    Intended Use

    RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments.

    The system functionality can be configured based on user needs.

    Device Description

    RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments.

    The system functionality can be configured based on user needs.

    RayStation consists of multiple applications:

    • The main RayStation application is used for treatment planning.
    • The RayPhysics application is used for commissioning of treatment machines to make them available for treatment planning and used for commissioning of imaging systems.
    • The RayTreat application is used for sending plans to treatment delivery devices for treatment and receiving records of performed treatments.

    The device to be marketed, RayStation 2024A SP3, adds the RayTreat application compared with last cleared version, the predicate RayStation 2024A SP3 (without RayTreat), K240398.

    The RayTreat application was previously cleared with RayStation 11B, K220141. Since then some RayTreat functions have been changed:

    • RayTreat is now session focused
    • Usability improvements
    • Bug fixes

    The RayStation applications are built on a software platform, containing the radiotherapy domain model and providing GUI, optimization, dose calculation and storage services. The platform uses three Microsoft SQL databases for persistent storage of the patient, machine and clinic settings data.

    As a treatment planning system, RayStation aims to be an extensive software toolbox for generating and evaluating various types of radiotherapy treatment plans. RayStation supports a wide variety of radiotherapy treatment techniques and features an extensive range of tools for manual or semi-automatic treatment planning.

    The RayStation applications are divided into modules, which are activated through licensing.

    The RayTreat application
    RayTreat manages treatment delivery. An approved plan can be assigned to fractions in a treatment course and sent to the treatment delivery device. Treatment records from the treatment delivery device are recorded and sent to RayCarePACS.

    Note that all real-time monitoring of actual delivery is handled by treatment delivery device software, not by RayStation.

    Scientific concepts that form the basis for the device and significant performance characteristics:

    RayStation is a stand-alone software medical device intended for radiation therapy. Input to the device is patient, disease and treatment unit information, output from the device is one or more treatment plans. The treatment plans include treatment unit parameter settings for optimal beam arrangements, energies, field sizes, and ultimately fluence patterns to produce as safe and effective radiation dose distribution as the predicate.

    The scientific concepts of a treatment planning system are patient and beam modeling, and algorithms for dose calculation and plan parameter optimization.

    The patient model is a computerized representation of the patient tissue and densities, identifying the target regions and particular organs at risk. The model is based on medical images of the patient and must have the desired level of accuracy. Likewise, the beam modeling is a computerized representation of the treatment unit, defined by fluence type, energy distribution, machine specific geometry, and beam modifiers such as MLC, flattening filters, wedges etc. The algorithms for dose calculation and plan parameter optimization must take into account all geometries and materials that affect irradiation transport through the treatment unit and the patient. The optimization algorithm iterates treatment plan parameters until the desired treatment plan and dose distribution have been obtained. Also here, all steps must be done to the desired level of accuracy.

    Significant physical characteristics of the device, material used, and physical properties:

    The device is a standalone software medical device. It has no physical properties or materials. The device design information can be found in the subsection above "Device design information".

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K242992

    Validate with FDA (Live)

    Date Cleared
    2025-04-17

    (203 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RayCare is an oncology information system intended to provide information which is used to take decisions for diagnosis, treatment management, treatment planning, scheduling, treatment and follow-up of radiation therapy, medical oncology and surgical oncology.

    For these disciplines, as applicable, RayCare enables the user to define the clinical treatment intent, prescribe treatment, specify the detailed course of treatment delivery, manage the treatment course and monitor the treatment course.

    In the context of radiation therapy, the RayCare image viewer can be used for viewing images, annotating images, performing and saving image registrations as well as image fusion to enable offline image review of patient positioning during treatment delivery.

    RayCare is not intended for use in diagnostic activities.

    Device Description

    RayCare is an oncology information system intended to provide information which is used to take decisions for diagnosis, treatment management, treatment planning, scheduling, treatment and follow-up of radiation therapy, medical oncology and surgical oncology.

    For these disciplines, as applicable, RayCare enables the user to define the clinical treatment intent, prescribe treatment, specify the detailed course of treatment delivery, manage the treatment course and monitor the treatment course.

    In the context of radiation therapy, the RayCare image viewer can be used for viewing images, annotating images, performing and saving image registrations as well as image fusion to enable offline image review of patient positioning during treatment delivery.

    As an oncology information system, RayCare supports healthcare professionals in managing cancer care treatments. The system provides functionalities as described briefly in the sections below. These functionalities are not provided separately in different applications and have a joint purpose for the treatment of the patient.

    RayCare is a software-as a Medical Device with a client part that allows the user to interact with the system and a server part that performs the necessary processing and storage functions. Selected aspects of RayCare are configurable, such as adapting workflow templates to the specific needs of the clinic.

    AI/ML Overview

    This document describes the premarket notification for RayCare (2024A SP1), an oncology information system. The relevant sections for acceptance criteria and study details are primarily found under "VII. Non-Clinical and/or Clinical Tests Summary" and the tables within it.

    Based on the provided text, RayCare (2024A SP1) is not an AI/ML device in the sense of making autonomous diagnostic decisions or image-based classifications. It is an Oncology Information System that supports clinical workflows for radiation therapy and other oncology disciplines. The "acceptance criteria" and "study that proves the device meets the acceptance criteria" in this context refer to the software verification and validation (V&V) activities. Therefore, the information provided focuses on demonstrating the software's functional correctness, safety, and effectiveness compared to a predicate device, rather than performance metrics specifically for an AI model (e.g., sensitivity, specificity, AUC).

    Here's a breakdown of the requested information based on the provided document:


    Acceptance Criteria and Device Performance (Software V&V)

    The acceptance criteria for RayCare (2024A SP1) are implicitly defined by the successful completion of various software verification and validation activities designed to demonstrate that the device performs as intended and is as safe and effective as its predicate. These are primarily functional and system-level criteria.

    Table of Acceptance Criteria and Reported Device Performance:

    Since this is a software verification and validation summary for an oncology information system, the "performance" is demonstrated through successful compliance with system specifications and validated functionality. The "acceptance criteria" are the "Pass criteria" of the specific tests.

    Acceptance Criteria (from "Criteria" or "Pass criteria" of listed V&V)Reported Device Performance
    Treatment Course Management (TCM) Workspace: The TCM workspace shall show the treatment course and its related series, treatment fractions, and assigned beam sets for the care plan selected in the global care plan selector. Specific criteria: • The treatment series related to the selected care plan is displayed. • The fractions in the fractions table are only related to the treatment series related to the selected care plan. • The assigned beam set table only displays the beam set related to the selected care plan.Passed. "The successful validation of this feature demonstrates that the device is as safe and effective as the predicate device."
    Extended RayCare Scripting Support (Unit Testing): Queries shall only be available for scripting if explicitly declared as scriptable (whitelisted data).Passed. "The successful validation of this feature demonstrates that the device is as safe and effective as the predicate device."
    Extended RayCare Scripting Support (System Level Verification): It is possible to run a script by clicking a RayCare script task, and the script has performed the expected action within RayCare.Passed. "The successful validation of this feature demonstrates that the device is as safe and effective as the predicate device."
    Offline and Online Recording of Treatment Results: Offline import is requested, received, and possible to sign with device and radiotherapy record selected for import for a selected session. Specific criteria: • Verify treatment course table and beam delivery result table in TC overview gets updated with corresponding data for the first session. • Verify the device selected for offline import is the delivered device on the session.Passed. "The successful validation of this feature demonstrates that the device is as safe and effective as the predicate device."
    Treatment Delivery Integration Framework (Varian TrueBeam): The treatment flow for treatment delivery is verified. Specific criteria: • The fraction is fully delivered, and the status of the fraction, session, and beams is set to "Delivered". • Compare the delivered meterset, the couch positions and angles. They should be the same. • The online couch corrections are calculated as the difference between the planned and the delivered couch positions.Passed. "The successful validation of this feature demonstrates that the device is as safe and effective as the predicate device."

    Overall Conclusion:
    "From the successful verification and validation activities, the conclusion can be drawn that RayCare 2024A SP1 has met specifications and is as safe, as effective and performs as well as or better than the legally marketed predicate device."


    1. Sample sizes used for the test set and the data provenance:

      • Test Set Sample Size: The document does not specify a numerical sample size for "test sets" in the traditional sense of patient cases or images for evaluating an AI model. Instead, it refers to software verification and validation ("V&V") activities including unit testing, integration testing, system-level testing, cybersecurity testing, usability testing, and regression testing. These involve testing against requirements and specifications, often using simulated data, test cases, or specific user scenarios, rather than a fixed "dataset" of patient images.
      • Data Provenance: The document does not explicitly state the country of origin of testing data or if it was retrospective or prospective. Given it's software V&V for an oncology information system, the "data" would primarily be test inputs and expected outputs generated internally during the development process (e.g., test scripts, simulated patient data to exercise specific functionalities). It's not a study on real patient data for diagnostic performance.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This concept is not applicable as this is a software verification and validation summary for an oncology information system, not a study evaluating an AI model's diagnostic or prognostic performance against expert-determined ground truth. The "ground truth" for V&V activities is the system's specified behavior and functional requirements. Software engineers and QA professionals establish whether the software meets these pre-defined requirements.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable. Adjudication methods are typically used in clinical studies involving human readers to resolve discrepancies in annotations or diagnoses, especially when establishing ground truth for AI model evaluation. This document describes software V&V.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. The document explicitly states: "No Clinical trials were required to demonstrate substantial equivalence."
      • This type of study is relevant for AI-assisted diagnostic devices. RayCare is described as an oncology information system, not an AI diagnostic tool.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • While the system has automated functions, the concept of "standalone performance" as it relates to an AI algorithm making a clinical decision (e.g., classifying a lesion) is not directly applicable here. The V&V described focuses on the system's ability to correctly manage and process information, integrate with other systems, and support user workflows, which are inherent to its "standalone" operation as an information system. The "performance" is demonstrated through successful execution of its intended software functions as per its specifications.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • The "ground truth" for the software verification and validation described here is the functional specifications and requirements of the RayCare system. Successful "verification" means the design output meets the requirements, and "validation" means the software conforms to user needs and intended uses. This is established through internal testing against defined expected behaviors.
    7. The sample size for the training set:

      • Not applicable. RayCare (2024A SP1) is an oncology information system, and the document does not indicate that it incorporates a machine learning model that was "trained" on a dataset in the way an AI diagnostic or predictive algorithm would be. The device's "development" involved standard software engineering practices.
    8. How the ground truth for the training set was established:

      • Not applicable. As there is no mention of an AI/ML training set, the concept of establishing ground truth for it does not apply.

    In summary, this FDA review document pertains to the clearance of an Oncology Information System (OIS) through the 510(k) pathway, demonstrating substantial equivalence to a predicate device. The "acceptance criteria" and "proof" come from a robust set of software verification and validation activities (unit, integration, system, cybersecurity, usability, regression testing) rather than clinical studies or the evaluation of an AI model's diagnostic performance against a clinical ground truth. The device is not presented as an AI-driven diagnostic tool.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240398

    Validate with FDA (Live)

    Date Cleared
    2025-04-04

    (420 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments.

    The system functionality can be configured based on user needs.

    Device Description

    RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments.

    The system functionality can be configured based on user needs.

    RayStation consists of multiple applications:

    • The main RayStation application is used for treatment planning.
    • The RayPhysics application is used for commissioning of treatment machines to make them available for treatment planning and used for commissioning of imaging systems.

    The devices to be marketed, RayStation/RayPlan 2024A SP3, 2024A and 2023B, contain modified features compared to last cleared version RayStation 12A including:

    • Improved sliding window VMAT (Volumetric Modulated Arc Therapy) sequencing
    • Higher dose grid resolution for proton PBS (Pencil Beam Scanning)
    • Automated field in field planning
    • LET optimization (Linear Energy Transfer)

    These applications are built on a software platform, containing the radiotherapy domain model and providing GUI, optimization, dose calculation and storage services. The platform uses three Microsoft SQL databases for persistent storage of the patient, machine and clinic settings data.

    As a treatment planning system, RayStation aims to be an extensive software toolbox for generating and evaluating various types of radiotherapy treatment plans. RayStation supports a wide variety of radiotherapy treatment techniques and features an extensive range of tools for manual or semi-automatic treatment planning.

    The RayStation application is divided in modules, which are activated through licensing. A simplified license configuration of RayStation is marketed as RayPlan.

    AI/ML Overview

    The provided document is a 510(k) clearance letter for the RayStation/RayPlan 2024A SP3, 2024A, and 2023B devices. It discusses the substantial equivalence of these devices to a predicate device (RayStation 12A).

    However, the document does not contain specific acceptance criteria tables nor detailed study results for a single, comprehensive study proving the device meets acceptance criteria in the format typically requested (e.g., a specific clinical validation study with explicitly defined acceptance metrics like sensitivity, specificity, or AUC, and corresponding reported performance values).

    Instead, the document describes a broad software verification and validation process, stating that the software underwent:

    • Unit Testing
    • Integration Testing
    • System Level Testing
    • Cybersecurity Testing
    • Usability Testing (Validation in a clinical environment)
    • Regression Testing

    For several "Added/updated functions," the document provides a description of the verification and validation data used to demonstrate substantial equivalence and simply states "Yes" under the "Substantially Equivalent?" column if the validation was "successful." The acceptance criteria for these tests are described narratively within the text, not in a consolidated table format with numerical performance outcomes.

    Therefore, I cannot generate the requested table of "acceptance criteria and the reported device performance" as a single, consolidated table with numerical results for the entire device's performance against specific, pre-defined acceptance criteria for a single study. The document describes a process of demonstrating substantial equivalence through various verification and validation activities rather than a single, large-scale study with quantitative acceptance criteria for the overall device performance.

    However, I can extract the information related to the validation activities for specific features and the general approach to proving substantial equivalence.

    Here's a breakdown of the requested information based on the provided document, addressing each point to the best of my ability given the available details:


    Acceptance Criteria and Device Performance (Based on provided verification and validation descriptions)

    As noted, a single, consolidated table of quantitative acceptance criteria and overall device performance is not provided. Instead, the document describes various verification and validation activities with implicit or explicit pass criteria for individual features or system aspects to demonstrate substantial equivalence to the predicate device.

    Below are examples of how some "acceptance criteria" (pass criteria) and "reported performance" are described for specific features. These are not aggregated performance metrics for the entire device but rather success criteria for sub-components or changes.

    Feature/Aspect TestedAcceptance Criteria (Pass Criteria) DescribedReported Device Performance (as stated in the document)
    Dose compensation point computation for Tomo Synchrony1. Calculated values for the center point coordinates are equal to values from the version used in Accuray validation. 2. Calculated values are numerically equal to values obtained from calling the method (regression test). 3. Calculated values are exported correctly from RayStation to DICOM (equality between calculated and exported point, only for Helical Tomo Synchrony plans, only in correct DICOM item). 4. Calculated values are converted correctly from DICOM to Accuray's system format (equality of point coordinates, only for relevant plan types)."The successful validation of this feature demonstrates that the device is as safe and effective as the predicate device." (Implies all pass criteria were met).
    Point-dose optimization in brachy plans1. Position from the correct image set is used for point-dose objectives/constraints. 2. Possible to add optimization objective/constraint to a point, referring to the correct point. 3. When adding objective/constraint, selection of function type and dose level is possible and reflected in description. 4. Saving and loading an optimization function template containing point objectives/constraints works correctly (loaded functions are same as saved). 5. Results from single/multiple point optimization are as expected (dose in point(s) should be equal to specified dose in objective(s))."The successful validation of this feature demonstrates that the device is as safe and effective as the predicate device." (Implies all pass criteria were met).
    Electron Monte Carlo dose engine improvementsComparing calculated doses with: - Measured doses obtained from clinics, - Doses computed in independent, well-established TPS, - Doses computed with earlier versions of RayStation, - Doses computed in BEAMnrc/egs++ using Gamma evaluation criteria."The successful validation of this feature demonstrates that the device is as safe and effective as the predicate device." (Implies adequate agreement based on Gamma criteria).
    Evaluation on converted CBCT images for protonsFor proton MC/PB dose computation: - Gamma 2%/2mm pass rate above 90% - Gamma 3%/3mm pass rate above 95%"The successful validation of this feature demonstrates that the device is as safe and effective as the predicate device." (Implies specified Gamma pass rates were achieved).
    Overall Device (Software Verification/Validation)Software specifications conform to user needs and intended uses, and particular requirements implemented through software can be consistently fulfilled. Conformance to applicable requirements and specifications. Successful outcome of unit, integration, system, cybersecurity, usability, and regression testing. Safety and effectiveness validated."RayStation/RayPlan 2024A SP3, 2024A and 2023B have met specifications and are as safe, as effective and perform as well as the legally marketed predicate devices." All general software tests (unit, integration, system, cybersecurity, usability, regression) were acceptable/successful.

    Study Details (Based on the document)

    Given the nature of the 510(k) submission for a treatment planning system, the "study" is primarily a comprehensive software verification and validation effort to demonstrate substantial equivalence, rather than a single, standalone clinical trial or diagnostic accuracy study.

    1. Sample sizes used for the test set and the data provenance:

      • Test Set Sample Sizes: Not explicitly stated as a single numerical value for a global "test set." Testing was conducted at multiple levels (unit, integration, system, usability, regression) across various features.
        • For "Evaluation on converted CBCT images for protons," it mentions "Test cases consist of CBCTs from the MedPhoton imaging ring on a Mevion S250i system, as well as the on-board CBCT systems on a Varian ProBeam and an IBA P1," implying a set of patient or phantom imaging data, but the exact number of cases/patients is not specified.
        • For other features, it refers to "tests," "validation data," or "computed doses" but doesn't quantify the number of distinct data points/cases used.
      • Data Provenance:
        • Country of Origin: Not specified in the document. Likely internal RaySearch data and potentially data from collaboration with clinical sites, but no specific countries are mentioned.
        • Retrospective or Prospective: Not explicitly stated. The verification and validation activities appear to be primarily retrospective (using existing data, phantom measurements, or simulated scenarios) as part of the software development lifecycle, rather than prospective clinical data collection for a specific study.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in the document. The document refers to "measured doses obtained from clinics" and "doses computed in independent, well-established TPS" as part of the validation for dose engine improvements, suggesting some form of external or expert-derived ground truth, but the number and qualifications of experts involved are not detailed. For "Evaluation on converted CBCT images for protons," it states "For each case, a ground truth CT image has been prepared to serve as ground truth," implying expert or established reference standard, but again, no details on experts.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • This information is not provided. The document focuses on computational and functional verification rather than multi-reader clinical assessment.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not explicitly done or reported in this document. The device, RayStation/RayPlan, is a treatment planning system that assists users in creating treatment plans, not primarily an AI-driven image interpretation or diagnostic aid where human reader performance improvement is typically measured. The AI-related feature mentioned is "deep learning segmentation," but the document states, "(The model training is performed offline on clinical CT and structure data.)" It does not detail an MRMC study related to its performance or impact on human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, standalone (algorithm-only) performance was central to the validation. The document describes extensive "Unit Testing," "Integration Testing," "System Level Testing," and "Dose engine validation" which are all a form of standalone algorithmic evaluation. For example, the Gamma evaluation criteria for dose calculations or the numerical equality checks for dose compensation points are purely algorithmic performance assessments.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The ground truth varied depending on the feature being validated:
        • "Measured doses" from clinics / Independent TPS computations / BEAMnrc/egs++ calculations: For dose engine validation. This represents a highly accurate, often physical measurement or well-established computational standard.
        • "Ground truth CT image": For evaluation of converted CBCT images for protons. This implies a high-quality reference image.
        • Internal "expected results" and "specifications": For functional and system-level tests (e.g., for point-dose optimization, the expected result was that the dose in the point should equal the dose specified in the objective).
        • "Clinical objectives": Used for plan comparisons (e.g., in segment weight optimization validation), likely representing desired dose distributions defined by clinical experts.
    7. The sample size for the training set:

      • The document mentions "deep learning segmentation" and states that "The model training is performed offline on clinical CT and structure data." However, the sample size for this training set is not provided.
    8. How the ground truth for the training set was established:

      • For "deep learning segmentation," the ground truth for training would implicitly be the "clinical CT and structure data" mentioned. This typically means expert-delineated structures (ROIs) on clinical CT images, but the exact method (e.g., single expert, consensus, specific software tools) is not detailed.
    Ask a Question

    Ask a specific question about this device

    K Number
    K222312

    Validate with FDA (Live)

    Device Name
    RayStation 12A
    Date Cleared
    2023-03-29

    (240 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments.

    The system functionality can be configured based on user needs.

    Device Description

    RayStation is a treatment planning system for planning, analysis and administration of radiation therapy and medical oncology treatment plans. The device lets the user import patient images and data, identify treatment targets and organs at risk, create an optimal treatment plan taking into account patient anatomy, prescribe treatment dose and organ at risk sensitivity, review and approve the plan and then administer the treatment. A scientific basis for the device is the implementation of peer reviewed algorithms of plan parameter optimization and photon and particle dose calculation.

    RayStation consists of multiple applications:

    • The main RayStation application is used for treatment planning.
    • The RayPhysics application is used for commissioning of treatment machines to make them available for treatment planning and used for commissioning of imaging systems.
    • The RayTreat application is used for sending plans to treatment delivery devices for treatment and receiving records of performed treatments.

    These applications are built on a software platform, containing the radiotherapy domain model and providing GUI, optimization, dose calculation and storage services. The platform uses three Microsoft SQL databases for persistent storage of the patient, machine and clinic settings data.

    The RayStation application is divided in modules, which are activated through licensing. A simplified license configuration of RayStation is marketed as RayPlan has a limited set of modules, indicated in the following table.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for RayStation 12A, based on the provided text:

    Acceptance Criteria and Device Performance

    Acceptance CriteriaReported Device Performance
    Support for eye planning with wedgesThe SOBP distal fall-off of the central axis depth dose curve meets accuracy requirements. 95% and 98% of the computed depth dose values meet Gamma pass rates.
    Automatic field in field planningFor a 3D-CRT plan, the merged beams' MU agrees with original beams' MU. Merged beams' segments maintain original shapes. MU and segment weights after split are subdivided correctly, and split beams are managed correctly in terms of ordering and ROI handling.
    Brachy Therapy support for Elekta Flexitron® afterloadersThe dose engine accurately reproduces dose for a variety of sources when compared to OA along-away data. Measured doses (EQUAL-ESTRO) relate computed dose to delivered dose correctly. Comparison to an independent and TG43 compliant treatment planning system validates correct superposition of dose. Comparison to an independent Monte Carlo system provides fully independent validation of complete treatment plans.
    Electron Monte Carlo dose engine updateFor comparison with previous RayStation dose, the calculated doses fail for less than 2% of the data points for gamma 2%/2mm. The fraction of calculated dose data points failing comparison to BEAMnrc/egs++ has been evaluated. Two different gamma criteria for comparison with another TPS or measurement are evaluated with specified requirements on agreement level.
    Overall Software Safety and Performance (Major Concern Level)Software verification and validation demonstrate that RayStation 12A performs as intended in specified use conditions and comparably to the predicate device.

    Study Information:

    2. Sample sizes used for the test set and the data provenance

    • Eye Planning with Wedges: Specific sample size for test cases is not provided, but they cover "line doses in homogeneous phantoms using a square aperture and a wedge mounted with varying opening angles and positions." The origin is internal testing/validation.
    • Automatic Field in Field Planning: Specific sample size for test cases is not provided. The origin is internal testing/validation.
    • Brachy Therapy Support for Elekta Flexitron® afterloaders: Specific sample size for test cases is not provided. Reference doses consist of "point doses, line doses, as well as 2D and 3D doses." Origin includes published consensus data, measured doses, doses computed in two major competing TPS, and doses computed with an independent Monte Carlo software.
    • Electron Monte Carlo Dose Engine Update: Specific sample size for test cases is not provided. Reference doses include measured doses, doses computed in a well-established competing TPS, doses computed with earlier versions of RayStation, and doses computed in BEAMnrc/egs++. Data provenance is thus mixed, including internal comparisons, external commercial TPS, and external academic/research software.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The document does not specify the number or qualifications of experts used to establish ground truth for any of the test sets. The ground truth for dose calculation is primarily based on:

    • Published consensus data
    • Measured doses
    • Comparison to other established commercial Treatment Planning Systems (TPS)
    • Comparison to independent Monte Carlo software

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    The document does not describe any adjudication method involving human experts for resolving discrepancies in the test sets. The comparisons are quantitative against established reference data or other computational models.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not mentioned or performed. This device is a treatment planning system, not an AI-assisted diagnostic tool that would typically involve human "readers" in that context. The improvements are in the accuracy and functionality of the dose calculation and planning algorithms, not in improving human interpretation of an AI output.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the studies described are standalone performance evaluations of the algorithm's accuracy in specific calculation tasks (dose computation for various modalities and planning features). The "device performance" described in the table above refers to the algorithm's performance against reference standards.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used is primarily based on:

    • Physical Measurements: "Measured doses," "depth-dose profiles along the central axis were acquired with a plane-parallel chamber in a water tank."
    • Published Consensus Data: For brachytherapy, "Published consensus data."
    • Established Computational Models: Doses computed in well-established competing TPS, independent Monte Carlo software (BEAMnrc/egs++), and comparison to previous versions of RayStation.

    8. The sample size for the training set

    The document does not provide any information regarding the sample size for a training set. The descriptions focus on the validation of specific algorithms and features, which often rely on physics models and deterministic calculations rather than machine learning training sets. While "deep learning segmentation" is mentioned as a module (available in RayStation 11B and 12A), the provided validation details for the new features in 12A do not describe training data for this specific version's changes.

    9. How the ground truth for the training set was established

    As no training set information is provided for the features specifically validated in K222312, there is no information on how its ground truth would have been established. For the general mention of "deep learning segmentation," the text indicates "The model training is performed offline on clinical CT and structure data," but no further details on ground truth establishment for this training.

    Ask a Question

    Ask a specific question about this device

    K Number
    K220141

    Validate with FDA (Live)

    Device Name
    RayStation 11B
    Date Cleared
    2022-04-15

    (87 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments.

    The system functionality can be configured based on user needs.

    Device Description

    RayStation is a treatment planning system for planning, analysis and administration of radiation therapy and medical oncology treatment plans. It has a modern user interface and is equipped with fast and accurate dose and optimization engines.

    RayStation consists of multiple applications:

    • . The main RayStation application is used for treatment planning.
    • . The RayPhysics application is used for commissioning of treatment machines to make them available for treatment planning and used for commissioning of imaging systems.
    • The RayTreat application is used for sending plans to treatment delivery devices for treatment and receiving records of performed treatments.

    The device to be marketed, RayStation 11B, contains modified features compared to RayStation 11.0 as indicated below:

    A simplified license configuration of RayStation is marketed as RayPlan. RayPlan has a limited set of purchasable licenses and some modules will not be accessible. RayPlan is marketed as RayPlan 11B.

    EQD2 dose computation (new) - From photon and/or braction doses, it is possible to compute, deform and accumulate the two Gray equivalent (EQD2) dose. The computation of the EQD2 dose uses the biological linear quadratic model, which is also the basis for the already released biological optimization and evaluation functionality.

    Generation of synthetic CT from CBCT (new) - Two new methods (algorithms) for synthetic CT generation will be included. The synthetic CT images are created by combining information in the CBCT image and a CT image for the specific patient to allow for dose computation using the HU values in the image, as for regular CT images. In RayStation 11.0 it is possible to compute dose on CBCT images for photons using bulk density assignments. The added functionality will improve the photon dose calculation accuracy on CBCT images. Handling of LET and other RBE components (new) - This functionality enables possibility to compute and evaluate the dose weighted LET (Linear Energy Transfer) for proton and light ion plans. LET is an additional dosimetric measure that can be used to assess the radiobiological effect of the proton and light ion radiation. Radiobiological equivalent (RBE) dose is a derived quantity with dependence on both the physical dose and the LET. In RayStation 11.0, it is possible to compute and evaluate RBE doses.

    AI/ML Overview

    The provided text describes a 510(k) submission for RayStation 11B, a radiation therapy treatment planning system. However, the document does not contain the detailed information required to answer many of the questions about acceptance criteria and the study that proves the device meets them, especially regarding the performance of a machine learning component if one exists and its specific acceptance criteria.

    The document states: "Related to machine learning, there is no change compared to the predicate device." This implies that while the RayStation product family might have machine learning components, RayStation 11B does not introduce new machine learning features that would necessitate specific performance studies for acceptance as part of this 510(k) submission. The focus of the provided text is on the validation of new dose computation methods and synthetic CT generation.

    Therefore, many sections below will indicate "Information Not Provided in Text" or state that the question is not applicable given the document's content.

    Acceptance Criteria and Reported Device Performance

    Given the information provided, the acceptance criteria are largely related to the successful verification and validation of the new features (EQD2 dose computation, synthetic CT generation from CBCT, and handling of LET and other RBE components) demonstrating that the dose computations are "adequate for clinical use" and that the system "has met specifications and is as safe, as effective and performs as well as or better than the legally marketed predicate device."

    Without specific numerical thresholds or performance metrics in the provided text, a formal table of acceptance criteria and reported performance cannot be fully constructed for these new features. The document suggests that the previous predicate device's performance benchmarks likely served as the implicit standard for "as well as or better than."

    Implicit Acceptance Criteria and Reported Performance (based on text):

    Acceptance CriterionReported Device Performance
    EQD2 dose computation is adequate for clinical use.Validation for photon and/or braction doses were validated as part of the Clinical Evaluation for Brachy and User Site Validation demonstrates that the dose computation adequate for clinical use.
    Improved photon dose calculation accuracy on CBCT images for synthetic CT from CBCT is adequate for clinical use.Validation of improved photon dose calculation accuracy on CBCT images was performed in CBCT Conversion validation demonstrates that the dose computation adequate for clinical use.
    Handling of LET and other RBE components for proton and light ion plans is adequate for clinical use.Validation of dose weighted LET (Linear Energy Transfer) for proton and light ion plans were performed as part of the Proton PBS Monte Carlo validation. The validation demonstrates that the dose computation adequate for clinical use.
    Overall system meets specifications and is as safe, as effective, and performs as well as or better than the predicate device."The data obtained from the verification show that system tests, unit and subsystem tests have passed, and the validations been completed successfully. The reviews of design, code and labeling are also passed."

    Study Information

    1. A table of acceptance criteria and the reported device performance
      See table above. Specific numerical thresholds are not provided in the text.

    2. Sample sizes used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • Sample Sizes: Not explicitly stated for any of the validations (e.g., Clinical Evaluation for Brachy, CBCT Conversion validation, Proton PBS Monte Carlo validation). The text mentions "User Site Validation," which implies real-world data, but details on the number of cases or patients are absent.
      • Data Provenance: Not specified (e.g., country of origin).
      • Retrospective or Prospective: Not specified. "User Site Validation" might imply prospective or real-world use data, but this is not confirmed.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • Number of Experts: Not specified.
      • Qualifications of Experts: Not specified. Mention of "User Site Validation in cooperation with cancer clinics" suggests involvement of clinical professionals, but their specific roles, number, or qualifications are not detailed.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Adjudication Method: Not specified.
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • An MRMC study is not mentioned for the new features. The document explicitly states: "Related to machine learning, there is no change compared to the predicate device." This indicates that no new AI-assisted workflows requiring human reader improvement studies were part of this submission. The validation efforts focus on the accuracy of the new computational functions themselves, not human-AI interaction.
    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

      • Yes, the validation activities described (e.g., "EQD2 dose computation (new) – Validation," "Generation of synthetic CT from CBCT (new) – Validation," "Handling of LET and other RBE components (new) - Validation") are, by their nature, evaluations of the algorithm's performance in generating accurate dose calculations or synthetic CTs. These are standalone evaluations of the new features. The "User Site Validation" is likely a real-world system test, but the core 'algorithm-only' performance is implied by the specific validation names.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

      • The term "ground truth" is not explicitly used, but the validation implies comparisons against established or expected physical and radiobiological models and potentially clinical standards of care.
        • For dose computation, validation would typically involve comparing computed doses against known physics models, phantom measurements, or existing clinical systems deemed accurate.
        • For synthetic CT, the "ground truth" would likely be the actual CT data that the synthetic CT is intended to replicate or approximate for dose calculation purposes.
      • The text doesn't specify if clinical outcomes were used as ground truth for these specific validations.
    8. The sample size for the training set

      • Information Not Provided in Text. The document states "Related to machine learning, there is no change compared to the predicate device," and the new features are described as "algorithms" or "functionality," not explicitly machine learning models that would require a distinct training set for this submission. If the original RayStation product had ML components, their training set information is not part of this document.
    9. How the ground truth for the training set was established

      • Information Not Provided in Text. (See point 8).

    In summary, the provided FDA 510(k) summary focuses on demonstrating the substantial equivalence of RayStation 11B to its predicate device by verifying and validating new computational features rather than new or modified AI/ML features requiring human-in-the-loop or specific training/test set performance metrics commonly associated with AI/ML device clearances. The acceptance criteria are implicitly tied to the successful completion of these verification and validation activities, proving the new functions are "adequate for clinical use" and the system performs "as well as or better than" the predicate.

    Ask a Question

    Ask a specific question about this device

    K Number
    K211867

    Validate with FDA (Live)

    Device Name
    RayStation 11.0
    Date Cleared
    2021-09-08

    (84 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments.

    Device Description

    RayStation is a treatment planning system for planning, analysis and administration of radiation therapy and medical oncology treatment plans. It has a modern user interface and is equipped with fast and accurate dose and optimization engines.

    RayStation consists of multiple applications:

    • . The main RayStation application is used for treatment planning.
    • . The RayPhysics application is used for commissioning of treatment machines to make them available for treatment planning and used for commissioning of imaging systems.
    • . The RayTreat application is used for sending plans to treatment delivery devices for treatment and receiving records of performed treatments.

    The device to be marketed, RayStation 11.0, marketing name "RayStation 11A", contains modified features compared to version 10.1. The main change is planning for the CyberKnife linear accelerator treatment unit.

    The device to be marketed supports planning and dose calculation for M6/S7 CyberKnife treatment machines. Older CyberKnife versions are not supported. The CyberKnife has one delivery technique, and different node sets. The node set defines the possible robot positions for different collimations (cones or MLC) and patient geometry (head or body). All node sets available for patient treatments are included in the validation.

    AI/ML Overview

    The document provided does not contain specific acceptance criteria for a device in the traditional sense of a medical diagnostic or AI-powered device. Instead, it describes a software system, RayStation 11.0, which is a radiation therapy treatment planning system. The document focuses on demonstrating substantial equivalence to a predicate device (RayStation 10.1) and validating its functionality for a new feature: planning for CyberKnife linear accelerator treatment units.

    Therefore, the "acceptance criteria" discussed are more about the software's performance meeting its specifications and being suitable for clinical use, rather than quantitative performance metrics for a diagnostic output.

    Based on the provided text, here's an attempt to answer your questions by interpreting "acceptance criteria" as the measures taken to validate the software's functionality and safety:

    Table of Acceptance Criteria and Reported Device Performance

    Given that RayStation is a treatment planning system and not a diagnostic device with specific output metrics like sensitivity/specificity, the "acceptance criteria" are related to the accuracy and safety of its dose calculations and overall system functionality for treatment planning. The document broadly states the "validation shows that the dose computation is suited for clinical use" and "adequate for clinical use."

    Acceptance Criteria CategoryReported Device Performance
    Dose Computation AccuracyCyberKnife specific photon dose engine validation (collapsed cone and photon Monte Carlo) performed. Validation shows "dose computation is suited for clinical use" and "adequate for clinical use."
    System Functionality"System Tests of RayStation" passed."Risk analysis-based tests for use error mitigation verification" passed.
    Software Quality"Unit and subsystem testing for low-level testing" passed."Reviews of design, code and Master Labeling" passed.
    User Validation"User validation in cooperation with cancer clinics" completed successfully.

    Study that proves the device meets the acceptance criteria:

    The provided document describes the verification and validation activities performed for RayStation 11.0. This is not a single "study" in the sense of a clinical trial with a defined sample size of patients with a specific condition. Instead, it's a comprehensive software development and quality assurance process.

    1. Sample sizes used for the test set and the data provenance:

    • Test Set Sample Size: The document does not specify a numerical "sample size" for a typical test dataset as one would expect in an AI/diagnostic device study. Instead, it refers to various testing activities:
      • CyberKnife Validation: "All node sets available for patient treatments are included in the validation." This implies comprehensive testing across different configurations of the CyberKnife treatment unit. It doesn't quantify the number of patient plans or variations tested.
      • System Tests, Unit/Subsystem Tests, Risk Analysis-based Tests: These are internal software testing activities, and their "sample size" would be defined by the number of test cases executed, which is not detailed.
      • User Validation: No specific number of cases or users is given.
    • Data Provenance: Not explicitly stated for specific test cases. Given it's a software for radiation therapy planning, the "data" would consist of simulated patient anatomies, treatment plans, and test configurations. The "user validation in cooperation with cancer clinics" implies real-world clinical context, but details are not provided. It's implicitly retrospective as it's part of a software release validation. No country of origin is specified for the data used in testing.

    2. Number of experts used to establish the ground truth for the test set and their qualifications:

    • CyberKnife Validation: The validation was performed internally by RaySearch Laboratories, likely by their physicists and engineers. The document does not specify the number of external experts or their qualifications for establishing a "ground truth" for the dose calculations. The "ground truth" for dose calculations would be established through established physics models and comparisons to known phantoms or reference data, not necessarily expert consensus in the typical sense of image interpretation.
    • User Validation: Conducted "in cooperation with cancer clinics." This indicates involvement of clinical experts (e.g., radiation oncologists, medical physicists) who would evaluate the usability and clinical appropriateness of the plans generated. The number and qualifications of these experts are not specified.
    • General: The document does not mention the use of experts to establish 'ground truth' in the context of diagnostic interpretation, which is common for AI-powered image analysis tools. For a treatment planning system, 'ground truth' relates to the accuracy of physics calculations and the clinical utility of the generated plans.

    3. Adjudication method for the test set:

    • No explicit adjudication method (e.g., "2+1" or "3+1") is described, as the validation is not focused on subjective interpretation or a diagnostic outcome requiring such consensus. For dose calculations, discrepancies would be resolved by identifying and correcting calculation errors. For user validation, feedback would be gathered and addressed.

    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and its effect size:

    • No MRMC study was performed or described. The study type is focused on validating the software's internal performance and its capability to generate treatment plans, not on how human readers (e.g., radiologists) improved their performance using the software. RayStation is a planning tool, not an AI-assisted diagnostic tool for human readers.

    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Yes, standalone performance was assessed for core functionalities. The "Dose engine validation" for the CyberKnife is a standalone assessment of the algorithm's accuracy in calculating dose. "System Tests," "Unit and subsystem testing," and "Risk analysis-based tests" also represent assessments of the algorithm's performance without direct human intervention during the test execution. The "user validation" component does involve humans in the loop to assess usability and clinical utility.

    6. The type of ground truth used:

    • For Dose Calculations: The ground truth for dose engine validation would be based on physical principles, established dose measurement techniques (e.g., phantom measurements), and validated physics models. It's about the objective accuracy of numerical calculations.
    • For System Functionality: The ground truth for system tests would be the expected output or behavior based on the software's design specifications and requirements.
    • For Clinical Appropriateness/Usability (User Validation): This would be based on expert clinical judgment from the cancer clinics involved, but it's not a "ground truth" for a diagnostic outcome, rather an assessment of the tool's utility.

    7. The sample size for the training set:

    • Not applicable / Not provided. RayStation is a physics-based treatment planning system, not typically an AI/machine learning model in the sense of being "trained" on a large dataset of patient images or outcomes to learn patterns. While it might contain machine learning features (the document mentions "Related to machine learning, there is no change compared to the predicate device," implying some existing ML, but not as the primary function or new feature for the 11.0 update), the core CyberKnife planning feature validated here is based on physics algorithms, not a trained model. Therefore, there's no "training set" in the context of supervised machine learning.

    8. How the ground truth for the training set was established:

    • Not applicable / Not provided. As explained above, the core functionality validated (CyberKnife planning) is algorithm-driven, not learned from a training set. If there are existing ML components, the document doesn't detail their training or ground truth establishment.
    Ask a Question

    Ask a specific question about this device

    K Number
    K210645

    Validate with FDA (Live)

    Device Name
    RayStation 10.1
    Date Cleared
    2021-06-29

    (118 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RayStation is a software system for radiation therapy and medical oncology. Based on user input, RayStation proposes treatment plans. After a proposed treatment plan is reviewed and approved by authorized intended users, RayStation may also be used to administer treatments.

    The system functionality can be configured based on user needs.

    Device Description

    RayStation is a treatment planning system for planning, analysis and administration of radiation therapy and medical oncology treatment plans. It has a modern user interface and is equipped with fast and accurate dose and optimization engines.

    RayStation consists of multiple applications:

    • The main RayStation application is used for treatment planning.
    • . The RayPhysics application is used for commissioning of treatment machines to make them available for treatment planning and used for commissioning of imaging systems.
    • . The RayTreat application is used for sending plans to treatment delivery devices for treatment and receiving records of performed treatments.
    • o The RayCommand application is used for treatment session management including treatment preparation and sending instructions to the treatment delivery devices.
    AI/ML Overview

    The provided text details the 510(k) summary for RayStation 10.1, a software system for radiation therapy and medical oncology. The document indicates that the determination of substantial equivalence to the primary predicate device (RayStation 9.1) is not based on an assessment of non-clinical performance data. Instead, it relies on the entire system verification and validation specifications and reports.

    However, the document does describe the performance data for several new features and explicitly states that these features have been "successfully validated for accuracy in clinically relevant settings according to specification" or "successfully validated according to specification." While these statements imply acceptance criteria were met, the specific numerical acceptance criteria and the reported device performance values are not explicitly provided in a comparative table format within the given text.

    Therefore, the following response will extract the implied acceptance criteria and reported performance from the descriptions provided, and note where specific numerical values are absent.


    Acceptance Criteria and Device Performance Study for RayStation 10.1

    The provided document, K210645 for RayStation 10.1, indicates that the determination of substantial equivalence to the primary predicate device (RayStation 9.1) is not based on non-clinical performance data directly comparing the existing features of 10.1 to 9.1. Instead, it relies on the comprehensive system verification and validation reports.

    However, for the new features introduced in RayStation 10.1, the document states that these features have undergone validation and met their respective specifications. While specific, quantifiable acceptance criteria and reported performance values are not presented in a direct comparative table within the provided text, the descriptions imply the following:

    1. A table of acceptance criteria and the reported device performance

    FeatureImplied Acceptance Criteria (from text)Reported Device Performance (from text)
    Brachytherapy TG43 Dose CalculationAccurately models output from single and combined brachytherapy sources in clinical plans. All doses reported as dose-to-water (DWw).Successfully validated for accuracy in clinically relevant settings according to specification.
    Medical Oncology Dose Calculation FunctionsAppropriate for supporting medical oncology planning workflows when used by qualified users according to IFU.Validated to be appropriate for supporting medical oncology planning workflows.
    Proton Ocular Treatment Dose CalculationAccurately models proton dose calculation for ocular treatments using the single scattering (SS) delivery technique (modeled as double scattering).Successfully validated for accuracy in clinically relevant settings according to specification.
    Robust Planning of Organ MotionCorrectly generates deformed image sets to simulate organ motion and uses them for robust planning against intra-fractional or inter-fractional organ motion.Successfully validated according to specification.

    Note: The provided text does not contain specific numerical acceptance criteria (e.g., "accuracy within X%") or quantitative reported performance data for any of these features. The reported performance essentially states that the criteria were "met" or "validated."

    2. Sample size used for the test set and the data provenance

    The document indicates that RayStation 10.1's test specification is a further developed version of RayStation 9.1's, supported by requirements specification. The verification activities included "User validation in cooperation with cancer clinics." However, no specific sample sizes for test sets (e.g., number of patient cases) or data provenance (e.g., country of origin, retrospective/prospective nature) are provided in the given text for any of the validations.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    The text mentions "User validation in cooperation with cancer clinics" but does not specify the number of experts, their qualifications, or how ground truth was established for the "test set" (if a distinct clinical test set was used for ground truth establishment).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    No information on adjudication methods is provided in the supplied text.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    The text does not mention any multi-reader multi-case (MRMC) comparative effectiveness study, nor does it discuss human readers improving with or without AI assistance. The device is a treatment planning system, and while it states it "proposes treatment plans" based on user input, it does not describe AI-assisted diagnostic or interpretation tasks. It explicitly states, "Related to machine learning, there is no change compared to the primary predicate device." suggesting limited or no direct machine learning components in the new features where such a study would typically be relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The validations described for brachytherapy, medical oncology, proton ocular treatment, and robust planning of organ motion appear to be standalone algorithm performance assessments against defined specifications. These validations verify the accuracy and appropriateness of the software's calculations and functionalities independently, assuming "intended qualified user" interaction for medical oncology, but not as part of a human-in-the-loop performance study.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    For the dose calculation features (brachytherapy, proton ocular treatment), the "ground truth" implicitly refers to theoretical models and established physical principles (e.g., accurate modeling of TG43 formalism, proton dose calculations) as compared against the output of the software. For medical oncology functions, the "ground truth" for validation appears to be whether the functions are "appropriate" for planning workflows, likely assessed against clinical guidelines or expert workflows. For robust planning of organ motion, the ground truth relates to the correct generation and application of deformed image sets according to specifications. The document does not explicitly state that ground truth was established through pathology or outcomes data.

    8. The sample size for the training set

    The document refers to the system as "built on the same software platform" and "developed under the same quality system, by the same development teams." It mentions that "related to machine learning, there is no change compared to the primary predicate device." Given this, and the nature of treatment planning software, the concept of a "training set" in the context of machine learning (e.g., for image classification or prediction models) is not directly applicable or discussed for the validations mentioned. The system's development would involve software engineering and clinical validation rather than machine learning training sets for the described functionalities.

    9. How the ground truth for the training set was established

    As there is no mention of a training set or machine learning components for the new features, information on how its ground truth was established is not provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200569

    Validate with FDA (Live)

    Device Name
    RayStation
    Date Cleared
    2020-09-24

    (204 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RayStation is a software system designed for treatment planning and analysis of radiation therapy. The treatment plans provide treatment unit set-up parameters and estimates of dose distributions expected during the proposed treatment, and may be used to administer treatments after review and approval by the intended user.

    The system functionality can be configured based on user needs.

    The intended users of RayStation shall be clinically qualified radiation therapy staff trained in using the system.

    Device Description

    RayStation 9.1/RayStation 9.2 are radiation therapy treatment planning systems, i.e. a software program for planning and analysis of radiation therapy. The functionality includes fusion capabilities (CT, PET and MRI), contouring, collapsed cone convolution dose computation, 4D data compatibility and treatment console interfacing, as well as unique features such as machine learning planning and segmentation, multi-criteria optimization, dose tracking, treatment adaptation and deformable registration, all available in one platform.

    AI/ML Overview

    The provided text primarily focuses on the FDA 510(k) clearance for RayStation 9.1/9.2, comparing it to a predicate device (RayStation 8.1). While it mentions "test specifications," "successful verification and validation," and "non-clinical tests," it does not contain the detailed information required to answer the questions about acceptance criteria, study design, sample sizes, expert involvement, or specific performance metrics related to a device's performance against acceptance criteria.

    The document states:

    • "The test specifications of RayStation 9.2 are further developed versions of the test specifications of RayStation 8.1."
    • "The successful verification and validation of RayStation 9.2 therefore support the substantial equivalence of the above RayStation versions."
    • "The summary of the performed non-clinical tests shows that RayStation 9.1/RayStation 9.2 are as safe and effective and performs as well as the predicate device."

    These statements indicate that testing was performed to demonstrate substantial equivalence, but they do not provide the details of those tests. The document is about regulatory clearance based on substantial equivalence, not a clinical performance study with specific acceptance criteria and detailed results.

    Therefore, I cannot extract the requested information from the provided text. To answer your questions, I would need a description of a study specifically designed to prove the device meets acceptance criteria, including quantitative performance data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200487

    Validate with FDA (Live)

    Device Name
    RayCare 3.1
    Date Cleared
    2020-06-10

    (104 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RayCare is an oncology information system used to support workflows and scheduling, as well as clinical information, planning and treatment management for oncology care. RayCare may be used in the transfer, storage, conversion and display of patient data in all areas of cancer care including radiation therapy, medical oncology and surgical oncology. RayCare is not intended for use in diagnostic activities.

    Device Description

    RayCare is an oncology information system that supports healthcare professionals in managing cancer care. The system is structured in functionality modules that are listed below.

    RayCare is a software-only system with a client part that allows the user to interact with the system and a server part that performs the necessary processing and storage functions. Selected aspects of RayCare are configurable, such as adapting workflow templates to the specific needs of the clinic.

    • Patient clinical data
    • Care administration
    • Clinical notes and documents
    • Workflow and task management
    • Charge capture
    • Scheduling and calendar functions
    • PACS
    • Image viewer
    • Simulation and imaging
    • System administration
    • Authentication, authorization and audit logging
    • Hospital system integrations
    AI/ML Overview

    The provided text is a 510(k) summary for the RayCare 3.1 device, an oncology information system. However, it does not contain specific acceptance criteria, performance metrics, or detailed study information that would allow a comprehensive answer to your request.

    The document primarily focuses on demonstrating substantial equivalence to a predicate device (RayCare 2.3) by highlighting similar technological characteristics and stating that verification and validation were performed according to relevant standards.

    Therefore, I cannot provide a complete answer to your request based solely on the provided input.

    However, I can extract the available information and indicate where details are missing:

    1. A table of acceptance criteria and the reported device performance

    Acceptance CriteriaReported Device Performance
    Not specified in the document.The document states, "The summary of the performed non-clinical tests shows RayCare 3.1 is as safe and effective and performs as well as the predicate device." However, no specific metrics or thresholds are provided for "safe," "effective," or "performs as well."

    2. Sample size used for the test set and the data provenance

    • Sample size for the test set: Not specified in the document.
    • Data provenance: Not specified in the document.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of experts: Not specified in the document.
    • Qualifications of experts: Not specified in the document.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not specified in the document.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC study: Not specified in the document.
    • Effect size: Not applicable, as no MRMC study or AI assistance is described. The device is an oncology information system, not an AI-powered diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • The document implies that verification and validation were done on the device's functionalities, but it does not detail specific "standalone" performance studies in the context of an algorithm's diagnostic or analytical capability. RayCare 3.1 is an information system, not primarily an AI algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not specified in the document. The document refers to "intended use and functionalities included in a standard oncology information system" for its verification and validation, implying functional rather than diagnostic ground truth.

    8. The sample size for the training set

    • Not applicable/Not specified. This device is an oncology information system, not an AI/ML model that typically requires a training set in the conventional sense.

    9. How the ground truth for the training set was established

    • Not applicable/Not specified, as it's not an AI/ML model with a typical training set.

    Summary of what is available in the document regarding testing:

    • Assessment of non-clinical performance data: "The verification and validation of RayCare 3.1 is based on the intended use and functionalities included in a standard oncology information system, and on the standards ISO 14971, IEC 61217, IEC 62304 and IEC 62366."
    • Test conclusion: "The summary of the performed non-clinical tests shows RayCare 3.1 is as safe and effective and performs as well as the predicate device."

    This document focuses on demonstrating substantial equivalence to an existing predicate device by showing that the new device has the same technological characteristics and has undergone standard verification and validation processes applicable to medical software. It does not contain the kind of detailed performance study data often found for AI/ML-driven diagnostic devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191384

    Validate with FDA (Live)

    Device Name
    RayCare 2.3
    Date Cleared
    2019-07-08

    (46 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RayCare is an oncology information system used to support workflows, scheduling and clinical information management for oncology care and follow-up. RayCare may be used in the transfer, storage, conversion and display of patient data in all areas of cancer care including radiation therapy, medical oncology and surgical oncology. RayCare is not intended for use in diagnostic activities.

    Device Description

    RayCare is an oncology information system that supports healthcare professionals in managing cancer care. The system is structured in functionality modules that are listed below. RayCare is a software-only system with a client part that allows the user to interact with the system and a server part that performs the necessary processing and storage functions. Selected aspects of RayCare are configurable, such as adapting workflow templates to the specific needs of the clinic.

    • Patient clinical data
    • Care administration
    • Clinical notes and documents
    • Workflow and task management
    • Charge capture
    • Scheduling and calendar functions
    • PACS
    • Image viewer
    • Simulation and imaging
    • System administration
    • Authentication, authorization and audit logging
    • Hospital system integrations
    AI/ML Overview

    The provided text is a 510(k) summary for the RayCare 2.3 device. It describes the device, its intended use, and its technological characteristics. However, it does not provide specific acceptance criteria or a detailed study proving the device meets those criteria with performance metrics.

    The document states that "The verification and validation of RayCare 2.3 is based on the intended use and functionalities included in a standard oncology information system, and on the standards ISO 14971, IEC 61217, IEC 62304 and IEC 62366. The successful verification and validation of RayCare 2.3 therefore support the substantial equivalence to the MOSAIQ Oncology Information System predicate device."

    This indicates that the device's performance was assessed against its intended use and relevant industry standards, demonstrating "substantial equivalence" to a predicate device (MOSAIQ Oncology Information System) rather than meeting specific, quantitative acceptance criteria with reported data.

    Therefore, I cannot provide the requested information. The document focuses on regulatory approval through substantial equivalence, which does not typically include the level of detail regarding acceptance criteria and performance studies that would be found in a clinical trial report or a more comprehensive technical document.