Search Filters

Search Results

Found 13 results

510(k) Data Aggregation

    K Number
    K080981

    Validate with FDA (Live)

    Date Cleared
    2008-05-02

    (25 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device measures the body temperature from the auditory canal of of a patient by means of an infrared sensor coupled with an electronic signal amplification, conditioning and digital LCD (display) unit. The device is reusable and intended for home use on people of all ages.

    Device Description

    The Mesure ear thermometer, models ST66, and ST67, are the handheld electronic thermometers that measures the temperature through the opening of the auditory canal by using a thermopile as the temperature sensor. The signal of sensor is calculated and displayed by an ASIC (Application Specific IC) - controlled circuit, which is considered the hard-wire control instead of programmable control. From the construction point of view, the ear thermometer comprises of a thermopile for the measuring sensor, a reference thermistor for comparison of temperature, a buzzer for sounding effect, an ASIC for calculating, and LCD for displaying the measuring temperature digitally for which the thermopile sensor detect the ear canal temperature through the infrared. This system uses a 3.0 V DC battery for operation of complete system whenever the battery is low, the ASIC circuit will detect the low battery condition automatically, and displays 'Low battery' in LCD display. Regarding the performance of ST66 and ST67, they were designed and verified according to the US standard ASTM E 1965-00.

    AI/ML Overview

    Here’s a breakdown of the acceptance criteria and study information based on the provided document:

    Acceptance Criteria and Device Performance

    The document indicates that the device (Ear Thermometers Models ST66 and ST67) was designed and verified according to the US standard ASTM E 1965-00. It also states compliance with EN 60601-1 and EN 60601-1-2. However, the document does not explicitly list specific numerical acceptance criteria or reported device performance metrics from these standards. Instead, it makes a general statement of compliance and states that the devices maintain the "same safety and effectiveness" as the predicate device.

    Table of Acceptance Criteria and Reported Device Performance:

    Acceptance Criteria (from ASTM E 1965-00, EN 60601-1, and EN 60601-1-2)Reported Device Performance (Models ST66 & ST67)
    Specific numerical criteria for accuracy, precision, stability, etc., as defined by ASTM E 1965-00, EN 60601-1, and EN 60601-1-2 are implied but not explicitly stated in this document."Designed and verified according to the US standard ASTM E 1965-00." "Compliance to applicable voluntary standards includes ASTM E1965: 2000, as well as EN 60601-1, and EN 60601-1-2 requirement." "Verification and validation tests contained in this submission demonstrate that the difference in the submitted demonstrate could maintain the same safety and effectiveness as that of cleared device."
    Specific criteria regarding functional safety, electromagnetic compatibility, electrical safety, etc., as defined by EN 60601-1 and EN 60601-1-2 are implied but not explicitly stated."Compliance to applicable voluntary standards includes ASTM E1965: 2000, as well as EN 60601-1, and EN 60601-1-2 requirement."

    Study Information

    1. Sample size used for the test set and the data provenance:

      • The document does not specify the sample size used for the test set.
      • The document does not specify the data provenance (e.g., country of origin, retrospective or prospective).
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in the document. The study described is an engineering verification and validation against standards, not a clinical study involving human readers or expert-established ground truth in the traditional sense of AI/medical imaging applications.
    3. Adjudication method for the test set:

      • This information is not applicable as the document does not describe a clinical study requiring expert adjudication.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC comparative effectiveness study was not done. The device is a standalone ear thermometer, not an AI-assisted diagnostic tool that would typically involve human readers.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, the evaluation described is essentially a standalone performance assessment against engineering standards (ASTM E1965-00, EN 60601-1, EN 60601-1-2). The device itself operates as a standalone measurement tool.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • The ground truth would be established by calibrated reference standards and controlled environmental conditions as specified by the engineering standards (e.g., blackbody radiators for temperature, controlled ambient temperatures, etc.). It's not clinical diagnosis-based ground truth.
    7. The sample size for the training set:

      • This information is not applicable. The device is an electronic thermometer with an ASIC (Application Specific IC) - controlled circuit, not a machine learning or AI-based device that requires a "training set" in the conventional sense. The ASIC is described as "hard-wire control instead of programmable control."
    8. How the ground truth for the training set was established:

      • This information is not applicable as there is no training set for this device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K071019

    Validate with FDA (Live)

    Date Cleared
    2007-04-27

    (17 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device measures the body temperature of a patient by means of a sensor (transducer) coupled with electronic signal amplification, conditioning and digital LCD (display) unit. The device is reusable and intended for oral, axillary or rectal temperature measurements.

    Device Description

    Digital Clinical Thermometer, Models ST8C Series

    AI/ML Overview

    The provided text is a letter from the FDA regarding a 510(k) premarket notification for a Digital Clinical Thermometer. It does not contain information about acceptance criteria for a device's performance, a study proving it meets those criteria, or any details related to AI algorithms, ground truth, expert reviews, or sample sizes for testing or training sets.

    Therefore, I cannot provide the requested information based on the input. The document merely states that the device is substantially equivalent to legally marketed predicate devices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K063542

    Validate with FDA (Live)

    Date Cleared
    2007-04-10

    (137 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The wireless temperature monitor, model ST323C, ST323F are the battery-operated electronic devices with intended use of measuring human body temperature precisely. This device is reusable and intended for armpit temperature measurement of the person of all age.

    Device Description

    The ST323C and ST323F are the wireless temperature monitor intended to be worn at left arm to monitor the armpit temperature continuously. ST323C and ST323F is composed of two operational parts, the receiver and transmitter. The receiver is the main operation unit on which, the LCD display control circuit and the main operation keys are included. And the transmitter was constructed with the thermo sensor, measuring circuit, and the signal communication unit and is to be worn at left arm for the continuous armpit temperature monitor. For the monitoring operation, both receiver and transmitter shall be switched on. Sooner after these two parts are switched on, the wireless signal communication will be set up between receiver and transmitter. The temperature monitoring signal measured at armpit will be continuously indicated on the LCD of receiver every 15 sec. This system uses a 3.0V DC battery for operation of complete system. Whenever the battery is low, the ASIC circuit will detect the low battery condition automatically, and displays 'Low battery' in LCD display. Regarding the performance of ST323C and ST323F, it was designed and verified according to the US standard ASTM E1112-00.

    AI/ML Overview

    The provided 510(k) summary describes a wireless temperature monitor (ST323C and ST323F). This device is a thermometer, not an AI/ML device, and therefore the majority of the requested information (related to AI acceptance criteria, training/test sets, expert adjudication, MRMC studies, etc.) is not applicable.

    However, I can extract and present the information that is relevant to the device's accuracy and the study conducted.

    Here's the relevant information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The submission refers to compliance with ASTM E1112: 2000 for performance verification. This standard is for electronic thermometers for intermittent determination of patient temperature. Specific acceptance criteria in terms of accuracy (e.g., ±0.1°C) and the device's reported performance against these numerical criteria are not explicitly stated in this 510(k) summary. However, the summary states: "Regarding the performance of ST323C and ST323F, it was designed and verified according to the US standard ASTM E1112-00."

    2. Sample Size Used for the Test Set and Data Provenance

    A clinical report for measurement accuracy was conducted.

    • Sample Size: Not explicitly stated in the document.
    • Data Provenance: The measurement was taken from "children less than eight years old." The location of the study (country) is not specified, but the applicant company is based in Taiwan. The study appears to be prospective as it involves taking new measurements for comparison.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This is not applicable as the device measures a physical parameter (temperature). The "ground truth" is established by a reference thermometer, not by human expert interpretation.

    4. Adjudication Method for the Test Set

    Not applicable. The "ground truth" for temperature measurement is established by a reference thermometer.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a physical measurement device, not an AI/ML diagnostic or assistive tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an AI/ML algorithm. The device's performance is inherently "standalone" in that it directly measures and displays temperature.

    7. The Type of Ground Truth Used

    The "ground truth" for the accuracy study was established by comparison with a "Mesure digital thermometer ST8A3C" which served as a reference. The method followed ASTM-E1965, which specifies standards for clinical thermometers.

    8. The Sample Size for the Training Set

    Not applicable. This is a non-AI device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. This is a non-AI device.


    Summary of Relevant Information from the provided text:

    Criteria/Study AspectDetails from 510(k) Summary
    Acceptance CriteriaCompliance with ASTM E1112: 2000 for performance. (Specific numerical accuracy criteria from the standard are not detailed in the summary).
    Reported Device PerformanceThe clinical report indicates that "ST323C is suitable for the indication of intended use" based on comparison with a predicate digital thermometer. Specific numerical performance data (e.g., mean difference, standard deviation, or percentage within a certain error margin) are not provided in this summary.
    Test Set Sample SizeNot explicitly stated.
    Test Set Data ProvenanceMeasurements taken from "children less than eight years old." Prospective study. Country of origin not specified for the clinical study itself, though the manufacturer is Taiwanese.
    Number of Experts for Ground TruthNot applicable (physical measurement).
    Qualifications of ExpertsNot applicable (physical measurement).
    Adjudication MethodNot applicable (physical measurement).
    MRMC Comparative Effectiveness StudyNot applicable (non-AI device).
    Standalone Performance StudyThe device's direct measurement is its standalone performance. The clinical study evaluated its accuracy against a predicate.
    Type of Ground Truth UsedComparison with a "Mesure digital thermometer ST8A3C" which served as a reference, following the method recommended in ASTM-E1965.
    Training Set Sample SizeNot applicable (non-AI device).
    How Training Set Ground Truth EstablishedNot applicable (non-AI device).
    Ask a Question

    Ask a specific question about this device

    K Number
    K050303

    Validate with FDA (Live)

    Date Cleared
    2005-02-25

    (17 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device measures the body temperature of a patient by means of a sensor (transducer) coupled with an electronic signal amplification, conditioning and digital 1.CD (display) unit. The device is reusable and intended for oral, axillary or rectal temperature measurements.

    Device Description

    The 5 Sec Digital Thermometer, models ST8630X/ST8640X, ST8730X/ST8740X, ST8930X/ST8940X, ST8030X/ST8040X, and ST8B30X/ST8B40X, are the electronic thermometers by using a thermistor as the temperature sensor. The signal of sensor is calculated and displayed by an ASIC (Application Specific IC) -- controlled circuit, which is considered the hard-wire control instead of programmable control.

    From the construction point of view, the digital thermometer comprises of a thermistor for measuring sensor, a reference resistor for comparison of temperature, a buzzer for sounding effect, an ASIC for calculating, and LCD for displaying the measuring temperature digitally for which the thermistor contacts.

    This system uses a 1.5V DC battery for operation of complete system whenever the battery is low, the ASIC circuit will detect the low battery condition automatically, and displays 'Low battery' in LCD display. Regarding the performance of 5 Sec Digital Thermometer, models ST8630X/ST8640X, ST8730X/ST8740X, ST8930X/ST8940X, ST8030X/ST8040X and ST8B30X/ST8B40X it was designed and verified according to the US standard ASTM E 1112-00.

    AI/ML Overview

    The provided text describes the 510(k) summary for the Mesure 5 Sec Digital Thermometer. It focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study proving the device meets specific acceptance criteria with a novel AI algorithm. Therefore, many of the requested details related to AI-specific studies (e.g., multi-reader multi-case studies, standalone algorithm performance, AI training/test set details) are not applicable or findable in this document.

    Here's an analysis based on the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of acceptance criteria with corresponding performance data in the typical sense of a clinical trial for an AI-powered device. Instead, it states that the device was designed and verified according to the US standard ASTM E 1112-00. This standard defines performance requirements for electronic thermometers.

    Acceptance Criteria (from ASTM E 1112-00, as implied by compliance):

    Acceptance Criteria CategorySpecific Criteria (Implied by ASTM E 1112-00 compliance)Reported Device Performance
    Accuracy (In Vivo)Expected to meet stated accuracy requirements for clinical thermometers (e.g., ±0.1°C or ±0.2°F for certain temperature ranges, ±0.2°C or ±0.3°F for others). Precision, repeatability, and response time are also parts of this standard.The document states, "Regarding the performance of 5 Sec Digital Thermometer... it was designed and verified according to the US standard ASTM E 1112-00." This implies that the device meets the accuracy and other performance specifications defined by ASTM E 1112-00. Specific numerical performance values beyond compliance with the standard are not provided in this summary.
    Safety and EMC (Non-Clinical)Compliance with EN 60601-1 (general safety), EN 60601-1-1 (safety requirements for medical electrical systems), and EN 60601-1-2 (electromagnetic compatibility)."Compliance to applicable voluntary standards includes ASTM E1112: 2000, as well as EN 60601-1, EN 60601-1-1, and EN 60601-1-2 requirement." This indicates the device meets these safety and EMC standards.
    Technological CharacteristicsThe new device should maintain safety and effectiveness and not alter the fundamental scientific technology compared to the predicate device."The Mesure 5 Sec Digital Thermometer... have the same intended use and technological characteristics as the cleared device... Moreover, verification and validation tests contained in this submission demonstrate that the difference... could maintain the same safety and effectiveness as that of cleared device."
    Intended UseThe device should measure body temperature for oral, axillary, or rectal measurements for home use on people of all ages.The stated intended use is identical to the predicate device, and the changes (sensor wire material and wiring construction) are described as small and not affecting intended use.

    Study Proving Acceptance Criteria:

    The study proving the device meets the acceptance criteria is referred to as "verification and validation tests" conducted according to the US standard ASTM E 1112-00 and other relevant safety standards (EN 60601-1 series). These are typically non-clinical laboratory tests to ensure accuracy, precision, and safety parameters.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not specified. For a digital thermometer conforming to ASTM E 1112-00, the "test set" would typically refer to the number of individual thermometer units tested, the number of temperature measurements taken under controlled conditions (e.g., water bath), and potentially a small number of human subjects for limited clinical validation (though in-depth clinical trials are not usually required for substantial equivalence of such devices). The document does not provide these numbers.
    • Data Provenance: Not specified. However, the company is based in Taiwan (Mesure Technology Co., Ltd., Taipei Hsien, Taiwan, R.O.C.), suggesting testing likely occurred in Taiwan or through certified labs. The submission is to the FDA for the US market. The term "retrospective or prospective" is not applicable as this concerns a medical device's physical performance, not an AI model's data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This query is typically relevant for AI/imaging devices where ground truth is established by expert consensus (e.g., radiologists). For a digital thermometer, the "ground truth" for temperature measurement is established by reference standards (e.g., highly accurate temperature baths traceable to national standards or calibration standards). Experts involved would be metrology engineers, technicians, and quality control personnel within the manufacturing and testing facilities, ensuring compliance with standards like ASTM E 1112-00. The document does not specify the number or qualifications of such experts.

    4. Adjudication Method for the Test Set

    Not applicable in the context of a digital thermometer's performance testing. Adjudication methods (e.g., 2+1, 3+1) are for resolving discrepancies among human readers interpreting medical images or data. For a thermometer, performance is objectively measured against a reference standard.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No. An MRMC study is relevant for AI systems assisting human readers in interpreting medical data (e.g., radiology). This is a simple digital thermometer, not an AI-powered diagnostic device.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    Yes, in essence. The "device" (thermometer) itself is a standalone measurement tool. Its performance is evaluated intrinsically through laboratory testing against metrological standards, not in conjunction with human interpretation in the way an AI algorithm might be. However, this is not an "algorithm" in the sense of a complex AI model. The "algorithm" here is simply the ASIC (Application Specific IC) controlling the circuit to calculate and display the temperature. Its standalone performance is what's verified against ASTM E 1112-00.

    7. The Type of Ground Truth Used

    The ground truth for a digital thermometer's accuracy and precision testing is established using calibrated reference standards, such as highly stable and accurate temperature-controlled water baths, whose temperatures are known with high certainty and are traceable to national/international measurement standards.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI algorithm that undergoes a "training phase." Its function is determined by its hardware design (thermistor, ASIC circuit) which translates a physical measurement into a digital reading based on fundamental physics and electronics, not machine learning from a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable, as explained in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K042627

    Validate with FDA (Live)

    Date Cleared
    2004-10-18

    (21 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device measures the body temperature from the auditory canal of of a patient by means of an infrared sensor coupled with an electronic signal amplification, conditioning and digital LCD (display) unit. The device is reusable and intended for home use on people of all ages.

    Device Description

    The Mesure ear thermometer, models ST61, ST62, ST63, ST64, and ST64, are the handheld electronic thermometers that measures the temperature through the opening of the auditory canal by using a thermopile as the temperature sensor. The signal of sensor is calculated and displayed by an ASIC (Application Specific IC) - controlled circuit, which is considered the hard-wire control instead of programmable control.

    From the construction point of view, the ear thermometer comprises of a thermopile for the measuring sensor, a reference thermistor for comparison of temperature, a buzzer for sounding effect, an ASIC for calculating, and LCD for displaying the measuring temperature digitally for which the thermopile sensor detect the ear canal temperature through the infrared.

    This system uses a 3.0 V DC battery for operation of complete system whenever the battery is low, the ASIC circuit will detect the low battery condition automatically, and displays 'Low battery' in LCD display. Regarding the performance of ST61, ST62, ST63,ST64 and ST65, they were designed and verified according to the US standard ASTM E 1965-00.

    AI/ML Overview

    This 510(k) submission describes the Mesure Technology Co., Ltd. Ear Thermometer, Models ST61, ST62, ST63, ST64, and ST65.

    Here's an analysis based on the provided text:

    1. Acceptance Criteria and Reported Device Performance

    The device was designed and verified according to the US standard ASTM E 1965-00. While the specific acceptance criteria values of ASTM E 1965-00 are not explicitly stated in the provided text, compliance with this standard implies that performance metrics such as accuracy, precision, and response time were evaluated and met the established thresholds within that standard for infrared thermometers.

    The document states that the devices (ST61-ST65) were developed from a cleared predicate device (ST613C & ST613F) with "only the small change in dimension, device housing, and some non-performance and safety related features." The conclusion is that these engineering differences do not affect intended use or alter the fundamental scientific technology, thereby maintaining the same safety and effectiveness as the predicate.

    Acceptance Criteria (Based on ASTM E 1965-00)Reported Device Performance (Implied by Compliance)
    Accuracy of temperature measurementCompliant with ASTM E 1965-00
    Precision/ReproducibilityCompliant with ASTM E 1965-00
    Response TimeCompliant with ASTM E 1965-00
    Safety requirements (e.g., electrical safety)Compliant with EN 60601-1, EN 60601-1-1, EN 60601-1-2
    Electromagnetic compatibility (EMC)Compliant with EN 60601-1-2

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the exact sample size used for the test set. It mentions "verification and validation tests," but the number of subjects or measurements is not provided.

    The data provenance is not explicitly stated. Given that the company is based in Taiwan (Mesure Technology Co., Ltd., Taipei Hsien, Taiwan, R.O.C.), it's plausible the testing was conducted there, but this is not confirmed. There is no information on whether the study was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the submission. For a medical device like a thermometer, the "ground truth" for temperature measurement would typically be established by highly accurate reference thermometers and controlled environments, rather than human experts, though clinicians might be involved in evaluating performance in clinical settings. The document focuses on compliance with a technical standard, ASTM E 1965-00.

    4. Adjudication Method for the Test Set

    This information is not provided. Given the nature of a thermometer performance test, an adjudication method in the sense of expert consensus on ambiguous cases is not typically applicable. Performance is usually assessed against a known reference.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size

    No, an MRMC comparative effectiveness study was not conducted. This type of study is relevant for image interpretation or diagnostic tasks where human readers' performance is influenced by AI assistance. For a standalone thermometer, this is not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, a standalone performance evaluation was conducted. The assessment focused on the device itself against established standards (ASTM E 1965-00, EN 60601-1, EN 60601-1-1, EN 60601-1-2). The device is a "hard-wire control" system, not programmable, implying that its performance is inherent to its design and hardware, without a human-in-the-loop for its basic function of temperature measurement.

    7. The Type of Ground Truth Used

    The ground truth for temperature measurement would be established by reference standards (e.g., highly accurate standard thermometers, temperature-controlled environments) as specified within ASTM E 1965-00. The document indicates compliance with this standard, which dictates how temperature accuracy and other metrics are determined.

    8. The Sample Size for the Training Set

    Not applicable. The device described uses an ASIC (Application Specific IC) - controlled circuit, which is characterized as "hard-wire control instead of programmable control." This implies it's not an AI/ML device that requires a "training set" in the conventional sense. The device's logic is fixed in hardware.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As stated above, this device does not use a training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K042013

    Validate with FDA (Live)

    Date Cleared
    2004-09-08

    (43 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device measures the body temperature of a patient by means of a sensor (transducer) coupled with an electronic signal amplification, conditioning and digital LCD (display) unit. The device is reusable and intended for oral, axillary or rectal temperature measurements.

    Device Description

    The digital clinical thermometer, models ST-8A3C, and ST-8A3F, are the electronic thermometers by using a thermistor as the temperature sensor. The signal of sensor is calculated and displayed by an ASIC (Application Specific IC) – controlled circuit, which is considered the hard-wire control instead of programmable control. From the construction point of view, the digital thermometer comprises of a thermistor for measuring sensor, a reference resistor for comparison of temperature, a buzzer for sounding effect, an ASIC for calculating, and LCD for displaying the measuring temperature digitally for which the thermistor contacts. This system uses a 1.5V DC battery for operation of complete system whenever the battery is low, the ASIC circuit will detect the low battery condition automatically, and displays 'Low battery' in LCD display.

    AI/ML Overview

    The Mesure Technology Co., Ltd. Digital Clinical Thermometer, models ST-8A3C and ST-8A3F, demonstrated compliance with ASTM E 1112-98 and other international standards to meet its acceptance criteria.

    1. Table of Acceptance Criteria and Reported Device Performance:

      Acceptance Criterion (Standard)Reported Device Performance
      ASTM E 1112-98 ComplianceCompliant
      EN 60601-1 ComplianceCompliant
      EN 60601-1-1 ComplianceCompliant
      EN 60601-1-2 ComplianceCompliant
      Low Battery DetectionDisplays 'Low battery'
    2. Sample Size Used for the Test Set and Data Provenance:
      The provided summary does not explicitly state the sample size used for the test set or the data provenance. It indicates that the device was "designed and verified according to the US standard ASTM E 1112-98," which implies testing was conducted. However, specific details about the number of measurements, patients, or the origin of the test data (e.g., country of origin, retrospective or prospective) are not provided.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:
      This information is not applicable and is not provided in the 510(k) summary. The device's performance is assessed against recognized standard specifications (e.g., ASTM E 1112-98), rather than requiring expert adjudication of results.

    4. Adjudication Method for the Test Set:
      This information is not applicable and is not provided in the 510(k) summary. Adjudication methods like 2+1 or 3+1 are typically used in studies involving human interpretation or subjective assessment, which is not the case for a digital clinical thermometer.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance:
      This is not applicable as the device is a standalone clinical thermometer measuring body temperature, not an AI-assisted diagnostic tool that would involve human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done:
      Yes, the performance evaluation described would be considered standalone. The "device performance" refers to the thermometer's ability to accurately measure and display temperature according to the specified standards without human interaction influencing the measurement itself. The device uses an ASIC for calculation, which functions as an algorithm.

    7. The Type of Ground Truth Used:
      The ground truth for the performance evaluation of the thermometer would be established by the reference standards and methodologies outlined in ASTM E 1112-98 and other applicable standards. This typically involves comparison against known, precisely controlled temperatures (e.g., from a calibrated temperature bath) or highly accurate reference thermometers, rather than expert consensus, pathology, or outcomes data.

    8. The Sample Size for the Training Set:
      The provided 510(k) summary does not mention a "training set" in the context of machine learning or AI. As a medical device that measures temperature using a thermistor and ASIC, its development and verification rely on engineering principles and adherence to standards rather than a machine learning training paradigm.

    9. How the Ground Truth for the Training Set Was Established:
      Not applicable, as a "training set" in the machine learning sense is not mentioned or relevant to the development and testing of this type of digital clinical thermometer. The "ground truth" for its accuracy and performance is established through calibration against physical standards as defined by test methodologies in ASTM E 1112-98.

    Ask a Question

    Ask a specific question about this device

    K Number
    K023956

    Validate with FDA (Live)

    Date Cleared
    2004-06-17

    (568 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The wireless temperature monitor, model ST313C/F, ST314C/F are the battery-operated electronic devices with intended use of measuring human body temperature precisely. This device is reusable and intended for wrist and abdomen temperature measurement of the baby less than eight years old, and the baby of eight years old.

    Device Description

    The Wireless temperature monitor, model ST-314C/F, are the electronic thermometers by using a thermistor as the temperature sensor. The signal of sensor is calculated and displayed by an ASIC (Application Specific IC) - controlled circuit, which is considered the hard-wire control instead of programmable control. From the construction point of view, the digital thermometer comprises of a thermistor for measuring sensor, a reference resistor for comparison of temperature, a buzzer for sounding effect, an ASIC for calculating, and LCD for displaying the measuring temperature digitally for which the thermistor contacts. This system uses a 3.0V DC battery for operation of complete system whenever the battery is low, the ASIC circuit will detect the low battery condition automatically, and displays 'Low battery' in LCD display. Reqarding the performance of ST-314C/F, it was designed and verified according to the US standard ASTM E1112-98.

    AI/ML Overview

    The provided text describes the 510(k) summary for a Wireless temperature monitor, models ST-313C/F and ST-314C/F. The information related to acceptance criteria and the study proving device performance is extracted below.

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Compliance with ASTM E1112:1998 for electronic thermometers.Device was designed and verified according to ASTM E1112-98. "The result of clinical report indicates that ST313C is suitable for the indication of intended use."
    Compliance with EN 60601-1, EN 60601-1-1, and EN 60601-1-2 requirements (Electrical safety and EMC standards).Device met these standards.
    Measurement accuracy (relative to a predicate device, Funai RT-9101, for wrist and abdomen temperature measurements in children less than eight years old)."The result of clinical report indicates that ST313C is suitable for the indication of intended use." (Specific numerical accuracy metrics are not provided in this summary, but a clinical test was performed to compare accuracy with the predicate device).

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set Sample Size: Not explicitly stated as a number of subjects. The text mentions "clinical report for measurement accuracy" and "The measurement was taken by both ST313C and RT-9101 at wrist and abdomen location for the children less than eight years old."
    • Data Provenance: The study was a "clinical report," suggesting prospective data collection. The location is not specified, but the submitter is based in Taiwan, R.O.C.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided in the given text. For a temperature monitor, the "ground truth" would likely be established by a reference thermometer or highly accurate direct measurement, rather than human experts.

    4. Adjudication Method for the Test Set

    This information is not provided in the given text. It is unlikely to be relevant for a device measuring a physical parameter directly.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and effect size.

    • MRMC Study: No, this was not an MRMC study. It was a comparison of device performance against a predicate device and standards, not a study evaluating human readers' effectiveness with or without AI assistance.
    • Effect Size: Not applicable.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, the primary evaluation appears to be a standalone performance assessment of the device. The "clinical report for measurement accuracy" compared the new device (ST313C) directly against a predicate device (Funai RT-9101). The device itself is an electronic thermometer, and its accuracy is assessed independently of human interpretation.

    7. The Type of Ground Truth Used

    The ground truth for accuracy was established by:

    • Compliance with ASTM E1112:1998 standard for electronic thermometers.
    • Comparison of measurements against a predicate device (Funai model RT-9101). The predicate device's measurements would serve as the "ground truth" for comparative accuracy.

    8. The Sample Size for the Training Set

    The provided text does not mention a training set as this device is a hardware electronic thermometer and not an AI/machine learning algorithm that requires training data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this type of device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K041056

    Validate with FDA (Live)

    Date Cleared
    2004-05-12

    (19 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device measures the body temperature of a patient by means of a sensor coupled with an electronic signal amplification, conditioning and digital LCD (display) unit. The device is reusable and intended for oral, axillary or rectal temperature measurements for home use on people of all ages.

    Device Description

    The 10-second digital thermometer, ST87, ST89, ST80, and ST8B, are the electronic thermometers by using a the temperature sensor. The signal of sensor is calculated and displayed by an ASIC (Application Specific IC) - controlled circuit, which is considered the hard-wire control instead of programmable control. From the construction point of view, the digital thermometer comprises of a thermistor for measuring sensor, a reference resistor for comparison of temperature, a buzzer for sounding effect an ASIC for calculating, and LCD for displaying the measuring temperature digitally for which the thermistor contacts. This system uses a 1.5V DC battery for operation of complete system whenever the battery is low, the ASIC circuit will detect the low battery condition automatically, and displays 'Low battery' in LCD display. Regarding the performance of ST87, ST89, ST80, and ST8B, it was designed and verified according to the US standard ASTM E 1112-98.

    AI/ML Overview

    This is a 510(k) summary for a digital thermometer, which is a medical device. The information provided heavily references standards for medical devices and does not describe AI/ML software. Therefore, most of the requested information regarding AI/ML clinical studies is not available in the provided text.

    Here's an analysis based on the provided text:

    Acceptance Criteria and Device Performance (for a digital thermometer)

    Acceptance Criteria (Standard Reference)Reported Device Performance (Compliance)
    ASTM E1112: 1998Designed and verified according to ASTM E1112-98.
    EN 60601-1Compliance stated.
    EN 60601-1-1Compliance stated.
    EN 60601-1-2Compliance stated.

    Note: The provided document is for a digital thermometer, not an AI/ML device. Therefore, the acceptance criteria are based on medical device standards for thermometers, focusing on accuracy, safety, and electromagnetic compatibility.


    The following information is not available in the provided text as it pertains to AI/ML device studies, which this device is not.

    • Sample size used for the test set and the data provenance: Not applicable. The device is a digital thermometer, not an AI/ML algorithm. Testing would typically involve physical measurements against reference standards, not data sets.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a thermometer is established through calibration against known temperature standards.
    • Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This is not an AI-assisted device.
    • If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is a standalone physical device, not an algorithm.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc.): For a digital thermometer, the ground truth would be established by calibrated reference thermometers or temperature sources traceable to national or international standards. This is not mentioned in detail, but implied by adherence to ASTM E1112.
    • The sample size for the training set: Not applicable. There is no AI/ML model to train.
    • How the ground truth for the training set was established: Not applicable. There is no AI/ML model to train.
    Ask a Question

    Ask a specific question about this device

    K Number
    K013779

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2002-03-29

    (136 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device measures the body temperature of a patient by means of a sensor coupled with an electronic signal amplification, conditioning and digital LCD (display) unit. The with an Cicottonio orginal and intended for oral, axillary or rectal temperature measurements for home use on people of all ages.

    Device Description

    Not Found

    AI/ML Overview

    Here's an analysis of the provided text regarding the Clinical Electronic Thermometer and its acceptance criteria and study:

    It's important to note that the provided documents (K013779) are from an FDA 510(k) clearance letter for a Clinical Electronic Thermometer. This type of device is typically cleared based on substantial equivalence to a predicate device, and the details of clinical studies, as commonly understood for AI/ML or more complex diagnostic devices, are often not extensively detailed in the clearance letter itself. Instead, performance data is usually submitted in a separate 510(k) summary or report.

    Based on the provided text, a comprehensive study description as requested for AI/ML devices is not present. However, I can extract the available information and highlight what is not present.


    Acceptance Criteria and Device Performance (Based on available information and common thermometer standards):

    For a clinical electronic thermometer, the primary acceptance criteria revolve around accuracy and precision according to recognized standards. While the specific numerical acceptance criteria and reported performance are NOT EXPLICITLY STATED in the provided documents, I will infer them based on typical requirements for this type of device and what would generally be tested to establish substantial equivalence. The document primarily focuses on the regulatory clearance itself.

    Acceptance CriteriaReported Device PerformanceComments from Document
    Accuracy (e.g., within ±0.1°C or ±0.2°F)Not explicitly stated in the provided text. However, for substantial equivalence, the device's accuracy would need to meet or exceed that of the predicate device and relevant industry standards (e.g., ASTM E1112 for electronic thermometers). The 510(k) submission would have contained this data.The document indicates "We have reviewed your Section 510(k) premarket notification... and have determined the device is substantially equivalent...". This implies that the performance data (including accuracy) was found to be comparable to a legally marketed predicate device.
    Repeatability/PrecisionNot explicitly stated in the provided text. This would also be a critical parameter, demonstrating consistent readings under identical conditions.Similar to accuracy, this would have been part of the technical data submitted for substantial equivalence.
    Response TimeNot explicitly stated in the provided text. The time taken to display a stable reading.
    Operating Range (e.g., 32-43°C)Not explicitly stated in the provided text. The range of temperatures the device can accurately measure.
    Environmental Conditions (Temperature, Humidity)Not explicitly stated in the provided text. Performance under varying environmental conditions.
    BiocompatibilityNot explicitly stated in the provided text. Materials in contact with the patient are safe.This is a general control requirement for medical devices.
    Electrical Safety (e.g., IEC 60601-1)Not explicitly stated in the provided text. Ensures safe operation regarding electrical hazards.This is a general control requirement for medical devices.
    Electromagnetic Compatibility (EMC)Not explicitly stated in the provided text. Ensures the device does not interfere with or is not interfered by other electronic devices.This is a general control requirement for medical devices.
    Durability/LifetimeNot explicitly stated in the provided text. The expected lifespan and robustness of the device.
    Readability of DisplayNot explicitly stated in the provided text. The clarity and ease of reading the digital display.The "digital LCD (display) unit" is mentioned, implying a visual display for temperature readings.
    Intended Use Fulfilled (Oral, Axillary, Rectal for home use)"The device measures the body temperature of a patient... for oral, axillary or rectal temperature measurements for home use on people of all ages."This is directly from the Indications for Use statement, confirming the device is intended for these specific measurement sites and user environment.

    Detailed Study Information (Based on provided text):

    1. Sample size used for the test set and the data provenance:

      • Sample Size: Not specified in the provided documents.
      • Data Provenance: Not specified in the provided documents (e.g., country of origin, retrospective or prospective). For a thermometer, testing is typically prospective under controlled laboratory and simulated clinical conditions.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • Not Applicable / Not Specified: For a clinical electronic thermometer, ground truth is typically established by physical reference standards (e.g., certified temperature baths, black body sources) or comparison to highly accurate reference thermometers, rather than expert human interpretation.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not Applicable / Not Specified: Adjudication methods like 2+1 or 3+1 are used for tasks involving human interpretation or subjective assessment (e.g., image analysis, diagnosis). For a thermometer, "adjudication" is not typically relevant as the output is a direct numerical measurement.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, not specified/applicable. This type of study is relevant for AI-assistant devices or diagnostic tools where human interpretation is involved. This device is a standalone temperature measurement device, not an AI assistant.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

      • Yes, implicitly. The device itself is standalone. Its "performance" refers to its accuracy, precision, and other technical specifications, independent of human interpretation or assistance in the measurement process. The operator simply places the device and reads the displayed temperature. The 510(k) submission would include data on the device's standalone performance characteristics compared to standards and the predicate device.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

      • Not explicitly stated, but typically highly accurate reference thermometers in controlled environments. For a thermometer, ground truth is established by comparing the device's readings against well-calibrated, highly accurate reference thermometers or certified temperature standards in controlled laboratory settings (e.g., water baths at precise temperatures).
    7. The sample size for the training set:

      • Not Applicable / Not Specified: This device does not typically involve a "training set" in the context of machine learning or complex algorithms. Its performance is based on its physical design, sensors, and electronic components, which are calibrated and verified.
    8. How the ground truth for the training set was established:

      • Not Applicable / Not Specified: As there is no "training set" in the AI/ML sense, this question is not relevant to this device. Calibration and verification processes (using reference standards) ensure the device's intrinsic accuracy.

    Summary of Limitations based on the provided text:

    The provided documents are primarily the FDA 510(k) clearance letter and the Indications for Use statement. These documents confirm regulatory clearance and the intended applications of the device. They do not contain the detailed technical report, performance data, or study methodologies that would typically be found in the 510(k) summary or supporting documentation submitted to the FDA. Therefore, specific numerical acceptance criteria, reported performance values, study sample sizes, and detailed methodology are not available in this text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K011254

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2001-07-10

    (77 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 2