510(k) Data Aggregation

    K Number
    K253339

    Validate with FDA (Live)

    Device Name
    Theracor
    Manufacturer
    Date Cleared
    2025-12-22

    (83 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Theracor is intended to cover, protect, and provide a moist wound environment.

    Theracor is indicated for use in the management of the following wounds:

    • Partial and full-thickness wounds
    • Pressure ulcers
    • Venous ulcers
    • Diabetic ulcers
    • Chronic vascular ulcers
    • Tunneled/undermined wounds
    • Surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence)
    • Trauma wounds (abrasions, lacerations, partial-thickness burns, and skin tears)
    • Draining wounds
    Device Description

    Theracor is derived from human umbilical cord extracellular matrix (ECM) and is indicated for the management of a range of acute and chronic wounds. The device is intended to cover, protect, and provide a moist wound environment. As a resorbable device, Theracor is aseptically processed and lyophilized to remove moisture while preserving the structural components of the umbilical cord matrix. The device is packaged in sheet form and should be rehydrated prior to applying to the wound bed.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K242828

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-10-18

    (29 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Corplex P/Theracor P/Allacor P is indicated for use in the management of the following wounds:

    • · Partial and full-thickness wounds
    • Pressure ulcers
    • Venous ulcers
    • · Diabetic ulcers
    • · Chronic vascular ulcers
    • · Tunneled/undermined wounds
    • · Surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence)
    • · Trauma wounds (abrasions, lacerations, partial-thickness burns, and skin tears)
    • · Draining wounds
    Device Description

    Corplex P/Theracor P/Allacor P is derived from human umbilical cord extracellular matrix (ECM) and is indicated for the management of a range of acute and chronic wounds. As a resorbable particulate device, Corplex P/Theracor P/Allacor P is lyophilized and packaged in a sterile vial, allowing the device to be rehydrated and applied directly to the wound.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a medical device called Corplex P/Theracor P/Allacor P. It states that the device is identical to its predicate device (K231325) in most aspects, with a minor modification being the addition of an 8 cc configuration. The document explicitly states that this modification does not raise new questions of safety or effectiveness and therefore, no new clinical testing, biocompatibility testing, or extensive performance studies were required to demonstrate substantial equivalence.

    Instead, the submission relies on the established performance of the predicate device and verification testing related to the minor design change.

    Therefore, many of the requested details about acceptance criteria, detailed study design, sample sizes, expert involvement, and ground truth establishment, which are typically associated with performance studies, are not explicitly present in this summary document because such extensive studies were deemed unnecessary for this 510(k) submission.

    Here's an attempt to answer your questions based on the provided text, indicating where information is not available due to the nature of this submission:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state acceptance criteria in a quantitative format for all characteristics. Instead, it relies on the device being "identical" or having "same" or "does not raise new questions of safety and effectiveness" compared to the predicate device. For the one specific test mentioned (Bioburden Testing), the outcome is a simple "PASS".

    CharacteristicAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance (K242828)
    Size/VolumeMust not raise new questions of safety and effectiveness compared to 1 cc, 2 cc, and 4 cc configurations.1 cc, 2 cc, 4 cc, 8 cc (The additional 8cc configuration was deemed to "not raise new questions of safety and effectiveness")
    Nominal Particle SizesSame as predicate: Particles ranging from 0.1 mm to 2.0 mm (must pass through 2.00 mm aperture sieve and not pass through 0.106 mm aperture sieve, ASTM E11).Particles ranging from 0.1 mm to 2.0 mm (Same as predicate: Particles must be able to pass through a 2.00 mm aperture calibrated sieve and not pass through a 0.106 mm aperture calibrated sieve (ASTM E11) when manually sieved.)
    Intended UseSame as predicate: To cover, protect, and provide a moist wound environment.Same
    Indications for UseSame as predicate for specified wound types.Same
    ConfigurationSame as predicate: Particles.Same
    Material SourceSame as predicate: Human umbilical cord (recovered per 21 CFR 1271).Same
    ComponentsSame as predicate: Collagen and associated ECM components (collagen I, collagen III, collagen V).Same
    Collagen (% total weight)Same as predicate: 46% (Mean Value).46% (Mean Value)
    Total Glycosaminoglycans (mg/g)Same as predicate: 20.3 mg/g (Mean Value).20.3 mg/g (Mean Value)
    Endotoxin (EU/device)Same as predicate: <20 EU/device.<20 EU/device
    PackagingMust not raise new questions of safety or effectiveness compared to 6 mL glass vial.10 mL Glass vial with bromobutyl stopper and aluminum crimped lid, inside single peel-open pouch (Deemed to "not raise new questions of safety or effectiveness" when compared to the 6 mL vial of the predicate)
    SterilizationSame as predicate: Sterilized by electron beam irradiation.Same
    Moisture ContentSame as predicate: <15%.<15%
    Sterility Assurance LevelSame as predicate: 10-6.10-6
    Shelf LifeSame as predicate: One Year.One Year
    Bioburden TestingAcceptable result based on methods used for predicate clearance.PASS

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document primarily relies on the substantial equivalence to a predicate device (K231325). For the specific modification (8 cc configuration), verification testing was performed.

    • Test Set Sample Size: Not specified for the bioburden testing or other verification tests mentioned. The document states "Testing was performed based on the risk assessment for the proposed change using the same methods used to support the clearance of the predicate device."
    • Data Provenance: Not explicitly stated. Given it's a 510(k) submission to the US FDA by a US-based company (StimLabs, LLC, Roswell, Georgia), the data is most likely from the US, and refers to prior data from the predicate device. It is not a clinical study, so terms like "retrospective" or "prospective" are not applicable to the verification testing conducted for this submission. The predicate device's data would have been part of its original 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This submission is for a material device, not an AI or diagnostic device that requires expert ground truth for image interpretation or similar tasks. The "ground truth" for the characteristics listed would be defined by standard laboratory testing methodologies and specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. The verification testing mentioned (e.g., bioburden) involves laboratory results rather than expert adjudication of interpretations.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. This is a wound dressing made from human umbilical cord extracellular matrix, not an AI or diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    No. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the verification testing related to the new device, the ground truth would be defined by standard laboratory test results and established acceptance criteria for physical and chemical properties (e.g., bioburden levels, particle size distribution, moisture content, collagen content, endotoxin levels). For the predicate substantial equivalence, the "ground truth" refers to the established safety and effectiveness of the existing legally marketed device.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231325

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-02-02

    (270 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Corplex P/Theracor P/Allacor P is indicated for use in the management of the following wounds:

    • Partial and full-thickness wounds
    • Pressure ulcers
    • Venous ulcers
    • Diabetic ulcers
    • Chronic vascular ulcers
    • Tunneled/undermined wounds
    • Surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence)
    • Trauma wounds (abrasions, lacerations, partial-thickness burns, and skin tears)
    • Draining wounds
    Device Description

    Corplex P/Theracor P/Allacor P is derived from human umbilical cord extracellular matrix (ECM) and is indicated for the management of a range of acute and chronic wounds. As a resorbable particulate device, Corplex P/Theracor P/Allacor P is lyophilized and packaged in a sterile vial, allowing the device to be rehydrated and applied directly to the wound.

    AI/ML Overview

    This document describes a 510(k) premarket notification for a wound dressing device, Corplex P/Theracor P/Allacor P. It focuses on demonstrating substantial equivalence to a predicate device, Myriad Particles, rather than proving the device meets specific performance criteria through a comparative effectiveness study in the context of an AI-powered medical device.

    Therefore, many of the requested categories related to AI device performance, such as sample size for test sets, data provenance, expert ground truth establishment, MRMC studies, standalone performance, and training set details, are not applicable to the information provided in this 510(k) submission.

    This submission is about a traditional medical device (wound dressing) aiming for clearance based on substantial equivalence, not a novel AI/ML device that requires extensive clinical validation of its algorithm's performance against human readers or a robust ground truth.

    Here's an attempt to answer the applicable questions based on the provided text:

    1. Acceptance Criteria and Device Performance

    The submission does not outline specific, quantified performance "acceptance criteria" in the way one would for an AI algorithm's diagnostic accuracy (e.g., "sensitivity must be >X%, specificity >Y%"). Instead, it demonstrates "substantial equivalence" to a predicate device by comparing various technological characteristics and presenting results of non-clinical (performance and biocompatibility) and clinical (human repeat insult patch test, skin prick test) testing to show that differences do not raise new safety or effectiveness concerns.

    Table of Performance Testing (Non-Clinical):

    TestResult
    Pour Test – DryPASS
    Solution Compatibility TestPASS
    Digestion AssayPASS
    Pour Test – WetPASS
    Absorption TestPASS
    Evaporation TestPASS
    Extracellular Matrix CharacterizationPASS

    Table of Biocompatibility Testing Results (All Met Requirements):

    • Cytotoxicity: ISO 10993-5:2009
    • Materials Mediated Pyrogenicity: ISO 10993-11:2017
    • Sensitization: ISO 10993-10:2010
    • Acute Systemic Toxicity: ISO 10993-11:2017
    • Intracutaneous Reactivity: ISO 10993-10:2021
    • Implantation: ISO 10993-6:2016
    • Chemical Characterization and Toxicological Risk Assessment: ISO 10993-18:2020, ISO 10993-17:2002, ISO 21726:2019
    • Genotoxicity: 10993-3:2014, ISO 10993-33:2015
    • Viral Risk Assessment and Clearance Study
    • Endotoxin: ANSI/AAMI/ST72
    • Packaging System Cytotoxicity: ISO 10993-5:2009

    Clinical Testing:

    • Human Repeat Insult Patch Test
    • Skin Prick Test

    The text states that these tests were conducted to "demonstrate substantial equivalence... or to mitigate any potential performance risks" and that results "meet the requirements" (for biocompatibility). No specific numerical thresholds or performance metrics are provided for these tests, as the goal is conformance to standards and equivalence, not a quantitative measure of superior performance an AI device might exhibit.

    2. Sample Size for the Test Set and Data Provenance

    Not applicable in the context of an AI/ML device's test set. The clinical testing mentioned (Human Repeat Insult Patch Test, Skin Prick Test) would have used human subjects, but their sample sizes are not disclosed in this document. These are typical safety evaluations for skin-contacting devices, not performance evaluations like those for AI.

    3. Number of Experts used to establish the Ground Truth for the Test Set and the Qualifications of those Experts

    Not applicable. Ground truth, in the AI context of expert consensus, is not relevant here. The "ground truth" for this device's testing would be defined by the results of the specific ASTM, ISO, or other standardized tests performed (e.g., a certain level of endotoxin, or a negative cytotoxicity finding). These are objective measurements from laboratory tests, not subjective interpretations by human experts.

    4. Adjudication Method for the Test Set

    Not applicable. There's no subjective interpretation requiring adjudication in the tests described.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done

    No, this is not an MRMC study. This is a submission for a wound dressing device, not an AI diagnostic tool.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. There is no algorithm or AI component in this device.

    7. The Type of Ground Truth Used

    The "ground truth" for this medical device's clearance is a combination of:

    • Conformance to established standards: e.g., ISO 10993 series for biocompatibility.
    • Comparison to a predicate device: showing that the new device's characteristics and performance are "substantially equivalent" and do not raise new safety or effectiveness concerns compared to a legally marketed device.
    • Laboratory test results: demonstrating properties like moisture content, dissolution, etc., meet internal specifications or are comparable to the predicate.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI device that requires a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable.

    Ask a Question

    Ask a specific question about this device