Search Filters

Search Results

Found 1 results

510(k) Data Aggregation

    K Number
    K955719

    Validate with FDA (Live)

    Device Name
    SICKLE-2000
    Date Cleared
    1996-08-14

    (240 days)

    Product Code
    Regulation Number
    864.7415
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SICKLE-2000 is an in-vitro centrifugal solubility test that detects Hemoglobin S, the abnormal hemoglobin found in sickle cell disease and sickle cell trait.

    Device Description

    SICKLE-2000 is an in-vitro centrifugal solubility test that detects Hemoglobin S, the abnormal hemoglobin found in sickle cell disease and sickle cell trait.

    AI/ML Overview

    Acceptance Criteria and Device Performance Study for SICKLE-2000

    This document outlines the acceptance criteria for the SICKLE-2000 device and details the studies conducted to demonstrate its performance.

    1. Table of Acceptance Criteria and Reported Device Performance

    Based on the provided information, the acceptance criteria are implicitly derived from the comparative and blind studies, focusing on accurate identification of Hemoglobin S.

    Acceptance Criteria CategorySpecific CriteriaSICKLE-2000 Device Performance
    Clinical Performance (Accuracy)Correctly identify samples with and without Hemoglobin S (Hb S).- In comparison study with Sicklequik (N=87), results were identical for 66 non-Hb S, 15 heterozygous (AS), and 6 homozygous (SS) samples. - In blind study (N=99), correctly identified all 59 normal samples and all 40 heterozygous S samples.
    Detection LimitDetect Hemoglobin S at clinically relevant concentrations.Detected Hemoglobin S at a concentration of 2.0g/dl using 10 heterozygous S samples.
    InterferenceNo significant interference from common blood components (lipemia, bilirubin, hemolysis).- No triglyceride interference up to 1000mg/dl. - Hemolysis does not interfere. - Bilirubin interference:      - Normal (AA) blood: No interference up to 60mg/dl.      - Homozygous (SS) blood: No interference up to 60mg/dl.      - Heterozygous (AS) blood: No interference up to 15mg/dl. At 30 mg/dl and 60 mg/dl, the aqueous phase shows color changes (orange/darker) but implied non-interference to result.
    StabilityMaintain performance over a specified shelf life.- Accelerated aging indicated no significant change in activity over a two-year period. - Real-time stability studies indicated product is good for at least 12 months.

    2. Sample Size Used for the Test Set and Data Provenance

    • Test Set 1 (Comparison Study):
      • Sample Size: 87 patients.
      • Data Provenance: Not explicitly stated, but implies clinical samples used for comparison with a predicate device. Retrospective or prospective nature is not specified.
    • Test Set 2 (Blind Study):
      • Sample Size: 99 samples (59 normal, 40 heterozygous S).
      • Data Provenance: Not explicitly stated, but implied to be clinical samples previously tested by hemoglobin electrophoresis. Retrospective or prospective nature is not specified.
    • Test Set 3 (Detection Limit Study):
      • Sample Size: 10 heterozygous S samples.
      • Data Provenance: Not explicitly stated.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    The document does not explicitly mention the number or qualifications of experts used to establish the ground truth for the test sets.

    • For Test Set 1 (Comparison Study), the ground truth was implicitly established by the results of the predicate device, Sicklequik, and blood typing for Hemoglobin S (66 contained no Hb S, 15 AS, 6 SS).
    • For Test Set 2 (Blind Study), the ground truth for the 99 samples was established by "hemoglobin electrophoresis," which is a standard laboratory method, rather than expert consensus on device results. This implies a gold standard laboratory method was the ground truth.

    4. Adjudication Method for the Test Set

    Not applicable. The ground truth was established by laboratory methods (predicate device results, hemoglobin electrophoresis) rather than subjective expert interpretation requiring adjudication.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This device is an in-vitro diagnostic test, not an AI-assisted diagnostic tool that involves human readers interpreting images or data with and without AI assistance to improve their performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies described are standalone performance studies of the SICKLE-2000 device itself. The device is an in-vitro centrifugal solubility test, and its results are read directly, not requiring human interpretation in conjunction with an algorithm's output. The "blind study" specifically tests the device's ability to correctly identify samples without human bias.

    7. The Type of Ground Truth Used

    • For Test Set 1 (Comparison Study): The ground truth was primarily the results obtained from the predicate device, Sicklequik, and the classification of blood samples based on the presence/absence and type of Hemoglobin S (no Hb S, AS, SS). This can be considered a comparative ground truth against an established method.
    • For Test Set 2 (Blind Study): The ground truth was established by hemoglobin electrophoresis, which is a definitive laboratory method for identifying hemoglobin types.

    8. The Sample Size for the Training Set

    The document does not mention a "training set" in the context of device development or machine learning. The studies described are performance validation studies.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set (in the context of machine learning) is mentioned.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1