Search Filters

Search Results

Found 5 results

510(k) Data Aggregation

    K Number
    K244049

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-05-28

    (148 days)

    Product Code
    Regulation Number
    892.1720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Europa (Alternative: AiRTouch) portable X-ray system is intended for use by trained or qualified doctors to produce diagnostic X-ray images of extremities in adult and pediatric(over 12 years old) patients. These images are obtained using anatomical structures captured by film or image processing systems (workstation) after an examination involving radiation exposure with cassette IP, CR, or DR (portable flat panel). Only intended for stand-mounted use.

    Device Description

    EUROPA(Alternative: AiRTouch) portable X-ray System generates X-ray with variable tube current and voltage (kVp) to take diagnostic X-rays of extremities for adult and pediatric patients. It operates on 22.2VDC supplied by a rechargeable Lithium-Ion Polymer battery pack. The X-ray tube head, X-ray controls and power source are assembled into a single portable X-ray enclosure. EUROPA(Alternative: AiRTouch) portable X-ray System includes high voltage generator, X-ray tube, a control board (PCB), rechargeable battery, LCD user interface, X-ray beam limiting device, and a remote-control switch (hand switch). Operating principle is that x-ray generated by high voltage electricity into x-ray tube, which penetrates patients' extremities and makes x-ray images on receptor. EUROPA(Alternative: AiRTouch) Portable X-ray System is intended to be used by trained clinicians or technicians for both adult and pediatric(over 12 years old) patients.

    EUROPA(Alternative: AiRTouch) portable X-ray is intended to be mounted on a tripod stand.

    AI/ML Overview

    The provided FDA 510(k) clearance letter for the Europa Portable X-ray System is primarily a regulatory document affirming substantial equivalence to a predicate device. It details the device's technical specifications and how they compare to the predicate, as well as the non-clinical testing performed to meet electrical safety and radiation protection standards.

    However, the document does NOT contain the information typically found in a study proving a device meets acceptance criteria related to its diagnostic performance (e.g., accuracy, sensitivity, specificity, or human perception improvement with AI assistance). The "clinical images" mentioned appear to be for general image quality assessment, not a structured clinical study with quantifiable metrics against defined acceptance criteria.

    Therefore, I cannot extract the detailed information requested regarding acceptance criteria and a study proving the device meets performance acceptance criteria. The document focuses on showing the device is substantially equivalent to a predicate device based on its technical specifications and adherence to safety and performance standards for X-ray equipment.

    The relevant section related to "performance" in this document refers to:

    • Non-clinical testing: Adherence to various IEC and CFR standards related to electrical safety, electromagnetic disturbances, radiation protection, and X-ray tube assemblies.
    • "Clinical images" assessment: A qualitative statement that "Clinical images taken with EUROPA 85 and EUROPA 90 have presented overall appropriate image quality of the anatomical structures, both bony and soft tissues of the upper and lower extremities." This is not a quantitative performance study.

    To answer your request based solely on the provided text, the following points are all "Not Provided" or "Not Applicable" because the document is a regulatory clearance letter focused on substantial equivalence rather than a detailed performance study report.


    Summary of Device Acceptance Criteria and Performance (Based on the provided K244049 document):

    The provided document, an FDA 510(k) clearance letter, primarily serves to demonstrate that the Europa (Alternative: AiRTouch) Portable X-ray System is substantially equivalent to a predicate device (EZER Portable X-ray System, K193535). The "acceptance criteria" discussed are largely related to technical specifications, electrical safety, radiation performance, and adherence to recognized standards, rather than a specific clinical performance study measuring diagnostic accuracy or reader improvement.

    No clinical study details regarding diagnostic performance metrics (e.g., sensitivity, specificity, AUC) are provided in this document. The "clinical images" mentioned appear to have been part of a qualitative assessment of overall image quality for regulatory submission, not a formal quantitative performance study.


    Detailed Breakdown of Requested Information:

    FeatureDescription (Based on Provided Document)
    1. Table of Acceptance Criteria & Reported PerformanceAcceptance Criteria (Implicit/Inferred from Non-Clinical Testing):
    - Compliance with IEC 60601-1 (Basic safety & essential performance)
    - Compliance with IEC 60601-1-2 (Electromagnetic disturbances)
    - Compliance with IEC 60601-1-3 (Radiation protection in diagnostic X-ray equipment)
    - Compliance with IEC 60601-2-28 (X-ray tube assemblies)
    - Compliance with IEC 60601-2-54 (X-ray equipment for radiography and radioscopy)
    - Compliance with IEC 62133 (Secondary cells and batteries safety)
    - Compliance with 21 CFR 1020 Subchapter J (Performance Standards for Ionizing Radiation Emitting Products), 21 CFR 1020.30 (Diagnostic x-ray system and their major components), and 21 CFR 1020.31 (Radiographic Equipment).
    - Qualitatively "appropriate image quality of the anatomical structures, both bony and soft tissues of the upper and lower extremities" (for clinical images taken with EUROPA 85 and EUROPA 90).
    Reported Device Performance (against the above):
    - "Testing was performed successfully according to the following standards:" (followed by the list above).
    - "Clinical images taken with EUROPA 85 and EUROPA 90 have presented overall appropriate image quality of the anatomical structures, both bony and soft tissues of the upper and lower extremities."
    - "Performance evaluation (test) reports and the device inspection report confirmed that the EUROPA(Alternative: AiRTouch) Portable X-ray Systems are suitable for its intended use and the instructions for use."
    2. Sample size and data provenance (test set)Not Provided. The document mentions "Clinical images taken," but does not specify the sample size (number of images, number of patients), or their provenance (country of origin, retrospective/prospective collection method). It's likely these were internal test images for qualitative assessment.
    3. Number/Qualifications of experts for ground truthNot Provided. No details on experts used for "ground truth" (if any beyond qualitative assessment by internal personnel) are given.
    4. Adjudication method for test setNot Provided. Since no formal clinical performance study with clear "ground truth" establishment is detailed, no adjudication method is mentioned.
    5. MRMC comparative effectiveness study? Effect size?Not Applicable/Not Provided. The device is an X-ray system, not an AI-assisted diagnostic tool. Therefore, an MRMC study comparing human readers with/without AI assistance is not relevant to this type of device based on the information provided.
    6. Standalone (algorithm only) performance?Not Applicable. This is a hardware X-ray system, not a diagnostic algorithm. Therefore, "standalone" performance in the sense of an algorithm operating independently is not relevant. The performance is the device's ability to produce quality X-ray images.
    7. Type of ground truth usedImplicit/Qualitative. For the image quality assessment, the "ground truth" seems to be a qualitative judgment by evaluators that the images showed "appropriate image quality of the anatomical structures." There is no mention of pathology, clinical outcomes, or expert consensus in a formal, quantifiable sense. For the safety and performance standards (IEC/CFR), the ground truth is adherence to the specified technical parameters and safety limits defined by those standards.
    8. Sample size for training setNot Applicable. This document describes an X-ray system, not an AI/Machine Learning model that would typically have a "training set" for an algorithm.
    9. How ground truth for training set was establishedNot Applicable. As above, no AI/ML model is described, so no training set or its associated ground truth establishment process is relevant to this document.
    Ask a Question

    Ask a specific question about this device

    K Number
    K193535

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-05-07

    (139 days)

    Product Code
    Regulation Number
    892.1720
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EZER Portable X-Ray system is a portable x-ray source with a fixed tube current and voltage for producing diagnostic x-ray images of extremities using digital or film image receptors. Its use is intended to be used by trained clinician or technicians for both adult and pediatric subjects age 2 and above.

    It is not intended to replace a radiographic system with variable tube current and voltage (KVp) which may be required for full optimization of image quality and radiation exposure for different exam types.

    Device Description

    EZER Portable X-ray System generates and controls X-ray with a fixed tube current and voltage (kVp) in order to take diagnostic X-rays of extremities for adult and pediatric patients. It operates on 22.2VDC supplied by a rechargeable Lithum-Ion Polymer battery pack. The X-ray tube head. X-ray controls and power source are assembled into a single hand-held enclosure. EZER Portable X-ray System includes high voltage generator, X-ray tube, a control board (PCB), rechargeable battery, LCD user interface, X-ray beam limiting cone, and a remote control switch (hand switch). Operating principle is that x-ray generated by high voltage electricity into x-ray tube, which penetrates extremities and makes x-ray images on receptor. INTEL stick PC is integrated with EZER so the user can see X-ray image from LCD display without a computer. EZER Portable X-ray System is intended to be used by trained clinicians or technicians for both adult and pediatric patients.

    The embedded 7" TFT display in EZER Portable X-ray is not intended to be used for diagnosis.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the EZER Portable X-ray System, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a table of "acceptance criteria" with specific numerical thresholds for performance metrics. Instead, it describes a subjective comparison study against a predicate device. The primary "acceptance criterion" appears to be demonstrating similar overall quality to the predicate device.

    Performance AspectAcceptance Criteria (Implied)Reported Device Performance
    Image QualitySimilar overall quality compared to the predicate device (NOMAD MD Handheld X-ray System)Images taken from EZER have similar quality overall compared with Nomad MD. Images from EZER show no other radiographic abnormalities or issues with diagnostic images.
    Safety - Operators/PatientsConformance to relevant safety standards and guidance documents (IEC, EN, CFR standards)Testing performed successfully according to relevant standards; Device development utilized "Radiation Safety Consideration for X-ray Equipment Designed for Hand-Held Use" and "The Content of Premarket Submissions for Software Contained in Medical Devices" guidance documents.
    Intended UseMeeting the stated Indications for Use (diagnostic x-ray images of extremities for adults and pediatrics age 2+)Confirmed as suitable for its intended use and user instruction.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size: Not explicitly stated in terms of number of images or cases. The document mentions "Images taken from the predicate and subject devices were reviewed." This implies a set of images was used, but the quantity is not specified.
    • Data Provenance: Not explicitly stated. The study details do not mention the country of origin of the data or whether it was retrospective or prospective.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    • Number of Experts: One (1) expert.
    • Qualifications: An American board-certified radiologist. No specific number of years of experience is provided.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. Only a single expert reviewed and rated the images, so there was no multi-reader consensus or adjudication process described. The expert directly compared and rated the images.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    • MRMC Study: No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. Only a single radiologist performed the image comparison.
    • Effect Size: Not applicable, as no MRMC study was conducted.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    • Standalone Performance: Not applicable. The EZER Portable X-ray System is an X-ray imaging device, not an AI algorithm. The study described is an evaluation of the image output generated by the device, which is then interpreted by a human expert. The device does include an embedded display for reference, but it's explicitly stated "not intended to be used for diagnosis."

    7. The Type of Ground Truth Used

    • Type of Ground Truth: "Expert comparison/rating." The ground truth was established by a single American board-certified radiologist who reviewed and rated images from both the subject device and the predicate device. This is a form of subjective expert evaluation.

    8. The Sample Size for the Training Set

    • Training Set Sample Size: Not applicable. The EZER Portable X-ray System is a hardware device (X-ray system) and not an AI or machine learning algorithm that requires a training set in the typical sense.

    9. How the Ground Truth for the Training Set Was Established

    • Training Set Ground Truth: Not applicable, as the device is not an AI/ML system requiring a training set with established ground truth.
    Ask a Question

    Ask a specific question about this device

    K Number
    K200022

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2020-04-03

    (88 days)

    Product Code
    Regulation Number
    892.1650
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The FLUSION-9001 fluoroscopic C-arm Mobile X-ray System is intended to provide fluoroscopic and radiographic imaging of the patient during diagnostic, surgical and interventional procedures. Clinical applications may include but are not limited to digital subtraction angiography, orthopedical, abdominal, vascular, cardias, critical care and emergency room procedures.

    Device Description

    The FLUSION-9001 fluoroscopic C-arm Mobile X-ray System consists of a high voltage (HV) inverter generator, a tube support unit, an X-ray beam limiting device, mobile cart, a detector, operating software, and a tube, and is primarily used in a hospital for diagnosis of diseases in skeletal. respiratory and urinary systems such as the skull, spinal column, chest, abdomen, extremities, and other body parts. This device is not intended to be used for mammography applications.

    FLUSION-9001 fluoroscopic C-arm Mobile X-ray System is a solution to produce radiological images of patient during medical operations. This inverter control X-ray unit visualizes the anatomical structure on screen, which is obtained by X-ray fluoroscopy and a flat panel detector. This system can be applied in emergency room, operation room, cast room or etc. of a hospital.

    AI/ML Overview

    The provided document is a 510(k) summary for the FLUSION-9001 Fluoroscopic C-arm Mobile X-ray System. It focuses on demonstrating substantial equivalence to predicate devices based on technical specifications and adherence to recognized standards.

    Therefore, the document does not contain information about acceptance criteria for device performance in a clinical study, or a study specifically designed to prove the device meets such criteria in terms of diagnostic accuracy or impact on human readers.

    The information provided describes the device's technical specifications and compares them to predicate and reference devices, along with compliance with relevant industry standards and regulations for medical devices.

    Here's a breakdown of what is available and what is not:

    1. A table of acceptance criteria and the reported device performance

    • Not Available: The document does not provide a table of acceptance criteria for diagnostic performance outcomes (e.g., sensitivity, specificity, accuracy) or reported performance metrics from a clinical study. The tables present technical specifications of the device and its comparators (e.g., X-ray tube heat capacity, detector DQE, pixel matrix).

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Not Available: There is no mention of a test set, sample size, or data provenance related to a clinical performance study. The testing performed focuses on non-clinical aspects like electrical, mechanical, software, and risk management.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • Not Available: Since there is no mention of a clinical test set or ground truth establishment, this information is not provided.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • Not Available: There is no information on any adjudication method for a clinical test set.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not Available: The document describes a fluoroscopic C-arm Mobile X-ray system, which is an imaging device, not an AI-powered diagnostic tool. Therefore, an MRMC comparative effectiveness study involving AI assistance for human readers would not be applicable or expected for this device nor is it mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not Available: As noted above, this is an imaging acquisition device, not an algorithm, so standalone performance is not relevant in the context of diagnostic algorithms.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • Not Available: There is no mention of ground truth as it pertains to clinical diagnostic performance.

    8. The sample size for the training set

    • Not Available: This device is a hardware system, not a machine learning model, so there is no training set in the context of AI/ML.

    9. How the ground truth for the training set was established

    • Not Available: See point 8.

    What the document does describe (Non-clinical testing for substantial equivalence):

    The document details the non-clinical testing performed to establish substantial equivalence (refer to section VIII. Non clinical testing and IX. Conclusion):

    • Standards Adherence: The device successfully underwent testing according to various IEC and EN standards related to medical electrical equipment, including safety, electromagnetic compatibility, and specific requirements for X-ray equipment (e.g., EN 60601-1, EN 60601-1-2, IEC 60601-1-3, IEC 60601-2-43, IEC 60601-2-54).
    • Guidance Document Utilization: The development informed by FDA guidance documents regarding software in medical devices, 510(k) submissions for solid-state X-ray imaging devices, cybersecurity, and pediatric information for X-ray imaging devices.
    • Regulatory Compliance: Conformance to 21 CFR 1020 Subchapter J (Performance Standards for Ionizing Radiation Emitting Products), 21 CFR 1020.30 (Diagnostic x-ray system and their major components), and 21 CFR 1020.32 (Fluoroscopic Equipment).

    The conclusion states that based on these tests and comparisons to predicate devices, the FLUSION-9001 system is considered substantially equivalent in its design, mechanical and electrical performance, making it suitable for its intended use.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190935

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-12-19

    (253 days)

    Product Code
    Regulation Number
    872.1800
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EZER Portable X-Ray system is intended to be used by trained dental technicians as an extraoral x-ray source for producing diagnostic x-ray intraoral image receptors. Its use is intended for both adult and pediatric subjects.

    Device Description

    EZER Portable X-ray System generates and controls X-ray in order to diagnose of tooth and iaw. It operates on 22.2VDC supplied by a rechargeable Lithum-Ion Polymer battery pack. The X-ray tube head, X-ray controls and power source are assembled into a single hand-held enclosure. EZER Portable X-ray System includes high voltage generator, X-ray tube, a control board (PCB), rechargeable battery, LCD user interface, X-ray beam limiting cone, and a remote control switch (hand switch). Operating principle is that x-ray generated by high voltage electricity into x-ray tube, which penetrates tooth and makes x-ray images on receptor. INTEL stick PC is integrated with EZER so the user can see X-ray image from LCD display without a computer. EZER Portable X-ray System is intended to be used by trained dentists or dental technicians. Its use is intended for both adult and pediatric subjects.

    The embedded 7' TFT display in EZER Portable X-ray is not intended to be used for diagnosis.

    AI/ML Overview

    The provided text is a 510(k) Summary for the EZER Portable X-ray System, which is a device for producing diagnostic dental X-ray images. The document focuses on demonstrating substantial equivalence to a predicate device rather than presenting a clinical study with detailed acceptance criteria and performance metrics for an AI/algorithm-based diagnostic device.

    Therefore, the following information regarding acceptance criteria and a study proving device performance will be limited by the nature of the provided text, as it describes a hardware device (X-ray system) and not an AI-driven diagnostic tool that would typically have the requested performance metrics.

    There is no information in the provided document about the acceptance criteria and study that proves the device meets specific performance criteria related to the diagnostic accuracy of an AI algorithm.

    The document primarily focuses on:

    • Safety and Efficacy of the X-ray Hardware: This is demonstrated through adherence to various IEC and EN standards for medical electrical equipment, radiation safety, and electromagnetic compatibility, as well as compliance with 21 CFR regulations for diagnostic X-ray systems.
    • Substantial Equivalence: The primary goal of the 510(k) submission is to show that the EZER Portable X-ray System is substantially equivalent to a legally marketed predicate device (Dexcowin ADX 4000W; iRay D4) in terms of indications for use and technological characteristics.

    Based on the provided text, I cannot complete the requested tables and sections with specific AI-related performance metrics. However, I can extract the relevant information regarding the device's technical specifications and the type of non-clinical testing performed.

    Here's a breakdown of what can be extracted and what cannot be, based on your request and the provided text:


    1. A table of acceptance criteria and the reported device performance

    The document does not specify "acceptance criteria" in terms of diagnostic accuracy (e.g., sensitivity, specificity for detecting a condition) because EZER is an X-ray source, not a diagnostic interpretation algorithm. Instead, it confirms compliance with safety and performance standards for an X-ray device.

    The performance metrics mentioned pertain to the X-ray output and electrical characteristics for the device itself.

    Acceptance Criteria Category (Derived from non-clinical testing)Reported Device Performance (EZER Portable X-ray System)
    Physical Dimensions & Weight
    Size233.5(L) x 162.9(W) x 116.9(H) [mm] (9.2" x 6.4" x 4.6")
    Weight2.6 kg
    X-ray Output & Control
    Total filteringOver 1.5mmAl (inherent filtration : 0.8mmAl)
    Cone diameter6 cm
    Exposure time range0.03 ~ 1.30 seconds in 0.01 increments
    Time Accuracy$\pm$ (10%)
    mA (current)2 mA fixed
    kVp (voltage)60 kV fixed
    WaveformConstant Potential (DC)
    Electrical Power
    Energy sourceRechargeable 22.2 V DC Lithium Ion Polymer battery pack
    Safety & Standards Compliance(Compliance confirmed by testing against standards)
    Electrical SafetyIEC 60601-1:2006/A1:2013
    Radiation Safety (Diagnostic X-ray equipment)IEC 60601-1-3:2008+A1:2013, IEC 60601-2-65:2012, EN 60601-2-65:2013, 21 CFR 1020 Subchapter J, 21 CFR 1020.30, 21 CFR 1020.31
    Battery SafetyIEC 62133:2012, EN 62133:2013
    Electromagnetic Compatibility (EMC)EN 60601-1-2:2015, IEC 61000-4-3: 2006 + A2: 2010

    Study Proving Device Meets Acceptance Criteria:

    The document states: "Testing was performed successfully according to the following standards" and lists multiple IEC, EN, and CFR standards. This non-clinical testing demonstrates that the device itself (the X-ray system) meets the technical and safety performance criteria required for such a device. There is no mention of a diagnostic performance study in the context of disease detection or image interpretation, as the device is an X-ray source.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This information is not applicable and therefore not provided in the document. The testing described is non-clinical performance testing of the hardware device against engineering standards, not a clinical study on diagnostic accuracy involving a test set of patient data.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and therefore not provided. No clinical "ground truth" was established as this is not a diagnostic algorithm.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and therefore not provided.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    This information is not applicable and therefore not provided. The device is an X-ray source, not an AI assistance tool for image interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    This information is not applicable and therefore not provided. The device is a standalone X-ray system; it does not perform algorithmic diagnosis. The embedded 7" TFT display is not intended to be used for diagnosis.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    This information is not applicable and therefore not provided.

    8. The sample size for the training set

    This information is not applicable and therefore not provided. The device is not an AI algorithm that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable and therefore not provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K180663

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2018-05-18

    (65 days)

    Product Code
    Regulation Number
    892.1680
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LaVid FMTS System is a stationery X-ray system intended for obtaining radiographic images of various of the human body, both pediatrics and adults, in a clinical environment.

    The LaVid FMTS System is not intended for mammography, interventional, or fluoroscopy use.

    Device Description

    The LaVid FMTS System is a stationery X-ray system intended for obtaining radiographic images of various of the human body, both pediatrics and adults, in a clinical environment.

    AI/ML Overview

    This document is an FDA 510(k) clearance letter for the LaVid FMTS DIAGNOSTIC X-RAY SYSTEM. It indicates that the device is substantially equivalent to legally marketed predicate devices and outlines general regulatory provisions. However, the provided text does not contain information regarding:

    • Acceptance criteria for device performance.
    • A study proving the device meets acceptance criteria.
    • Reported device performance metrics.
    • Sample sizes used for test or training sets.
    • Data provenance.
    • Number or qualifications of experts for ground truth establishment.
    • Adjudication methods.
    • Multi-reader multi-case (MRMC) comparative effectiveness studies.
    • Effect size of human reader improvement with AI.
    • Standalone algorithm performance studies.
    • Type of ground truth used.
    • How ground truth for the training set was established.

    The document primarily focuses on the regulatory clearance for the device, its indications for use, and general regulatory compliance information. It does not delve into the technical performance evaluation that would typically include the requested details.