510(k) Data Aggregation

    K Number
    K180767

    Validate with FDA (Live)

    Date Cleared
    2019-01-22

    (305 days)

    Product Code
    Regulation Number
    880.2910
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FEMON Banana Thermometer is a battery operated electronic device with intended use of measuring human body temperature. This device is intended for armpit temperature measurement from ages 29 days and older.

    Device Description

    The FEMON Banana Thermometer is a thermometer composed of a body patch to be directly attached to the patient's skin on armpit. The temperature information is then to be retrieved from the patch to the mobile phone application via Bluetooth.

    AI/ML Overview

    The FEMON Banana Thermometer is intended for armpit temperature measurement for patients aged 29 days and older.

    1. Table of Acceptance Criteria and Reported Device Performance:

    Acceptance CriteriaReported Device Performance
    Accuracy (°C): ±0.1°C between 35°C and 42°C±0.1°C: 35~42°C

    Important Note: The provided document states that the subject device's accuracy criteria are ±0.1°C for the entire range of 35-42°C, which is different and stricter than the predicate device's accuracy criteria (±0.1°C for 37-39°C and ±0.2°C for 35-37°C and 39-42°C). However, the document does not explicitly state an "acceptance criteria" but rather reports the device's performance. For the purpose of this table, the reported device performance for accuracy is assumed to be the acceptance criterion it aims to meet.

    2. Sample size used for the test set and the data provenance:

    The document states that "Clinical Accuracy Performance Testing (per ISO80601-2-56)" was performed. However, it does not specify the sample size used for this test set or the data provenance (e.g., country of origin, retrospective or prospective).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document for the clinical accuracy testing.

    4. Adjudication method for the test set:

    This information is not provided in the document for the clinical accuracy testing.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done:

    No, an MRMC comparative effectiveness study is not mentioned in the provided document. The study performed was a clinical accuracy performance test for the device itself against a standard, not a comparison of human readers with and without AI assistance.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    Yes, a standalone performance study was done in the form of "Clinical Accuracy Performance Testing (per ISO80601-2-56)." This tests the device's accuracy in measuring temperature directly, without human interpretation or intervention in the measurement process itself.

    7. The type of ground truth used:

    The ground truth for the clinical accuracy performance testing was established by following the "ISO80601-2-56" standard. This standard typically dictates methods for determining the true or reference temperature, often involving a reference thermometer or a calibrated temperature bath to provide the gold standard for comparison.

    8. The sample size for the training set:

    This information is not applicable or not provided. The FEMON Banana Thermometer is an electronic thermometer, not an AI/machine learning algorithm that requires a training set in the conventional sense. Its performance is validated through direct measurement accuracy testing.

    9. How the ground truth for the training set was established:

    This information is not applicable as there is no training set for an electronic thermometer in the context of an AI/ML algorithm.