510(k) Data Aggregation
(273 days)
MatriDerm pluS+ is indicated for the management of wounds including full thickness and partial thickness wounds, chronic wounds (e.g. pressure ulcers, venous ulcers, diabetic ulcers, chronic ulcers), surgical wounds (donor sites/grafts, post-Mohs surgery, post-laser surgery, podiatric, wound dehiscence), partial thickness burns, trauma wounds (abrasions, lacerations and skin tears) and draining wounds.
MatriDerm pluS+ Bi-Layer is a non-pyrogenic, single use, prescription use three-dimensional dermal matrix consisting of two layers. The first layer is composed of collagen fibers, a key component of the native extracellular matrix, and hydrolyzed elastin. The second layer consists of a medical grade silicone grid to control moisture loss from the wound providing a flexible adherent covering for the wound surface and adding increased mechanical strength to the device. The device conforms in the defect space / wound bed and includes a fibrous, porous structure that allows for fluid absorption. The device serves as a scaffold for cellular invasion and capillary growth and promotes a moist environment for the body's natural healing process. The device is supplied sterile and is provided in different sizes providing flexibility of choice based on the treatment protocol, wound location, size, and depth.
N/A
Ask a specific question about this device
(210 days)
MatriDerm is indicated for the management of wounds including:
- Full thickness and partial thickness wounds
- Chronic wounds (e.g. pressure ulcers, venous ulcers, diabetic ulcers, chronic ulcers)
- Surgical wounds (donor sites/grafts, post-Moh's surgery, post-laser surgery, podiatric, wound dehiscence)
- Partial thickness burns
- Trauma wounds (abrasions, lacerations and skin tears)
- Draining wounds
MatriDerm is a single-use three-dimensional acellular dermal matrix composed of bovine collagen fibers and bovine elastin. The device is supplied sterile and is available in different sizes providing flexibility of choice based on the treatment protocol, wound location, size and depth. The device can be cut to fit the wound.
This document describes the MatriDerm device, a collagen-elastin wound matrix, and its substantial equivalence to a predicate device (PriMatrix Dermal Repair Scaffold). The information provided focuses on the device's characteristics and safety testing rather than a comparative effectiveness study with AI assistance.
Here's a breakdown of the requested information based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly state "acceptance criteria" in a quantitative, pass/fail manner with specific numerical thresholds for performance metrics. Instead, it demonstrates substantial equivalence by comparing the MatriDerm device to a legally marketed predicate device (PriMatrix Dermal Repair Scaffold) and referencing successful completion of standard biological and bench tests.
However, based on the Performance Test Data section, we can infer some criteria and reported performance:
| Acceptance Criteria (Inferred from testing) | Reported Device Performance |
|---|---|
| Biocompatibility | MatriDerm was determined to be biocompatible based on ISO 10993-1 and FDA Guidance. Evaluated biological endpoints included physical and chemical characterization, cytotoxicity, sensitization or intracutaneous reactivity, acute systemic toxicity, subchronic and subacute toxicity, implantation, chronic toxicity, carcinogenicity, material-mediated pyrogenicity. The recommended tests were successfully performed/evaluated on the final, finished, and sterile subject device. |
| Fluid Absorbency | MatriDerm: 22.8 g/g, 11.77 g/100cm² |
| Time to Full Absorbency | MatriDerm: 21 seconds |
| Sterilization Validation | Successfully validated (terminally sterilized by gamma irradiation). |
| Bacterial Endotoxin | No specific value reported, but "successfully performed/evaluated" implies meeting acceptable limits. |
| Viral Inactivation | Compliant with ISO 22442 and FDA Guidance on Medical Device Containing Materials from Derived Animal Sources (Except for In Vitro Diagnostics). |
| Device Characterization | Successfully completed. |
| Irritation/Immunogenicity (Clinical) | Human Repeat Insult Patch Testing (HRIPT): In 60 subjects, MatriDerm was not irritating or immunogenic for a type 4/delayed type hypersensitivity. Skin Prick Testing (SPT): One out of 22 subjects had a positive wheal response at 15 minutes (product test site wheal was bigger than negative control and about the same size as positive control). At 6 hours and 1-2 day time points, all sites were negative, demonstrating a low immunogenic potential. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Biocompatibility: The text states "recommended tests were successfully performed/evaluated on final, finished and sterile subject device." It does not specify sample sizes for each individual biocompatibility test (e.g., cytotoxicity, implantation), nor the provenance of the data.
- Clinical (HRIPT): 60 subjects.
- Clinical (SPT): 22 subjects.
- Data Provenance: The document does not specify the country of origin for the clinical study data or if it was retrospective or prospective. It is implied to be prospective for HRIPT and SPT as these are typically conducted specifically for regulatory submissions.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
This section is not applicable. The evaluations for MatriDerm are primarily based on laboratory testing (biocompatibility, bench tests) and human clinical tests (HRIPT, SPT) that assess biological responses or physical properties, not expert interpretations of images or clinical cases that require "ground truth" establishment in the context of diagnostic AI.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
This section is not applicable as the studies described (biocompatibility, bench, and clinical skin tests) do not involve expert adjudication methods typically used in diagnostic assessment studies.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device is a wound dressing, not a diagnostic AI system, and therefore does not involve human readers or AI assistance in diagnostic interpretation.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
This question is not applicable as MatriDerm is a medical device (wound matrix) and not an algorithm or AI system.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
The concept of "ground truth" as typically defined for diagnostic device evaluation (e.g., pathology for cancer diagnosis) is not directly applicable here. For MatriDerm, the "truth" is established by:
- Biocompatibility: Adherence to ISO standards and FDA guidance, with confirmation through specific biological tests.
- Bench tests: Measurements against material specifications and validated test methods.
- Clinical (HRIPT/SPT): Direct physiological responses in human subjects (e.g., presence/absence of irritation, wheal formation), observed and documented by qualified personnel (not specified in detail, but standard for such clinical tests).
8. The sample size for the training set
This question is not applicable as MatriDerm is a physical medical device, not a machine learning algorithm that requires a training set.
9. How the ground truth for the training set was established
This question is not applicable as MatriDerm is a physical medical device, not a machine learning algorithm.