Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K132701

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-09-26

    (28 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Primi™ Support Catheter is intended to be used in conjunction with guidewires in order to access discreet regions of the peripheral vasculature. It may be used to facilitate placement and exchange of guidewires and other interventional devices. The support catheter also may be used to selectively infuse/deliver physician specified diagnostic and/or therapeutic agents, such as sterile saline, contrast agents, vaso-dilators, anti-blastics, lytics, and anti-thrombotics, with or without a guidewire in position.

    Device Description

    The ThermopeutiX Support Catheter that is the submission is a single lumen catheter with a radiopaque marker that aids in positioning the catheter and a tapered tip to facilitate entry. The catheter utilizes an integral spine wire. The catheter is available in several diameters and lengths which are 0.014", 0.018", or 0.035" guidewire compatible. The working lengths are 90 cm thru 170 cm. Reference Table 1 for model number availability. The distal portion of the catheter shaft has a hydrophilic coating to facilitate insertion. The device has a luer hub connector to allow infusion of diagnostic and/or therapeutic agents. The device is supplied sterile and is intended for single use only. Reference Table 2 below for physical and performance characteristics.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the ThermopeutiX Primi™ Support Catheter with an advanced hydrophilic coating. This submission aims to demonstrate substantial equivalence to a previously cleared version of the same catheter (K130850) by ThermopeutiX, Inc., with the primary difference being the advanced hydrophilic coating.

    Here's an analysis of the acceptance criteria and study information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the physical and performance characteristics of the predicate device and the new device. The study demonstrates that the new device's performance is either identical or comparable, and meets internal protocol requirements and product specifications.

    Element/FeatureAcceptance Criteria (from Predicate Device)Reported Device Performance (New Device)
    Working Length (s)90 cm – 170 cm; polymer shaft with integral spine wire90 cm – 170 cm; polymer shaft with integral spine wire
    Outside Diameter (s)Distal:0.045" – 0.064"Proximal: 0.060" – 0.080"Distal:0.045" – 0.064"Proximal: 0.060" – 0.080"
    LumenOneOne
    Guide wire Compatibility0.014"-0.035"0.014"-0.035"
    Guide wire UseOver the WireOver the Wire
    Distal End ConfigurationTapered tip for ease of crossingTapered tip for ease of crossing
    Fluid connectionLuer hub connectorLuer hub connector
    Flow rate @ 150 psi w/Saline0.014" wire models > 0.80 cc/sec0.035" wire models > 0.50 cc/sec0.014" wire models > 0.80 cc/sec0.035" wire models > 0.50 cc/sec
    Flow rate @ 150 psi w/Contrast0.014" wire models > 0.30 cc/sec0.035" wire models > 0.15 cc/sec0.014" wire models > 0.30 cc/sec0.035" wire models > 0.15 cc/sec
    Maximum Injection PressureDo not exceed 300 psiDo not exceed 300 psi
    CoatingYes; hydrophilic, distal portionYes; advanced hydrophilic, distal portion (provides comparable lubricity with a reduction in particulates)
    Coating Length65 cm, 115 cm or 140 cm65 cm, 115 cm or 140 cm
    Coating ThicknessProprietarySame as predicate (considered equivalent)
    Radiopaque MarkersOneOne
    Sheath Compatibility5 Fr and 6 Fr5 Fr and 6 Fr
    Materials UsedIdentical list of materials as the new device, except for the coating.Hub-PebaxOuter Shaft-PebaxInner Shaft-HDPEDistal Shaft-Pebax w/radiopaque fillerCore Wire-Stainless Steel tapered wireCoating-Advanced hydrophilic proprietary coatingMarker Band Platinum/Iridium

    Study Proving Device Meets Acceptance Criteria:

    The study conducted is a non-clinical performance and safety study comparing the new device with the advanced hydrophilic coating to the predicate device.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify an exact numerical sample size for each individual test. It states "The non-clinical tests performed on sample devices are listed below," and "All performance testing was repeated on samples with a minimum of 10 month aging." This indicates that multiple samples were used for testing, including aged samples, but the exact number isn't quantified.

    • Sample Size for Test Set: Not explicitly stated as a number, but "sample devices" were used, including aged samples.
    • Data Provenance: The data is from in-vitro (bench) testing performed by the manufacturer, ThermopeutiX, Inc., in the United States. This is prospective testing performed for the purpose of this 510(k) submission.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    This type of submission (510(k) for a medical device with minor modification) typically relies on engineering and biological testing against established standards and product specifications, rather than expert interpretation of a test set in the way AI/ML algorithms use ground truth. Therefore, the concept of "experts establishing ground truth for the test set" is not applicable here. The "ground truth" is the objective physical and performance measurements determined by standardized testing protocols.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, this study is based on objective physical and performance measurements via bench testing, not subjective assessment requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging AI, not for performance testing of a physical medical device like a catheter. The aim here is to show "substantial equivalence" of device characteristics and safety, not to measure human reader improvement with AI.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. This is a physical medical device, not an algorithm or AI system. The "standalone" performance is the device's measured performance during bench testing without human intervention in the device's function.

    7. The Type of Ground Truth Used

    The ground truth used for this study is based on objective physical and performance measurements derived from standardized engineering and biocompatibility tests (e.g., flow rate, tensile strength, lubricity, cytotoxicity, etc.) and comparison to the known characteristics of the predicate device. It's essentially test results against predetermined specifications and accepted standards (e.g., ISO, ASTM).

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device, not an AI/ML algorithm that requires a training set. The device itself is the product, and its performance is being evaluated, not an algorithm's learning capability.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K130850

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-06-25

    (90 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K112219

    Validate with FDA (Live)

    Device Name
    TAPAS CATHETER
    Manufacturer
    Date Cleared
    2011-11-18

    (108 days)

    Product Code
    Regulation Number
    870.4450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ThermopeutiX TAPAS™ Catheter is intended for general intravascular use in the peripheral vasculature in arteries 1.8 mm and larger. Once placed in the selected region, the catheter can be used for infusion of diagnostic and/or therapeutic agents, and for controlling blood flow to the selected region. The diagnostic and/or therapeutic agents are to be used in accordance with specifications outlined by the manufacturer.

    The TAPAS™ Catheter is contraindicated for use in the coronary and intracranial arteries.

    The TAPAS™ Catheter is not intended for embolic protection or as an aspiration catheter.

    Device Description

    Other than deletion of an optional extracorporeal circuit connection and the addition of a second balloon, the TAPAS™ catheter that is the subject of this submission is a substantially equivalent to the predicate ThermopeutiX DuoFlo catheter. The TAPAS™ catheter, as does the DuoFlo™ catheter, consists of two coaxial tubular shafts (inflow and outflow lumens) that can infuse diagnostic and/or therapeutic agents to selected regions, as well as direct arterial blood or other solutions to a specific region or organ. The lumens may be accessed using external devices connected to these lumens. There are three additional smaller lumen is used for inflation of the distal balloon on the inflow lumen. one is for optional pressure monitoring and the third lumen permits inflation of the balloon on the outflow lumen.

    AI/ML Overview

    This document is a 510(k) summary for the ThermopeutiX TAPAS™ Catheter. It describes the device, its intended use, and its substantial equivalence to a predicate device.

    Here's an analysis of the acceptance criteria and the study information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test/CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    BiocompatibilityIn accordance with ISO 10993-1 standards for patient contact materials.Demonstrated through biocompatibility studies.
    Tip and Shaft DimensionsIn compliance with relevant standards and product specifications.Results were acceptable.
    Balloon Compliance & Inflation VolumeIn compliance with relevant standards and product specifications.Results were acceptable.
    Balloon Inflation & Deflation TimesIn compliance with relevant standards and product specifications.Results were acceptable.
    Balloon FatigueIn compliance with relevant standards and product specifications.Results were acceptable.
    Balloon BurstIn compliance with relevant standards and product specifications.Results were acceptable.
    Balloon Bond StrengthIn compliance with relevant standards and product specifications.Results were acceptable.
    Shaft/Lumen Joint Bond StrengthIn compliance with relevant standards and product specifications.Results were acceptable.
    Sheath CompatibilityIn compliance with relevant standards and product specifications.Results were acceptable.
    Shaft KinkIn compliance with relevant standards and product specifications.Results were acceptable.
    Shaft TorqueIn compliance with relevant standards and product specifications.Results were acceptable.
    Shaft Burst PressureIn compliance with relevant standards and product specifications.Results were acceptable.
    Tip Bond StrengthIn compliance with relevant standards and product specifications.Results were acceptable.
    Concentric Shaft MovementIn compliance with relevant standards and product specifications.Results were acceptable.
    Manifold Lumen Joint StrengthIn compliance with relevant standards and product specifications.Results were acceptable.
    Stopcock Lumen Joint StrengthIn compliance with relevant standards and product specifications.Results were acceptable.
    SterilitySterility Assurance Level (SAL) of 10⁻⁶.Conforms to SAL of 10⁻⁶ (adopted into existing validated ethylene oxide sterilization cycle).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify numerical sample sizes for the mechanical and biocompatibility tests performed. It states "Physical testing was performed and results were in compliance with relevant standards and the product specification." and "The biological safety of the device has been demonstrated through biocompatibility studies of patient contact materials in accordance with the standards outlined in ISO 10993-1."

    There is no information regarding data provenance (e.g., country of origin, retrospective or prospective) as these are non-clinical, in-vitro/bench tests.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The tests performed are mechanical and biological (biocompatibility) in nature, not involving expert interpretation of data like medical image analysis. The "ground truth" for these tests would be established by the physical and chemical properties of the materials and device components, measured against established standards.

    4. Adjudication Method for the Test Set

    Not applicable. This is not a study involving human interpretation or subjective assessments that would require an adjudication method. The outcomes of the tests are objective measurements against predefined specifications or standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This device is a medical catheter, not an AI-powered diagnostic tool. Therefore, no MRMC study or AI assistance evaluation was performed or is relevant.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) Was Done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

    The ground truth for the performance and safety tests described (biocompatibility, dimensions, balloon characteristics, bond strengths, etc.) is based on established industry standards (e.g., ISO 10993-1) and ThermopeutiX's internal product specifications. These serve as objective benchmarks.

    8. The Sample Size for the Training Set

    Not applicable. This is a physical medical device submission, which does not involve machine learning models or training sets.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093847

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2010-08-25

    (252 days)

    Product Code
    Regulation Number
    870.4450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DuoFlo™ Catheter is intended for general intravascular use in the peripheral vasculature in arteries 3.5 mm and larger. Once placed in the selected region, the catheter can be used for infusion of diagnostic and/or therapeutic agents, and for controlling blood flow to the selected region when connected to an extracorporeal circuit. The diagnostic and/or therapeutic agents are to be used in accordance with specifications outlined by the manufacturer. The DuoFlo™ Catheter is contraindicated for use in the coronary and intracranial arteries. The DuoFlo™ Catheter is not intended for embolic protection or as an aspiration catheter.

    Device Description

    The DuoFlo™ Catheter is a sterile single use device that consists of concentric shafts with four lumens with access via Luer connectors as follows, one for balloon inflation, one for pressure monitoring and two concentric lumens for infusate injection or extracorporeal circuit connections. The central through lumen accepts up to a 0.038" guidewire.

    AI/ML Overview

    The provided text is a 510(k) summary for the ThermopeutiX DuoFlo Catheter (Heparin Coated). This document is primarily focused on demonstrating substantial equivalence to a predicate device rather than providing a detailed study that proves a device meets specific acceptance criteria in the way a diagnostic software or AI model might.

    Here's an analysis based on the available information, addressing your specific points where possible, and noting where information is not present in the provided text:

    1. A table of acceptance criteria and the reported device performance

    The document does not present specific quantitative acceptance criteria or detailed performance results in the format typically used for a diagnostic device's clinical study. The device is a "Temporary Intravascular Occluding Catheter" and the focus is on its safety and equivalence.

    Performance CharacteristicAcceptance Criteria (Implied)Reported Device Performance
    BiocompatibilityMeet ISO 10993-1 standards for patient contact materials."Demonstrated through biocompatibility studies"
    SterilityConform to a Sterility Assurance Level (SAL) of 10⁻⁶"Conforms to a Sterility Assurance Level (SAL) of 10⁻⁶"
    Design/ConstructionIdentical to predicate (K080700) with added heparin coating."Identical in design and construction to the currently marketed DuoFlo™ catheter with the exception that bound heparin has been included in the catheter coating."
    Thrombus MinimizationHeparin coating helps to minimize thrombus formation."The heparin in the coating helps to minimize thrombus formation on the catheter surfaces." (This is a design claim rather than a quantifiable performance criterion with specific data in this document).
    Safety/EffectivenessNo new safety or effectiveness issues compared to predicate."There are no new safety or effectiveness issues related to this device."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided in the given text. The document refers to "biocompatibility studies" but does not detail their methodology, sample sizes, or data provenance. This is a medical device clearance, not typically a clinical trial for a diagnostic algorithm.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/provided. The ground truth establishment by experts is relevant for diagnostic AI/software evaluations, not for the clearance of a mechanical catheter based on substantial equivalence and biocompatibility.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/provided for the same reasons as point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    An MRMC study is not applicable/provided. This device is a catheter, not an AI or imaging diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A standalone performance study is not applicable/provided. This device is a physical medical instrument, not a software algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" in this context revolves around:

    • Biocompatibility: Adherence to ISO standards, likely demonstrated through in vitro and in vivo testing (e.g., cytotoxicity, sensitization, irritation, genotoxicity, hemocompatibility studies).
    • Sterility: Verification of the sterilization process to achieve the specified SAL.
    • Physical/Mechanical Properties: Verification that the design and construction are identical to the predicate (except for the coating), which implies similar mechanical performance.
    • Heparin Coating Efficacy (implied): The claim that the heparin helps minimize thrombus formation would ideally be supported by in vitro or in vivo studies demonstrating reduced thrombus compared to the uncoated predicate. However, detailed data for this is not presented in the summary, only the claim is made.

    8. The sample size for the training set

    This information is not applicable/provided. A "training set" is relevant for machine learning algorithms, not for the type of device being described.

    9. How the ground truth for the training set was established

    This information is not applicable/provided. (See point 8).


    Summary of the Study Mentioned (as much as can be gleaned):

    The primary "study" mentioned for the ThermopeutiX DuoFlo™ Catheter (Heparin Coated)'s performance and safety is:

    • Biocompatibility Studies: These were conducted on the patient contact materials in accordance with ISO 10993-1 standards. The exact details (e.g., specific tests, sample sizes, animal models if any) are not provided in this summary, but the claim is that biological safety has been "demonstrated."
    • Sterility Validation: The device's sterility "conforms to a Sterility Assurance Level (SAL) of 10⁻⁶." This indicates that sterilization processes were validated to meet this standard.

    The overall approach for this 510(k) is demonstrating substantial equivalence to a predicate device (K080700). The "study" largely consists of showing that the device is identical in design and construction, with the only change being the addition of a bound heparin coating. The safety of this change hinges on the biocompatibility studies of the new material and the claim that the heparin helps minimize thrombus, implying it's a beneficial addition without new safety concerns. No comparative clinical effectiveness study against the predicate is described in this document to quantify improvement due to the heparin coating.

    Ask a Question

    Ask a specific question about this device

    K Number
    K080700

    Validate with FDA (Live)

    Device Name
    DUOFLO CATHETER
    Manufacturer
    Date Cleared
    2008-09-26

    (199 days)

    Product Code
    Regulation Number
    870.4450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DuoFlo™ Catheter is intended for general intravascular use in the perioheral vasculature in arteries 3.5 mm and larger. Once placed in the selected region, the catheter can be used for infusion of diagnostic and/or therapeutic agents, and for controlling blood flow to the selected region when connected to an extracorporeal circuit. The diagnostic and/or therapeutic agents are to be used in accordance with specifications outlined by the manufacturer.

    The Duo-Flo Catheter is contraindicated for use in the coronary and intracranial arteries.

    The Duo-Flo Catheter is not intended for embolic protection or as an aspiration catheter.

    Device Description

    The DuoFlo™ Catheter is a sterile single use device that consists of concentric shafts with four lumens with access via Luer connectors as follows, one for balloon inflation, one for pressure monitoring and two concentric lumens for infusate injection or extracorporeal circuit connections. The central through lumen accepts up to a 0.038" guidewire.

    AI/ML Overview

    The provided text is a 510(k) summary for the ThermopeutiX DuoFlo™ Catheter, and it does not contain the level of detail typically found in a study report for an AI/ML-based medical device. Therefore, I cannot extract all the requested information. This document focuses on demonstrating substantial equivalence to predicate devices based on design, materials, and basic performance/safety validation for a physical medical device (a catheter), not a software algorithm.

    However, I can provide information based on what is available in the document regarding the performance and safety aspects of the physical device.

    Here's a breakdown of what can be extracted and what cannot:

    1. A table of acceptance criteria and the reported device performance

    The document mentions Performance and Safety validation but does not provide specific quantitative acceptance criteria or detailed reported performance in a tabular format as would be expected for an AI/ML study. It generally states that the device meets safety and integrity requirements.

    Acceptance Criteria CategoryReported Device Performance
    BiocompatibilityDemonstrated through biocompatibility studies of materials in accordance with ISO 10993-1.
    Physical IntegrityPhysical testing was performed to assure catheter integrity, including verification of catheter body and balloon burst pressure.
    SterilityDevice is supplied sterile with a Sterility Assurance Level (SAL) of 10⁻⁶.
    Safety/EffectivenessThere are no new safety or effectiveness issues related to this device.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not provided as the document describes validation for a physical catheter through biocompatibility and physical integrity testing, not a clinical study on a test set of data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable/provided. The validation described does not involve establishing ground truth by human experts for a test set, as it's not a diagnostic or AI/ML device requiring such evaluation.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable/provided for the same reasons as item 3.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable/provided. The device is a physical catheter, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    This information is not applicable/provided. There is no algorithm for this device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not applicable/provided. The "ground truth" for a physical device like a catheter would be its physical properties, material safety, and functional performance metrics, not clinical outcomes derived from expert assessment or pathology.

    8. The sample size for the training set

    This information is not applicable/provided. There is no training set for a physical catheter.

    9. How the ground truth for the training set was established

    This information is not applicable/provided. There is no training set for a physical catheter.


    Summary of Study (Based on available information):

    The "study" described in the 510(k) summary is primarily comprised of:

    • Biocompatibility studies: Conducted on the raw materials in accordance with ISO 10993-1 to demonstrate the safety of materials in contact with the body.
    • Physical testing: Performed to assure the integrity of the catheter, including verification of catheter body and balloon burst pressure. The document states that these tests were done to "assure catheter integrity."
    • Sterility validation: The device is supplied with a Sterility Assurance Level (SAL) of 10⁻⁶.

    The overall approach for this device relies on demonstrating substantial equivalence to already legally marketed predicate devices in terms of general indications, design, features, construction materials, and safety profile. The validation tests (biocompatibility, physical integrity, sterility) confirm that the DuoFlo™ Catheter meets established safety and performance standards for its type.

    Ask a Question

    Ask a specific question about this device

    K Number
    K073025

    Validate with FDA (Live)

    Date Cleared
    2007-11-09

    (14 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VascuTraK™ II PTA Dilatation Catheter is intended for dilatation of stenoses in the peripheral arteries including the iliac, femoral, popliteal, ilio-femoral, infrapopliteal and renal arteries and for the treatment of obstructive lesions of native or synthetic arteriovenous dialysis fistulae.

    Device Description

    The VascuTraK™ II Catheter is a sterile single use device that consists of a flexible shaft with a single through lumen terminating in a semi-compliant balloon. The catheter has a 24 cm lumen proximal to the balloon that accepts a 0.014" guidewire and an 18 mm lumen distal to the balloon that accepts a 0.014" guidewire. The distal balloon is inflated via a central lumen terminated in a luer fitting at the proximal end. The balloon is available in various diameters and lengths,

    AI/ML Overview

    The provided text is a 510(k) summary for the VascuTraK™ II PTA Dilatation Catheter. It describes a medical device, its intended use, and the comparison to a predicate device. However, it does not include a study that proves the device meets specific acceptance criteria in the manner requested by the prompt (e.g., performance metrics like sensitivity, specificity, accuracy, or human reader improvement for an AI/CAD system).

    This document is a regulatory submission for a physical medical device (a catheter), not a software or AI-based diagnostic device. Therefore, many of the requested categories like "multi reader multi case (MRMC) comparative effectiveness study," "standalone (algorithm only) performance," "number of experts used to establish ground truth," "adjudication method," "sample size for training set," and "how ground truth for training set was established" are not applicable to this type of device.

    The "acceptance criteria" and "device performance" in this context refer to engineering and biocompatibility testing for the catheter itself.

    Here's an attempt to extract and reframe the information based on the provided text, acknowledging the limitations for a physical device:


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria & TestReported Device Performance
    Material Biocompatibility (ISO 10993-1 requirements)Previously demonstrated for patient contact materials. No new materials used.
    Sterilization Assurance Level (SAL)Conforms to SAL of 10⁻⁶. (Prior sterilization testing is valid as modification doesn't affect sterilization.)
    Balloon PerformanceResults for modified device were consistent with those of the original device.
    Tip Tensile TestingResults for modified device were consistent with those of the original device.

    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided in a quantifiable manner for the specific tests mentioned. The text states "Testing of the modified catheter was performed," but does not detail the sample sizes for balloon performance or tip tensile testing. Data provenance is not applicable in the context of clinical data for AI/CAD, as this is a physical device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. "Ground truth" in this context would refer to established engineering standards and material science, not expert interpretation of diagnostic images or patient outcomes.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used for resolving disagreements among human readers in diagnostic studies.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/CAD device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/CAD device.

    7. The Type of Ground Truth Used

    For biocompatibility, the ground truth is established by ISO 10993-1 standards.
    For sterilization, the ground truth is established by regulatory requirements for Sterility Assurance Level (SAL).
    For balloon and tip performance, the ground truth is established by engineering specifications and performance data of the original (predicate) device.

    8. The Sample Size for the Training Set

    Not applicable. This refers to a training set for an algorithm, which is not relevant for this physical device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This refers to ground truth for an algorithm's training data.


    Summary of the Study that Proves the Device Meets Acceptance Criteria (as described in the document):

    The document details a Special 510(k) Device Modification for the VascuTraK™ II PTA Dilatation Catheter, which is a modification of the previously cleared VascuTraK™ PTA Dilatation Catheter (cleared under K063657).

    The primary modification is to the distal tip configuration, changing from a short lumen tip to permit the use of an external guidewire for easier catheter exchange.

    The study that proves the device meets acceptance criteria consists of:

    • Reliance on prior demonstrations:
      • Biocompatibility of patient contact materials was previously demonstrated in accordance with ISO 10993-1, and no new materials were used in the modification.
      • The modification does not affect the ability to sterilize the device, so prior sterilization testing establishing a Sterility Assurance Level (SAL) of 10⁻⁶ is considered valid.
    • Specific testing of the modified catheter:
      • Balloon performance testing was conducted.
      • Tip tensile testing was conducted.
      • The "results for the modified device were consistent with those of the original device," indicating that the modification did not negatively impact these critical performance characteristics and that the device still met the established performance benchmarks set by its predicate.

    The document concludes that there are "no new safety or effectiveness issues related to this device" based on these tests and comparisons. This type of submission relies heavily on the equivalence to a legally marketed predicate device, especially when the changes are minor and do not introduce new technology or materials.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1