510(k) Data Aggregation

    K Number
    K192291

    Validate with FDA (Live)

    Date Cleared
    2020-08-20

    (363 days)

    Product Code
    Regulation Number
    880.5965
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The TidalPort-AP™ Implantable Apheresis Vascular Access Port is indicated for therapies requiring repeated access to the vascular system. The port system can be used for long-term therapeutic apheresis, withdrawal of blood, and infusion of medications, I.V. fluids, parenteral nutrition solutions, blood, and blood products.

    The TidalPort-AP™ Implantable Apheresis Vascular Access Port is indicated for power injection of contrast media. For power injection of contrast media, the maximium recommended infusion rate is 5 mL/s.

    Device Description

    The TidalPort-AP™ has one port size (Standard) and one catheter size (9.6F Polyurethane). It is designed to provide repeated access to the vascular system without the need for repeated venipuncture or the daily care of an external catheter. The TidalPort-APTM is available as a standard profile totally implantable, titanium port-based design and is accessed perpendicularly to the skin like a typical, conventional access port. For the purpose of apheresis, high-flow procedures, it is accessed with using the FDA Cleared 16G or 18G high-flow, non-coring needle Tidal High-Flow Non-coring Needle (K151341).

    The TidalPort-APTM can also be used for routine vascular access infusion or withdrawal using Huber point needle. For power injection infusion procedures, the subject device can be accessed with a power injection rated needle to create a power-injectable system.

    The design of the TidalPort-AP™ utilizes a spherical reservoir and an elongated radius contoured septum to achieve the design purpose of creating a port with a smaller reservoir and less clearance volume. The TidalPort-APTM comes with a number of kit components to aid in the implantation procedure and/or access of the device once implanted. The TidalPort-AP™ and necessary kit components are provided sterile (EtO).

    The overall implanted port system consists of three primary components: the titanium port body with a silicone septum, an attachable radiopaque polyurethane catheter, and a catheter lock which secures the catheter to the port body stem. Once implanted, the method of accessing the subject TidalPort-APTM device is the exact same as the predicate TidalPort™ device. After the implanted device has been identified and access is prepped per institutional policy, the user palpates the uniquely shaped port. Once the port is palpated, providing the location of the septum, the 16G or 18G high-flow needle (K151341) is inserted into the reservoir for apheresis procedure use. After the reservoir floor is reached, the stylet is unlocked and pulled back slightly, and the needle is once again advanced forward until contacting the reservoir floor again. The stylet is then completely removed, leaving the hollow cannula with luer lock fitting in place. After stylet removal, the cannula is attached to the appropriate extension set and secured for the necessary infusion or withdrawal procedure.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the TidalPort-AP™ Implantable Apheresis Vascular Access Port. This document details the device's indications for use, technological characteristics, and non-clinical tests performed to demonstrate substantial equivalence to predicate devices. However, it does not contain the specific information requested in the prompt regarding acceptance criteria, study details, sample sizes, expert qualifications, or ground truth establishment in a format typically used for evaluating the performance of AI/ML-driven devices.

    The document focuses on engineering and material performance testing of a physical medical device, not a diagnostic or prognostic AI/ML algorithm. Therefore, many of the requested fields are not applicable in this context.

    Here's an attempt to extract the relevant information based on the provided text, while acknowledging the limitations:

    1. A table of acceptance criteria and the reported device performance

    The document lists performance tests but does not explicitly state quantitative "acceptance criteria" and "reported performance" in a structured table for each test. It states that "The TidalPort-AP™ Implantable Apheresis Vascular Access Port has met all predetermined acceptance criteria of design verification evaluations through testing examination." This implies that the device did meet the criteria for each performed test.

    Key Performance Tests Mentioned:

    • Catheter to Port Connection
    • Catheter Tensile Strength
    • Radiopacity
    • Gravity Flow Rate
    • Port System Burst under Power Injection
    • Stem Tensile Strength
    • Corrosion Resistance
    • Septum Puncture / Port Leak (w/both Huber point and 2-part large bore needles)
    • Port Clearance Volume
    • Power Injection / Multi-Power Injection
    • Simulated Apheresis Testing- comparison to predicate (PowerFlow)
    • Recirculation, Forward and Reverse Flow- comparison to predicate (PowerFlow)
    • Hemolysis, Forward and Reverse Flow- comparison to predicate (PowerFlow)
    • MRI Safety Testing
    • Packaging Ship Testing
    • Pyrogenicity Testing

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    Not applicable. The document describes bench testing and biocompatibility assessments of a physical device, not a test set of medical images or patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is not an AI/ML diagnostic or prognostic device requiring expert-established ground truth from a test set of data.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is not an AI-assisted diagnostic or prognostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable in the context of AI/ML ground truth. For this physical device, the "ground truth" would be established by validated engineering specifications, accepted international standards (ISO, ASTM), and FDA guidance documents (e.g., FDA Guidance on 510(k) Submissions for Implanted Infusion Ports, standards listed in the document). Performance was evaluated against these predefined physical and functional requirements.

    8. The sample size for the training set

    Not applicable. This is a physical medical device, not an AI/ML algorithm that undergoes a "training set."

    9. How the ground truth for the training set was established

    Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K151341

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2015-08-20

    (93 days)

    Product Code
    Regulation Number
    880.5570
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tidal High-Flow Non-coring Needle is intended for the access of totally implantable vascular access ports (VAPs) to administer I.V. fluids, infusions drugs, and other fluids, and for the administration of blood products, and/or withdrawal of blood as part of the therapy regimen

    Device Description

    The 16Ga and 18Ga Tidal High-Flow Non-coring Needles are an extension of our 19, 20, and 22 gauge non-coring (Huber) Needle line used to access our line of totally implantable access devices. These large gauge non-coring needles incorporate a stylet to render them non-coring. The needle is made from 304 series stainless steel and the hubs are molded from K-Resin and LDPE. All materials are certified USP Class VI and are the exact same materials used to manufacture the products approved under K863721 and K111101. The needles are a disposable product that will be packaged sterile and non-pyrogenic in a single use peel pouch. ETO sterilization will be used. The sterilization will be done in-house using a validated sterilization procedure that is identical to the sterilization method used in previous 510(k) clearances.

    AI/ML Overview

    The provided text describes a 510(k) summary for a medical device called the Tidal High-Flow Non-coring Needle. This document focuses on demonstrating substantial equivalence to predicate devices rather than presenting a study to prove the device meets specific acceptance criteria in the manner that an AI/ML device would be evaluated.

    Therefore, many of the requested details about acceptance criteria, performance studies, sample sizes, expert involvement, and ground truth establishment are not applicable to this type of regulatory submission and device.

    However, I can extract information related to the closest equivalents to "acceptance criteria" and "performance data" as presented in this document.


    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance metrics in the format typically seen for AI/ML device evaluations. Instead, it states that "The Tidal High-Flow Non-coring Needles have met all predetermined acceptance criteria of design verification evaluations through testing examination." The tests performed are listed as:

    Acceptance Criteria (Implied)Reported Device Performance
    Performance with 16Ga high-flow, noncoring needles for septum puncture / port leak to failure (worst case scenario) using a representative sample of ports found in the marketplace.The data collected from the non-clinical tests demonstrated that the functionality and performance characteristics of the Tidal High-Flow Non-coring Needles are comparable to the currently marketed needles based on available literature.
    Performance following 25 punctures with the 16Ga high-flow needle for septum puncture / port leak to failure with a 19Ga Huber point needle using a representative sample of ports found in the marketplace.The data collected from the non-clinical tests demonstrated that the functionality and performance characteristics of the Tidal High-Flow Non-coring Needles are comparable to the currently marketed needles based on available literature. (This is a general statement. Specific numerical results are not provided in this summary.)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Sample size: The document mentions "a representative sample of ports found in the marketplace" was used for the septum puncture/port leak tests. No specific number is provided.
    • Data provenance: Not explicitly stated, however, the testing was conducted by Norfolk Medical, a US-based company, and the FDA guidance referenced is US-based. It is implied to be prospective bench testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. This is a physical device being evaluated via bench testing for mechanical properties and biocompatibility, not an AI/ML device requiring expert consensus for ground truth. The "ground truth" here is the physical performance as observed in the tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. This is not an image review or interpretative task requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This is a physical medical device (needle), not an AI/ML diagnostic or assistive device for human readers.

    6. If a standalone (i.e. algorithm only, without human-in-the-loop performance) was done

    Not applicable. This is a physical medical device. The "standalone" performance refers to the device's inherent mechanical and material properties.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    The "ground truth" for the performance evaluation in this context is based on bench testing results (septum puncture/port leak to failure) and material characterization against established standards (e.g., ISO, USP Class VI for biocompatibility).

    8. The sample size for the training set

    Not applicable. This is not an AI/ML device, so there is no "training set."

    9. How the ground truth for the training set was established

    Not applicable. There is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112713

    Validate with FDA (Live)

    Date Cleared
    2011-10-27

    (38 days)

    Product Code
    Regulation Number
    880.5965
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The SportPort™ is indicated for use when the patient requires the following: repeated access to the vascular system for injections, infusion of drugs, administration of blood or blood products, and/or withdrawal of blood as part of the therapy regimen.

    When used with a power injectable needle infusion set, the SportPort™ is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s with a 19 or 20 gauge non-coring power injectable needle and 2 ml/s with a 22-gauge non-coring power injectable needle.

    Device Description

    The devices are to be used by or on the order of a physician. The SportPort™ line of ports has a body/septum retainer made of either commercially pure grade 2 titanium or medical grade polysulfone and has a body/septum chamber with a silicone rubber septum designed for repeated needle puncturing. The SportPort™ comes in a standard design to enable regular use as an infusion/withdrawal port and has high-pressure injection capability to aid to the delivery of large volumes or special fluids such as contrast media. The shape of the base plate of the port body is triangular and is compatible with most imaging systems. The top of the port (septum retainer) has a raised triangle to secure the septum in place during high-pressure use and to enable palpable or image system identification. The port has six (6) elongated holes for suture fixation to the deep fascia tissue during implantation. The fixation is required to prevent migration or flipping of the port. A catheter comes with the port and is inserted into the vascular system. The catheter is assembled to the port and fed into a venous vessel and fed down into the vascular system. The catheter is radiopaque to enable visualization for proper placement. The kit provided to aid in insertion of the catheter and placement of the port may include items like needles, sheaths, vein picks, guidewires, straighteners, and dilators. Likewise, the cut down kit for port placement may contain most infusion sets with Lucent Non-Coring needles, straight pointed Huber needles, blunt needles, and syringes along with the port and catheter.

    The Norfolk Medical SportPort™ family of ports provides a simple method for the delivery of volumes of medications, fluids and special fluids like chemotherapy agents via a chamber leading to a catheter and opening into a large vessel in the body. The catheter is inserted into a large vessel that ideally terminates in the superior vena cava/high right atrial junction. The port is surgically implanted subcutaneously in the soft tissue near the clavicle on the patient's right upper chest wall. Medications, fluids, nutritional liquids or chemotherapy agents can then be administered as necessary. The ports are intended for long-term placement.

    The Norfolk Medical SportPort™ family of Vascular Access Devices is a group of ports and catheter sets with infusion and withdrawal capabilities via a port and an access catheter to the large vessels in the vascular system. These ports must be properly attached to a catheter during a procedure.

    When used with a legally marketed power capable needle infusion set, the SportPort™ is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/sec with a 19 or 20 gauge needle and 2ml/sec with a 22 gauge needle set.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the SportPort™ medical device, which is an implanted port and catheter system. The submission focuses on demonstrating substantial equivalence to a predicate device (NorPort CT-PC Port - K111101) rather than presenting a detailed study with acceptance criteria and performance data in the format typically seen for novel AI/diagnostic devices.

    Based on the information provided, here's a breakdown:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission does not specify numerical "acceptance criteria" in the typical sense (e.g., sensitivity, specificity, accuracy targets) because it's for a physical medical device (implantable port) and focuses on mechanical and material equivalence to a predicate. Instead, the "performance" is demonstrated through various mechanical tests to show the device functions as intended and is comparable to the predicate.

    Acceptance Criteria (Implied)Reported Device Performance
    Material EquivalenceMade from the same biocompatible materials as the predicate device (titanium, polysulfone, silicone rubber septum).
    Manufacturing Process EquivalenceUtilizes the same manufacturing process as the NorPort product line.
    Mechanical Integrity (Leaks)Mechanically tested for leaks.
    Mechanical Integrity (Septum Puncturing)Mechanically tested for septum puncturing.
    Mechanical Integrity (Catheter Tensile Strength)Catheter tensile strength testing performed.
    Mechanical Integrity (Catheter to Port Connection)Catheter to port connection test performed.
    Mechanical Integrity (Dynamic Failure)Dynamic failure testing performed.
    Mechanical Integrity (Patency Verification)Patency verification testing performed.
    Mechanical Integrity (Static Burst)Static burst testing performed.
    High-Pressure Injection CapabilityEngineered to withstand high internal pressures in the chamber, catheter securing mechanism, and port outlet for "pressure injectable" clinical applications (power injection of contrast media).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of patient data or clinical studies. The "testing" appears to be entirely bench testing (mechanical testing) of the device prototypes. Therefore, there is no information on:

    • Sample size used for a test set (as it relates to patients/data).
    • Data provenance (country of origin, retrospective/prospective), as no patient data was used.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    Not applicable. The ground truth for mechanical performance is established through engineering standards and testing protocols, not by expert medical review of a test set.

    4. Adjudication Method for the Test Set

    Not applicable, as no human adjudication for a test set (e.g., imaging review) is mentioned.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    No. This type of study is specifically relevant for diagnostic devices where human readers interpret data (e.g., images), often with and without AI assistance, to assess the AI's impact on human performance. The SportPort™ is a physical medical device, not a diagnostic or AI-powered system, so an MRMC study is not relevant or mentioned.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable. The SportPort™ is a physical medical device, not an algorithm or AI system.

    7. The Type of Ground Truth Used

    The ground truth for the device's performance is established by engineering specifications, material science, and mechanical testing standards. The success of the tests (e.g., no leaks, ability to withstand burst pressure, appropriate tensile strength) serves as the ground truth for validating the device's physical and functional integrity.

    8. The Sample Size for the Training Set

    Not applicable. There is no "training set" as this is not an AI or machine learning device. The development process for the SportPort™ relies on engineering design, material selection, and iterative mechanical testing rather than data-driven training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K111101

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-08-25

    (127 days)

    Product Code
    Regulation Number
    880.5965
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NorPort CT PC Family of Implanted Vascular Access Devices is indicated for use when the patient requires repeated access to the vascular system for injections, infusion drugs, administration of blood or blood products, and/or withdrawal of blood as part of the therapy regimen.

    The NorPort CT PC Port is indicated for power injection of contrast media. For rne ron ection of contrast media, the maximum recommended infusion rate is 5 power injection of contained on coring power injectable needle and 2 ml/s with a 22-gauge non-coring power injectable needle.

    The Lucent Non-Coring needle Infusion Set is intended for the administration of fluids and drugs as well as blood sampling through implanted vascular ports.

    When used with the NorPort CT·PC Family of Implanted Vascular Access Devices, the Lucent Non-Coring Needle Infusion Set is also indicated for power injection of contrast media into the central venous system.

    Device Description

    The Norfolk Medical NorPort CT-PC Port is composed of a port reservoir with an attachable catheter system. The Port reservoir consists of a titanium or polysulfone chamber with a silicone septum and an outlet pin (510 (k)#102385). The silicone septum is designed to allow multiple needle puncturing while maintaining the leak-tight integrity.

    The NorPort CT·PC comes in standard design to enable regular use as an infusion/withdrawal port and has high-pressure injection capability to aid in the delivery of large volumes of specialty fluids such as contrast media. The NorPort CT·PC is triangular shaped with a recessed septum ring. The body port has a "CT" symbol machine etched on the bottom end to enable easy recognition of the port with most imaging systems. The port has 3 holes for suture fixation to the tissue. A catheter comes with the port and is inserted in the vascular system. The catheter is usually inserted into a venous vessel and fed down into the superior vena cava. The placement of the catheter is checked by radiographic technique. The catheter is radiopaque to enable visualization for placement. The kit provided to aid in insertion of the catheter and placement of the port may include introducer items like needles, sheathes, vein picks, guidewires, straighteners, and dilators. Likewise the, "cut-down", kit for port placement may contain tunnelers, infusion sets with 90 degree Huber point needles, and syringes along with the port and catheter.

    The Lucent Non-Coring Needle Infusion Set is similar to the following predicate devices: Smiths Medical Gripper Plus Power P.A.C Needle (510 (k)# 070116), the Angiodynamics Lifeguard CT Safety Infusion Set (510 (k)#072375), the Bard the rater of the Safety Infusion Set (510 (k)# 082306), and the MEDCOMP CT Injectable Safety Huber Needle (510 (k)# 080544). The Lucent Non-coring and the predicated devices are supplied sterile and non-pyrogenic and are intended for the administration into or withdrawal of fluids from the Norfolk Medical line of implanted ports. When a "Power Injectable" NorPort CT-PC Port is used, the Lucent Non-coring and the predicate devices are indicated for power injection of contrast media.

    AI/ML Overview

    The Norfolk NorPort CT-PC Port and the Lucent Non-coring Needle Infusion Set underwent non-clinical testing to demonstrate substantial equivalence to predicate devices.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test CategoryAcceptance CriteriaReported Device Performance
    Catheter to Port ConnectionFunctionality and sealing integrity of the connection between the catheter and the port."demonstrated that the functionality and performance characteristics of the NorPort CT-PC implanted vascular port and the Lucent Non-coring needle infusion set are comparable to the currently marketed vascular ports and power capable needles." Specific details on acceptance criteria (e.g., pressure limits, leakage rates) are not provided in the document, but the overall conclusion is that performance is comparable.
    Septum Puncture/Port LeakAbility of the silicone septum to withstand multiple punctures without leaking."The silicone septum is designed to allow multiple needle puncturing while maintaining the leak-tight integrity." "demonstrated that the functionality and performance characteristics of the NorPort CT-PC implanted vascular port and the Lucent Non-coring needle infusion set are comparable to the currently marketed vascular ports and power capable needles." Specific details on acceptance criteria (e.g., number of punctures, maximum leakage) are not explicitly stated, but the device is reported to maintain leak-tight integrity and be comparable to predicates.
    Patency VerificationConfirmation of open flow through the device after implantation and use."demonstrated that the functionality and performance characteristics of the NorPort CT-PC implanted vascular port and the Lucent Non-coring needle infusion set are comparable to the currently marketed vascular ports and power capable needles." Specific details on how patency was verified or its acceptance criteria are not provided, but the overall conclusion is that performance is comparable.
    High Pressure Static TestingAbility of the port to withstand high-pressure injections, particularly for contrast media power injection. Maximum recommended infusion rates: 5 ml/s (19 and 20 gauge needle), 2 ml/s (22 gauge needle).The device "has high-pressure injection capability to aid in the delivery of large volumes of specialty fluids such as contrast media." "When used with the NorPort CT·PC Family of Implanted Vascular Access Devices, the Lucent Non-Coring Needle Infusion Set is also indicated for power injection of contrast media into the central venous system. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s with a 19 and 20 gauge non-coring power injectable needle and 2 ml/s with a 22-gauge non-coring power injectable needle." This indicates the device met the stated high-pressure criteria.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample size used for the test set. It mentions "verification testing was performed according to protocols" and "positive performance examination results available as objective evidence in each category."

    The data provenance is not specified. It is a non-clinical study, so it likely involves laboratory testing rather than data from human subjects or a particular country. It is retrospective in the sense that the testing has been completed.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    Not applicable. This was a non-clinical performance study involving laboratory testing, not human-read interpretations or clinical expert evaluations to establish ground truth.

    4. Adjudication Method for the Test Set:

    Not applicable. As a non-clinical performance study, there was no expert adjudication involved. Performance was measured against pre-defined engineering and functional criteria.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:

    Not applicable. This is a submission for a medical device (implanted port and needle infusion set), not an AI-powered diagnostic or assistive tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:

    Not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.):

    The "ground truth" for this non-clinical study was based on established engineering standards and design specifications, and the performance characteristics of legally marketed predicate devices. The listed standards include:

    • FDA Guidance on 510(k) Submissions for Implanted Infusion Ports, dated October 1990
    • BS EN ISO 10555-1:2009 (Sterile, single-use intravascular catheters Part I: General requirements)
    • ISO 11135-1:2007 (Sterilization of health care products -Ethylene oxide-)
    • ISO 10993-7:2008 (Biological Evaluation of Medical Devices- Part 7: Ethylene Oxide Sterilization Residuals)
    • ISO 14971:2007 (Medical Devices- Application of Risk Management to Medical Devices)

    8. The Sample Size for the Training Set:

    Not applicable. This is a physical medical device, not a machine learning algorithm that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K102385

    Validate with FDA (Live)

    Date Cleared
    2011-01-14

    (144 days)

    Product Code
    Regulation Number
    880.5965
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NorPort Family of Implanted Vascular Access Devices is indicated for use when the patient requires repeated access to the vascular system for injections, infusion drugs, administration of blood or blood products, and/or withdrawal of blood as part of the therapy regimen.

    Device Description

    Our device is composed of a port reservoir with an attachable catheter system. The Port reservoir consists of a titanium (510 (k)# 871192) or polysulfone (510 (k)# 840788) chamber with a silicone septum and an outlet section. The silicone septum is designed to allow multiple needle puncturing while maintaining the leak-tight integrity. The port comes in two different sizes to accommodate different body frames. The catheter that is attached to the port is long enough to be inserted in the superior vena cava/right atrium junction to allow for fluid infusion into the heart and large blood vessels. The NorPort will also be available as part of an implantable kit. The kit contains all the tools needed to aid in catheter placement, insertion and port implantation.

    AI/ML Overview

    The Norfolk Medical NorPort is an implanted drug delivery device (port) consisting of a reservoir with an attachable catheter system. The port is intended for repeated access to the vascular system for injections, infusion drugs, administration of blood or blood products, and/or withdrawal of blood.

    Here's an analysis of the acceptance criteria and supporting study information:

    1. Acceptance Criteria and Reported Device Performance

    The document states that "The NorPort family of products meets all established acceptance criteria for performance testing and design verification testing" (Page 7). While the specific quantitative acceptance criteria are not explicitly listed in a table, the document indicates that the device's performance was compared to established standards and predicate devices.

    Table of Acceptance Criteria and Reported Device Performance (Inferred from Text):

    Acceptance Criterion (Inferred)Reported Device PerformanceStudy Type
    Catheter to Port ConnectionComparable to currently marketed vascular portsNon-clinical testing
    Septum Puncture/Port Leak IntegrityComparable to currently marketed vascular portsNon-clinical testing
    Patency VerificationComparable to currently marketed vascular portsNon-clinical testing
    Static Burst StrengthComparable to currently marketed vascular portsNon-clinical testing
    Tensile StrengthComparable to currently marketed vascular portsNon-clinical testing
    Biocompatibility (Implied by standard use for implantable devices)Not explicitly detailed, but implied as part of "design verification" and comparison to predicate devices, which are already approved for implantation. This would typically involve material safety standards.Not explicitly detailed, but usually evaluated through material testing and predicate device history.
    Sterilization EfficacyMeets ISO 11135-1:2007 requirements (Ethylene oxide sterilization)Sterilization Validation
    Leak-tight integrity after multiple punctures (for silicone septum)Designed to allow multiple needle puncturing while maintaining leak-tight integrity.Non-clinical testing
    Same Indications for Use"Yes, the Norfolk Medical NorPort and catheter have the same indications for use statement to legally marketed predicate devices."Comparison to predicate devices
    No alteration of therapeutic/diagnostic effect"No, the differences do not alter the intended use of the port."Comparison to predicate devices
    No new significant characteristics affecting safety and effectiveness"No, there are no new significant characteristics that would affect safety and effectiveness in the new device."Comparison to predicate devices
    Substantial EquivalenceDemonstrated by performance data gathered in design verification testing.Performance Data Comparison to Predicate Devices

    The overarching acceptance criterion is Substantial Equivalence to legally marketed predicate devices, which the study concluded was met.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: The document does not specify the exact number of devices or components used for each non-clinical test (catheter to port connection, septum puncture/port leak, patency verification, static burst, tensile strength). The phrasing "data collected from the non-clinical tests" implies a test set was used, but the quantity is not provided.
    • Data Provenance: The tests are described as "non-clinical tests" and "design verification testing," suggesting these were conducted in a laboratory setting, likely by the manufacturer, rather than on human or animal subjects. No information is provided about the country of origin or whether it was retrospective or prospective, but as it's non-clinical, these terms are less applicable.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • Number of Experts: Not applicable. For non-clinical performance and design verification tests of a medical device like an implantable port, the "ground truth" is typically established by engineering specifications, recognized industry standards (e.g., ISO, FDA guidance), and the performance characteristics of predicate devices. There wouldn't be human experts establishing "ground truth" in the same way as, for example, reading medical images.
    • Qualifications of Experts: Not applicable for establishing ground truth in this context. The testing would be conducted by qualified engineers and technicians.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies, particularly for diagnostic devices where multiple human readers interpret results and discrepancies are resolved by a senior expert. For non-clinical performance testing, the results are objectively measured against defined criteria, standards, or predicate device performance.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study Done: No. An MRMC study is designed to assess the performance of a diagnostic device or AI algorithm when used by multiple human readers across multiple cases, often comparing performance with and without AI assistance. This device is an implanted drug delivery port, not a diagnostic tool or an AI-assisted interpretation system. Therefore, such a study would not be relevant or possible for this type of device.
    • Effect Size of Human Readers Improvement with AI: Not applicable, as no MRMC study was conducted.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Standalone Study Done: No. This term also typically applies to AI or diagnostic algorithms. For a physical medical device like an implantable port, performance is inherently "standalone" in mechanical and material terms, but it functions as part of a human-in-the-loop system (physician implants, nurse accesses, patient lives with it). The non-clinical tests evaluated the device's inherent physical performance characteristics.

    7. Type of Ground Truth Used

    The ground truth used for evaluating the NorPort's performance was based on:

    • Recognized Industry Standards: FDA Guidance on 510(k) Submissions for Implanted Infusion Ports (October 1990), BS EN ISO 10555-1:2009 (Sterile, single-use intravascular catheters Part 1: General requirements), and ISO 11135-1:2007 (Sterilization of health care products - Ethylene oxide).
    • Predicate Device Performance: The NorPort's functionality and performance characteristics were compared to "currently marketed vascular ports" and explicitly to specific predicate devices: Norfolk Medical Vascular Access Port (K830000), PHS C-Port (K091099), Angiodynamics Vortex VX Port Implantable Infusion Port (K081472), and Bard PowerPort Implanted Titanium Port Chronoflex with 8 Fr. Catheter (K060812).

    8. Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable. This device is a physical medical implant, not an AI algorithm, so there is no "training set" in the context of machine learning. The design and manufacturing process would involve iterative development, prototyping, and testing, but not a formal "training set" for an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    • Ground Truth for Training Set: Not applicable, as there is no training set for this type of device. The design specifications and performance objectives are established through engineering design processes, regulatory requirements, and historical data from similar devices.