510(k) Data Aggregation

    K Number
    K152665

    Validate with FDA (Live)

    Date Cleared
    2015-10-26

    (39 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The PolarCath Peripheral Dilatation System's intended use is for the dilatation of stenoses in the peripheral vasculature (iliac, femoral, popliteal, infrapopliteal, renal and subclavian arteries) and for the treatment of obstructive lesions of polytetrafluoroethylene (PTFE) access grafts or native arteriovenous dialysis fistulae. The PolarCath Peripheral Dilatation System is also indicated for post-deployed stent expansion of self-expanding peripheral vascular stents.

    Device Description

    The PolarCath™ Peripheral Dilatation System (PolarCath System) consists of a disposable catheter, a reusable Cryoplasty inflation unit, a disposable nitrous oxide cartridge, and a disposable catheter extension.

    The PolarCath System is designed for dilatation of stenotic lesions in peripheral arteries. The procedure consists of inserting a catheter over a quidewire to the target lesion, attaching the catheter to the catheter extension, attaching the catheter extension to the Inflation Unit, inserting a nitrous oxide cartridge into the Inflation Unit, and inflating the balloon catheter for a set time at a set pressure, using the controls on the Inflation Unit.

    The Cryoplasty inflation unit functions to: a) deliver the inflation media (liquid nitrous oxide) to the balloon; b) control the pressure inside the inner balloon; c) provide a vacuum between the inner and outer balloons; and d) control the length of time for delivery of the inflation media which is manually extracted at the end of the inflation time. The inflation unit monitors the temperature inside the balloon. Excess pressure is prevented by the presence of a relief valve in the inflation unit.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device called the PolarCath Peripheral Dilatation System. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a study proving device performance against specific acceptance criteria in a clinical context.

    Therefore, many of the requested sections regarding clinical study design (sample size, data provenance, expert ground truth, adjudication, MRMC studies) cannot be fulfilled from this document. The document describes non-clinical testing.

    Here's an analysis of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance & The Study that Proves the Device Meets Acceptance Criteria:

    Testing TypeAcceptance Criteria (Implied)Reported Device Performance
    Performance Bench TestingPass all functional testing; meet all product specification requirements; demonstrate equivalent performance to the predicate device.The PolarCath System passed all functional testing and met all product specification requirements and demonstrated equivalent performance to the predicate device.
    Software Verification Validation TestingMeet all requirements of the Software Requirements Specification (SRS).PolarCath System Software met all requirements of the SRS.
    Electromagnetic Compatibility & Electrical SafetyMeet all acceptance criteria in accordance with IEC 60601-1:2005 and IEC 60601-1-2:2007.The PolarCath System met all acceptance criteria in accordance with: IEC 60601-1:2005 and IEC 60601-1-2:2007.

    The Study that Proves the Device Meets Acceptance Criteria:

    The document describes "non-clinical testing" as the study. This involved "Performance Bench Testing," "Software Verification Validation Testing," and "Electromagnetic Compatibility and Electrical Safety" testing. These tests were conducted to demonstrate that the materials chosen, manufacturing processes, and design meet established specifications and do not raise new questions of safety or effectiveness compared to the predicate device. No animal or clinical testing was required or performed for this submission.

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not specified for any of the non-clinical tests. For bench testing, it refers to multiple aspects (Catheter Extension, Inflation Unit, Simulated Use, Transit Testing), implying a series of tests on various components/systems, but specific quantities are not given.
    • Data Provenance: Not applicable as these are non-clinical, bench-level tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable as this was non-clinical and did not involve expert-established ground truth in a clinical diagnostic sense. Performance was measured against engineering specifications and industry standards.

    4. Adjudication method for the test set:

    • Not applicable as this was non-clinical and did not involve expert adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This is a medical device for physical intervention (dilatation system), not an AI-powered diagnostic or assistive tool for human readers.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable as this is a physical medical device, not an algorithm.

    7. The type of ground truth used:

    • For the non-clinical tests, the "ground truth" was established by engineering specifications, product requirements, and recognized industry standards (e.g., IEC 60601-1, IEC 60601-1-2). For the software, it was the Software Requirements Specification (SRS). For performance, it was also equivalent performance to the predicate device.

    8. The sample size for the training set:

    • Not applicable. This is not a machine learning or AI device that would have a training set.

    9. How the ground truth for the training set was established:

    • Not applicable.