Search Filters

Search Results

Found 33 results

510(k) Data Aggregation

    K Number
    K251507

    Validate with FDA (Live)

    Device Name
    INTEGRE LIO
    Manufacturer
    Date Cleared
    2026-02-06

    (266 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    K Number
    K232302

    Validate with FDA (Live)

    Device Name
    Pocket III
    Manufacturer
    Date Cleared
    2024-04-26

    (269 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Pocket III is intended to be used in ophthalmology clinical applications for measurement of corneal thickness by ultrasonic means with pachymetry. This device should be operated by doctors or other appropriately-trained healthcare professionals and should be used in health institutions. The device is not intended for fetal use.

    Device Description

    The Pocket III is a small hand-held device in pen-shaped design. Measurement results appear numerically on the LCD graphic display. The user can perform a basic cable transfer (USB-C) or activate Bluetooth to export data to a computer. The Pocket III is equipped with an integrated 30 MHz pachymetry probe. The ultrasonic transducer makes contact with, and transmits ultrasonic pulses through, the surface cornea. Echoes are returned from the anterior and posterior surfaces of the cornea.

    AI/ML Overview

    Acceptance Criteria and Device Performance Study for Quantel Medical Pocket III

    Here's an analysis of the provided information regarding the acceptance criteria and the study proving the device meets them:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriterionReported Device Performance
    Accuracy+/- 5 microns
    Resolution1 micron
    Corneal thickness range150-1200 microns
    Ultrasonic Transducer Frequency30 MHz
    Tip Diameter1.2 mm
    Focal Point0.5 mm from tip
    Angle35 degrees
    Thermal Index<1
    Electrical SafetyComplies with IEC 60601-1
    EMC (Electromagnetic Compatibility)Complies with IEC 60601-1-2
    UsabilityComplies with IEC 60601-1-6
    Ultrasound SafetyComplies with IEC 60601-2-37
    Battery SafetyComplies with IEC 62133-2 Ed. 1.0 2017
    BiocompatibilityComplies with ISO 10993-1 for surface device, intact skin, limited duration
    Software PerformancePerforms as intended, appropriate for Moderate level of concern

    2. Sample Size Used for the Test Set and Data Provenance

    The document states: "Pachymetry measurements recorded with test blocks were verified for all available freeze acquisition measuring modes (Easy, Medium and Hard) from 208 um to 1042 um of theoretical value of test block."

    • Sample Size for Test Set: Not explicitly stated as a number of individual measurements or "patients." The testing involved "test blocks" covering a range of corneal thickness values (208 µm to 1042 µm). The number of individual measurements taken on these blocks is not specified.
    • Data Provenance: The data is from non-clinical testing using "test blocks." This indicates an in-vitro or simulated environment, not human subject data. Therefore, there is no country of origin or retrospective/prospective status as it is not derived from human patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Not applicable. The ground truth for the test set (the "theoretical value of test block") appears to be based on the known, calibrated values of the test blocks themselves, rather than expert interpretation of patient data.
    • Qualifications of Experts: Not applicable, as expert consensus was not used to establish the ground truth for these test blocks.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. As the ground truth was based on the known theoretical values of the test blocks, there was no need for expert adjudication. The device's measurements were directly compared to these known values.

    5. Was a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Done?

    • No, an MRMC comparative effectiveness study was not done. The document describes non-clinical testing focused on the device's technical specifications and accuracy against test blocks, not a comparison of human reader performance with and without AI assistance.

    6. Was a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Done?

    • Yes, in essence, a standalone performance evaluation was done for the device's core pachymetry function. The statement "Pachymetry measurements recorded with test blocks were verified... all results are acceptable and compliant with an accuracy of ±5μm" describes the device (algorithm and hardware) operating independently to produce a measurement, which was then compared to a known standard (the test block). While it's a physical device rather than purely an algorithm, the evaluation focuses on its direct output accuracy.

    7. The Type of Ground Truth Used

    • Type of Ground Truth: The ground truth used was based on the known, theoretical values of calibrated test blocks.

    8. The Sample Size for the Training Set

    • Sample Size for Training Set: The document does not provide information regarding a training set. This is not a machine learning or AI device that typically requires a large training dataset for its primary function of ultrasonic pachymetry measurement. The device's operation relies on established physical principles of ultrasound.

    9. How the Ground Truth for the Training Set Was Established

    • Not applicable. As no training set information is provided, the method for establishing its ground truth is also not mentioned.
    Ask a Question

    Ask a specific question about this device

    K Number
    K220430

    Validate with FDA (Live)

    Device Name
    Capsulo
    Manufacturer
    Date Cleared
    2022-05-12

    (86 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CAPSULO is indicated for photodisruption of ocular tissue using light energy emitted by a Nd: YAG laser, including discission of the posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy.

    Device Description

    The Yttrium Aluminum Garnet (YAG) ophthalmic laser system produces short pulses of focused infrared light at a wavelength of 1.064 nm. These pulses can be precisely positioned using a slit lamp microscope and a dual-point aiming system. The energy contained in a short pulse is concentrated by focusing a very small spot (10 um) so that the formation of plasma is at the focal point. The firing causes an acoustic wave that radiates from the focal point and breaks the near tissue which is known as disruptive photo-effect. Once formed, plasma absorbs and further diffuses incident light, protecting deeper structures. Furthermore, the divergence of the beam after the focal point protects the retina from damage that could otherwise occur by the absorption of concentrated energy in the Nd: YAG treatment beam. As the treatment energy is increased, the size of the plasma formed also increases, until a larger and stronger acoustic wave is produced. For increased energy, it is necessary to focus the treatment beam further behind the ocular structure being treated. The Nd: YAG 1064 nm laser operates at a nano-second pulse rate and the laser energy is delivered to the treatment site using the integrated slit lamp.

    AI/ML Overview

    The provided text is a 510(k) summary for the Quantel Medical CAPSULO device, which is an ophthalmic laser. It focuses on demonstrating substantial equivalence to a predicate device rather than providing a detailed study proving the device meets specific performance acceptance criteria for an AI/ML-driven device.

    Therefore, the information required to answer the prompt regarding acceptance criteria, study details, sample sizes, expert involvement, ground truth, and training set information for an AI/ML device is not present in the provided document.

    The document discusses the following:

    • Device Type: Nd: YAG ophthalmic laser (not an AI/ML device).
    • Purpose: Photodisruption of ocular tissue (posterior capsulotomy, pupillary membranectomy, iridotomy).
    • Equivalence: The submission claims substantial equivalence to a predicate laser device (Quantel Medical Optimis Fusion).
    • Performance Data: States non-clinical testing was performed to verify design specifications and substantial equivalence, and lists the standards adhered to (electrical safety, EMC, usability, laser safety, biocompatibility). It also mentions software verification and validation activities for a "Major level of concern," but this refers to the software controlling the laser, not a diagnostic AI algorithm.

    In summary, there is no information in the provided text about acceptance criteria for an AI/ML device's performance, nor any study details that would typically be conducted for verifying an AI/ML model's diagnostic accuracy or effectiveness.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213254

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2021-11-15

    (46 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The AXIALIS Ophthalmic Ultrasound System and the probes that are indicated for diagnostic imaging and biometric measurement of the eye including:

    • · Axial Length measurement of the eye by ultrasonic means.
    • · Implanted IOL power calculation, using the Axial Length measurement.
    • · Measurement of corneal thickness by ultrasonic means.
    Device Description

    The AXIALIS Ophthalmic Ultrasound System is an ultrasonic system for ophthalmology which consists of a base and probes. The base performs the same calculations and displays as the predicate device, the Quantel Medical COMPACT TOUCH. The A-scan probe for biometry and pachymetry is identical to the predicate device.

    AI/ML Overview

    The provided text is a 510(k) summary for the Quantel Medical AXIALIS Ophthalmic Ultrasound System. It details the device's indications for use, its substantial equivalence to a predicate device (Quantel Medical Compact Touch Ophthalmic Ultrasound System), and the performance data submitted for its clearance.

    However, the document focuses on non-clinical testing related to electrical safety, EMC, usability, and specific requirements for ultrasonic medical diagnostic equipment, as well as hardware and software validation. It does not contain information about clinical performance studies regarding diagnostic accuracy, sensitivity, specificity, or how the device's measurements compare to a "ground truth" or to human readers' performance.

    Therefore, I cannot fully answer your request based on the provided text, especially regarding:

    • Acceptance criteria and reported device performance related to diagnostic accuracy. The text mentions "biometric measurement of the eye," but no data on the accuracy of these measurements is provided.
    • Sample size used for a clinical test set or data provenance.
    • Number of experts, qualifications, or adjudication methods for establishing ground truth.
    • Any multi-reader multi-case (MRMC) comparative effectiveness study.
    • Standalone algorithm performance.
    • Type of ground truth used (clinical/pathology/outcomes data).
    • Training set size or ground truth establishment for a training set.

    The document primarily focuses on demonstrating substantial equivalence based on technological characteristics and non-clinical performance (safety, EMC, usability, software validation) rather than direct clinical accuracy studies with human participation.

    To answer your questions, clinical study data would be required, which is not present in this 510(k) summary.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201502

    Validate with FDA (Live)

    Device Name
    Vitra 810
    Manufacturer
    Date Cleared
    2020-08-06

    (62 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vitra 810 is intended for retinal photocoagulation, laser trabeculoplasty, transscleral cyclophotocoagulation, transscleral retinal photocoagulation, iridotomy, including the following:
    Retinal photocoagulation for the treatment of:
    Diabetic retinopathy including:
    Nonproliferative retinopathy
    Macular edema
    Proliferative retinopathy
    Retinal Tears, Detachments and Holes
    Lattice degeneration
    Age-related macular degeneration (AMD) with choroidal neovascularization (CNV)
    Retinopathy of prematurity
    Sub-retinal (choroidal) neovascularization
    Central and Branch Retinal Vein Occlusion
    Laser Trabeculoplasty, Iridotomy, Transscleral Cyclophotocoagulation for the treatment of glaucoma, including:
    Primary open angle
    Closed angle
    Refractory Glaucoma (recalcitrant/uncontrolled)

    Device Description

    The Vitra 810 is a laser system which emits a treatment beam at 810 nm and is intended for use in photocoagulation of ocular tissues in the treatment of diseases of the eye. The laser is used to treat the retinal diseases and glaucoma.
    The Vitra 810 is a laser system which emits a treatment beam at 810 nm and is intended for use in photocoagulation of ocular tissues in the treatment of diseases of the eye. The laser is used to treat the anterior and posterior segments of the eye. The laser is particularly well suited for treating the eye because it has minimal effect on transparent tissues and materials. This means that the laser can be efficiently delivered to opaque structures of the eye through the transparent cornea, aqueous humor, lens, and vitreous humor. This allows many conditions to be treated by non-invasive techniques.
    Laser energy is delivered to opaque structures within the eye by means of laser delivery systems including slit lamp, indirect ophthalmoscope, operating microscope, cyclophotocoagulation probe and laser probe. The standard delivery system includes a lens system to focus the laser energy and vary the size of the laser spot in the plane of observation of the slit lamp, for example. The laser energy is delivered to the delivery system by the means of a flexible fiber optic. For most procedures, a laser contact lens is used to direct the laser energy to the part of the eye being treated. The contact lens may have mirrors so that laser energy can be delivered to areas of the retina behind the iris, or into the angle so that the trabecular meshwork can be treated. The contact lens also helps to hold the eye open and still so that the laser energy can be delivered effectively.

    AI/ML Overview

    The Quantel Medical Vitra 810 is an ophthalmic laser system intended for photocoagulation of ocular tissues in the treatment of diseases of the eye, specifically retinal diseases and glaucoma.

    Here's an analysis of the acceptance criteria and the study information provided in the document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of quantitative acceptance criteria for device performance. Instead, substantial equivalence is claimed based on comparing the technological characteristics and intended use of the Vitra 810 with two predicate devices. The "Performance Data" section primarily lists compliance with recognized consensus standards related to safety and essential performance, as well as hardware and software validation. These can be considered the implicit acceptance criteria, demonstrating that the device meets established regulatory and technical benchmarks.

    Acceptance Criteria CategoryDetails / Reported Device Performance
    Intended UsePhotocoagulation of ocular tissue including the retina and pigmented tissue (Matches predicates). Specific indications for use (retinal photocoagulation for diabetic retinopathy, retinal tears, AMD, etc., and laser trabeculoplasty, iridotomy, transscleral cyclophotocoagulation for glaucoma) are identical to the secondary predicate (Iridex IQ 810).
    Technological Characteristics (Comparison)Laser Source: Laser Diode (Matches predicates)
    Wavelength: 810 nm (Matches predicates)
    Treatment Power: 50-3000 mW (depending on delivery device) (Matches predicates)
    Laser Safety Class: 4/IV (Matches predicates)
    Pulse Duration (Continuous Wave): 10 ms to continuous (up to 60 seconds) (Matches predicates)
    Pulse Duration (Subthreshold Mode - Ton): 0.1 ms to 1.0 ms (5% to 35% duty cycle) (Within range of Iridex IQ 810's 0.025 ms to 1.0 ms with 0.5% to 50% duty cycle)
    Repeat Interval (Continuous Wave): 50 ms to 1000 ms and one pulse (Similar to Iridex IQ 810's 50 ms to 1000 ms and one pulse; predicate Supra Twin states "Repeat mode: 0.1 - 1 s")
    Repeat Interval (Subthreshold Mode - Toff): 0.3 ms to 19 ms (duty cycle: 5% to 35%) (Matches Iridex IQ 810's 1.0 ms to 9.5 ms, 0.5% to 50% duty cycle within acceptable parameters)
    Aiming Beam Source: Red Laser Diode, 1 mW maximum (Matches predicates)
    Aiming Beam Wavelength: 635 - 650 nm (Matches predicates)
    Compliance with StandardsIEC 60601-1:2005 + Corr. 1:2006+Corr. 2:2007+A1:2012 (General requirements for basic safety and essential performance)
    IEC 60601-1-2:2014 (Electromagnetic compatibility)
    IEC 60601-1-6:2010+A1:2013 (Usability)
    IEC 60601-2-22: 2007 (Third Edition) + A1:2012 (Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment)
    IEC 60825-1 (Safety of Laser Products – Equipment Classification and Requirements)
    Software and Hardware ValidationHardware and software validation activities were performed to ensure the device performed as intended and software documentation appropriate for the major level of concern was provided.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not describe a clinical test set with a specific sample size, nor does it mention data provenance (e.g., country of origin, retrospective/prospective). The assessment for substantial equivalence is based on a comparison of technical specifications and intended use against legally marketed predicate devices, along with compliance to international safety standards. It does not appear to involve a new clinical study with a patient test set in the traditional sense.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not applicable as the documentation does not describe a clinical study with a test set requiring expert-established ground truth. The device is a medical laser, and its substantial equivalence is determined by engineering and regulatory comparisons, not by diagnostic performance against an expert-derived ground truth.

    4. Adjudication Method for the Test Set

    This information is not applicable, for the same reasons as point 3.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This information is not applicable. The Vitra 810 is a surgical laser device (a treatment tool), not a diagnostic device or an AI-assisted diagnostic tool. Therefore, an MRMC study assessing human reader improvement with AI assistance is irrelevant for this type of medical device.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) Was Done

    This information is not applicable. The Vitra 810 is a medical laser system operated by trained medical personnel. It is not an algorithm-only or AI-based device, so standalone algorithm performance testing is irrelevant.

    7. The Type of Ground Truth Used

    This information is not applicable. As discussed, the substantial equivalence decision is based on technical specifications, intended use, and compliance with safety standards, rather than the performance of a diagnostic algorithm against a clinical ground truth like pathology or outcomes data.

    8. The Sample Size for the Training Set

    This information is not applicable. There is no mention of a training set as the device is not an AI/ML-based system requiring data for training.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191962

    Validate with FDA (Live)

    Device Name
    Vitra 2
    Manufacturer
    Date Cleared
    2019-12-05

    (135 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vitra 2 is for use in the treatment of ocular pathology of anterior segments including retinal photocoagulation and pan retinal photocoagulation of vascular and structural abnormalities of the retina and choroids including:

    • · Proliferative and nonproliferative diabetic retinopathy
    • Choroidal neovascularization
    • · Branch retinal vein occlusion
    • · Treatment of choroidal neovascularization associated with wet Age-related macular degeneration
    • · Retinal tears and detachments
    • · Macular edema
    Device Description

    The Vitra 2 is a laser system which emits a treatment beam at 532 nm and is intended for use in photocoagulation of ocular tissues in the treatment of diseases of the eye.Laser energy is delivered to opaque structures within the eye by means of delivery systems including slit lamp, indirect ophthalmoscope, operating microscope and endocular probe. The standard delivery system includes a lens system to focus the laser energy and vary the size of the laser spot in the plane of observation of the slit lamp, for example. It also includes a Multispot delivery system made by scanner motors which can deliver several spots to the retina in a pattern. The laser energy is delivery system by the means of a flexible fiber optic. For most procedures, a laser contact lens is used to direct the laser energy to the part of the eye being treated. The contact lens also helps to hold the eye open and still so that the laser energy can be delivered effectively.

    The Quantel Medical VITRA 2 includes a Scanning Laser Delivery System adaptor with scanner controls that may be coupled to a slit lamp. Once activated by the user, the VITRA 2 delivers a predetermined pattern by sequentially scanning the placement of the laser spots and the emission of the individual pulses of laser light. Treatment is initiated by pressing the footswitch and may be aborted by releasing the footswitch.

    AI/ML Overview

    The provided text is a 510(k) summary for the Quantel Medical Vitra 2 device. It describes the device, its intended use, and argues for its substantial equivalence to a predicate device (Quantel Medical Vitra Multispot).

    However, the information requested in your prompt (Acceptance Criteria and a study proving the device meets the criteria, particularly in the context of an AI/ML-driven device) is not present in this document. The Vitra 2 is a laser surgical instrument, and its performance data focuses on compliance with electrical and laser safety standards (IEC 60601 series, IEC 60825-1) and hardware/software validation. There is no mention of AI/ML components, a test set, expert readers, ground truth establishment for a diagnostic algorithm, or MRMC studies.

    Therefore, I cannot fulfill your request using the provided text. The requested information pertains to a different type of device evaluation (typically for AI/ML-based diagnostic or prognostic tools) than what is described for the Vitra 2 laser system.

    To illustrate, if this were an AI/ML device document, the "Performance Data" section would ideally contain details such as:

    • Acceptance Criteria Table: Specific metrics (e.g., accuracy, sensitivity, specificity, AUC) with a pre-defined threshold that the device's performance must meet.
    • Study Design: Description of the test set, sample size, data provenance, and how ground truth was established.
    • Expert Consensus: Details on the number and qualifications of experts involved in ground truth labeling and any adjudication process.
    • MRMC Study: Information on comparative effectiveness studies, if human readers were assisted by the AI.
    • Standalone Performance: Results when the algorithm performs without human intervention.
    • Training Set Details: Information on the size and ground truth establishment for the data used to train the AI model.
    Ask a Question

    Ask a specific question about this device

    K Number
    K183414

    Validate with FDA (Live)

    Device Name
    ABSOLU
    Manufacturer
    Date Cleared
    2019-03-20

    (100 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Quantel Medical ABSOLU is intended to be used for:

    -Axial Length measurement of the eye by ultrasonic means;

    -Implanted IOL power calculation, using the Axial Length measurement;

    • Visualization of the interior of the eye and the orbit by A and B scans.
    Device Description

    The ABSOLU is a high definition multifunction ophthalmic ultrasound system used for the following intended use:

    • -Axial Length measurement of the eye by ultrasonic means;
    • Implanted IOL power calculation, using the axial length measurements; -
    • Visualization of the interior of the eve and the orbit by A and B scans. -
    AI/ML Overview

    The Quantel Medical ABSOLU device is intended for axial length measurement of the eye, implanted IOL power calculation using axial length measurements, and visualization of the interior of the eye and orbit via A and B scans. The performance data and acceptance criteria are outlined below, primarily focusing on safety and compliance with recognized standards rather than specific diagnostic accuracy metrics of a standalone AI algorithm.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria / Performance AspectReported Device Performance
    Electrical SafetyComplies with AAMI ANSI ES60601-1:2005 + Corr. 1:2006+Corr. 2:2007+A1:2012 (Medical electrical equipment-Part 1: General requirements for basic safety and essential performance).
    Electromagnetic Compatibility (EMC)Complies with IEC 60601-1-2:2014 (Medical electrical equipment-Part 1-2: General requirements for basic safety and essential performance-Collateral standard: Electromagnetic compatibility-Requirements and tests).
    UsabilityComplies with IEC 60601-1-6:2010+A1:2013 (Medical electrical equipment-Part 1-6: General requirements for safety-Collateral Standard Usability). This is supported by the improved, more intuitive user interface.
    Safety of Ultrasonic Medical Diagnostic and Monitoring EquipmentComplies with IEC 60601-2-37: "Ed2 + AM1 with 60601-1 (ed.3), AM1 with correc 1 & correc 2 Particular requirements for the safety of ultrasonic medical diagnostic and monitoring equipment". This specifically addresses performance of the B-scan probes.
    Hardware and Software ValidationHardware and software validation activities were performed to ensure the device performs as intended. Software documentation appropriate for the Moderate level of concern was provided.
    Technological Changes (e.g., B-scan probe improvements, frequency changes)The 20 MHz annular B-scan probe technology increases depth of field by 70%, allowing simultaneous examination of vitreous, retina, and orbit pathologies without compromising image quality. The 15 MHz B-scan probe offers superior resolution and improved measurement accuracy compared to the previous 10 MHz version, with a focal point change from 25 mm to 24 mm. These changes do not affect safety or performance.
    Measurement Accuracy (Axial Length, IOL power calculation)Improved with the 15 MHz B-scan probe due to increased frequency. (No specific quantitative metric provided in the document for comparison against a threshold.)

    2. Sample Size Used for the Test Set and Data Provenance

    The provided document describes tests performed for regulatory compliance (safety, EMC, usability, and ultrasonic equipment safety standards) and validation of technological changes. It does not mention a specific "test set" in the context of clinical data for diagnostic or measurement accuracy of a standalone AI algorithm. The performance data refers to instrumental performance, safety, and functional validation against standards for the device as a whole and its components (probes, software). Therefore, information regarding sample size, country of origin, or retrospective/prospective nature of a clinical test set is not available in this document.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the document, as the evaluations described are primarily for device safety, EMC, usability, and technical performance against engineering standards, not a clinical diagnostic performance study with expert-adjudicated ground truth.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set

    This information is not provided in the document, as the evaluations described are for device safety and technical performance, not for clinical diagnostic performance adjudicated by experts.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study is present in the document. The device is an ultrasound system with specific measurement and visualization capabilities, and the performance data focuses on its compliance with operational, safety, and technical standards, rather than an AI-driven diagnostic accuracy study comparing human readers with and without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    The document does not describe a standalone performance study for an AI algorithm without human intervention. The ABSOLU is described as an "ultrasonic pulsed echo imaging system" intended for use by "trained medical personnel" where "the use of the different parameters of the ABSOLU® and the predicate device are under control of the physician for the different phases of the diagnostic." This strongly suggests a human-in-the-loop system, not a standalone AI.

    7. Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

    The document does not detail specific "ground truth" for clinical diagnostic or measurement accuracy in the context of expert consensus, pathology, or outcomes data. The "performance data" mentioned pertains to compliance with recognized consensus standards for electrical safety, electromagnetic compatibility, usability, and ultrasonic medical device safety. These standards serve as the "ground truth" for the device's technical and safety performance.

    8. Sample Size for the Training Set

    No information regarding a training set sample size is provided. The document outlines a regulatory submission for an ultrasound imaging system, not an AI algorithm that typically requires a large training dataset for its development. The "software" mentioned as being revised is general device software, not an AI component trained on data.

    9. How the Ground Truth for the Training Set Was Established

    Since no training set for an AI algorithm is mentioned, there is no information on how ground truth for such a set was established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180265

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-03-02

    (30 days)

    Product Code
    Regulation Number
    892.1560
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The COMPACT TOUCH Ophthalmic Ultrasound System and the probes that are used with it are indicated for diagnostic imaging and biometric measurement of the eye including:

    • Visualization of the interior of the eye and the orbit by A and B scans.
    • Axial Length measurement of the eye by ultrasonic means.
    • Implanted IOL power calculation, using the Axial Length measurement.
    • Measurement of corneal thickness by ultrasonic means.
    Device Description

    The COMPACT TOUCH Ophthalmic Ultrasound System is an ultrasonic system for ophthalmology. It is a modification to the COMPACT TOUCH® cleared in K094038. Similar to the previous version of the device, it consists of a base and probes. The Base performs the same calculations and displays as the predicate COMPACT TOUCH®. The A-scan probe for biometry and pachymeter is identical to the predicate while the B-scan probe has an increased transducer frequency of 15 Hz which allows for improved measurement accuracy of ± 0.115 mm.

    AI/ML Overview

    The provided text does not contain detailed acceptance criteria and a study proving the device meets those criteria in the typical format of a clinical trial or performance study report. Instead, it is a 510(k) summary for a medical device (COMPACT TOUCH® Ophthalmic Ultrasound System) seeking FDA clearance based on substantial equivalence to an existing predicate device (K094038).

    The document focuses on explaining that the new device has the same indications for use and similar technological characteristics to the predicate device, with some minor modifications. The "performance data" section primarily lists compliance with recognized consensus standards for electrical safety, electromagnetic compatibility, usability, and particular requirements for ultrasonic medical diagnostic equipment. It also mentions hardware and software validation activities.

    However, I can extract the relevant information regarding the performance improvement stated and the basis for equivalence as outlined in the document.

    Here's an attempt to structure the information based on your request, acknowledging the limitations of the provided text for a full "acceptance criteria and study results" breakdown.


    Device Name: COMPACT TOUCH® Ophthalmic Ultrasound System

    Predicate Device: Quantel Medical Compact Touch (K094038)


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state acceptance criteria in the form of specific thresholds (e.g., sensitivity, specificity, accuracy targets) for its diagnostic functions or biometric measurements, as would be typical for a clinical performance study. Instead, the "acceptance criteria" are implied by showing equivalence to the predicate and highlighting an improvement in measurement accuracy for one specific probe.

    Acceptance Criterion (Implied by Equivalence / Improvement)Reported Device Performance (New Device)
    B-Scan Probe Measurement AccuracyImproved to ± 0.115 mm
    (Predicate Device Performance)(± 0.2 mm)
    Safety and Effectiveness(Not explicitly quantified, but stated as unimpaired despite modifications)
    Compliance with Consensus StandardsDemonstrated compliance with listed IEC and AAMI ANSI standards
    Software ValidationPerformed, appropriate for Moderate level of concern
    Electrical SafetyComplies with electrical safety standards (degree of protection changed from BF to B)
    Electromagnetic Compatibility (EMC)Complies with EMC standards
    UsabilityComplies with usability standards
    Indications for UseSame as predicate device
    Technological CharacteristicsSimilar to predicate device (A-scan/pachymeter identical, B-scan frequency increased, other minor changes)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a sample size for a test set or test subjects for evaluating the improved measurement accuracy (± 0.115 mm). It states that the improvement is "due to the increase of the frequency of the probe" and that "IEC 60601-2-37 tests have been followed...". This suggests the performance claim for accuracy is likely derived from laboratory testing or phantom measurements, rather than a clinical study on a patient population.

    Data provenance (country of origin, retrospective/prospective) and details of the "test set" are not provided, as it seems no clinical human subject test set was used for the accuracy claim or overall equivalence beyond technical standard compliance.


    3. Number of Experts Used to Establish Ground Truth and Qualifications

    Since no clinical test set with human subjects is described for evaluating the stated performance improvement, there is no mention of experts being used to establish ground truth in such a context. The evaluation appears to be based on technical specifications and compliance with standards.


    4. Adjudication Method for the Test Set

    As no clinical test set for human subject performance validation is described, there is no adjudication method mentioned.


    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    There is no mention of a Multi-Reader Multi-Case (MRMC) comparative effectiveness study or any assessment of human readers improving with or without AI assistance. This device is an ultrasound system purely for diagnostic imaging and biometric measurement, not an AI-powered diagnostic interpretation tool.


    6. Standalone (Algorithm Only) Performance Study

    The document describes the device as an "Ophthalmic Ultrasound System" which performs imaging and measurements. It mentions "calculations" and "software revised to accommodate the other device changes". However, there is no standalone algorithm-only performance study described in the context of diagnostic interpretation or classification. The performance discussed relates to the physical measurement capabilities of the ultrasound system.


    7. Type of Ground Truth Used

    For the improved measurement accuracy of ± 0.115 mm, the ground truth is implicitly derived from technical specifications and calibration procedures of the ultrasound system, likely using phantoms or reference standards in a laboratory setting, rather than expert consensus, pathology, or outcomes data from human subjects. The primary "ground truth" for the overall device's safety and effectiveness in the context of this 510(k) is its substantial equivalence to the predicate device, which itself would have undergone its own validation.


    8. Sample Size for the Training Set

    This document pertains to a hardware device with associated software for measurement and display. It is not an AI/Machine Learning device that undergoes a "training" process with a large dataset. Therefore, there is no training set sample size mentioned.


    9. How the Ground Truth for the Training Set Was Established

    As it is not an AI/Machine Learning device, there is no training set and consequently no ground truth establishment process for a training set mentioned.

    Ask a Question

    Ask a specific question about this device

    K Number
    K170067

    Validate with FDA (Live)

    Device Name
    Easyret
    Manufacturer
    Date Cleared
    2017-07-25

    (197 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Easyret® is indicated for use in the treatment of ocular pathology of anterior and posterior segments including, retinal photocoagulation, pan retinal photocoagulation for vascular and structural abnormalities of the retina and uvea including:

    • Proliferative and severe nonproliferative diabetic retinopathy,
    • Choroidal neovascularization
    • Branch retinal vein occlusion
    • Treatment of choroidal neovascularization associated with wet age-related macular degeneration
    • Retinal tears and detachments,
    • Certain forms of macular edema
    • Retinopathy of Prematurity
    • Iridotomy in angle closure glaucoma, and trabeculoplasty in open angle glaucoma
    Device Description

    Easyret® is an integrated photocoagulation laser system including a slit lamp and a single column table. It provides a yellow laser beam (577nm).

    Easyret® laser is available with the following configurations: - A single slit lamp delivery system: With this model two different slit lamps manufactured by CSO (CSO SL 9800 5x and CSO SL 9900 5x) are available. These slit lamps models have been cleared under K992836. - A slit lamp delivery system and a laser indirect ophthalmoscope (LIO) optional port In addition to the slit lamp delivery system, an optional LIO (laser indirect ophthalmoscope) can be used with the laser. Quantel Medical recommends the use of the following LIO models: Heine Omega 500 and Keeler Vantage +.

    The touch screen includes a user interface where laser treatment settings such as power, pulse duration, spot size treatment and aiming beam intensity level can be programmed.

    The user interface allows the physician to quickly select and access the following 3 treatment modes:

    • Monospot treatment: This selection allows the user to carry out thermal "traditional" laser treatment using single spot via 4 delivery modes: Single, Repeat, Painting or Continuous.
    • -MultiSpot treatment: The MultiSpot mode allows automatically consecutive several laser spots according a pattern selected and the treatment patterns are customizable: square, circle, triple arc and macular grid.
    • SubLiminal treatment: The sequencing (SubLiminal mode, or sub-threshold) is used to split the energy delivery into several successive very short pulses (microseconds) separated by a cooling time (interval).

    Feedback to the user, such as laser operating state, number of performed laser spots, etc. is provided through the user interface. The laser treatment can be initiated by pressing on a footswitch and aborted by releasing it.

    AI/ML Overview

    This document is a 510(k) premarket notification for a medical device called Easyret®, an ophthalmic laser system. It focuses on demonstrating the device's substantial equivalence to a legally marketed predicate device (Quantel Medical Supra Scan K120825). Therefore, the document does not contain information related to acceptance criteria, device performance details from a study proving it meets those criteria, sample sizes, expert qualifications, or comparative effectiveness studies in the way typically associated with clinical performance validation of diagnostic AI/ML devices.

    The document primarily focuses on:

    • Indications for Use: The Easyret® is indicated for use in the treatment of ocular pathology of anterior and posterior segments including, retinal photocoagulation, pan retinal photocoagulation for vascular and structural abnormalities of the retina and uvea, as listed in pages 2 and 4-5.
    • Device Description: Easyret® is an integrated photocoagulation laser system including a slit lamp and a single column table, providing a yellow laser beam (577nm). It supports Monospot, MultiSpot, and SubLiminal treatment modes (page 4).
    • Substantial Equivalence Argument: The core of the submission is to demonstrate that Easyret® is substantially equivalent to the predicate device (Quantel Medical Supra Scan). This is argued based on:
      • Same intended use/indications for use (page 5).
      • Similar technological characteristics: Both use a 577nm laser. Differences primarily relate to the integration of components (scanner and zoom integrated into the slit lamp in Easyret® vs. added to the slit lamp in the predicate) and the type of laser cavity (fiber laser cavity in Easyret® vs. Optically Pumped Semiconductor Laser (OPSL) cavity in the predicate) (page 5-6).
      • Similar performance parameters: While some ranges (exposure time, spot size) have been slightly adjusted, the document states that the delivered power, spot size, and exposure time can be adjusted in the same ranges as the predicate, resulting in the same range of fluence delivered (page 6).
    • Performance Data: This section only lists the recognized consensus standards to which performance testing was conducted (e.g., electrical safety, electromagnetic compatibility, usability, particular requirements for surgical laser equipment, laser product safety). It also states that hardware and software validation activities were performed and appropriate software documentation was provided. It does not provide specific performance results, acceptance criteria, or a study report demonstrating the device's clinical performance against acceptance criteria.

    Therefore, based on the provided document, I cannot answer the requested questions regarding acceptance criteria and performance data for a study proving the device meets those criteria, including sample sizes, ground truth establishment, or any comparative effectiveness studies, because this information is not present in the 510(k) summary.

    The document's purpose is a regulatory submission for substantial equivalence to a predicate laser device, not a detailed clinical performance study of a diagnostic or AI/ML device. It focuses on engineering and regulatory compliance rather than clinical trial results with performance metrics against predefined acceptance criteria for AI algorithms.

    Ask a Question

    Ask a specific question about this device

    K Number
    K140336

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-06-25

    (135 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    OPTIMIS FUSION YAG: photodisruption of ocular tissue using light energy emitted by a Nd: YAG Laser, including discission of posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy.
    OPTIMIS FUSION SLT: Selective Laser Trabeculoplasty
    OPTIMIS FUSION YAG/SLT: photodisruption of ocular tissue using light energy emitted by a Nd: YAG Laser, including discission of posterior capsule of the eye (posterior capsulotomy), and discission of pupillary membranes (pupillary membranectomy) in aphakic and pseudophakic patients, and iridotomy; and Selective Laser Trabeculoplasty

    Device Description

    The QUANTEL MEDICAL OPTIMIS FUSION is an ophthalmic surgical laser designed for performing photodisruption of ocular tissue using laser energy emitted by a Nd:YAG laser including discission of the posterior capsule of the eye (posterior capsulotomy), discission of the pupillary membranes (pupillary membranectomy), and iridotomy/iridectomy; and selective laser Trabeculoplasty.

    The Family of OPTIMIS FUSION Ophthalmic Lasers. Delivery Device and Accessories consists of the following models:

    1. OPTIMIS FUSION YAG a Nd: YAG Laser providing-switched laser pulses at a wavelength of 1064 nm for use in photodisruption of ocular tissue (posterior capsulotomy, pupillary membranectomy, iridotomy/iridectomy). The 1064 treatment beam delivers 4 ns, 0.3 -10mJ adjustable and selectable single, double or triple pulse of energy. It is conditioned trough beam shaping optics to generate a photodisruption micro pulse of plasma at a precision adjustable location relative to the visual focal plane (located at slit lamp center of rotation) and along the slit lamp objective lens axis. A twin aiming beam is also focused by the slit lamp objective to a converging 10 um spot located at the focal point of the lens. The focal point of photodisruption is adjustable 150um in the posterior direction and -150um in the anterior direction by the physician relative to this convergence of the twin aiming beams.
    2. OPTIMIS FUSION SLT a Nd: YAG Laser providing Q-switched frequency doubled pulses at a wavelength of 532 nm for use in Selective Laser Trabeculoplasty. The treatment beams delivers a 4 nsec. 0.3-2mJ adjustable single pulse of energy. The aiming and treatment beams are coaxial with each other and focussed by the slit lamp objective to a 400um spot at the focal point of the lens.
    3. OPTIMIS FUSION YAG/SLT a Nd: YAG Laser providing Q-switched laser pulses at a wavelength of 1064 nanometers for use in photodisruption or O-switched frequency doubled pulses at a wavelength of 532 nm for use in Selective Laser Trabeculoplasty. depending upon the mode selected. The OPTIMIS FUSION YAG/SLT contains two aiming beam modules that produce a single beam for the 532nm mode and a dual beam for the 1064nm mode, respectively.

    For each OPTIMIS FUSION model, the physician controls delivery of laser energy from the OPTIMIS FUSION control display unit and activates the treatment laser beam with a footswitch or joystick pushbutton. In addition, a laser slit lamp adaptor may be coupled to each of the above OPTIMIS FUSION models and connected to a currently cleared QUANTEL MEDICAL 532nm photocoagulator (VITRA (K04236), SUPRA (K07776), VITRA MULTISPOT (K122251)) to allow to the physician to use the Slit Lamp Adaptor to deliver 532 nm continuous wave laser energy for retinal photocoagulation.

    The OPTIMIS FUSION produces short, individual pulses of focused laser light with wavelengths of either 1064 nm or 532 nm. depending on the selected operational mode. Using a slit lamp microscope and aiming beam, the pulsed light is accurately targeted on a structure within the patient's eye.

    When the photodisruptor mode is selected, the treatment wavelength is 1064nm. A twinaiming beam targets the area of tissue disruption. The energy contained within a single short pulse is concentrated by focusing to a very small spot size so that plasma formation occurs at the focal point. This creates an acoustic wave which disrupts nearby tissue.

    When the SLT mode is selected, the treatment wavelength is 532nm. A coaxial aiming beam targets the trabecular meshwork via a contact lens. The SLT treatment laser provides a low energy, short pulse of laser light that produces a thermal effect in pigmented cells in the trabecular meshwork.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting study for the Quantel Medical OPTIMIS FUSION Ophthalmic Laser Systems:

    Disclaimer: This 510(k) summary focuses on demonstrating substantial equivalence to existing predicate devices, rather than proving novel clinical efficacy or accuracy against a defined ground truth in a clinical study. Therefore, many of the typical elements sought for evaluating AI/diagnostic device performance will not be explicitly present.


    Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary does not delineate specific, quantifiable "acceptance criteria" in the format of a diagnostic performance study (e.g., sensitivity, specificity, AUC thresholds). Instead, the acceptance criteria are implicitly tied to demonstrating substantial equivalence to predicate devices. This means the device is deemed acceptable if its technological characteristics and intended use are similar enough to previously cleared devices that it raises no new questions of safety or effectiveness.

    The "device performance" reported is also in terms of technological characteristics and similarity to predicate devices, rather than clinical outcome or diagnostic accuracy metrics.

    Feature / CriterionOPTIMIS FUSION Performance (vs. Predicate)
    Indications for UseIdentical to Lumenis Family of Selecta Ophthalmic Laser Systems. YAG model indications are similar to Carl Zeiss Meditec VISULAS YAG III (Optimis Fusion YAG also indicates pupillary membranectomy). SLT model indications are identical to Quantel Medical Solutis.
    Technological Characteristics (General)Same basic design: Ophthalmic surgical laser for photodisruption (YAG) and Selective Laser Trabeculoplasty (SLT). Emits Nd:YAG laser energy. Multiple configurations (SLT, YAG, YAG/SLT combined).
    SLT Model (OPTIMIS FUSION SLT)- Laser Type: Q-Switched frequency doubled Nd: YAG laser - Wavelength: 532 nm (Same as Selecta SLT, Solutis) - Pulse Mode: Single pulse (Same as Selecta SLT, Solutis) - Energy Range: 0.3-2.0 mJ (Same as Selecta SLT; Solutis is 0.2-2.0 mJ - considered minor difference) - Spot Size in Air: 400 microns (Same as Selecta SLT, Solutis) - Focus Angle: < 3 degrees (Same as Selecta SLT, Solutis) - Max Energy Density at Min Exposure: 1.6 J/cm² (Same as Selecta SLT, Solutis) - Repetition Rate: 2.0 Hz (Same as Selecta SLT, Solutis) - Pulse Duration: 4 ns (Same as Solutis; Selecta SLT is 3 ns - considered minor difference) - Max Average Power: 4 mW (Same as Selecta SLT, Solutis)
    YAG Model (OPTIMIS FUSION YAG)- Laser Type: Q-Switched Nd: YAG laser - Wavelength: 1064 nm (Same as Selecta 1064, VISULAS YAG III) - Pulse Mode: Pulsed with single spot, user selectable bursts of 1, 2, or 3 pulses (Same as Selecta 1064, VISULAS YAG III) - Energy Range: 0.3-10.0 mJ (Same as Selecta 1064, VISULAS YAG III) - Spot Size in Air: 10 microns (Same as Selecta 1064, VISULAS YAG III) - Focus Angle: 16 degrees (Same as Selecta 1064, VISULAS YAG III)
    YAG/SLT Model (OPTIMIS FUSION YAG/SLT)Combines features/specs of both OPTIMIS FUSION SLT and OPTIMIS FUSION YAG systems, with unchanged individual specs, similar to Lumenis Selecta Duet.
    Aiming System635 nm red diode (Same as predicate systems).
    Delivery SystemSlit lamp (Same as predicate systems). Uses CSO SL980 slit lamp (cleared in K992836).
    Validation/Verification (General Statement)Laboratory testing was conducted to validate and verify that the Family of OPTIMIS FUSION Ophthalmic Laser Systems, Delivery Device and Accessories met all design specifications and was substantially equivalent to the predicate devices.

    Study Details (Based on the Provided 510(k) Summary)

    The provided document describes a 510(k) Premarket Notification submission, which is primarily focused on demonstrating "substantial equivalence" to legally marketed predicate devices, not on a de novo clinical study to prove efficacy or accuracy. As such, the study details you've requested regarding AI/diagnostic device performance are largely not applicable (N/A) or not explicitly detailed in this type of submission.

    Here's what can be inferred or stated as N/A:

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
      • N/A. This submission relies on "laboratory testing" to verify design specifications and demonstrate substantial equivalence comparing technological characteristics. It does not describe a clinical "test set" of patient data for evaluating diagnostic performance.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
      • N/A. As there is no clinical "test set" with a diagnostic ground truth, expert consensus for ground truth establishment is not relevant to this submission.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
      • N/A. No clinical test set.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      • No. This is a medical device (laser system) without an AI component for interpretation or diagnosis. Therefore, an MRMC study comparing human readers with and without AI assistance is not applicable.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
      • N/A. This is a physical laser system, not an algorithm, so "standalone algorithm performance" is not relevant. The device operates under direct physician control ("human-in-the-loop" in the sense that a human physically operates it).
    6. The type of ground truth used (expert concensus, pathology, outcomes data, etc):
      • N/A. The "ground truth" for this type of submission is the performance and safety established for the predicate devices. The new device demonstrates "ground truth" of its own performance through "laboratory testing" measured against its own design specifications, and the comparison of these specifications to the predicates.
    7. The sample size for the training set:
      • N/A. This device does not use machine learning, therefore there is no "training set" in the AI sense.
    8. How the ground truth for the training set was established:
      • N/A. No training set for an AI algorithm.

    Summary of the Study Conducted (Based on the 510(k) Text):

    The "study" conducted for the Quantel Medical OPTIMIS FUSION Ophthalmic Laser Systems, as described in the 510(k) summary, consists of:

    • Laboratory Testing: This was performed to validate and verify that the device met its own design specifications (e.g., wavelength, energy range, spot size, pulse duration, repetition rate, maximum average power).
    • Substantial Equivalence Analysis: A comparative analysis of these design specifications and the Indications for Use against legally marketed predicate devices (Lumenis Selecta Family, Carl Zeiss Meditec VISULAS YAG III, Quantel Medical Solutis). The primary goal was to demonstrate that the new device's technological characteristics and intended use are sufficiently similar to the predicates, and that any differences do not raise new questions of safety or effectiveness. The conclusion was that differences were "extremely minor in significance and don't affect the safety or efficacy of the devices."
    Ask a Question

    Ask a specific question about this device

    Page 1 of 4