Search Results
Found 63 results
510(k) Data Aggregation
(131 days)
The Philips Lumify Diagnostic Ultrasound System is intended for diagnostic ultrasound imaging in B (2D), Color Doppler, Combined (B+Color), Pulsed Wave Doppler (PWD), and M-modes.
It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: Fetal/Obstetric, Abdominal, Pediatric, Cephalic, Urology, Gynecological, Cardiac Fetal Echo, Small Organ, Musculoskeletal, Peripheral Vessel, Carotid, Cardiac, Lung.
The Lumify system is a transportable ultrasound system intended for use in environments where healthcare is provided by healthcare professionals.
The Lung Application 3 is intended to assist healthcare professionals by providing automated image processing to analyze ultrasound images for lung-related conditions. Specifically, it evaluates the adequacy of ultrasound frames for clinical interpretation and assesses the appearance of pleural lines as normal or irregular.
The Lung Application 3 is a software-only functionality integrated into the Philips Lumify Diagnostic Ultrasound System, designed to support lung ultrasound examinations. It introduces two key features: pleural line assessment and lung image view quality assessment. The Pleural Line feature identifies and assesses the appearance of pleural lines as normal or irregular (defined as thickened, interrupted, fragmented, jagged, uneven, or otherwise non-smooth appearance on ultrasound). The lung view quality tool assesses the adequacy of ultrasound frames based on overall image appearance and the presence of any pleural lines. The application operates on a compatible Android-based commercial off-the-shelf device (e.g., tablet or smartphone) connected to Lumify transducers (C5-2, S4-1, and L12-4 models). It utilizes machine learning algorithms trained on a large dataset of expert-annotated lung ultrasound images to ensure accurate analysis. The workflow includes zone selection, image acquisition, navigation, review, and editing of results, with real-time feedback provided via visual indicators for image quality and pleural line analysis. The Lung Application 3 is intended for use by trained professionals in clinical settings to assist in evaluations of adult patients (18 years and older) with various pulmonary conditions. It does not introduce any new contraindications and is designed to comply with existing safety and operational standards.
Key Features:
- Software-based functionality for lung ultrasound enhancement.
- Pleural line classification as normal or irregular appearance.
- Lung view quality assessment for diagnostic adequacy.
- Real-time feedback via visual indicators.
- Machine learning-based algorithms for accurate image analysis.
- Compatibility with existing Lumify transducers and Android devices.
The Philips Lumify Diagnostic Ultrasound System (Lumify) is a mobile, durable, and reusable, software-controlled medical device, which is intended to acquire high-resolution ultrasound data and to display the data in B mode (2D), Pulsed Wave Doppler, Color Doppler, Combined (B+ Color), and M modes. The Lumify system is compatible with iOS and Android operating systems.
The Lumify Diagnostic Ultrasound System (iOS) utilizes:
- A commercial off-the-shelf (COTS) iOS mobile item (smart phone or tablet)
- The Philips Ultrasound Lumify software running as a medical device application on the COTS device
- The Philips C5-2 Curved array USB transducer
- The Philips L12-4 Linear array USB transducer
- The Philips S4-1 Sector array USB transducer
- Lumify Micro B Transducer Cable
- Lumify Micro C Transducer Cable
- Lumify USB-C to USB-C Transducer Cable
- Lumify Power Module
Here's a breakdown of the acceptance criteria and the study proving the device's adherence, based on the provided FDA 510(k) clearance letter for the Philips Lumify Diagnostic Ultrasound System with Lung Application 3:
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Feature | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Pleural Line Assessment (Binary Classification) | One-sided 97.5% Lower Confidence Limit for Prevalence-Adjusted Bias-Adjusted Kappa (PABAK) > 0.35 (indicating at least fair agreement with ground truth). | PABAK: 0.71 (95% CI: 0.67–0.76). Concordance: 85.6% Cohen's Kappa: 0.66 (95% CI: 0.61–0.71) Consistency across transducers: curved 0.72, sector 0.70, linear 0.71 (PABAK) |
| Lung View Quality Assessment (Binary Classification) | One-sided 97.5% Lower Confidence Limit for Prevalence-Adjusted Bias-Adjusted Kappa (PABAK) > 0.35 (indicating at least fair agreement with ground truth). | PABAK: 0.76 (95% CI: 0.72–0.80) Concordance: 87.9% Cohen's Kappa: 0.67 (95% CI: 0.61–0.72) Consistency across transducers: curved 0.76, sector 0.75, linear 0.77 (PABAK) |
Study Details
1. Sample size used for the test set and the data provenance:
- Test Set Sample Size: The document does not explicitly state the exact numerical sample size for the test set. It mentions that the machine learning algorithms were trained on a "large dataset of expert-annotated lung ultrasound images" and that the retrospective data analysis evaluated the performance on a set of images to assess agreement with ground truth. More specific numbers for the test set are not provided.
- Data Provenance: The data was described as "retrospective data analysis study evaluated the performance of two artificial intelligence algorithms integrated into the Philips Lumify Diagnostic Ultrasound System for automated classification of lung view quality and pleural line appearance during clinical LUS examinations." The country of origin for the data is not specified.
2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Number of Experts: The document does not explicitly state the number of experts used to establish ground truth. It refers to "expert-annotated lung ultrasound images" and "qualified clinical experts" when establishing acceptance criteria based on inter-rater agreement.
- Qualifications of Experts: The experts are referred to as "qualified clinical experts." Specific qualifications (e.g., "radiologist with 10 years of experience") are not provided.
3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- The document does not specify the adjudication method used for establishing ground truth for the test set. It mentions "expert-annotated," implying multiple experts, but the process for resolving disagreements (if any) is not detailed.
4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- A MRMC comparative effectiveness study involving human readers with vs. without AI assistance was not explicitly described in this document as part of the performance evaluation for this 510(k) clearance. The study focused on the standalone performance of the AI algorithms against expert-established ground truth.
5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Yes, a standalone study was done. The performance evaluation described in Section 8, "Non-Clinical Performance Data," is a standalone assessment of the AI algorithms. It evaluated "algorithm agreement with ground truth labels." The results presented for PABAK, concordance, and Kappa are all measures of the algorithm's performance independent of real-time human interaction.
6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- The ground truth was established by expert annotation/consensus. The document states, "machine learning algorithms trained on a large dataset of expert-annotated lung ultrasound images" and "evaluated algorithm agreement with ground truth labels." The acceptance criteria were also "established based on published inter-rater agreement ranges for lung view quality and pleural line irregularity among qualified clinical experts."
7. The sample size for the training set:
- The document states, "It utilizes machine learning algorithms trained on a large dataset of expert-annotated lung ultrasound images." A specific numerical sample size for the training set is not provided.
8. How the ground truth for the training set was established:
- The ground truth for the training set was established through expert annotation. The document explicitly mentions "machine learning algorithms trained on a large dataset of expert-annotated lung ultrasound images."
Ask a specific question about this device
(133 days)
EPIQ:
The intended use of EPIQ Ultrasound Diagnostic System is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use:
Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging.
The clinical environments where EPIQ Series Diagnostic ultrasound Systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients.
When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live echo guidance.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed.
However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
Affiniti:
The intended use of Affiniti Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use:
Abdominal, Cardiac Adult, Cardiac Other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging.
The clinical environments where the Affiniti diagnostic ultrasound systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
The purpose of this special 510(k) Pre-Market Notification is to introduce EPIQ and Affiniti Series Diagnostic Ultrasound Systems with Custom Electromagnetic Interference (EMI) filter. The proposed EPIQ Series Diagnostic Ultrasound System and Affiniti Series Diagnostic Ultrasound Systems with new Custom EMI filter spare part, will reduce the electrical interferences induced by the external environments and feed the clean electrical signals to the ultrasound device, equivalent to predicate device, K250177.
Ultrasound systems are very sensitive in nature. These electrical interference couplings result in Image distortions or artifacts in Images, using this Custom EMI filter will reduce such artifact effects appearing on Ultrasound systems.
N/A
Ask a specific question about this device
(73 days)
EPIQ Series Diagnostic Ultrasound System: The intended use of EPIQ Ultrasound Diagnostic System is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging.
The clinical environments where EPIQ Series Diagnostic ultrasound Systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients.
When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live echo guidance.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
Affiniti Series Diagnostic Ultrasound Systems: The intended use of Affiniti Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use: Abdominal, Cardiac Adult, Cardiac Other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging.
The clinical environments where the Affiniti diagnostic ultrasound systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients. The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
The purpose of this 510(k) Pre-market Notification is to introduce the SVS v2 Contrast software application with the EPIQ and Affiniti Series Diagnostic Ultrasound Systems as part of the VM13.0 program.
SVS v2 Contrast is an automated software feature that assists in the selection of images for analysis with the new Philips license options 2D Auto EF Advanced (Adv) developed by DiA Imaging Analysis (K243235), now part of Philips Ultrasound LLC, and existing Philips license options AutoStrain LV; or 2D Auto LV (both K240850) license application in Adult Echo Transthoracic (TTE) examinations.
As described in 510(k) K240850, this feature automatically classifies each acquired image by view and selects an appropriate set of images for Left ventricle (LV) analysis. The classification is based on a Deep Learning AI inference engine; the selection is a non-AI algorithm that considers the view classification and image depth to select the optimal set of images. The difference in SVS v2 Contrast from the predicate SVS v1 (K240850) is that the algorithm is revised to include the selection of optimal images for analysis when contrast is used in routine TTE exams.
Users can launch existing 2D Auto EF; 2D Auto EF Adv; AutoStrain LV; or 2D Auto LV with the set of images that have been automatically selected without the need to review the acquired images and manually select the views. Users may select 2D Auto EF Adv, to process the selected views by SVS v2 Contrast. SVS v2 Contrast prioritizes the contrast image pair; however, if an appropriate pair of contrast images is not found, then Auto EF Adv may select non-contrast images. The manual approach to select views is still available and the user can override automatically selected images from SVS v2 Contrast.
No hardware changes to the EPIQ or Affiniti Series Diagnostic Ultrasound Systems are required when using SVS v2 Contrast, and existing, cleared Philips TTE transducers.
The SVS v2 Contrast feature is supported by all EPIQ models running software version 13.0 or higher including EPIQ CVx/CVxi, EPIQ Elite Advanced, EPIQ Elite, EPIQ 7, EPIQ 5. The SVS v2 Contrast feature is supported by the Affiniti models running software version VM13.0 or higher, including Affiniti CVx, Affiniti 70, Affiniti 50, and Affiniti 30. The SVS v2 Contrast feature is associated with the cardiac adult indication.
Acceptance Criteria and Study Details for SVS v2 Contrast Software
This document outlines the acceptance criteria and details of the study conducted to demonstrate that the Philips Ultrasound systems (EPIQ and Affiniti Series Diagnostic Ultrasound System) with the SVS v2 Contrast software application meet the established performance benchmarks.
1. Table of Acceptance Criteria and Reported Device Performance
| Criteria | Acceptance Criteria | Reported Device Performance (SVS Selected vs Manually Selected Ground Truth) |
|---|---|---|
| Primary Endpoint: Pearson's correlation coefficient (r) for Biplane EF (Automated Selection vs. Manual Selection) | Lower Confidence Bound for the Pearson's correlation coefficient (r) to be > 0.8. (Informed by published literature (Maret, E. B. (2008)) and previous regulatory submissions (K130779, K232500)). Note: The manual selection here refers to "manual contrast clip selection," which was automatically processed within the Contrast EF application, with the automated EF results serving as ground truth. | N = 46: r = 0.953 (95% CI: 0.917, 0.974), p-value < 0.0001. The lower bound of the 95% CI (0.917) is > 0.8, meeting the acceptance criteria. |
| Agreement Assessment: Bland-Altman Analysis for Biplane EF (Automated Selection vs. Manual Selection) | Acceptable performance for automatic selection of appropriate clips by SVS software and subsequent output of Contrast Enhanced Biplane EF feature compared to the results of Contrast Enhanced Biplane EF from manually selected contrast clips. (No specific numerical thresholds provided for Bland-Altman, but indicated as demonstrating "acceptable performance" if the primary endpoint is met). | N = 46: Mean Difference ± SD = 0.71 ± 4.36 (95% CI for Mean Difference: -0.58, 2.01). Lower LoA = -7.83 (95% CI: -10.06, -5.60), Upper LoA = 9.26 (95% CI: 7.03, 11.49). Reported as demonstrating acceptable performance. |
| Agreement Assessment: Pearson's correlation coefficient (r) for Biplane EF (Automated Selection vs. Manual Tracing Ground Truth) | No explicit independent acceptance criterion for this comparison; serves as an additional supporting analysis. | N = 46: r = 0.938 (95% CI: 0.890, 0.965), p-value < 0.0001. |
| Agreement Assessment: Bland-Altman Analysis for Biplane EF (Automated Selection vs. Manual Tracing Ground Truth) | No explicit independent acceptance criterion for this comparison; serves as an additional supporting analysis. | N = 46: Mean Difference ± SD = 1.22 ± 5.22 (95% CI for Mean Difference: -0.33, 2.77). Lower LoA = -9.01 (95% CI: -11.67, -6.34), Upper LoA = 11.45 (95% CI: 8.78, 14.11). |
2. Sample Size Used for the Test Set and Data Provenance
- Initial Cohort: 60 subjects
- Evaluable Study Sample (Test Set): 46 subjects (after 10 samples were non-reportable)
- Data Provenance: Retrospective data study, utilizing previously acquired exams (including contrast and non-contrast clips) with TTE transducers.
- Country of Origin: The sample included data from two sites: one outside the USA and one in the USA.
- 7 patient exams collected from the first site (outside USA).
- Remaning 39 patient exams from the second site (USA).
- Country of Origin: The sample included data from two sites: one outside the USA and one in the USA.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications
- Ground Truth (Primary): The ground truth for the primary endpoint (Biplane EF - Manually Selected (Ground Truth)) was generated by one sonographer (reviewer) who manually selected contrast clips from subject exams. These clips were then automatically processed within the Contrast EF application, with the automated EF results serving as this ground truth. This ground truth was created as part of the K243235 submission.
- Ground Truth (Supporting): For the comparison against "Manual Tracing (Ground Truth)," three sonographers manually measured the previously selected clips (by the single sonographer) following a physician's review. This clinical ground truth was also created as part of the K243235 submission.
- Qualifications of Experts: The document specifies "sonographer" for those involved in manual clip selection and measurements, and "physicians review" for the support ground truth. Specific details on years of experience or board certification are not provided within this document.
4. Adjudication Method for the Test Set
The document does not explicitly state an adjudication method (e.g., 2+1, 3+1).
- For the primary ground truth, it appears one sonographer made the manual selection, which then fed into an automated process.
- For the supporting ground truth ("Manual Tracing"), three sonographers performed measurements following a physician's review. It is not clear if these three sonographer measurements were averaged, or if disagreements were arbitrated by the physician or another expert.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was explicitly mentioned in the provided text, nor was an effect size of human readers improving with AI vs. without AI assistance. The study focused on the agreement between the device's automated selection and manually established ground truth, rather than human reader performance.
6. Standalone (Algorithm Only) Performance
Yes, a standalone performance of the algorithm (SVS v2 Contrast) was performed. The study evaluates the algorithm's ability to automatically select appropriate clips and its subsequent impact on Biplane EF calculation. The comparison is made between the "Biplane EF - SVS Selected (Automated)" and the "Biplane EF - Manually Selected (Ground Truth)," where the latter represents a ground truth derived with manual input for clip selection but still using an automated EF calculation. Additionally, a comparison was made against "Biplane EF – by Manual Tracing (Ground Truth)," which reflects a more direct human-traced ground truth reference.
7. Type of Ground Truth Used
The ground truth used was primarily a hybrid approach:
- Expert consensus/manual selection combined with automated processing: For the primary ground truth, a single sonographer manually selected clips, which were then processed by the Contrast EF application to yield the ground truth EF.
- Expert manual tracing (with physician review): For the supporting analysis, three sonographers performed manual tracings/measurements following a physician's review.
This suggests a "clinical ground truth" established by human experts, sometimes in conjunction with other software tools. It is not pathology or direct outcomes data.
8. Sample Size for the Training Set
The document does not specify the sample size for the training set. It mentions "To ensure data separation between development and testing data, the data set to which each exam belongs was identified," indicating that there were distinct training and test sets, but the training set size is not provided.
9. How the Ground Truth for the Training Set Was Established
The document does not provide details on how the ground truth for the training set was established. It only refers to the test set's ground truth and mentions that the algorithm is based on a Deep Learning AI inference engine, implying the use of a training set.
Ask a specific question about this device
(85 days)
Intended Use / Indications for Use
EPIQ:
The intended use of Philips EPIQ series diagnostic ultrasound systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use:
Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intra-cardiac Echo, Intra-luminal, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging.
The clinical environments where Philips EPIQ diagnostic ultrasound systems can be used include clinics, hospitals, and clinical point-of-care for diagnosis of patients.
When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live echo guidance.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgement and best clinical procedure.
Affiniti:
The intended use of the Affiniti Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body with the following Indications for Use:
Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
Modes of operation include: B Mode, M Mode, PW Doppler, CW Doppler, Color Doppler, Color M Mode, Power Doppler, and Harmonic Imaging.
The clinical environments where the Affiniti Diagnostic Ultrasound Systems can be used include Clinics, Hospitals, and clinical point-of-care for diagnosis of patients.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces your responsibility for sound clinical judgment and best clinical procedure.
The purpose of this Special 510(k) Pre-Market Notification is to expand the Smart View Select (SVS) software application onto the Affiniti Series Diagnostic Ultrasound Systems and to modify the Smart View Select software application onto both the EPIQ Series Diagnostic Ultrasound Systems.
Smart View Select is an automated software feature that assists the user in selection of images for analysis with the existing Philips AutoStrain LV or 2D Auto LV application in Adult Echo Transthoracic (TTE) examination. This feature automatically classifies each acquired image by view and selects an appropriate set of images for left ventricle (LV) analysis. The classification is based on a Deep Learning AI interface engine; the selection is a non-AI algorithm that considers the view classification and image depth to select the optimal set of images.
No hardware changes to the EPIQ or Affiniti systems are required when using SVS, and existing, cleared Philips TTE transducers are used with these software applications.
The SVS v2 feature is supported by all EPIQ models running software version 13.0 or higher including EPIQ CVx/CVxi, EPIQ Elite Advanced, EPIQ Elite, EPIQ 7, EPIQ 5. The SVS v2 feature is supported by the Affiniti models running software version VM13.0 or higher, including Affiniti CVx, Affiniti 70, Affiniti 50, and Affiniti 30. The SVS and SWM software features are associated with the cardiac adult indication.
The provided document details the 510(k) clearance for Philips Ultrasound's EPIQ and Affiniti Series Diagnostic Ultrasound Systems with the Smart View Select (SVS) v2 software. The SVS feature is an automated software that assists users in selecting images for Left Ventricle (LV) analysis using existing Philips AutoStrain LV or 2D Auto LV applications in Adult Echo Transthoracic (TTE) examinations. It classifies acquired images by view using a Deep Learning AI interface and then uses a non-AI algorithm to select an optimal set of images.
Here's an analysis of the acceptance criteria and the study that proves the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance
The study aimed to demonstrate agreement between LV analysis outputs (Ejection Fraction - EF, and Global Longitudinal Strain - GLS) derived from manually selected clips (ground truth) and automatically selected clips by the SVS software.
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Co-primary Endpoint: Lower Confidence Bound for Pearson's correlation coefficient to be > 0.8 for both GLS and EF measurements. | Pearson's Correlation Coefficient for EF: 0.891 (95% CI: 0.851, 0.920) Lower Confidence Bound: 0.851 (Met criteria) |
| Pearson's Correlation Coefficient for GLS: 0.906 (95% CI: 0.871, 0.931) Lower Confidence Bound: 0.871 (Met criteria) |
2. Sample Size and Data Provenance
- Test Set Sample Size: The exact number of patients or cases in the test set is not explicitly stated as a single number. However, the demographic characteristics section indicates
n=71for various demographic variables like Sex, Age, Height, Weight, BMI, Race, LV systolic function, RWMA, Known CAD, Previous reported MI Location, and LV Hypertrophy, suggesting the test dataset comprised data from at least 71 unique patients. - Data Provenance: The data was collected from a US-based medical center. The study was a retrospective data analysis.
3. Number of Experts and Qualifications for Ground Truth
- Number of Experts: Three (3) reviewers (clinical experts) participated in establishing the ground truth.
- Qualifications of Experts: They are referred to as "clinical experts." No further specific qualifications (e.g., years of experience, specific board certifications like radiologist/cardiologist) are provided in the document.
4. Adjudication Method for the Test Set
The adjudication method used to establish ground truth was consensus/averaging. For each output (GLS and EF), the average across the three reviewers was used as the ground truth.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was described. The study focused on comparing the AI-selected clips' performance (GLS and EF) against the ground truth established by human experts using manually selected clips. There is no information provided regarding how much human readers improve with AI vs. without AI assistance. The study evaluates the AI's ability to select optimal clips, a step that precedes human analysis with AutoStrain LV.
6. Standalone Performance Study (Algorithm Only)
Yes, a standalone performance study was done. The performance of the SVS software was evaluated solely based on its ability to automatically select appropriate clips, and then the subsequent, un-edited evaluation of GLS and EF through AutoStrain LV using these SVS-selected clips was compared against the human-selected clips. This represents the algorithm's performance in its specific task (clip selection) without direct human intervention in the selection process.
7. Type of Ground Truth Used
The ground truth used was expert consensus. Specifically, it was the average of GLS and EF measurements obtained from clips manually selected and subsequently semi-automatically processed by three clinical experts using the AutoStrain LV software.
8. Sample Size for the Training Set
The document does not provide the sample size used for the training set of the deep learning AI model for image classification. It only states that "the neural network of the subject device was revised as described in attachment 002," implying updates to an existing model rather than a completely new one, but no training data details are given.
9. How Ground Truth for Training Set was Established
The document does not explicitly state how the ground truth for the training set was established. It mentions that the "classification is based on a Deep Learning AI interface engine," but details regarding the training data annotation and ground truth establishment are not provided within the excerpt.
Ask a specific question about this device
(162 days)
EPIQ: The intended use of EPIQ Ultrasound Diagnostic System is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
Affiniti: The intended use of Affiniti Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use: Abdominal, Cardiac Adult, Cardiac Other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
The R-Trigger algorithm software feature on Philips EPIQ and Affiniti Ultrasound System is intended to support detection of R-wave peak (R-trigger) as an input to certain TTE clinical applications, initially including AutoStrain LV, AutoEF, 2D Auto LV (collectively referred to as "AutoStrain"), and AutoMeasure applications. The R-trigger algorithm is planned to be implemented as workflow enhancement for transthoracic clinical applications on EPIQ and Affiniti Ultrasound Systems in the VM13 software release. The Auto-Measure and AutoStrain features support users during B-mode (2D), CW-, PW- and TDI-Doppler measurements by automating some of the measurements needed to complete a routine transthoracic echo (TTE) exam for adult patients. The R-trigger feature (non-ECG-based) has been developed to enable clinical users to use AutoMeasure and AutoStrain application without the R-trigger (ECG based) input, which is currently required. There are no hardware changes to the EPIQ and Affiniti systems due to change to the introduction of the R-Trigger software application. The software application is supported by all EPIQ and Affiniti models running software version 13.0 or higher.
The provided FDA 510(k) clearance letter describes the R-Trigger software application on Philips EPIQ and Affiniti Ultrasound Systems, which aims to provide an alternative method for detecting R-wave peaks (R-triggers) for cardiac clinical applications like AutoStrain and AutoMeasure, especially when the ECG signal is unavailable or unusable.
Here's an analysis of the acceptance criteria and the study proving the device meets them:
1. Table of Acceptance Criteria and Reported Device Performance:
The primary acceptance criteria for the R-Trigger algorithm are related to the agreement of its R-wave time stamp detection with the ground truth (ECG-based R-trigger) and the subsequent impact on the clinical outputs of AutoMeasure and AutoStrain. These are evaluated using Bland-Altman analysis (for agreement, specifically the Upper and Lower Limits of Agreement, LoA) and Pearson's correlation (for correlation, specifically the Lower Confidence Bound, LCB).
| Endpoint / Outcome Comparison | Measurement Type | Acceptance Criteria (Upper/Lower LoA or LCB) | Reported Device Performance (Upper/Lower LoA or Pearson's r with 95% CI) | Met Criteria? |
|---|---|---|---|---|
| Endpoint 1: R-trigger | ||||
| R-wave peak time stamp | Time Stamp | [-99.5ms, 99.5ms] | -58.06ms (-59.34, -56.78) to 69.69ms (68.41, 70.97) | Yes (-58.06 > -99.5, 69.69 < 99.5) |
| Endpoint 3: AutoStrain | ||||
| EF | EF (Correlation) | LCB > 0.8 | 0.892 (0.853, 0.922) | Yes (0.853 > 0.8) |
| GLS | GLS (Correlation) | LCB > 0.8 | 0.992 (0.990, 0.994) | Yes (0.990 > 0.8) |
| Endpoint 2: AutoMeasure | ||||
| MV E Vel | Pw/cw Doppler velocity | [-25%, 25%] | -12.00 % (-13.17%, -10.84 %) to 12.98 % (11.81 %, 14.14%) | Yes (-12.00 > -25, 12.98 < 25) |
| LVIDd | Distance | [-30%, 30%] | -9.33 % (-9.84%, -8.81 %) to 9.33 % (8.82 %, 9.85%) | Yes (-9.33 > -30, 9.33 < 30) |
| RVLd | Distance | [-30%, 30%] | -8.60 % (-9.57 %, -7.62 %) to 5.35 % (4.38 %, 6.33%) | Yes (-8.60 > -30, 5.35 < 30) |
| TR VTI | Pw/cw Doppler VTI | [-29%, 29%] | -12.26 % (-15.15 %, -9.37 %) to 17.44 % (14.54 %, 20.33%) | Yes (-12.26 > -29, 17.44 < 29) |
| PV VTI | Pw/cw Doppler VTI | [-29%, 29%] | -13.73 % (-15.87 %, -11.59 %) to 17.91 % (15.78 %, 20.05%) | Yes (-13.73 > -29, 17.91 < 29) |
| LA Diameter a.p. (PLAX) | Distance | [-30%, 30%] | -8.05 % (-9.41 %, -6.70 %) to 9.39 % (8.03 %, 10.74%) | Yes (-8.05 > -30, 9.39 < 30) |
| MV A Vel | Pw/cw Doppler velocity | [-25%, 25%] | -15.95 % (-17.44 %, -14.47 %) to 14.69 % (13.20 %, 16.17%) | Yes (-15.95 > -25, 14.69 < 25) |
| Ao SV diam | Distance | [-30%, 30%] | -7.60 % (-8.42 %, -6.77 %) to 9.30 % (8.47 %, 10.12%) | Yes (-7.60 > -30, 9.30 < 30) |
| LVOT diam | Distance | [-30%, 30%] | -9.84 % (-10.51 %, -9.16 %) to 9.55 % (8.87 %, 10.22%) | Yes (-9.84 > -30, 9.55 < 30) |
| LVOT VTI | Pw/cw Doppler VTI | [-29%, 29%] | -10.97 % (-12.66 %, -9.28 %) to 14.41 % (12.72 %, 16.09%) | Yes (-10.97 > -29, 14.41 < 29) |
| AoR Diam(2D) = Ao Annlus diam | Distance | [-30%, 30%] | -13.02 % (-14.60 %, -11.44 %) to 13.00 % (11.42 %, 14.58%) | Yes (-13.02 > -30, 13.00 < 30) |
| RV S'(l) | TDI velocity | [-28%, 28%] | -16.11 % (-17.59 %, -14.63 %) to 17.97 % (16.49 %, 19.45%) | Yes (-16.11 > -28, 17.97 < 28) |
| LV A'(s) | TDI velocity | [-28%, 28%] | -15.85 % (-18.53 %, -13.17 %) to 17.53 % (14.85 %, 20.22%) | Yes (-15.85 > -28, 17.53 < 28) |
| TR Vmax | Pw/cw Doppler velocity | [-25%, 25%] | -9.66 % (-11.95 %, -7.38 %) to 13.83 % (11.54 %, 16.12%) | Yes (-9.66 > -25, 13.83 < 25) |
| AV VTI | Pw/cw Doppler VTI | [-29%, 29%] | -12.85 % (-14.06 %, -11.64 %) to 14.77 % (13.56 %, 15.98%) | Yes (-12.85 > -29, 14.77 < 29) |
| Ao Asc diam | Distance | [-30%, 30%] | -9.63 % (-10.49 %, -8.76 %) to 10.64 % (9.77 %, 11.50%) | Yes (-9.63 > -30, 10.64 < 30) |
| TAPSE | Distance (TAPSE/MAPSE) | [-34%, 34%] | -19.65 % (-22.85 %, -16.44 %) to 18.23 % (15.02 %, 21.43%) | Yes (-19.65 > -34, 18.23 < 34) |
| Ao STJ diam | Distance | [-30%, 30%] | -7.61 % (-8.26 %, -6.95 %) to 8.83 % (8.17 %, 9.49%) | Yes (-7.61 > -30, 8.83 < 30) |
| RA Volume (A4Cs) | Volume Contour | [-46%, 46%] | -27.08 % (-29.52 %, -24.63 %) to 32.89 % (30.45 %, 35.33%) | Yes (-27.08 > -46, 32.89 < 46) |
| LA Vol (A2Cs) | Volume Contour | [-46%, 46%] | -21.72 % (-23.61 %, -19.82 %) to 20.47 % (18.58 %, 22.37%) | Yes (-21.72 > -46, 20.47 < 46) |
| LA Vol (A4Cs) | Volume Contour | [-46%, 46%] | -23.80 % (-25.88 %, -21.73 %) to 21.51 % (19.44 %, 23.58%) | Yes (-23.80 > -46, 21.51 < 46) |
| LV A'(l) | TDI velocity | [-28%, 28%] | -17.60 % (-20.15 %, -15.06 %) to 21.44 % (18.90 %, 23.98%) | Yes (-17.60 > -28, 21.44 < 28) |
| LV E'(s) | TDI velocity | [-28%, 28%] | -19.45 % (-21.25 %, -17.65 %) to 20.89 % (19.08 %, 22.69%) | Yes (-19.45 > -28, 20.89 < 28) |
| LV E'(l) | TDI velocity | [-28%, 28%] | -19.21 % (-21.04 %, -17.37 %) to 21.62 % (19.78 %, 23.45%) | Yes (-19.21 > -28, 21.62 < 28) |
| LVPWd | Distance Short | [-40%, 40%] | -14.13 % (-14.89 %, -13.36 %) to 13.59 % (12.82 %, 14.35%) | Yes (-14.13 > -40, 13.59 < 40) |
| MV Dec. Time | Doppler Time Interval | [-35%, 35%] | -24.95 % (-26.87 %, -23.03 %) to 23.62 % (21.70 %, 25.54%) | Yes (-24.95 > -35, 23.62 < 35) |
| IVSd | Distance Short | [-40%, 40%] | -15.82 % (-16.65 %, -15.00 %) to 14.21 % (13.38 %, 15.04%) | Yes (-15.82 > -40, 14.21 < 40) |
| TV Ann diam ant-post | Distance | [-30%, 30%] | -14.91 % (-17.46 %, -12.37 %) to 14.74 % (12.20 %, 17.29%) | Yes (-14.91 > -30, 14.74 < 30) |
| LVIDs | Distance | [-30%, 30%] | -13.31 % (-14.27 %, -12.34 %) to 12.98 % (12.02 %, 13.95%) | Yes (-13.31 > -30, 12.98 < 30) |
| RVDd base (RVD1) | Distance | [-30%, 30%] | -12.60 % (-13.27 %, -11.92 %) to 12.68 % (12.00 %, 13.36%) | Yes (-12.60 > -30, 12.68 < 30) |
| RVDd mid (RVD2) | Distance | [-30%, 30%] | -20.45 % (-21.80 %, -19.11 %) to 16.86 % (15.51 %, 18.21%) | Yes (-20.45 > -30, 16.86 < 30) |
| MR VTI | Pw/cw Doppler VTI | [-29%, 29%] | -14.36 % (-15.90 %, -12.82 %) to 17.04 % (15.50 %, 18.58%) | Yes (-14.36 > -29, 17.04 < 29) |
Conclusion on Acceptance Criteria: The study results demonstrate that the R-Trigger AI-based algorithm meets all the pre-defined acceptance criteria across all three primary endpoints (R-trigger time stamp detection, AutoStrain outputs, and AutoMeasure outputs), as indicated by the reported confidence intervals for the limits of agreement and correlation.
2. Sample Size Used for the Test Set and Data Provenance:
- Test Set Sample Size: A total of 7309 cardiac clips from 3964 subjects were used for the analysis of the R-trigger AI-based algorithm.
- Data Provenance: The data consisted of retrospective cardiac TTE clips acquired with Philips Ultrasound systems. The subject demographics indicate data from various regions, including North America (USA1, USA2, Canada, Mexico) and "Rest of world". This suggests a diverse, multi-national data set.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
The document states that the R-Trigger algorithm's performance was compared against "ECG-based R-trigger" as the ground truth. It does not explicitly mention the number or qualifications of human experts used to establish this ground truth or to review the R-trigger AI's output. The ECG signal from the on-cart physio board is presented as the existing and preferred method for R-trigger detection, implying it is the established clinical standard.
4. Adjudication Method for the Test Set:
The document does not describe a formal adjudication method (e.g., 2+1, 3+1) for the test set. Instead, the study compares the R-trigger AI's output directly against the "ECG-based R-trigger" as the established ground truth. It also mentions that "users are generally expected to review and concur with the initialization and generated results" and can "edit the application(s) generated measurements and outputs based on their clinical expertise," implying a human-in-the-loop for clinical use, but not necessarily for the ground truth establishment in this particular retrospective study.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
No, a MRMC comparative effectiveness study was not explicitly stated or performed in this context. The study focuses on the standalone performance of the R-Trigger AI algorithm compared to the existing ECG-based R-trigger, and its impact on automated measurements, rather than a direct comparison of human readers' performance with and without AI assistance on R-trigger detection itself. The AI is positioned as a "back-up" or "workflow enhancement" rather than a tool to directly improve human reader accuracy in R-trigger detection, as the ECG is the preferred method when available.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:
Yes, a standalone performance study was conducted. The primary endpoint #1 specifically evaluated "the performance of the R-Trigger (non-ECG) algorithm in isolation" by comparing its detected R-wave time stamp against the ECG-based R-trigger (ground truth). Endpoints #2 and #3 then evaluated the impact of this standalone algorithm's output when used as input for AutoMeasure and AutoStrain. The study reports the performance of the AI algorithm directly, without human intervention in the R-trigger detection process for the test clips.
7. The Type of Ground Truth Used:
The primary ground truth used for evaluating the R-Trigger AI algorithm was the ECG-based R-trigger signal obtained from the ultrasound system's on-cart physio board. This is considered the established clinical standard for R-wave detection in cardiac cycles.
8. The Sample Size for the Training Set:
The document does not provide the sample size used for the training set for the R-Trigger AI algorithm. It only details the test set's sample size and demographics.
9. How the Ground Truth for the Training Set Was Established:
The document does not specify how the ground truth for the training set was established. Given that the ground truth for the test set was the "ECG-based R-trigger," it is highly probable that similar ECG signals were used to establish the ground truth for the training data, allowing the AI to learn to identify R-wave peaks from ultrasound clips in correlation with known ECG R-waves.
Ask a specific question about this device
(265 days)
Indications for Use
Focal seizure detection (with or without secondary generalization) in patients with epilepsy age 6 years and older.
160 MHz 802.11ax Wi-Fi and Bluetooth 5.0 (Dual Band)
The provided FDA 510(k) clearance letter and summary for the Philips Lumify Diagnostic Ultrasound System (K242519) primarily focus on demonstrating substantial equivalence to predicate and reference devices, rather than detailing a specific study to prove the device meets acceptance criteria in a comprehensive clinical performance sense for a new AI/software feature.
The submission is for the addition of a new transducer (C9-4ec) and a "Fertility Package" feature. The primary change is the addition of the Trans-vaginal clinical indication, enabled by the C9-4ec transducer.
Therefore, many of the requested points regarding acceptance criteria, specific studies, sample sizes, and ground truth establishment for a clinical performance study are not explicitly detailed in the provided documents, as the application relies more on non-clinical performance data and comparison to already cleared devices.
Here's a breakdown of the information that can be extracted and what is not available:
1. Table of Acceptance Criteria and Reported Device Performance
The provided document does not contain a table of specific quantitative acceptance criteria (e.g., sensitivity, specificity, accuracy targets) for clinical performance of the Fertility Package or the C9-4ec transducer, nor does it report specific clinical performance metrics. Instead, "acceptance criteria were met" is a general statement from non-clinical testing.
The document indicates that non-clinical verification testing was conducted to address system-level requirements and design specifications, and that the device complies with referenced standards.
| Acceptance Criterion (General) | Reported Device Performance (General) |
|---|---|
| Compliance with Philips internal procedures | Met (non-clinical verification testing) |
| Assurance of continued safe and effective performance | Met (non-clinical verification testing) |
| Compliance with IEC 62304 (Medical device software) | Met |
| Compliance with IEC 62366-1 (Usability engineering) | Met |
| Compliance with ISO 14971 (Risk management) | Met |
| Compliance with IEC 60601-1 (Basic safety & essential performance) | Met |
| Compliance with IEC 60601-2-37 (Ultrasonic medical diagnostic equipment) | Met |
| Compliance with IEC 60601-1-2 (EMC) | Met |
| Compliance with IEC 62359 (Thermal/mechanical indices) | Met |
| Compliance with ISO 10993-1 (Biological evaluation) | Met |
| Meets intended use | Met (based on non-clinical testing) |
| Does not raise new questions of safety or effectiveness compared to predicate | Demonstrated (based on substantial equivalence argument) |
2. Sample size used for the test set and the data provenance
The document explicitly states: "The proposed Lumify Diagnostic Ultrasound System with C9-4ec transducer did not require clinical data for determination of substantial equivalence."
Therefore, there is no mention of a clinical "test set" sample size or data provenance (country of origin, retrospective/prospective) for a clinical performance study. The evaluation focused on non-clinical performance and substantial equivalence.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
Not applicable, as no clinical test set requiring expert-established ground truth for performance metrics is described.
4. Adjudication method for the test set
Not applicable, as no clinical test set requiring adjudication is described.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The document does not describe an AI-assisted diagnostic feature that would warrant an MRMC study comparing human readers with and without AI assistance. The "Fertility Package" provides measurement tools and a summary page, which are productivity/workflow enhancements rather than AI-driven diagnostic interpretations.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Not applicable. The "Fertility Package" appears to be a set of measurement and reporting tools, not a standalone diagnostic algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
Not applicable for a clinical performance study since none was conducted or reported. For the non-clinical testing, the "ground truth" would be the engineering specifications and recognized industry standards against which the device's hardware and software performance were verified.
8. The sample size for the training set
Not applicable, as no clinical data-driven "training set" for a new algorithm is described or implied. The Fertility Package seems to be functionality built upon established ultrasound principles and measurement techniques found in reference devices.
9. How the ground truth for the training set was established
Not applicable, as no training set for a new algorithm is described or implied.
Ask a specific question about this device
(28 days)
EPIQ: The intended use of Philips EPIQ series diagnostic ultrasound systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Intra-luminal, Intra-cardiac Echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
Affiniti: The intended use of the Affiniti Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body with the following Indications for Use: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
The purpose of this Special 510(k) Pre-Market Notification is to modify the Segmental Wall Motion and introduce Auto EF with contrast software application onto the Philips EPIQ Series Diagnostic Ultrasound System and Affiniti Series Diagnostic Ultrasound Systems. SWM algorithm was previously cleared (K243862) and Auto EF contrast algorithm(K243235) was also previously cleared. The current software release version VM 13, replaces the previous version of the segmental wall motion algorithm included in version 11 of the software release with the current cleared version (K243862) without any modifications, and integrates Auto EF contrast algorithm(K243235) without any modifications into one software release.
The SWM v2 software automatically evaluates the segmental (regional) function of the left ventricle (LV) from adult TTE echo examinations. For input, the SWM algorithm uses 3 apical views (A4C, A2C and A3C) and performs border detection and tracking to identify each of the LV segments based on the three views. The SWM v2 software is a semi-automated quantitative imaging algorithm, as users are generally expected to review and concur with the initial and generated results. The users can also manually edit algorithm-generated segmental wall motion scores for individual segments based on their clinical expertise.
Auto EF with contrast is an AI based automated software feature that assists clinicians with assessing Left Ventricular Ejection Fraction (LVEF) from images acquired with contrast agent as part of an adult transthoracic echo (TTE) examination. The feature is located within the Philips Auto EF Adv application. Auto EF with contrast uses 2 apical views (Apical four chamber view (A4C), Apical two chamber (A2C)) and provides EF for each view and a biplane Simpsons EF.
No hardware changes to the EPIQ or Affiniti systems were made due to the SWM v2 and Auto EF with contrast, and existing, cleared Philips TTE transducers are used with these software applications.
This FDA 510(k) clearance (K251110) modifies the Segmental Wall Motion (SWM) algorithm and integrates an existing Auto EF with contrast software into Philips EPIQ and Affiniti Series Diagnostic Ultrasound Systems. The study details for each software component are primarily based on retrospective data analyses.
Here's a breakdown of the acceptance criteria and study information for each feature:
SWM v2 (Segmental Wall Motion v2)
| Information Requested | Description |
|---|---|
| 1. Acceptance Criteria and Reported Device Performance | Acceptance Criteria: Sensitivity and specificity of WMSI > 0.675 (Lower Bound of the Confidence Interval) based on a predefined prevalence of abnormal cases of 65%. The expected value of the correlation coefficient is 0.8, and the lower bound is 0.675, computed for Sensitivity, with a power of 80%, a one-sided alpha of 0.025, a prevalence of 65%, and a target sensitivity of 0.80. No acceptance criteria were defined specifically for Philips data, but the objective was to ensure similar trends in performance. Reported Performance: "The SWM v2 software applied to Philips data shows performance that aligns with previously reported assessment of the algorithm (K243862)." "The reported sensitivity and specificity of the SWM v2 software as compared to visual estimation of WMSI, indicate that the safety and effectiveness of the proposed subject software, SWM v2, is acceptable and aligns with the previously reported performance." (Specific numerical values for sensitivity and specificity for this clearance are not provided, only a statement of alignment with previous clearance K243862). |
| 2. Sample Size & Data Provenance (Test Set) | Sample Size: Not explicitly stated for this particular evaluation, but the study was a retrospective data analysis. Data Provenance: Transthoracic (TTE) ultrasound clips obtained from subjects referred for clinical TTE exam. No country of origin is specified, but the data originates from "Philips data." The study was retrospective. |
| 3. Number & Qualifications of Experts (Ground Truth) | Number of Experts: 3. Qualifications: "expert cardiologists." (No specific years of experience or subspecialty mentioned other than "cardiologist"). |
| 4. Adjudication Method (Test Set) | Two approaches for sensitivity and specificity analyses were tested: 1. Comparison of the average WMSI from ground truthers to the SWM v2 software's WMSI. 2. Majority agreement: A minimum of 2 out of 3 ground truthers were required to agree on WMSI = 1 (normal) or WMSI > 1 (abnormal) for the exam, which was then compared to the software's WMSI. |
| 5. Multi-Reader Multi-Case (MRMC) Study | No, an MRMC comparative effectiveness study was not explicitly described. The study compared the algorithm's performance to expert visual estimation as the ground truth. |
| 6. Standalone Performance | Yes, this was a standalone (algorithm only) performance evaluation against an expert-established ground truth. Although it is a "semi-automated quantitative imaging algorithm" where "users are generally expected to review and concur," the performance assessment described here focuses on the algorithm's output prior to user editing. |
| 7. Type of Ground Truth Used | Expert consensus/visual estimation by 3 expert cardiologists. |
| 8. Training Set Sample Size | Not provided in this document. This clearance refers to a previous clearance (K243862) for the SWM v2 algorithm. |
| 9. Training Set Ground Truth Establishment | Not provided in this document. This clearance refers to a previous clearance (K243862). |
Auto EF with Contrast
| Information Requested | Description |
|---|---|
| 1. Acceptance Criteria and Reported Device Performance | Acceptance Criteria: Lower Confidence Bound for the Pearson's correlation to be > 0.75. The expected value of the correlation coefficient is 0.85, and the lower bound is 0.75, computed with 80% power to detect a difference of -0.10 between the null hypothesis correlation of 0.75 and the alternative hypothesis correlation of 0.85 using a one-sided hypothesis test with a significance level of 0.025. Reported Performance: The acceptance criteria for the primary endpoint were met. The Pearson's correlation coefficient and associated confidence intervals are 0.952 (95%CI 0.919, 0.972). |
| 2. Sample Size & Data Provenance (Test Set) | Sample Size: Not explicitly stated. Data Provenance: Transthoracic (TTE) contrast-enhanced clips obtained from subjects referred for clinical TTE exam. A sub-analysis was performed on data acquired with Philips systems. No country of origin is specified. The study was retrospective. |
| 3. Number & Qualifications of Experts (Ground Truth) | Number of Experts: 3 sonographers for initial assessment and 1 cardiologist for final review. Qualifications: "qualified sonographers" and "Cardiologist specializing in echo." (No specific years of experience are provided). |
| 4. Adjudication Method (Test Set) | Consensus with physician review: 1. Manual selection of clips by an experienced sonographer. 2. Manual EF assessment using Simpson's method of disks by 3 qualified sonographers. 3. "Shoulder to shoulder" consensus meetings among the 3 sonographers to review manual measurements, jointly select best traced images ("Best Trace"), and adjust traces if needed. 4. A cardiologist specializing in echo subsequently reviewed the selected A4C and A2C clips with "Best Trace" and the resulting LV EF measurements, performing manual adjustments if needed. These final measurements were used as ground truth. |
| 5. Multi-Reader Multi-Case (MRMC) Study | No, an MRMC comparative effectiveness study was not explicitly described. This was a direct comparison of the algorithm's standalone performance against a consensus-derived ground truth. |
| 6. Standalone Performance | Yes, the study describes the "Contrast EF application to an EF output" in an automated fashion, and "The ground truth EF measurements were compared to the Contrast EF automated results for EF (biplane)." This indicates a standalone performance evaluation. |
| 7. Type of Ground Truth Used | Expert consensus (between 3 sonographers and 1 cardiologist) using manual EF assessment via Simpson's method of disks. |
| 8. Training Set Sample Size | Not provided. This clearance refers to a previous clearance (K243235) for the Contrast EF algorithm. |
| 9. Training Set Ground Truth Establishment | Not provided. This clearance refers to a previous clearance (K243235). |
Overall Context:
It's important to note that this 510(k) is a Special 510(k). This type of submission is usually for modifications to a currently cleared device where the changed device does not raise new questions of safety and effectiveness. Both SWM v2 and Auto EF with contrast were previously cleared as standalone applications (K243862 and K243235, respectively). This clearance focuses on integrating these previously cleared algorithms into the Philips EPIQ and Affiniti ultrasound systems and replacing an older SWM algorithm. Therefore, the performance studies described here are primarily to demonstrate that the algorithms maintain their previously established performance when integrated into the new platform, or to provide updated performance data for the newer version of the SWM algorithm.
Ask a specific question about this device
(245 days)
The Philips Lumify Diagnostic Ultrasound System is intended for diagnostic ultrasound imaging in B (2D), Color Doppler, Combined (B+Color), Pulsed Wave Doppler (PWD), and M modes.
It is indicated for diagnostic ultrasound imaging and fluid flow analysis in the following applications: Fetal/Obstetric,Abdominal, Pediatric, Cephalic, Urology, Gynecological, Cardiac Fetal Echo, Small Organ,Musculoskeletal, Peripheral Vessel, Carotid, Cardiac, Lung, Ophthalmic.
The Lumify system is a transportable ultrasound system intended for use in environments where healthcare is provided by healthcare professionals.
The Philips Lumify Diagnostic Ultrasound System (Lumify) is a mobile, durable, and reusable, software-controlled medical device, which is intended to acquire high-resolution ultrasound data and to display the data in B mode (2D), Pulsed Wave Doppler, Color Doppler, Combined (B+ Color), and M modes. The Lumify system is compatible with iOS and Android operating systems.
The Lumify Diagnostic Ultrasound System (iOS) utilizes:
- A commercial off-the-shelf (COTS) iOS mobile item (smart phone or tablet)
- The Philips Ultrasound Lumify software running as a medical device application on the COTS device
- The Philips C5-2 Curved array USB transducer
- The Philips L12-4 Linear array USB transducer
- The Philips S4-1 Sector array USB transducer
- Lumify Micro B Transducer Cable
- Lumify Micro C Transducer Cable
- Lumify USB-C to USB-C Transducer Cable
- Lumify Power Module
The purpose of this Traditional 510k pre-market notification is to add Ocular preset to currently commercialized L12-4 transducer available with Lumify Diagnostic Ultrasound System. Addition of Ocular preset will be available under newly added Ophthalmic clinical indication, as part of this 510k submission.
The Ocular Preset is an imaging setting to image the eye. It supports 2D (B Mode) and Color Doppler Mode and is available under "Ophthalmic" clinical indication.
The ocular preset is available to download with the Lumify 5.1 software and is available with the previously commercialized L12-4 Lumify transducer.
Based on the provided FDA 510(k) clearance letter for the Philips Lumify Diagnostic Ultrasound System (K242670), here's an analysis regarding acceptance criteria and the study that proves the device meets them:
It's crucial to understand that this 510(k) application is for an addition of an "Ocular preset" to an already cleared device (Lumify Diagnostic Ultrasound System, K192226), utilizing an already cleared transducer (L12-4). The core technology and device itself are established. Therefore, the "study" described focuses on validating the new Ocular preset and Ophthalmic indication against existing standards and a reference device, rather than a full de novo clinical trial for a completely new device.
Key takeaway: The document explicitly states, "The proposed Lumify Diagnostic Ultrasound System with Ocular preset did not require clinical data for determination of substantial equivalence" and "For testing, all pre-determined acceptance criteria were met." This suggests the acceptance criteria were primarily met through non-clinical performance testing and substantial equivalence arguments based on technical similarity to a predicate and reference device that already had the Ophthalmic indication.
Acceptance Criteria and Device Performance
Since no specific quantitative acceptance criteria or clinical study results are detailed in the provided document (as it was deemed unnecessary for this 510(k)), we infer the acceptance criteria from the information presented:
Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category (Inferred from regulatory requirements and submission context) | Description (Inferred) | Reported Device Performance (From K242670) | Proving Study/Method |
|---|---|---|---|
| Safety | Device must not introduce new or significantly modified risks to patients or users with the addition of the Ocular preset. | "The changes made to the subject device do not affect the use of the device, nor do they introduce any new or significantly modified risks." | Risk Analysis and Management Review (Non-clinical), Compliance with IEC 62304, IEC 62366-1, ISO 14971. |
| Effectiveness/Performance (Non-Clinical) | The Ocular preset must function as intended for diagnostic ultrasound imaging of the eye (2D and Color Doppler Modes). Image quality and performance metrics for Ocular imaging should be comparable to established standards and the reference device. | "The proposed modification... was tested in accordance with Philips internal procedures." "Non-clinical verification testing was conducted to address the system level requirements according to system and design specifications, and risk control measures." "All pre-determined acceptance criteria were met." | Requirements Review, Product Specification Review, Design Reviews, Internal Non-clinical verification testing. Comparison to Sparq Diagnostic Ultrasound System (K162329) with L12-4 transducer and Ophthalmic indication. |
| Substantial Equivalence | The device with the Ocular preset must be substantially equivalent to a legally marketed predicate device (K192226) and/or a reference device (K162329) for the Ophthalmic indication. | "The proposed device is substantially equivalent to the predicate and reference devices in terms of indications for use, design, technological characteristics, modes of operations, safety, and effectiveness." | Comparison of design features, indications for use, fundamental scientific technology, non-clinical performance testing, safety and effectiveness. (Table 6.1: Technological Comparison) |
| Compliance with Standards & Guidance | The device must comply with relevant medical device standards and FDA guidance documents. | "The proposed Lumify Diagnostic Ultrasound System with Ocular preset is Track 3 device and comply with the reference standards and with FDA ultrasound guidance document, 'Marketing Clearance of Diagnostic Ultrasound Systems and Transducers: Guidance for Industry and Food and Drug Administration Staff, February 21, 2023'." | Internal verification of compliance. |
Study Details (Based on the supplied document)
Given that the submission category is a "Traditional 510(k)" for an addition to an already cleared device, and not a de novo submission or one requiring extensive clinical data, the "study" primarily consists of non-clinical performance testing and a technological comparison to establish substantial equivalence.
-
Sample Size used for the test set and the data provenance:
- Test Set Sample Size: Not explicitly stated as a number of cases or images. The testing conducted was "non-clinical verification testing" to address system-level requirements, design specifications, and risk control measures. This implies testing against engineering specifications and possibly phantom studies rather than a clinical dataset of patients.
- Data Provenance: Not applicable in the context of a clinical test set from patients. The testing was internal to Philips Ultrasound, conducted according to their internal procedures.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- This information is not provided as no clinical test set requiring expert ground truth establishment was deemed necessary for this 510(k) submission. The substantial equivalence argument relies on the prior clearance of the predicate and reference devices for similar indications.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- None is explicitly described. As no clinical test set was required for human review with an established ground truth, adjudication methods are not applicable here.
-
If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC study was done. This 510(k) is for a diagnostic ultrasound system and its imaging preset, not an AI/CADe/CADx device that assists human readers. Therefore, the concept of human readers improving with AI assistance is not applicable to this submission.
-
If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:
- Not applicable. This is a diagnostic imaging system, not an algorithm. Its performance is inherent to its ability to acquire and display ultrasound data.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the non-clinical performance testing, the "ground truth" would be the engineering and performance specifications the device was designed to meet. For the substantial equivalence argument, the effective "ground truth" is the demonstrated safe and effective performance of the predicate and reference devices for similar imaging capabilities. No clinical ground truth (e.g., pathology, outcomes) was required for this specific 510(k) submission.
-
The sample size for the training set:
- Not applicable. This is a traditional diagnostic ultrasound system, not an AI/machine learning product that requires a "training set" in the conventional sense. The "training" of the system refers to its design and engineering to meet diagnostic imaging requirements.
-
How the ground truth for the training set was established:
- Not applicable, as there is no "training set" in the machine learning context for this device. Ground truth for the underlying ultrasound technology would be established through physics principles, engineering design, and prior clinical validation of similar devices.
In summary, the provided document explicitly states that clinical data was not required for this 510(k) submission. The acceptance criteria were primarily met through non-clinical performance testing (design reviews, risk analysis, system requirements verification) and demonstrating substantial equivalence to existing, cleared devices (the predicate Lumify system and the Sparq system which already has an Ophthalmic indication for the L12-4 transducer), ensuring the new Ocular preset did not introduce new safety or effectiveness concerns.
Ask a specific question about this device
(112 days)
The Flash Ultrasound System 5100 Point of Care (5100 POC and 5100 POC Pro) is indicated for diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use:
Abdominal, Adult Cephalic, Cardiac Adult, Cardiac Pediatric, Cerebral Vascular, Fetal/OB, Lung, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Neonatal Cephalic, Ophthalmic, Other: Carotid, Other: GYN, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Trans-Esophageal (Cardiac), Trans-Vaginal.
Modes of operation include B Mode (or 2D), Color Doppler, Color Power Angio Doppler, Continuous Wave Doppler, Pulse Wave Doppler, Tissue Doppler, M-Mode (including anatomical M-Mode), Harmonics (Tissue), and Combined modes.
The clinical environments where the Flash Ultrasound System 5100 POC may be used are: hospitals, surgery centers, clinics, physicians' offices, diagnostic centers, critical care and emergency room environments, and point-of-care clinical settings for diagnosis of patients.
The systems are intended to be installed, used, and operated only in accordance with the safety procedures and operating instructions given in the product user information. Systems are to be operated only by appropriately trained healthcare professionals for the purposes for which they were designed. However, nothing stated in the user information reduces the user's responsibility for sound clinical judgment and best clinical procedure.
The Flash Ultrasound System 5100 Point of Care (POC) device is constructed with assets from the previously cleared 5000 Compact Series hardware platform (K222648) and a new Rhythm software platform that is new to the POC environment.
The Flash Ultrasound System 5100 POC supports a full touch screen interface on the main display as the primary means of directing the system, leveraging customer familiarity with smart devices and the ease of cleaning and maintenance of a pure glass surface.
Additionally, Flash Ultrasound System 5100 POC is designed to fit into tight spaces in a crowded Emergency Room (ER) with its small footprint, sleek design, and a large vertical tablet with ergonomic controls. The new Rhythm software offers point-of-contact workflow enhanced with assisted needle visualization, streamline point-of-care workflows such as flexible patient data management and AutoStrain EF. In addition to these POC workflows, the Rhythm software provides enhanced responsiveness with an easy-to-learn user interface as all controls are located in the same plane as the user's eye and hand for working more efficiently with the patient. The Flash Ultrasound System 5100 POC has two models, the 5100 POC and the 5100 POC Pro. The 5100 POC version does not support two transducers, C5-1 and X8-2t; however, the 5100 POC Pro supports all the transducers listed for the system.
The provided FDA 510(k) clearance letter and summary for the Flash Ultrasound System 5100 Point of Care do not include detailed information about specific acceptance criteria related to its performance as a diagnostic imaging device (e.g., sensitivity, specificity, resolution), nor does it describe a study specifically designed to prove these performance metrics.
Instead, the document focuses on demonstrating substantial equivalence to predicate devices through a combination of design features, intended use, fundamental scientific technology, and a comprehensive suite of non-clinical performance and safety testing.
Here's an analysis based on the provided text, addressing the requested information to the extent possible:
Acceptance Criteria and Study Proving Device Meets Criteria
The submission primarily emphasizes compliance with recognized standards and verification testing against system and design specifications, rather than clinical performance metrics in a sense that would typically involve sensitivity, specificity, or human reader improvement with AI assistance.
1. Table of Acceptance Criteria and Reported Device Performance
Based on the provided text, the "acceptance criteria" are broad compliance with international standards and internal design specifications, and the "reported device performance" is that these criteria were met.
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| Safety and Effectiveness | Demonstrated through non-clinical performance testing and compliance with various IEC and ISO standards. All pre-determined acceptance criteria were met. |
| System Level Requirements | Addressed according to system and design specifications. |
| Risk Control Measures | Implemented and verified through Risk Analysis and Risk Management activities (ISO 14971). |
| Electromagnetic Compatibility (EMC) | Complied with IEC 60601-1-2. |
| Usability | Complied with IEC 60601-1-6 and IEC 62366-1. User interface found safe and effective. |
| Software Life Cycle Process | Complied with IEC 62304. |
| Ultrasonics Field Characterization (Thermal/Mechanical Indices) | Complied with IEC 62359. |
| Biocompatibility | Not needed as patient-contacting materials and manufacturing processes were not impacted, and common parts are used in other cleared systems. |
| Design Controls | Followed (Requirements Review, Risk Analysis, Product Specifications, Design Reviews, Verification). |
2. Sample Size Used for the Test Set and Data Provenance
- Not explicitly stated for clinical performance. The document states: "The proposed Flash Ultrasound System 5100 POC did not require clinical data for determination of substantial equivalence." This implies that no clinical test set (patient data) was used in a comparative effectiveness study for this submission. The testing was primarily non-clinical.
- Data Provenance: N/A for clinical performance. For non-clinical testing, the provenance would be internal Philips Ultrasound LLC laboratory testing.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Not applicable. Since clinical data was not required for substantial equivalence, there was no independent expert panel establishing ground truth on patient cases for performance evaluation.
4. Adjudication Method for the Test Set
- Not applicable. No clinical test set requiring adjudication was used.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done
-
No. The document explicitly states: "The proposed Flash Ultrasound System 5100 POC did not require clinical data for determination of substantial equivalence." Therefore, no MRMC study was conducted to evaluate human reader improvement with AI assistance (as AI assistance is not mentioned as a feature of the device's diagnostic performance, but rather "assisted needle visualization" is a workflow enhancement).
-
Effect size of human readers improve with AI vs without AI assistance: Not applicable, as no such study was performed or needed for this clearance. The "assisted needle visualization" is a feature mentioned, but its impact on reader performance is not quantified in this submission.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done
- Not applicable for diagnostic performance metrics. The device is an ultrasound system with various imaging modes and "assisted needle visualization." While there is a "new Rhythm software platform," the submission does not present it as a standalone diagnostic algorithm requiring performance metrics like sensitivity/specificity on its own. Its performance is integrated into the ultrasound system's overall functionality.
7. The Type of Ground Truth Used
- For non-clinical testing: The ground truth would be defined by the specifications and expected outputs of the system based on physical and engineering principles, verified through testing against known inputs or reference standards (e.g., phantom measurements, electrical signals, safety limits).
- For diagnostic clinical performance: No clinical ground truth (e.g., expert consensus, pathology, outcomes data) was used or required for this 510(k) submission.
8. The Sample Size for the Training Set
- Not applicable. The submission does not describe the development or training of a de-novo AI algorithm that would require a "training set" in the context of generating diagnostic performance metrics. The "Rhythm software" is mentioned, and "assisted needle visualization" is a feature, but the document focuses on its equivalence and compliance, not on a machine learning training paradigm.
9. How the Ground Truth for the Training Set was Established
- Not applicable. As no training set for a diagnostic AI algorithm in the context of performance evaluation was described, the method for establishing its ground truth is not provided.
Ask a specific question about this device
(78 days)
EPIQ: The intended use of EPIQ Ultrasound Diagnostic System is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use: Abdominal, Cardiac Adult, Cardiac other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), intra-luminal, intra-cardiac echo, Musculoskeletal (Conventional), Musculoskeletal (Superficial), Ophthalmic, Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung. When integrated with Philips EchoNavigator, the systems can assist the interventionalist and surgeon with image guidance during treatment of cardiovascular disease in which the procedure uses both live X-ray and live echo guidance.
Affiniti: The intended use of Affiniti Series Diagnostic Ultrasound Systems is diagnostic ultrasound imaging and fluid flow analysis of the human body, with the following indications for use: Abdominal, Cardiac Adult, Cardiac Other (Fetal), Cardiac Pediatric, Cerebral Vascular, Cephalic (Adult), Cephalic (Neonatal), Fetal/Obstetric, Gynecological, Intraoperative (Vascular), Intraoperative (Cardiac), Musculoskeletal (Conventional), Musculoskeletal (Superficial), Other: Urology, Pediatric, Peripheral Vessel, Small Organ (Breast, Thyroid, Testicle), Transesophageal (Cardiac), Transrectal, Transvaginal, Lung.
The purpose of this Traditional 510(k) Pre-Market Notification is to introduce Remote Software Management (RSM) software feature onto the EPIQ and Affiniti Series Diagnostic Ultrasound Systems. The proposed EPIQ Series Diagnostic Ultrasound System and Affiniti Series Diagnostic Ultrasound Systems with new RSM software feature, which is a workflow-enhancement software feature, is equivalent to predicate device, K242020. The Philips RSM feature provides the capabilities to remotely upgrade EPIQ/Affiniti Systems, with the existing RSM capabilities, only patch upgrade was possible, and full upgrade (including OS components) was not supported. The full RSM feature provides the capabilities to perform a remote upgrade to EPIQ/Affiniti Systems. The RSM feature is capable of deploying following packages onto the system: OS Packages, Device Drivers, Printer Drivers, BIOS/UEFI version, Application Packages, Service Application Packages, Automatic backup/restore of existing system settings. Along with the above functionalities, it also supports Automatic and Scheduled software downloads/Installation. No hardware changes to the EPIQ or Affiniti Systems are required when using RSM software feature, and existing, commercialized Philips transducers are used for the RSM software feature. The feature is supported by EPIQ and Affiniti models running software version 12.0.
The provided FDA 510(k) clearance letter (K250177) describes the Philips EPIQ and Affiniti Series Diagnostic Ultrasound Systems with a new feature: Remote Software Management (RSM). The submission focuses on demonstrating substantial equivalence to a predicate device (K242020), specifically highlighting the enhancements to the RSM feature.
Here's an analysis of the acceptance criteria and study information, based on the provided document:
Acceptance Criteria and Device Performance
The document describes the acceptance criteria in terms of the RSM feature's capabilities and its performance in comparison to the predicate device's existing RSM. The key acceptance criteria revolve around the ability to perform comprehensive remote software upgrades without compromising safety or effectiveness.
Table 1: Acceptance Criteria and Reported Device Performance for Remote Software Management (RSM)
| Acceptance Criteria (Derived from "Application Description" and "Application performance" under Comparison) | Reported Device Performance (as stated in the document) |
|---|---|
| Ability to remotely upgrade to major software releases (full upgrade including OS components). | The full RSM feature provides the capabilities to perform a remote upgrade to EPIQ/Affiniti Systems, including OS Packages, Device Drivers, Printer Drivers, BIOS/UEFI version, Application Packages, and Service Application Packages. |
| Support for automatic backup/restore of existing system settings during upgrades. | Supports "Automatic backup/restore of existing system settings." |
| Support for automatic and scheduled software downloads/installation. | "Along with the above functionalities, it also supports Automatic and Scheduled software downloads/Installation." |
| User interface facilitates download and installation process. | "The user can go into Philips Support Connect and navigate the download and installation process for system updates and upgrades. Additionally, Remote Software Management supports automatic and scheduled software downloads and installations." |
| Verification of retention of all system settings and customizations post-installation. | "Remote Service Engineer performs verification post installation, verify all settings and customizations are retained." |
| Installation verification via Philips Remote Service portal. | "Installation is verified through the Philips Remote Service portal." |
| Overall application performance is equivalent to the predicate device. | "Verification and Validation Testing has been conducted on the RSM feature to support the Application performance is equivalent to the predicate device." |
Study Details for Demonstrating Equivalence and Meeting Criteria
The provided document does not describe a clinical study or a comparative diagnostic performance study in the traditional sense, as the submission focuses on a software workflow enhancement feature (Remote Software Management). Instead, the "study" for this feature is a series of non-clinical performance tests and a demonstration of substantial equivalence.
1. Sample sizes used for the test set and the data provenance:
- Test Set Sample Size: Not explicitly stated as a numerical sample size of patient data or images. The "test set" in this context refers to the simulated testing environments or actual ultrasound systems used to validate the RSM feature.
- Data Provenance: Not applicable in the context of diagnostic image data. The validation would have involved testing the software functionality on the EPIQ and Affiniti platforms. The document mentions "Philips Internal Procedures" for testing.
2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable as this submission is for a software management feature, not a diagnostic algorithm that requires ground truth established by medical experts. The validation would have been performed by software/system engineers and subject matter experts in software deployment and system functionality.
3. Adjudication method for the test set:
- Not applicable for this type of submission. The "adjudication" would be internal verification and validation against functional and performance requirements by the development and quality assurance teams.
4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC comparative effectiveness study was done. This submission is for a Remote Software Management feature, which is a workflow enhancement for system maintenance, not a diagnostic AI tool that assists human readers.
5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- The "standalone" performance here refers to the RSM feature functioning correctly on its own in deploying software updates. The document states:
- "All software verification tests met the acceptance criteria."
- "Results of these tests show that the proposed Remote Software Management software feature meets the intended use."
- This implies that the algorithm/feature was tested to ensure it performs its intended function (remote upgrades, backups, scheduled downloads) accurately and reliably.
6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- Not applicable in the medical/diagnostic sense. The "ground truth" for the RSM feature would be the successful execution of its defined functional and performance requirements (e.g., a software package is correctly installed, system settings are retained, the system remains operational after an upgrade).
7. The sample size for the training set:
- Not applicable. This is not a machine learning or AI algorithm in the context of diagnostic interpretation that requires a "training set" of patient data.
8. How the ground truth for the training set was established:
- Not applicable as there is no training set for a diagnostic algorithm.
Summary of the Study and Equivalence Argument:
The "study" described in K250177 for the Remote Software Management (RSM) feature primarily consisted of non-clinical performance testing and software verification/validation.
- Approach: Philips evaluated the RSM feature against internal procedures, adhering to software lifecycle processes (IEC 62304) and risk management (ISO 14971).
- Comparison Basis: The proposed RSM feature was compared to the predicate device's existing, more limited RSM capabilities (which only supported bug fixes and patch upgrades). The new feature expands this to include major software releases, OS components, drivers, BIOS/UEFI versions, application packages, service application packages, and automatic backup/restore.
- Conclusion of Testing: "All software verification tests met the acceptance criteria." and "Results of these tests show that the proposed Remote Software Management software feature meets the intended use."
- Substantial Equivalence Argument: The document argues that this enhancement is substantially equivalent to the predicate device because it does not introduce new questions of safety or effectiveness. The core function of the ultrasound system (diagnostic imaging) remains unchanged, and the RSM feature is purely a workflow enhancement for system maintenance. No hardware changes are required, and existing transducers are used. The comparison table directly addresses the differences and highlights how the improved RSM capabilities are an extension of the predicate's feature, not a fundamentally new diagnostic function.
Ask a specific question about this device
Page 1 of 7