Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K243795

    Validate with FDA (Live)

    Date Cleared
    2025-09-05

    (269 days)

    Product Code
    Regulation Number
    870.4450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The pREBOA-PRO Catheter is intended for temporary partial or complete occlusion of large vessels and blood pressure monitoring including patients requiring emergency control of hemorrhage.

    Device Description

    The pREBOA-PRO™ Catheter is a large vessel occlusion catheter with a dedicated lumen for pressure monitoring. The device consists of a semi-compliant balloon, in-line pressure relief valve (safety valve), atraumatic distal tip (P-tip™), dual lumen catheter shaft and a hub with extension lines to provide access to each lumen.

    The pREBOA-PRO™ Catheter is designed with flow channels to enable precise, smooth control of inflation and deflation of the balloon. Instructions for how to titrate the balloon for partial occlusion, if so desired by the clinician, are included in the Instructions for Use.

    The device has an effective length of 72 cm and is compatible with 7 Fr introducer sheaths as shown in the compatibility section of the IFU. The catheter has a unibody design and is intended to be placed and advanced without a guidewire. In addition, the catheter is compatible with guidewires up to 0.025" that may be used to facilitate subsequent vascular procedures after a REBOA procedure. An orange peel-away sheath is preloaded on the catheter shaft covering the balloon to ease insertion of the catheter's P-tip™ into an introducer sheath hemostasis valve. A catheter shaft provides appropriate stiffness with a distal tip (P-tip™) designed for atraumatic advancement of the catheter in a blood vessel. Pad printed marks on both sides of the outer catheter shaft indicate distance to the center of the balloon as well as average insertion depth ranges to the center of aortic Zones 1 and 3 to facilitate proper placement. Radiopaque platinum iridium marker bands are located at the functional ends of the balloon to facilitate and visualize accurate balloon placement when used with imaging. The proximal end of the catheter has a hub and extension lines. The arterial line lumen is used to monitor blood pressure above the balloon. Below the balloon core aortic pressure monitoring is possible through the side arm of the sheath. The balloon lumen is used to inflate and deflate the balloon as needed. An in-line safety valve is connected to the balloon lumen stopcock and is designed to help prevent over-inflation of the balloon. The stopcocks provide control to each of the catheter's two lumens. The orange peel-away sheath can be separated from the catheter shaft after insertion if needed. The device is a single-use, sterile device.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K200459

    Validate with FDA (Live)

    Date Cleared
    2020-06-03

    (99 days)

    Product Code
    Regulation Number
    870.4450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The pREBOA-PRO™ Catheter is intended for temporary occlusion of large vessels and blood pressure monitoring including patients requiring emergency control of hemorrhage.

    Device Description

    The pREBOA-PRO™ Catheter is a large vessel occlusion catheter with a dedicated lumen for pressure monitoring. The device consists of a semi-compliant balloon, in-line pressure relief valve (safety valve), atraumatic distal tip (P-tip™), dual lumen catheter shaft and a hub with extension lines to provide access to each lumen. The pREBOA-PRO™ Catheter balloon is designed with flow channels to enable the clinician precise, smooth control during inflation and deflation. Instructions for how to titrate the balloon for partial occlusion, if so desired by the clinician, are included in the Instructions for Use. The device has an effective length of 72 cm and is compatible with 7 Fr introducer sheaths as shown in the compatibility section of the IFU. The catheter has a unibody design and is intended to be placed and advanced without a quidewire. In addition, the catheter is compatible with guidewires up to 0.025" that can be used to facilitate subsequent vascular procedures affer a REBOA procedure. A peel-away sheath is preloaded on the catheter shaft covering the balloon to ease insertion of the catheter's P-tip™ into an introducer sheath hemostasis valve. A catheter shaft provides appropriate stiffness with a distal tip (P-tip™) designed for atraumatic advancement of the catheter in a blood vessel. Pad printed marks on both sides of the outer catheter shaft indicate distance to the center of the balloon as well as average insertion depth ranges to the center of aortic Zones 1 and 3 to facilitate proper placement. Radiopaque platinum iridium marker bands are located at the functional ends of the balloon to facilitate and visualize accurate balloon placement when used with imaging. The proximal end of the catheter has a hub and extension lines. The arterial line lumen is used to monitor blood pressure above the balloon. The balloon lumen is used to inflate and deflate the balloon as needed. An in-line Pressure Relief (safety) Valve is connected to the balloon lumen stopcock and is designed to help prevent over-inflation of the balloon. The stopcocks provide control to each of the catheter's two lumens. The peel-away sheath can be separated from the catheter shaft affer insertion if needed. The device is a single-use, sterile device.

    AI/ML Overview

    The pREBOA-PRO Catheter is a medical device intended for temporary occlusion of large vessels and blood pressure monitoring, particularly in patients requiring emergency control of hemorrhage. The provided document is a 510(k) summary, which indicates that the device is seeking substantial equivalence to a legally marketed predicate device (ER-REBOA™ Catheter, K172790).

    The document details performance data including various bench and in vivo animal tests to demonstrate the safety and effectiveness of the modifications made to the pREBOA-PRO Catheter compared to its predicate. Additionally, a retrospective clinical study using the predicate device is summarized to support the concept of partial REBOA, which the pREBOA-PRO Catheter is designed to facilitate with improved control.

    Here's an analysis of the acceptance criteria and study information provided:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't explicitly present a table of acceptance criteria with corresponding performance values for the pREBOA-PRO Catheter in a pass/fail format. Instead, it lists various "in vitro bench tests" and "in vivo animal tests" that were performed to demonstrate that the device "meets applicable design and performance requirements."

    However, based on the types of tests mentioned, we can infer some of the performance aspects that were evaluated. For a medical device submission seeking substantial equivalence, "acceptance criteria" are often implicitly linked to demonstrating performance comparable to the predicate device and meeting recognized standards.

    Here's a generalized table reflecting the types of performance assessed, acknowledging that specific numerical criteria are not provided in this summary:

    Performance AspectReported Device Performance (Implied)
    Mechanical Integrity & Durability
    Balloon Burst / Freedom from fragmentationMeets requirements
    Fatigue (inflation/deflation cycle)Meets requirements
    Tensile StrengthMeets requirements
    Kink DiameterMeets requirements
    Corrosion ResistanceMeets requirements
    Functional Performance
    Balloon Inflation/Deflation TimeMeets requirements
    Occlusion TimeMeets requirements
    Peel-away Sheath PeelabilityMeets requirements
    Compatibility (Guidewire, Sheath, Luer)Compatible as specified
    TorqueMeets requirements
    Dimensional AccuracyMeets requirements
    Balloon Occlusion Volume to Inflation Dia.Meets requirements
    Freedom from LeakageMeets requirements
    Printed Mark LegibilityMeets requirements
    Pressure Relief Valve (Max. pressure, Dislodgement)Meets specifications and functions
    Pressure ResponseAccurate pressure monitoring
    Simulated Use (ex-vivo porcine, over-inflation, burst)Meets safety and performance requirements
    Repeatability & Reliability of Flow Char.Consistent and controlled flow
    Balloon Working Length CharacterizationCharacterized as designed
    Balloon Internal Pressure AnalysisSupports PRV specifications
    Aortic Depth / Diameter Range AnalysisSupports zone depth markings and occlusion capabilities
    BiocompatibilityMeets ISO 10993-1 and FDA Guidance
    SterilizationAchieves SAL of 1 x 10^-6
    In Vivo Performance (Animal Studies)
    Performance Evaluation (porcine aorta)Safe and effective in animal model
    Thrombogenicity Study (ovine aorta)Safe (low thrombogenicity)

    It's crucial to note that the specific numerical targets or thresholds for "meets requirements" are not detailed in this summary. These would typically be found in the full submission to the FDA.

    2. Sample Size Used for the Test Set and the Data Provenance:

    The document describes two main categories of studies:

    • Bench and Animal Tests (for pREBOA-PRO Catheter itself):

      • Sample Size: Not explicitly stated for individual bench tests (e.g., how many catheters were tested for burst strength, fatigue, etc.).
      • Data Provenance: Not explicitly stated for the bench tests. For the in vivo animal tests, they were conducted using porcine and ovine models. These are prospective animal studies designed to evaluate the new device.
    • Clinical Study (Retrospective, using the predicate ER-REBOA™ Catheter):

      • Sample Size: 46 patients were included in the retrospective clinical study.
      • Provenance: This was a retrospective analysis of data collected from a local prospectively collected database at a major US trauma center. The data was collected between January 2016 and May 2019. The information indicates the data is from the United States.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts:

    This information is not provided in the document.

    For the retrospective clinical study, the "ground truth" would generally be established by clinical outcomes and diagnoses within the patient medical records. The study refers to "electronic data and local REBOA registry", implying that medical professionals (e.g., trauma surgeons, intensivists) made the clinical decisions and recorded outcomes. However, the exact number and qualifications of experts involved in data abstraction, adjudication, or "ground truth" determination for the study are not specified.

    For the animal studies, the "ground truth" is measured physiological responses and observations, typically assessed by veterinarians and research personnel. Their specific qualifications are not mentioned.

    4. Adjudication Method for the Test Set:

    This information is not provided in the document.

    For the retrospective clinical study, it's unclear if any formal adjudication process was used to resolve discrepancies in patient data or outcomes. For bench and animal studies, "adjudication" in the sense of expert consensus on outcomes is less common; results are typically based on objective measurements and observations by the study team.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device (pREBOA-PRO Catheter) is a physical medical device, not an AI or imaging diagnostic tool that would typically involve human "readers" interpreting cases with or without AI assistance. Therefore, the concept of "effect size of how much human readers improve with AI" is not applicable here.

    The document discusses the clinical effectiveness of a technique (REBOA, specifically partial REBOA) and the device's role in facilitating that technique, but not cognitive assistance.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    No, this refers to a physical medical device, not a standalone algorithm. The "performance data" section details the device's mechanical, functional, and biological performance characteristics, evaluated through bench and animal testing. These are assessments of the device itself, akin to "standalone" performance for a physical product.

    7. The Type of Ground Truth Used:

    • For Bench Tests: The "ground truth" for bench tests is based on engineering specifications, recognized standards, and objective measurements of the device's physical and functional properties. For example, "balloon burst" would have an objective standard for acceptable pressure.
    • For Animal Studies: The "ground truth" is based on physiological measurements (e.g., blood pressure, blood flow), observed performance (e.g., occlusion success, ease of deployment), and pathological findings from the animal models. This is considered outcomes data from an experimental setting.
    • For Retrospective Clinical Study (using predicate device): The "ground truth" was derived from patient medical records, clinical outcomes data (e.g., mortality, organ failure, ventilator days, dialysis, vasopressor requirements), and the local REBOA registry. This is essentially retrospective clinical outcomes data.

    8. The Sample Size for the Training Set:

    This information is not applicable in the conventional sense for this type of device submission.

    • The pREBOA-PRO Catheter is a physical medical device. It does not employ machine learning algorithms that require a "training set" of data.
    • The "Performance Data" describes validation and verification testing, not a training process.
    • The "Real World Evidence" clinical study using the predicate device is an observational study of clinical outcomes, not a dataset used for "training" an algorithm.

    9. How the Ground Truth for the Training Set was Established:

    This information is not applicable as there is no "training set" in the context of an AI algorithm described for this device. The "ground truth" for the device's design and testing (as described in point 7) was established through engineering principles, adherence to standards, and direct measurement in laboratory and animal settings, and for the retrospective study, through recorded clinical outcomes.

    Ask a Question

    Ask a specific question about this device

    K Number
    K172790

    Validate with FDA (Live)

    Date Cleared
    2017-11-08

    (54 days)

    Product Code
    Regulation Number
    870.4450
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ER-REBOA Catheter is intended for temporary occlusion of large vessels and blood pressure monitoring including patients requiring emergency control of hemorrhage.

    Device Description

    The ER-REBOA™ Catheter is a large vessel occlusion catheter with a dedicated lumen for pressure monitoring. The device consists of a compliant occlusion with an atraumatic distal tip (P-tip®), a dual lumen catheter shaft and a hub with extension lines to provide access to each lumen. The balloon lumen is used to inflate and deflate the balloon. The arterial line lumen is used to monitor blood pressure. The cather has a unibody design and is designed to be used without a guidewire. A peel away sheath is pre-loaded on the catheter shaft to ease insertion of the catheter's P-tip® into an introducer sheath hemostasis valve. The device has an effective length of 72 cm and is compatible with 7 Fr or larger introducer sheaths. The device is a single use sterile device. The distal tip (P-tip®) eases advancement of the catheter in a blood vessel. The compliant occlusion balloon is capable of occluding vessels up to 32 mm in diameter. Radiopaque platinum iridium marker bands are located at the functional ends of the balloon to facilitate accurate balloon placement. A co-axial catheter shaft provides appropriate stiffness and a dedicated lumen for pressure monitoring distal to the balloon. Pad printed marks on the outer catheter shaft indicate distance to the balloon to facilitate proper placement. The proximal end of the catheter has a hub and extension lines. The stopcocks provide control to each of the catheter's two lumens. The peel-away sheath can be separated from the catheter shaft after insertion if needed.

    AI/ML Overview

    Here's an analysis of the provided text regarding the ER-REBOA Catheter (K172790), focusing on acceptance criteria and supporting study details:

    Summary of Acceptance Criteria and Device Performance (as inferable from the document):

    The document does not explicitly state quantitative acceptance criteria in terms of specific performance metrics for the device itself (e.g., burst pressure, inflation time, etc.). However, the primary "acceptance criteria" discussed relate to equivalence to a predicate device and the justification for removing a pregnancy contraindication.

    The performance data section focuses on demonstrating the safety and effectiveness of using the REBOA catheter in pregnant patients, which implicitly means the device performs acceptably in this population.

    Acceptance Criteria CategoryReported Device Performance / Justification
    Equivalence to Predicate DeviceThe ER-REBOA™ Catheter is identical to the predicate device (ER-REBOA™ Catheter, K170411) in terms of technological characteristics. Conclusion: Substantially Equivalent.
    Safety and Effectiveness in Pregnant Patients (justification for removing contraindication)1. Clinical Rationale: Radiation risk of fluoroscopy (often a concern for pregnant patients during interventional procedures) is not applicable to the ER-REBOA™ Catheter. 2. Clinical Study: A retrospective case series study of 36 patients treated with fluoroscopy-free REBOA for severe postpartum hemorrhage. - REBOA success rate: 100% - REBOA-related mortality: 0%

    Detailed Analysis of the Study and Ground Truth:

    The document describes a single study related to the change in labeling regarding pregnancy.

    1. Sample Size Used for the Test Set and Data Provenance:

      • Sample Size: 36 patients
      • Data Provenance: Retrospective case series study. The country of origin is not specified.
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications:

      • Not specified. The study is a "case series," which implies clinical outcomes were observed and recorded by treating physicians, but the process of establishing a formal "ground truth" by independent experts for the purpose of the submission is not detailed. The "REBOA success rate" and "REBOA related complications" likely refer to the clinical assessment of the treating medical teams.
    3. Adjudication Method for the Test Set:

      • Not specified. Given it's a retrospective case series, a formal adjudication method by a separate panel for the purpose of this submission is unlikely to have been carried out or documented in this summary.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

      • No. This type of study (MRMC) is typically used for diagnostic or imaging-based devices where different readers interpret data. The ER-REBOA Catheter is an interventional therapeutic device. The study described focuses on clinical outcomes in a specific patient population.
    5. If a Standalone Study Was Done:

      • Yes, implicitly. The retrospective case series assessed the outcomes of the REBOA procedure using the device in a specific clinical context (postpartum hemorrhage) without comparing it directly to human readers or other diagnostic methods. It evaluated the device's performance in a clinical setting.
    6. The Type of Ground Truth Used:

      • Clinical Outcomes/Expert Judgment (Implicit): The "REBOA success rate" and "no patients died from REBOA related complications" are based on clinical assessments and outcomes reported by the medical professionals involved in the care of those 36 patients. This is essentially ground truth established by the treating clinicians through observation of patient response and medical records.
    7. The Sample Size for the Training Set:

      • Not applicable. The described study is a case series for proving clinical benefit/safety in a specific population for labeling purposes, not a study to train a machine learning algorithm. The ER-REBOA Catheter is a physical medical device, not an AI/ML algorithm.
    8. How the Ground Truth for the Training Set Was Established:

      • Not applicable, as there is no training set for an AI/ML algorithm.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1