Found 7 results

510(k) Data Aggregation

    K Number
    K233273

    Validate with FDA (Live)

    Date Cleared
    2024-11-27

    (425 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The C&B Ceramic material is used for fabrications such as inlays, onlays, onlays, veneers and full crown restorations.

    The C&B Permanent material is used for fabrications such as inlays, onlays, onlays, veneers and full crown restorations

    C&B Interim material is indicated for the fabrication of temporary dental restorations in conjunction with extra-oral lightcuring equipment.

    Device Description

    Prevest C&B, Ceramic Resin is a light-curing material with ceramic fillers used for the 3D printing/ fabrication of Ceramic Crown & Bridge dental prosthesis for use in DLP, LCD/MSLA 3D printers. It is reactive to wavelength of light between 385nm and 405nm. Recommended printers and light curing unit are Asiga Max UV, Ackuretta SOL and Acuretta Curie. It is stored in 500 and 1000ml HDPE bottles and is available in several shades: A1 /A2 /A3 /A3.5 /B1 /B2 /B3 /C2 /D3 /Bleach/Extra Bleach based on the shade guide.

    Prevest C&B, Interim resin is a light-curing material with glass fillers used for the 3D printing/ fabrication of long term temporary Crown & Bridge dental prosthesis for use in DLP, LCD/MSLA 3D printers. It is reactive to wavelength of light between 385nm and 405nm. Recommended printers and light curing unit are Asiga Max UV, Ackuretta SOL and Acuretta Curie. It is stored in 500 and 1000ml HDPE bottles and is available in several Shades: A1 /A2 /A3 /A3.5 /B1 /B2 /B3/ C2 /D3/Bleach/Extra Bleach based on the shade guide.

    Prevest C&B, Permanent resin is a light-curing material with glass fillers for the 3D printing/fabrication of Permanent crown and bride dental prosthesis for use in DLP, LCD/MSLA 3D printers. It is reactive to wavelength of light between 385nm and 405nm. Recommended printers and light curing unit are Asiga Max UV, Ackuretta SOL and Acuretta Curie. It is stored in 500 and 1000ml HDPE bottles and is available in several Shades: A1 / A2 / A3 / A3.5 / B1 / B2 / B3 / C2 / D3/Bleach/Extra Bleach based on the shade guide.

    These materials are used for the additive manufacturing of crown and bridge prosthesis following the workflow as depicted below:

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for three dental resin materials: C&B Ceramic, C&B Permanent, and C&B Interim. The document states that these devices are substantially equivalent to predicate devices based on various testing results.

    Here's the breakdown of the acceptance criteria and the study information as requested:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document directly compares the subject devices (C&B Ceramic, C&B Permanent, C&B Interim) with predicate/reference devices against established ISO standards for certain physical parameters.

    Performance MetricAcceptance Criteria (from ISO Standard)C&B Ceramic PerformanceC&B Permanent PerformanceC&B Interim Performance
    Flexural StrengthISO 4049 ≥ 100 MPa ISO 10477 ≥ 50 MPa> 100 MPa> 100 MPa> 100 MPa
    Water SorptionISO 4049 ≤ 40 µg/mm³ ISO 10477 ≤ 40 µg/mm³< 40 µg/mm³< 40 µg/mm³< 40 µg/mm³
    Water SolubilityISO 4049 ≤ 7.5 µg/mm³ ISO 10477 ≤ 7.5 µg/mm³< 7.5 µg/mm³< 7.5 µg/mm³< 7.5 µg/mm³

    Note: The document also mentions testing for "appearance, sensitivity to ambient light, color stability, shade stability," and "biocompatibility assessment according to ISO 10993-1," but does not provide specific numerical acceptance criteria or performance results for these tests beyond stating "All test results met the criteria in standard."

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). It only mentions that "All test results met the criteria in standard" after testing protocols based on ISO 4049 and ISO 10477. These standards would typically define the required sample sizes for testing.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The tests performed are physical and chemical property tests against ISO standards, not evaluations that require human expert judgment for "ground truth" in the typical sense of clinical or diagnostic studies.

    4. Adjudication Method (e.g., 2+1, 3+1, none) for the Test Set:

    An adjudication method is not applicable or mentioned in this context. The testing involves laboratory measurements of material properties against predefined international standards.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    An MRMC comparative effectiveness study was not performed or mentioned. This type of study is relevant for diagnostic medical devices where multiple human readers interpret cases with and without AI assistance. The devices described are dental resin materials, not diagnostic imaging or AI devices.

    6. Standalone (i.e., algorithm only without human-in-the-loop performance) Study:

    A standalone performance study is not applicable as these are physical materials, not algorithms. The "testing summary" describes the evaluation of the materials' physical and chemical properties.

    7. Type of Ground Truth Used:

    The ground truth used for performance evaluation is based on established international standards (ISO 4049 and ISO 10477) for physical and chemical properties of dental restorative materials. These standards define the acceptable ranges for metrics like flexural strength, water sorption, and water solubility. Additionally, a biocompatibility assessment according to ISO 10993-1 was conducted.

    8. Sample Size for the Training Set:

    The concept of a "training set" is not applicable to this submission. These are physical materials, not machine learning algorithms that require training data.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no training set for these physical materials, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231263

    Validate with FDA (Live)

    Date Cleared
    2024-01-05

    (248 days)

    Product Code
    Regulation Number
    872.3690
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fusion Core DC Flo is indicated for use as core build up material on vital teeth, sealing root canal openings, cavity floor liner, post cementation, restorative material, and as a dentin replacement material.

    Fusion Flo is indicated for direct filling of class V (cervical caries, root erosions, and wedge shaped defects), anterior restorations( Class III and IV), small posterior restorations, repair of composite/ceramic veneers, and blocking out of undercuts.

    Fusion Flo SE is indicated for direct filling of class V (cervical caries, root erosion, wedge shaped defects), anterior restorations (Class III and IV), small posterior restorations,

    Fusion I-Seal is indicated for Class I and Class II restorations, as a liner under composite restorations, and as a pit and fissure sealant.

    Magma NT is indicated for direct restoration in anterior and posterior teeth classes I, III, IV and V, and porcelain/ composite repairs.

    Device Description

    Fusion Core DC Flo is dual-curing, radiopaque, flowable restoration for core build-up composite supplied in an auto-mix delivery system.

    Fusion Flo is a flowable composite of medium viscosity used for fillings in anterior teeth and small fillings in posterior teeth. Fusion Flo can be used separately or with Magma-NT composites.

    Fusion Flo SE is a self-etch, self-adhesive flowable composite. The flowable composite is ready for use and doesn't require the preparatory steps of etching, priming and bonding. Fusion Flo SE can be used separately or with Magma-NT composites.

    Fusion I-Seal is a single component, light cured flowable composite.

    Magma-NT is a visible light cured, radio opaque highly filled hybrid composite restorative material for anterior and posterior restoration of all teeth.

    AI/ML Overview

    The provided text describes the acceptance criteria and performance of several dental composite resins from Prevest Denpro Limited (Fusion Core DC Flo, Fusion Flo SE, Fusion I-Seal, and Magna NT) as part of a 510(k) premarket notification to the FDA. The study aims to demonstrate substantial equivalence to legally marketed predicate devices.

    Here's an analysis of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are generally based on meeting the requirements of ISO 4049:2019 and ISO 13116:2014, with specific values mentioned for some parameters. The reported device performance indicates that "All test results met the criteria in standards."

    Device PropertyAcceptance Criteria (Standard / Value)Reported Device Performance
    Fusion Core DC FloBased on ISO 4049:2019 & ISO 13116:2014Met criteria in standards. Specific values for Flexural Strength (≥50 MPa), Radio Opacity (3mm compared with aluminium wedge), Water Sorption (maximum of 40 µg/mm³), Water Solubility (maximum of 7.5 µm/mm³), Working Time (1:25), Setting Time (3:20).
    Fusion FloBased on ISO 4049:2019 & ISO 13116:2014Met criteria in standards. Specific values for Flexural Strength (≥ 80 MPa), Depth of Cure (1.00 mm opaque / 1.5 mm others minimum), Radio Opacity (3mm compared with aluminium wedge), Water Sorption (≤40 µg/mm³), Water Solubility (maximum of 7.5 µm/mm³).
    Fusion Flo SEBased on ISO 4049:2019 & ISO 13116:2014, USP 971 (for pH)Met criteria in standards. Specific values for Flexural Strength (≥ 80 MPa), Depth of Cure (1.00 mm opaque / 1.5 mm others minimum), Radio Opacity (3mm compared with aluminium wedge), Water Sorption (<40 µg/mm³), Water Solubility (maximum of 7.5 µm/mm³).
    Fusion I-SealBased on ISO 4049:2019 & ISO 13116:2014Met criteria in standards. Specific values for Flexural Strength (≥ 25 MPa), Depth of Cure (1.5 mm minimum), Radio Opacity (3mm compared with aluminium wedge), Water Sorption (≤40 µg/mm³), Water Solubility (maximum of 7.5 µm/mm³).
    Magma NTBased on ISO 4049:2019 & ISO 13116:2014Met criteria in standards. Specific values for Flexural Strength (≥ 80 MPa), Depth of Cure (2.59 mm opaque / 2.89 mm non-opaque), Radio Opacity (3mm compared with aluminum wedge), Water Sorption (≤40 µg/mm³), Water Solubility (maximum of 7.5 µm/mm³).
    Shelf Life (all composites)Testing protocols based on ISO 4049:2019Shelf life of 3 years (except Fusion Core DC Flo: 2 years). Determinations use the same testing protocols and met the criteria.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not explicitly state the sample sizes used for the test sets for each specific test (e.g., flexural strength, water sorption). It also does not specify the country of origin of the data or whether the studies were retrospective or prospective. The manufacturer, Prevest Denpro Limited, is based in India, so it's possible the testing was conducted there, but this is not confirmed.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This information is not applicable to this type of device and study. The "ground truth" for dental composite resins is established through objective physical and chemical testing against international standards (ISO standards), rather than expert interpretation of images or clinical outcomes that require human assessment.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    Not applicable. As noted above, the "truth" is determined by objective laboratory measurements against predefined standards.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This study concerns the physical and chemical properties of dental composite resins, not an AI-powered diagnostic device requiring evaluation of human reader performance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The ground truth used for evaluating these dental composite resins is defined by international standard specifications for physical and chemical properties as outlined in ISO 4049:2019 (Dentistry - Polymer-based restorative materials) and ISO 13116:2014 (Dentistry - Dental impression materials - Radiopacity). For Fusion Flo SE, USP 971 was also referenced for pH testing. The "truth" is whether the measured properties of the device meet the quantitative limits set by these standards.

    8. The sample size for the training set

    Not applicable. This is not a machine learning or AI device that requires a training set.

    9. How the ground truth for the training set was established

    Not applicable. As this is not an AI device, there is no training set or ground truth established in that context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K231696

    Validate with FDA (Live)

    Date Cleared
    2023-08-11

    (60 days)

    Product Code
    Regulation Number
    872.3200
    Panel
    Dental
    Age Range
    All
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Fusion Bond 5 is indicated for direct composite or compomer restorations, adhesive cementation, and composite repair.

    Fusion Bond 7 is indicated for bonding of composites to tooth structure, core build up, and adhesive cementation of crown & bridges, including inlays and onlays.

    Fusion Bond DC is indicated for direct light cure composite or compomer restorations, core build up, and adhesive cementation of crown & bridges, including inlays and onlays.

    Renew MDP is indicated for bonding of dual cure, light cure or self cure composite or compomer restorations to tooth structure. treatment of hypersensitive teeth, and intraoral repairs of fractured restorations.

    Renew Universal is indicated for direct bonding of light-cured compomers to tooth structure, bonding of dual-cured core build up composites to tooth structure as long as these materials are light-cured. intraoral repair of composite, metal-based and zirconia /alumina-based restorations, intraoral repair of ceramic restorations in combination with a silane coupling agent, treatment of hypersensitive teeth, and cavity sealing as a pretreatment for indirect restorations.

    Device Description

    Prevest Denpro Tooth Bonding Agents attach different types of restorations to teeth.

    Fusion Bond 5 is a light curing single component bonding agent. It primes and bonds in one step to attach restorations of composites, copomers, amalgams, and porcelains to tooth structures. Tooth structures should have a 37% phosphoric acid etchant application of Fusion Bond 5.

    Fusion Bond 7 is a light curing single component bonding agent. It etches, primes, and bonds in one step to attach restorations of composites, copomers, amalgams, and porcelains to tooth structures.

    Fusion Bond DC is a dual cure bonding direct and indirect restorations. Enamel and dentin surfaces should be conditioned with 37% phosphoric acid etchant before application of Fusion Bond DC. It is a two component adhesive with bonding agent and activator.

    Renew MDP is a light curing single component bondines etching, priming and bonding in one bottle. It is an ethanol-water based dental adhesive containing 10-Methacryloyloxydececyl Dihydrogen Phosphate (10-MDP) a functional monomer which helps in bonding to dentin and cut and un-cut enamel. Renew MDP works with light-cured and dual-cured composite materials.

    Renew Universal is a light curing single component bondines etching, priming and bonding. Renew Universal contains a combination of functional monomers such as 10-MDP and 4-META. Renew Universal works with light-cured and dual-cured composite materials.

    AI/ML Overview

    This document describes a 510(k) premarket notification for a series of dental bonding agents manufactured by Prevest Denpro Limited. The purpose of the 510(k) is to demonstrate substantial equivalence to legally marketed predicate devices, not to prove clinical efficacy through a comparative effectiveness study with human readers and AI assistance. Therefore, many of the requested criteria often associated with AI/ML medical devices are not applicable to this submission.

    Based on the provided text, here's a breakdown of the acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance:

    The acceptance criteria are established by the ISO 29022:2013 standard, specifically for "Dentistry - Adhesive - Notched-edge shear bond strength test." The reported device performance is stated as "Complies with ISO 29022" for both shear bond strength and curing time, with specific values provided as well.

    Device NameAcceptance Criteria (from ISO 29022:2013)Reported Device Performance (Shear Bond Strength)Reported Device Performance (Curing Time)
    Fusion Bond 5Complies with ISO 29022 (Shear Bond Strength)10 MPa20-30 sec
    Complies with ISO 29022 (Curing Time)
    Fusion Bond 7Complies with ISO 29022 (Shear Bond Strength)25 MPa20-30 sec
    Complies with ISO 29022 (Curing Time)
    Fusion Bond DCComplies with ISO 29022 (Shear Bond Strength)19 MPa20-30 secs
    Complies with ISO 29022 (Curing Time)
    Renew MDPComplies with ISO 29022 (Shear Bond Strength)21 MPa20-30 secs
    Complies with ISO 29022 (Curing Time)
    Renew UniversalComplies with ISO 29022 (Shear Bond Strength)21 MPa20-30 secs
    Complies with ISO 29022 (Curing Time)

    Note: The document explicitly states "All test results met the criteria in standard." The specific numeric values (e.g., 10 MPa for Fusion Bond 5) are the reported performance, which is implicitly accepted as meeting the standard's criteria.

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    The document does not specify the exact sample sizes used for each test. It states "Fusion Bond 5, Fusion Bond 7, Fusion Bond DC, Renew MDP and Renew Universal were tested for appearance, shear bond strength and curing time according to protocols based on ISO 29022:2013."

    The data provenance is from Prevest Denpro Limited, located in Bari Brahmana, Jammu 181133 India. The tests are presented as part of a regulatory submission, implying they were prospective tests conducted specifically for this 510(k).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    This is a physical testing submission, not a diagnostic imaging AI/ML device. Therefore, the concept of "experts establishing ground truth for a test set" in the context of medical image interpretation (e.g., radiologists) is not applicable. The ground truth is established by the physical properties measured according to an international standard (ISO 29022:2013).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. Adjudication methods like 2+1 or 3+1 are used for establishing consensus among human readers for image interpretation. This submission involves physical, objective measurements.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    No, an MRMC comparative effectiveness study was not done. This is not an AI/ML medical device where human reader performance with and without AI assistance is relevant. This is a dental material.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. There is no algorithm or AI component in these dental bonding agents. The "performance" is the physical properties of the material.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The ground truth for the device's performance is based on objective physical measurements conducted according to the ISO 29022:2013 standard for dental adhesives, specifically for shear bond strength and curing time. In essence, the ground truth is the scientific and engineering standard.

    8. The sample size for the training set

    Not applicable. There is no "training set" in the context of machine learning for this type of device. The products are dental materials, and their formulation and manufacturing are based on established chemical and material science principles, not on iterative training data.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set (in the ML sense), there is no ground truth to establish for it. The "ground truth" for the development of these materials would stem from principles of chemistry, material science, and prior knowledge in dental adhesive development.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Accusil Dental Impression Material is intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums.

    Device Description

    Accusil Dental Impression Materials are designed for dental applications to define and reproduce the structure of a patient's teeth and gums for producing crowns, bridges, occlusal models and dental implant restorative devices. Base and catalyst components are mixed in an equal ratio 1:1, placed into an impression tray and inserted into the patient's mouth. The material will conform to the patient's dentition and when set will produce a reproduction of the patient's teeth and occlusion.

    AI/ML Overview

    The document describes the acceptance criteria and performance of Accusil Dental Impression Materials as part of a 510(k) premarket notification for FDA clearance.

    Here's an analysis of the provided information, structured according to your request:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document indicates that Accusil Dental Impression Materials were tested against specific ISO standards (ISO 4823 and ISO 48-4) and internal protocols. The "Acceptance Criteria" are implicitly derived from the requirements of these standards. The "Reported Device Performance" for Accusil is provided in the comparative table with predicate and reference devices.

    Performance CharacteristicAccusil Dental Impression Materials (Reported Performance)Acceptance Criteria (Implicit from ISO standards and comparison with predicate/references)
    Working/Processing Time40-150 secSimilar to predicate (90 sec) and reference devices (35-120 sec).
    Setting time/Time in the mouth60-300 secSimilar to predicate (90 sec) and reference devices (120-330 sec).
    Hardness63-70 Shore ASimilar to predicate (46-70 Shore A) and reference devices (42-75 Shore A).
    Working Humidity50%Matches predicate (50%).
    Dimensional Accuracy99.9%-99.2%Similar to predicate (99.9%-99.2%) and reference devices (98%-99.9%).
    Stability (Linear Dimensional Change)<0.8% typicalSimilar to predicate (<0.2% typical) and reference devices (<0.1% to <0.50% typical).
    ConsistencyType 0-type 3 ISO 4823Matches predicate (Type 0-type 3 ISO 4823) and similar to reference (Type 1-type 3 ISO 4823).
    Compatibility with GypsumAll tests results met the criteria in standards.Criteria specified in ISO 4823 for impression materials.
    Elastic RecoveryAll tests results met the criteria in standards.Criteria specified in ISO 4823 for impression materials.
    Strain in CompressionAll tests results met the criteria in standards.Criteria specified in ISO 4823 for impression materials.
    Shelf Life3 yearsEstablished through relevant protocols, meeting specified criteria.
    BiocompatibilityNo biocompatibility testing was done.Assumed to meet criteria due to similar ingredients and contact type as predicate/references, introducing no new risks.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not explicitly stated in the provided text. The document mentions "Accusil light, heavy, monophase and putty were tested" and "Accusil bite registration was tested," implying samples of these product lines were used. However, the exact number of units or measurements for each test is not provided.
    • Data Provenance: The testing was performed by Prevest Denpro Limited, based in India ("Export Promotion Industrial Park Bari Brahmana, Jammu 181133 India"). The nature of the testing (physical property measurements against international standards) suggests these were prospective tests conducted on manufactured samples, rather than retrospective data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications:

    This information is not applicable to this type of device and study. The "ground truth" for dental impression materials is established by objective physical and chemical measurements against internationally recognized standards (ISO 4823, ISO 48-4), not by expert consensus or interpretations.

    4. Adjudication Method for the Test Set:

    This information is not applicable. Since the evaluation involves objective physical property measurements against standardized criteria, there is no need for an adjudication method in the context of expert review. The results are quantitative and compared directly to standard specifications.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. This type of study (comparing human readers with and without AI assistance) is relevant for diagnostic imaging AI devices, not for physical property testing of dental impression materials.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    No, a standalone AI algorithm performance study was not done. This is not an AI-driven device. The testing focused on the physical and chemical properties of the material itself.

    7. The Type of Ground Truth Used:

    The "ground truth" used for this device is based on objective standards and measurements of physical and chemical properties. Specifically, performance was evaluated against the requirements outlined in:

    • ISO 4823 (Dentistry - Elastomeric impression materials)
    • ISO 48-4 (Rubber, vulcanized or thermoplastic - Determination of hardness - Part 4: Indentation hardness by means of a durometer)
    • Internal protocols for setting time (as ISO 4823 does not specify setting time).

    8. The Sample Size for the Training Set:

    This information is not applicable. As this is not an AI/machine learning device, there is no "training set."

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable for the same reason as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212563

    Validate with FDA (Live)

    Date Cleared
    2022-05-23

    (283 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    0 - 21
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    CalApex is a calcium hydroxide, polymeric resin, root canal filling material that is used in conjunction with gutta percha or silver endodontic points.

    Calplus is a temporary or permanent root canal filling material for use following pulpectomy, or for apexogenesis or apexification.

    Cerafill RCS is a MTA (mineral trioxide aggregate) based root canal sealer that provides complete and permanent sealing of root canals. It can be used with or without root canal obturation materials.

    Endoseal is used for permanent obturation of the root canal space with the aid of obturating points in accordance with ISO 6876 for Dental Root Canal Sealing materials.

    Nanoseal S is indicated in patients for permanent obturations of root canal after vital extirpation or after treatment of pulpal gangrene and temporary filling.

    Zical Ultra Paste is used for permanent obturation of the root canal space with the aid of obturating points.

    Zical Ultra Powder/Liquid is used for permanent obturation of the root canal space with the aid of obturating points.

    Device Description

    CalApex is a non-eugenol, radiopaque, calcium hydroxide polymeric resin root canal filling material. It is a two part, base/catalyst – paste/paste system that is mixed in equal portions. The mixture is then carried to the root canal with endodontic obturation points or directly dispensed in the root canal depending on delivery method, e.g. tube or dual-barrel syringe.

    Calplus can be used as a temporary or permanent root canal filling material.

    Cerafill RCS is an endodontic sealer based on MTA, providing a biocompatible and effective root canal filling. It is premixed and pre-loaded in a syringe, which allows a complete filling of the entire root canal including accessory and lateral canals. The product is eugenol-free and will not impede adhesion inside the root canal.

    Endoseal is a Zinc Oxide/Eugenol root canal sealant. The product is made up of two separate components, a powder base and a liquid catalyst, to form the final device.

    Nanoseal S is a permanent root canal filling material which is silicone based (polydimethylsiloxane). It is cold flowable and is in a dual barrel cartridge.

    Zical Ultra Paste is a two-part, base/accelerator, paste/paste system that is mixed in equal portions. The mixture is carried to the root canal with endodontic obturation points. The product comes in a dual barrel cartridge.

    Zical Ultra Powder/Liquid is a two-part, powder/liquid system for permanent filling of root canals. The mixture is carried to the root canal with endodontic obturation points.

    AI/ML Overview

    The document describes the acceptance criteria and performance of several root canal sealers developed by Prevest Denpro Limited. The study involved a series of bench tests conducted according to ISO standards to demonstrate substantial equivalence to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test ParameterAcceptance Criteria (ISO Standard & Specific Values)Reported Device Performance (Summary for all tested sealers)
    AppearancePass/Fail (based on acceptable appearance)All passed
    FlowBased on ISO 6876:2012, Section 5.2
    - CalApex, Cerafill RCS, Endoseal, Nanoseal S, Zical Ultra Paste, Zical Ultra Powder/LiquidNot less than 17mm> 17mm
    - Calplus25-28mm25-28mm
    Film ThicknessISO 6876:2012, Section 5.5, not more than 50µm<50µm
    Water SolubilityISO 6876:2012, Section 5.6, not more than 3% by massCalApex: 1.31%, Cerafill RCS: 1.61%, Endoseal: 1.09%, Nanoseal S: 1.06%, Zical Ultra Paste: 1.31%, Zical Ultra Powder/Liquid: 0.75% (All within <3% limit)
    Working TimeBased on ISO 6876:2012, Section 5.3
    - CalApex16-18min16-18min
    - Cerafill RCS, Endoseal, Zical Ultra Paste, Zical Ultra Powder/Liquid25-30min25-30min
    - Nanoseal S8-10min8-10min
    Setting TimeBased on ISO 6876:2012, Section 5.4
    - CalApex20-40min20-40min
    - Cerafill RCS24 hoursWithin 24 hours
    - Endoseal, Zical Ultra Paste, Zical Ultra Powder/Liquid45-60min45-60min
    - Nanoseal S10-15min10-15min
    DisintegrationISO 6876:2012, Section 4.3.5, no evidence of disintegrationNo evidence of disintegration (All passed)
    Radio-opacityISO 13116:2014, above or equal to 3 mm Al6 mm Al (All tested sealers, which is above 3mm Al)
    BiocompatibilityAssessment according to ISO 10993Yes (assessment provided for all subject devices)
    Shelf LifeSame as bench testing pass criteria (ISO 6876:2012)CalApex: 2 years, Others: 3 years (All met criteria)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    The document does not specify the exact sample sizes used for each bench test. It states that the tests were conducted "according to protocols based on ISO 6876:2012 and tested for radio-opacity according to a protocol based on ISO 13116:2014". The data provenance is implied to be from the manufacturer's testing, but no specific country of origin for the data or whether it was retrospective or prospective is mentioned. The manufacturer, Prevest Denpro Limited, is located in India.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

    This type of information is not applicable to this study. The tests performed are benchtop physical and chemical property evaluations rather than evaluations requiring expert human interpretation of medical images or patient outcomes. The "ground truth" for these tests is defined by the objective measurement protocols and acceptance criteria of the specified ISO standards.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    Not applicable. The study involves objective bench testing, not subjective assessments requiring adjudication by multiple experts.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This study focuses on the physical and chemical properties of root canal sealers and does not involve AI or human readers for diagnostic interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This study does not involve an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    The ground truth for this study is defined by the objective measurement standards and acceptance criteria set forth by international ISO standards (ISO 6876:2012 for most physical properties and ISO 13116:2014 for radio-opacity). For biocompatibility, the ground truth is established by the assessment according to ISO 10993.

    8. The sample size for the training set:

    Not applicable. This study involves bench testing of medical devices, not training of machine learning models.

    9. How the ground truth for the training set was established:

    Not applicable. There is no training set for this type of device submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K212475

    Validate with FDA (Live)

    Date Cleared
    2022-05-13

    (280 days)

    Product Code
    Regulation Number
    872.3250
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Apacal ART and is suitable for several indications including:

    • · Liner to be applied to the interior of a prepared cavity prior to the insertion of restorative materials.
    • · Indirect pulp capping or management of deep caries lesions, or
    • · Direct pulp capping

    CAL-LC is indicated for use as a cavity liner and pulp capping material. It is self-adhering and highly filled.

    Calcigel is a calcium hydroxide paste that has a creamy consistency and is suitable for several indications including;

    • · Indirect pulp capping or management of deep caries lesions, or
      · Direct pulp capping

    CalUltra is a hard-setting calcium hydroxide cavity liner and pulp capping agent to be used in conjunction with all permanent restorative techniques.

    Device Description

    Apacal ART is a light cured single-component paste reinforced with tricalcium phosphate and hydroxyapatite. It is sold in a syringe with accompanying applicator tips.

    CAL-LC is a highly filled cavity liner and pulp capping material. It is sold in a syringe with accompanying applicator tips.

    Calcigel is a creamy, ready -to -use calcium hydroxide paste with a pH above 11. It is syringe with accompanying applicator tips.

    CalUltra is a hard-setting calcium hydroxide cavity liner and pulp capping agent. CalUltra is a two-part base/catalyst paste/paste system. It is sold in a tubes with accompanying spatula and mixing pad.

    AI/ML Overview

    The provided text describes the substantial equivalence of Prevest Denpro Cavity Liners to predicate devices, focusing on physical parameters and material composition rather than AI performance. Therefore, many of the requested categories related to AI studies and expert review are not applicable to this document.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that "All tests results met the criteria in standards." The acceptance criteria are implicit in the ISO and USP standards referenced for each test. The performance is reported as meeting these standards.

    Device NameTestStandard Reference (Implicit Acceptance Criteria)Reported Device Performance
    Apacal ART & Cal LCAppearanceISO 9917-2:2017, ISO 4049:2019, ISO 13116:2014Met standards
    Flexural StrengthISO 9917-2:2017 (e.g., for Apacal ART: 99.1 MPa; for CAL-LC: 96.3 MPa)Met standards
    Depth of CureISO 4049:2019Met standards
    Water SorptionISO 4049:2019 (e.g., for Apacal ART: 13.6 µg/mm³; for CAL-LC: Data not specified but implied to meet standards)Met standards
    Water SolubilityISO 4049:2019 (e.g., for Apacal ART: 4.5 µg/mm³; for CAL-LC: Data not specified but implied to meet standards)Met standards
    Radio-opacityISO 13116:2014Met standards
    CalcigelAppearanceISO 6876:2012Met standards
    FlowISO 6876:2012Met standards
    pHUSP 971 (e.g., 11-13)Met standards
    Radio-opacityISO 13116:2014Met standards
    CalUltraAppearanceISO 9917-1:2017Met standards
    Compressive StrengthISO 9917-1:2017Met standards
    Water SolubilityISO 6876:2012 (e.g., 3.8%)Met standards
    Setting TimeISO 6876:2012 (e.g., 1-3 minutes)Met standards
    Radio-opacityISO 13116:2014Met standards
    pHUSP 971 (e.g., 9.2)Met standards
    All DevicesShelf LifeRelevant protocols (not specified in detail, but implied to meet standards)3 years

    2. Sample size used for the test set and the data provenance

    The document does not specify the sample sizes used for the testing of these physical and chemical properties. The data provenance is presumed to be from the manufacturer's testing conducted in India (Prevest Denpro Limited, Jammu, India). The testing is described as being performed "according to protocols based on" relevant ISO and USP standards, indicating manufacturer-conducted in vitro/bench testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. This is a medical device clearance based on material properties and performance against established international standards, not an AI or diagnostic device requiring expert ground truth for image interpretation or similar tasks.

    4. Adjudication method for the test set

    Not applicable. This type of adjudication is typically used in clinical or image-based studies. The "adjudication" here is the assessment against specified ISO/USP standard requirements.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The device is a dental material, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. The device is a dental material, not an algorithm.

    7. The type of ground truth used

    The "ground truth" for the device's performance is based on the objective measurements against established international and national standards (ISO and USP) for material properties (e.g., flexural strength, depth of cure, water sorption/solubility, pH, setting time, radio-opacity).

    8. The sample size for the training set

    Not applicable. There is no training set mentioned, as this is not an AI or machine learning device.

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200555

    Validate with FDA (Live)

    Date Cleared
    2021-05-13

    (436 days)

    Product Code
    Regulation Number
    872.3275
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Micron Bioactive Class III and V. Restoration of cervical erosions and root surface caries, Core buildup. Base/Liner Class I, limited Class II, temporary fillings Restoration of primary teeth

    Micron Superior Shades A1. A2. A3 Class III, Class V, limited Class I Restoration of primary teeth, Core buildup

    Micron Superior Capsules Shades A1, A2, A3 Class III, Class V, limited Class I Restoration of primary teeth, Core buildup

    Micron Luting

    Base/Liner

    Cementation of all types of metal, porcelain fused to metal, and resin crowns, inlays, and bridges Cementation of orthodontic bands Cementation of stainless steel crowns or orthodontic appliances retained with stainless steel crowns

    Micron Dentin Conditioner Dentin pre-treatment prior to filling with glass ionomer cement

    Device Description

    Micron Bioactive is a radiopaque, hydroxyapatite modified-glass ionomer cement.

    Micron Superior is a radiopaque glass ionomer cement. It comes in shades A1, A2 and A3.

    Micron Superior Capsules is a radiopaque glass ionomer cement. Capsules come in shades A1, A2 and A3.

    Micron Luting is a radiopaque, self-curing translucent, particle glass-ionomer luting material.

    Micron Dentin Conditioner is a mild polyacrylic acid solution designed to remove the dentinal smear layer and to condition dentine, thus enhancing the bond between glass ionomer cement and the dentine.

    AI/ML Overview

    The document describes the Prevest Dental Cements, a line of dental materials including Micron Bioactive, Micron Superior, Micron Superior Capsules, Micron Luting, and Micron Dentin Conditioner. The acceptance criteria and the study conducted to prove the device meets these criteria are detailed in the "Testing Summary" and comparison tables.

    Here's a breakdown of the requested information:

    1. A table of acceptance criteria and the reported device performance

    The acceptance criteria for the Prevest Dental Cements are based on ISO 9917-1. The study reports that all test results met the criteria in Table 1 of ISO 9917-1.

    Acceptance Criteria (based on ISO 9917-1)Reported Device Performance (Prevest Dental Cements)
    Appearance of powder and liquidMet criteria in Table 1 of ISO 9917-1
    OpacityMet criteria in Table 1 of ISO 9917-1
    Acid-soluble lead contentMet criteria in Table 1 of ISO 9917-1
    Setting timeMet criteria in Table 1 of ISO 9917-1 (Specific values provided for each product type in the comparison tables, e.g., Micron Bioactive: 1:30 – 2 min; Micron Superior: A1 3:30-3:35 min, A2 5:20-5:30 min, A3 3:15-3:20 min; Micron Luting: 2:30 min)
    Compressive strengthMet criteria in Table 1 of ISO 9917-1
    Film thicknessMet criteria in Table 1 of ISO 9917-1
    pH (for Micron Dentin Conditioner)Met criteria in Table 1 of ISO 9917-1
    Shelf life3 years (Met relevant protocols for physical characteristics)

    2. Sample size used for the test set and the data provenance

    The document does not explicitly state the sample sizes used for the test set or the data provenance (e.g., country of origin, retrospective or prospective). It only states that "Micron line products were tested" and "Shelf life testing was completed on all devices."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    This information is not provided in the document. The testing was based on international standards (ISO 9917-1) for physical and chemical properties, not on expert clinical evaluation for ground truth.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    This information is not applicable/provided as the testing involved laboratory measurements against a standard, not human adjudication of subjective assessments.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study is typically relevant for AI-powered diagnostic devices involving human interpretation, which is not the nature of these dental cements.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This concept is not applicable to this device. These are dental cements, not an algorithm or AI system. The "standalone" testing refers to the physical and chemical property tests performed on the materials themselves, without human interpretation as part of the performance metric.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance evaluation of these dental cements is established by the specified physical and chemical properties outlined in ISO 9917-1. These are objective, measurable parameters (e.g., setting time, compressive strength, opacity) for dental materials.

    8. The sample size for the training set

    This information is not applicable as this is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable as this is not an AI/ML device that requires a training set.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1