Thales 3D Mr Scanner
K200907 · Lap GmbH Laser Applikationen · IYE · Jun 3, 2020 · Radiology
Device Facts
| Record ID | K200907 |
| Device Name | Thales 3D Mr Scanner |
| Applicant | Lap GmbH Laser Applikationen |
| Product Code | IYE · Radiology |
| Decision Date | Jun 3, 2020 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 892.5050 |
| Device Class | Class 2 |
Intended Use
The THALES 3D MR SCANNER system is designed to measure the characteristic radiation beam for an irradiation unit. The beam data is stored in the software program and can be exported from there for use during commissioning in a treatment planning system (TPS) or for quality assurance. The THALES 3D MR SCANNER system can be used with irradiation units with integrated magnetic resonance imaging and a magnetic field of less than or equal to 0.35 tesla.
Device Story
System comprises water phantom, carriage, manual control unit, and software. Measures radiation beam characteristics for radiotherapy commissioning and QA. Operates by moving sensors through water basin via motor controller; electrometer captures radiation data. Software records, analyzes, and exports beam data libraries for TPS. Used in clinical settings by medical physicists. Output informs radiation therapy planning and quality assurance, ensuring accurate dose delivery and patient safety.
Clinical Evidence
Bench testing only. Performance verified via electrical/mechanical safety (IEC 61010), EMC (IEC 60601-1-2), dosimeter requirements (IEC 60731), and MRI safety (ASTM standards).
Technological Characteristics
Water phantom with PMMA walls; 380mm x 380mm x 242mm scanning range. Motorized 3-axis carriage (0.1mm step size). Electrometer with 1 fA resolution. Connectivity for data export to TPS. Complies with IEC 60731, IEC 61010, and IEC 60601-1-2:2015.
Indications for Use
Indicated for use by medical physicists for dosimetry measurements, beam data acquisition, commissioning, and periodic quality assurance of radiation therapy systems integrated with MRI units having static magnetic fields ≤ 0.35 Tesla.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
Related Devices
- K161807 — BEAMSCAN · Ptw-Freiburg Physikalisch Technische-Werksraetten · May 19, 2017
- K191646 — Beamscan MR · Ptw-Freiburg Physikalisch-Technische-Werkstaetten · Feb 21, 2020
- K011763 — OMNIDOS · Scanditronix Medical AB · Sep 5, 2001
- K103193 — DOSEVIEW 3D · Standard Imaging, Inc. · Dec 27, 2010
- K030175 — 2 BASIC · Capintec, Inc. · Apr 17, 2003
Submission Summary (Full Text)
{0}------------------------------------------------
June 3, 2020
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
LAP GmbH Laser Applikationen % Mr. Martin Pfabel Director Quality Management & Regulatory Affairs Zeppelinstrasse 23 21337 Lueneburg. Lower Saxony GERMANY
#### Re: K200907
Trade/Device Name: Thales 3D MR Scanner Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: Class II Product Code: IYE Dated: March 31, 2020 Received: April 6, 2020
Dear Mr. Pfabel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K200907
Device Name THALES 3D MR SCANNER
#### Indications for Use (Describe)
The THALES 3D MR SCANNER system is designed to measure the characteristic radiation unit. The beam data is stored in the software program and can be exported from there for use during commissioning in a treatment planning system (TPS) or for quality assurance. The THALES 3D MR SCANNER system can be used with irradiation units with integrated magnetic resonance imaging and a magnetic field of less than or equal to 0.35 tesla.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
In accordance with the requirements of the Safe Medical Device Act, LAP GmbH Laser Applikationen herewith submits a Summary of Safety and Effectiveness. This 510(k) summary for the THALES 3D MR Scanner meets the requirements of 21 CFR 807.92.
| Submitter information: | LAP GmbH Laser Applikationen<br>Zeppelinstrasse 23<br>21337 Lueneburg<br>Phone: +49 4131 951167<br>Fax: +49 4131 951196 |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Establishment Registration Number: 968113-<br>Owner/Operator Number: 9019434 |
| Official Correspondent: | Martin Pfabel<br>Zeppelinstrasse 23<br>21337 Lueneburg<br>Germany<br>Phone: +49 4131 951167<br>Email: m.pfabel@lap-laser.com |
| US Agent (Contact): | Trent Van Arkel<br>LAP of America Laser Applications, L.C.<br>161 Commerce Rd Ste 3<br>Boynton Beach, FL 33426<br>Phone: 561 4169250<br>Fax: 561 4169263<br>Email: Tava@Lap-Laser.com |
| Date Prepared: | 31 March 2020 |
{4}------------------------------------------------
# Device(s) Identification:
Device trade name: Common name:
THALES 3D MR scanner 3D water phantom
## Classification of the device:
Device Classification Name: Product Code: C.F.R. Section .: Classification Panel: Device Class:
Medical charged-particle radiation therapy system IYE 892.5050 Radiology devices Class II
# Device Description:
The THALES 3D MR SCANNER system consists of a water phantom with a carriage system, a manual control unit and a software program. A water basin with the corresponding pump, a control unit with an electrometer, and a motor controller are integrated in the carriage system. The software program is used to control the three axes of the water basin, perform measurements and analyse measured data.
The software program has the following properties: Acquisition of measured data, analysis, and recording of photon beams with the water phantom. Generation and export of beam data libraries for a treatment planning system (TPS) for commissioning of irradiation units. The software program is equipped with a central database and/or a file system for central storage of all recorded data for different irradiation units within a large hospital network with different sites. The software program offers the option of defining different user levels to support the typical clinical workflows.
#### Indications for use:
The THALES 3D MR SCANNER system is designed to measure the characteristic radiation beam for an irradiation unit. The beam data is stored in the software program and can be exported from there for use during commissioning in a treatment planning system (TPS) or for quality assurance. The THALES 3D MR SCANNER system can be used with irradiation units with integrated magnetic resonance imaging and a magnetic field of less than or equal to 0.35 tesla.
#### Predicate device:
| Device Name: | BEAMSCAN MR |
|-----------------------------|---------------------------------------------------|
| 510k number: | K191646 |
| Device Classification Name: | Medical charged-particle radiation therapy system |
| Product Code: | IYE |
| C.F.R. Section .: | 892.5050 |
| Classification Panel: | Radiology devices |
| Device Class: | Class II |
{5}------------------------------------------------
# Comparison to predicate device:
The following table presents the comparison of technological characteristics, functions and parameters of the identified predicate device and the proposed device.
| Parameter | Predicate device | Proposed device | Evaluation |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | BEAMSCAN MR | THALES 3D MR | |
| | | Scanner | |
| Intended Use | The PTW water<br>phantom system<br>BEAMSCAN MR is<br>intended for dosimetry<br>measurements in<br>radiotherapy systems.<br>The device is<br>intended to determine<br>the beam<br>characteristics of the<br>radiotherapy system<br>(beam data<br>acquisition) during the<br>commissioning and/or<br>for periodic quality<br>assurance procedures<br>according to the QA<br>plan of the<br>responsible medical<br>physicist. The system<br>can also be used at<br>combined MRI-<br>Radiation Therapy<br>systems with static<br>magnetic fields of up<br>to 1.5T. | The THALES 3D MR<br>SCANNER system is<br>designed to measure<br>the characteristic<br>radiation beam for an<br>irradiation unit. The<br>beam data is stored<br>in the software<br>program and can be<br>exported from there<br>for use during<br>commissioning in a<br>treatment planning<br>system (TPS) or for<br>quality assurance.<br>The THALES 3D MR<br>SCANNER system<br>can be used with<br>irradiation units with<br>integrated magnetic<br>resonance imaging<br>and a magnetic field<br>of less than or equal<br>to 0.35 tesla. | Similar<br>The intended use of<br>both systems covers<br>the quality assurance<br>for radiation therapy<br>systems as part of<br>regular quality checks,<br>commissioning or<br>other occasions.<br>Both systems are<br>further indicated for<br>use with integrated<br>magnetic resonance<br>imaging of magnetic<br>field.<br>The predicate device<br>is indicated to be used<br>in combined MRI-<br>Radiation therapy<br>systems with static<br>magnetic fields up to<br>1.5 Tesla.<br>The proposed device<br>can be used with<br>irradiation units with<br>integrated magnetic<br>resonance imaging<br>and a magnetic field<br>of less than or equal<br>to 0.35 tesla. |
| Scanning range | 408mm * 355mm *<br>248mm<br>(model for ViewRay) | 380mm * 380mm *<br>242mm | Similar<br>The scanning ranges<br>are nearly identical |
| Parameter | Predicate device<br>BEAMSCAN MR | Proposed device<br>THALES 3D MR<br>Scanner | Evaluation |
| Tank shape | Square | Square | Identical |
| Wall material | PMMA | PMMA | Identical |
| Chamber holder<br>orientation | Horizontal and vertical | Horizontal and<br>vertical | Identical |
| Motor scanning mode | Step by step | Step by step | Identical |
| Step size | 0.1mm | 0.1mm | Identical |
| Maximum speed | 30mm/s | 40mm/s | Similar<br>The THALES system<br>has a higher<br>maximum speed<br>which allows a faster<br>scanning process |
| Position accuracy | +/-0.1mm | +/- 0.25mm<br>(at 0.35Tesla) | Similar<br>The difference in<br>accuracy does not<br>affect the safety or the<br>effectiveness of the<br>proposed system.<br>The construction<br>requirements on<br>phantoms in<br>accordance with IEC<br>60731 have been met. |
| Reproducibility | +/-0.1mm | +/- 0.25mm<br>(at 0.35Tesla) | Similar<br>The difference in<br>reproducibility does<br>not affect the safety or<br>the effectiveness of<br>the proposed system.<br>The construction<br>requirements on<br>phantoms in<br>accordance with IEC<br>60731 have been met. |
| MRI compatible | Up to 1.5 Tesla | Up to 0.35 Tesla | Different<br>The LAP system is<br>indicated for use with<br>integrated magnetic<br>resonance imaging of<br>magnetic fields less or<br>equal than 0.35T |
| Parameter | Predicate device<br>BEAMSCAN MR | Proposed device<br>THALES 3D MR<br>Scanner | Evaluation |
| | | | while the predicate<br>device is indicated to<br>be used in a static<br>magnetic field up to<br>1.5 Tesla. The<br>difference in MR<br>compatibility does not<br>affect the safety or<br>effectiveness of the<br>system while the<br>target system of the<br>proposed device<br>(ViewRay MRidian)<br>only provides a static<br>magnetic field of 0.35<br>Tesla or less.<br>LAP has tested the<br>performance and<br>safety aspects with<br>the target system and<br>can confirm that the<br>proposed device can<br>be used in the<br>respective MRI<br>environment. |
| Electrometer channels | 2 entries | 2 entries | Identical |
| Electrometer<br>resolution | 10 fA | 1 fA | Different<br>The electrometer<br>resolution for the<br>proposed device is<br>better (10x better)<br>than the predicate<br>device's. |
| Electrical safety | IEC 61010 | IEC 61010 | Identical |
| Dosimeter testing | IEC 60731 | IEC 60731 | Identical |
| Electromagnetic<br>compatibility | IEC 60601-1-2:2007 | IEC 60601-1-2:2015 | Identical |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
# Summary of performance testing:
The proposed device has been tested in respect to electrical and mechanical safety in accordance with IEC 61010, EMC in accordance with IEC 60601-1-2 and dosimeter requirements in accordance with IEC 60731. MRI safety has been tested in accordance with applicable ASTM standards.
## Conclusion:
LAP GmbH Laser Applikationen believes that the THALES 3D MR Scanner is substantially equivalent to currently legally marketed devices. The device does not introduce new indications for use, has the same technological characteristics and does not introduce new potential hazards or safety risks.