    K Number
    K031206

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-05-21

    (35 days)

    Product Code
    Regulation Number
    892.1750
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    RadioMed™ Soft Tissue Marker is indicated for use to radiographically mark soft tissue for future therapeutic procedures.

    Device Description

    The RadioMed™ Soft Tissue Marker is a non-sterile device, in the form of a gold coil that ranges in OD between 0.75mm and 1.2mm.

    The RadioMed™ Soft Tissue Marker is packaged non-sterile, single use, and is to be sterilized by the end user in accordance with a validated sterilization process. Sterilization is achieved by exposure to steam autoclave.

    The RadioMed™ Soft Tissue Marker will be manufactured, labeled, and packaged in accordance with the current FDA QSR. To ensure compliance to specifications, upon completion of the manufacturing process the device will be inspected and tested in accordance with RadioMed standard operating procedures.

    The RadioMed™ Soft Tissue Marker will be delivered using either a 17 gauge or 18 gauge needle and stylet.

    AI/ML Overview

    This document describes a 510(k) submission for the RadioMed™ Soft Tissue Marker, which is a medical device. The submission focuses on demonstrating substantial equivalence to predicate devices rather than proving specific performance characteristics against acceptance criteria in a comprehensive study.

    Based on the provided text, there is no specific acceptance criteria table or a study proving the device meets distinct performance criteria in terms of accuracy, sensitivity, specificity, etc. This is because 510(k) submissions for devices like markers often rely on demonstrating that the new device is as safe and effective as a legally marketed predicate device. The performance testing mentioned is related to manufacturing standards and visibility, not a clinical effectiveness study with defined performance metrics.

    Here's a breakdown of the information that can be extracted or inferred, and what is not present:

    Missing Information (Not Applicable to this 510(k) submission):

    • 1. A table of acceptance criteria and the reported device performance: Not provided. The submission focuses on substantial equivalence to predicate devices and manufacturing quality, not a comparison against specific clinical performance metrics.
    • 2. Sample size used for the test set and the data provenance: Not applicable. There's no "test set" in the context of a clinical performance study as described in the prompt.
    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No clinical ground truth establishment is described.
    • 4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    • 5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done: No.
    • 6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: No. This is a physical marker, not an AI algorithm.
    • 7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable.
    • 8. The sample size for the training set: Not applicable. This is not an AI/algorithm-based device requiring a training set.
    • 9. How the ground truth for the training set was established: Not applicable.

    Information that is present or can be inferred:

    Overview of Evidence Presented:

    The submission focuses on comparing the RadioMed™ Soft Tissue Marker to existing predicate devices (RadioMed™ Marker, Implanter/Adjustable/Anderson's Marker, Extracranial Marker) to demonstrate substantial equivalence. The key aspects of this comparison are:

    • Intended Use: The intended use and indications for use have not changed from the predicate device.
      • Intended Use: "RadioMed™ Soft Tissue Marker is indicated for use to radiographically mark soft tissue for future therapeutic procedures."
    • Technological Characteristics: The fundamental scientific technology of the modified device has not changed.
      • Material: The material (metallic gold) is identical to the Anderson Marker (K940121) and Extracranial Marker, and is widely used in other approved medical devices with a "safe history of use." This obviates the need for additional biocompatibility testing.
      • Design: The design is identical to the predicate RadioMed™ Marker (K022326) except for the outside diameter (OD) and material (gold vs. unspecified material in K022326). The new marker ranges in OD between 0.75mm and 1.2mm.
    • Performance Testing (Non-Clinical):
      • Summary of standards achieved:
        • FDA QSR 21 CFR Part 820 Good Manufacturing Practices
        • AAMI Standard 11134-1994 Recommended practice for Steam Autoclave
      • Visibility Studies: "Visibility Studies (see section H - Performance Testing)" are mentioned, implying that the marker's visibility under radiographic imaging was assessed. However, no specific data, acceptance criteria, or experimental setup for these visibility studies are provided in this summary.

    Conclusion from a 510(k) Perspective:

    The FDA's letter (K031206) confirms that they have "determined the device is substantially equivalent...to legally marketed predicate devices." This determination is based on the information provided in the 510(k) submission, primarily focusing on the safety and effectiveness as demonstrated through comparison to predicates and adherence to manufacturing and sterilization standards, rather than direct clinical performance against novel acceptance criteria.

    In summary, for this specific 510(k) submission, the "acceptance criteria" and "study" are intrinsically linked to demonstrating substantial equivalence through comparison with predicate devices and adherence to established manufacturing and safety standards, rather than a clinical trial proving specific performance metrics of an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K001070

    Validate with FDA (Live)

    Device Name
    RADIOMED SOURCE
    Manufacturer
    Date Cleared
    2000-10-24

    (204 days)

    Product Code
    Regulation Number
    892.5730
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The RadioMed"10 Source with activities from 0.1 to 5.0mCi per centimeter is indicated for temporary or permanent interstitial implantation or surface application to treat selected localized turnors. They can be used either as primary treatment (such as for prostate cancer or unresectable tumors) or as treatment for residual disease after excision of primary or recurrent tumors. The Pd-103 sources may be used concurrently with or following treatment with other interventions, such as external beam therapy, hyperthermia, or chemotherapy. Tumors of the head, neck, lung, pancreas, prostate and other accessible tumors are commonly treated.

    Device Description

    The RadioMed™ source utilizes Pd-103 as the radionuclide for brachytherapy. The RadioMed The Source exists in the form of a coiled rhodium wire. The rhodium matrix also provides the radiopacity of the device. The RadioMed™ Source is a naturally sealed source in the form of a wire.

    The RadioMed The Source is packaged non-sterile, single use, and is intended to be sterilized by the end user in accordance with a validated sterilization process. Sterilization is accomplished by exposure to steam autoclave.

    The RadioMed" Source will be manufactured. labeled and packaged under GMP controls. Upon completion of the manufacturing and assembly process the device will be inspected to assure compliance to specifications. The devices will be tested in accordance with Standard Operating Procedures.

    The sources are delivered in the same manner as brachytherapy seeds currently on the market, i.e. a 17 or 18-gauge needle and stylet.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the "RadioMed™ Source," a brachytherapy radionuclide source. It primarily focuses on demonstrating substantial equivalence to existing predicate devices. Consequently, the document does not detail specific acceptance criteria for performance metrics or present a study proving the device meets those criteria in the way one might expect for a diagnostic or AI-driven device.

    Instead, the "Performance Testing" section outlines the standards achieved for different aspects of the device, implying that compliance with these standards serves as the "acceptance criteria" and the "study" is the demonstration of adherence to these standards.

    Here's an attempt to structure the information based on your request, acknowledging the limitations of the provided text for certain categories:


    Acceptance Criteria and Device Performance Study for RadioMed™ Source

    The RadioMed™ Source is a brachytherapy radionuclide source. Its "performance" is subdivided into four categories, and the device's acceptance is based on its compliance with established standards and specifications for these categories and general regulatory requirements.

    1. Table of Acceptance Criteria and Reported Device Performance

    Category / Performance AspectAcceptance Criteria (Standard / Description)Reported Device Performance
    General Manufacturing & QualityFDA QSR 21 CFR Part 820 Good Manufacturing PracticesManufactured, labeled, and packaged under GMP controls. Device will be inspected to assure compliance to specifications. Devices will be tested in accordance with Standard Operating Procedures.
    BiocompatibilityISO 10993-1 1992 (E) Biological Evaluation of Medical DevicesAchieved (Summary states "Summary of standards achieved: ISO 10993-1 1992 (E) Biological Evaluation of Medical Devices")
    Radioactive Source ClassificationANSI N43.6-1997: Classification of sealed radioactive sourcesAchieved (Summary states "Summary of standards achieved: ANSI N43.6-1997: Classification of sealed radioactive sources")
    Integrity & Test SpecificationsANSI N44.1-1973 Integrity and Test Specifications for selected Brachytherapy SourcesAchieved (Summary states "Summary of standards achieved: ANSI N44.1-1973 Integrity and Test Specifications for selected Brachytherapy Sources")
    Leak TestingANSI N44.2-1973 Leak testing radioactive brachytherapy sources / ISO 9978: 1992(E) "Radiation protection - Sealed radioactive sources - Leakage test methods".Achieved (Summary states "Summary of standards achieved: ANSI N44.2-1973 Leak testing radioactive brachytherapy sources" and "ISO 9978: 1992(E) "Radiation protection - Sealed radioactive sources - Leakage test methods".") The RadioMed™ Source is described as a naturally sealed source.
    Sterilization (by end user)AAMI Standard 11134-1994 Recommended practice for Steam AutoclaveIntended to be sterilized by the end user in accordance with a validated sterilization process (steam autoclave).
    Radiation EnergyEnergy emitted for 103Pd (20-23 keV x-rays)The energy emitted for the RadioMed™ Source and the Theragenics 103Pd Seed is exactly the same: 20-23 keV x-rays.
    Source Strength MeasurementCalibration and Calibration AccuracyPerformance is subdivided into this category, implying that established calibration and accuracy standards are met (details not provided in the text).
    Dose Comparison(Implicit comparison to predicate devices for therapeutic effect)Performance is subdivided into this category, implying that dose characteristics are comparable to predicate devices (details not provided in the text).
    Sealed Source Testing(Implicit integrity and safety testing for sealed sources)Performance is subdivided into this category, implying that all necessary tests for sealed sources are conducted (details not provided in the text). The device is described as a "naturally sealed source."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of a typical clinical trial or performance study for a diagnostic device. The performance testing appears to refer to engineering and quality assurance tests conducted on manufactured units.

    • No specific sample size for a "test set" is mentioned.
    • Data provenance (country of origin, retrospective/prospective) is not applicable or provided, as the testing relates to manufacturing and product specifications rather than clinical data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts

    Not applicable. The ground truth for this device is based on objective measurements against engineering standards, physical properties of the radionuclide, and established regulatory benchmarks, rather than subjective expert interpretation of data.

    4. Adjudication Method for the Test Set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used for expert consensus on subjective data (e.g., image interpretation). For this device, compliance is determined by objective measurements and adherence to technical standards.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a brachytherapy source, not a diagnostic imaging or AI-driven device intended to assist human readers. Therefore, an MRMC study is irrelevant to its evaluation.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. The RadioMed™ Source is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the RadioMed™ Source's performance and safety is:

    • Engineering Specifications: Compliance with design blueprints and material properties.
    • Physical Standards: Adherence to established physical properties of the radionuclide (Pd-103 energy emission).
    • Regulatory Standards: Compliance with national and international standards for medical devices, manufacturing (GMP, QSR), radiation safety (ANSI, ISO), and biocompatibility (ISO).
    • Predicate Device Equivalence: The comparison to legally marketed predicate devices serves as a "ground truth" for demonstrating substantial equivalence in function and safety.

    8. The Sample Size for the Training Set

    Not applicable. This device does not involve machine learning or AI, so there is no "training set."

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as there is no training set.