Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K232031

    Validate with FDA (Live)

    Device Name
    LUNA 3D
    Date Cleared
    2024-03-29

    (266 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LAP LUNA 3D system is used to support reproducible patient positioning and monitor patient surface motion during radiotherapy and radiosurgery treatments and during CT simulation for radiation therapy planning. LAP LUNA 3D may also be used to guide patients through breath-holds. The LAP LUNA 3D system can be used for all patients, undergoing radiotherapy and radiosurgery treatments and radiosurgery treatments and CT simulation for radiation therapy planning.

    Device Description

    The LUNA 3D optical camera system captures the current 3D patient skin surface with one or multiple camera pods. The software calculates the spacial deviations between the captured live surface and a reference surface within a selected region of interest using a registration algorithm. Reference surfaces may be generated with the optical camera system or may be imported using data received from Treatment Planning Systems (TPS). Based on the registration results the user can adjust the patient position for reproducible patient positioning relative to the treatment isocenter. During the imaging (simulation) and treatment delivery process, the system continuously calculates the deviations between live- and reference surface for patient motion monitoring to ensure that the patient position remains within pre-defined tolerances.

    AI/ML Overview

    Here's an analysis of the acceptance criteria and study information for the LUNA 3D device based on the provided FDA 510(k) summary:

    1. Table of Acceptance Criteria and Reported Device Performance

    The FDA 510(k) summary provides limited detailed acceptance criteria and mostly focuses on comparison to a predicate device. The performance claims primarily relate to accuracy.

    Acceptance Criteria / Performance MetricReported Device Performance (LUNA 3D)
    Positioning AccuracyTarget registration errors (as measured using calibration phantom) < 0.5mm for all couch angles.
    Respiratory Tracking (RMS Errors)Surface displacements can be tracked with RMS errors < 0.5mm over 10 or more breathing cycles.
    Electrical SafetyIEC 60601-1 compliant.
    EMC StandardIEC 60601-1-2 compliant.
    Photobiological SafetyCompliant with IEC 60601-2-57 and IEC 62471.
    Warm-up DriftTested (compliance to established technical specifications).
    LatencyTested (compliance to established technical specifications).
    Multi-camera AccuraciesTested (compliance to established technical specifications).

    2. Sample Size Used for the Test Set and Data Provenance

    The provided summary does not specify a sample size for any clinical test set or the data provenance (e.g., country of origin, retrospective/prospective). The performance testing described is "non-clinical performance testing" which typically refers to bench testing rather than studies involving human subjects or clinical data.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    The provided summary does not mention any test set requiring expert ground truth or the involvement of experts for this purpose. The performance testing relies on a "calibration phantom."

    4. Adjudication Method for the Test Set

    As no expert-derived ground truth or clinical test set is described, no adjudication method is mentioned or applicable in the provided document.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A MRMC comparative effectiveness study was not performed or described in the provided summary. The summary focuses on the device's standalone performance and comparison of technical characteristics to a predicate device, not on human reader improvement with or without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Yes, a standalone performance assessment was conducted. The "non-clinical performance testing in terms of warm-up drift, latency, and multi-camera accuracies of the camera pods" and the reported "positioning accuracy" and "respiratory tracking" metrics (using a calibration phantom) reflect the algorithm's performance without direct human interaction during the measurement process. The device's function is to provide information for humans to use, but the core measurements are automated.

    7. Type of Ground Truth Used

    The ground truth for the non-clinical performance testing was established using a calibration phantom. This is a physical object with known dimensions and properties used to assess the accuracy of the imaging system.

    8. Sample Size for the Training Set

    The provided summary does not specify a sample size for any training set. Given that this is a 510(k) for a medical charged-particle radiation therapy system accessory and not a purely AI/ML enabled diagnostic device, details on training data are often less emphasized in the summary unless substantial AI model development is a primary feature. The LUNA 3D is described as an "optical camera system" with "software calculates the spacial deviations... using a registration algorithm," which implies algorithms but not necessarily deep learning that would require a large training set in the typical sense.

    9. How the Ground Truth for the Training Set Was Established

    As no training set is described, no information on how its ground truth was established is provided.

    Ask a Question

    Ask a specific question about this device

    K Number
    K200907

    Validate with FDA (Live)

    Date Cleared
    2020-06-03

    (58 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The THALES 3D MR SCANNER system is designed to measure the characteristic radiation beam for an irradiation unit. The beam data is stored in the software program and can be exported from there for use during commissioning in a treatment planning system (TPS) or for quality assurance. The THALES 3D MR SCANNER system can be used with irradiation units with integrated magnetic resonance imaging and a magnetic field of less than or equal to 0.35 tesla.

    Device Description

    The THALES 3D MR SCANNER system consists of a water phantom with a carriage system, a manual control unit and a software program. A water basin with the corresponding pump, a control unit with an electrometer, and a motor controller are integrated in the carriage system. The software program is used to control the three axes of the water basin, perform measurements and analyse measured data.

    The software program has the following properties: Acquisition of measured data, analysis, and recording of photon beams with the water phantom. Generation and export of beam data libraries for a treatment planning system (TPS) for commissioning of irradiation units. The software program is equipped with a central database and/or a file system for central storage of all recorded data for different irradiation units within a large hospital network with different sites. The software program offers the option of defining different user levels to support the typical clinical workflows.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the THALES 3D MR Scanner:

    Important Note: The provided FDA 510(k) summary focuses on demonstrating substantial equivalence to a predicate device, not necessarily on a comprehensive efficacy study with predefined acceptance criteria for its performance metrics as one might see in a clinical trial. The "acceptance criteria" here are more about meeting safety, electromagnetic compatibility, and dosimeter standards, and showing that its operational parameters are comparable or superior without introducing new risks. There is no information about a comparative effectiveness study (MRMC) with human readers, nor a standalone algorithm performance study as would be applicable for AI/CADe devices.


    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Criteria (from predicate comparison/standards)Reported Device Performance (THALES 3D MR Scanner)
    Intended UseQA for radiation therapy systems, commissioning, and other occasions; use with integrated MRI (up to 1.5T for predicate)QA for radiation therapy systems, commissioning, and other occasions; use with integrated MRI with magnetic field ≤ 0.35 Tesla
    Scanning Range408mm * 355mm * 248mm (predicate)380mm * 380mm * 242mm
    Tank ShapeSquareSquare
    Wall MaterialPMMAPMMA
    Chamber Holder OrientationHorizontal and verticalHorizontal and vertical
    Motor Scanning ModeStep by stepStep by step
    Step Size0.1mm0.1mm
    Maximum Speed30mm/s (predicate)40mm/s
    Position Accuracy+/-0.1mm (predicate)+/- 0.25mm (at 0.35Tesla)
    Reproducibility+/-0.1mm (predicate)+/- 0.25mm (at 0.35Tesla)
    MRI CompatibilityUp to 1.5 Tesla (predicate)Up to 0.35 Tesla
    Electrometer Channels2 entries2 entries
    Electrometer Resolution10 fA (predicate)1 fA
    Electrical SafetyIEC 61010IEC 61010
    Dosimeter TestingIEC 60731IEC 60731
    EMCIEC 60601-1-2:2007 (predicate)IEC 60601-1-2:2015
    MRI SafetyN/A (implied by predicate's MRI compatibility)Tested in accordance with applicable ASTM standards

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a "test set" in the context of an AI/ML algorithm or a specific patient-data based study. Instead, it details performance characteristics of a physical measurement device (water phantom) for radiation therapy. The performance testing refers to evaluating the device's physical and electrical characteristics against standards (IEC, ASTM). Therefore, there is no information about a sample size or data provenance related to patient data. The "test set" implies physical testing of the device itself.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. The ground truth for this device's performance is established by engineering and physics standards (IEC, ASTM) for safety, compatibility, and measurement accuracy, not by human expert consensus on clinical data.

    4. Adjudication Method for the Test Set

    Not applicable. There's no clinical data or expert review process described that would require an adjudication method.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The document makes no mention of an MRMC study or any study involving human readers or comparative effectiveness in a clinical workflow with or without AI assistance. This device is a measurement tool, not an AI diagnostic or assistive tool in the traditional sense.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    No. This is not an AI algorithm but a physical device with integrated software to control movement, acquire, analyze, and export measurement data. Its "performance" refers to how accurately it physically measures radiation beams, not its ability to interpret images or suggest diagnoses.

    7. The Type of Ground Truth Used

    The ground truth used for validating the device's performance comes from established engineering and medical device standards:

    • Electrical and mechanical safety: IEC 61010
    • EMC: IEC 60601-1-2
    • Dosimeter requirements: IEC 60731
    • MRI safety: Applicable ASTM standards
    • Performance metrics (e.g., position accuracy, electrometer resolution) are likely validated against manufacturer-defined specifications derived from engineering principles and comparison to the predicate device.

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/ML model that requires a "training set" of data in the conventional sense. Its software controls hardware and processes raw measurement data based on established physics, not learned patterns from a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI/ML algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K173926

    Validate with FDA (Live)

    Device Name
    DORADOnova MR3T
    Date Cleared
    2018-09-14

    (262 days)

    Product Code
    Regulation Number
    892.5050
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The DORADOnova MR3T system assists in reproducible patient positioning during radiation therapy planning and is utilized during treatment planning workflow. The DORADOnova MR3T system projects visual laser beams which can be positioned at specified coordinates via a control system. The position coordinates are typically chosen by the operator or provided by an external system (e.g. treatment planning system). The DORADOnova MR3T system is allowed to be used in MR environment with a magnetic field, less or equal to 3 Tesla.

    Device Description

    The LAP moving laser positioning system contains laser devices and a control system. The laser system is provided in a configuration that enables the movement of two projection planes by using an appropriate composition of motorized and fixed laser devices. The installation of the system is realized in a combined bridge solution.

    AI/ML Overview

    This FDA 510(k) summary (K173926) describes the DORADOnova MR3T system, a medical device used for reproducible patient positioning during radiation therapy planning. The document primarily focuses on demonstrating substantial equivalence to predicate devices rather than providing a detailed clinical study with acceptance criteria for an AI/algorithm-driven device.

    Therefore, many of the requested elements for describing the acceptance criteria and the study that proves the device meets them (especially those related to AI/algorithm performance, ground truth establishment, expert adjudication, and MRMC studies) are not applicable to this specific device and document, as it is a laser positioning system, not an AI-powered diagnostic or therapeutic device.

    However, based on the provided text, I can extract the relevant performance specifications and how they were tested.


    Device Name: DORADOnova MR3T (LAP moving laser positioning system)

    Device Type: Medical charged-particle radiation therapy system (21 CFR 892.5050, Class II) - Specifically, a laser positioning system.


    1. Table of Acceptance Criteria and Reported Device Performance

    Since this is a physical positioning device and not an AI/algorithm, "acceptance criteria" are tied to its physical specifications and safety standards, largely demonstrated through bench testing and compliance with recognized standards.

    Feature / CriterionAcceptance Criteria (Stated or Implied)Reported Device Performance / Characteristics
    Intended UseMust assist in reproducible patient positioning during radiation therapy planning and be utilized during treatment planning workflow, by projecting visual laser beams positionable at specified coordinates. Must be usable in MR environment (≤ 3 Tesla).The DORADOnova MR3T system "assists in reproducible patient positioning during radiation therapy planning and is utilized during treatment planning workflow. The DORADOnova MR3T system projects visual laser beams which can be positioned at specified coordinates via a control system... The DORADOnova MR3T system is allowed to be used in MR environment with a magnetic field, less or equal to 3 Tesla." (Identical to intended use).
    Functional PrincipleMust use projection of visual light/laser lines on the patient's skin surface, with reflection aiding medical staff to determine patient's position."Projection of a visual laser lines on the patient's skin surface. Reflection of the laser lines aide's professional medical staff to determine the patient's position." (Similar to predicate/reference devices).
    Patient Reference Position InformationMust allow coordinates for 3 laser intersection points to be manually input or imported from a treatment planning system for reproducible patient positioning."The patient reference position information is the coordinates for 3 laser intersection points. They are either manually input into the device by the operator or imported from the treatment planning system." (Similar to predicate/reference devices).
    Positioning AccuracyMust be comparable to or better than predicate/reference devices. Reference device: +/- 0.1 mm. Predicate device: max mean: 0.27mm; max std dev: 0.65mm.+/- 0.1 mm (Matches reference device performance, superior to predicate device in terms of stated accuracy).
    Resolution / Line WidthMust be comparable to or better than predicate/reference devices. Reference device: <1 mm at 4m distance. Predicate device: <1 mm.0.5 mm (blue) at 4m distance; <1 mm (red&green) at 4m distance (Meets or exceeds criteria, blues is better than reference).
    Compatibility with MR EnvironmentMust be safe and perform as intended in MR environments up to 3 Tesla."For use in hospital environment, including MRI environment up to and including 3 Tesla. MRI safety testing and labeling of the proposed device are in accordance with applicable ASTM standards. Bench testing in MR environment up to three Tesla did not show any negative impact on the performance of the proposed device." (Meets criterion).
    Electrical SafetyMust comply with IEC 60601-1."The proposed device has been tested in respect to electrical and mechanical safety in accordance with IEC 60601-1..." (Complies).
    EMCMust comply with IEC 60601-1-2."...EMC in accordance with IEC 60601-1-2..." (Complies).
    Laser SafetyMust comply with IEC 60825-1 and U.S. performance standard 21CFR 1040.10. Laser Class 2."...laser safety in accordance with IEC 60825-1. The proposed device complies with U.S. performance standard 21CFR 1040.10." and "Laser Class: 2." (Complies).
    Bio-compatibilityMust have no contact with patient."No contact with patient." (Identical to predicate/reference devices, indicating compliance with the implicit "no patient contact" requirement for biocompatibility).

    Study Details for DORADOnova MR3T Performance:

    This device's performance demonstration relies on engineering specifications, adherence to international safety standards, and bench testing, rather than a clinical study in the typical sense for AI/diagnostic devices.

    1. Sample Size Used for the Test Set and Data Provenance:

      • Not applicable in the typical AI/diagnostic study sense.
      • Performance testing described is focused on physical properties, safety, and compatibility. Tests were likely conducted on a defined number of device units or prototypes. Data provenance is implied to be from internal company testing conducted in Germany (location of LAP GmbH Laser Applikationen).
    2. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

      • Not applicable. Ground truth for device performance (e.g., accuracy, resolution, safety compliance) is established through standardized engineering measurements and compliance with regulatory standards, not expert clinical interpretation of data.
    3. Adjudication Method for the Test Set:

      • Not applicable. This falls under engineering and regulatory compliance testing, not clinical diagnosis where adjudication is needed.
    4. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

      • No. This is not an AI-assisted diagnostic or CAD device requiring MRMC studies. Its function is to project light for patient positioning.
    5. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done:

      • Not applicable. This is a physical system with an operator, not an AI algorithm. Its performance directly relates to its ability to project lasers with specified accuracy.
    6. The Type of Ground Truth Used:

      • Engineering/Physical Measurement Ground Truth: For parameters like accuracy, resolution, and line width, the ground truth is established by precise metrological measurements using calibrated equipment.
      • Standard Compliance Ground Truth: For safety (electrical, laser, EMC, MRI), the "ground truth" is adherence to the specified international and national standards (IEC 60601-1, IEC 60601-1-2, IEC 60825-1, 21 CFR 1040.10, ASTM standards for MRI safety).
    7. The Sample Size for the Training Set:

      • Not applicable. This device does not use machine learning or AI models that require a "training set."
    8. How the Ground Truth for the Training Set Was Established:

      • Not applicable, as there is no training set.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1