Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K070589

    Validate with FDA (Live)

    Device Name
    C!
    Date Cleared
    2007-04-24

    (54 days)

    Product Code
    Regulation Number
    892.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    For use in the acquisition of Single Photon Emission Computed Tomography (SPECT) and Planar gamma camera images.

    Device Description

    The C! SPECT Imaging System is an open gantry (with no external moving parts), upright seated device with an integrated patient chair, an acquisition station, two detectors, and hardware and software components. The patient chair is mechanized to accommodate patient loading and centering in the detector field of view.

    AI/ML Overview

    This document describes the C! SPECT Imaging System. Here's a breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Stated)Reported Device Performance (as per submission)
    Performance met or exceeded predicate devicesPerformance of the C! device met or exceeded that of predicate devices (using NEMA NU1 phantoms and NEMA Standard test protocols).
    Tomographic image quality at least equal to predicate devicesClinical images obtained in human subjects demonstrated tomographic image quality at least equal to images produced by reference predicate devices.
    Electrical safety met applicable standardsElectrical safety testing has been performed and found to meet applicable standards and defined acceptance criteria.

    2. Sample Size Used for the Test Set and Data Provenance

    The document mentions "Clinical images were obtained using the C! in human subjects," but it does not specify the sample size for this clinical test set.

    The data provenance is prospective as clinical images were "obtained ... in human subjects." The document does not specify the country of origin of the data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    The document does not provide information on the number of experts used, nor their qualifications, for evaluating the clinical images. The evaluation states "Tomographic image quality was at least equal to images produced by reference predicate devices," implying an expert comparison, but details are absent.

    4. Adjudication Method for the Test Set

    The document does not specify any adjudication method for the clinical image evaluation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study, nor any effect size for human reader improvement with or without AI assistance.

    6. If a Standalone (Algorithm Only) Performance Study Was Done

    The C! SPECT Imaging System is a hardware and software system for image acquisition. The performance testing described focuses on the overall system's image quality and does not specifically detail a standalone algorithm-only performance assessment separate from the imaging system. The performance testing refers to the "C! device" as a whole.

    7. The Type of Ground Truth Used

    For the phantom studies, the ground truth is based on the known physical properties of NEMA NU1 phantoms and NEMA Standard test protocols, which provide objective benchmarks.

    For the clinical images, the ground truth appears to be a qualitative comparison to "reference predicate devices" by unspecified evaluators. It is not explicitly stated if this was expert consensus, pathology, or outcomes data.

    8. The Sample Size for the Training Set

    The document does not mention a training set or its sample size. The device is a SPECT imaging system, and its performance evaluation focuses on its ability to acquire images, not on an algorithm trained on a dataset.

    9. How the Ground Truth for the Training Set Was Established

    Since no training set is mentioned (as the device is an imaging system, not an AI algorithm trained on data for interpretation), this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K033001

    Validate with FDA (Live)

    Date Cleared
    2003-12-24

    (90 days)

    Product Code
    Regulation Number
    892.1200
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Apex XL-4 is intended for use as an acquisition of Nuclear Medicine SPECT and Planar Gamma Camera images.

    Device Description

    The APEX XL-4 is a Gamma Camera - SPECT Imaging System. It is intended to acquire nuclear medicine gamma camera output data for previously-approved Elscint APEX Gamma Camera systems. This acquisition is performed in the same manner as the corresponding components of the equivalent device, K021823. The detector electronics, mechanical components, software, and the manufacturing specifications of the applicant system (the APEX XL-4) are substantially equivalent to the CardioSPECT SC, the legally marketed device.

    AI/ML Overview

    The provided document for K033001 describes the Photon2 APEX XL-4, a gamma camera - SPECT imaging system. This 510(k) submission is for demonstrating substantial equivalence to a predicate device, not for establishing novel performance criteria through a new clinical study. Therefore, the information requested about specific acceptance criteria, device performance tables, sample sizes for test sets, expert involvement, and comparative effectiveness studies as typically presented for AI/ML device submissions is not applicable in this context.

    This 510(k) relies on demonstrating that the new device has substantially equivalent specifications, safety, effectiveness, and intended use as the legally marketed predicate device, the NC Systems, Inc. CardioSPECT SC (K021823). The document primarily focuses on comparing technical specifications and intended use.

    Here's a breakdown of what can be extracted from the document based on your request, with an emphasis on what is not present:

    1. A table of acceptance criteria and the reported device performance

    • Not Applicable. This type of table is typically found in submissions for novel devices or those with new performance claims where specific metrics (e.g., sensitivity, specificity, AUC) are being established against a defined ground truth. For a substantial equivalence claim for a gamma camera, the acceptance criteria are generally met by demonstrating that the device meets existing performance standards and is comparable to the predicate device across various technical and functional specifications. The document states: "The conclusion of tests performed and clinical images obtained demonstrate that the subject device (APEX XL-4) is as safe and effective as the legally marketed device (CardioSPECT SC)." However, no specific performance metrics or acceptance criteria are presented in a table form.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not Applicable/Not Provided. The document does not describe a specific "test set" in the context of an AI/ML algorithm evaluation. The "tests performed and clinical images obtained" mentioned are likely part of the validation process to ensure the device performs as expected and similarly to the predicate, but detailed information about the number of patients/cases, their origin, or study design (retrospective/prospective) is not provided.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not Applicable/Not Provided. Since there's no described "test set" and ground truth establishment in the context of AI/ML performance evaluation, there's no mention of experts or their qualifications for this purpose.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not Applicable/Not Provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Applicable. This device is a gamma camera hardware system, not an AI-assisted diagnostic tool. Therefore, an MRMC study comparing human readers with AI assistance is not relevant to this submission.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Applicable. This is a hardware device; no standalone algorithm performance is being evaluated.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    • Not Applicable/Not Provided. As there is no "test set" requiring ground truth for performance evaluation in the AI/ML sense, no specific type of ground truth is mentioned. The comparison is based on the technical specifications and functional equivalence to the predicate device.

    8. The sample size for the training set

    • Not Applicable/Not Provided. This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established

    • Not Applicable/Not Provided.

    Summary of the study that proves the device meets the acceptance criteria (as understood in a 510(k) for a hardware device):

    The "study" presented is primarily a comparison of specifications and intended use between the APEX XL-4 and its predicate device, the NC Systems, Inc. CardioSPECT SC (K021823). The document states:

    • "The intended use of the applicable components in the two devices is identical."
    • "The specifications for each device are compared in Section C." (This section is referenced but not provided in the input, but implies detailed technical comparison).
    • "The labeling for each device are equivalent."
    • "The conclusion of tests performed and clinical images obtained demonstrate that the subject device (APEX XL-4) is as safe and effective as the legally marketed device (CardioSPECT SC)."

    This demonstrates that the device achieves substantial equivalence by meeting the implied "acceptance criteria" of being functionally and technically similar to a device that has already been cleared by the FDA for safety and effectiveness. The proof is in the direct comparison of the new device to the predicate across various attributes including design, materials, performance claims, and intended use as outlined in Section 6.0 "Determination of Substantial Equivalence." There is no detailed clinical study described with patient cohorts or statistical performance endpoints in the provided text.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1