Found 2 results

510(k) Data Aggregation

    K Number
    K222790

    Validate with FDA (Live)

    Device Name
    F65 Laser System
    Manufacturer
    Date Cleared
    2023-06-14

    (272 days)

    Product Code
    Regulation Number
    890.5500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The subject F65 has connection capability with the following available treatment handpieces, for multi-application treatment options. All handpieces are designed for aesthetic and dermatological skin procedure applications, as follows:

    The Intense Pulsed Light (IPL) handpiece with a spectrum of 400-1200 nm (with 9 different filters) is indicated for:

    • o Benign epidermal lesions, including dyschromia, hyperpigmentation, melasma, and ephelides (freckles)
    • o Cutaneous lesions, including warts, scars and striae
    • o Benign cutaneous vascular lesions, including port wine stains, hemoangiomas, facial, truncal and leg telangiectasias, erythema of rosacea, angiomas and spider angiomas, poikiloderma of Civatte, leg veins and venous malformations
    • o Removal of unwanted hair and to effect stable long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles
    • o Mild to moderate inflammatory Acne (Acne vulgaris)

    The Nd: YAG Laser handpiece with a wavelength of 1064 nm (Multi-Spot Nd: YAG) is indicated for:

    • o The coagulation and hemostasis of vascular lesions and soft tissue, including the treatment and clearance of superficial and deep telangiectasias (venulectasias) and reticular veins (0.1-4.0 mm. diameter) of the leg
    • o The removal of unwanted hair and to effect table long term, or permanent* hair reduction in skin types I-V through selective targeting of melanin in hair follicles
    • o The non-ablative treatment of facial wrinkles

    The F65 module and handpiece, with wavelength of 1565 nm, are indicated for:

    • o Use in dermatological procedures requiring fractional skin resurfacing and coagulation of soft tissue
    • o Improving the appearance of scalp hair in adult males with Fitzpatrick skin type I to IV who are seeking treatment for hair loss.

    The Q-Switched Nd:YAG Laser Handpiece with a wavelength of 1064 nm is indicated for:
    o Removal of dark tattoos
    o Treatment of pigmented lesions

    *Note
    Permanent hair reduction is defined as long-term, stable reduction in the number of hairs regrowing when measured at 6, 9, and 12 months after completion of treatment regime.

    Device Description

    The Lumenis Be F65 Laser System is intended to be used for the new indication for Improving the appearance of scalp hair in adult males with Fitzpatrick skin type I to IV seeking treatment for hair loss, as well as, for the identical set of functions as the predicate device. The description here is limited to the system module required to support the new indication which is the subject of this 510(k):

    • The laser console includes a power supply unit, a Laser module (+ aiming beam), a control unit and a PC module. The laser beam is transferred to the handpiece via fiber optics.
    • . The handpiece includes a scanner module – built of two mirrors that are controlled by two motors directing the Laser beam to desired target tissue.
    • . The handpiece uses a tip (user may choose from two types) to align the Laser beam focus with the target tissue, enabling a treatment shape size and a tip thermo-electric cooler (TEC).

    The F65 is intended for professional use only.

    AI/ML Overview

    The provided FDA 510(k) summary only describes the clinical studies for the F65 module and handpiece with 1565 nm wavelength, specifically for the new indication of "Improving the appearance of scalp hair in adult males with Fitzpatrick skin type I to IV who are seeking treatment for hair loss." It does not provide detailed acceptance criteria and performance data for all the device's indications, nor does it present the information in a clear table format for acceptance criteria vs. device performance.

    Based on the information provided for the 1565nm module and its new indication:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria for the new indication ("Improving the appearance of scalp hair in adult males with Fitzpatrick skin type I to IV who are seeking treatment for hair loss") are implicitly derived from the study results presented in the document. The document states a primary efficacy endpoint of "demonstrating improvement in more than 70% of subjects" for the retrospective study.

    Acceptance Criteria (for 1565nm module, hair loss indication)Reported Device Performance (Retrospective Study)
    No reported adverse eventsNo adverse events were reported or documented in patient files during treatment or follow-up period across all 98 subjects.
    Primary efficacy endpoint: Improvement in scalp hair appearance in >70% of subjects with confirmed Androgenic Alopecia (implicit)For 44 subjects with a confirmed diagnosis of androgenic alopecia, the assessment of improvement in scalp hair appearance yielded a correct identification in 97.7% (95% Cl: 87.8% -99.7%) of the cases by a group of 3 blinded reviewers. This demonstrates improvement in more than 85% of patients, supporting the primary efficacy endpoint of demonstrating improvement in more than 70% of subjects.
    Overall efficacy endpoint: Improvement in scalp hair appearance across all subjects (explicit/observed)For the total of 98 subjects included in the performance analysis, the assessment of improvement in scalp hair appearance yielded a correct identification in 96.9% (95% CI: 91.38 -98.5%) of the cases by a group of 3 blinded reviewers.

    2. Sample Sizes and Data Provenance

    • Preclinical Study (Animal Testing): Sample size for the preclinical study is not specified, only that it "was conducted to evaluate the Lumenis F65 subject module for hair follicle regeneration capabilities using both Low Level Laser Therapy (LLLT) and untreated controls."
    • Clinical Studies:
      • Prospective, Single Arm Study:
        • Sample Size: 32 subjects enrolled, 25 completed.
        • Diagnosis Breakdown: 18 with Androgenic Alopecia, 7 with other diagnoses (evaluated separately).
        • Hair Count Data: 5 male subjects with androgenic alopecia, 18 female subjects with androgenic alopecia.
        • Data Provenance: Not explicitly stated, but clinical studies are generally prospective. No country of origin is mentioned.
      • Retrospective, Single Arm Study:
        • Sample Size: 98 subjects evaluated for performance.
        • Diagnosis Breakdown: 44 with confirmed Androgenic Alopecia, 54 with no documented diagnosis (presumably hair loss but not specifically diagnosed as androgenic alopecia).
        • Data Provenance: Retrospective. No country of origin is mentioned.

    3. Number of Experts and Qualifications for Ground Truth

    • Retrospective Study: "a group of 3 blinded reviewers" were used for the assessment of improvement in scalp hair appearance.
    • Qualifications of Experts: The specific qualifications (e.g., dermatologist, years of experience) of these 3 blinded reviewers are not specified in the provided document. They are only referred to as "reviewers."

    4. Adjudication Method for the Test Set

    • For the retrospective study, where 3 blinded reviewers assessed improvement, the adjudication method is not explicitly stated. It's mentioned they "yielded a correct identification" in a certain percentage of cases. This implies either:
      • All 3 reviewers had to agree (unlikely for such high percentages given typical inter-reader variability without prior agreement).
      • A majority vote (2 out of 3) was used.
      • One expert's assessment was deemed primary, or they collectively discussed and reached a consensus.
        The document does not detail the process.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • No Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done comparing human readers with AI assistance vs. without AI assistance.
    • The studies described are for the device's direct effect on hair growth, not for evaluating human reader performance with or without an AI. The "blinded reviewers" in the retrospective study were assessing the clinical outcome (improvement in scalp hair appearance), not the performance of an AI algorithm.

    6. Standalone (Algorithm Only) Performance

    • This document describes a medical device (laser system) that directly treats patients, not an AI algorithm. Therefore, there was no standalone (algorithm only) performance study as it's not an AI device. The "performance" refers to the clinical outcome achieved by the laser system.

    7. Type of Ground Truth Used

    • For the clinical studies, particularly the retrospective study, the "ground truth" for showing improvement in scalp hair appearance was established by expert consensus/review (3 blinded reviewers).
    • For the prospective study, quantifiable metrics like mean increase in hair count per cm² from baseline were reported, which can be considered objective outcomes data.

    8. Sample Size for the Training Set

    • This document describes physical medical device studies (laser system) for a 510(k) submission, not an AI model requiring a training set. Therefore, there is no concept of a "training set" in the context of this submission. The "training" here refers to the device's design and manufacturing through engineering principles and prior knowledge, not machine learning.

    9. How the Ground Truth for the Training Set Was Established

    • As there is no AI model or "training set" in the context of this device submission, this question is not applicable.
    Ask a Question

    Ask a specific question about this device

    K Number
    K220467

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2022-05-18

    (90 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The UltraPulse Alpha CO2 Laser System, Delivery Devices and Accessories (members of the UltraPulse CO2 Laser Systems Family) are indicated for use in surgical or aesthetic applications requiring: ablation, vaporization, excision, or coagulation of soft tissue in medical specialties including: dermatology and plastic surgery (aesthetic), podiatry , gynecology, general and oral surgery, dental and oral surgery and genitourinary surgery as follows:

    The UltraPulse SurgiTouch and UltraPulse Encore CO2 Laser System, Delivery Devices and Accessories (members of the UltraPulse CO2 Laser Systems Family) are indicated for use in surgical or aesthetic applications requiring: ablation, vaporization, excision, or coagulation of soft tissue in medical specialties including: aesthetic (dermatology and plastic surgery), podiatry, otolaryngology (including laparoscopy), neurosurgery, orthopedics (soft tissue), arthroscopy (knee), general and thoracic surgery (including open and endoscopic), dental and oral surgery and genitourinary surgery as follows:

    Device Description

    The Lumenis Be Family of UltraPulse CO2 Surgical and Aesthetic Laser Systems, Delivery Devices and Accessories (subject devices) consist of the following devices:

    Laser Systems

    • o UltraPulse Alpha (also written as UltraPulse α)
    • UltraPulse Encore o
    • UltraPulse SurgiTouch o

    Delivery Devices and Accessories

    • o DeepFX Microscanner and disposable tips
    • UltraScan CPG Microscanner o
    • o TrueSpot 2.0 mm Collimated Handpiece
    • o 0.2 mm and 1.0 mm Focused Incisional Handpieces (also called Standard Handpieces)

    Each UltraPulse laser system is an advanced computer-controlled device with RF - modulated CO2 laser tube technology that emits laser beams at a wavelength of 10,600 or 11,100 nm. Each has an additional diode laser in the red visible spectrum emitting along the same optical path as the CO2 to serve as an aiming beam.

    Attached to each laser system are the delivery devices and accessories that direct the laser beams to the intended treatment site in the format as selected by the user.

    AI/ML Overview

    This document is a 510(k) premarket notification decision letter from the FDA regarding a CO2 Laser System for surgical and aesthetic applications. It does not contain information about the acceptance criteria or a study proving the device meets those criteria in the context of an AI/algorithm-based medical device.

    The document states:

    • "For the labeling changes described in the 510(k), no testing was required." (Page 12)
    • The entire submission is focused on "labeling changes" and a new manufacturer name, not performance evaluation of a new or modified device functionality that would require clinical or technical studies demonstrating performance metrics.
    • The device being cleared is a traditional medical laser system, not an AI or algorithm-based device. Therefore, the concepts of "ground truth," "expert consensus," "training set," "test set," "MRMC study," and "human readers improve with AI assistance" are not applicable to this document.

    Therefore, I cannot extract the requested information (acceptance criteria table, sample size, expert information, adjudication methods, MRMC study, standalone performance, ground truth types, training set size, and ground truth establishment for training set) from the provided text. The document pertains to a regulatory clearance based on substantial equivalence for a non-AI device with labeling changes, not a performance study for an AI-enabled device.