510(k) Data Aggregation

    K Number
    K170050

    Validate with FDA (Live)

    Device Name
    Z-CAD smile
    Manufacturer
    Date Cleared
    2018-08-23

    (595 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    METOXIT Z-CAD® smile blanks are indicated for the preparation of full ceramic crowns, inlays, veneers and 3 unit bridges in the anterior area. All white Z-CAD® smile blanks are indicated to be coloured with Z-CAD® Liquid smile-TC.

    The Z-CAD® Liquid smile-TC will be used to dye white pre-sintered zirconia restorations milled from Z-CAD® smile blanks in the dental field.

    Device Description

    Z-CAD® smile ceramic blanks are indicated for the production of dental-prosthetic restorations. The blanks named above can be machined with CAM, CAD/CAM or copy milling systems commonly used in dental medical technology to produce dental-prosthetic restorations made from zirconium oxide. This material offers excellent properties with regards to bio-compatibility, stability and aesthetics, which make it the material of choice for the production of restorations for dental crowns, inlays, onlays, veneers, and up to three unit bridges, which in turn combine the advantages of conventional casting technology and full ceramics. Dental technicians are therefore in a position to offer patients the extra added value when finishing the prosthesis.

    Metoxit Z-CAD® smile Blanks also includes Z-CAD® Liquids smile-TC. Z-CAD® Liquids smile-TC are ready for use water-based products, which are applied to the milled restauration prior to sintering to achieve individual shades. These liquids are specifically formulated for the dyeing of Z-CAD® smile blanks from Metoxit AG.

    AI/ML Overview

    The provided text describes the 510(k) summary for the Z-CAD® Smile device, which is a dental ceramic blank used for prosthetic restorations. The document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a comprehensive study report with detailed acceptance criteria and performance data in a traditional clinical trial format.

    However, I can extract the relevant information regarding performance testing and acceptance criteria for the physical properties of the device as presented, and address the other points based on the available text.


    Acceptance Criteria and Device Performance Study for Z-CAD® Smile

    The Z-CAD® Smile device is a dental ceramic blank. The performance data presented focuses on physical and biocompatibility properties, demonstrating its safety and effectiveness by comparison to a predicate device and established standards.

    1. Table of Acceptance Criteria and Reported Device Performance (Physical Properties):

    Physical PropertyUnitsZ-CAD® Smile Acceptance CriteriaReported Z-CAD® Smile PerformancePrimary Predicate Device PerformanceTesting Method
    Bending strength (biaxial)MPa$\ge 500$610> 500ASTM C1161
    Solubility (acetic acid)µg/cm$^2$$\le 100$$\le 50$< 100ISO 6872
    Radioactivity (raw material)Bq/kg$\le 200$< 45< 200GS
    Coefficient of Thermal Expansion*$10^{-6}$/KN/A (Similar)10$10.1 \pm 0.5$N/A

    *Note: The Coefficient of Thermal Expansion is listed as "Similar, does not influence safety or effectiveness" rather than a specific acceptance criterion with a pass/fail value.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the sample sizes used for each physical property test. The study appears to be a retrospective comparison based on the characteristics of the manufactured device and existing predicate devices, rather than a prospective clinical study involving patient data. The provenance of the data is from the manufacturer, Metoxit AG, Switzerland.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    Not applicable in this context. The "ground truth" for the physical properties are established by standardized testing methods (ASTM C1161, ISO 6872, GS) which yield objective measurements. Expert consensus would not be used for these types of material property evaluations.

    4. Adjudication Method for the Test Set:

    Not applicable. Physical property tests are objective measurements and do not typically involve adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done:

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for diagnostic devices that involve human interpretation of medical images or data. The Z-CAD® Smile is a material used for dental restorations, and its effectiveness is assessed through physical properties and biocompatibility, not reader interpretation.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done:

    Not applicable. There is no algorithm or AI component to the Z-CAD® Smile device described in this submission. The performance assessment is standalone in the sense that it relies on laboratory testing of the material itself.

    7. The Type of Ground Truth Used:

    The "ground truth" for the performance claims are based on:

    • Standardized Physical Property Measurements: Using methods like ASTM C1161, ISO 6872, and GS to quantify material properties (Bending Strength, Solubility, Radioactivity, Coefficient of Thermal Expansion).
    • Biocompatibility Testing Standards: Adherence to ISO 10993-1 and ISO 10993-5 (for cytotoxicity).
    • Historical Data/Predicate Device Comparison: Demonstrating that the device's characteristics are "equal or better" or "similar" to a legally marketed predicate device (Zolid FX, K152383).

    8. The Sample Size for the Training Set:

    Not applicable. There is no concept of a "training set" for physical material properties or for a device that does not involve machine learning or AI.

    9. How the Ground Truth for the Training Set Was Established:

    Not applicable, as there is no training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K072569

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-10-31

    (49 days)

    Product Code
    Regulation Number
    872.6660
    Panel
    Dental
    Age Range
    All
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Metoxit blanks are indicated for use as a substructure for porcelain fused ceramic fixed dental restorations. Limitations are listed in Table 1.

    Device Description

    Metoxit CAM-Blanks are dental ceramics, composed of yttrium-oxide stabilized tetragonal zirconium-oxide polycrystals. Metoxit CAM-Blanks are designed for the manufacturing of substructures for all-ceramic dental appliances. The dental appliance is machined either by CAD/CAM machining technique or the copying technique, then sintered to full density and strength and for the sole use of the particular patients. Metoxit CAM-Blanks are designed for the use as single tooth restorations or bridgeworks with up to two pontics in the anterior as well as in the posterior teeth region.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for Metoxit CAM-Blanks, a dental ceramic device. It seeks to demonstrate substantial equivalence to previously marketed devices rather than establishing novel performance criteria through a clinical study. Therefore, the information you've requested about acceptance criteria and a study proving the device meets those criteria, a test set, ground truth, MRMC study, and training set is not applicable in the context of this 510(k) submission.

    Here's why and what is available in the document:

    Purpose of a 510(k) Submission:
    A 510(k) submission aims to demonstrate that a new device is "substantially equivalent" to a legally marketed predicate device. This typically involves comparing technological characteristics, intended use, and materials. It often does not require new clinical studies to establish safety and effectiveness if the device is sufficiently similar to existing ones.

    Regarding the Absence of Requested Information:

    1. Acceptance Criteria and Reported Device Performance (Table): Not applicable. The submission does not define new performance acceptance criteria or report specific device performance metrics in comparison to such criteria. Instead, it claims identical technological characteristics and intended use to predicate devices.

    2. Sample Size and Data Provenance (Test Set): Not applicable. There is no "test set" in the sense of a set of cases used for evaluating an AI algorithm's performance. The "data" here refers to the device's material properties and manufacturing process, which are compared to predicate devices. There's no clinical data or image data provenance mentioned.

    3. Number of Experts and Qualifications (Ground Truth for Test Set): Not applicable. No ground truth was established by experts for a test set in this kind of submission.

    4. Adjudication Method: Not applicable. There was no need for adjudication as no expert ratings for a test set were generated.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study: Not applicable. This type of study would be relevant for evaluating AI-driven diagnostic or interpretative devices, which Metoxit CAM-Blanks are not. There is no mention of human readers or AI assistance.

    6. Standalone Performance Study: Not applicable. The device itself is a physical dental material, not a software algorithm that would have standalone performance.

    7. Type of Ground Truth Used: Not applicable. No "ground truth" (e.g., pathology, outcomes data) was established for an evaluation set. Safety and effectiveness are inferred from equivalence to predicate devices and their known performance.

    8. Sample Size for Training Set: Not applicable. This device is not an AI algorithm requiring a training set.

    9. How Ground Truth for Training Set was Established: Not applicable.


    What the Document Does Provide (Relevant to a 510(k) Summary):

    • Device Trade Name: Metoxit CAM-Blanks
    • Device Common Name: CAM-Blanks
    • Device Classification Name: Porcelain Powder for Clinical Use
    • Intended Use: Metoxit blanks are indicated for use as a substructure for porcelain fused ceramic fixed dental restorations.
    • Limitations/Indications of Use (Table 1 on page 4): This table specifies the types of restorations (single unit crowns, bridges with varying pontics) the blanks can be used for, categorized by material (TZP-A, Al-999, TZP) and process chain (porous blank or dense blank). For example, TZP-A porous blanks can be used for bridges with up to 2 pontics in both anterior and posterior regions.
    • Legally Marketed Predicate Devices: K013230, K051462 - Cercon base; K051462 Cercon base colored; K050903 XAWEX G100; K022996 Vita YZ cubes; K051705 Ivoclar ZIRCAD; K052130 Vita AL cubes; K062506 Sirona InCoris AL; K001815 DCS DC-Zirkon.
    • Technological Characteristics Comparison: The document explicitly states: "The technological characteristics between the predicate and proposed devices are identical. There is no difference in fundamental scientific technology. They are made from the same materials and have the same intended use."
    • Conclusion of Substantial Equivalence: The submission concludes that Metoxit CAM-Blanks are as safe and effective as the predicate devices and are substantially equivalent because they share the same intended use and technological characteristics, raising no new questions regarding safety and effectiveness.

    This 510(k) demonstrates "substantial equivalence" based on direct comparison of materials and intended use with existing, approved devices, rather than through new clinical performance studies against defined acceptance criteria.