Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K242397

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-10-25

    (73 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Norlase LYNX Photocoagulator is intended to be used in ophthalmic laser procedures including retinal and macular photocoagulation, iridotomy, and trabeculoplasty.

    Device Description

    Not Found

    AI/ML Overview

    I apologize, but the provided text from the FDA 510(k) clearance letter for the LYNX Photocoagulator does not contain information on the acceptance criteria and the study that proves the device meets those criteria.

    This document is a clearance letter, which confirms that the FDA has found the device to be substantially equivalent to a legally marketed predicate device. It briefly touches on device classification, regulations, and mentions the intended use of the device.

    To describe the acceptance criteria and the study that proves the device meets them, I would need access to the actual 510(k) submission summary or a more detailed technical report which would typically include sections on performance testing, design specifications, and clinical or non-clinical data. The provided document only details the regulatory clearance, not the underlying performance data.

    Therefore, I cannot fulfill your request with the given input.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222157

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-10-11

    (83 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Norlase ECHO Photocoagulator is intended to be used in ophthalmic laser procedures including retinal and macular photocoagulation, iridotomy, and trabeculoplasty.

    Device Description

    The Norlase ECHO device is a modification to the Norlase Leaf Photocoagulator (K190083) and also consists of a laser unit, a wireless control unit and a foot switch. The laser unit must be attached to the Doctors own compatible slit lamp to work as a full system. The Norlase ECHO Device is controlled by the wireless control unit that allows the Doctor to set the desired parameters for treatment. A shrouded foot switch is connected to the laser unit to control the emission of laser light. The Doctor will use the user supplied slit lamp to identify the area to be treated and after setting the appropriate parameters, will target the ophthalmic tissue utilizing a visible red aiming laser. The foot switch will be pressed by the Doctor to deliver therapeutic green laser light to the target tissue. A mechanically fixed eve safety filter is built into the Laser Unit to protect the Doctor from any stray or reflected green laser light travelling along the visual path of the slit lamp. The Norlase Echo device consists of the following modules: Laser unit + power supply + power cable, Foot switch + cable, Control unit (Tablet computer with installed Software) + power supply + USB cable. The wavelength of the treatment beam is centered around ~520 nm, and the maximum optical output power is 1.5W (to tissue). The laser is operated in continuous wave (CW) mode and electronically pulsing the laser output power to achieve durations from 50 usec to 1 second. Laser parameters are controlled by a Control Unit (tablet) User Interface that utilizes a touchscreen control to select or change a setting. Voice Control of selected parameters is an optional feature that can assist in the increment or selection of a parameter. The ECHO modification consists of capability for pattern scanning.

    AI/ML Overview

    The provided text is a 510(k) summary for the Norlase ECHO Photocoagulator. This document focuses on demonstrating substantial equivalence to a predicate device, rather than proving a device meets specific acceptance criteria based on clinical study performance. Medical devices cleared via the 510(k) pathway often rely on bench testing, engineering verification, and comparison to legally marketed predicate devices, especially when the changes are related to technological characteristics rather than new indications for use or significant safety/efficacy concerns.

    The document explicitly states: "Clinical data was not required for this type of device." This means there was no clinical study conducted to evaluate device performance against specific acceptance criteria in a human population. Instead:

    • Engineering verification and validation were performed.
    • Bench testing was conducted to demonstrate compliance with various safety and performance standards.
    • Software testing was performed.
    • Human factors evaluation was carried out in a simulated use environment.
    • The key finding for substantial equivalence is that the differences (primarily the addition of pattern scanning) do not introduce new issues related to safety and efficacy.

    Therefore, I cannot provide an answer that aligns with the structure of your requested output (acceptance criteria and proof from a clinical study) because the provided document does not contain that information. The performance data discussed is limited to engineering and bench testing, not clinical performance against ground truth established by experts.

    To directly address your points based on the provided text, while acknowledging the lack of clinical study data:

    1. A table of acceptance criteria and the reported device performance:
      The document does not present "acceptance criteria" in the sense of a clinical study endpoint (e.g., sensitivity, specificity, accuracy). Instead, it lists standards the device complies with and types of testing performed. The "performance" is stated as successful verification and validation according to documented procedures.

      Acceptance Criteria (as implied by testing)Reported Device Performance (as stated in submission)
      Compliance with electrical and mechanical safety standards (e.g., IEC 60601-1)Successful bench testing for electrical and mechanical safety in compliance with cited standards.
      Compliance with laser performance standards (e.g., IEC 60825-1)Successful bench testing for laser performance in compliance with cited standards.
      Compliance with ophthalmic device standards (e.g., ISO 15004-2)Successful bench testing specific to ophthalmic devices in compliance with the cited standard.
      Specified performance of pattern scanning featureSuccessful bench testing for verification and validation of the specified performance of the pattern scanning feature.
      Software compliance (e.g., IEC 62304), including cybersecuritySuccessful software testing (verification and validation), including test cases for off-the-shelf software and cybersecurity features.
      Usability in simulated environmentSuccessful human factors evaluation to demonstrate usability in a simulated use environment when used by healthcare professionals.
      Risk mitigation effectiveness (ISO 14971)All risk mitigations satisfactorily verified and validated.
      Device meets design requirements and user needsResults of performance testing demonstrated that the device met the design requirements and user needs.
    2. Sample size used for the test set and the data provenance: Not applicable. No clinical test set. The data provenance is from internal company verification and validation testing, not from a patient population.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No clinical test set requiring expert ground truth in this context.

    4. Adjudication method for the test set: Not applicable. No clinical test set.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a laser photocoagulator, not an AI diagnostic or assistance tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This is a medical device, not an AI algorithm. Its "performance" is related to its physical and software functionality, which was tested during engineering verification.

    7. The type of ground truth used: For the engineering and bench testing, "ground truth" would be established by validated measurement equipment, engineering specifications, and compliance with industry standards. For instance, a power meter would provide the "ground truth" for laser power output, and test protocols would define the "ground truth" for software functionality. There is no clinical "ground truth" (e.g., pathology, outcomes data) relevant to this 510(k) submission.

    8. The sample size for the training set: Not applicable. This is not an AI/ML device that requires a training set.

    9. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K201623

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-08-11

    (57 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator is intended to be used in ophthalmic laser procedures including retinal and macular photocoagulation, iridotomy, and trabeculoplasty.

    Device Description

    The Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator system was developed to utilize advancements in visible green semiconductor (diode) laser technology and to provide a compact design to be incorporated in a lightweight headset worn by the physician. The Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator system consists of an indirect ophthalmoscope headpiece, a headset laser enclosure/casing, a wireless user interface tablet, a footswitch with a battery, and a cable extending from the footswitch to the headset. The Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator system is controlled by the wireless user interface tablet that allows the physician to set the desired parameters for treatment. A shrouded footswitch is connected to the headset to control the emission of laser light and the battery to power the device illumination and laser source. The physician uses the headset to locate the area to be treated and after setting the appropriate parameters, targets the desired tissue utilizing a visible red aiming beam. The foot switch is pressed by the physician to deliver therapeutic green laser light to the targeted tissue. Eye safety filters are built into the laser unit to protect the physician's eyes while providing a clear view of the target area.

    AI/ML Overview

    The document provided is a 510(k) summary for the Norlase Laser Indirect Ophthalmoscope (LIO) Photocoagulator. It focuses on demonstrating substantial equivalence to predicate devices through engineering and bench testing, rather than clinical studies involving human patients or complex algorithms requiring extensive training and test sets.

    Therefore, many of the requested categories related to medical AI/ML device studies (such as sample sizes for test and training sets, data provenance, expert qualifications, adjudication methods, MRMC studies, standalone performance, and specific ground truth types like pathology or outcomes data) are not applicable to the information provided in this document.

    Here's a summary of the information available based on the provided document:

    Acceptance Criteria and Device Performance

    The device underwent verification and validation testing against several international standards. The acceptance criteria were met if the device demonstrated compliance with these standards.

    Acceptance Criteria (Standard Compliance)Reported Device Performance
    ISO 14971: Application of risk management to medical devicesAll identified potential risks were analyzed, and mitigations were implemented, verified, and validated satisfactorily.
    ISO 15004-2: Ophthalmic instruments - Fundamental requirements and test methods - Part 1: Light hazard protectionIncluded in bench testing specific to ophthalmic devices; results demonstrated predetermined acceptance criteria were met.
    IEC 60601-1: Medical electrical equipment - General requirements for basic safety and essential performanceBench testing for electrical and mechanical safety was successfully completed.
    IEC 60601-1-2: Electromagnetic disturbancesBench testing for electrical and mechanical safety was successfully completed.
    IEC 60601-2-22: Medical electrical equipment - Part 2: Particular requirements for the safety of diagnostic and therapeutic laser equipmentBench testing for laser performance was successfully completed.
    IEC 60825-1: Safety of laser products - Part 1: Equipment classification, and requirementsBench testing for laser performance was successfully completed.
    IEC 62304: Medical device software - Software life-cycle processesSoftware testing, including verification and validation, was conducted in compliance with this standard, covering test cases related to off-the-shelf software and cybersecurity features.
    IEC 62366-1: Application of usability engineering to medical devicesHuman factors testing was conducted to demonstrate usability in a simulated use environment when used by healthcare professionals.
    ANSI Z136-1: American National Standard for Safe Use of LasersIncluded in bench testing for laser performance; results demonstrated predetermined acceptance criteria were met.
    ANSI Z136-3: American National Standard for Safe Use of Lasers in Health CareIncluded in bench testing for laser performance; results demonstrated predetermined acceptance criteria were met.
    21 CFR Part 820.30: Design Control processVerification and validation testing was completed in accordance with the company's Design Control process; results demonstrated that predetermined acceptance criteria were met.
    Overall Safety and Efficacy (via Substantial Equivalence)The device demonstrated substantial equivalence to predicate devices (Norlase LEAF Green Laser Photocoagulator K190083 and Iridex TruFocus LIO K170718) with no new issues of safety or efficacy identified.

    Study Details:

    1. Sample size used for the test set and the data provenance: Not applicable. The submission relies on engineering bench tests and simulated use environments, not a clinical test set with patient data.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for expert consensus is not relevant for this type of engineering and human factor testing.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI/ML device requiring MRMC studies for diagnostic or interpretive tasks.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done: Not applicable. The device is a physical laser photocoagulator, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable in the context of clinical gold standard. The "ground truth" here refers to the specifications and safety requirements defined by the various engineering and medical device standards (e.g., laser power output accuracy, electrical safety, software functionality, usability in a simulated environment).
    7. The sample size for the training set: Not applicable. This device does not use a "training set" in the context of machine learning.
    8. How the ground truth for the training set was established: Not applicable.

    Note: The 510(k) summary explicitly states that "Clinical data was not required for this type of device." This further confirms that the assessment of this device for substantial equivalence did not involve clinical studies with human participants that would generate the types of data requested in many of the prompts. The approval is based on engineering, safety, software, and human factors testing demonstrating compliance with established standards and equivalence to predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190083

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-05-03

    (106 days)

    Product Code
    Regulation Number
    886.4390
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Norlase Leaf Photocoagulator is intended to be used in ophthalmic laser procedures including retinal and macular photocoagulation, iridotomy, and trabeculoplasty.

    Device Description

    The Norlase Leaf device consists of a laser unit, a wireless control unit and a foot switch. The laser unit must be attached to the Doctors own compatible slit lamp to work as a full system. The Norlase Leaf Device is controlled by the wireless control unit that allows the Doctor to set the desired parameters for treatment. A shrouded foot switch is connected to the laser unit to control the emission of laser light.

    The Doctor will use the user supplied slit lamp to identify the area to be treated and after setting the appropriate parameters, will target the ophthalmic tissue utilizing a visible red aiming laser. The foot switch will be pressed by the Doctor to deliver therapeutic green laser light to the target tissue. A mechanically fixed eye safety filter is built into the Laser Unit to protect the Doctor from any stray or reflected green laser light travelling along the visual path of the slit lamp.

    The Norlase Leaf device consists of the following modules:

    • -Laser unit + power supply + power cable
    • Foot switch + cable -
    • Control unit (Tablet computer with installed Software) + power supply + USB cable -

    The wavelength of the treatment beam is centered around ~520 nm, and the maximum optical output power is 1.5W (to tissue). The laser is operated in continuous wave (CW) mode and electronically pulsing the laser output power to achieve durations from 50 usec to 1 second. Laser parameters are controlled by a Control Unit (tablet) User Interface that utilizes a touchscreen control to select or change a setting. Voice Control of selected parameters is an optional feature that can assist in the increment or selection of a parameter,

    AI/ML Overview

    The provided text describes the Norlase Leaf Photocoagulator, a medical device, and its acceptance criteria as part of a 510(k) premarket notification to the FDA. The study presented focuses on demonstrating substantial equivalence to a predicate device through verification and validation testing against recognized standards.

    Here's an breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document states that "Every specification of the Leaf Photocoagulator device has been verified and validated according to the company's documented development and test procedures." It then lists the standards used for testing without explicitly detailing specific numerical acceptance criteria. However, the completion of these tests with "successful results" implies that the device met the requirements outlined within those standards.

    Acceptance Criteria CategoryReported Device Performance
    General Safety and Performance- Verified and validated according to Norlase ApS's documented development and test procedures.- Satisfactorily verified and validated all risk mitigations.- Demonstrated that technological differences from the predicate did not result in new issues of safety or efficacy.- Met design requirements and user needs.
    Risk Management- Identified potential risks according to ISO 14971.- Analyzed risks with regard to risk/benefit category.- Implemented and tested mitigations as part of performance testing.
    Electrical and Mechanical Safety- Bench testing for electrical and mechanical safety completed successfully in compliance with ISO 15004-2, IEC 60601-1+, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, ANSI Z136-1, and ANSI Z136-3.
    Laser Performance- Bench testing for laser performance completed successfully in compliance with ISO 15004-2, IEC 60601-1+, IEC 60601-1-2, IEC 60601-2-22, IEC 60825-1, ANSI Z136-1, and ANSI Z136-3.
    Ophthalmic Device Specifics- Bench testing specific to ophthalmic devices completed successfully in compliance with ISO 15004-2.
    Software Performance- Software testing completed successfully, including verification and validation in compliance with IEC 62304.- Included test cases related to off-the-shelf software and cybersecurity features.
    Usability (Human Factors)- Human factors testing completed successfully to demonstrate usability in a simulated use environment when used by healthcare professionals, in compliance with IEC 62366-1.

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not specify a "sample size" in the context of patients or clinical data for testing the device's performance. It states that "Clinical data was not required for this type of device." The testing described is primarily laboratory-based (bench testing) and human factors testing in a simulated environment. Therefore, the "test set" consists of the device itself and its components, subjected to various engineering and usability tests, rather than a dataset of patient images or outcomes.

    The "data provenance" is implied to be internal testing conducted by Norlase ApS, as described by "the company's documented development and test procedures." No specific country of origin or retrospective/prospective nature is applicable in the clinical sense, as no clinical data was used.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    Not applicable. Since no clinical data was used and the testing consisted of bench testing and human factors in a simulated environment, there was no need for expert ground truthing of clinical cases. The "ground truth" for these types of tests would be the specifications and requirements defined by the standards and the device's design documentation.

    4. Adjudication Method for the Test Set

    Not applicable. As no clinical data requiring expert adjudication was used, there was no adjudication method like 2+1 or 3+1. The assessment of test results would be based on comparison against predefined technical specifications and standards (e.g., whether a certain parameter falls within a specified range, or whether a safety feature functions as intended).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The document explicitly states: "Clinical data was not required for this type of device." Therefore, an MRMC comparative effectiveness study was not performed.

    6. If a Standalone (i.e., Algorithm Only Without Human-in-the-Loop Performance) Was Done

    This question seems to be more relevant to AI/CAD devices. The Norlase Leaf Photocoagulator is a direct treatment device, not an AI or diagnostic algorithm. Its performance is evaluated on its technical specifications and safety rather than algorithmic accuracy. The device's "standalone" performance would be its ability to emit laser light at specified parameters and its safety functions, which were evaluated through bench testing. Human factors testing, however, did involve human users (healthcare professionals) interacting with the device to assess usability, but this is distinct from "human-in-the-loop performance" in the context of an AI-assisted diagnostic task.

    7. The Type of Ground Truth Used

    The ground truth for this device's evaluation is primarily based on:

    • Technical specifications and design requirements: The device's ability to operate within specified ranges for wavelength, power output, exposure time, etc.
    • International and national standards: Compliance with standards like ISO 14971, ISO 15004-2, IEC 60601 series, IEC 60825-1, IEC 62304, IEC 62366-1, ANSI Z136-1, and ANSI Z136-3. These standards define acceptable performance and safety criteria.
    • Usability metrics: Outcomes from human factors testing demonstrating that the device can be used safely and effectively by healthcare professionals in a simulated environment.

    8. The Sample Size for the Training Set

    Not applicable. This is not an AI/machine learning device, so there is no training set as would be used for an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1