510(k) Data Aggregation

    K Number
    K213373

    Validate with FDA (Live)

    Date Cleared
    2022-03-10

    (149 days)

    Product Code
    Regulation Number
    880.6250
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Powder-Free Green Biodegradable Nitrile Examination Glove Tested for Use with Chemotherapy Drugs and Fentany] Citrate is a disposable device intended for medical purposes that is worn on the examiner's hand to prevent contamination between patient and examiner. It is also tested to be used against Chemotherapy Drugs and Fentanyl Citrate.

    The gloves were tested with chemotherapy drugs and fentanyl citrate as per ASTM D6978-05, Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    Device Description

    Powder-Free Green Biodegradable Nitrile Examination Glove Tested for Use with Chemotherapy Drugs and Fentanyl Citrate

    AI/ML Overview

    This FDA 510(k) clearance letter pertains to medical examination gloves and does not describe an AI/ML powered medical device. Therefore, the questions related to AI/ML specific criteria (such as sample sizes for test and training sets, expert qualifications, ground truth establishment methods, MRMC studies, or standalone algorithm performance) are not applicable to this document.

    The document focuses on the performance of a physical medical device (gloves) against specific chemical permeation tests.

    Here's the relevant information based on the provided text, focusing on the acceptance criteria and study for the gloves:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device performance is evaluated based on its resistance to permeation by chemotherapy drugs and Fentanyl Citrate. The acceptance criterion is implied by the testing methodology using ASTM D6978-05, aiming for a long breakthrough detection time. While an explicit "acceptance criteria" value isn't stated in the table for each drug, the "Average Breakthrough Detection Time (mins)" provides the reported performance directly. For most drugs, a breakthrough time of "> 240 minutes" is presented, indicating good resistance for at least 4 hours.

    Test drug and concentrationAcceptance Criteria (Implied: High DBT)Reported Device Performance (Average Breakthrough Detection Time (mins))
    Carmustine (BCNU), 3.3 mg/ml (3,300 ppm)High DBT13.5
    Cisplatin, 1.0 mg/ml (1,000 ppm)High DBT> 240
    Cyclophosphamide (Cytoxan), 20.0 mg/ml (20,000 ppm)High DBT> 240
    Dacarbazine, 10.0 mg/ml (10,000 ppm)High DBT> 240
    Doxorubicin HCL, 2.0 mg/ml (2,000 ppm)High DBT> 240
    Etoposide, 20.0 mg/ml (20,000 ppm)High DBT> 240
    Fluorouracil, 50.0 mg/ml (50,000 ppm)High DBT> 240
    Ifosfamide, 50.0 mg/ml (50,000 ppm)High DBT> 240
    Methotrexate, 25.0 mg/ml (25,000 ppm)High DBT> 240
    Mitomycin, 0.5 mg/ml (500 ppm)High DBT> 240
    Mitoxantrone, 2.0 mg/ml (2,000 ppm)High DBT> 240
    Paclitaxel, 6.0 mg/ml (6.000 ppm)High DBT> 240
    ThioTepa, 10.0 mg/ml (10,000 ppm)High DBT19.0
    Vincristine Sulfate, 1.0 mg/ml (1,000 ppm)High DBT> 240
    Fentanyl Citrate Injection, 100 mcg/2mlHigh DBT> 240

    Note: The cautionary statement "CAUTION: Do not use with Carmustine and ThioTepa" implies that the breakthrough times for these two drugs (13.5 and 19.0 minutes respectively) are considered insufficient for safe use, effectively setting an implicit acceptance criterion that these drugs fail it for prolonged exposure.

    2. Sample size used for the test set and the data provenance

    The document does not specify the exact sample size for the test set (number of gloves tested per drug). It only refers to a standard test method. The data provenance is not explicitly stated beyond being the result of testing the manufacturer's "Powder-Free Green Biodegradable Nitrile Examination Glove." The testing is prospective for the purpose of this submission. The country of origin of the data is not specified, but the manufacturer is based in Malaysia.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    Not applicable. This is a physical device performance test (chemical permeation), not an AI/ML diagnostic or image interpretation. "Ground truth" in this context is the objective measurement of chemical breakthrough time, typically conducted in a laboratory by trained technicians following a standardized protocol.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. This is a physical device performance test, not involving human interpretation or adjudication in the sense of comparing opinions. The results are quantitative measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/ML powered device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML powered device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this type of device is the objective measurement of chemical permeation/breakthrough time determined by laboratory testing protocols, specifically ASTM D6978-05, Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

    8. The sample size for the training set

    Not applicable. This is not an AI/ML powered device.

    9. How the ground truth for the training set was established

    Not applicable. This is not an AI/ML powered device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181047

    Validate with FDA (Live)

    Date Cleared
    2019-01-09

    (265 days)

    Product Code
    Regulation Number
    878.4460
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    A Patient examination glove is a disposable device intended for medical purposes that worn on the examiner's hand or finger to prevent containation between patient and examiner. Tested for use with chemotherapy drugs as listed below.

    Device Description

    Powder Free White, Blue Sterilized Nitrile Copolymer Surgical Gloves - Tested for Use with Chemotherapy Drugs.

    AI/ML Overview

    The document describes a medical device called "Powder Free White, Blue Sterilized Nitrile Copolymer Surgical Gloves - Tested for Use with Chemotherapy Drugs" and its clearance for marketing. The acceptance criteria and the study proving the device meets these criteria are specifically related to the gloves' permeation resistance to various chemotherapy drugs.

    Here's the breakdown of the requested information:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The implicit acceptance criterion for gloves tested with chemotherapy drugs is to demonstrate a reasonable "Breakthrough Detection Time," meaning the time it takes for the chemotherapy drug to permeate through the glove material. While not explicitly stated as a pass/fail threshold in the provided text, the standard practice in evaluating such gloves is typically referenced against established guidelines (e.g., ASTM standards for barrier protection). The performance reported is the actual breakthrough detection time for each drug.

    Chemotherapy Drugs:Acceptance Criteria (Implied)White Glove Reported Performance (Breakthrough Detection Time)Blue Glove Reported Performance (Breakthrough Detection Time)
    1. Carmustine (BCNU) 3.3 mg/ml (3,300 ppm)High breakthrough time79.0 min48.5 min
    2. Cisplatin 1.0 mg/ml (1,000 ppm)High breakthrough time>240 min>240 min
    3. Cyclophosphamide (Cytoxan) 20mg/ml (20,000ppm)High breakthrough time>240 min>240 min
    4. Dacarbazine (DTIC) 10.0 mg/ml (10,000ppm)High breakthrough time>240 min>240 min
    5. Doxorubicin Hydrochloride 2.0 mg/ml (2,000ppm)High breakthrough time>240 min>240 min
    6. Etoposide (Toposar) 20.0 mg/ml (20,000ppm)High breakthrough time>240 min>240 min
    7. Fluorouracil 50.0 mg/ml (50,000ppm)High breakthrough time>240 min>240 min
    8. Ifosfamide 50.0 mg/ml (50,000ppm)High breakthrough time>240 min>240 min
    9. Methotrexate 25 mg/ml (25,000ppm)High breakthrough time>240 min>240 min
    10. Mitomycin C 0.5 mg/ml (500ppm)High breakthrough time>240 min>240 min
    11. Mitoxantrone 2.0 mg/ml (2,000ppm)High breakthrough time>240 min>240 min
    12. Paclitaxel (Taxol) 6.0 mg/ml (6,000ppm)High breakthrough time>240 min>240 min
    13. Thio-Tepa 10.0 mg/ml (10,000ppm)High breakthrough time79.2 min169.4 min
    14. Vincristine Sulfate 1.0 mg/ml (1,000ppm)High breakthrough time>240 min>240 min

    Note: The document explicitly states: "Please Note that Carmustine and Thio-Tepa have low permeation time of less than 240 minutes." This indicates that while higher numbers are generally better, specific breakthrough times are reported and acknowledged. The "acceptance criteria" here is simply the demonstration and reporting of these values to show the glove's performance against specific drugs, allowing users to make informed decisions.

    2. Sample Size Used for the Test Set and the Data Provenance:

    The document does not explicitly state the sample size used for testing each glove type with each chemotherapy drug. The provenance of the data is also not explicitly stated in terms of country of origin or whether it was retrospective or prospective. However, given that it's a 510(k) submission to the FDA, it implies that the testing was conducted in a controlled, prospective manner to meet regulatory requirements, likely in a laboratory setting adhering to recognized standards (e.g., ASTM D6978).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    This type of information is not applicable (N/A) to the clearance of surgical gloves for chemotherapy drug permeation. The "ground truth" here is objective scientific measurement (breakthrough detection time) determined through laboratory testing, not subjective expert assessment as would be the case for image interpretation or diagnostic accuracy.

    4. Adjudication Method for the Test Set:

    This is N/A. Since the "ground truth" is established by objective laboratory measurements, there is no need for expert adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done:

    This is N/A. Surgical gloves are not diagnostic devices that involve "readers" or "cases" in the context of comparative effectiveness studies against AI. The document focuses on the material's barrier properties.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    This is N/A. The device is a physical product (gloves), not an algorithm or AI system.

    7. The Type of Ground Truth Used:

    The ground truth used is objective laboratory measurement of Breakthrough Detection Time obtained through standardized chemical permeation testing (e.g., following a standard like ASTM D6978), which measures the time it takes for a chemotherapy drug to permeate through the glove material.

    8. The Sample Size for the Training Set:

    This is N/A. The device is a physical product, not an AI model, so there is no "training set."

    9. How the Ground Truth for the Training Set Was Established:

    This is N/A for the same reason as point 8.