™ is constructed of a nonresorbable mesh comprised of knitted filaments of polypropylene and a biorsonable polyarylate coating on the mesh containing the antimicrobial agents, rifampir and minocycline. The purpose of the resorbable coating is to provide additional stiffness to the nesh in order to facilitate interoperative handling during placement and act a carrier for the imicrobial agents. Once placed, the polymer resorbs in approximately 90 days leaving a ermanent mesh incorporated into the tissue.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the PivitAB™ surgical mesh. This document focuses on demonstrating the substantial equivalence of the PivitAB™ to legally marketed predicate devices, rather than establishing specific acceptance criteria and proving performance through a study designed to meet those criteria.

    Therefore, many of the requested categories are not applicable or cannot be extracted from this type of regulatory submission. The 510(k) summarizes existing data to show equivalence, rather than detailing a de novo study to establish new performance metrics against pre-defined acceptance criteria.

    Here's an attempt to address your request based on the provided text, with explanations for what cannot be found:

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/TestAcceptance CriteriaReported Device Performance
    Biocompatibility(Implicitly, comparable to predicate devices and acceptable for human implantation as per FDA guidance for surgical mesh.)"The PivitAB™ and the polyarylate polymer have been demonstrated to be biocompatible."
    Tissue Ingrowth(Implicitly, satisfactory and comparable to commercial surgical mesh, as per FDA guidance for surgical mesh.)"animal testing demonstrated that PivitAB™ will achieve satisfactory tissue ingrowth compared to control as evidenced by histopathology."
    Antimicrobial Effectiveness (In Vivo)(Implicitly, significantly fewer colonized implants compared to a non-antimicrobial control mesh.)"Result demonstrated that there were significantly fewer colonized implants with PivitAB (13%) than with the PROLENE Mesh control (43%)."
    Mechanical Properties / Hernia Repair Performance(Implicitly, "comparable to standard surgical mesh devices that are indicated for hernia repair and other abdominal fascial or muscular deficiencies" as per FDA guidance for surgical mesh.)"Non-clinical laboratory testing was performed in accordance with the FDA guidance document 'Guidance for the Preparation of a Premarket Notification for a Surgical Mesh' demonstrating that the device is comparable to standard surgical mesh devices that are indicated for hernia repair and other abdominal fascial or muscular deficiencies."

    Explanation: The 510(k) submission doesn't explicitly state numerical acceptance criteria in the way a clinical trial protocol would. Instead, it relies on demonstrating comparability to predicate devices and adherence to relevant FDA guidance for surgical mesh. The "reported device performance" is a summary of findings from various non-clinical and animal tests designed to support this comparability and safety.

    2. Sample size used for the test set and the data provenance

    • Test Set Sample Size:
      • For the antimicrobial effectiveness study: The text states "PivitAB™ and a non-antimicrobial coated Control (PROLENE, Ethicon, Sommerville, NJ) were inoculated with 10^5 CFUs of S aureus and implanted into rabbits." While it provides the result (13% vs 43%), it does not specify the number of rabbits or implants used in this study.
      • For tissue ingrowth and other non-clinical tests: Not specified.
    • Data Provenance: The studies mentioned are referred to as "Non-clinical laboratory testing" and "animal testing."
      • Country of Origin: Not specified in the provided text.
      • Retrospective or Prospective: These would typically be prospective studies, especially the animal model for antimicrobial effectiveness.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This information is not provided in the 510(k) summary. "Histopathology" for tissue ingrowth implies expert review by pathologists, but details on their number or qualifications are absent. For the antimicrobial effectiveness, the "ground truth" was the percentage of colonized implants, likely determined through microbiological culture and counting, not expert consensus on an image or clinical observation.

    4. Adjudication method for the test set

    • This information is not explicitly provided. For the antimicrobial effectiveness study, the method was likely objective measurement (microbial counts), not an adjudication process among human observers.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices (e.g., AI for medical imaging) where human readers interpret data. The PivitAB™ is a surgical mesh; its effectiveness is assessed through physical, biological, and antimicrobial performance, not human interpretation of diagnostic output. Therefore, there's no "AI assistance" or "human reader improvement" to measure.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. PivitAB™ is a physical surgical implant, not an algorithm. Its performance is inherent to the device itself (materials, design, antimicrobial properties).

    7. The type of ground truth used

    • For Biocompatibility: Likely based on standardized assays and histopathological examination in animal models, comparing responses to established benign materials.
    • For Tissue Ingrowth: "Histopathology" is mentioned, indicating microscopic examination of tissue sections as the ground truth.
    • For Antimicrobial Effectiveness: The "ground truth" was the presence and extent of microbial colonization on the implants, quantified through microbiological methods (e.g., CFUs - Colony Forming Units) from the explanted devices. This is an objective, laboratory-derived ground truth.
    • For Mechanical Properties: Likely engineering tests and physical measurements according to standards.

    8. The sample size for the training set

    • Not applicable as this is not an AI/ML device that requires a training set. The data presented are from pre-clinical studies and animal models.

    9. How the ground truth for the training set was established

    • Not applicable, as there is no training set for this type of medical device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K052864

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-12-22

    (72 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Is intended for the repair of hernias and other abdominal fascial deficioncies requiring the addition of a reinforcing or bridging material to obtain the desired surgical result. . Examples of applications where TyRx Surgical Mesh may be used include, but are not limited to: inguinal, femoral, umbilical, abdominal, incisional and intramuscular hernias and muscle flap reinforcement.

    Device Description

    TyRx Surgical Mesh is dual component (resorbable and non-resorbable), sterile prosthesis designed for the reconstruction of soft tissue deficiencies. TyRx Surgical Mesh is constructed of a non-resorbable mesh comprised of knitted filaments of polypropylene and a bioresorbable polyarylate coating on the mesh. The resorbable coating represents approximately 10% of the total weight of the device. The purpose of the resorbable coating is to provide additional stiffness to the mesh in order to facilitate interoperative handling during placement. Once placed, the polymer resorbs in approximately 90 days leaving a lighter permanent mesh incorporated into the tissue.

    AI/ML Overview

    This document describes a 510(k) premarket notification for a new surgical mesh and does not contain detailed information about specific acceptance criteria or an analytical study with performance metrics in the way a diagnostic algorithm or AI device submission would. The submission focuses on demonstrating substantial equivalence to predicate devices for its intended use.

    However, I can extract the relevant information that is available and indicate when specific details are not provided in the document.

    1. A table of acceptance criteria and the reported device performance

    Based on the provided text, specific numerical acceptance criteria and corresponding performance values for the TyRx Surgical Mesh are not explicitly stated in terms of a table with quantitative metrics. The submission focuses on demonstrating comparability and biocompatibility.

    Acceptance Criteria (Implied)Reported Device Performance
    BiocompatibilityDemonstrated to be biocompatible (TyRx Surgical Mesh and polyarylate polymer)
    Satisfactory tissue ingrowthAnimal testing demonstrated satisfactory tissue ingrowth compared to control, evidenced by histopathology.
    Comparability to predicate devices for intended useNon-clinical laboratory testing performed in accordance with FDA guidance document "Guidance for the Preparation of a Premarket Notification for a Surgical Mesh" demonstrated that the device is comparable to standard surgical mesh devices.
    Proper handling (stiffness facilitation)Resorbable coating provides additional stiffness to the mesh to facilitate intraoperative handling during placement.
    Resorption propertiesPolymer resorbs in approximately 90 days.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Test Set Sample Size: The document only mentions "animal testing" and "non-clinical laboratory testing." It does not specify the sample size (number of animals or test replicates) for either of these.
    • Data Provenance: The document does not specify the country of origin of the data. It indicates the company is based in Monmouth Junction, NJ, USA, but this doesn't confirm where the testing took place. The testing was "non-clinical laboratory testing" and "animal testing," which are typically considered prospective in nature for evaluating a new device.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document mentions "histopathology" for evaluating tissue ingrowth. This implies expert assessment. However, the number of experts and their specific qualifications (e.g., "veterinary pathologist with 10 years of experience") are not specified in the provided text.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    The document does not provide any information regarding an adjudication method for establishing ground truth or evaluating the test results.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study Done? No. This device is a surgical mesh, not a diagnostic imaging device or an AI-assisted diagnostic tool. Therefore, a multi-reader multi-case (MRMC) comparative effectiveness study as described (human readers with/without AI assistance) is not applicable and was not performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance? No. This device is a physical surgical mesh, not an algorithm. Therefore, "standalone (algorithm only)" performance is not applicable and was not performed.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the animal testing evaluating tissue ingrowth, the ground truth was established through histopathology, which typically involves examination and interpretation by a pathologist.

    8. The sample size for the training set

    This is not applicable as the device is a physical surgical mesh and not an algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This is not applicable as the device is a physical surgical mesh and not an algorithm that requires a training set.