Search Results
Found 9 results
510(k) Data Aggregation
(138 days)
The FIREFLY® Pedicle Screw Navigation Guide can be used with any 510(k) cleared, legally marketed, pedicle screw spinal system (for its cleared indications for use) and its respective compatible components for non-cervical open, posterior spinal fixation procedures (T1-S2/S2AI and ilium) intended for fusion, with the additional conditions listed below:
• Pedicle screw's shank is straight along its longitudinal axis (i.e. not curved)
• Pedicle screw's major and minor thread diameters are centered about the longitudinal axis
• Pedicle screw's longitudinal axis matches the direction of insertion
• Pedicle screw is intended to be inserted into a pilot hole
• Pedicle screw's diameter is larger than the pilot hole created with FIREFLY®
• Patient's pedicle must be dimensionally adequate to accommodate a pedicle screw, as determined on preoperative scan
• Compatible pedicle screw spinal system instruments may be used with the FIREFLY® Pedicle Screw Navigation Guide
- Pedicle sounding probes (a.k.a. feeler/ball-tip probes) may be used to confirm pedicle integrity
- Only OEM pedicle screw spinal system taps specified in the Approved Patient-Specific Surgical Plan may be guided to tap pilot holes
- All other pedicle screw spinal system components and accessories (including non-guided taps) are to be used, after removal of the FIREFLY® Pedicle Screw Navigation Guide, as directed by the pedicle screw spinal system's instructions for use
This device is intended for single use only.
The FIREFLY® Pedicle Screw Navigation Guide is intended to assist in the accurate placement of pedicle screws. It consists of single-use components designed for treatment of a specific patient as well as reusable non-patient-specific components.
The FIREFLY® Pedicle Screw Navigation Guide uses Patient-Specific Pedicle Screw Guides that fit on the patient's anatomy to guide surgical instruments in line with trajectories chosen presurgically, by the surgeon, based on the patient's CT imaging or MRI-Derived Synthetic CT data. Patient-Specific Bone Models may also be provided.
N/A
Ask a specific question about this device
(152 days)
Acorn 3D Software is a modular image processing software intended for use as an interface for visualization of medical images, segmentation, treatment planning, and production of an output file.
The Acorn 3D Segmentation module is intended for use as a software interface and image segmentation system for the transfer of CT or CTA medical images to an output file. Acorn 3D Segmentation is also intended for measuring and treatment planning. The Acorn 3D Segmentation output can also be used for the fabrication of physical replicas of the output file using additive manufacturing methods, Acorn 3DP Models. The physical replica can be used for diagnostic purposes in the field of musculoskeletal and craniomaxillofacial applications.
The Acorn 3D Alignment and Measurement module contains registration capabilities and measurement functionality based on anatomical reference geometry. It is intended to allow the user to align anatomical structures between datasets, perform spinopelvic measurements on 3D models of anatomy, and plan surgical procedures in pediatric and adult patients.
Acorn 3D Software and 3DP Models should be used in conjunction with expert clinical judgment.
Acorn 3D Software is an image processing software that allows the user to import, visualize and segment medical images, check and correct the segmentations, and create digital 3D models. The models can be used in Acorn 3D Software for measuring, treatment planning and producing an output file to be used for additive manufacturing (3D printing). Acorn 3D Software is structured as a modular package.
This includes the following functionality:
- Importing medical images in DICOM format
- Viewing images and DICOM data
- Selecting a region of interest using generic segmentation tools
- Segmenting specific anatomy using dedicated semi-automatic tools or fully automatic algorithms
- Verifying and editing a region of interest
- Calculating a digital 3D model and editing the model
- Measuring on images and 3D models
- Exporting 3D models to third-party packages
- Image registration
The Acorn 3D Segmentation module contains both machine learning based auto segmentation as well as semi-automatic and manual segmentation tools. The auto-segmentation tool is only intended to be used for thoracic and lumbar regions of the spine (T1-T12 and L1-L5) and the pelvis (sacrum). Semi-automatic and manual segmentation tools are intended to be used for all musculoskeletal anatomy.
Acorn 3DP Model is an additively manufactured physical replica of the virtual 3D model generated in Acorn 3D Segmentation. The output file from Acorn 3D Segmentation is used to additively manufacture the Acorn 3DP Model.
The Acorn 3D Alignment and Measurement module contains registration capabilities and spinopelvic measurement functionality. It is intended to align spinopelvic anatomical structures between datasets. The module allows the user to perform spinopelvic measurements on 3D models of anatomy, and to plan surgical procedures.
Here's a breakdown of the acceptance criteria and study details for the Acorn 3D Software, based on the provided FDA 510(k) clearance letter:
Acceptance Criteria and Reported Device Performance
| Acceptance Criteria Category | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Alignment Accuracy (Target Registration Error - TRE) for Slot Beam X-Ray Technique | Not explicitly stated (implied to be equivalent to predicate device) | Median TRE: 1.70 mm3rd Quartile TRE: 2.74 mm |
| Alignment Accuracy (Target Registration Error - TRE) for Cone Beam X-Ray Technique | Not explicitly stated (implied to be equivalent to predicate device) | Median TRE: 4.33 mm3rd Quartile TRE: 5.58 mm |
| Machine Learning Model Performance (DICE Similarity Coefficient - DSC) for Vertebral Model (T1-T12, L1-L5) | > 0.88375 | In-House (Mighty Oak Medical): 0.9331VERSE '20: 0.94451 |
| Machine Learning Model Performance (DICE Similarity Coefficient - DSC) for Sacral Model | > 0.96045 | In-House (Mighty Oak Medical): 0.96630 |
| General Performance | Device performance and substantial equivalence to the predicate device. | All performance testing conducted demonstrated device performance and substantial equivalence to the predicate device. |
Study Details
1. Sample Sizes and Data Provenance
Test Set:
- Vertebral Model (T1-T12, L1-L5):
- In-House (Mighty Oak Medical): 35 cases (450 segments - 139 lumbar / 311 thoracic).
- VERSE '20: 36 cases (401 segments - 144 lumbar / 257 thoracic).
- Total Test Set for Vertebral Model: 71 cases.
- Sacral Model:
- In-House (Mighty Oak Medical): 40 cases.
- Total Test Set for Sacral Model: 40 cases.
Data Provenance (Test Set):
- In-House (Mighty Oak Medical): Retrospective (implied, as it's an internal database). Country of origin not explicitly stated.
- VERSE '20: Publicly available dataset. Inclusion criteria: Patients > 30 years of age, no history of bone metastases. Specific country of origin not explicitly stated, but typically these are multi-institutional datasets from various countries. Retrospective.
- CTPelvic1K: Publicly available dataset used only for training and tuning the Sacral model. Country of origin not explicitly stated.
2. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Mighty Oak Medical (In-House): "Highly-trained experts" and "operators who have undergone extensive classroom training and tests for efficacy, with an established track record of creating accurate segmentations." Specific number not mentioned. Qualifications are described as having an "established track record of creating accurate segmentations in support of the FIREFLY Pedicle Screw Navigation Guide."
- VERSE '20: "Widely accepted as ground truth, with three levels of annotation and expert review (by medical and/or graduate students, radiology fellows, and finally a senior consultant)." Specific number not mentioned. The qualifications range from medical/graduate students to radiology fellows and senior consultants.
3. Adjudication Method for the Test Set
- Mighty Oak Medical (In-House): "Verified by feedback loops of expert review and manual editing and correction."
- VERSE '20: "Three levels of annotation and expert review (by medical and/or graduate students, radiology fellows, and finally a senior consultant)." This implies a multi-level consensus approach, but a specific method like "2+1" or "3+1" is not explicitly stated.
4. MRMC Comparative Effectiveness Study
No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was mentioned where human readers' improvement with AI vs. without AI assistance was evaluated. The performance data focuses on the standalone algorithm's accuracy, and the device is intended to be used "in conjunction with expert clinical judgment."
5. Standalone Performance Study
Yes, standalone (algorithm only) performance was done. The "Machine-Learning Model Performance Testing" section directly compares the model's output (segmentations) with ground truth segmentations using the DICE Similarity Coefficient (DSC). The "Alignment Accuracy" evaluation also assesses the algorithm's performance directly against ground truth.
6. Type of Ground Truth Used
- Expert Consensus: For both In-House (Mighty Oak Medical) and VERSE '20 datasets, the ground truth was established through expert review and refinement of segmentations.
- Pathology/Outcomes Data: Not explicitly mentioned as being used for ground truth for the segmentation or alignment accuracy.
7. Sample Size for the Training Set
- Vertebral Model (T1-T12, L1-L5): 147 cases (from Mighty Oak Medical In-House dataset).
- Sacral Model: 204 cases (104 from Mighty Oak Medical In-House dataset, 100 from CTPelvic1K dataset).
8. How the Ground Truth for the Training Set Was Established
- Mighty Oak Medical (In-House): Similar to the test set, segmentations are "produced on clinical cases by operators who have undergone extensive classroom training and tests for efficacy, with an established track record of creating accurate segmentations." These were "verified by feedback loops of expert review and manual editing and correction."
- CTPelvic1K: "Widely accepted as ground truth, with three levels of annotation and expert review (by medical and/or graduate students, radiology fellows, and finally a senior consultant), and being used in support of public machine-learning segmentation research."
Ask a specific question about this device
(168 days)
Acorn 3D Software is a modular image processing software intended for use as an interface for visualization of medical images, segmentation, treatment planning, and production of an output file.
The Acorn 3D Segmentation module is intended for use as a software interface and image segmentation system for the transfer of CT or CTA medical images to an output file. Acorn 3D Segmentation is also intended for measuring and treatment planning. The Acorn 3D Segmentation output can also be used for the fabrication of physical replicas of the output file using additive manufacturing methods, Acorn 3DP Models. The physical replica can be used for diagnostic purposes in the field of musculoskeletal and craniomaxillofacial applications.
The Acorn 3D Trajectory Automation module may be used to plan pedicle screw placement in the thoracic and lumbar regions of the spine in pediatric and adult patients.
Acorn 3D Software and 3DP Models should be used in conjunction with expert clinical judgment.
Acorn 3D Software is an image processing software that allows the user to import, visualize and segment medical images, check and correct the segmentations, and create digital 3D models. The models can be used in Acorn 3D Software for measuring, treatment planning and producing an output file to be used for additive manufacturing (3D printing). Acorn 3D Software is structured as a modular package.
This includes the following functionality:
- Importing medical images in DICOM format
- Viewing images and DICOM data
- Selecting a region of interest using generic segmentation tools
- Segmenting specific anatomy using dedicated semi-automatic tools or fully automatic algorithms
- Verifying and editing a region of interest
- Calculating a digital 3D model and editing the model
- Measuring on images and 3D models
- Exporting 3D models to third-party packages
- Planning pedicle screw placement
The Acorn 3D Segmentation module contains both machine learning based auto segmentation as well as semi-automatic and manual segmentation tools. The auto-segmentation tool is only intended to be used for thoracic and lumbar regions of the spine (T1-T12 and L1-L5) and the pelvis (sacrum). Semi-automatic and manual segmentation tools are intended to be used for all musculoskeletal anatomy.
| Automatic | Semi-Automatic | Manual | |
|---|---|---|---|
| Definition | Algorithmic with little or no direct human control | A combination of algorithmic and direct human control | Directly controlled by a human |
| Tool Type | Machine Learning algorithm used to automatically segment individual vertebrae and the pelvis | Algorithmic based tools that do not incorporate machine learning. | Manual tools requiring user input. |
| Anatomical Location(s) | Spinal anatomy:• Thoracic (T1-T12)• Lumbar (L1-L5)• Sacrum | Musculoskeletal & craniomaxillofacial bone:• Short• Long• Flat• Sesamoid• Irregular | Musculoskeletal & craniomaxillofacial bone:• Short• Long• Flat• Sesamoid• Irregular |
Acorn 3DP Model is an additively manufactured physical replica of the virtual 3D model generated in Acorn 3D Segmentation. The output file from Acorn 3D Segmentation is used to additively manufacture the Acorn 3DP Model.
The Acorn 3D Trajectory Automation module contains dedicated fully automatic algorithms for planning pedicle screw trajectories. The algorithms are only intended to be used for the thoracic and lumbar regions of the spine (T1-T12 and L1-L5). The output file from Acorn 3D Trajectory Automation contains information relevant to pedicle screw placement surgery, including entry points, end points, and screw sizes of planned screws.
The provided FDA 510(k) clearance letter describes the Acorn 3D Software, specifically focusing on the new Acorn 3D Trajectory Automation module. However, the document is quite sparse on detailed descriptions of the acceptance criteria and the specifics of the study proving the device meets these criteria. The information below is extracted from the provided text, and where details are missing, it's explicitly stated.
1. Table of Acceptance Criteria and Reported Device Performance
The document mentions "deviations were within the acceptance criteria" without specifying the numerical acceptance criteria themselves. It also doesn't provide specific numerical results of the device's performance, only a qualitative statement of accuracy.
| Acceptance Criteria (Not explicitly stated in numerical terms) | Reported Device Performance |
|---|---|
| Accuracy of pedicle screw geometry: Deviations within acceptance criteria | Deviations were within the acceptance criteria. |
| Accuracy of pedicle screw trajectories: Deviations within acceptance criteria | Deviations were within the acceptance criteria. |
| Substantial equivalence to predicate device for planning pedicle screws and trajectories | Performance testing demonstrated substantial equivalence to the predicate device. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: Not specified. The document only mentions "clinical data" was used.
- Data Provenance: Not specified. It's unclear if the clinical data was retrospective or prospective, or the country of origin.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
- The document does not specify the number of experts used or their qualifications for establishing ground truth for the test set.
4. Adjudication Method for the Test Set
- The document does not specify any adjudication method (e.g., 2+1, 3+1, none) used for the test set.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- The document does not mention that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done. Therefore, no effect size of human readers improving with AI vs. without AI assistance is provided.
6. Standalone Performance Study
- Yes, a standalone performance study was done for the Acorn 3D Trajectory Automation module. The document states:
- "The accuracy of pedicle screw geometry as well as pedicle screw trajectories created in the subject device, Acorn 3D Trajectory Automation, was assessed via bench testing."
- This implies the algorithm's performance was evaluated independently without human intervention during the trajectory planning process itself, as it's a "fully automatic algorithm."
7. Type of Ground Truth Used
- The ground truth for the "accuracy of pedicle screw geometry" and "pedicle screw trajectories" was established by comparing the device's output to "clinical data." While the nature of this "clinical data" is not explicitly defined (e.g., expert consensus on images, surgical outcomes, or pathology reports), it serves as the reference standard.
8. Sample Size for the Training Set
- The document does not specify the sample size used for the training set of the machine learning algorithms. It mentions "Using a collection of images and masks as a training dataset for machine-learning segmentation algorithm" but no numbers.
9. How the Ground Truth for the Training Set Was Established
- The document states that the machine learning segmentation algorithm uses "a collection of images and masks as a training dataset." It doesn't explicitly describe how these "masks" (which represent the ground truth segmentations) were created. It can be inferred that these masks would have been generated by human experts, likely through manual or semi-automatic segmentation, but this is not confirmed in the text.
Ask a specific question about this device
(206 days)
Acorn Segmentation is intended for use as a software interface and image segmentation system for the transfer of CT or CTA medical images to an output file. Acorn Segmentation is also intended for measuring and treatment planning. The Acorn Segmentation output can also be used for the fabrication of the output file using additive manufacturing methods, Acorn 3DP Models. The physical replica can be used for diagnostic purposes in the field of musculoskeletal and craniomaxillofacial applications.
Acorn Segmentation and 3DP Models should be used in conjunction with expert clinical judgment.
Acorn Segmentation is an image processing software that allows the user to import, visualize and segment medical images, check and correct the segmentations, and create digital 3D models. The models can be used in Acorn Segmentation for measuring, treatment planning and producing an output file to be used for additive manufacturing (3D printing). Acorn Segmentation is structured as a modular package. This includes the following functionality:
- Importing medical images in DICOM format
- Viewing images and DICOM data
- Selecting a region of interest using generic segmentation tools
- Segmenting specific anatomy using dedicated semi-automatic tools or automatic algorithms
- Verifying and editing a region of interest
- Calculating a digital 3D model and editing the model
- Measuring on images and 3D models
- Exporting 3D models to third-party packages
Acorn Segmentation contains both machine learning based auto-segmentation as well as semi-automatic and manual segmentation tools. The auto-segmentation tool is only intended to be used for thoracic and lumbar regions of the spine (T1-T12 and L1-L5). Semi-automatic and manual segmentation tools are intended to be used for all musculoskeletal and craniomaxillofacial anatomy.
Acorn 3DP Model is an additively manufactured physical replica of the digital 3D model generated in Acorn Segmentation. The output file from Acorn Segmentation is used to additively manufacture the Acorn 3DP Model.
The provided text describes the acceptance criteria and the study that proves the device meets those criteria for the Acorn 3D Software (AC-SEG-4009) and Acorn 3DP Model (AC-101-XX).
Here's the breakdown:
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria (Metric) | Target Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Geometric accuracy of digital models (Dice-Sorensen Coefficient) - Automatic Segmentation (Thoracic and Lumbar spine) | Average Dice-Sorensen Coefficient ≥ 0.93 | Average Dice-Sorensen Coefficient > 0.93 |
| Geometric accuracy of digital models (Dice-Sorensen Coefficient) - Semi-automatic and Manual Segmentation (Musculoskeletal & Craniomaxillofacial bone) | Average Dice-Sorensen Coefficient ≥ 0.93 | Average Dice-Sorensen Coefficient > 0.93 |
| Geometric accuracy of physical replicas (Mean Deviation) | Mean Deviation < 1mm | Mean Deviation < 1mm |
2. Sample size(s) used for the test set and the data provenance
The document does not specify the exact sample size for the test set nor the data provenance (e.g., country of origin, retrospective or prospective). It only states that software verification and validation included bench testing for geometric accuracy.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document does not explicitly state the number of experts or their qualifications used to establish the ground truth for the test set.
4. Adjudication method for the test set
The document does not mention any specific adjudication method (e.g., 2+1, 3+1) for the test set.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
A Multi-Reader, Multi-Case (MRMC) comparative effectiveness study was not explicitly mentioned or described. The study primarily focused on the algorithmic performance against predicate devices and physical models, rather than human reader improvement with AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done
Yes, a standalone (algorithm only) performance study was conducted. The "Accuracy of automatic segmentation" was evaluated, which by definition means the algorithm's performance without direct human control. The bench testing for geometric accuracy directly assesses the algorithm's output against a reference.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth for the geometric accuracy of digital models was established by comparing the Acorn Segmentation output against "predicate device segmentations." This implies that the segmentations generated by previously cleared and accepted devices served as the reference standard. For physical replicas, the ground truth was the "digital models" themselves.
8. The sample size for the training set
The document states that the "Acorn Segmentation contains both machine learning based auto-segmentation... Using a collection of images and masks as a training dataset for machine-learning segmentation algorithm". However, the specific sample size for the training set is not provided in this document.
9. How the ground truth for the training set was established
The document states that the ground truth for the training set was established by "using a collection of images and masks as a training dataset for machine-learning segmentation algorithm." This implies that pre-existing segmented images (masks) were used as the ground truth for training the machine learning model. The method by which these initial "masks" were generated (e.g., manual segmentation by experts, semi-automatic segmentation, etc.) is not detailed in this document.
Ask a specific question about this device
(281 days)
The FIREFLY® Cervical Navigation Guide system is intended to provide mechanical guidance for the preparation and drilling of pilot holes for the placement of posterior cervical spine (C1-C7) and the upper thoracic spine (T1-T3). Pilot holes are created using the FIREFLY® Cervical Navigation Guide in the pedicles (C2-T3), Pars (C2), or lateral masses (C1) during open, posterior spinal fixation procedures, on skeletally mature patients, that are intended for fusion. The patient's pedicles, pars, or lateral masses must be dimensionally adequate to safely accommodate a posterior cervical screw, as determined on a preoperative CT/CTA scan.
The FIREFLY® Cervical Navigation Guide system is compatible with FDA cleared, legally marketed, posterior cervical screw systems (and their respective components) that are specified in the precautions. Pedicle sounding probes (a.k.a. feeler/ball-tip probes) may be used to confirm each pedicle's integrity. Only qualified compatible OEM posterior cervical screw system taps may be used to visually guide the tapping of pilot holes. All other posterior cervical screw system components and accessories (including non-visually guided taps) are to be used, after removal of the FIREFLY @ Cervical Navigation Guide, as directed by the posterior cervical screw system's instructions for use.
The FIREFLY® Cervical Navigation Guide system is only compatible with consoles systems (attachments and burs) listed in the precautions.
This system (guide, bone model, drill bit, and depth stop) are intended for single use only.
The FIREFLY® Cervical Navigation Guide is intended to assist in the accurate placement of posterior cervical screws. It consists of single-use components designed for treatment of a specific patient.
The FIREFLY® Cervical Navigation Guide uses Patient-Specific Cervical Guides that fit on the patient's anatomy to guide surgical instruments in line with trajectories chosen presurgically, by the surgeon, based on the patient's CT/CTA imaging data. Navigation guides are intended to guide instruments to create pilot holes in the pedicles (C2-T3), Pars (C2), or lateral masses (C1) for placing screws following the Approved Patient-Specific Surgical Plan.
Patient-Specific Bone Models are also provided.
The provided text describes a 510(k) premarket notification for the FIREFLY® Cervical Navigation Guide. It outlines the device's indications for use, technological characteristics, and a summary of performance data from cadaveric accuracy testing.
Here's the breakdown of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document states that "Cadaveric accuracy testing of the FIREFLY® Cervical Navigation Guide was performed. The results demonstrated that the acceptance criteria were met and that the FIREFLY® Cervical Navigation Guide's performance is adequate to perform as intended." However, the specific quantitative acceptance criteria and the corresponding reported device performance values (e.g., accuracy in angle or translation) are not explicitly provided in the document.
Therefore, the table cannot be fully constructed.
2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)
- Sample Size: Not explicitly stated in the document. The text only mentions "Cadaveric accuracy testing." It does not specify the number of cadavers or the number of pilot holes tested.
- Data Provenance: The study was "Cadaveric accuracy testing." The country of origin is not specified, and it is a type of prospective study on cadavers.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Number of Experts: Not explicitly stated.
- Qualifications of Experts: Not explicitly stated.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Adjudication Method: Not explicitly stated.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study: No, an MRMC comparative effectiveness study involving human readers and AI assistance was not conducted or described. The device is a "Cervical Navigation Guide" which provides mechanical guidance for instrument placement. It is not an AI-assisted diagnostic or interpretation tool for human readers but rather a surgical guidance tool.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- This question is less directly applicable to a mechanical surgical navigation guide. The "performance" described is the accuracy of the guide itself in directing instruments. Since it's a mechanical guide, its function is inherently "standalone" in guiding the physical instrument, although a human surgeon is always "in the loop" operating the instruments through the guide. The cadaveric testing assesses the mechanical accuracy of the guide as intended.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- For the cadaveric accuracy testing, the ground truth would typically be established by highly precise measurements of the actual pilot hole trajectory and depth in relation to the pre-surgically planned trajectory, often using imaging (e.g., post-procedure CT scans) and/or metrology. The document does not specify the exact method for ground truth establishment.
8. The sample size for the training set
- The document describes a cadaveric performance study, not a machine learning model that relies on a training set. Therefore, this question is not applicable.
9. How the ground truth for the training set was established
- As this is not a machine learning device, this question is not applicable.
Ask a specific question about this device
(82 days)
The FIREFLY® Midline Navigation Guide can be used with any 510(k) clearly marketed, pedicle screw spinal system (for its cleared indications for use) and its respective components for non-cervical open, posterior spinal fixation procedures (T10-S2) intended for fusion, with the additions listed below:
- o Pedicle screw's shank is straight along its longitudinal axis (i.e. not curved)
- . Pedicle screw's major and minor thread diameters are centered about the longitudinal axis
- Pedicle screw's longitudinal axis matches the direction of insertion
- Pedicle screw is intended to be inserted into a pilot hole
- Pedicle screw's diameter is larger than the pilot hole created with FIREFLY®
- Patient's pedicle must be dimensionally adequate to accommodate a pedicle screw, as determined on ● preoperative scan
- . Compatible pedicle screw spinal system instruments may be used with the FIREFLY® Midline Screw Navigation Guide
- Pedicle sounding probes (a.k.a. feeler/ball-tip probes) may be used with the FIREFLY® Midline Navigation O Guide to confirm pedicle integrity
- All other pedicle screw spinal system components and accessories [including non-guided taps] are to be O used, after removal of the FIREFLY® Midline Navigation Guide, as directed by the pedicle screw spinal system's instructions for use
This device is intended for single use only.
The FIREFLY® Midline Navigation Guide is intended to assist in the accurate placement of pedicle screws. It consists of single-use components designed for treatment of a specific patient.
The FIREFLY® Midline Navigation Guide uses Patient-Specific Guides that fit on the patient's anatomy to guide surgical instruments in line with trajectories chosen presurgically, by the surgeon, based on the patient's CT imaging data. Navigation Guides are intended to guide instruments to create pilot holes in the pedicles for placing pedicle screws following the Approved Surgical Plan.
Patient-Specific Bone Models are also provided.
Here's a breakdown of the acceptance criteria and study information for the FIREFLY® Midline Navigation Guide, based on the provided FDA 510(k) summary:
1. Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance | Comments |
|---|---|---|
| Not explicitly stated in terms of specific numerical thresholds for accuracy (e.g., angular deviation, translational error). | "The results demonstrated that the acceptance criteria were met and that the FIREFLY® Midline Navigation Guide's performance is adequate to perform as intended." | The document confirms that acceptance criteria were met, but does not detail the specific numerical values of these criteria or the test results. |
2. Sample Size and Data Provenance for Test Set
- Sample Size Used for Test Set: Not specified. The document only mentions "Cadaveric accuracy testing."
- Data Provenance: Cadaveric. The document does not specify the country of origin or if the cadavers were retrospectively or prospectively selected.
3. Number of Experts and Qualifications for Ground Truth (Test Set)
Not specified. The document does not mention the involvement of experts in establishing ground truth for the cadaveric accuracy testing. It's likely that the ground truth for pedicle screw placement in cadaveric studies is often established by comparing the guided placement to anatomical landmarks or established safe zones, which may involve expert anatomical assessment, but the details are not provided.
4. Adjudication Method (Test Set)
Not specified. No information is provided regarding any adjudication method used for the cadaveric accuracy testing.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No. The document does not mention a multi-reader multi-case (MRMC) comparative effectiveness study, nor does it discuss human reader performance with or without AI assistance. The device is a "Navigation Guide," implying it assists in physical placement rather than image interpretation by human readers.
6. Standalone Performance Study (Algorithm Only)
Yes. The "Cadaveric accuracy testing" appears to be a standalone performance study of the device's ability to guide instruments accurately, without human-in-the-loop performance being assessed in a comparative manner with human readers. The device guides the surgeon to create pilot holes, meaning its performance is inherently linked to its interaction with a surgeon, but the "accuracy testing" usually refers to the precision of the guide itself.
7. Type of Ground Truth Used
The type of ground truth used for the cadaveric accuracy testing is implicitly anatomical/physical measurement. The device guides instrument placement, and "accuracy testing" in this context would likely involve measuring the deviation of the guided trajectories from the presurgically planned trajectories or known correct anatomical paths within the cadaveric pedicles.
8. Sample Size for Training Set
The document does not mention a training set. The FIREFLY® Midline Navigation Guide uses "Patient-Specific Guides" based on "the patient's CT imaging data" and "trajectories chosen presurgically, by the surgeon." This suggests a patient-specific design approach rather than a machine learning model that requires a large training set.
9. How Ground Truth for Training Set Was Established
Not applicable, as no training set for an AI/machine learning model is described. The device's operation is based on pre-surgical planning by a surgeon using patient CT data, not on a trained algorithm.
Ask a specific question about this device
(56 days)
The FIREFLY® Pedicle Screw Navigation Guide can be used with any 510(k) cleared, legally marketed, pedicle screw spinal system (for its cleared indications for use) and its respective compatible components for non-cervical open, posterior spinal fixation procedures (TI-S2/S2Al and ilium) intended for fusion, with the additional conditions listed below:
- . Pedicle screw's shank is straight along its longitudinal axis (i.e. not curved)
- . Pedicle screw's major and minor thread diameters are centered about the longitudinal axis
- Pedicle screw's longitudinal axis matches the direction of insertion ●
- Pedicle screw is intended to be inserted into a pilot hole ●
- Pedicle screw's diameter is larger than the pilot hole created with FIREFLY® ●
- Patient's pedicle must be dimensionally adequate to accommodate a pedicle screw, as determined on ● preoperative scan
- . Compatible pediale screw spinal system instruments may be used with the FIREELY® Pedicle Screw Navigation Guide
- Pedicle sounding probes (a.k.a. feeler/ball-tip probes) may be used to confirm pedicle integrity o
- Only OEM pedicle screw spinal system taps specified in the Approved Patient-Specific Surgical Plan may be O quided to tap pilot holes
- All other pedicle screw spingl system components and accessories (including non-quided faps) are to be O used, after removal of the FIREFLY® Pedicle Screw Navigation Guide, as directed by the pedicle screw spinal system's instructions for use
This device is intended for single use only.
The FIREFLY® Pedicle Screw Navigation Guide is intended to assist in the accurate placement of pedicle screws. It consists of single-use components designed for treatment of a specific patient as well as reusable non-patient-specific components.
The FIREFLY® Pedicle Screw Navigation Guide uses Patient-Specific Pedicle Screw Guides that fit on the patient's anatomy to guide surgical instruments in line with trajectories chosen presurgically, by the surgeon, based on the patient's CT imaging data. Navigation guides are intended to guide instruments to create pilot holes in the pedicles for placing pedicle screws following the Approved Patient-Specific Surgical Plan.
Patient-Specific Bone Models may also be provided.
The provided text describes the 510(k) summary for the FIREFLY® Pedicle Screw Navigation Guide (K181399). Based on the text, here's a breakdown of the requested information:
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| Not explicitly stated in the document. The document mentions "the acceptance criteria were met" but does not detail what those criteria specifically were. | The results of "Additional cadaveric accuracy testing" demonstrated that the device met its unspecified acceptance criteria and its performance is adequate to perform as intended. |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document simply states "Additional cadaveric accuracy testing" was performed. It does not specify the number of cadavers or pedicles used in this test set.
- Data Provenance: The study was "cadaveric accuracy testing," implying it was conducted on human cadavers. The country of origin and whether it was retrospective or prospective are not mentioned.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of those Experts
This information is not provided in the document. The text does not describe how ground truth was established for the cadaveric accuracy testing, nor does it mention the involvement or qualifications of experts for this purpose.
4. Adjudication Method for the Test Set
This information is not provided in the document.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Was an MRMC study done? No, the document does not mention an MRMC comparative effectiveness study involving human readers with and without AI assistance.
- Effect size: Not applicable, as no such study is described.
6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study
The FIREFLY® Pedicle Screw Navigation Guide is a physical device intended to guide surgical instruments. It is not an AI algorithm in the traditional sense that would have "standalone" performance without human-in-the-loop. The "performance data" section refers to cadaveric accuracy testing of the physical guide's ability to direct instrument placement, which inherently involves human surgeons using the device.
7. Type of Ground Truth Used
The document does not explicitly state the type of ground truth used for the cadaveric accuracy testing. For such a device, ground truth would typically be established by precise measurements of actual screw placement in relation to planned trajectories, possibly using highly accurate imaging techniques (e.g., post-operative CT scans with fiducial markers, or direct anatomical measurements). However, the specific method is not detailed.
8. Sample Size for the Training Set
The document states, "The FIREFLY® uses Patient-Specific Pedicle Screw Guides that fit on the patient's anatomy to guide surgical instruments in line with trajectories chosen presurgically, by the surgeon, based on the patient's CT imaging data." This indicates that each guide is patient-specific and generated from individual patient CT data. There isn't a "training set" in the conventional machine learning sense for the device itself, as it's a patient-specific physical guide based on pre-operative imaging rather than a learned algorithm trained on a large dataset.
9. How the Ground Truth for the Training Set Was Established
As explained above, there isn't a traditional "training set" for an AI algorithm here. The "training" for the patient-specific guide comes from the individual patient's CT imaging data and the surgeon's presurgical planning. The "ground truth" for this process would be the surgeon's intended screw trajectory based on their anatomical assessment from the CT.
Ask a specific question about this device
(59 days)
The FIREFLY® Pedicle Screw Navigation Guide can be used with any 510(k) cleared, legally marketed, pedicle screw spinal system (for its cleared indications for use) and its respective compatible components for non-cervical open, posterior spinal fixation procedures (T1-S2/ilium) intended for fusion, with the additional conditions listed below:
- Pedicle screw's shank is straight along its longitudinal axis (i.e. not curved)
- Pedicle screw's major and minor thread diameters are centered about the longitudinal axis
- Pedicle screw's longitudinal axis matches the direction of insertion ●
- Pedicle screw is intended to be inserted into a pilot hole ●
- Pedicle screw's diameter is larger than the pilot hole created with FIREFLY® ●
- Patient's pedicle must be dimensionally adequate to accommodate a pedicle screw, as determined on ● preoperative scan
- Compatible pediale screw spinal system instruments may be used with the FIREELY® Pedicle Screw Navigation Guide .
- Pedicle sounding probes (a.k.a. feeler/ball-tip probes) may be used to confirm pedicle integrity o
- Only OEM pedicle screw spinal system taps specified in the Approved Patient-Specific Surgical Plan may be 0 guided to tap pilot holes
- All other pedicle screw spinal system components and accessories lincluding non-quided faps) are to be O used, after removal of the FIREFLY® Pedicle Screw Navigation Guide, as directed by the pedicle screw spinal system's instructions for use
This device is intended for single use only.
The FIREFLY® Pedicle Screw Navigation Guide is intended to assist in the accurate placement of pedicle screws. It consists of single-use components designed for treatment of a specific patient as well as reusable non-patient-specific components.
The FIREFLY® Pedicle Screw Navigation Guide uses Patient-Specific Pedicle Screw Guides that fit on the patient's anatomy to guide surgical instruments in line with trajectories chosen presurgically, by the surgeon, based on the patient's CT imaging data. Navigation guides are intended to guide instruments to create pilot holes in the pedicles for placing pedicle screws following the Approved Patient-Specific Surgical Plan.
Patient-Specific Bone Models may also be provided.
The purpose of this traditional 510(k) is to expand the FIREFLY® indications for use to allow compatibility with all pedicle screw systems (for their cleared indications for use) meeting certain criteria and use at an additional spinal level. The FIREFLY® design is the same as was originally cleared in K143222.
The provided document, K162419, describes the substantial equivalence determination for the "FIREFLY® Pedicle Screw Navigation Guide". While it mentions that "Additional cadaveric accuracy testing of the FIREFLY® Pedicle Screw Navigation Guide was performed. The results demonstrated that the acceptance criteria were met and that the FIREFLY® Pedicle Screw Navigation Guide's performance is adequate to perform as intended," it does not fully detail the acceptance criteria or the specifics of the study in a manner that allows for a comprehensive answer to all parts of your request.
Here's an attempt to answer based on the available information, noting where details are not provided:
1. Table of Acceptance Criteria and Reported Device Performance
Acceptance Criteria: The document states that acceptance criteria were met but does not explicitly list the specific quantitative acceptance criteria. It only generally refers to "accuracy."
Reported Device Performance: The document states that "Additional cadaveric accuracy testing... demonstrated that the acceptance criteria were met and that the FIREFLY® Pedicle Screw Navigation Guide's performance is adequate to perform as intended." Specific performance metrics (e.g., mean angular deviation, mean translational error) are not provided in this document.
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size for Test Set: The document simply mentions "Additional cadaveric accuracy testing" without specifying the number of cadavers, pedicles, or screws tested.
- Data Provenance: "Cadaveric accuracy testing" implies the data was derived from cadaveric studies. The country of origin is not specified but given the FDA submission, it's likely a study conducted in the USA or adhering to US standards. It's inherently a retrospective analysis of the device's accuracy on cadaveric specimens.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
The document does not provide information on the number of experts used or their qualifications for establishing ground truth in the cadaveric study. For cadaveric studies involving pedicle screws, ground truth often involves post-operative CT scans or direct measurement by skilled researchers/surgeons.
4. Adjudication Method for the Test Set
The document does not specify an adjudication method.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not mentioned or described in the provided document. This type of study is more common for diagnostic imaging AI systems to assess human performance with and without AI assistance. This device is a surgical navigation guide.
6. Standalone (Algorithm Only) Performance Study
The document describes "accuracy testing," which is inherently a standalone performance evaluation of the device as it guides surgical instruments. However, the exact methodology and quantitative results for this standalone performance are not provided beyond the statement that acceptance criteria were met.
7. Type of Ground Truth Used
For "cadaveric accuracy testing," the ground truth typically involves:
- Post-operative imaging: High-resolution CT scans of the cadaveric spine with inserted pedicle screws, where the actual position of the screws relative to the pedicle canal is measured.
- Direct measurement: Physical dissection and measurement, though less common for pedicle screw accuracy compared to imaging.
The document does not explicitly state how the ground truth was established, but the context of pedicle screw accuracy testing points to methods that objectively verify screw placement against anatomical goals.
8. Sample Size for the Training Set
The document describes a "Pedicle Screw Navigation Guide," which is a mechanical guidance device, not an AI/machine learning algorithm that requires a "training set" in the traditional sense. The device's design is based on patient-specific CT imaging for pre-surgical planning, but this is a process of individual case design, not an algorithm being "trained" on a large dataset. Therefore, the concept of a training set sample size is not applicable here.
9. How Ground Truth for the Training Set Was Established
As noted above, the concept of a "training set" and its associated ground truth is not applicable to this type of mechanical surgical navigation guide.
Ask a specific question about this device
(396 days)
The FIREFLY™ Pedicle Screw Navigation Guide is a patient-specific system intended to guide the drilling and tapping of pilot holes for placement of pedicle screws according to surgeon-prescribed pre-operatively planned trajectories during open, posterior, instrumented spinal surgery (T1-S1/Ilium). The FIREFLY™ Pedicle Screw Navigation Guide is intended for use with the pedicle screw spinal systems specified in the instructions for use and in patients consistent with the selected system's cleared indications for use.
Use of the FIREFLY™ Pedicle Screw Navigation Guide involves surgical planning software used pre-operatively to plan the surgical placement of the pilot holes on the basis of patient CT radiological images with identifiable placement of anatomical landmarks. Only compatible OEM taps that are supplied with the pedicle screw spinal systems specified in the instructions for use may be used through the FIREFLY™ Pedicle Screw Navigation Guide to tap pilot holes. All other pedicle screw spinal system components and accessories (including non-guided taps) are to be used after removal of the FIREFLY™ Pedicle Screw Navigation Guide, as directed by the pedicle screw spinal system's instructions for use.
This device is intended for single use only.
The FIREFLY ™ Pedicle Screw Navigation Guide is a patient-specific system intended to assist in the accurate placement of pedicle screws. It consists of single-use components designed for treatment of a specific patient as well as reusable non-patient-specific components.
The FIREFLY ™ Pedicle Screw Navigation Guide uses Patient-Specific Pedicle Screw Guides that fit on the patient's anatomy to guide surgical instruments in line with trajectories chosen presurgically, by the surgeon, based on the patient's CT imaging data. Navigation guides are intended to guide instruments to create pilot holes in the pedicles for placing pedicle screws following the Approved Patient-Specific Surgical Plan.
Patient-Specific Bone Models may also be provided.
The provided text describes the FIREFLY™ Pedicle Screw Navigation Guide and its performance data. However, it does not contain specific acceptance criteria, detailed study results, sample sizes for test/training sets, or information on ground truth establishment, expert consensus, or MRMC studies. The document is a 510(k) summary for FDA clearance, which typically presents a high-level overview rather than detailed scientific study breakdowns.
Based on the available information, here's what can be extracted and what is missing:
1. Table of Acceptance Criteria and Reported Device Performance
| Performance Metric | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Accuracy | Not specified | Met acceptance criteria |
| Sterilization stability | Not specified | Met acceptance criteria |
Missing Information: The specific numerical or qualitative acceptance criteria for accuracy and sterilization stability are not detailed in this document.
2. Sample Size Used for the Test Set and Data Provenance
The document mentions "Cadaveric accuracy and sterilization stability testing."
- Sample Size for Test Set: Not specified.
- Data Provenance: Cadaveric testing. (Country of origin is not specified, and it's a prospective study as it was performed for this specific device).
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
- Number of Experts: Not specified.
- Qualifications of Experts: Not specified.
4. Adjudication Method for the Test Set
- Adjudication Method: Not specified.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done
- MRMC Study: No, an MRMC comparative effectiveness study is not mentioned. The document describes accuracy testing, but not a comparison of human readers with and without AI assistance to measure improvement effect size.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done
The device is a "Pedicle Screw Navigation Guide," which implies it is a physical guide used by a surgeon during surgery. It is not an AI algorithm in the traditional sense that would have "standalone" performance independent of human interaction in the operating room. Therefore, the concept of "standalone algorithm performance" as typically applied to AI image analysis is not relevant here. The accuracy refers to the guide's precision in directing the surgical instruments.
7. The Type of Ground Truth Used
- Ground Truth: For accuracy, the "ground truth" would likely be the actual placement of the pilot holes or screws in the cadavers, as measured post-procedure (e.g., via CT scanning and comparison to pre-operative plans). This is implied by the term "accuracy testing," but the specific method of establishing this ground truth (e.g., using a gold-standard imaging technique and detailed measurements) is not specified.
8. The Sample Size for the Training Set
The device is a physical navigation guide, not a machine learning algorithm that requires a "training set" in the computational sense. The planning software uses patient CT images, but these are for individual patient planning, not for training a generalized model. Therefore, this question is not applicable.
9. How the Ground Truth for the Training Set Was Established
Not applicable, as there is no "training set" in the context of an AI algorithm.
Summary of Study that Proves Device Meets Acceptance Criteria:
The document states: "Cadaveric accuracy and sterilization stability testing of the FIREFLY Pedicle Screw Navigation Guide was performed. The results demonstrated that the acceptance criteria were met and that the FIREFLY Pedicle Screw Navigation Guide performance is adequate to perform as intended."
This indicates that an experimental study was conducted using cadaveric specimens to evaluate the precision of the guide in directing surgical instruments and to confirm its sterility after processing. However, the details of these studies, including specific methodologies, data collected, statistical analyses, and the actual numerical results against defined acceptance criteria, are not provided in this 510(k) summary.
Ask a specific question about this device