Found 2 results

510(k) Data Aggregation

    K Number
    K160466

    Validate with FDA (Live)

    Date Cleared
    2016-03-23

    (33 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Response 5.5 Spine System is intended for immobilization of the posterior, noncervical spine in skeletally mature patients as an adjunct to the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.

    When used for posterior non-cervical pedicle screw fixation in pediatic patients, the Response 5.5 Spine System implants are indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis. The Response 5.5 Spine System is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    Device Description

    The Response 5.5 Spine System consists of longitudinal members (rods), anchors (hooks and screws), interconnection components (rod-to-rod and anchor-to-rod connectors) and fasteners in a variety of sizes to accommodate differing anatomic requirements. No accessories are offered with the system. The purpose of this Pre-Market Notification is to expand the Indications of Use of the system per product codes NKB, OSH, MNH, MNI, KWP.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    Acceptance Criteria and Device Performance for OrthoPediatrics Corp. Response 5.5 Spine System

    Device Name: Response 5.5 Spine System
    Submission Type: 510(k) Premarket Notification (K160466)

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    Mechanical Testing
    Axial Grip (ASTM F1717)Not explicitly detailed, but implied to be within acceptable ranges for spinal implant constructs.The subject devices met the pre-determined acceptance criteria for all tests.
    Axial Torsion (ASTM F1717)Not explicitly detailed, but implied to be within acceptable ranges for spinal implant constructs.The subject devices met the pre-determined acceptance criteria for all tests.
    Flexion Extension Static Testing (ASTM F1717)Not explicitly detailed, but implied to be within acceptable ranges for spinal implant constructs.The subject devices met the pre-determined acceptance criteria for all tests.
    Flexion Extension Fatigue (ASTM F1717)Not explicitly detailed, but implied to be within acceptable ranges for spinal implant constructs.The subject devices met the pre-determined acceptance criteria for all tests.
    Construct Static Compression Bending (ASTM F1798)Not explicitly detailed, but implied to be within acceptable ranges for interconnection mechanisms.The subject devices met the pre-determined acceptance criteria for all tests.
    Construct Static Torsion (ASTM F1798)Not explicitly detailed, but implied to be within acceptable ranges for interconnection mechanisms.The subject devices met the pre-determined acceptance criteria for all tests.
    Construct Compression Bending Fatigue (ASTM F1798)Not explicitly detailed, but implied to be within acceptable ranges for interconnection mechanisms.The subject devices met the pre-determined acceptance criteria for all tests.
    BiocompatibilityAdherence to "Flow Chart for the Selection of Toxicity Tests for 510(K)s per FDA Guidance #G95-1 'Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing', May 1, 1995" or demonstration of identical materials and processing to existing biocompatible predicates.Biocompatibility testing was not required because the materials are identical to previously cleared predicate systems (Response 5.5/6.0 Spine System (K150600) and Response Spine System (K130655)) and have a well-characterized history of clinical use.

    Summary of Acceptance: The device met the pre-determined acceptance criteria for all mechanical tests and biocompatibility was deemed not required due to material equivalence to established predicates and FDA guidance.

    Study Details

    The provided document describes a 510(k) Premarket Notification, which primarily focuses on demonstrating substantial equivalence to legally marketed predicate devices, rather than a clinical effectiveness study. Therefore, the "study" referred to here is the set of tests and comparisons conducted to support this substantial equivalence.

    1. Sample sized used for the test set and the data provenance:

    • Test Set Sample Size: The document does not specify a "sample size" in terms of patient data. For mechanical testing, the sample size would refer to the number of device components or constructs tested. This information is not explicitly provided in the excerpt.
    • Data Provenance: The data is primarily from mechanical testing conducted in a lab environment. The document does not mention any human or animal studies, or data from specific countries of origin for test results. It's an engineering and material science assessment.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. This submission relies on established international standards (ASTM) and FDA guidance for mechanical testing and biocompatibility rather than expert consensus on a test set's ground truth. The "ground truth" here is the adherence to these accepted scientific and regulatory standards.

    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable. Adjudication methods like 2+1 or 3+1 typically apply to clinical studies where human reviewers assess ambiguous cases. In this context, the assessment is based on objective measurements against pre-defined engineering standards.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. An MRMC comparative effectiveness study was not performed. This type of study is relevant for AI-powered diagnostic or assistive devices, which is not what the Response 5.5 Spine System (a pedicle screw spinal system) is.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • No. This is a physical implantable device, not a software algorithm or AI device. Therefore, the concept of "standalone performance" for an algorithm is not applicable.

    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For Mechanical Testing: The "ground truth" is defined by the acceptance criteria within the international standards ASTM F1717 and ASTM F1798 and presumably internal specifications aligned with these standards. These standards prescribe methodologies and expected performance characteristics for spinal implants.
    • For Biocompatibility: The "ground truth" is based on FDA guidance (FDA Guidance #G95-1 and ISO-10993) and the historical clinical safety record of the specific materials (Titanium Alloy and CoCr) that match those used in legally marketed predicate devices.

    7. The sample size for the training set:

    • Not applicable. This is a physical medical device, not a machine learning model. Therefore, there is no "training set."

    8. How the ground truth for the training set was established:

    • Not applicable. As there is no training set for a machine learning model, this question is not relevant.
    Ask a Question

    Ask a specific question about this device

    K Number
    K150600

    Validate with FDA (Live)

    Date Cleared
    2015-05-22

    (73 days)

    Product Code
    Regulation Number
    888.3070
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Response 5.5/6.0 Spine System is intended for immobilization of the posterior, non-cervical spine in skeletally mature patients as an adjunct to the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.

    When used for posterior non-cervical pedicle screw fixation in pediatric patients, the Response 5.5/6.0 Spine System implants are indicated as an adjunct to treat adolescent idiopathic scoliosis. The Response 5.5/6.0 Spine System is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

    Device Description

    The Response 5.5/6.0 Spine System consists of longitudinal members (rods), anchors (hooks and screws), interconnection components (rod-to-rod and anchor-to-rod connectors) and fasteners in a variety of sizes to accommodate differing anatomic requirements.

    No accessories are offered with the system.

    AI/ML Overview

    The provided document is a 510(k) Pre-Market Notification for a medical device called the "Response 5.5/6.0 Spine System." This document focuses on demonstrating substantial equivalence to predicate devices, primarily through mechanical testing and a review of material biocompatibility. It is not a study proving the device meets acceptance criteria related to AI or algorithmic performance, as the device itself is a physical spinal implant system, not a software algorithm.

    Therefore, the requested information about AI/algorithmic performance, sample sizes for test/training sets, data provenance, number of experts, adjudication methods, MRMC studies, standalone performance, and ground truth types and establishment are not applicable to this document. These questions are designed for evaluating AI/software as a medical device (SaMD), which this spinal system is not.

    However, I can extract information regarding the device's mechanical acceptance criteria and the study (testing) performed to meet them.

    Here's the relevant information paraphrased from the document:

    1. Table of Acceptance Criteria and Reported Device Performance (Mechanical Testing)

    The document states that the acceptance criteria for all mechanical tests were pre-determined. While the specific numerical acceptance values are not provided in this summary, the document explicitly states:

    Acceptance Criteria TypeReported Device Performance
    Mechanical PerformanceThe subject devices met the pre-determined acceptance criteria for all tests, including Axial Grip, Axial Torsion, Flexion Extension Static Testing, Flexion Extension Fatigue, Construct Static Compression Bending, Construct Static Torsion, and Construct Compression Bending Fatigue.

    2. Sample size used for the test set and the data provenance

    • Sample Size for Test Set: Not explicitly stated in terms of number of implants tested, but testing was performed on "subject devices" as per ASTM F1717 and ASTM F1798.
    • Data Provenance: Not applicable in the context of clinical data provenance (e.g., country of origin, retrospective/prospective clinical study) as this refers to mechanical performance testing of the device itself, not clinical data from patients. The testing was conducted by OrthoPediatrics Corp.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • This question is not applicable as there is no "ground truth" established by human experts for mechanical testing of a physical medical device in the way it would be for an AI algorithm evaluating medical images. The "ground truth" for mechanical testing is defined by the standardized test methods (ASTM F1717 and ASTM F1798) and their associated performance specifications.

    4. Adjudication method for the test set

    • This question is not applicable for mechanical testing. Performance is evaluated against objective, pre-defined engineering standards established by ASTM.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This question is not applicable. This is a physical spinal implant, not an AI software.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • This question is not applicable. This is a physical spinal implant, not an AI software.

    7. The type of ground truth used

    • For mechanical performance: The "ground truth" (or standard of performance) is defined by the requirements of the recognized industry standards: ASTM F1717 - "Standard Test Methods for Spinal Implant Constructs in a Vertebrectomy Model" and ASTM F1798 - "Standard Guide to Evaluating the Static and Fatigue Properties of Interconnection Mechanisms and Subassemblies Used in Spinal Arthrodesis Implants."

    8. The sample size for the training set

    • This question is not applicable. There is no "training set" for a physical medical device. The design, materials, and manufacturing processes are developed based on engineering principles and prior knowledge, not machine learning training.

    9. How the ground truth for the training set was established

    • This question is not applicable for the same reasons as above.