Search Filters

Search Results

Found 235 results

510(k) Data Aggregation

    K Number
    K260766

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2026-04-08

    (30 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K251666

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2026-04-08

    (313 days)

    Product Code
    Regulation Number
    880.6885
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K252680

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-09-24

    (30 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Celerity™ 20 HP Biological Indicator (BI) is intended for routine monitoring of the following sterilizer cycles:

    • Lumen, Non Lumen, Fast Non Lumen, Fast, Flexible and Specialty Cycles of the V-PRO 1, 1 Plus, maX, maX 2, 60 and s2 Low Temperature Sterilization Systems.
    • STERRAD® 100S Sterilizer (Default Cycle)
    • Standard and Advanced Cycles of the STERRAD® NX® Sterilizer with or without ALLClear
    • Standard, FLEX, Express and DUO Cycles of the STERRAD® 100NX® Sterilizer with or without ALLClear® Technology

    When used in conjunction with the Celerity™ HP Incubator or the Celerity™ Incubator, the Celerity™ 20 HP BI provides a fluorescent result within 20 minutes.

    Device Description

    The product is intended to monitor the vapor phased hydrogen peroxide sterilization cycles described in the indications for use. It produces an optical change (signal) that is detected by the STERIS proprietary reader, STERIS Celerity™ Incubator, within 20 minutes to confirm the viability of the biological indicator at the end of a sterilization process. The product consists of Geobacillus stearothermophilus spores and a defined nutrient media in a plastic vial. A reporter enzyme, which is produced by the organism, reacts with a fluorogenic substrate within the defined nutrient media to produce a fluorescent moiety.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K251452

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-06-06

    (28 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Celerity 5 HP Biological Indicator:
    The Celerity 5 HP Biological Indicator (BI) is intended for routine monitoring of the following sterilizer cycles:

    • Lumen, Non Lumen, Fast Non Lumen, Fast, Flexible and Specialty Cycles of the V-PRO 1, 1 Plus, maX, maX 2, 60 and s2 Low Temperature Sterilization Systems
    • Standard and Advanced Cycles of the STERRAD® NX® Sterilizer with or without ALLClear®
    • Standard, FLEX, Express and DUO Cycles of the STERRAD® 100NX® Sterilizer with or without ALLClear®.

    When used in conjunction with the Celerity® Incubator, the Incubator provides a fluorescent result within 5 minutes.

    Celerity 5 HP Challenge Pack:
    The Celerity 5 HP Challenge Pack is intended for qualification testing of the V-PRO Low Temperature Sterilization System following installation, relocation, malfunctions or major repairs and for routine requalification testing.

    The Challenge Pack is placed in an otherwise empty sterilizer chamber; a hospital-defined challenge load is not included.

    The Challenge Pack is not intended for routine monitoring of V-PRO Sterilizers. It has been tested and validated solely for use in periodic testing of the Sterilizers.

    Celerity 20 HP Biological Indicator
    The Celerity 20 HP Biological Indicator (BI) is intended for routine monitoring of the following sterilizer cycles:

    • Lumen, Non-Lumen, Fast Non Lumen, Fast, Flexible and Specialty Cycles of the V-PRO 1, 1 Plus, maX, maX 2, 60 and s2 Low Temperature Sterilization Systems.
    • Standard and Advanced Cycles of the STERRAD® NX® Sterilizer with or without ALLClear
    • Standard, FLEX, Express and DUO Cycles of the STERRAD® 100NX® Sterilizer with or without ALLClear®

    When used in conjunction with the Celerity HP Incubator, the Celerity 20 HP BI provides a fluorescent result within 20 minutes.

    Celerity 20 HP Challenge Pack
    The Celerity 20 HP Challenge Pack is intended for qualification testing of the V-PRO® Low Temperature Sterilization System following installation, relocation, malfunctions or major repairs and for routine requalification testing.

    The challenge pack is placed in an otherwise empty sterilizer chamber; a hospital-defined challenge load is not included.

    The challenge pack is not intended for routine monitoring of V-PRO Sterilizers. It has been tested and validated solely for use in periodic testing of the Sterilizers.

    Device Description

    Celerity 5 HP BI
    The product is intended to monitor the vapor phased hydrogen peroxide sterilization cycles described in the indications for use. It produces an optical change (signal) that is detected by the STERIS proprietary reader, STERIS Celerity Incubator, within 5 minutes to confirm the viability of the biological indicator at the end of a sterilization process. The product consists of Geobacillus stearothermophilus spores and a defined nutrient media in a plastic vial. A reporter enzyme, which is produced by the organism, reacts with a fluorogenic substrate within the defined nutrient media to produce a fluorescent moiety.

    Celerity 5 HP Challenge Pack
    The Celerity 5 HP Challenge Pack is used in the sterile processing department, usually by service technicians or biomedical technicians to qualify the unit following installation, relocation, malfunctions or major repairs and for routine requalification testing. It consists of a Celerity 5 HP Biological Indicator, a Celerity HP Chemical Indicator, and a layer of absorptive foam within a Tyvek pouch.

    The intended change to the devices is to update the secondary container and product labeling. Specifically, the clear cellophane wrapper around the product carton will be replaced by a clear, laminated pouch. Labeling on the carton and the Instructions for Use (IFU) will be updated to specify that the product must remain stored in its original laminated pouch to maintain the claimed shelf-life. Additionally, a storage disclaimer label will be added to the exterior of the pouch to notify users of this change. Finally, this submission is intended to establish a 3-month in-use shelf-life following pouch opening, not to exceed the original expiration date.

    Celerity 20 HP BI
    The product is intended to monitor the vapor phased hydrogen peroxide sterilization cycles described in the indications for use. It produces an optical change (signal) that is detected by the STERIS proprietary reader, STERIS Celerity Incubator, within 20 minutes to confirm the viability of the biological indicator at the end of a sterilization process. The product consists of Geobacillus stearothermophilus spores and a defined nutrient media in a plastic vial. A reporter enzyme, which is produced by the organism, reacts with a fluorogenic substrate within the defined nutrient media to produce a fluorescent moiety.

    Celerity 20 HP Challenge Pack
    The Celerity 20 HP Challenge Pack is used in the sterile processing department, usually by service technicians or biomedical technicians to qualify the unit following installation, relocation, malfunctions or major repairs and for routine requalification testing. It consists of a Celerity 5 HP Biological Indicator, a Celerity HP Chemical Indicator, and a layer of absorptive foam within a Tyvek pouch.

    The intended change to these devices is to update the secondary container and product labeling. Specifically, the clear cellophane wrapper around the product carton will be replaced by a clear, laminated pouch. Labeling on the carton and the Instructions for Use (IFU) will be updated to specify that the product must remain stored in its original laminated pouch to maintain the claimed shelf-life. Additionally, a storage disclaimer label will be added to the exterior of the pouch to notify users of this change. Finally, this submission is intended to establish a 3-month in-use shelf-life following pouch opening, not to exceed the original expiration date.

    AI/ML Overview

    The provided text describes premarket notification for biological indicators and doesn't involve an AI/ML powered medical device. Therefore, many of the requested criteria, such as "multi-reader multi-case (MRMC) comparative effectiveness study" or "number of experts used to establish ground truth", which are specific to AI/ML powered devices, are not applicable here.

    However, I can extract information related to the acceptance criteria and the study proving the device meets those criteria, as much as applicable for a non-AI medical device.

    Device: Celerity 5 HP Biological Indicator (LCB052) & Challenge Pack (LCB059), and Celerity 20 HP Biological Indicator (LCB044) & Challenge Pack (LCB045)

    Type of Device: Sterilization Process Indicator (Biological Indicator - BI). These devices contain a specific type of spore (Geobacillus stearothermophilus) and media to confirm the effectiveness of a sterilization process. A positive result (fluorescent signal) indicates sterilization failure, while a negative result indicates successful sterilization.

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria and results are provided for two distinct product lines: Celerity 5 HP and Celerity 20 HP. Note that the "Result" for all tests is simply "PASS", indicating that the criteria were met.

    For Celerity 5 HP Biological Indicator and Challenge Pack:

    TestAcceptance CriteriaResult
    Reduced Incubation Time TestingAll BI lots will demonstrate 97% or greater positive growth results at 7 days as compared to the fluorescent result at less than or equal to 5-minutesPASS
    Simulated Use TestingAll processed BIs shall be sterile following full cycle exposure sterilizer cycles.PASS
    Cap Media TestingAll BIs will demonstrate a positive fluorescent signal in the fluorescent incubators.PASS
    Stability Testing: PopulationThe mean initial population at the start of the stability study must be 1.0-4.0 x 10⁶ CFU/BI. At each subsequent time point, the mean population shall be 50-300% of the initial mean population.PASS
    Stability Testing: D-valueD-value between 6 and 20 seconds.PASS
    Stability Testing: Survival/KillDemonstration of one all survive time and one all kill time point.PASS
    Stability Testing: RITAll BI lots will demonstrate 97% or greater positive growth results at 7 days as compared to the fluorescent result at less than or equal to 5 minutes.PASS
    Stability Testing: Media TestingAll BIs will demonstrate a positive fluorescent signal in the fluorescent incubators.PASS

    For Celerity 20 HP Biological Indicator and Challenge Pack:

    TestAcceptance CriteriaResult
    Reduced Incubation Time TestingAll BI lots will demonstrate 97% or greater positive growth results at 7 days as compared to the fluorescent result at less than or equal to 20 minutesPASS
    Simulated Use TestingAll processed BIs shall be sterile following full cycle exposure sterilizer cycles.PASS
    Cap Media TestingAll BIs will demonstrate a positive fluorescent signal in the fluorescent incubators.PASS
    Stability Testing: PopulationThe mean initial population at the start of the stability study must be 1.0-4.0 x 10⁶ CFU/BI. At each subsequent time point, the mean population shall be 50-300% of the initial mean population.PASS
    Stability Testing: D-valueD-value between 6 and 20 seconds.PASS
    Stability Testing: Survival/KillDemonstration of one all survive time and one all kill time point.PASS
    Stability Testing: RITAll BI lots will demonstrate 97% or greater positive growth results at 7 days as compared to the fluorescent result at less than or equal to 20 minutes.PASS
    Stability Testing: Media TestingAll BIs will demonstrate a positive fluorescent signal in the fluorescent incubators.PASS

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the numerical sample size (e.g., number of BIs or challenge packs) used for each specific test (Reduced Incubation Time, Simulated Use, Cap Media, Stability testing). It refers to "All BI lots" for Reduced Incubation Time and "All processed BIs" for Simulated Use, and implicitly "BIs" for Stability Testing. This suggests a statistically relevant number of units from manufactured lots were tested.

    • Data Provenance: The document does not specify the country of origin of the data. Given the "U.S. Food & Drug Administration" letterhead and the manufacturer's address in Ohio, USA, it is highly likely that the testing was conducted in the US.
    • Retrospective or Prospective: The testing appears to be prospective as it's a pre-market submission to demonstrate the device meets performance criteria for clearance. Stability testing also inherently involves prospective observation over time.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This concept is not applicable to this type of device. The "ground truth" for a biological indicator is determined by the biological and chemical reactions themselves (i.e., presence or absence of viable spores, or a color change in the chemical indicator), not by human expert interpretation of images or data. The determination of "sterile" or "non-sterile" from a biological indicator is a direct, objective measurement.

    4. Adjudication Method for the Test Set

    Not applicable. As noted above, the results of biological indicators are typically objective and determined by the physical/chemical properties and reactions, not by subjective human interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device nor one that involves human "readers" interpreting medical images or data.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical, biological and chemical indicator device, not a software algorithm. The "Celerity Incubator" is an accessory that reads the fluorescent signal, it is not an AI algorithm.

    7. The Type of Ground Truth Used

    The ground truth for these devices is based on biological viability (presence or absence of viable Geobacillus stearothermophilus spores) for the biological indicator and chemical reaction (color change) for the chemical indicator, validated against recognized sterilization parameters (e.g., D-value, Survival Time, Kill Time). This is a direct measure of the effectiveness of the sterilization process.

    8. The Sample Size for the Training Set

    Not applicable. These are not AI/ML devices that require training sets. The "training" of these devices (i.e., their manufacturing and design) is based on established microbiological and sterilization science.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no "training set" in the context of an AI/ML algorithm. The performance of these biological indicators relies on their inherent design and manufacturing quality, ensuring the spores exhibit the specified resistance characteristics and the media accurately detects their viability. This is established through rigorous quality control and adherence to relevant standards for biological indicators.

    Ask a Question

    Ask a specific question about this device

    K Number
    K251048

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-05-02

    (29 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RAPICIDE™ PA Test Strips are chemical indicators used after the disinfection cycle to ensure that the RAPICIDE™ PA High-Level Disinfectant solution is met the minimum recommended concentration of 850ppm peracetic acid. The RAPICIDE PA Test Strips are designed for use by trained personnel who reprocess endoscopes and their accessories in facilities where endoscopies are performed.

    Device Description

    Rapicide PA High-Level Disinfectant Test Strip has the ability to measure the disinfectant use solution concentration above 850 ppm Peracetic Acid (PAA). This is the minimum recommended concentration (MRC) of PAA for Rapicide PA high-level disinfectant. If the solution is at or below MRC, the test strip pad will indicate a failure by turning dark grey, blue grey, violet grey, light grey, or white (no change in color). A passing result will be indicated by the test strip pad turning solid black after contact with solution at concentrations of 1100 ppm ± 20 ppm PAA.

    AI/ML Overview

    The provided FDA 510(k) clearance letter pertains to a chemical indicator device, "Rapicide PA High-Level Disinfection Test Strip Model Ref # = ML02-0118," not an AI/ML-enabled medical device. Therefore, the information regarding the acceptance criteria and study proving its performance is focused on the chemical and physical properties of the test strips, rather than AI model performance metrics like sensitivity, specificity, MRMC studies, or ground truth establishment relevant to AI/ML devices.

    Based on the document, I can extract the following information relevant to the device's acceptance criteria and studies:

    Device Description and Intended Use:

    The Rapicide PA High-Level Disinfection Test Strips are chemical indicators used after the disinfection cycle to verify that the RAPICIDE™ PA High-Level Disinfectant solution has met the Minimum Recommended Concentration (MRC) of 850ppm peracetic acid. A "pass" is indicated by the test strip pad turning solid black when exposed to solutions at concentrations of 1100 ppm ± 20 ppm PAA. A "failure" (solution at or below MRC) is indicated if the pad turns dark grey, blue grey, violet grey, light grey, or white (no color change).

    Here's the breakdown of the acceptance criteria and the studies that prove the device meets these criteria, based on the provided text:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Minimum Recommended Concentration (MRC) RecognitionThe device must reliably indicate that the RAPICIDE™ PA HLD solution has met the MRC of 850ppm peracetic acid.
    "Pass" Endpoint SpecificationThe test strip pad must turn solid black when the solution is at concentrations above 850ppm and specifically at 1100 ppm ± 20 ppm PAA.
    "Fail" Endpoint SpecificationThe test strip pad must indicate failure (e.g., turn dark grey, blue grey, violet grey, light grey, or white) when the solution is at or below the MRC of 850ppm PAA.
    Dynamic RangeThe device must perform within its specified detection range for peracetic acid concentration.
    Comparative Sensitivity and SpecificityThe device must demonstrate appropriate sensitivity and specificity in detecting the target peracetic acid concentration against a known standard.
    Analytic Specificity – ContaminantsThe device's performance should not be adversely affected by common contaminants that might be present in the use environment.
    Analytic Specificity – Other GermicidesThe device should not cross-react or provide false readings due to the presence of other germicides.
    Shelf LifeThe device must maintain its specified performance characteristics for the stated shelf life (18 Months).
    In-Use (Open Bottle) StabilityThe device must maintain its specified performance characteristics for the stated open bottle shelf life (4 months).

    Study Details (Based on available information from the 510(k) Summary)

    The provided document is a 510(k) clearance letter and an abbreviated 510(k) summary, which focuses on demonstrating substantial equivalence to a predicate device. It does not contain the detailed study protocols, raw data, or comprehensive statistical analyses that would be found in a full study report or clinical trial documentation. Therefore, many of the requested details for AI/ML device studies (sample size, data provenance, expert adjudication, MRMC studies, training set details) are not applicable or not provided for this type of chemical indicator device.

    Here's a breakdown of the requested information based on the provided text:

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not explicitly stated in the provided document. The 510(k) summary states that "Testing to assess and demonstrate performance of the subject device is summarized below," listing categories like "Dynamic Range," "Comparative Sensitivity and Specificity," etc., and concluding they "Met Acceptance Criteria." However, the exact number of test strips or solution samples used for each test is not disclosed.
      • Data Provenance: Not specified. It can be inferred that the testing was conducted internally by STERIS Corporation, likely in a laboratory setting, to evaluate the chemical performance of the test strips. Retrospective or prospective nature is not mentioned for this type of chemical performance testing.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable / Not Provided: For a chemical indicator device, "ground truth" is typically established by precisely formulated chemical solutions (e.g., 850ppm PAA, 1100ppm PAA ± 20ppm PAA) with known concentrations, verified by analytical chemistry methods, rather than by human expert consensus or annotations as would be the case for medical imaging AI. The document does not mention the use of human experts for establishing ground truth.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • Not Applicable / None: Adjudication methods like 2+1 or 3+1 are used to resolve discrepancies in human expert annotations for complex data (e.g., medical images). For a chemical indicator, the output (color change) is objective and directly correlated with the chemical concentration. The "reading" of the strip is a visual interpretation of a chemical reaction against a defined color standard (solid black for pass, various greys for fail). Discrepancies would likely be addressed through re-testing or instrument calibration, not expert adjudication.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Not Applicable: MRMC studies are specific to evaluating the impact of AI assistance on human reader performance in tasks like image interpretation. This device is a chemical indicator test strip, not an AI-assisted diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

      • Not Applicable: This device is a physical chemical strip, not a software algorithm. Its performance is inherent in its chemical design.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Known Chemical Concentrations: The ground truth for this device is established by precisely prepared chemical solutions with known, verified concentrations of peracetic acid (e.g., 850ppm MRC, 1100ppm pass threshold). This is a direct physical or chemical measurement, not dependent on expert consensus, pathology, or outcomes data.
    7. The sample size for the training set:

      • Not Applicable: This device is a chemical indicator, not an AI/machine learning model that requires a training set. The "design" is based on chemistry and materials science, not data-driven learning.
    8. How the ground truth for the training set was established:

      • Not Applicable: As there is no "training set" for a chemical indicator, this question is irrelevant. The "design" and "calibration" involve chemical engineering and materials science, where the components are selected and optimized to react predictably to specific chemical concentrations.
    Ask a Question

    Ask a specific question about this device

    K Number
    K243876

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-04-25

    (128 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The integrating indicator is designed to chemically react over time with the critical parameters of a steam sterilization cycle within a specified tolerance. The integrating indicator strip is intended to be placed in each pack, pouch, container, tray or other containment device to function as an independent monitor of critical parameters for the following sterilization cycles:

    Steam Sterilization Cycles:

    • 250°F/121°C, 30 minutes Gravity
    • 270°F/132°C, 4 minutes Dynamic Air Removal
    • 270°F/132°C, 5 minutes Dynamic Air Removal
    • 270°F/132°C, 6 minutes Dynamic Air Removal
    • 270°F/132°C, 7 minutes Dynamic Air Removal
    • 270°F/132°C, 8 minutes Dynamic Air Removal
    • 270°F/132°C, 9 minutes Dynamic Air Removal
    • 270°F/132°C, 10 minutes Dynamic Air Removal
    • 270°F/132°C, 15 minutes Gravity
    • 273°F/134°C, 4 minutes Dynamic Air Removal
    • 275°F/135°C, 3 minutes Dynamic Air Removal
    • 275°F/135°C, 10 minutes Gravity

    Steam Sterilization Cycles (IUSS):

    • 270°F/132°C, 4 minutes Dynamic Air Removal
    • 270°F/132°C, 3 minutes Gravity
    • 270°F/132°C, 10 minutes Gravity
    • 275°F/135°C, 3 minutes Dynamic Air Removal
    • 275°F/135°C, 3 minutes Gravity
    • 275°F/135°C, 10 minutes Gravity
    Device Description

    The VERIFY STEAM Integrating Indicator is a single use device used by healthcare providers to monitor steam sterilization cycles. The VERIFY STEAM Integrating Indicator is included in a pack, pouch, container, tray or other containment device in a steam sterilizer and the load is processed in accordance with the sterilizer's manufacturer's directions. Prior to the use of the processed items, the integrator is observed. If the dark bar on the device enters the ACCEPT (OK) window, the integrator is read as a PASS to indicate that the steam sterilization criteria for the cycle have been met. If the dark bar on the device does not enter the ACCEPT (OK), the integrator is read as a FAIL, indicating that sufficient steam sterilization criteria has not been met and processed materials should be subjected to another steam sterilization cycle prior to use.

    AI/ML Overview

    Here's an analysis of the provided FDA 510(k) clearance letter, focusing on the acceptance criteria and the study proving the device meets them:

    The VERIFY STEAM Integrating Indicator (K243876) is a chemical indicator designed to monitor steam sterilization cycles. It is intended to be placed within containment devices (packs, pouches, etc.) during sterilization to independently verify that critical parameters have been met. The device is considered to "pass" if a dark bar on the indicator enters the "ACCEPT (OK)" window.

    Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Stated Value Testing per ANSI/AAMI/ISO 11140-1:2014 section 11• All integrators processed will demonstrate no visible deterioration of the film• All integrators processed in a dry heat cycle will demonstrate a failing result• The Stated Values for time at 121 °C and 135 °C shall be specified and shall not be less than 16.5 min at 121 °C and 1.2 min at 135 °C• The temperature coefficient shall not be less than 10°C and not more than 27°C.• The correlation coefficient shall not be less than 0.9PASS
    Simulated Use Testing• All integrators processed in a full cycle will demonstrate a passing result with no physical deterioration of the film• All integrators processed in a partial cycle will demonstrate a passing result with no physical deterioration of the film• No integrators will exhibit a leak in the viewing windowPASS
    Stability Study• All CIs processed in pass/full cycles in the BIER vessel will demonstrate pass results and shall not be under minimum specification limits as per the standard ISO 11140-1• All CIs processed in fail/partial cycles in BIER vessel will demonstrate fail results• All CIs processed for dry heat testing will demonstrate fail results.PASS
    Comparison Testing with Biological Indicator• The Chemical Indicator does not reach its endpoint before the biological indicator is inactivated.PASS

    Study Details

    The provided document describes non-clinical testing performed to evaluate the performance of the VERIFY STEAM Integrating Indicator and demonstrate its substantial equivalence to a predicate device (K213412).

    1. Sample sized used for the test set and the data provenance:

    • The document does not specify the exact sample sizes (number of indicators) used for each individual test (Stated Value, Simulated Use, Stability, Comparison with Biological Indicator).
    • The data provenance is not explicitly stated in terms of country of origin or whether it was retrospective or prospective. However, given the context of a 510(k) submission for a medical device, it is highly likely that these were prospective, laboratory-based studies conducted by the manufacturer (STERIS Corporation, Franklin Park, IL, USA) to evaluate the device's performance under controlled conditions.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This device is a chemical indicator that produces a visual change (dark bar entering an "ACCEPT (OK)" window) based on exposure to sterilization parameters. The "ground truth" for these tests is established by the controlled sterilization cycle conditions (temperature, time, presence of steam) and the objective physical change of the indicator, rather than subjective expert interpretation.
    • Therefore, no human experts were used to establish the ground truth for the test set in the way radiologists or pathologists would interpret an image. The performance is assessed against defined physical and chemical criteria.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • As the "ground truth" is determined by objective physical and chemical changes under controlled conditions, and the outcome (pass/fail) is a direct observation of the indicator's state, no adjudication method involving human experts was necessary for the test set.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging devices where human readers interpret results, sometimes with AI assistance. The VERIFY STEAM Integrating Indicator is a simple chemical indicator, not an AI-powered diagnostic tool, and its primary function is to react chemically, not to be "read" in a complex interpretive sense by multiple human readers.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This question is not applicable in the context of a chemical indicator. The device is entirely standalone in its function – it reacts chemically without any human intervention or an algorithm. Its "performance" is its physical color change, which is then observed by a human. There is no separate "algorithm" to evaluate.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth for these tests is based on objective physical and chemical properties and predefined standards. Specifically:
      • Controlled Sterilization Parameters: The cycles are run at precisely defined temperatures and times, and the presence of steam is controlled. The "truth" is whether these parameters were met or not.
      • ANSI/AAMI/ISO 11140-1:2014: This international standard defines the performance requirements for chemical indicators, providing the objective benchmarks against which the device is evaluated.
      • Biological Indicator Inactivation: For comparison testing, the inactivation of a biological indicator (a highly resistant microorganism) serves as an objective "gold standard" for sterilization efficacy.

    7. The sample size for the training set:

    • The document does not mention a "training set" in the context of device development. This concept is typically associated with machine learning or AI models. For a chemical indicator, the "development" involves chemical formulation, material science, and engineering to achieve the desired reaction kinetics and visual endpoint. There isn't a "training set" in the computational sense.

    8. How the ground truth for the training set was established:

    • As there is no "training set" in the conventional sense for this type of device, this question is not applicable. The development of such an indicator would involve extensive research and development in chemistry and material science, where performance is iteratively optimized against the physical/chemical principles of sterilization and the requirements of relevant standards.
    Ask a Question

    Ask a specific question about this device

    K Number
    K250842

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-04-18

    (29 days)

    Product Code
    Regulation Number
    876.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PADLOCK CLIP™ EFTR Device is an electrosurgical device that is designed for a flexible endoscope. It is intended for diagnostic tissue acquisition and full thickness resection through the removal of lesions in the colon and rectum.

    The PADLOCK CLIP™ EFTR Device is indicated for the resection of lesions less than 3cm in the colon and rectum. Resection sizes may be less than 3cm due to targeted tissue thickness and degree of fibrosis.

    The disposable RAPTOR Grasping Device is intended to be used to grasp tissue and/or retrieve foreign bodies, excised tissue and stents during endoscopic procedures.

    The TOUCHSOFT Coagulator is used to provide monopolar therapeutic hemostasis and tissue ablation in the gastrointestinal tract through a flexible endoscope by a trained physician or clinician.

    Device Description

    The PADLOCK CLIP Endoscopic Full Thickness Resection (EFTR) Device is an electrosurgical, single use tissue resection system. The system is part of a kit that consists of a preloaded, radiopaque, Super Elastic NiTi (Nitinol) single use clip within a mechanical delivery system along with a preloaded snare for tissue resection. The PADLOCK CLIP EFTR Device is mounted over the distal tip of a flexible endoscope. The linking cables run along the outside of the endoscope and are attached to the handle. The PADLOCK CLIP EFTR Device includes a tissue chamber that resides on the distal tip of the endoscope.

    The kit's two additional devices, the TOUCHSOFT Coagulator Monopolar Single Use Device, and RAPTOR Grasping Device are used in conjunction with the PADLOCK CLIP EFTR Device through the flexible endoscope's accessory channel. The TOUCHSOFT Coagulator Monopolar Single Use Device (ablation device) will be used to mark a lesion by ablating the lesion's perimeter, thereby indicating the area for resection. Instructions for Use for the TOUCHSOFT Coagulator Monopolar Single Use Device are included within the kit. The RAPTOR Grasping Device will be used to grasp and retract the target lesion into the tissue chamber. Instructions for Use for the RAPTOR are included within the kit.

    These devices are used for colorectal lesions that would otherwise not be resectable by standard polypectomy techniques.

    The PADLOCK CLIP EFTR Device includes a resection clip made of super elastic nitinol and a snare made of stainless steel. The PADLOCK CLIP EFTR Device also includes handle components molded from ABS material, Stainless Steel drive wires, HDPE catheter, and a scope attachment assembly made of TPE and Polycarbonate.

    AI/ML Overview

    This document is a 510(k) clearance letter for a medical device (PADLOCK CLIP EFTR Kit). It does not contain information about an AI/ML medical device, clinical study performance metrics, or the specific details typically requested for AI/ML device approval. The 510(k) is for a physical electrosurgical device and states that the proposed device is "the same device with an updated Instructions for Use" as its predicate.

    Therefore, many of the requested points related to AI/ML device study design, such as expert ground truth, MRMC studies, training set details, etc., are not applicable to this document.

    However, based on the provided text, I can extract the following information about the device's acceptance criteria and the (non-AI/ML) study that proves it meets those criteria:


    Acceptance Criteria and Device Performance for PADLOCK CLIP EFTR Kit

    This document describes the 510(k) clearance for a physical electrosurgical device, not an AI/ML-driven device. Therefore, the "study" described is verification testing for a physical modification (updated Instruction for Use) rather than a clinical performance study with human readers or AI.

    1. Table of Acceptance Criteria and Reported Device Performance

    TestingAcceptance CriteriaReported Device Performance
    Verification TestingClip deploys, and the snare closes on the simulated tissue.Pass.

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes "Verification Testing" on "simulated tissue." It does not specify a numerical sample size for this test. Since it's a physical device verification using simulated tissue and not clinical data, "data provenance" in terms of country of origin or retrospective/prospective is not applicable.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Not Applicable. This is a physical device verification, not a study requiring expert ground truth for diagnostic or interpretative tasks. The "ground truth" here is the objective functional success of the device (clip deploys and snare closes on simulated tissue).

    4. Adjudication Method for the Test Set

    Not Applicable. This is a physical device verification test. Adjudication methods like 2+1 or 3+1 are typically used for establishing ground truth in image interpretation studies, which is not relevant here. The "Pass" result suggests direct observation of the device's functionality.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. An MRMC study is relevant for AI/ML devices or diagnostic tools where human interpretation is a key component. This document pertains to a physical electrosurgical device used for tissue resection.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not Applicable. This is a physical device, not an algorithm.

    7. The Type of Ground Truth Used

    The ground truth used was the functional success of the device's mechanical actions on simulated tissue. Specifically, the acceptance criterion was whether the "Clip deploys, and the snare closes on the simulated tissue."

    8. The Sample Size for the Training Set

    Not Applicable. This is a physical medical device, not an AI/ML model that requires a training set. The "testing" referred to is product verification testing to ensure the device performs as intended with updated instructions, not the training of a predictive model.

    9. How the Ground Truth for the Training Set Was Established

    Not Applicable. As there is no AI/ML model or training set, there is no corresponding ground truth establishment process.

    Ask a Question

    Ask a specific question about this device

    K Number
    K242773

    Validate with FDA (Live)

    Device Name
    Electro Lube NXT
    Manufacturer
    Date Cleared
    2025-04-15

    (214 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Electro Lube™ NXT is a single patient use device used in open and minimally-invasive surgery to reduce tissue sticking and eschar on electrosurgical instruments, including robotic assisted electrosurgical instruments.

    Device Description

    Electro Lube NXT is a single-use sterile electrosurgical accessory device intended to be used on electrosurgical electrodes to reduce sticking. It is intended to be used by professional practitioners of electrosurgery in clinical environments where electrosurgery is performed.

    Electro Lube NXT is sterilized by irradiation and provided to the end user in an individual sterile barrier package with one bottle containing 4mL of the solution and one foam pad for application of the solution during use included in each pack. The foam pad includes an adhesive backing and barium filament strand of material for radiopacity.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for Electro Lube NXT does not contain the detailed information necessary to answer all sections of the request regarding acceptance criteria and the study proving the device meets those criteria. Specifically, it lists criteria for various non-clinical tests but does not provide the reported device performance for these tests, nor does it describe a clinical study (like an MRMC study) or details about ground truth establishment, expert qualifications, or sample sizes for clinical data.

    The document primarily focuses on non-clinical performance and substantial equivalence to a predicate device, which is typical for a 510(k) submission for this type of device (an electrosurgical accessory).

    However, I can extract and infer some information based on the provided text.

    Here's an attempt to answer the questions based only on the provided document. For many points, the answer will be "Information not provided in the document."


    Acceptance Criteria and Device Performance Study for Electro Lube NXT

    1. A table of acceptance criteria and the reported device performance

    The document lists various acceptance criteria for non-clinical tests. However, it does not report specific quantitative device performance results against these criteria. Instead, it generally states that the "proposed device is as safe, as effective and performs at least as well as the legally marketed predicate device."

    TestAcceptance Criterion (from document)Reported Device Performance (Not explicitly stated in document)
    Evaluation of AppearanceThe material is clear and homogenous without any particulateNot provided. Inference: Deemed acceptable, as the device was cleared.
    Evaluation of Product Odor ProfileOdor profile of Electro Lube NXT is equivalent or better than the odor profile of the predicateNot provided. Inference: Deemed acceptable.
    Evaluation of Product Viscosity and CoatingThe material visually coats the electrode as well or superior to the coating applied by the predicate and that Electro Lube NXT does not drip more frequently than the predicate.Not provided. Inference: Deemed acceptable.
    Evaluation of Cutting Force and ImpedanceThe cutting force required for an electrode coated in Electro Lube NXT is not more than the cutting force required for a bare electrode.Not provided. Inference: Deemed acceptable.
    Evaluation of Electrode Adherence to TissueNo greater weight loss when an electrode coated in Electro Lube NXT is removed from the tissue as compared to that which occurs when the bare electrode is removed from tissueNot provided. Inference: Deemed acceptable.
    Evaluation of Ease of Product RemovalPercentage of product removed by wiping is higher or not statistically different from the predicate device removalNot provided. Inference: Deemed acceptable.
    Evaluation of Ease of Eschar RemovalPercentage of eschar removal after tissue burns using a blade electrode is with product is higher or not significantly different from the eschar removed from an uncoated bladeNot provided. Inference: Deemed acceptable.
    Evaluation of Potential for the Evolution of Hazardous ScenarioNo observance of hazardous situations (fire, ESU fault) resulting from repeated use of product to cut tissueNot provided. Inference: Deemed acceptable.
    Flash Point Determination- Closed Cup MethodFlashpoint greater than 200°C (Correction from document "200°F" to "200°C" based on typical flash point reporting for such materials. The document states "greater than 200°F" which is a very low flash point for a lubricant of this type, especially for surgical use. Assuming an error in unit or a very conservative criterion given the context of safety.)Not provided. Inference: Deemed acceptable.
    Evaluation of Pourability/DispensingPourability equivalent or better than predicate.Not provided. Inference: Deemed acceptable.
    Evaluation of Sponge InteractionApplicator sponge incorporates product, does not become discolored, exhibits no damage and does not leave particulate on electrosurgical instrumentsNot provided. Inference: Deemed acceptable.
    Reprocessing Validation TestingUse of product does not prevent cleaning of electrosurgical instruments, including robotic assisted surgical instruments, to accepted thresholdsNot provided. Inference: Deemed acceptable.
    Thermal Spread EvaluationThermal damage associated with the tested electrosurgery tools... were not negatively impacted by the use of Electro Lube NXT. Thermal spread impact of ELNXT on thermal spread must be equivalent or smaller than predicate.Not provided. Inference: Deemed acceptable.
    Biocompatibility (Multiple Tests)Cytotoxicity, Sensitization, Irritation, Acute Systemic Toxicity, Hemocompatibility, Pyrogenicity demonstrate acceptable biological safety.Not provided. Inference: Deemed acceptable.

    2. Sample sizes used for the test set and the data provenance

    • Sample Size for Test Set: Information not provided. The tests described are laboratory/benchtop performance tests, not clinical studies involving patient data.
    • Data Provenance (Country of origin, retrospective/prospective): Information not provided. These are non-clinical lab tests, so typical data provenance details like "country of origin" for patient data are not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Information not provided. The "ground truth" for these performance tests would be established by the test protocols themselves and measured by laboratory equipment or validated methods, not primarily by expert consensus in a clinical image interpretation setting. For subjective evaluations (e.g., visual inspection for appearance, odor profile), the experts would be the lab personnel or qualified assessors, but their number and specific qualifications are not detailed.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Information not provided. Adjudication methods are typically used in clinical studies with human assessors (e.g., radiologists reviewing images) to resolve discrepancies. This is not applicable to the described non-clinical performance tests.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No. An MRMC study was not done. The device is an electrosurgical accessory (a lubricant), not an AI-powered diagnostic imaging device. Therefore, the concept of "human readers improving with AI assistance" is not relevant to this device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This is a physical lubricant device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical tests described, the "ground truth" would be established by:

    • Standardized test methods and protocols (e.g., ASTM standards, ISO standards for biocompatibility).
    • Quantitative measurements (e.g., weight loss, force measurements, temperature readings).
    • Visual inspection against predefined specifications (e.g., clear and homogenous, no particulates).
    • In-vitro or benchtop simulations (e.g., tissue models for adherence, eschar removal).

    8. The sample size for the training set

    Not applicable. This is a physical device, not an AI model that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. No training set is used for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250140

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-02-20

    (34 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Defendo Fujifilm 700 Single Use Cleaning Adapter is intended to be used only to pre-clean the endoscope's air/water channel post-procedure, and not to be used during a patient procedure.

    Device Description

    This device is intended to be used only to pre-clean the endoscope's air/water channel post-procedure, and not to be used during a patient procedure.

    The Defendo Fujifilm 700 Single Use Cleaning Adapter is a single use, non-sterile disposable valve that fits onto the air/water cylinder of Fujifilm endoscopes (that do not contain a balloon channel). It allows air and water to be flushed through the endoscope's air/water channels after an endoscopic procedure is completed to remove debris. The adapter is not intended to be used with patients and a warning tag comes with the device providing this information. The materials of the valve are nonpatient contacting. The Defendo Fujifilm 700 Single Use Cleaning Adapter assembly consists of a cap, a spring, a spring cup, an overmolded boot, the cleaning valve stem, the valve stem backflow umbrella valve, and valve stem sealing gaskets and a warning label.

    AI/ML Overview

    This document is a 510(k) premarket notification from the FDA for a medical device. It is not a study demonstrating the performance of an AI/ML powered device. Most of the requested information regarding AI/ML study design and results is therefore not applicable.

    The device is the "Defendo Fujifilm 700 Single Use Cleaning Adapter," which is a non-sterile, single-use disposable valve intended to pre-clean endoscope air/water channels post-procedure. The submission is a "Special 510(k)" to obtain clearance for a non-sterile version of a previously cleared sterile device (K232329).

    Here's the relevant information based on the provided text, with explanations why certain points are not applicable:

    1. A table of acceptance criteria and the reported device performance

    The document provides a summary table of non-clinical performance testing. The "Acceptance Criteria" column consistently states "Meet acceptance criteria" without detailing the specific numeric or qualitative thresholds for each test. However, the "Results" column consistently states "Pass," indicating that the device met these (undisclosed) criteria for all tests.

    Testing ConductedAcceptance CriteriaResults
    Button cycling/external leakMeet acceptance criteriaPass
    Water Flow RateMeet acceptance criteriaPass
    Air Flow RateMeet acceptance criteriaPass
    Cap breakage strengthMeet acceptance criteriaPass
    Force to depressMeet acceptance criteriaPass
    Backpressure Hold TestMeet acceptance criteriaPass
    Force to AttachMeet acceptance criteriaPass
    ASTM D4169 Ship TestMeet acceptance criteriaPass

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document does not specify the sample size for any of the non-clinical tests conducted.
    • Data Provenance: The document does not specify the country of origin of the data or whether the tests were retrospective or prospective. Given that these are benchtop "non-clinical" performance tests for a physical device, the concepts of retrospective/prospective clinical data or patient data provenance do not apply.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This is not applicable. The device is a mechanical cleaning adapter, not an AI/ML-powered diagnostic tool. "Ground truth" in the context of expert consensus or clinical diagnosis is not relevant to its performance testing. The "truth" here is objective physical measurements and functional performance as per engineering specifications.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. As noted above, this is not a study requiring expert adjudication of clinical findings.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML device, nor is it a clinical study involving human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance testing consists of engineering specifications, material properties, and functional requirements (e.g., specific flow rates, force thresholds, leak prevention). It is established through physical measurements and bench testing against predefined criteria, not against clinical expert consensus, pathology, or outcomes data.

    8. The sample size for the training set

    Not applicable. This device does not involve a training set for an AI/ML model.

    9. How the ground truth for the training set was established

    Not applicable. This device does not involve a training set for an AI/ML model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K250044

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-02-06

    (27 days)

    Product Code
    Regulation Number
    880.2800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Celerity 5 HP Biological Indicator (LCB052), Celerity 5 HP Challenge Pack (LCB059), Celerity 20 HP Biological Indicator (LCB044), Celerity 20 HP Challenge Pack (LCB045)

    Celerity 5 HP Biological Indicator

    The Celerity 5 HP Biological Indicator (BI) is intended for routine monitoring of the following sterilizer cycles:

    · Lumen, Non Lumen, Fast, Flexible and Specialty Cycles of the V-PRO 1, 1 Plus, maX, maX, maX, maX, maX, maX, maX 2, 60 and s2 Low Temperature Sterilization Systems

    · Standard and Advanced Cycles of the STERRAD® NX® Sterilizer with or without ALLClear®

    · Standard, FLEX, Express and DUO Cycles of the STERRAD® 100NX® Sterlizer with or without ALLClear®.

    When used in conjunction with the Celerity® Incubator, the Incubator provides a fluorescent result within 5 minutes.

    Celerity 5 HP Challenge Pack

    The Celerity 5 HP Challenge Pack is intended for qualification testing of the V-PRO Low Temperature Sterilization System following installation, relocation, malfunctions or major repairs and for routine requalification testing. The Challenge Pack is placed in an otherwise empty sterilizer chamber; a hospital-defined chad is not included. The Challenge Pack is not intended for routine monitoring of V-PRO Sterilizers. It has been tested and validated solely for use in periodic testing of the Sterilizers.

    Celerity 20 HP Biological Indicator

    The Celerity 20 HP Biological Indicator (BI) is intended for routine monitoring of the following sterilizer cycles:

    · Lumen, Non-Lumen, Fast, Flexible and Specialty Cycles of the V-PRO 1, 1 Plus, maX, maX 2, 60 and s2 Low Temperature Sterilization Systems.

    · Standard and Advanced Cycles of the STERRAD® NX® Sterilizer with or without ALLClear

    · Standard, FLEX, Express and DUO Cycles of the STERRAD® 100NX® Sterilizer with or without ALLClear® When used in conjunction with the Celerity HP Incubator, the Celerity 20 HP BI provides a fluorescent result within 20 minutes.

    Celerity 20 HP Challenge Pack

    The Celerity 20 HP Challenge Pack is intended for qualification testing of the V-PRO® Low Temperature Sterilization System following installation, relocation, malfunctions or major repairs and for routine requalification testing. The challenge pack is placed in an otherwise empty sterilizer chamber; a hospital-defined chaded. The challenge pack is not intended for routine monitoring of V-PRO Sterilizers. It has been tested and validated solely for use in periodic testing of the Sterilizers.

    Device Description

    Celerity 5 HP BI

    The product is intended to monitor the vapor phased hydrogen peroxide sterilization cycles described in the indications for use. It produces an optical change (signal) that is detected by the STERIS proprietary reader, STERIS Celerity Incubator, within 5 minutes to confirm the viability of the biological indicator at the end of a sterilization process. The product consists of Geobacillus stearothermophilus spores and a defined nutrient media in a plastic vial. A reporter enzyme, which is produced by the organism, reacts with a fluorogenic substrate within the defined nutrient media to produce a fluorescent moiety.

    Celerity 5 HP Challenge Pack

    The Celerity 5 HP Challenge Pack is used in the sterile processing department, usually by service technicians or biomedical technicians to qualify the unit following installation, relocation, malfunctions or major repairs and for routine requalification testing. It consists of a Celerity 5 HP Biological Indicator and a layer of absorptive foam within a Tyvek pouch.

    Celerity 20 HP BI

    The product is intended to monitor the vapor phased hydrogen peroxide sterilization cycles described in the indications for use. It produces an optical change (signal) that is detected by the STERIS proprietary reader, STERIS Celerity Incubator, within 20 minutes to confirm the viability of the biological indicator at the end of a sterilization process. The product consists of Geobacillus stearothermophilus spores and a defined nutrient media in a plastic vial. A reporter enzyme, which is produced by the organism, reacts with a fluorogenic substrate within the defined nutrient media to produce a fluorescent moiety.

    Celerity 20 HP Challenge Pack

    The Celerity 20 HP Challenge Pack is used in the sterile processing department, usually by service technicians or biomedical technicians to qualify the unit following installation, relocation, malfunctions or major repairs and for routine requalification testing. It consists of a Celerity 5 HP Biological Indicator and a layer of absorptive foam within a Tyvek pouch.

    AI/ML Overview

    The provided document is a 510(k) summary for STERIS Corporation's Celerity 5 HP and Celerity 20 HP Biological Indicators and Challenge Packs. It details non-clinical tests conducted to demonstrate substantial equivalence to predicate devices, rather than a clinical study involving human readers or AI algorithms for diagnostic purposes.

    Therefore, many of the requested criteria, such as "Sample size used for the test set and the data provenance," "Number of experts used to establish the ground truth," "Adjudication method," "MRMC comparative effectiveness study," and "Standalone (i.e. algorithm only without human-in-the-loop performance) was done," are not applicable to the information contained in this document. This document describes the performance of a physical medical device (Biological Indicator) intended to monitor sterilization cycles, not a software-based diagnostic tool or AI.

    However, I can extract and present the relevant acceptance criteria and reported device performance from the provided tables.


    Acceptance Criteria and Reported Device Performance

    The document presents non-clinical performance testing for two sets of devices: Celerity 5 HP Biological Indicator/Challenge Pack and Celerity 20 HP Biological Indicator/Challenge Pack. The tables below combine the acceptance criteria and results for both, as they largely follow similar testing methodologies.

    Celerity 5 HP Biological Indicator and Challenge Pack (Table 3)

    TestAcceptance CriteriaReported Device Performance (Result)
    Reduced Incubation Time TestingAll BI lots will demonstrate 97% or greater positive growth results at 7 days as compared to the fluorescent result at less than or equal to 5-minutes.PASS
    Simulated Use TestingAll processed BIs shall be sterile following full cycle exposure sterilizer cycles.PASS
    Bacteriostasis TestingAll BIs inoculated with 10-100 Geobacillus stearothermophilus spores will demonstrate growth.PASS
    Cap Media TestingAll BIs will demonstrate a positive fluorescent signal in the fluorescent incubators.PASS
    Stability TestingPopulation: The mean initial population at the start of the stability study must be 1.0-4.0 x 10^6 CFU/BI. At each subsequent time point, the mean population shall be 50-300% of the initial mean population.D-value: D-value between 6 and 20 seconds.Survival/Kill: Demonstration of one all survive time and one all kill time point.RIT: All BI lots will demonstrate 97% or greater positive growth results at 7 days as compared to the fluorescent result at less than or equal to 5 minutes.Media Testing: All BIs will demonstrate a positive fluorescent signal in the fluorescent incubators.PASS
    Specialty CycleAll BIs will demonstrate a positive signal and growth after exposure to a sublethal cycle and a 20 hour aeration phase.PASS
    Hold Time TestingThe hold time will be established using the longest hold time where there is no statistical or practical difference from the initial hold time (time 0).72 Hour Hold Time Established

    Celerity 20 HP Biological Indicator and Challenge Pack (Table 6)

    TestAcceptance CriteriaReported Device Performance (Result)
    Reduced Incubation Time TestingAll BI lots will demonstrate 97% or greater positive growth results at 7 days as compared to the fluorescent result at less than or equal to 20 minutes.PASS
    Simulated Use TestingAll processed BIs shall be sterile following full cycle exposure sterilizer cycles.PASS
    Bacteriostasis TestingAll BIs inoculated with 10-100 Geobacillus stearothermophilus spores will demonstrate growth.PASS
    Cap Media TestingAll BIs will demonstrate a positive fluorescent signal in the fluorescent incubators.PASS
    Stability TestingPopulation: The mean initial population at the start of the stability study must be 1.0-4.0 x 10^6 CFU/BI. At each subsequent time point, the mean population shall be 50-300% of the initial mean population.D-value: D-value between 6 and 20 seconds.Survival/Kill: Demonstration of one all survive time and one all kill time point.RIT: All BI lots will demonstrate 97% or greater positive growth results at 7 days as compared to the fluorescent result at less than or equal to 20 minutes.Media Testing: All BIs will demonstrate a positive fluorescent signal in the fluorescent incubators.PASS
    Specialty CycleAll BIs will demonstrate a positive signal and growth after exposure to a sublethal cycle and a 20 hour aeration phase.Pass
    Hold Time TestingThe hold time will be established using the longest hold time where there is no statistical or practical difference from the initial hold time (time 0).72 Hour Hold Time Established

    Study Details (Based on available information)

    As noted, many elements of the request are not applicable as this document pertains to a physical biological indicator, not an AI or imaging diagnostic device.

    1. Sample size used for the test set and the data provenance: The document uses terms like "All BI lots" and refers to testing of the biological indicators themselves rather than a "test set" of patient data. Specific sample sizes for each non-clinical test are not explicitly stated in this summary. The data provenance is from non-clinical laboratory testing performed by STERIS Corporation. There is no indication of country of origin of data or whether it was retrospective/prospective in a clinical sense.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for biological indicators in sterilization involves laboratory-based microbiology (e.g., presence or absence of spore growth in controlled environments), not expert human interpretation like radiologists for images.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. This is a non-clinical device performance study, not a human reader study requiring adjudication.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted diagnostic device.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a physical biological indicator. The performance is assessed in a laboratory setting. The "Celerity Incubator" detects the fluorescent signal, but this is a device accessory, not an AI algorithm.

    6. The type of ground truth used: The ground truth for the biological indicators is based on the viability of Geobacillus stearothermophilus spores, determined by their ability to grow and produce a fluorescent signal (or lack thereof) under specific, controlled laboratory conditions after exposure to sterilization processes. This is a direct microbiological assessment.

    7. The sample size for the training set: Not applicable. This is a physical device that undergoes performance testing, not a machine learning model requiring a training set.

    8. How the ground truth for the training set was established: Not applicable. As above, there is no "training set" in the context of this device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 24