Found 26 results

510(k) Data Aggregation

    K Number
    K221502

    Validate with FDA (Live)

    Date Cleared
    2022-07-26

    (64 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    0.1 - 100
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Parcus Synd-EZ SS is intended to be used as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as adjuncts in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting.

    The Parcus Synd-EZ SS is intended to provide fixation during the healing process following syndesmotic trauma, such as fixation of the syndesmosis (syndesmosis disruptions) in connection with Weber B and Weber C ankle fractures.

    Device Description

    The Parcus Synd-EZ SS consists of two buttons with suture connecting them and are designed to be used for stabilizing the syndesmosis during healing. Building upon the technology of the Parcus Synd-EZ Ti, the Synd-EZ SS is easily adjusted for length and resists elongation without requiring knots to secure the fixation. The device is made from medical grade titanium, stainless steel and UHMWPE suture and is provided sterile.

    AI/ML Overview

    The provided text is a 510(k) premarket notification summary for a medical device called "Parcus Synd-EZ SS." This document is a regulatory filing that demonstrates substantial equivalence to a predicate device, and as such, it does not contain acceptance criteria or detailed study information for new device performance.

    The primary purpose of this 510(k) is a labeling change for MR conditional parameters, indicating that the design, materials, and manufacturing processes of the device itself have not changed since its previous clearance (K192964). Therefore, the document explicitly states: "No new testing was completed, apart from the testing required to support MR Conditional parameters that was provided in this submission."

    Given this, I cannot extract the information you requested about acceptance criteria and proof of device performance for the medical functions of the Parcus Synd-EZ SS from this document. The document primarily focuses on the MR Conditional testing.

    However, I can provide the available information regarding the MR Conditional testing:

    Regarding MR Conditional Testing:

    1. Table of Acceptance Criteria and Reported Device Performance (for MR Conditional Parameters):

    Acceptance Criteria (MR Conditional)Reported Device Performance
    Magnetically induced forceDevice was tested for magnetically induced force and the results were used to establish MR system conditions suitable for safe use. (Specific values or thresholds not provided in the summary).
    Magnetically induced torqueDevice was tested for magnetically induced torque and the results were used to establish MR system conditions suitable for safe use. (Specific values or thresholds not provided in the summary).
    Heating by RF fieldsDevice was tested for heating by RF fields and the results were used to establish MR system conditions suitable for safe use. (Specific values or thresholds not provided in the summary). Note: The document indicates "The Synd-EZ SS was evaluated for use in the MR Environment and were determined to fit the definition of MR Conditional." This implies it met the relevant safety thresholds.
    Image artifactDevice was tested for image artifact and the results were used to establish MR system conditions suitable for safe use. (Specific values or thresholds not provided in the summary).

    2. Sample size used for the test set and the data provenance:

    • Sample Size: "worst-case device was selected for testing." The exact number of devices tested is not specified, but it implies a single representative device or a small number to characterize the "worst-case."
    • Data Provenance: Not explicitly stated, but assumed to be from internal testing conducted by Parcus Medical, LLC (or a contracted lab) to comply with FDA guidance and ASTM standards. This would be prospective testing specifically for this submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable for MR Conditional testing. MR Conditional testing relies on standardized physical measurements and engineering evaluations against defined limits, not expert clinical consensus on a "ground truth" derived from patient data.

    4. Adjudication method for the test set:

    • Not applicable. MR Conditional testing involves objective measurements rather than adjudication of interpretations.

    5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done:

    • No, an MRMC study was not done. This type of study is for evaluating human performance with and without AI assistance, which is not relevant to MR Conditional testing of a physical implant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable. This is not an AI/algorithm-based device.

    7. The type of ground truth used:

    • For MR Conditional testing, the "ground truth" is defined by the safety limits and thresholds established in regulatory guidance (FDA Guidance Document - Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment) and applicable ASTM standards.

    8. The sample size for the training set:

    • Not applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable.

    In summary, this document is a regulatory submission focused on demonstrating substantial equivalence and providing MR Conditional labeling. It does not provide detailed performance data or clinical study results for the device's primary indications, as those would have been covered in the original 510(k) clearances (K191783 and K192964) to which this device is substantially equivalent.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202662

    Validate with FDA (Live)

    Date Cleared
    2021-02-22

    (161 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Parcus Titanium Interference Screw is indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair

    Device Description

    The Parcus Titanium Interference Screw is a cannulated, threaded, tapered fastener for use in interference fixation of ligaments and tendons in patients requiring ligament or tendon repair. The device is made from a Titanium alloy, Ti-6AI-4V ELI (ASTM F136) and is available in sizes ranging from 7-12mm in diameter and 20-35mm in length.

    AI/ML Overview

    The provided text pertains to the 510(k) premarket notification for the "Parcus Titanium Interference Screws." This is a medical device for orthopedic fixation, specifically for ligaments and tendons. The document focuses on regulatory approval based on demonstrating substantial equivalence to a predicate device, rather than a study evaluating the performance of a diagnostic AI device against acceptance criteria. Therefore, most of the requested information regarding AI device-specific acceptance criteria and studies (such as MRMC studies, human-in-the-loop performance, expert qualifications for ground truth, sample sizes for training/test sets concerning AI, etc.) is not applicable to this document.

    However, I can extract information related to the device's performance testing where described, as it serves a similar purpose to acceptance criteria demonstration for physical medical devices.

    Here's an analysis based on the closest applicable information in the provided text:

    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria / Performance MetricReported Device Performance
    Pull-out StrengthDemonstrated no significant differences compared to competitive devices (based on K083619 submission).
    Insertion TorqueDemonstrated no significant differences compared to competitive devices (based on K083619 submission).
    Pyrogenicity (LAL testing)Concluded that the device does not raise any additional concerns regarding pyrogenicity.
    Substantial Equivalence (to Predicate Device K083619)Determined to be the exact same device as the predicate, with the only changes being the inclusion of MR Conditional parameters into the Instructions for Use.

    2. Sample size used for the test set and the data provenance

    • Sample Size: The document mentions that pull-out strength and insertion torque were measured for the smallest (7mm), largest (12mm), and an intermediate size of the Parcus Titanium Interference Screws. The exact number of samples for each size is not specified in this document.
    • Data Provenance: The performance data for pull-out strength and insertion torque is referenced as being included with the submission of K083619 (the predicate device). This suggests the performance data was likely generated through in-vitro mechanical testing. The country of origin for this data is not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    N/A - This medical device is a physical fixation screw, not an AI diagnostic device. Ground truth as typically understood for AI studies (e.g., expert consensus on image interpretation) is not applicable here. The "ground truth" for this device's performance would be the direct measurements of mechanical properties.

    4. Adjudication method for the test set

    N/A - As above, adjudication methods for expert disagreement on ground truth are not applicable to the mechanical testing of a physical medical device.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    N/A - This is not an AI-assisted diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    N/A - This is not an AI algorithm.

    7. The type of ground truth used

    For the mechanical performance testing (pull-out strength, insertion torque), the ground truth is the direct physical measurement of these properties, likely using laboratory equipment. For pyrogenicity, the ground truth is established through LAL testing laboratory results. For substantial equivalence, the "ground truth" is the physical and functional characteristics of the device itself compared to a legally marketed predicate device.

    8. The sample size for the training set

    N/A - This is not an AI device, so there is no "training set."

    9. How the ground truth for the training set was established

    N/A - As above, no training set for an AI model.

    Ask a Question

    Ask a specific question about this device

    K Number
    K202259

    Validate with FDA (Live)

    Date Cleared
    2020-10-28

    (79 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Parcus GFS II and GFS Mini are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair.

    The Parcus GFS Naked devices are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair.

    The GFS Ultimate devices are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair.

    The GFS BTB devices are indicated for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair.

    The Parcus ATLAS is indicated for adjunct fixation of the 1st and 2nd metacarpals in CMC arthroplasty.

    The Parcus Actiflip Naked, Actiflip Cinch, and Actiflip Whip are used for fixation of bone or soft tissue to bone, and is intended as fixation post, a distributing suture tension over areas of ligament or tendon repair in the knee, shoulder, and elbow and may include the following indications: anterior cruciate ligament, pectoralis repair (minor), biceps tendon repair and reattachment (distal/proximal), acromioclavicular repair, and ulnar collateral ligament reconstruction.

    The Parcus GFS Ultimate Hip is used for fixation of soft tissue to bone in the hip and is indicated for ligamentum teres reconstruction.

    The Parcus Synd-EZ Ti are intended to be used as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as adjuncts in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting. The Parcus Synd-EZ Ti are intended to provide fixation during the healing process following syndesmotic trauma, such as fixation of the syndesmosis disruptions) in connection with Weber B and Weber C ankle fractures.

    Device Description

    The Parcus GFS II, GFS Mini, GFS Naked, GFS Ultimate, GFS BTB, ATLAS, Synd-EZ Ti, Actiflip Naked, Actiflip Cinch, Actiflip Whip, and GFS Ultimate Hip are a range of devices design for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair. Each device is made from a medical grade titanium alloy, Ti-6AI4V ELI, and may also include an UHMWPE polyblend suture component. Each device is provided sterile.

    AI/ML Overview

    This document is a 510(k) Premarket Notification from Parcus Medical, LLC for several orthopedic fixation devices. It is a submission to the FDA for market clearance, asserting substantial equivalence to previously cleared predicate devices. The primary focus of the performance data in this submission is related to MR Conditional safety, as the devices themselves are largely unchanged from their predicates.

    Here's a breakdown of the requested information based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    FeatureAcceptance Criteria (Implicit for MR Conditional)Reported Device Performance
    Magnetically Induced ForceDevices should not pose a significant hazard due to magnetic forces in specified MR environments.Evaluated; worst-case devices selected for testing; results used to establish MR system conditions for safe use.
    Magnetically Induced TorqueDevices should not pose a significant hazard due to magnetic torque in specified MR environments.Evaluated; worst-case devices selected for testing; results used to establish MR system conditions for safe use.
    Heating by RF FieldsDevices should not heat to an unsafe level in specified MR environments.Evaluated; worst-case devices selected for testing; results used to establish MR system conditions for safe use.
    Image ArtifactDevices should produce acceptable levels of image artifact in specified MR environments, allowing for diagnostic interpretation where relevant.Evaluated; results used to establish MR system conditions for safe use.
    BiocompatibilityShould be biocompatible as per predicate devices.Unchanged from predicate devices.
    PackagingShould have appropriate packaging as per predicate devices.Unchanged from predicate devices.
    Shelf-lifeShould have appropriate shelf-life as per predicate devices.Unchanged from predicate devices.
    Sterilization ProcessShould have an effective sterilization process as per predicate devices.Unchanged from predicate devices.
    LAL Testing (Endotoxin)Endotoxin levels within acceptable limits.Tested on representative samples; concluded no additional concern.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document states that "worst-case devices were selected for testing" for the MR environment evaluation. It does not provide a specific number for the sample size.
    • Data Provenance: The document does not explicitly state the country of origin for the data or whether it was retrospective or prospective. It implies the testing was conducted by or for Parcus Medical, LLC, which is based in Sarasota, Florida, USA. The testing was likely prospective for the specific purpose of this submission (determining MR Conditional status).

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This type of information is not applicable to this submission. The "ground truth" here is based on objective physical measurements of the devices' interaction with an MR environment and chemical testing (LAL), not on expert subjective interpretation or clinical outcomes. The evaluation against standards like the "FDA Guidance Document - Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment and applicable ASTM standards" serves as the "ground truth" criteria.

    4. Adjudication Method for the Test Set

    This is not applicable as the evaluation is based on objective physical and chemical testing against established standards, not on expert consensus or adjudication of subjective data.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is relevant for diagnostic imaging AI algorithms, not for the safety assessment of orthopedic implants.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No, a standalone algorithm performance study was not done. This submission is for medical devices (implants), not for an algorithm or AI system.

    7. The Type of Ground Truth Used

    The "ground truth" for the performance evaluation in this submission is based on:

    • Physical measurements and compliance with established standards: For MR Conditional safety, this refers to measurements of magnetic force, torque, and heating in MR environments, assessed against the "FDA Guidance Document - Establishing Safety and Compatibility of Passive Implants in the Magnetic Resonance (MR) Environment and applicable ASTM standards."
    • Chemical testing results: For LAL (Limulus Amebocyte Lysate) testing, the ground truth is the measured endotoxin level against a defined acceptable limit.
    • Predicate device characteristics: For biocompatibility, packaging, shelf-life, and sterilization, the "ground truth" is that these aspects remain unchanged and were previously deemed acceptable based on predicate devices.

    8. The Sample Size for the Training Set

    This is not applicable. This submission is for medical devices (implants) and does not involve an AI algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This is not applicable for the reason stated above.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201083

    Validate with FDA (Live)

    Date Cleared
    2020-07-23

    (91 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Parcus Miti Suture Anchors are indicated for attachment of soft tissue to bone. This product is intended for the following indications:
    Shoulder: Rotator Cuff Repair, Acromioclavicular Separation Repair, Bankart Lesion Repair, Biceps Tenodesis, Capsular Shift or Capsulolabral Reconstruction, Deltoid Repair, SLAP Lesion Repair.
    Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Extra Capsular Reconstruction, Iliotibial Band Tenodesis, Patellar Ligament and Tendon Avulsion Repair.
    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Midfoot Reconstruction, Achilles Tendon Repair, Hallux Valgus Reconstruction, Metatarsal Ligament Repair
    Elbow: Tennis Elbow Repair, Biceps Tendon Reattachment.
    Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, TFCC.

    The Parcus V-lox™ Titanium Suture Anchors are indicated for attachment of soft tissue to bone. This product is intended for the following indications:
    Shoulder: Rotator Cuff Repair, Acromioclavicular Separation Repair, Bankart Lesion Repair, Biceps Tenodesis, Capsular Shift or Capsulolabral Reconstruction, Deltoid Repair, SLAP Lesion Repair.
    Knee: Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Extra Capsular Reconstruction, Iliotibial Band Tenodesis, Patellar Ligament and Tendon Avulsion Repair.
    Foot/Ankle: Lateral Stabilization, Medial Stabilization, Midfoot Reconstruction, Achilles Tendon Repair, Hallux Valgus Reconstruction, Metatarsal Ligament Repair
    Elbow: Tennis Elbow Repair, Biceps Tendon Reattachment.
    Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, TFCC.
    Hip: Acetabular Labral Repair

    Device Description

    The Parcus Miti and V-lox Titanium Suture Anchors are a family of threaded, tapered fasteners for use in attachment of soft tissue to bone. The device is made from a Titanium alloy, Ti-6AI-4V ELI (ASTM F136). It comes preloaded with either one, two, or three strands of sutures between 3-0 and #2 in size or suture tape between 1.4mm and 2.0mm wide. The suture and suture tape is available either with or without attached needles, and is provided sterile and attached to a driver. The Miti Suture Anchor is available in diameters of 2.0mm, 2.5mm and the V-lox Titanium Suture Anchor is available in diameters of 4.5mm, 5.0mm, 5.5mm and 6.5mm.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for Parcus V-lox Titanium Suture Anchors and Parcus Miti Suture Anchors. It does not contain information about acceptance criteria or a study proving the device meets specific performance criteria in the way typically associated with a new AI/software device. Instead, this document is a regulatory submission demonstrating substantial equivalence to previously cleared devices.

    Here's an analysis of why the requested information cannot be fully provided based on the input:

    • Type of Device: The devices are physical implants (suture anchors), not an AI or software device. The "performance data" section focuses on MR (Magnetic Resonance) compatibility, which is a safety evaluation for implanted medical devices, not an evaluation of diagnostic or treatment efficacy based on AI algorithms.
    • Regulatory Pathway: A 510(k) submission primarily aims to demonstrate "substantial equivalence" to a predicate device, meaning it is as safe and effective as a device already on the market. This often involves comparing device specifications, materials, and intended use, and sometimes includes performance testing for specific safety attributes (like MR compatibility here), rather than clinical efficacy studies against acceptance criteria for a new mode of action.

    Given this context, I will address the requested points as much as possible, clarifying where the information is not present due to the nature of the submission.


    Acceptance Criteria and Device Performance

    • 1. A table of acceptance criteria and the reported device performance

    The document does not provide a table of acceptance criteria for diagnostic performance or treatment efficacy. The only performance data presented relates to MR compatibility.

    Acceptance Criterion (MR Compatibility)Reported Device Performance (MR Conditional)
    Magnetically Induced ForceEvaluated
    Magnetically Induced TorqueEvaluated
    Heating by RF FieldsEvaluated
    Image ArtifactEvaluated
    Overall ConclusionThe devices were determined to fit the definition of MR Conditional, establishing suitability for safe use under specific MR system conditions. (Specific numerical limits not provided in this summary)
    • 2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document mentions "worst-case devices were selected for testing" for MR compatibility. However, it does not specify the exact sample size (number of devices tested) for the MR evaluation. There is no mention of data provenance (e.g., country of origin, retrospective/prospective) as this was a laboratory-based physical test on the device itself, not a clinical data study.

    • 3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This point is not applicable. The MR compatibility testing is a physical measurement of the device's interaction with an MR environment, not a diagnostic or clinical assessment requiring expert consensus or ground truth in the medical sense.

    • 4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This point is not applicable. Adjudication methods are relevant for studies involving human interpretation or clinical outcomes, not for physical performance testing like MR compatibility.

    • 5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This point is not applicable. The device is a physical medical implant (suture anchor), not an AI or software system that would involve human readers or AI assistance.

    • 6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This point is not applicable. The device is a physical medical implant, not an algorithm.

    • 7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    For the MR compatibility testing, the "ground truth" would be the scientifically established physical limits and behaviors for MR safety as defined by FDA guidance and ASTM standards. It's a technical standard, not a medical "ground truth" derived from patient data.

    • 8. The sample size for the training set

    This point is not applicable. There is no training set mentioned or implied as this is not an AI/machine learning device.

    • 9. How the ground truth for the training set was established

    This point is not applicable. There is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193295

    Validate with FDA (Live)

    Date Cleared
    2020-01-23

    (57 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Parcus Knotless PEEK and PEEK CF Push-In Suture Anchors are indicated for attachment of soft tissue to bone. This product is intended for the following indications:
    Shoulder Rotator Cuff Repair, Acromioclavicular Separation Repair, Bankart Lesion Repair, Biceps Tenodesis, Capsular Shift or Capsulolabral Reconstruction, Deltoid Repair, SLAP Lesion Repair.
    Knee Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Extra Capsular Reconstruction, lliotibial Band Tenodesis, Patellar Ligament and Tendon Avulsion Repair.
    Foot/Ankle Lateral Stabilization. Medial Stabilization. Midfoot Reconstruction. Achilles Tendon Repair, Hallux Valgus Reconstruction, Metatarsal Ligament Repair.
    Elbow Tennis Elbow Repair, Biceps Tendon Reattachment.
    Hand/Wrist Scapholunate Ligament Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, TFCC.
    Hip Acetabular Labral Repair

    Device Description

    The Parcus Knotless PEEK and PEEK CF Push-In Suture Anchors are tapered fasteners with an eyelet at the distal tip and circumferential barbs used to secure the implant in bone and provide suture fixation without the need for knot tying. After passing through the target tissue, sutures are passed through the distal eyelet of the anchor, it is pounded into place which secures both the anchor and the suture in place. This is the same design and insertion technique as the predicate Knotless PEEK CF Push-In Suture Anchors. The predicate device was originally provided in 3.5mm and 4.5mm diameters and this 510(k) will expand that size range to include 2.8mm and 5.5mm options. The Knotless PEEK Push-In Suture Anchor is made from polyetheretherketone (PEEK) and the Knotless PEEK CF Push-In Suture Anchors are made from carbon fiber polyetheretherketone (PEEK CF). The Parcus predicate device is comprised of carbon fiber polyetheretherketone (PEEK CF). The Parcus reference devices consist of devices comprised of either PEEK or PEEK CF. The Parcus Knotless PEEK and PEEK CF Push-In Suture Anchors are provided sterile with a single use suture passer and either with or without a single use inserter.

    AI/ML Overview

    This document describes the Parcus Knotless PEEK and PEEK CF Push-In Suture Anchors. The provided text is a 510(k) summary for a medical device and, as such, does not contain information about acceptance criteria or a detailed study proving the device meets those criteria in the context of diagnostic performance or AI model evaluation.

    The "Summary Performance Data" section briefly mentions that the device was evaluated for "strength and elongation under cycle loading and ultimate failure conditions" and that "Results were compared with test data for the predicate device and demonstrated substantial equivalency." This indicates mechanical testing was performed to demonstrate that the new device performs comparably to an existing, legally marketed device (the predicate).

    However, the document does not provide specific numerical acceptance criteria for these mechanical tests, nor does it detail the specific results. It also does not involve any AI, image analysis, or human reader performance evaluation, and therefore, many of the requested points are not applicable.

    Below is an attempt to address the request based only on the provided text.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Implied)Reported Device Performance
    Mechanical strength under cycle loading comparable to predicateEvaluation of strength and elongation under cycle loading demonstrated substantial equivalency with the predicate device.
    Mechanical strength under ultimate failure conditions comparable to predicateEvaluation of ultimate failure conditions demonstrated substantial equivalency with the predicate device.
    Biocompatibility comparable to predicate and reference devicesAssessment of biocompatibility did not raise concerns, as materials and manufacturing are same/similar to predicate/reference devices.
    Endotoxin levels meet specifications (non-pyrogenic)LAL testing concluded the device meets endotoxin limit specifications and does not raise concerns regarding pyrogenicity.

    2. Sample size used for the test set and the data provenance
    The document does not specify the sample size for the mechanical performance tests. The data provenance is not explicitly stated in terms of country of origin, but it is implied to be from internal testing conducted by Parcus Medical, LLC. The testing is for a new device attempting to prove substantial equivalence to a predicate, so it is a prospective evaluation of the new device's performance against existing data or concurrent testing of the predicate.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
    Not applicable. This is a mechanical device performance study, not a study requiring expert interpretation of diagnostic data.

    4. Adjudication method for the test set
    Not applicable. This is a mechanical device performance study.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not applicable. This document pertains to a physical medical device (suture anchors) and does not involve AI, image analysis, or human reader studies.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not applicable. This document pertains to a physical medical device.

    7. The type of ground truth used
    For the mechanical tests, the "ground truth" (or reference) was the performance data of the predicate device (Parcus Knotless PEEK CF Suture Anchors, K113730) for "strength and elongation under cycle loading and ultimate failure conditions." For biocompatibility, the ground truth was based on the known materials and manufacturing processes of the predicate and reference devices. For pyrogenicity, it was an endotoxin limit specification.

    8. The sample size for the training set
    Not applicable. This document describes the testing of a physical medical device and does not involve a "training set" in the context of machine learning or AI.

    9. How the ground truth for the training set was established
    Not applicable. This document describes the testing of a physical medical device and does not involve a "training set" in the context of machine learning or AI.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192750

    Validate with FDA (Live)

    Date Cleared
    2020-01-16

    (108 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Parcus AcTiFlip is used for fixation of bone or soft tissue to bone, and is intended as fixation post, a distribution bridge, or for distributing suture tension over areas of ligament or tendon repair in the knee, shoulder, and elbow and may include the following indications; anterior cruciate ligament, posterior cruciate ligament, pectoralis repair (minor/major), biceps tendon repair and reattachment (distal/proximal), acromioclavicular repair, and ulnar collateral ligament reconstruction.

    The Parcus GFS Ultimate Hip is used for fixation of soft tissue to bone in the hip and is indicated for ligamentum teres reconstruction.

    Device Description

    The Parcus AcTiFlip and GFS Ultimate Hip devices are designed for use in the fixation of ligaments and tendons in patients requiring ligament or tendon repair. The devices are made from medical grade titanium and UHMWPE suture and are provided sterile.

    AI/ML Overview

    This FDA 510(k) summary provides limited information regarding the acceptance criteria and study data for the Parcus Medical AcTiFlip and GFS Ultimate Hip devices.

    Here's an analysis based on the provided text, highlighting where information is available and where it's missing for your specific requests:

    Summary of Acceptance Criteria and Device Performance (Limited Information):

    Acceptance Criteria CategoryAcceptance Criteria (from document)Reported Device Performance (from document)
    Mechanical Strength- Meet or exceed predicate device performance for strength- "Results were compared with test data for the predicate device and demonstrated substantial equivalency." (Specifically for strength under cycle loading and ultimate failure conditions)
    Elongation- Meet or exceed predicate device performance for elongation- "Results were compared with test data for the predicate device and demonstrated substantial equivalency." (Specifically for elongation under cycle loading and ultimate failure conditions)
    Pyrogenicity / Endotoxin- Meet endotoxin limit specifications- "It was concluded that both the Parcus AcTiFlip and the GFS Ultimate Hip devices meet endotoxin limit specifications and do not raise any additional concerns regarding pyrogenicity."

    Detailed Responses to Specific Questions:

    1. A table of acceptance criteria and the reported device performance:

      • See the table above. The document primarily states that the devices demonstrate "substantial equivalency" to predicate devices for mechanical properties (strength and elongation) and meet endotoxin specifications. Specific numerical acceptance criteria for strength and elongation are not
        provided
        in this summary, nor are the precise numerical results of the testing.
    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

      • Sample Size: The document states that "testing was conducted on the worst case configurations," but does not specify the sample size (N) for these tests.
      • Data Provenance: Not specified. The document does not mention the country of origin of the data or whether the study was retrospective or prospective. It describes laboratory testing, not clinical studies.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):

      • This is not applicable as the summary describes mechanical and biocompatibility laboratory testing of a medical device (fixation fasteners), not a diagnostic or AI-driven device requiring expert-established ground truth.
    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

      • This is not applicable for mechanical and biocompatibility testing. Adjudication methods are typically used in clinical studies or studies involving human assessment of data.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • This is not applicable. The device (fixation fasteners) is not an AI/diagnostic device that would involve human readers or AI assistance.
    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • This is not applicable. The device is a physical medical implant, not an algorithm.
    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the mechanical tests, the "ground truth" (or more accurately, the reference standard for comparison) was the performance of the predicate device. The testing aimed to show that the new device's performance was "substantially equivalent" to this established predicate.
      • For pyrogenicity, the "ground truth" was established based on endotoxin limit specifications, which are regulatory standards.
    8. The sample size for the training set:

      • This is not applicable. There is no "training set" as this is not an AI/machine learning device. The testing described is for physical properties and biocompatibility.
    9. How the ground truth for the training set was established:

      • This is not applicable for the reasons stated above.

    In summary, the provided FDA 510(k) letter and summary focus on demonstrating substantial equivalence of a physical medical device (bone/soft tissue fixation fasteners) to already cleared predicate devices through mechanical testing and biocompatibility assessments. It does not involve AI, diagnostic algorithms, or human interpretation studies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192964

    Validate with FDA (Live)

    Date Cleared
    2019-11-22

    (30 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Parcus Synd-EZ SS is intended to be used as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as adjuncts in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting.

    The Parcus Synd-EZ SS is intended to provide fixation during the healing process following syndesmotic trauma, such as fixation of the syndesmosis (syndesmosis disruptions) in connection with Weber B and Weber C ankle fractures.

    Device Description

    The Parcus Synd-EZ SS consists of two buttons with suture connecting them and are designed to be used for stabilizing the syndesmosis during healing. Building upon the technology of the Parcus Synd-EZ Ti, the Synd-EZ SS is easily adjusted for length and resists elongation without requiring knots to secure the fixation. The device is made from medical grade titanium, stainless steel and UHMWPE suture and is provided sterile.

    AI/ML Overview

    The provided text describes the Parcus Synd-EZ SS, a medical device intended for fracture repair and syndesmotic fixation. However, the document does not contain the detailed information necessary to fully answer your request regarding acceptance criteria and a study proving the device meets them, especially in the context of AI/algorithm performance.

    Here's what can be extracted and what is missing:


    1. A table of acceptance criteria and the reported device performance

    The document provides a general statement about performance testing but does not offer a table of specific acceptance criteria or quantitative performance metrics.

    Acceptance CriteriaReported Device Performance
    Not explicit. The document states "Devices were evaluated for strength and elongation under cycle loading and ultimate failure conditions." The specific numerical thresholds for these evaluations are not provided."Results were compared with test data for the predicate device and demonstrated substantial equivalency." This indicates the device performed comparably to a predicate device, but specific performance numbers are absent.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not available in the provided text. The document mentions "Devices were assembled to test blocks," implying laboratory testing, but does not specify the sample size or any data provenance related to human subjects or clinical data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not available in the provided text. The device described is a physical implant, not an AI or diagnostic tool that would typically involve expert-established ground truth for performance evaluation in the way you've framed the question. The "testing" referred to is mechanical.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not available in the provided text. Adjudication methods are typically used in clinical trials or studies involving expert interpretation, which is not described here for this mechanical device.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not available in the provided text. The device is a surgical implant and does not involve AI assistance for human readers. Therefore, an MRMC study related to AI effectiveness is not relevant to this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not available in the provided text. The device is a physical implant and does not involve an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for this device's performance would be established through mechanical test standards and material properties. The document states that "Devices were evaluated for strength and elongation under cycle loading and ultimate failure conditions." This implies engineering and biomechanical testing, where the "ground truth" is defined by established engineering principles and benchmarks for such implants. It's not clinical "ground truth" in the sense of diagnosis or pathology.

    8. The sample size for the training set

    This information is not available in the provided text. The document describes a physical medical device, not an AI/machine learning model that would have a "training set."

    9. How the ground truth for the training set was established

    This information is not available in the provided text. This question is not applicable as there is no mention of a training set or AI model.


    Summary of what the document does indicate about performance:

    • Type of Study: Laboratory performance testing ("side by side with the predicate device").
    • Parameters Evaluated: Strength and elongation under cyclic loading and ultimate failure conditions.
    • Conclusion: The device "demonstrated substantial equivalency" to the predicate device (Parcus Synd-EZ Ti, K191783).
    • Other Testing: LAL (Limulus Amebocyte Lysate) testing for pyrogenicity and a biocompatibility assessment in alignment with ISO 10993-1 were also conducted.

    This document focuses on the regulatory clearance for a physical medical device based on substantial equivalence to a predicate device, primarily through mechanical testing and biocompatibility assessments. It does not pertain to the evaluation of AI systems or diagnostic performance as implied by many of your questions.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192824

    Validate with FDA (Live)

    Date Cleared
    2019-11-02

    (31 days)

    Product Code
    Regulation Number
    888.3030
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Parcus Knotless AP Suture Anchors are indicated for the attachment of soft tissue to bone. This product is intended for the following indications:

    ShoulderRotator Cuff Repair, Acromioclavicular Separation Repair,Bankart Lesion Repair, Biceps Tenodesis, Capsular Shift orCapsulolabral Reconstruction, Deltoid Repair, SLAP LesionRepair.
    KneeMedial Collateral Ligament Repair, Lateral Collateral LigamentRepair, Posterior Oblique Ligament Repair, Extra CapsularReconstruction, Iliotibial Band Tenodesis, Patellar Ligament andTendon Avulsion Repair.
    Foot/AnkleLateral Stabilization, Medial Stabilization, Midfoot Reconstruction,Achilles Tendon Repair, Hallux Valgus Reconstruction, MetatarsalLigament Repair.
    ElbowTennis Elbow Repair, Biceps Tendon Reattachment.
    Hand/WristScapholunate Ligament Reconstruction, Ulnar or Radial CollateralLigament Reconstruction, TFCC.
    Device Description

    The Parcus Knotless AP Suture Anchors consist of ribbed anchor bodies in 2.8mm, 4.5mm, and 5.5mm diameters. The anchor body is comprised of Parcus Medical's Advanced Polymer (AP) which is a formulation of ß-TCP and PLGA that has a proven record as a safe and effective absorbable material that has been used in the industry for over 10 years. This material has been developed such that it will maintain the necessary strength to insert successfully and provide the necessary resistance to pull-out long enough for the body to heal and then, over time, break down, be absorbed into the body and eventually replaced by bone. The presence of the osteoconductive ß-TCP will allow for boney replacement at the insertion site as the anchor is absorbed.

    Knotless AP Suture Anchors are provided with a single use suture threader and a single use driver. These devices are individually packaged and provided sterile.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called "Parcus Knotless AP Suture Anchors." It primarily focuses on demonstrating substantial equivalence to a predicate device based on material composition, intended use, indications for use, biocompatibility, and mechanical testing.

    This document describes a physical medical device (suture anchors) and does NOT contain information about an AI/ML-driven device. Therefore, it is impossible to extract the details requested in points 1-9, which relate specifically to the acceptance criteria and study design for an AI/ML device (e.g., test set sample size, expert ground truth, MRMC studies, standalone performance, training set details).

    The "Summary Performance Data" section mentions "pull-out strength and cyclic loading studies" and comparison with "published acceptance criteria," but these are engineering performance metrics for a physical implant, not diagnostic or clinical performance metrics for an AI/ML algorithm.

    In summary, the provided text does not describe an AI/ML device or its associated acceptance criteria and validation study as typically found in submissions for such technologies.

    Ask a Question

    Ask a specific question about this device

    K Number
    K191783

    Validate with FDA (Live)

    Date Cleared
    2019-10-09

    (98 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Parcus Synd-EZ Ti is intended to be used as an adjunct in fracture repair involving metaphyseal and periarticular small bone fragments where screws are not indicated, and as adjuncts in external and intramedullary fixation systems involving plates and rods, with fracture braces and casting.

    The Parcus Synd-EZ Ti is intended to provide fixation during the healing process following syndesmotic trauma, such as fixation of the syndesmosis (syndesmosis disruptions) in connection with Weber B and Weber C ankle fractures.

    Device Description

    The Parcus Synd-EZ Ti consists of two buttons with suture connecting them and are designed to be used for stabilizing the syndesmosis during healing. Building upon the technology of the Parcus GFS Ultimate, the Synd-EZ Ti is easily adjusted for length and resists elongation without requiring knots to secure the fixation. The device is made from medical grade titanium and UHMWPE suture and is provided sterile.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the Parcus Synd-EZ Ti device:

    Based on the provided FDA 510(k) summary (K191783), there are no mentions of AI assistance, expert readers, or ground truth establishment in a medical imaging context. This document describes a medical device, the Parcus Synd-EZ Ti, which is a hardware implant for syndesmosis fixation, and not a software device or an AI/ML algorithm.

    Therefore, many of the requested criteria related to AI/ML software performance are not applicable to this submission. The 510(k) summary focuses on the physical and mechanical properties of the device.

    Here's the information that can be extracted from the provided text, adapted to the closest relevant categories:


    1. A table of acceptance criteria and the reported device performance

    Acceptance Criteria CategorySpecific Criteria (Implied)Reported Device Performance
    Mechanical StrengthResist elongation under cyclic loading conditions.Evaluated for strength and elongation under cycle loading and ultimate failure conditions. Results "demonstrated substantial equivalency" to the predicate device.
    Ultimate FailureWithstand ultimate failure conditions.Evaluated for strength and elongation under cycle loading and ultimate failure conditions. Results "demonstrated substantial equivalency" to the predicate device.
    Material BiocompatibilityMade from medical grade materials.Made from medical grade titanium and UHMWPE suture.
    PyrogenicityNon-pyrogenic.LAL testing conducted on a representative device concluded that the Synd-EZ Ti "does not raise any additional concerns regarding pyrogenicity."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample Size: The document states "Devices were assembled to test blocks" and "testing was conducted on the worst case configurations." It does not specify a numerical sample size (e.g., number of devices tested).
    • Data Provenance: Not explicitly stated, but it's implied that the testing was conducted by the manufacturer, Parcus Medical, LLC, which is based in Sarasota, Florida, USA. The testing is prospective for the purpose of demonstrating substantial equivalence to the predicate.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This device is a mechanical implant, not a diagnostic imaging aid. Ground truth was established by direct physical measurement and material testing, not expert interpretation of medical images.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not a study involving human readers or interpretation. The performance data is derived from mechanical testing.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted device, nor does the document describe any human reader effectiveness studies.


    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    Not applicable. This device is not an algorithm. Performance refers to the physical characteristics of the implant.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for this device's performance would be the direct physical and mechanical properties measured during the testing (e.g., actual tensile strength, elongation, pyrogenicity test results). These are objective, quantifiable measurements from material science and engineering, not medical interpretation categories.


    8. The sample size for the training set

    Not applicable. This is not an AI/ML device that requires a training set.


    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190375

    Validate with FDA (Live)

    Device Name
    Parcus ATLAS
    Date Cleared
    2019-04-24

    (64 days)

    Product Code
    Regulation Number
    888.3040
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Parcus ATLAS is indicated for CMC Arthroplasty.

    Device Description

    The Parcus ATLAS is designed for use in CMC arthroplasty. The ATLAS is comprised of medical grade titanium, PEEK-OPTIMA™ from Invibio™, and UHMWPE. The ATLAS is provided sterile.

    AI/ML Overview

    The documents provided do not contain information related to software or AI/ML-driven devices. The submission (K190375) describes a mechanical medical device, the Parcus ATLAS, which is indicated for CMC (Carpometacarpal) arthroplasty. This device is composed of titanium, PEEK-OPTIMA™, and UHMWPE.

    Therefore, the requested information regarding acceptance criteria and studies proving the device meets those criteria, specifically concerning AI/ML performance metrics, sample sizes for test and training sets, expert qualifications, adjudication methods, MRMC studies, or standalone algorithm performance, cannot be extracted from this text. The provided text primarily focuses on substantial equivalence to predicate devices, material composition, manufacturing, sterilization, and basic mechanical performance testing (tensile loading).