Found 2 results

510(k) Data Aggregation

    K Number
    K213771

    Validate with FDA (Live)

    Date Cleared
    2022-01-27

    (56 days)

    Product Code
    Regulation Number
    870.5150
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    · Merlin Aspiration Catheter
    As a part of the Merlin Aspiration System, the Merlin Aspiration Catheter is indicated for the removal of fresh, soft emboli and thrombi in the peripheral arterial system.
    · High Flow Tubing
    As a part of the Merlin Aspiration System, the High Flow Tubing is indicated to connect the guide catheter to an aspiration pump.

    Device Description

    The Merlin Aspiration System is comprised of two devices:

    • Merlin Aspiration Catheter
    • High Flow Tubing
      The Merlin Aspiration Catheter is introduced through a guide catheter or long sheath into the peripheral vasculature and guided over a guidewire to the site of primary occlusion. The Merlin Aspiration Catheter is used in conjunction with an aspiration pump, which is connected to High Flow Tubing. These devices are visible under fluoroscopy.
      Merlin Aspiration Catheter
      The Merlin Aspiration Catheter consists of single lumen, coiled construction, variable stiffness shaft with radiopaque markers on distal and proximal end of the catheter portion for angiographic visualization. The Merlin Aspiration Catheter is identical in design to the MIVI Q Distal Access Catheter (K192558).
      High Flow Tubing
      The MIVI High Flow Tubing is a sterilized and disposable (onetime use) product consisting of high-pressure PVC tubing with end fittings and an on/off switch. The MIVI High Flow Tubing connects an aspiration pump to a reperfusion catheter, providing a means for introducing vacuum during procedures. The MIVI High Flow Tubing has a flow valve that allows the physician to start and stop the flow of aspiration.
    AI/ML Overview

    The provided text describes the MIVI Merlin Aspiration System and its premarket notification (K213771) to the FDA, demonstrating substantial equivalence to a predicate device. However, the document does not contain specific acceptance criteria, reported device performance (in quantitative metrics), human reader studies, or detailed information about the ground truth establishment in the manner typically associated with studies proving device performance against acceptance criteria for AI/ML-based devices.

    The provided text focuses on the mechanical and material performance of a medical device (aspiration catheter and tubing), not an AI/ML diagnostic or therapeutic device. Therefore, the questions related to AI/ML device testing (e.g., sample size for test set, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details) are not applicable to the content provided.

    The document instead provides information about engineering bench testing, animal testing, and biocompatibility, which are standard for physical medical devices.

    Here's a summary of the information that is available, formatted to address the relevant parts of your request, and explicitly stating where information is not applicable (N/A) based on the nature of the device described:


    Acceptance Criteria and Device Performance (for a physical medical device)

    1. Table of Acceptance Criteria and Reported Device Performance

    For this physical medical device, the "acceptance criteria" are implied by the "established criteria" that the device was tested against. The "reported device performance" is simply that the device "Met established criteria" for each test. Specific quantitative criteria and performance values are not provided in this summary.

    TestAcceptance CriteriaReported Device Performance
    Merlin Aspiration Catheter
    Coating Particulate TestingMet established criteriaMet established criteria
    Coating AdhesionMet established criteriaMet established criteria
    Coating UniformityMet established criteriaMet established criteria
    Surface IntegrityMet established criteriaMet established criteria
    Tip InspectionMet established criteriaMet established criteria
    Liquid Leakage under PressureMet established criteriaMet established criteria
    Dimensional VerificationMet established criteriaMet established criteria
    Kink ResistanceMet established criteriaMet established criteria
    Push/TrackMet established criteriaMet established criteria
    Suction Flow RateMet established criteriaMet established criteria
    System Introduction / Aspiration of ThromboemboliMet established criteriaMet established criteria
    Tensile Strength-Push WireMet established criteriaMet established criteria
    Tensile Strength-Distal SectionMet established criteriaMet established criteria
    Catheter Torque TestingMet established criteriaMet established criteria
    Catheter Burst TestingMet established criteriaMet established criteria
    High Flow Tubing
    Dimensional VerificationMet established criteriaMet established criteria
    Vacuum PressureMet established criteriaMet established criteria
    Flow RateMet established criteriaMet established criteria
    Degree of CollapseMet established criteriaMet established criteria
    Simulated UseMet established criteriaMet established criteria
    Leak TestMet established criteriaMet established criteria
    Distal Male Luer Tubing Tensile StrengthMet established criteriaMet established criteria
    Female Luer Tubing Tensile StrengthMet established criteriaMet established criteria
    Proximal Male to Tubing Tensile StrengthMet established criteriaMet established criteria
    Connector to Tubing Tensile StrengthMet established criteriaMet established criteria

    Information for AI/ML Device Evaluation (Not Applicable for this Device)

    The following points are primarily relevant for AI/ML-based medical devices, which the Merlin Aspiration System is not. Therefore, specific details for these points are Not Applicable (N/A) based on the provided text. The device is a physical aspiration system, and its evaluation focuses on mechanical performance, biocompatibility, and animal testing.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • N/A. This device is a physical medical device (aspiration system), not an AI/ML device that uses a "test set" of data in the common sense for diagnostic or prognostic purposes. Performance was evaluated through bench testing and animal testing.
    • For animal testing: 8 animals (porcine carotid, renal, and subclavian vasculature). This is prospective testing in an animal model.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • N/A. Ground truth establishment for a "test set" of data points is not relevant for the physical device performance testing described.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • N/A. Not relevant for the type of device testing described.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • N/A. This is a physical medical device, not an AI-assisted diagnostic or therapeutic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • N/A. This is a physical medical device, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • N/A. For the bench testing, the "ground truth" would be established engineering specifications and physical measurements. For animal testing, it would be direct observation of the device's performance in vivo concerning safety and effectiveness in removing emboli/thrombi.

    8. The sample size for the training set

    • N/A. This device is not an AI/ML product developed using a training set.

    9. How the ground truth for the training set was established

    • N/A. Not applicable as there is no training set for this physical device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K192558

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-04-06

    (202 days)

    Product Code
    Regulation Number
    870.1250
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Q Catheter is indicated for use with compatible guide catheters in facilitating the insertion and guidance of microcatheters into a selected blood vessel in the peripheral, coronary and neuro vascular systems.

    Device Description

    The MIVI Q Catheter is a single-lumen, variable stiffness catheter with radiopaque markers on the distal and proximal end of the catheter portion for angiographic visualization. The catheter shaft has a hydrophilic coating to reduce friction during use. The proximal portion of the catheter is a stainless-steel control (push) wire. The Q catheter may be introduced via an 8F guide catheter/6F guide sheath and over a guidewire/microcatheter into the arterial vasculature until the desired vessel is reached. The catheter contains a pin vise threaded on the control wire, which may be used to advance the catheter.

    AI/ML Overview

    Acceptance Criteria and Study for MIVI Q Distal Access Catheter

    This document describes the acceptance criteria and study proving the MIVI Q Distal Access Catheter meets these criteria, based on the provided FDA 510(k) summary (K192558).

    It's important to note that this device is a medical catheter and not an AI/ML powered device, so many of the requested categories (e.g., sample size for test/training sets, number of experts, adjudication methods, MRMC studies, standalone algorithm performance, data provenance) are not directly applicable in the context of this traditional medical device submission. The study focuses on bench testing (non-clinical performance testing) to demonstrate substantial equivalence to a predicate device, rather than clinical trials or AI model validation.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Test)Criterion (Pass/Fail)Reported Device Performance (Result)
    Dimensional VerificationDevice meets specified dimensional requirements.Pass
    Corrosion ResistanceExposed metallic components show no signs of corrosion.Pass
    Tensile Strength – Pin ViseAdequate device tensile strength.Pass
    System Compatibility / Simulated UseSuccessful simulated use testing with "worst case" dimensional stack up.Pass

    2. Sample Size Used for the Test Set and Data Provenance

    1. Sample Size: Not explicitly stated for each test. For medical devices undergoing bench testing, samples are typically chosen based on statistical methods to ensure representativeness and confidence in the results, though the exact numbers are not detailed in this summary.
    2. Data Provenance: This is bench testing conducted by the manufacturer, MIVI Neuroscience, Inc., at their facilities. It is inherently prospective for the specific design verification of the modified device. Country of origin for the data is implicitly the United States, given the company's location (Eden Prairie, Minnesota).

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable. For this type of medical device submission focused on physical properties and performance, "ground truth" is established by engineering specifications, validated test methods, and industry standards, not by expert medical consensus on human-generated data. The "experts" would be the engineers and quality control personnel designing and executing the tests.


    4. Adjudication Method for the Test Set

    Not applicable. The tests are objective and quantitative (e.g., measuring dimensions, observing corrosion, measuring tensile strength). There is no "adjudication" in the sense of resolving disagreements between human observers. The outcomes are determined by adherence to pre-defined engineering and performance specifications.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done. This type of study is relevant for AI/ML-powered diagnostic devices where human readers (e.g., radiologists) interpret cases with and without AI assistance. The MIVI Q Distal Access Catheter is a physical medical device (catheter) with no AI/ML components or human diagnostic interpretation associated with its direct function.


    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was Done

    No, this is not applicable. The MIVI Q Distal Access Catheter is a physical instrument, not an algorithm.


    7. The Type of Ground Truth Used

    The "ground truth" for the performance testing is based on:

    • Engineering Specifications: Pre-defined dimensional requirements, material properties, and functional performance targets derived from the device's intended use and comparison to the predicate device.
    • Industry Standards: Relevant ISO or ASTM standards governing medical device testing (though not explicitly listed in this summary, they are implicitly followed in such submissions).
    • Predicate Device Performance: The established performance and safety profile of the legally marketed predicate device (MIVI Mi-EXT Extension Catheter, now branded as the MIVI Q Distal Access Catheter K163233) serves as a benchmark for substantial equivalence.

    8. The Sample Size for the Training Set

    Not applicable. This device does not involve machine learning or a "training set."


    9. How the Ground Truth for the Training Set was Established

    Not applicable. There is no training set for this device.