Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K242188

    Validate with FDA (Live)

    Device Name
    ClearRead CT CAC
    Date Cleared
    2024-12-03

    (131 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ClearRead CT CAC is image processing software designed to aid physicians in assessing coronary artery calcification on non-gated, non-contrast, standard, or low-dose chest CT scans of adult patients 30 years of age or older. The primary output of ClearRead CT CAC is the Agatston score. ClearRead CT CAC locates calcified coronary lesions and assigns them to one of several coronary arteries: the right coronary artery (RCA), the left main artery (LM), the left anterior descending artery (LAD), and the left circumflex artery (LCX). The software calculates the Agatston score as the sum for each artery while also providing the total across all coronary arteries. The total Agatston score is calibrated to 3.0mm slice thickness and is used to derive a CAC category and the arterial age. The provided segmentations are for illustrative purposes only and are not intended for diagnostic use.

    ClearRead CT CAC provides adjunctive information and is not intended to be used without clinical expert review.

    Device Description

    ClearRead CT CAC is a computer aided quantification system intended to aid in the assessment of coronary artery calcification. The system receives chest Computed Tomography (CT) studies as input, in DICOM® format, and generates output in DICOM (or other) format.

    AI/ML Overview

    Here's a detailed breakdown of the acceptance criteria and the study proving the device meets them, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Metric / EndpointAcceptance Criteria (Internal Investigation/Literature)Reported Device Performance (Clinical Study)
    Agatston Score Category AgreementWeighted Kappa > 0.85 (considered excellent)0.959 (95% CI: 0.943-0.975)
    CAC Negative Agreement (NPA)Not explicitly stated as acceptance criteria, but reported.99.1% (95% CI: 98.0%, 100.0%) for Category 0 vs. Not Category 0
    CAC Positive Agreement (PPA)Not explicitly stated as acceptance criteria, but reported.PPA for Category 0: 88.1% (95% CI: 82.5%, 93.2%)PPA for Category 1-99: 92.8% (95% CI: 87.7%, 96.9%)PPA for Category 100-399: 88.1% (95% CI: 80.5%, 94.4%)PPA for Category ≥ 400: 98.6%
    Overall Continuous Agatston Score AgreementBland-Altman analysis showing strong agreement (qualitative, based on Figure 2 from non-clinical). No specific numerical acceptance criteria provided for LoA.Limits of Agreement (LoA) of -12.20 ± 437.84
    Coronary Artery Calcification Segmentation AccuracyNot explicitly stated as a separate acceptance criterion with a specific numerical threshold.Average Dice score: 0.892 (0.871, 0.911) (Device vs. GT)

    2. Sample Size for Test Set and Data Provenance

    • Non-clinical testing: Approximately 400 cases. Data provenance is not specified, but it's referred to as "internal investigation."
    • Clinical testing: 491 cases (143 cases with no CAC, 348 with CAC). The data included samples from four CT manufacturers across five clinical sites. The specific country of origin is not mentioned, nor is it explicitly stated whether the data was retrospective or prospective, though the mention of "cases collected and held out from development" suggests retrospective.

    3. Number of Experts and Qualifications for Ground Truth

    • Clinical testing: Three radiologists established the consensus ground truth review. Their specific qualifications (e.g., years of experience, subspecialty) are not detailed in the provided text.

    4. Adjudication Method for the Test Set

    • Clinical testing: A "consensus ground truth review by three radiologists" was used. This implies that the three radiologists collectively determined the ground truth for each case. The specific mechanism (e.g., majority vote, discussion until agreement) is not specified, but it suggests a form of 3-reader consensus.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No MRMC comparative effectiveness study was done. The study performed was a standalone model performance assessment comparing the algorithm's output to ground truth measurements. The document does not describe human readers improving with AI vs. without AI assistance.

    6. Standalone Performance Assessment

    • Yes, a standalone (algorithm only without human-in-the-loop performance) study was done. The document explicitly states: "This study was a standalone model performance assessment of ClearRead CT CAC."

    7. Type of Ground Truth Used

    • Clinical testing: "Consensus ground truth review by three radiologists." This is a form of expert consensus.

    8. Sample Size for Training Set

    • The sample size for the training set is not provided in the given document. The document mentions that approximately 400 cases were "held out from development of any kind to estimate the test performance," implying these were for the test set, not training.

    9. How the Ground Truth for the Training Set was Established

    • The document does not provide information on how the ground truth for the training set was established, as the details focus on the "held out" test set for performance estimation. For the test set, "Regions associated with coronary artery calcification were manually outlined, including the particular artery, from which the ground truth Agatston score was obtained."
    Ask a Question

    Ask a specific question about this device

    K Number
    K221612

    Validate with FDA (Live)

    Device Name
    ClearRead CT
    Date Cleared
    2022-12-05

    (185 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ClearRead CT is comprised of computer-assisted reading tools designed to aid the radiologist in the detection and characterization of pulmonary nodules during the review of screening and surveillance (low-dose) CT examinations of the chest on a non-oncological patient population. ClearRead CT requires both lungs be in the field of view and is not intended for monitoring patients undergoing therapy for lung cancer or limited field of view CT scans. ClearRead CT provides adjunctive information and is not intended to be used without the original CT series.

    Device Description

    ClearRead CT Compare is a post-processing application which processes a prior chest CT to determine whether a nodule detected in the current exam was present in the prior exam using the same detection algorithm used on the current exam. ClearRead CT Compare requires both lungs to be in the field of view. ClearRead CT Compare provides adjunctive information and is not intended to be used without the original CT series and is only invoked on those patients where a prior exam exists and if a nodule is detected in the current exam. ClearRead CT Compare receives images according to the DICOM® protocol, processes the Lung CT series, and delivers the resulting information through the same DICOM network interface in conjunction with results provided for the current exam, specifically whether the nodule is present on the prior exam and if so, the percent volume change between the current and prior exam along with the volume doubling time. Series inputs are limited to Computed Tomography (CT). The ClearRead CT Compare Processor processes each prior series received. The ClearRead CT Compare output is sent to a destination device that conforms to the ClearRead CT DICOM Conformance Statement, such as a storage archive. ClearRead CT Compare does not support printing or DICOM media. ClearRead CT Compare is a product extension of our FDA cleared and marketed ClearRead CT device (K161201). The initial device contained ClearRead CT Vessel Suppress as well as ClearRead CT Detect. ClearRead CT (the base system), includes normalization, segmentation, and characterization of nodules, and provides the following key features: ● ClearRead CT Vessel Suppress aids radiologists by suppressing normal structures in the input chest CT series. . ClearRead CT Detect aids radiologists in the detection and characterizations of nodules in the input chest CT series. ● ClearRead CT Compare includes Scan Registration and Nodule Matching functions and adds the following key features: ClearRead CT Compare aids radiologists in ● tracking nodule changes over time, providing additional characterizations per nodule, including percent nodule change and volume doubling time.

    AI/ML Overview

    Here's a detailed breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Nodule Match Rate (Overall Target)Minimum of 90% for each selected stratum.
    Nodule Match Rate (Solid Nodule Type)0.961 (0.952, 0.978) - Exceeds 90% benchmark
    Nodule Match Rate (Part-solid Nodule Type)0.957 (0.942, 0.971) - Exceeds 90% benchmark
    Nodule Match Rate (Ground Glass Nodule Type)0.946 (0.934, 0.965) - Exceeds 90% benchmark
    Nodule Match Rate (Isolated Nodule Location)0.940 (0.934, 0.947) - Exceeds 90% benchmark
    Nodule Match Rate (Juxta-Vascular Nodule Location)0.969 (0.963, 0.975) - Exceeds 90% benchmark
    Nodule Match Rate (Juxta-Pleura Nodule Location)0.955 (0.949, 0.961) - Exceeds 90% benchmark
    Volume Doubling Time (VDT) and % Change Calculation AccuracyManual and automated calculations matched in every instance.
    Vessel Suppress Performance for Thicker Slices (3.5mm - 5mm)Performance across different slice thickness values (3.5mm, 4.0mm, 4.5mm, 5.0mm) had little impact compared to baseline (1mm data), remaining well within the predefined 10% significance threshold for rejecting a test (non-contrast and contrast cases). Average performance change ranged from -4.5% to 4.6%.
    Nodule Registration ErrorAverage registration error of 4.46mm, with a standard deviation of 2.69mm, well within the predefined 15mm tolerance.
    New Nodule Identification (Real Cases)All 3 new nodules were detected and correctly identified as new.

    2. Sample Size Used for the Test Set and Data Provenance

    • Quantitative Nodule Matching Performance: A total of 900 nodules were used for assessment.
    • Vessel Suppress Thicker Slice Evaluation: The same data previously used to assess vessel suppression performance for 1mm to 3mm slice thickness was used, extended to include 3.5mm, 4.0mm, 4.5mm, and 5.0mm.
    • Clinical Performance Testing (Real Nodules): A 25-patient cohort containing 40 real nodules (42 actionable nodules identified by radiologists, with 39 having prior counterparts and 3 being new).

    Data Provenance:
    The document does not explicitly state the country of origin for the data. It also does not specify whether the data was retrospective or prospective for the 900 nodules or the vessel suppress evaluation. For the "clinical performance testing," the use of "real nodules" from a "25 patient cohort" suggests existing patient data, which is typically retrospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Quantitative Nodule Matching Performance: The document does not explicitly state how the ground truth for the 900 nodules was established, nor the number or qualifications of experts involved. It only states that "detected nodules were split into three categories based on their attenuation pattern."
    • Clinical Performance Testing (Real Nodules): Ground truth for the 42 actionable nodules was established by "the radiologist." The document uses "radiologist" in the singular, implying one expert, but does not provide specific qualifications (e.g., years of experience).

    4. Adjudication Method for the Test Set

    The document does not describe a formal adjudication method (like 2+1 or 3+1) for establishing the ground truth of the test sets. For the "clinical performance testing," it states "42 actionable nodules were ground-truthed by the radiologist," implying a single observer established the ground truth.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No, an MRMC comparative effectiveness study involving human readers' improvement with AI vs. without AI assistance was not explicitly described or presented in the provided text. The studies focused on the performance of the ClearRead CT Compare algorithm itself, not its impact on human reader performance.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Yes, the studies presented primarily focus on standalone (algorithm only) performance.

    • The nodule matching performance tables (Tables 5.2 and 5.3) report the algorithm's match and mismatch rates.
    • The VDT and % change calculation accuracy was checked against "manual computation," implying a comparison of algorithm output to a reference, not human interaction.
    • The vessel suppress evaluation directly assesses the algorithm's output (residual changes).
    • The "clinical performance testing" assessed the algorithm's ability to detect and match nodules and identify new ones, again without a human-in-the-loop component for the reported metrics.

    The device's indication for use explicitly states it provides "adjunctive information and is not intended to be used without the original CT series," suggesting it is designed to aid radiologists, but the studies described focus on its internal performance metrics rather than its performance in an assisted reading workflow.

    7. The Type of Ground Truth Used

    • Quantitative Nodule Matching Performance: The type of ground truth used for the 900 nodules is not explicitly stated beyond classification by attenuation pattern.
    • Volume Doubling Time and % Change Calculation: Manual computation of these values was used as ground truth.
    • Vessel Suppress Performance for Thicker Slices: The ground truth for this evaluation appears to be based on a baseline performance (1mm data) and a predefined significance threshold for residual changes. The "residual analysis" implies comparing the algorithm's output to an expected or ideal output for vessel suppression.
    • Clinical Performance Testing (Real Nodules): Expert radiologist assessment ("ground-truthed by the radiologist") was used to identify "actionable nodules" and their presence in prior scans.

    8. The Sample Size for the Training Set

    The document does not provide information regarding the sample size used for the training set for any component of ClearRead CT or ClearRead CT Compare.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information regarding how the ground truth for the training set was established, as the training set details are not mentioned.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213566

    Validate with FDA (Live)

    Date Cleared
    2022-03-10

    (121 days)

    Product Code
    Regulation Number
    892.2080
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ClearRead Xray Pneumothorax is a notification-only triage workflow tool for use by trained professionals to help prioritize chest X-rays. The device operates in parallel to and independent of standard of care image interpretation workflow. Specifically, the device uses an artificial intelligence algorithm to analyze images for features suggestive of a pneumothorax 5 mm or larger; it makes case-level output available to a PACS/workstation for worklist prioritization or triage. Identification of cases suspected of containing a pneumothorax is not for diagnostic use beyond notification. ClearRead Xray Pneumothorax is limited to analysis of imaging data as a guide to possible urgency of adult chest X-ray image review and should not be used in lieu of full patient evaluation or relied upon to make or confirm diagnoses. The device does not replace review and diagnosis of the Xrays by trained professionals. The device is not intended to be used with plain film X-ray.

    Device Description

    ClearRead Xray Pneumothorax is comprised of a computer assisted triaging tool, designed to prioritize chest X-rays based on the suspected presence of a pneumothorax (PTX) 5mm or larger. ClearRead Xray Pneumothorax requires both lungs to be in the field of view. ClearRead Xray Pneumothorax provides adjunctive information and is not intended to be used for diagnosis. ClearRead Xray Pneumothorax receives images according to the DICOM® protocol (via a standard IEEE 802.3 network connection), processes the image, and delivers the resulting information through the same DICOM network interface. Image inputs are limited to adult, digital frontal chest radiographs. The output results are sent to facilitate prioritization of chest Xrays for radiologist review on one or more devices that conform to the ClearRead Xray Pneumothorax DICOM Conformance Statement. ClearRead Xray Pneumothorax does not support printing or DICOM media.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study proving the device meets them, based on the provided text:


    Device: ClearRead Xray Pneumothorax (K213566)

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a "table of acceptance criteria" per se. However, it indicates key performance metrics that were measured and met through the clinical evaluation. The primary objective was to demonstrate the device meets or exceeds expected performance. The specific metrics, measured on an independent test set, and their reported values are:

    MetricAcceptance Criterion (Implicitly Met)Reported Device Performance
    AUC (ROC Curve)> 0.97 (or similar high value)0.974
    SensitivityHigh (e.g., > 0.90)0.922
    SpecificityHigh (e.g., > 0.90)0.951
    Time to NotificationLow (e.g., < 10-15 seconds)9.73 seconds

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size:

      • Clinical Evaluation: A total of 1138 cases were used.
        • 1028 cases from MIMIC-CXR (419 PTX positive, 609 PTX negative)
        • 110 cases from Georgetown University Medical Center (GUMC) (40 PTX positive, 70 PTX negative)
      • Internal Test Set (Benchmarking): 600 cases (300 negatives and 300 positives). This test set was separate from the internal "development set."
    • Data Provenance:

      • MIMIC-CXR: Collected from 2011 to 2016 at Beth Israel Deaconess Medical Center, USA. This is a controlled online dataset.
      • Georgetown University Medical Center (GUMC): Collected between 2002 to 2013, USA.
      • The clinical evaluation used retrospective adult patient images from multiple sources. The data was explicitly stated to be independent data, not used for training or internal validation.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Number of Experts: Three (3)
    • Qualifications: Senior board-certified radiologists with expertise in thoracic radiology.

    4. Adjudication Method for the Test Set

    The ground truth for the test set was established by a majority voting rule from the expert panel. Each expert independently reviewed the PA/AP image and associated radiology reports. The final image label (PTX positive/negative) and associated annotations (bounding boxes for PTX) were derived from this majority vote. If a true positive case, the associated annotation bounding boxes were replaced with a single box that enclosed all bounding boxes agreed upon by the majority. This resembles a "3-reader majority" consensus.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC comparative effectiveness study was described in the provided text. The study focused on the standalone (algorithm-only) performance of the device as a triage tool. The "Indications for Use" explicitly state it is a "notification-only triage workflow tool" and "operates in parallel to and independent of standard of care image interpretation workflow." It also states, "The device does not replace review and diagnosis of the X-rays by trained professionals." Therefore, it's not designed for human-in-the-loop performance improvement but rather for workflow prioritization.

    6. Standalone (Algorithm Only) Performance

    Yes, a standalone performance study was done. The reported AUC, Sensitivity, Specificity, and Time to Notification are all measures of the algorithm's performance without human intervention. The study describes "The ClearRead Xray Pneumothorax system was run on all selected images... Machine indications were transferred to the statistical analysis team and used as the basis for performance assessment."

    7. Type of Ground Truth Used

    The ground truth used for the clinical evaluation (test set) was expert consensus.

    • This was based on the visual inspection of image data by three expert radiologists.
    • It was further supported by available radiology reports.
    • For true positive cases, annotations (bounding boxes) were included.

    8. Sample Size for the Training Set

    The document does not specify an exact numerical sample size for the training set. It mentions:

    • "All models used by the ClearRead Xray Pneumothorax system were trained with cases that were clinically validated as negative for pneumothorax, and cases labeled as positive via a simulated data construction process."
    • "This included the publicly available data from the NIH, as hosted by Kaggle, wherein experts outlined proven pneumothoraces for thousands of images."
    • "Importantly, this also included over 7000 confirmed negatives cases."

    Therefore, the training set included:

    • Thousands of images with expert-outlined pneumothoraces from NIH/Kaggle.
    • Over 7000 confirmed negative cases.
    • Simulated data: Synthetically generated positive cases by inserting pneumothoraces into negative cases.

    9. How Ground Truth for the Training Set Was Established

    The ground truth for the training set was established using a hybrid approach:

    • Real Data:
      • For real positive cases (pneumothoraces), ground truth was established by clinical experts who "hand-drew outlines" of proven pneumothoraces (e.g., from the NIH/Kaggle dataset).
      • For real negative cases (no pneumothoraces), ground truth was established by being "clinically validated as negative." (This likely implies expert review or clinical reports confirming absence of PTX).
    • Simulated Data:
      • For simulated positive cases, the ground truth was arrived at via construction. This means the synthetic pneumothoraces were digitally inserted into negative cases, and their location and size were precisely known by design.
    Ask a Question

    Ask a specific question about this device

    K Number
    K161201

    Validate with FDA (Live)

    Device Name
    ClearRead CT
    Date Cleared
    2016-09-09

    (134 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ClearRead CT™ is comprised of computer assisted reading tools designed to aid the radiologist in the detection of pulmonary nodules during review of CT examinations of the chest on an asymptomatic population. The ClearRead CT requires both lungs be in the field of view. ClearRead CT provides adjunctive information and is not intended to be used without the original CT series.

    Device Description

    ClearRead CT is a dedicated post-processing application that generates a secondary vessel suppressed Lung CT series with CADe marks and associated region descriptors intended to aid the radiologist in the detection of pulmonary nodules.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study details for the ClearRead CT device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Localization Receiver Operating Characteristic (LROC) AUCClearRead CT found to significantly increase the AUC compared to the unaided read, indicating superior performance for detecting nodules.
    Radiologists' Interpretation TimeClearRead CT found to decrease read times with and without outliers.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state the exact sample size for the test set (number of cases). It refers to a "multi-reader multi-case (MRMC) study."

    The data provenance (country of origin, retrospective/prospective) is not specified in the provided text.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not specify the number of experts used to establish ground truth or their qualifications. It mentions a "multi-reader multi-case (MRMC) study," implying multiple readers were involved in the evaluation, but not necessarily in establishing the initial ground truth.

    4. Adjudication Method for the Test Set

    The document does not explicitly describe the adjudication method used for the test set.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    Yes, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done.

    • Effect Size of Human Readers Improvement with AI vs. Without AI Assistance: The study found that using ClearRead CT "significantly increase[d] the AUC" of the LROC response. It also found that ClearRead CT "decrease[d] read times with and without outliers." While a specific numerical effect size (e.g., a percentage increase in AUC or specific time reduction) is not provided in this summary, the terms "significantly increase" and "decrease" indicate a positive and measurable improvement.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    The document focuses on the multi-reader multi-case study, explicitly stating that ClearRead CT is "designed to aid the radiologist" and "provides adjunctive information and is not intended to be used without the original CT series." This implies the primary evaluation was human-in-the-loop. It also mentions that the device "generates a secondary vessel suppressed Lung CT series with CADe marks," which could be considered a standalone function, but the performance metrics provided are for the combined human-AI workflow.

    7. The Type of Ground Truth Used

    The document does not explicitly state the type of ground truth used (e.g., expert consensus, pathology, outcomes data).

    8. The Sample Size for the Training Set

    The sample size for the training set is not mentioned in the provided text.

    9. How the Ground Truth for the Training Set Was Established

    The document does not provide information on how the ground truth for the training set was established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K123526

    Validate with FDA (Live)

    Date Cleared
    2012-12-27

    (42 days)

    Product Code
    Regulation Number
    892.2050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    ClearRead +Confirm is intended to generate an enhanced, secondary digital radiographic image of the chest to facilitate confirmation of line/tubes. The enhanced AP or PA image of the chest provides improved visibility of lines and tubes. The ClearRead +Confirm image provides adjunctive information and is not a substitute for the original PA/AP image. This device is intended to be used by trained professionals, such as physicians, radiologists, and technicians, on patients with lines and tubes and is not intended to be used on pediatric patients.

    Device Description

    ClearRead +Confirm is a dedicated post-processing application that generates an enhanced, secondary digital radiographic image of the chest to facilitate confirmation of line/tubes.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information based on the provided text, using the specified format:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Primary Endpoints: Reduction in time required by radiologists to localize the tips of tubes, lines, and electrical cardiac wires (TLW) when using ClearRead +Confirm.ClearRead +Confirm was found to significantly decrease read times in analyses both including and excluding outliers.
    Secondary Endpoints: Radiologists' accuracy in localizing the TLW when using ClearRead +Confirm.The difference from true locations in ClearRead +Confirm reads and unaided reads were not found to be statistically significant.
    Overall Conclusion: Is the device a useful adjunct?The results indicated that ClearRead +Confirm is a useful adjunct for the interpretation of chest radiographs when looking for TLW.

    Study Details

    1. Sample size used for the test set and the data provenance: Not explicitly stated in the provided text. The text mentions it was a "multi-reader multicase (MRMC) study," implying a collection of cases, but the specific number is not given. However, the study was "clinical validation," implying prospective or retrospective clinical data. No country of origin is specified.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not explicitly stated. The study involved "radiologists" as readers, but how many established ground truth, or their specific qualifications (e.g., years of experience), is not detailed.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not explicitly stated. The text mentions "true locations," implying that a ground truth was established, but the method of adjudication (e.g., consensus, by an expert panel) is not described.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • Yes, a multi-reader multicase (MRMC) study was done.
      • Effect Size: The study found a "significant decrease" in read times when using ClearRead +Confirm. The specific quantitative "effect size" (e.g., percentage reduction, time saved in seconds) is not provided, only that it was statistically significant. For accuracy, no statistically significant difference from true locations was found between ClearRead +Confirm reads and unaided reads.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The document does not explicitly state whether a standalone performance study was conducted. The focus is on the human-in-the-loop performance through the MRMC study, where the device provides "adjunctive information."

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The text refers to "true locations" of tubes, lines, and electrical cardiac wires (TLW). This broadly suggests an expert-determined ground truth, likely based on anatomical knowledge and possibly confirmed by other imaging or clinical data, but the exact method (e.g., expert consensus) is not specified.

    7. The sample size for the training set: Not provided in the document.

    8. How the ground truth for the training set was established: Not provided in the document.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1