Search Filters

Search Results

Found 9 results

510(k) Data Aggregation

    K Number
    K141816

    Validate with FDA (Live)

    Device Name
    HEMOPORE
    Manufacturer
    Date Cleared
    2014-10-23

    (108 days)

    Product Code
    Regulation Number
    874.4780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    HEMOPORE® is intended for use in patients undergoing nasal/simus surgery as a temporary wound dressing.

    HEMOPORE® functions as a topical hemostatic aid to control mild bleeding by tamponade effect, blood absorption, platelet activation and aggregation.

    It acts as an adjunct to aid in the natural healing process as a space occupying stent to separate and support tissues. It prevents adhesions and minimizes edema.

    Device Description

    HEMOPORE® is a sterile fragmentable nasal dressing of 8 cm composed of poly(DL-lactide-cos-caprolactone) urethane and blended with a chitosan derivate and a violet color additive.

    After insertion of HEMOPORE®in the nasal cavity, fluids will be absorbed by the dressing, which helps to control bleeding after surgery.

    The dressing fragments within several days by hydrolyzing ester bonds, whereafter it is drained from the nasal cavity via the natural mucus flow.

    The HEMOPORE® is sterilized in a blister package. The device is single-use, cannot be resterilized, and is a prescription product.

    AI/ML Overview

    This document is a 510(k) Pre-market Notification for the HEMOPORE® device. The provided text describes the device, its indications for use, and a summary of performance testing to demonstrate substantial equivalence to predicate devices. However, it does not contain specific acceptance criteria, detailed study results with reported device performance metrics against those criteria, or information regarding sample sizes, data provenance, expert involvement, or MRMC studies that would be typically found in a clinical study report or a more comprehensive technical document.

    Therefore, many of the requested items cannot be extracted from this document.

    Here's what can be extracted based on the provided text, and where information is missing:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria (Stated or Implied)Reported Device Performance
    Biocompatibility: Complies with ISO 10993 and FDA guidance (G95-1) for biological evaluation.HEMOPORE® complies with the biocompatibility requirements for their intended use.
    Degradation: (Implicit: Fragments within several days by hydrolyzing ester bonds)Degradation test performed, criteria met (specifics of criteria not provided).
    Shelf-life: (Implicit: Maintains functionality over stated shelf-life)Shelf-life test performed, criteria met (specifics of criteria not provided).
    Absorption: (Implicit: Absorbs fluids to control bleeding)Absorption test performed, criteria met (specifics of criteria not provided).
    Thrombogenicity/Hemostasis: (Implicit: Has appropriate hemostatic properties, at least equivalent to predicates)Thrombogenicity testing demonstrated HEMOPORE® has appropriate hemostatic properties to assure effectiveness and performs at least equivalent to predicate devices.
    Technological Characteristics: Similar to predicate devices, with color additive not raising safety/effectiveness concerns.HEMOPORE shares the same indications for use, device operation, and overall technical and functional capabilities. Performance test data demonstrated at least adequate device performance.
    Safety and Effectiveness: As safe and effective as predicate devices.Results of all testing demonstrate substantial equivalence and no concerns about safety and/or effectiveness.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size for test set: Not provided. The document mentions "in vitro testing" and "biocompatibility testing" but does not specify the number of samples used for these tests.
    • Data provenance: Not provided. The type of study (e.g., retrospective or prospective) is not mentioned beyond "in vitro testing."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable/Not provided. The testing described (biocompatibility, degradation, shelf-life, absorption, thrombogenicity) appears to be laboratory-based physical and chemical performance testing, not human-read clinical data that would require expert ground truth establishment.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/Not provided, for the same reasons as item 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. This is not an AI/imaging device. No MRMC study was conducted or mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is a medical device (intranasal splint/dressing), not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • For the performance testing mentioned (biocompatibility, degradation, shelf-life, absorption, thrombogenicity), the "ground truth" would be established by standardized laboratory methods, physical/chemical measurements, and potentially comparison to predicate devices' known performance characteristics. There is no mention of clinical ground truth such as pathology or outcomes data in this summary.

    8. The sample size for the training set

    • Not applicable. This is not an AI/machine learning device that would require a training set.

    9. How the ground truth for the training set was established

    • Not applicable. (See item 8).
    Ask a Question

    Ask a specific question about this device

    K Number
    K141423

    Validate with FDA (Live)

    Device Name
    NASOPORE-FD
    Manufacturer
    Date Cleared
    2014-08-07

    (69 days)

    Product Code
    Regulation Number
    874.4780
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    NASOPORE ® FD is a fragmentable nasal dressing and is indicated for use in patients undergoing nasal/sinus surgery as a space occupying stent to separate and prevent adhesions between mucosal surfaces; to help control minimal bleeding following surgery or nasal trauma by tamponade effect and blood absorption.

    Device Description

    NASOPORE "FD is composed of a bioresorbable poly(DL-lactide-co-s-caprolactone) urethane that fragments within several days after insertion in the nasal cavity, whereafter it is drained from the nasal cavity via the natural mucus flow.
    The NASOPORE® FD size and type are indicated on the label and are packed in a blister. NASOPORE "FD is indicated for single-use.
    The modification in this submission concerns a line extension to the predicate device NASOPORE "with a NASOPORE" FD that fragments faster than the cleared predicate device.

    AI/ML Overview

    The document provided describes the NASOPORE FD nasal dressing, a line extension of the NASOPORE nasal dressing. The purpose of this submission is to demonstrate substantial equivalence to the predicate device (K052099; NASOPORE nasal dressing). The performance data presented focuses on in vitro testing to show that the technological characteristics and performance criteria of the new NASOPORE FD are comparable to the currently cleared predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document outlines "Performance Data" from in vitro testing to demonstrate comparability. While not explicitly termed "acceptance criteria," these are the parameters used to establish substantial equivalence.

    PropertiesMethodCurrent NASOPORE® Performance (Predicate Device)New NASOPORE FD Performance (New Device)Acceptance Criteria (Implied: Comparable to Predicate)
    Shape integrity (shape intact in saline at 37°C)Visual>36 hours (in Saline)>36 hours (in Saline)>36 hours
    Fragmentation time (In saline at 37°C)Fragmentation test96 hours48 hoursN/A (Intentionally faster fragmentation)
    CompressionCompressive strength measurement>3 kPa (Firm)>3 kPa>3 kPa
    PorosityPorosity measurement95-98 %95-98 %95-98 %
    ColorVisualOff-whiteBlue/GreenN/A (Color is a difference, deemed not to raise safety/effectiveness issues)
    IV (intrinsic viscosity)IV measurementIV ≥ 0.8 dl/gIV ≥ 0.8 dl/gIV ≥ 0.8 dl/g
    BiocompatibilityISO 10993Biocompatible: Non-cytotoxic, Non-irritating (intracutaneous and oral), Non-sensitiveBiocompatible: Non-cytotoxic, Non-irritating (intracutaneous and oral), Non-sensitiveBiocompatible
    SterilityISO11135-1SAL 10-6 (half cycle validation)SAL 10-6 (rationale, identical blister pack and density)SAL 10-6
    EtO ResidualsISO10993-7< 2 mg/device< 2 mg/device< 2 mg/device
    CompositionIn process verificationConform Polyurethane specificationConform Polyurethane specification, additionally blended with PEG20.000 and D&C GreenConform Polyurethane specification
    Package integrityISO11607Qualified, Sterilized blisterpackQualified, Sterilized blisterpackQualified, Sterilized blisterpack
    Shelf-lifeASTM F1980: Real time aging18 months6 months (interim results)N/A (Shorter shelf-life is a difference, deemed not to raise safety/effectiveness issues)

    Note: The "Acceptance Criteria" column is implied based on the comparison to the predicate device. The goal of a 510(k) is to demonstrate substantial equivalence, meaning the new device performs as safely and effectively as a legally marketed predicate device. For most parameters, this means matching or exceeding the predicate's performance. For fragmentation time, the change is intentional and a defining characteristic of the "FD" (fragmentable) line extension.

    2. Sample size used for the test set and the data provenance:

    • Sample size: The document does not specify the exact sample sizes for each in vitro test. It generally refers to "in vitro testing" without providing specific numbers of samples tested for, for example, compressive strength or porosity measurements.
    • Data Provenance: The data is generated from in vitro testing of the device. The country of origin for the data is implicitly The Netherlands, as the submitter, Polyganics BV, is located in Groningen, The Netherlands. The study is a comparative in vitro study rather than clinical retrospective or prospective data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not applicable to the provided document. The performance data is based on in vitro physical and chemical property testing, not on clinical evaluation requiring expert interpretation or ground truth establishment by medical professionals.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    This information is not applicable as the testing is in vitro and does not involve adjudication by multiple human experts.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable. The device is an intranasal splint, a physical medical device, not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study with human readers and AI assistance is not relevant to this submission.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    This information is not applicable. The device is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    The "ground truth" for the in vitro tests is the direct measurement of the physical and chemical properties of the device according to established test methods and standards (e.g., ISO 10993 for biocompatibility, ISO 11135-1 for sterility, ASTM F1980 for shelf-life, and internal methods for fragmentation, compression, porosity, etc.). The reference point (or "ground truth" in terms of performance comparison) is the performance of the legally marketed predicate device, NASOPORE® (K052099).

    8. The sample size for the training set:

    This information is not applicable. The device is not an AI algorithm and therefore does not involve a training set.

    9. How the ground truth for the training set was established:

    This information is not applicable as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112267

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-10-20

    (73 days)

    Product Code
    Regulation Number
    882.5275
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neurolac® nerve guide is indicated for the reconstruction of a peripheral nerve discontinuity up to 20 mm in patients who have sustained a complete division of a nerve.

    Device Description

    Neurolac® is designed to be a flexible and transparent resorbable poly(DL-lactide- co-e-caprolactone) tube to provide a protective environment for peripheral nerve reqeneration after injury and to create a conduit to quide axonal growth across a nerve gap. Neurolac® nerve guides are provided sterile in Tyvek pouch packages and retainer in a variety of sizes. The modification in this submission concerns a line extension to the predicate devices with a thinner wall thickness.

    AI/ML Overview

    The provided document describes a 510(k) submission for a line extension to the Neurolac® Nerve guide, with the modification being a thinner wall thickness. The study is an in vitro comparison of the new thinner-walled device to the predicate device (K032115).

    Here's a breakdown of the acceptance criteria and study as requested, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state "acceptance criteria" with numerical targets for each property. Instead, it demonstrates comparability to the predicate device. The implicit acceptance criterion is that the thinner-walled Neurolac nerve guides perform equivalently to the current cleared Neurolac nerve guides.

    PropertyImplicit Acceptance Criteria / TargetReported Device Performance (Neurolac with 37% wall thickness reduction)
    Dimensional ChangesComparable to Cleared Neurolac
    Inner diameterLumen maintained till week 10-11Lumen maintained till week 10-11
    Outer diameterGradual increaseGradual increase
    Wall thicknessGradual increase in outward directionGradual increase in outward direction
    Weight ChangesComparable to Cleared Neurolac
    Swelling (water uptake)Stable in initial weeks, then gradual increaseStable in initial weeks and then gradual increase
    Weight lossAfter week 10-11After week 10-11
    AnalysisComparable to Cleared Neurolac
    NMR (degradation products)After week 12After week 12
    DSC (Tg)Gradual decrease in TgGradual decrease in Tg
    Intrinsic ViscositySlow and gradual decreaseSlow and gradual decrease
    Mechanical PropertiesComparable to Cleared Neurolac
    Mechanical propertiesStable for initial 10-11 weeksStable for initial 10-11 weeks
    Flexural propertiesStable for initial 10-11 weeksStable for initial 10-11 weeks
    Compression propertiesStable for initial 10-11 weeksStable for initial 10-11 weeks
    SuturabilityNot explicitly defined as "comparable" but notable improvementEasy suturing (compared to "Difficult suturing" for cleared device)

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify the sample size for the in vitro tests.
    The data provenance is presented as in vitro testing, not explicitly stating country of origin, though the submitter is based in The Netherlands. It is a retrospective comparison as it compares to an already cleared predicate device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. This is an in vitro test for material properties and device performance, not a study requiring expert clinical assessment for ground truth.

    4. Adjudication Method for the Test Set

    Not applicable. This is an in vitro test for material properties and device performance.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was Done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI/imaging device.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was Done

    Not applicable. This is not an algorithm/software device. The performance data presented are for the physical device itself.

    7. The Type of Ground Truth Used

    The "ground truth" here is the established performance characteristics of the predicate device (Neurolac nerve guides K032115), against which the new thinner-walled device is compared. This is based on previously established performance data for the predicate and fundamental material science.

    8. The Sample Size for the Training Set

    Not applicable. There is no concept of a "training set" in this type of in vitro device comparison study. The study focuses on demonstrating comparability to a predicate device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set mentioned or implied. The "ground truth" for comparison is the documented performance of the predicate device, which would have been established through its own preclinical studies and regulatory clearance process.

    Ask a Question

    Ask a specific question about this device

    K Number
    K070715

    Validate with FDA (Live)

    Device Name
    NASOPORE EAR
    Manufacturer
    Date Cleared
    2007-05-07

    (54 days)

    Product Code
    Regulation Number
    874.3620
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nasopore® Ear is a fragmentable ear packing and is indicated for use in patients, undergoing ear surgery, as a space occupying stent to separate and prevent adhesions between mucosal surfaces; to help control minimal bleeding, following ear surgery, by tamponade effect and blood absorption.

    Device Description

    Nasopore® Ear is composed of a bioresorbable poly(DL-lactide-co-s-caprolactone) urethane that fragments within several days after insertion in the outer and / or middle ear, whereafter it will fall out of the ear or leaves the body by the natural mucus flow. The type of Nasopore® Ear is indicated on the label. Nasopore® Ear is packed in a blister closed by a medical paper lid. Nasopore® Ear is indicated for single-use.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Nasopore® Ear device and focuses on its substantial equivalence to predicate devices rather than a detailed clinical study with specific acceptance criteria and performance metrics. Therefore, many of the requested elements (e.g., sample size for test set, number of experts, adjudication method, MRMC study, training set details) are not explicitly present in the provided document.

    However, based on the spirit of the request and what is available, here's a structured interpretation:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly define quantitative "acceptance criteria" in the way one might expect for a diagnostic AI device (e.g., target sensitivity/specificity). Instead, the "performance" verification focuses on design characteristics demonstrating safety and equivalence.

    Acceptance Criteria (Implied from Submission)Reported Device Performance (as demonstrated by design verification tests and analyses)
    Material Degradation: Device fragments and is bioresorbable.In-vitro Degradation Testing: Performed to demonstrate the material fragments and bioresorbable nature within several days after insertion. (Specific data not provided)
    Shelf Life: Device maintains its properties over its stated shelf life.Shelf Life Testing: Performed to confirm the device's stability over time. (Specific data not provided)
    Biocompatibility: Device material is safe for use in the body.Biocompatibility: Demonstrated by the fact that the material (poly(DL-lactide-co-s-caprolactone) urethane) is the same as the predicate Nasopore® Nasal Dressing (K052099), which has demonstrated satisfactory biocompatibility.
    Functional Equivalence (Physical Characteristics): Similar physical properties to predicate.Physical Characteristics: Stated to be equivalent to Nasopore® Nasal Dressing, as it's made from the same material and uses an identical manufacturing process.
    Functional Equivalence (Intended Use): Performs similarly to predicate for its indicated use.Intended Use: Stated to be substantially equivalent to Nasopore® Nasal Dressing for use as a space-occupying stent, preventing adhesions, and controlling minimal bleeding, differing only in anatomical location (outer/middle ear vs. nasal/sinus cavities). Also equivalent to Invotec Ear Tampon for its primary function.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not applicable in the context of this 510(k) submission, as no "test set" in the sense of patient data for evaluating diagnostic performance (like an AI model) was used. The evaluation was based on design verification tests (in-vitro) and comparison to a predicate device.
    • Data Provenance: The data primarily relates to in-vitro testing and manufacturing process descriptions. There is no mention of country of origin for patient data or whether it was retrospective or prospective, as no clinical study focused on patient outcomes was presented for this device. The biocompatibility claim relies on prior approval of the predicate device (Nasopore® Nasal Dressing).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    Not applicable. No "ground truth" establishment by medical experts for a diagnostic performance test set is described.

    4. Adjudication method for the test set

    Not applicable. No diagnostic test set or adjudication process is described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a medical implant (ear packing) and not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a medical implant and not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" in this context is the established safety and efficacy of the predicate devices. For biocompatibility, the ground truth is the previous determination of "satisfactory biocompatible" for the poly(DL-lactide-co-s-caprolactone) urethane material used in the Nasopore® Nasal Dressing. For functional aspects, the ground truth is the known performance and indications for use of the predicate devices (Nasopore® Nasal Dressing and Invotec Ear Tampon).

    8. The sample size for the training set

    Not applicable. No training set for an AI/algorithm is mentioned.

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K062540

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-10-18

    (49 days)

    Product Code
    Regulation Number
    874.5220
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nasopore® Ear is a fragmentable ear packing and is indicated for use in patients, undergoing outer ear surgery, as a space occupying stent to separate and prevent adhesions between mucosal surfaces; to help control minimal bleeding, following ear surgery, by tamponade effect and blood absorption.

    Device Description

    Nasopore® Ear is composed of a bioresorbable poly(DL-lactide-co-sscription caprolactone) urethane that fragments within several days after insertion in the outer ear, whereafter it is drained from the cavity. The type of Nasopore® Ear is indicated on the label. Nasopore® Ear is packed in a blister closed by a medical paper lid. Nasopore® Ear is indicated for single-use.

    AI/ML Overview

    The provided text describes the 510(k) summary for the Nasopore® Ear device, focusing on its substantial equivalence to predicate devices rather than a standalone performance study with specific acceptance criteria and detailed device performance metrics.

    Therefore, many of the requested details about acceptance criteria, a specific study proving it, sample sizes, expert involvement, and ground truth establishment are not available in the provided document. The submission relies on demonstrating substantial equivalence based on the device's material composition and intended use being similar to existing cleared devices, rather than an independent performance study with defined acceptance criteria.

    Here's a breakdown of what can be gleaned from the text, and what information is missing:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Not explicitly stated as numerical criteria for this deviceThe safety and effectiveness of the Nasopore® were demonstrated via Performance Data: data collected from design verification tests and analyses.
    In vitro degradation testing
    Shelf life testing
    Substantial equivalence to predicate devices in terms of physical characteristics and intended use.

    Explanation: The document does not set numerical acceptance criteria for the Nasopore® Ear device's performance in a clinical study. Instead, it relies on demonstrating substantial equivalence to predicate devices (Invotec Ear Tampon (Wick) w/String and Nasopore® Nasal Dressing). The "performance data" mentioned refers to design verification tests (in vitro degradation and shelf life), which are engineering tests, not clinical performance metrics against specific acceptance criteria.

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not applicable/Not mentioned. The document does not describe a clinical test set with human or animal subjects for performance evaluation against specific criteria.
    • Data Provenance: Not applicable/Not mentioned for a clinical test set. The "performance data" mentioned (in vitro degradation, shelf life testing) would typically be conducted in a lab setting.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Number of Experts: Not applicable/Not mentioned. No test set requiring expert ground truth is described.
    • Qualifications of Experts: Not applicable/Not mentioned.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Adjudication Method: Not applicable/Not mentioned. No clinical test set requiring adjudication is described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • MRMC Study: No. This is not an AI-assisted device. The Nasopore® Ear is a physical medical device.
    • Effect Size of AI Improvement: Not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Standalone Performance: Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Type of Ground Truth: Not applicable for a typical clinical performance study. The "ground truth" for the claims in this submission is the established safety and effectiveness of the predicate devices and the physical/chemical properties verified by design tests.

    8. The sample size for the training set

    • Sample Size for Training Set: Not applicable/Not mentioned. This is not an AI device that requires a training set.

    9. How the ground truth for the training set was established

    • Ground Truth for Training Set: Not applicable.

    In summary, the 510(k) submission for Nasopore® Ear relies on a demonstration of "substantial equivalence" to legally marketed predicate devices, rather than a de novo clinical performance study with explicit acceptance criteria. The performance data mentioned (in vitro degradation, shelf life) are design verification tests, not clinical performance metrics. The underlying rationale is that if the new device is made of the same material, uses the same manufacturing process, and is intended for a similar use as a previously cleared device, then its safety and effectiveness are considered substantially equivalent.

    Ask a Question

    Ask a specific question about this device

    K Number
    K052099

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-11-21

    (110 days)

    Product Code
    Regulation Number
    874.4780
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Nasopore is a fragmentable nasal dressing and is indicated for use in patients undergoing nasal/sinus surgery as a space occupying stent to separate and prevent adhesions between mucosal surfaces; to help control minimal bleeding following surgery or nasal trauma by tamponade effect and blood absorption.

    Device Description

    Nasopore is composed of a fragmentable poly(DL-lactide-co-e-caprolactone) urethane that fragments within several days after insertion in the nasal cavity, whereafter it is drained from the nasal cavity via the natural mucus flow. The Nasopore size and type are indicated on the label and are packed in a Tyvek pouch. Nasopore is indicated for single-use.

    AI/ML Overview

    The provided 510(k) submission for POLYGANICS' KOS2099 Nasopore® nasal dressing is a submission for a Class I device and relies primarily on demonstrating substantial equivalence to predicate devices rather than proving performance against specific acceptance criteria through a clinical study with defined endpoints.

    Here's a breakdown of the requested information based on the provided text, highlighting what is available and what is not typically required or present for a Class I device 510(k) relying on substantial equivalence:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    SafetyDemonstrated via biocompatibility testing per ISO 10993.
    No adverse events reported in summary, implied safe based on preclinical testing and equivalence to predicates.
    EffectivenessIn vitro fragmentation testing: Fragments within several days after insertion and is drained via natural mucus flow.
    Shelf life testing: Demonstrated acceptable shelf life.
    Space occupying stent: Functions to separate and prevent adhesions between mucosal surfaces (by comparison to predicates).
    Control minimal bleeding: Provides tamponade effect and blood absorption (by comparison to predicates).
    BiocompatibilityConforms to ISO 10993.
    Mechanical/Physical PropertiesTested (details not specified, but implied to be comparable to predicates).

    2. Sample Size Used for the Test Set and Data Provenance:

    • Sample Size: Not explicitly stated for specific tests. The submission mentions "design verification tests and analyses" and "biocompatibility, mechanical and physical property testing and in vitro fragmentation testing." These are typically conducted on a limited number of device samples in a lab setting, not a clinical "test set" in the context of human subjects.
    • Data Provenance: The device's performance data is derived from in-vitro testing and pre-clinical analyses (e.g., biocompatibility) conducted by the manufacturer, Polyganics BV, likely in The Netherlands (their submitted address). It is retrospective in the sense that it's laboratory testing and comparisons, not a prospective clinical trial on human subjects for this 510(k) submission.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not applicable and not provided in this type of 510(k) submission. For a Class I device demonstrating substantial equivalence, especially one relying on pre-clinical data and comparison to predicates, external experts determining ground truth for a clinical test set are not typically involved or required. The "ground truth" for the device's technical specifications and performance characteristics would be established by Polyganics' internal R&D and QA teams based on established test methods and standards (e.g., ISO 10993).

    4. Adjudication Method for the Test Set:

    This information is not applicable and not provided. Adjudication methods (like 2+1, 3+1) are relevant for clinical studies where multiple readers interpret clinical data (e.g., medical images) and their disagreements need resolution to establish a consensus ground truth. This 510(k) relies on lab-based testing and performance verification, not clinical interpretation.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size:

    No, an MRMC comparative effectiveness study was not done. This type of study is typically used for diagnostic devices involving human interpretation of results and is not relevant for this device (a nasal dressing). The submission's focus is on the device's physical and biological properties and its equivalence to existing products.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    This concept is not applicable to this device. "Standalone" performance is usually refers to the accuracy of an AI algorithm in performing a task without human intervention. The Nasopore® nasal dressing is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used:

    The "ground truth" for this device's safety and effectiveness is established through:

    • Pre-clinical verification testing: This includes in-vitro fragmentation testing, shelf-life testing, biocompatibility testing (ISO 10993 standards), and mechanical/physical property testing. The results of these tests and their adherence to accepted scientific methods and standards form the basis of the "ground truth" for the device's functional characteristics.
    • Comparison to predicate devices: The "ground truth" for its clinical utility (e.g., separating adhesions, controlling bleeding) is established through its substantial equivalence to legally marketed predicate devices with proven clinical histories.

    8. The Sample Size for the Training Set:

    This information is not applicable and not provided. The concept of a "training set" (used for machine learning algorithms) does not apply to this physical medical device. The device was designed and developed based on scientific and engineering principles, not trained on a dataset.

    9. How the Ground Truth for the Training Set Was Established:

    This information is not applicable and not provided for the same reasons as point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K050573

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-05-04

    (58 days)

    Product Code
    Regulation Number
    882.5275
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neurolac nerve guide is indicated for the reconstruction of a peripheral nerve discontinuity up to 20 mm in patients who have sustained a complete division of a nerve.

    Device Description

    Neurolac® (4-10 mm) is designed to be a flexible and transparent resorbable poly(DL-lactide-co-s-caprolactone) tube to provide a protective environment for peripheral nerve regeneration, for nerves with an internal diameter ≤ 9.5 mm, after injury and to create a conduit to guide axonal growth across a nerve gap. Neurolac® (4-10mm) nerve guides are provided sterile in Tyvek pouch packages and retainer in a variety of sizes (4, 5, 6, 7, 8, 10 mm).

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Neurolac® (4-10 mm) nerve guide. This type of submission focuses on demonstrating substantial equivalence to a legally marketed predicate device, rather than proving safety and effectiveness through extensive new clinical trials with specific acceptance criteria and detailed performance metrics.

    Therefore, many of the requested details about acceptance criteria, specific performance studies with sample sizes, expert involvement, and ground truth establishment are not typically part of a 510(k) submission and are not present in the provided documents.

    Here's a breakdown of the available information based on your request:

    1. Table of Acceptance Criteria and Reported Device Performance

    This information is not provided in the document because this is a 510(k) submission demonstrating substantial equivalence to predicate devices, rather than a de novo submission or PMA requiring explicit performance acceptance criteria and reporting. The "Performance Data" section in {1} outlines what kind of tests were performed (in vitro, in vivo), but doesn't list specific quantitative acceptance criteria or results in a table format.

    2. Sample Size Used for the Test Set and Data Provenance

    Again, this level of detail for specific test sets with sample sizes and data provenance (country, retrospective/prospective) is not explicitly provided in the context of the 510(k) submission. The document mentions "design verification tests and analyses" and "in vivo nerve function recovery," but doesn't detail the sample sizes or provenance for these tests. The primary "study" is the comparison to predicate devices and reliance on pre-existing data (K032115).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not provided. The 510(k) process for this device relies on demonstrating equivalence rather than establishing new clinical ground truth through expert consensus.

    4. Adjudication Method for the Test Set

    This information is not provided.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size

    This information is not provided. MRMC studies are typically for evaluating diagnostic imaging devices or other technologies where human readers interpret results, which is not directly applicable to a nerve guide.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This information is not applicable/not provided. The Neurolac® nerve guide is a medical device, not a software algorithm.

    7. The Type of Ground Truth Used

    The "ground truth" for the Neurolac® (4-10 mm) nerve guide's safety and effectiveness relies on:

    • Substantial Equivalence to Predicate Devices: The primary "ground truth" is that the Neurolac® (4-10 mm) nerve guide is substantially equivalent in design, fundamental technology, and intended use to previously cleared devices:
      • Salumedica™ nerve cuff, Salumedica, K002098
      • Neurogen™ nerve guide, Integra life science, K011168
      • Neurolac® nerve guide, Polyganics BV, K032115 (the 1-3 mm version)
    • Pre-clinical data: "Biocompatibility, mechanical and chemical and physical property testing, in vitro degradation testing and performance testing, clinical testing and testing in an animal model provide scientific evidence that Neurolac (1-3 mm) nerve quide (510(k) 032115) is safe for implantation." {1}
    • Literature and Comparison: "Evaluation of the Polyganics Neurolac (4-10 mm) nerve guide based on results from literature and the comparison of the Neurolac (4-10 mm) nerve guide with its predicate devices, shows that the material of which the Neurolac (4-10mm) nerve guide is made, is safe for implantation." {1}

    8. The Sample Size for the Training Set

    This information is not provided. The concept of a "training set" is typically associated with machine learning algorithms, which is not applicable here. The safety and effectiveness are established through reference to predicate devices and pre-clinical data, not an AI training process.

    9. How the Ground Truth for the Training Set Was Established

    As explained above, this concept of "training set" and associated ground truth establishment is not applicable to this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K042811

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-11-08

    (27 days)

    Product Code
    Regulation Number
    878.3300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The VivoSorb Sheet is indicated for the use as a temporary wound support, to reinforce soft tissues where weakness exists, or for the repair of hernia or other fascial defects that require the addition of a reinforcing material to obtain the desired surgical result. The resorbable protective film minimizes tissue attachment to the device in case of direct contact with the viscera.

    Device Description

    The VivoSorb Sheet is designed to be a flexible and transparent resorbable Device poly (DL-lactide-co-s-caprolactone) sheet to provide support to soft tissue Description where weakness exists. The VivoSorb Sheet is provided sterile in Tyvek pouch packages in a variety of sizes.

    AI/ML Overview

    This document is a 510(k) summary for the VivoSorb Sheet, a medical device. It does not describe an AI/ML powered device, therefore the standard acceptance criteria for such a device are not applicable.

    Here's an analysis of the provided information, focusing on the device's performance claims and the study used to support them, based on the document's content:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly define "acceptance criteria" in a quantitative manner as would be expected for an AI system, nor does it present specific numerical performance targets. Instead, it relies on demonstrating substantial equivalence to predicate devices through various tests.

    Acceptance Criterion (Implicit)Reported Device Performance (Summary)
    Safety- Biocompatibility testing provided reasonable scientific evidence.
    - Animal tests provided reasonable scientific evidence.
    - Evaluation based on literature comparison.
    - Compared favorably to predicate devices (Surgi Wrap K031955, IMMIX K032673) in terms of safety.
    Effectiveness (Functionality)- Design verification tests and analyses performed, including:
    - In vitro suture retention testing
    - In vitro degradation testing
    - Aging testing
    - Mechanical testing
    - Design, fundamental technology, and intended use are substantially equivalent to predicate devices.
    - Mechanical and physical property testing provided reasonable scientific evidence.
    Intended UseMeets indications for use as a temporary wound support, to reinforce soft tissues where weakness exists, or for the repair of hernia or other fascial defects.
    Resorbable protective film minimizes tissue attachment.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify the sample sizes used for the "design verification tests" or "animal tests." It only lists the types of tests performed (e.g., in vitro suture retention, degradation, aging, mechanical, biocompatibility, animal tests).
    • Data Provenance: The document does not provide details on the country of origin of the data or whether the studies were retrospective or prospective. It only mentions "design verification tests and analyses" and "animal tests."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This information is not applicable to this document as it describes a physical medical device and its performance, not an AI/ML system that requires ground truth established by experts. The "ground truth" here would be the physical properties and biological responses observed in the tests.

    4. Adjudication Method for the Test Set

    This information is not applicable for the same reason as point 3. No expert adjudication method is described or required for the device's physical and biological testing.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    This information is not applicable. The device is a physical surgical mesh, not an AI system intended to assist human readers. Therefore, an MRMC study is not relevant.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This information is not applicable. The device is a physical surgical mesh, not an algorithm.

    7. The Type of Ground Truth Used

    For this physical device, the "ground truth" was established through:

    • Objective Measurement/Pathology: This would include direct measurements from the "in vitro suture retention testing," "in vitro degradation testing," "aging testing," and "mechanical testing." For "biocompatibility testing" and "animal tests," "pathology" (histological analysis, observation of biological responses) would be used to assess safety and interaction with biological tissues.
    • Literature and Predicate Device Comparison: The determination of "substantial equivalence" also relies on the established performance and safety profiles of the predicate devices and existing scientific literature.

    8. The Sample Size for the Training Set

    This information is not applicable. This is a physical device, not an AI/ML model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable. As it's not an AI/ML model, there is no "training set" and associated ground truth in that context. The "ground truth" for the device's development would be based on engineering principles, material science, and biological understanding, informed by pre-clinical testing mentioned in point 7.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032115

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-10-10

    (93 days)

    Product Code
    Regulation Number
    882.5275
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Neurolac® nerve guide is indicated for the reconstruction of a peripheral nerve discontinuity up to 20 mm in patients who have sustained a complete division of a nerve.

    Device Description

    Neurolac® is designed to be a flexible and transparent resorbable poly(DL-lactide-co-ε-caprolactone) tube to provide a protective environment for peripheral nerve regeneration after injury and to create a conduit to guide axonal growth across a nerve gap.
    Neurolac® nerve guides are provided sterile in Tyvek pouch packages and retainer in a variety of sizes.

    AI/ML Overview

    The Neurolac® Nerve Guide is a medical device designed to assist in the regeneration of peripheral nerves after injury. The device manufacturer, Polyganics BV, submitted a 510(k) premarket notification to the FDA to demonstrate its substantial equivalence to previously marketed predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    The provided document does not specify quantified acceptance criteria (e.g., specific thresholds for suture retention strength or nerve function recovery). Instead, the performance data focused on demonstrating safety and effectiveness through various tests and analyses, indicating that the device performs comparably to predicate devices. The implicit acceptance criteria were that the device's performance data, alongside biocompatibility and comparison to predicate devices, would be sufficient to establish substantial equivalence.

    Acceptance Criteria (Implicit)Reported Device Performance
    Device biocompatibilityBiocompatibility testing was performed. Evaluation of the Polyganics Neurolac® Nerve guide based on biocompatibility testing, animal tests, results from literature, and comparison with predicate devices showed it to be safe for implantation.
    Adequate mechanical properties (e.g., suture retention)In vitro suture retention testing was conducted. No specific values provided, but the inclusion implies it met internal or comparative standards.
    Appropriate degradation profileIn vitro degradation testing was conducted. No specific values provided, but the inclusion implies it met internal or comparative standards suitable for nerve regeneration.
    Support for nerve function recoveryIn vivo nerve function recovery was evaluated in an animal model. No specific metrics or statistical results are provided within this summary, but the study implies positive outcomes demonstrating its ability to guide axonal growth and promote regeneration. Performance testing in an animal model provided reasonable scientific evidence.
    Substantial equivalence to predicate devices in design, technology, and intended useThe Neurolac® Nerve Guide's design, fundamental technology, and intended use are substantially equivalent to Neurotube™ (K983007) and NeuroGen™ Nerve Guide (K011168). This was supported by biocompatibility, mechanical and physical property testing, in vitro degradation testing, and performance testing in an animal model, indicating its safety and efficacy are comparable to the predicate devices.

    2. Sample Size and Data Provenance for Test Set:

    • Sample Size: Not explicitly stated for each test.
      • For the "In vivo nerve function recovery," it states "an animal model" was used, but the number of animals or specific experimental subjects is not provided.
      • For "In vitro suture retention testing" and "In vitro degradation testing," the number of samples or replicates used in these tests is not specified.
    • Data Provenance: The studies were conducted by Polyganics BV, a company based in The Netherlands.
      • The in vitro tests (suture retention, degradation) are laboratory-based.
      • The in vivo test was conducted using "an animal model." The specific type of animal, the location of the animal study, or if it was retrospective or prospective is not detailed, but it would have been a prospective study to evaluate the device's performance.

    3. Number of Experts and Qualifications for Ground Truth:

    Not applicable. This is a medical device for nerve repair, and the evaluation did not involve human-in-the-loop diagnostic performance or
    expert interpretation of images to establish ground truth in the context of AI/diagnostic device studies. Instead, its performance was assessed through physical, chemical, and biological testing.

    4. Adjudication Method for Test Set:

    Not applicable. As described above, the study does not involve human readers interpreting data that would require adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    Not applicable. This device is a passive nerve guide, not an AI or diagnostic tool that assists human readers. Therefore, an MRMC study comparing human readers with and without AI assistance was not performed.

    6. Standalone (Algorithm Only) Performance:

    Not applicable. The Neurolac® Nerve Guide is a physical medical device, not an algorithm. Its performance is inherent to its physical and biological properties.

    7. Type of Ground Truth Used:

    The "ground truth" for evaluating the Neurolac® Nerve Guide was established through:

    • Direct measurement and observation: For in vitro tests like suture retention and degradation, the "ground truth" was empirical data derived from standardized laboratory tests.
    • Biological outcomes: For in vivo nerve function recovery, the "ground truth" would be the observed biological response to the device in an animal model, such as nerve regeneration, functional recovery, and histological assessments. These would be determined by objective measurements and scientific observation in a controlled experimental setting.
    • Comparison to predicate devices: The "ground truth" for substantial equivalence was the established performance and safety profiles of the legally marketed predicate devices (Neurotube™ and NeuroGen™).

    8. Sample Size for the Training Set:

    Not applicable. This device is not an AI/machine learning system, so there is no "training set" in the conventional sense.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable. As there is no training set, there is no ground truth established for one.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1