510(k) Data Aggregation

    K Number
    K251305

    Validate with FDA (Live)

    Date Cleared
    2025-08-26

    (120 days)

    Product Code
    Regulation Number
    884.6180
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Ultra-Fast Vitri is indicated for use in the preparation, vitrification and storage of oocytes (MII).
    Ultra-Fast Warm is indicated for use in the preparation and warming of vitrified oocytes(MII).

    Device Description

    The Ultra-Fast Vitri and Ultra-Fast Warm is composed of a set of three media to vitrify and warm oocytes for assisted reproductive technology (ART) procedures.

    The Ultra-Fast Vitri includes two components, Equilibration Solution (ES) and Vitrification Solution (VS), containing the cryoprotectants ethylene glycol and dimethyl sulfoxide. There are two vitrification procedures to choose from. During the vitrification process, oocytes are first exposed to ES then in VS within several minutes. Using this methodology, permeating cryoprotectants can replace water in the oocytes prior to vitrification and storage in liquid nitrogen. The Ultra-Fast Vitri comes prepackaged with 1.5 mL vial or 4 mL vial of ES, three 1.5 mL vials or three 4 mL vials of VS.

    Ultra-Fast Warm is composed of one media used for warming and removing cryoprotectants from vitrified oocytes. It is composed of Thawing Solution (TS). The Ultra-Fast Warm comes pre-packaged with four 4.0 ml vials of TS.

    All the media in the Ultra-Fast Vitri and Ultra-Fast Warm contain Gentamicin. The media undergoes aseptic filtration, while the vials are sterilized by radiation.

    AI/ML Overview

    Based on the provided FDA 510(k) Clearance Letter, here's a description of the acceptance criteria and the study that proves the device meets them:

    Device: Ultra-Fast Vitri; Ultra-Fast Warm
    Description: A set of three media used for the vitrification (cryopreservation) and warming of oocytes (MII) for Assisted Reproductive Technology (ART) procedures.


    Acceptance Criteria and Reported Device Performance

    The core acceptance criteria for this device, as demonstrated through non-clinical and clinical performance data, revolve around its biological compatibility and effectiveness in preserving and recovering oocytes without compromising their viability or subsequent reproductive outcomes.

    Acceptance Criteria CategorySpecific Metric (Unit)Acceptance CriteriaReported Device Performance (Ultra-Fast Vitri/Warm)
    Non-Clinical Performance
    Color/AppearanceVisual inspectionAcceptable appearancePassed
    pH TestingpH value (range)7.20 – 7.60Passed (7.20 – 7.60)
    Endotoxin TestingEndotoxin level (EU/mL)<0.25 EU/mL (LAL methodology)Passed (<0.25 EU/mL LAL methodology)
    Osmolality TestingOsmolality value(Not explicitly stated in document, but implied as conforming to media standards)Passed
    Sterility TestingSterilityPasses USP <71>Passed (Passes USP <71>)
    Gentamicin TestGentamicin presence/level(Not explicitly stated, but implied as conforming to specification)Passed
    Initial Media Dispensing ValidationFunctional dispensing/packaging(Not explicitly stated, but implied as successful)Passed
    Mouse Embryo Assay (MEA)One-cell embryo development (96 hours)>80%Passed (>80%)
    BiocompatibilityBiocompatibility with cellsPassesPasses
    Storage StabilityTemperature range (°C)2 – 8°C2 – 8°C
    Shelf LifeDuration (months)12 months12 months
    Clinical Performance
    Oocyte Survival Rate%(Implied to be comparable to conventional protocol)100.0% (Ultra-Fast) vs. 90.9% (Conventional)
    Clinical Pregnancy Rate%(Implied to be comparable to conventional protocol)65.2% (Ultra-Fast) vs. 54.3% (Conventional)
    Live Birth Rate%(Implied to be comparable to conventional protocol)56.5% (Ultra-Fast) vs. 52.2% (Conventional)

    Study Proving Device Meets Acceptance Criteria

    The provided document describes both non-clinical (bench) and clinical performance studies to demonstrate the safety and effectiveness of the Ultra-Fast Vitri and Ultra-Fast Warm device and its substantial equivalence to the predicate device.

    1. Non-Clinical Performance Data (Bench Testing):

    • Description: A series of laboratory tests conducted directly on the media to confirm its physical, chemical, and biological properties.
    • Specific Tests: Color/Appearance, pH Testing, Endotoxin testing, Osmolality Testing, Sterility Testing, Gentamicin Test, Initial Media Dispensing Validation, Mouse Embryo Assay (MEA), and Biocompatibility.
    • Proof of Concept: The device passed all these tests, including achieving >80% one-cell development in the Mouse Embryo Assay, which is a critical biological performance indicator for reproductive media as per FDA guidance.

    2. Clinical Performance Data:

    • Study Design: A comparative study referenced from literature that evaluated the effectiveness of the ultra-fast vitrification and warming protocols using the subject device against conventional protocols (presumably using the predicate or similar conventional vitrification/warming solutions).
    • Study Objective: To demonstrate comparable outcomes (oocyte survival, clinical pregnancy, live birth rates) between the ultra-fast protocol and conventional protocols.

    Here's a breakdown of the specific requested information about the clinical study:

    • 2. Sample Size Used for the Test Set and Data Provenance:

      • Sample Size: 1,077 mature oocytes in total.
        • 519 oocytes for the conventional vitrification and warming protocols group.
        • 558 oocytes for the ultra-fast protocols (subject device) group.
      • Data Provenance: The document states "The referenced literature used Kitazato's vitrification and warming solutions (K171748 and K160864)".
        • It does not explicitly state the country of origin of the data.
        • It does not explicitly state if the study was retrospective or prospective. However, given it's a "study" comparing protocols, it's typically prospective, but this cannot be confirmed from the text.
    • 3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

      • This information is not provided in the document. The study focuses on clinical outcomes (survival, pregnancy, live birth rates) rather than human interpretation of images or other subjective assessments that would require expert consensus for ground truth.
    • 4. Adjudication Method for the Test Set:

      • This information is not applicable/not provided. Adjudication methods (like 2+1, 3+1) are typically relevant for studies involving human interpretation where reviewer disagreement needs to be resolved (e.g., radiology studies). This study measures biological/clinical outcomes.
    • 5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done:

      • No, an MRMC study was not done. MRMC studies are used to assess the impact of a device (often AI) on human reader performance, typically in diagnostic imaging. This study evaluated the direct clinical effectiveness of the media itself.
      • Therefore, an effect size of how much human readers improve with AI vs. without AI assistance is not applicable.
    • 6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study was Done:

      • Yes, in a sense, the clinical study assessed the "standalone" performance of the media with its associated protocols. It compared the outcomes from using the media (with its specific ultra-fast protocol) to conventional media/protocols. There isn't an "algorithm" in the traditional sense for this device; it's a chemical formulation and protocol. The "performance" is the biological outcome achieved by the oocytes.
    • 7. The Type of Ground Truth Used:

      • The ground truth was based on clinical outcomes data:
        • Oocyte survival rate after vitrification and thawing.
        • Clinical pregnancy rate (following embryo transfer resulting from these oocytes).
        • Live birth rate (following clinical pregnancy).
      • These are considered objective biological and clinical endpoints.
    • 8. The Sample Size for the Training Set:

      • This information is not applicable/not provided. This device is a media (consumable), not an AI algorithm that requires a separate "training set" of data. The "development" of the media and protocols would be based on laboratory research and refinement rather than a data training paradigm.
    • 9. How the Ground Truth for the Training Set Was Established:

      • This information is not applicable/not provided for the same reasons as #8.
    Ask a Question

    Ask a specific question about this device

    K Number
    K200249

    Validate with FDA (Live)

    Date Cleared
    2020-10-23

    (263 days)

    Product Code
    Regulation Number
    884.6180
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sequential Culture Media consists of Fertilization Medium, Cleavage Medium, and Blastocyst Medium that are intended to be used sequentially from fertilization to the blastocyst stage of development. The intended uses of the Fertilization Medium, Cleavage Medium, and Blastocyst Medium are as follows:

    Fertilization medium is intended for use during in vitro fertilization (IVF) and intracytoplasmic sperm insertion (ICSI) procedures and culture to the two pronuclei (zygote) stage of development.

    Cleavage Medium is intended for culture of embryos from the two pronuclei (zygote) stage to the 8cell stage of development. Cleavage Medium is not intended for transferring embryos to the uterine cavity.

    Blastocyst Medium is intended for culture from the 8-cell stage to the blastocyst stage of development. Blastocyst Medium is not intended for transferring embryos to the uterine cavity.

    Device Description

    Sequential Culture Media are intended for use sequentially from fertilization to late embryonic stages during assisted reproduction technology procedures for insemination and embryo culture.

    The Sequential Culture Media is provided in three variants: Fertilization Medium, Cleavage Medium, and Blastocyst Medium. Each variant is provided with or without protein (human serum albumin (HSA) or recombinant HSA (rHA)). All variants contain gentamicin, an antibiotic agent that suppresses bacterial growth. Each Sequential Culture Media solution is offered in three volumes (10mL, 50mL and 100mL).

    The Sequential Culture Media solution is a colorless, clear fluid, provided sterile-filtered into a container pre-sterilized by gamma irradiation. The primary container of Sequential Culture Media 10mL is a sterile non-pyrogenic PETG vial, and the primary container of the Sequential Culture Media 50mL and 100mL is a square, non-pyrogenic PETG bottle. The containers are manufactured and provided sterile (with a SAL of 10° ) by ThermoFisher Scientific, Inc. After sterilefilling, the top of the vial and bottle are sealed with tamper-evident shrink-wrap.

    AI/ML Overview

    The provided text is a 510(k) Summary for a medical device called "Sequential Culture Media." It details the device's characteristics, comparison to a predicate device, and non-clinical performance testing.

    Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance

    The device is a non-AI/ML product (culture media for IVF), so the acceptance criteria are not in terms of AI model performance metrics like accuracy, sensitivity, or specificity. Instead, they are related to the physical and biological properties of the culture media.

    Criteria (Internal Requirement / Standard)Acceptance CriteriaReported Device Performance
    AppearanceClear, particulate-freePassed (Clear, particulate-free)
    pH (per USP <791>)7.2 – 7.6Passed (7.2 – 7.6)
    Osmolarity (freezing depression method)270-295 mOsm/LPassed (270-295 mOsm/L)
    Endotoxin (per USP <85>)≤ 0.25 EU/mLPassed (≤ 0.25 EU/mL)
    Mouse Embryo Assay (MEA)≥ 80% of 1-cell mouse embryos developed to expanded blastocyst at 96 hoursPassed (≥ 80% of 1-cell mouse embryos developed to expanded blastocyst at 96 hours)
    Sterility (per USP <71>)No microbial growthPassed (No microbial growth)
    Stability TestingMaintain performance specifications (Appearance, pH, Osmolarity, Endotoxin, MEA, Sterility) for 4 monthsPassed (Real-time aged samples at baseline and 4 months)
    Container Seal (per USP <671>)≤ 5.0% permeability and ≤ 1 sample exceeding 2.50% over 14 daysPassed
    Sterile Filtration/Aseptic FillValidation per ISO 13408-1:2008/A1:2013 and ISO 13408-2:2018Passed
    Transportation TestingPackage integrity and device performance maintained (per ASTM D4169)Passed

    2. Sample Size for the Test Set and Data Provenance

    • Sample Size for Test Set: The document does not specify a distinct "test set" sample size in the way an AI/ML study would. Instead, performance testing applies to batches of the culture media. For the MEA, it mentions "1-cell mouse embryos," but the exact number of embryos or experimental replicates used is not provided.
    • Data Provenance: The studies were non-clinical performance testing conducted by Kitazato Corporation, the device manufacturer. The data provenance is internal to the manufacturer.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Given this is a non-clinical, laboratory-based product, the concept of "experts establishing ground truth" as it applies to image interpretation or diagnostic accuracy is not directly applicable. The "ground truth" here is based on analytical chemistry, biological assays, and sterility testing, which are measured objectively using established scientific methods and standards (e.g., USP, ISO). The qualifications of those performing these tests would be standard laboratory technicians/scientists, but no specific details on their number or qualifications are provided.

    4. Adjudication Method for the Test Set

    Not applicable. This is not a human-reader-based test where adjudication would be necessary. The results are based on objective laboratory measurements.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Not applicable. This is not a diagnostic device that involves human readers or AI assistance.

    6. Standalone Performance (Algorithm Only without Human-in-the-Loop Performance)

    Not applicable. This is not an algorithm or AI device. The device itself is the sequential culture media.

    7. Type of Ground Truth Used

    The "ground truth" in this context refers to the objectively measured characteristics of the culture media that are critical for its function and safety. This includes:

    • Analytical Chemistry/Physical Measurements: pH, Osmolarity, Endotoxin levels.
    • Biological Assay: Mouse Embryo Assay (MEA), which assesses the biological efficacy by observing embryo development.
    • Microbiology: Sterility testing.
    • Material Science/Engineering: Container seal integrity, package integrity.

    These are established scientific and regulatory standards.

    8. Sample Size for the Training Set

    Not applicable. As a non-AI/ML product, there is no "training set." The product's formulation and manufacturing processes are developed based on scientific understanding of embryo culture requirements and validated against performance specifications, not through machine learning training.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI/ML model for this device. The formulation and manufacturing parameters are likely established through R&D, chemical engineering, and biological testing, informed by existing scientific knowledge and regulatory requirements for reproductive media.

    Ask a Question

    Ask a specific question about this device

    K Number
    K193522

    Validate with FDA (Live)

    Date Cleared
    2020-08-25

    (250 days)

    Product Code
    Regulation Number
    884.6180
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    iMedium Single Step media are intended to be used as culture media for gametes and embryos from fertilization through day 5/6 of development. iMedium Single Step media are not intended for embryo transfer procedures.

    Device Description

    The Kitazato iMedium Single Step media are culture media used in assisted reproductive procedures. The iMedium Single Step media are designed for development of the embryo from fertilization through day 5/6 until the embryo transfer, in a continuous and uninterrupted culture system without the need to change media. The iMedium Single Step media are provided in three variants: without protein (Model: IM-S), with human serum albumin (HSA; Model: IM-SS), and with recombinant human albumin (rHA; Model: IM-SC). All variants contain gentamicin, an antibiotic agent that suppresses bacterial growth. Each iMedium Single Step solution is offered in three volumes (10mL, 50mL and 100mL). Each iMedium Single Step solution is a colorless and clear fluid. The product is single-use only, provided aseptically-filtered in a container sterilized by gamma irradiation.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and study for the iMedium Single Step device:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Performance Specification)Reported Device Performance
    pH7.2-7.6
    Endotoxin≤ 0.25 EU/mL
    1-Cell Mouse Embryo Assay (MEA)≥ 80 % embryos developed to expanded blastocyst at 96 hours
    SterilityNo growth
    AppearancePassed (Implicitly, as it passed all testing)
    OsmolarityPassed (Implicitly, as it passed all testing)
    Shelf-lifeMet at 4 months (real-time aged samples)
    Transportation TestingPassed
    Sterilization ValidationPassed

    2. Sample Size and Data Provenance

    The document does not specify the sample sizes used for the non-clinical tests (e.g., number of mouse embryos for MEA, number of batches for sterility, etc.).

    Data provenance: The testing was conducted by Kitazato Corporation, the manufacturer, as internal non-clinical tests. The country of origin for the data is Japan, based on the manufacturer's address (Kitazato Corporation, Japan). The data is retrospective as it was collected before the 510(k) submission.

    3. Number of Experts and their Qualifications

    This information is not provided in the document. The non-clinical tests described are laboratory-based and do not inherently involve human expert interpretation in the way a diagnostic imaging device might.

    4. Adjudication Method

    This information is not applicable as the non-clinical tests described do not involve human interpretation or subjective assessment that would require an adjudication method.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    A multi-reader multi-case (MRMC) comparative effectiveness study was not conducted. This device is a culture medium, not a diagnostic imaging or interpretive aid that would typically involve human readers.

    6. Standalone Performance Study

    Yes, a standalone performance study was done. The non-clinical performance data described (pH, endotoxin, MEA, sterility, etc.) represents the performance of the algorithm/device (the culture medium itself) without human intervention during the actual performance measurement. The purpose of these tests was to demonstrate the inherent safety and effectiveness characteristics of the iMedium Single Step media.

    7. Type of Ground Truth Used

    The ground truth for the non-clinical tests was established through:

    • Laboratory measurements/standards: For pH (USP <791>), Endotoxin (USP <85>), Sterility (USP <71>), and Osmolarity (freezing point depression method), the ground truth is based on established scientific and regulatory standards for these measurements.
    • Biological assay outcomes: For the 1-Cell Mouse Embryo Assay (MEA), the ground truth is the biological outcome of embryo development to the expanded blastocyst stage within 96 hours. This is a recognized biological endpoint for assessing the suitability of reproductive media.
    • Engineering/Quality standards: For shelf-life, transportation testing (ASTM D4169-16), and sterilization validation (ISO 13408-1:2008 / A1:2013, ISO 13408-2:2018), the ground truth refers to meeting the specified parameters and requirements outlined in these engineering and quality standards.

    8. Sample Size for the Training Set

    This information is not applicable/not provided. This device is a medical device (culture medium) and not an AI/ML algorithm that requires a training set in the conventional sense. The development of the medium would involve formulation scientists and biologists, rather than an AI training process.

    9. How the Ground Truth for the Training Set was Established

    This information is not applicable/not provided for the same reasons mentioned in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192845

    Validate with FDA (Live)

    Date Cleared
    2020-08-17

    (319 days)

    Product Code
    Regulation Number
    884.6180
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sperm Freeze is intended for the cryopreservation of human semen samples prior to use in assisted reproductive technology procedures.

    Sperm Fridge is intended to protect human semen samples during refrigerated storage prior to use in assisted reproductive technology procedures.

    Device Description

    Sperm Freeze and Sperm Fridge are preservation media intended for preserving sperm (short-term and long-term) for in vitro fertilization (IVF) and assisted reproduction technologies (ART). There are two methods for sperm preservation; freezing and refrigeration. Freezing is necessary for long-term sperm preservation, but may have disadvantages such as reduced survival rate and/or sperm motility after thawing. Refrigeration makes it possible to preserve sperm for several days without freezing and may be preferable for short-term preservation. Both products are intended to be used exclusively in a laboratory environment by trained professionals.

    Sperm Freeze is a is a Tris-buffered solution containing the cryoprotectants glycerol and trehalose that is used for long-term sperm cryopreservation procedures. Sperm Fridge is a buffered solution (TES and Tris) that includes a nutrient source for sperm (dextrose) that is used for short-term sperm preservation procedures. Sperm Freeze and Sperm Fridge are offered in two volumes (0.5 mL and 10 mL, and 0.5 mL is sold in a pack of 5).

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and study information for the "Sperm Freeze, Sperm Fridge" device, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test MethodAcceptance Criteria (Sperm Fridge)Reported Performance (Sperm Fridge)Acceptance Criteria (Sperm Freeze)Reported Performance (Sperm Freeze)
    AppearanceNot specified in tablePassedNot specified in tablePassed
    Endotoxin (per USP <85>)≤ 0.25 EU/mLMet Criteria≤ 0.25 EU/mLMet Criteria
    OsmolalityNot specified in tablePassedNot specified in tablePassed
    pH (per USP <791>)7.2–7.6Met Criteria7.2–7.6Met Criteria
    Sperm Refrigeration Survival≥ 80% (45 min)Met CriteriaN/A (for Sperm Freeze)N/A
    Sperm Cryo-survivalN/A (for Sperm Fridge)N/A≥ 80% (45 min)Met Criteria
    Sterility (per USP <71>)No growthMet CriteriaNo growthMet Criteria
    Shelf-life performance testingMet product specificationsMet CriteriaMet product specificationsMet Criteria

    Note: The "Reported Performance" column indicates that the device "passed" or "met criteria" based on the general statement in section 9: "Results confirm that the design inputs and performance specifications for the devices are met." Specific numerical results for each test (e.g., the exact percentage for sperm survival) are not provided in this summary.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify the sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). It only mentions that "Kitazato completed the following non-clinical tests" and refers to "internal requirements, national standards, and international standards."

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The studies described are non-clinical (laboratory testing of the media itself), not clinical studies involving human samples and expert evaluation of outcomes in the way one might see for an AI diagnostic device.

    4. Adjudication Method for the Test Set:

    This information is not applicable as the tests described are non-clinical and do not involve human interpretation or adjudication of results.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This is not applicable. The device (Sperm Freeze, Sperm Fridge) is a reproductive media product, not an AI or imaging device that would involve human readers or AI assistance in interpretation. No MRMC study was conducted.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This is not applicable. The device is a chemical medium, not an algorithm or software. No standalone performance testing in this context was performed.

    7. The Type of Ground Truth Used:

    The ground truth for the non-clinical tests appears to be laboratory measurements against established specifications and standards. For example, the pH measurement would be compared to the specified pH range (7.2-7.6), endotoxin levels compared to the maximum allowed (≤ 0.25 EU/mL), and sperm survival rates compared to the minimum required (≥ 80%). The "ground truth" is essentially the predefined acceptable range or threshold for each physical/chemical/biological property of the media.

    8. The Sample Size for the Training Set:

    This is not applicable. The device is a manufactured product, not a machine learning model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established:

    This is not applicable for the same reasons as #8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K192540

    Validate with FDA (Live)

    Date Cleared
    2020-05-01

    (228 days)

    Product Code
    Regulation Number
    884.6110
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Kitazato ET Catheters are intended for ultrasound guided introduction of embryos into the uterine cavity following in vitro fertilization.

    Device Description

    Kitazato ET Catheters are sterile, single-use catheters for use in embryo transfer procedures. The catheter is used to hold and deliver embryos to the uterine cavity during a procedure. Catheters are available in 3Fr and 4Fr sizes, both of which are provided in lengths of 220 mm and 250 mm, as well as one version that is 3Fr and 400 mm length that has the same uterine insertion depth as the other device versions, but allows the user to be positioned further from the patient during a procedure. Devices also include versions with and without an echo reflective chip in the tip of the catheter. The echo reflective chip aids in visualization of the catheter during a transfer procedure; all transfer procedures are to be done under ultrasound guidance. Both the Type 2 devices also include a stainless steel or polyimide core to provide additional rigidity to the catheter. The guide is used to navigate the pathway through the cervical canal to allow for easier catheter insertion Guides are provided with a 30° pre-curved distal shaft and are available in lengths of 170 mm and 200 mm. Guides also include a styrene elastomer stopper to aid in positioning the targeted uterine insertion depth. The guides are also available with a straight or bulb tip design, and with and without an echo reflective chip in the tip of the guide. An obturator with a length of 170 mm or 200 mm is provided with the Type 2 versions of the device to prevent ingress of fluid into the guide lumen during guide insertion. The Stylet is an optional accessory used when additional rigidity is needed during insertion of the guide through the cervix, and is provided in two lengths, 170 mm and 200 mm. The stainless steel core of the Stylet can be shaped to aid in guide delivery through the cervix.

    AI/ML Overview

    This document describes a 510(k) premarket notification for the Kitazato ET Catheters. The focus of the provided text is on demonstrating the substantial equivalence of the Kitazato ET Catheters to a legally marketed predicate device (Guardia™ Access Embryo Transfer Catheter Sets). As such, the information provided is geared towards comparing the new device to an existing one, rather than establishing the performance of a novel AI/software medical device against defined acceptance criteria in the way a traditional clinical study would for an AI-powered diagnostic.

    Therefore, the requested information about "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of an AI medical device (e.g., sample size for test set, data provenance, number of experts, adjudication method, MRMC study, standalone performance, type of ground truth, training set sample size and ground truth establishment) is largely not applicable to this submission.

    This submission is for a physical medical device (catheters) and relies on bench testing and biocompatibility studies to show substantial equivalence, not a clinical study involving human readers and AI performance.

    However, I can extract information related to relevant non-clinical performance and "acceptance criteria" where applicable to this physical device, as described in the document.


    Acceptance Criteria and Device Performance for Kitazato ET Catheters (Physical Device)

    This submission focuses on demonstrating substantial equivalence of a physical medical device (embryo transfer catheters) to a predicate device, primarily through non-clinical performance data. Therefore, the "acceptance criteria" are related to established norms for medical device safety and performance rather than AI-specific metrics like AUC, sensitivity, or specificity in a diagnostic context.

    Here's the relevant information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria CategorySpecific Test/CriterionReported Device Performance / Evaluation Method
    SterilizationValidated per ISO 11137-1:2006(R)2010 &A1:2013 and ISO 11137-2:2013Studies performed to support substantial equivalence.
    BiocompatibilityEvaluated per ISO 10993-1: 2018 and 2016 FDA GuidanceStudies performed to support substantial equivalence.
    - CytotoxicityMet ISO 10993-5:2009
    - SensitizationMet ISO 10993-10:2010
    - Intracutaneous reactivityMet ISO 10993-10:2010
    Endotoxin< 20 EU/device (per AAMI/ANSI ST72:2011; USP <85>)Met: < 20 EU/device
    Packaging & Transport- Transportation Simulation per ASTM D4169-16Assessment of packaging and contents after conditioning; bubble leak test per ASTM F2096-11 after conditioning.
    - Package Integrity (Visual Inspection)Per ASTM F1866/F1886M-09
    - Package Integrity (Dye Penetration)Per ASTM F1929-15
    - Package Integrity (Seal Strength)Per ASTM F88/F88M-15
    Embryo CompatibilityMouse Embryo Assay (MEA): 1-cell, ≥ 80% blastocyst at 96h (before and after aging)Met: 1-Cell MEA: ≥80% blastocysts at 96h
    Bench PerformanceAll predetermined acceptance criteria were metStudies performed before and after aging.
    - Dimensional Verification (against device input requirements)Met: Devices measured and verified against device input requirements.
    - Appearance (no burrs, scratches, or foreign objects)Met: Devices visually inspected.
    - Tensile Strength (greater than pre-determined acceptance criterion)Met: Testing demonstrated value greater than pre-determined acceptance criterion.
    - Luer Taper Inspection (assess connection)Met: Visual inspection to assess connection.
    - Colorfastness (legible depth marking after friction)Met: Testing demonstrated legible depth marking after friction testing (100 rubs at 200g load weight).
    - Echo Test (maintenance of echo chip band)Met: Testing demonstrated maintenance of echo chip band in accordance with pre-determined acceptance criterion.

    2. Sample Size Used for the Test Set and Data Provenance:

    • Not Applicable in the traditional sense of an AI/ML test set. The "tests" here are non-clinical bench and lab experiments on the physical device components and assembled units (e.g., sterilization validation, biocompatibility, mechanical testing, Mouse Embryo Assay).
    • Data Provenance: The studies are described as "performed to support substantial equivalence." The source of the materials and specific lab locations are not specified beyond the manufacturer (Kitazato Corporation) and their correspondent (Emergo Global Consulting, LLC) being based in Japan and the USA respectively. These are prospective tests of the manufactured device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    • Not Applicable. Ground truth, in the context of expert consensus for image or data interpretation, is not relevant here as this is a physical device submission. The "ground truth" for these tests comes from established international standards (ISO, ASTM, AAMI/ANSI, USP) and verified laboratory procedures.

    4. Adjudication Method for the Test Set:

    • Not Applicable. Adjudication methods are typically for resolving discrepancies in expert interpretations (e.g., in MRMC studies for AI). Here, the tests follow defined protocols with measurable outcomes.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done:

    • No. An MRMC study is not relevant for this type of physical medical device submission.

    6. If a Standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • No. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used:

    • For this physical device, the "ground truth" is derived from:
      • Established Industry Standards: Compliance with ISO, ASTM, AAMI/ANSI, and USP standards.
      • Laboratory Measurements and Biological Assays: Objective measurements (e.g., dimensions, tensile strength) and biological assays (e.g., Mouse Embryo Assay, cytotoxicity testing) with predefined acceptance limits.

    8. The Sample Size for the Training Set:

    • Not Applicable. This is not an AI/ML device where a "training set" would be used to develop an algorithm.

    9. How the Ground Truth for the Training Set was Established:

    • Not Applicable. See point 8.

    Summary for this specific FDA Submission (K192540):

    The FDA 510(k) clearance for the Kitazato ET Catheters is based on demonstrating substantial equivalence to an existing legally marketed predicate device. This is achieved through a comprehensive set of non-clinical performance tests, including sterilization validation, biocompatibility studies, endotoxin testing, packaging and transportation simulation, and specific bench performance tests (dimensional verification, appearance, tensile strength, Luer taper inspection, colorfastness, echo test). Additionally, a Mouse Embryo Assay (MEA) is performed to confirm the device's compatibility with embryos. The "acceptance criteria" are the successful completion of these tests in accordance with relevant international and national standards, with all predetermined criteria reported as met. The study design is not a clinical AI performance study, but a series of laboratory and bench tests on the physical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190199

    Validate with FDA (Live)

    Date Cleared
    2019-09-26

    (234 days)

    Product Code
    Regulation Number
    884.6180
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    SepaSperm Washing Solution is used for preparation and washing of sperm for use in assisted reprocedures. SepaSperm Washing Solution is not intended for use in intrauterine insemination procedures.

    SepaSperm Solution is used for separation of motile sperm from seminal fluid for use in assisted reproduction procedures.

    Device Description

    SepaSperm is used for preparation and washing of sperm and for separation of motile sperm from semen in assisted reproductive techniques.

    The device includes two types of solutions used for density-gradient centrifugation:

    • . SepaSperm Solution for sperm preparation and separation from semen, and
    • . SepaSperm Washing Solution for sperm washing and preparation

    Both solutions are provided in two sizes (100mL and 50mL bottles). This product is single-use only and includes no accessories. The provided sterile-filtered into a container that is sterilized by gamma irradiation.

    SepaSperm Solution is offered in four different densities, determined by the amount of silica present:

    • SepaSperm Solution 100% -
    • -Lower laver solution 80%
    • Middle layer solution 60% -
    • -Upper layer solution 40%

    SepaSperm Solution is composed of silica particles diluted with modified human tubal fluid (m-HTF) medium, with dextran and polyvinylpyrrolidone added. The m-HTF medium is a HEPES-buffered solution which is suitable for use in ambient conditions. All SepaSperm Solutions are provided with and without gentamicin. SepaSperm Solution has a shelf-life of six or 12 months for media without or with gentamicin, respectively.

    The SepaSperm Washing Solution consists of m-HTF medium supplemented with dextran and polyvinylpyrrolidone. It contains no silica. SepaSperm Washing Solution is used to wash the prepared sperm and suspend the final pellet. SepaSperm Washing Solutions are provided with and without gentamicin. SepaSperm Washing Solution has a shelf-life of six or 12 months for media without or with gentamicin, respectively.

    AI/ML Overview

    The provided text describes the acceptance criteria and a "Summary of Non-Clinical Performance Testing" for the SepaSperm Washing Solution and SepaSperm Solution, which are reproductive media. This document is a 510(k) summary for a medical device submitted to the FDA, indicating a comparison to a predicate device.

    Here's the breakdown of the information requested:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance Criteria (Specification)Reported Device Performance (Summary of Non-Clinical Performance Testing)
    AppearanceMet (implicitly, as it's part of "Shelf Life Performance Testing" and results confirm specifications are met)
    Specific gravity: 1.0600-1.1200Met (implicitly, as results confirm specifications are met)
    pH, per USP <791>: 7.2 - 7.6Met (implicitly, as results confirm specifications are met)
    Osmolality, per USP <785>: 270-300 mOsm/KgMet (implicitly, as results confirm specifications are met)
    Endotoxin, per USP <85>: < 0.25 EU/mLMet (implicitly, as results confirm specifications are met)
    HSSA (Sperm Survival Assay): ≥80% of control motility at 24hMet (implicitly, as results confirm specifications are met)
    Sterility, per USP <71>: No microbial growthMet (implicitly, as results confirm specifications are met)
    Sterile filtration and Aseptic fill validation, per ISO 13408-1:2008(R2013) and ISO 13408-2:2003(R)2013)Met (implicitly, as results confirm specifications are met)
    Transportation testing per ASTM D4169-16Met (implicitly, as results confirm specifications are met)
    Bottle seal strength testing and containers performance testing per <USP 671>Met (implicitly, as results confirm specifications are met)
    Effectiveness of device in three separation modes (sperm assessment for motility, morphology, viability, purity, and integrity before and after separation)Met (implicitly, as results confirm specifications are met)

    2. Sample size used for the test set and the data provenance:

    The document does not specify the sample size for individual tests (e.g., how many batches for shelf-life, how many sperm samples for HSSA or separation effectiveness) or the provenance (country of origin, retrospective/prospective nature) of the non-clinical performance test data. It only states that "The following studies have been performed to support substantial equivalence to the predicate devices."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. The studies listed are laboratory performance tests for the media itself, not clinical studies involving expert interpretation of medical images or patient data.

    4. Adjudication method for the test set:

    This information is not applicable and therefore not provided, as the studies are described as non-clinical performance testing of a medical device (media), not human-read clinical data requiring adjudication.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable and therefore not provided. This is a 510(k) submission for reproductive media, not an AI-powered diagnostic device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable and therefore not provided. This is a medical device (culture media), not an algorithm.

    7. The type of ground truth used:

    The "ground truth" for the performance testing of these media solutions is based on established laboratory standards and specifications (e.g., USP <791> for pH, USP <785> for osmolality, USP <85> for endotoxin, USP <71> for sterility, ISO 13408 standards for sterile filtration, ASTM D4169 for transportation, USP <671> for container performance). For sperm assessment, the "ground truth" would be the initial characteristics of the sperm samples and their characteristics after processing, compared against expected physiological ranges and performance standards.

    8. The sample size for the training set:

    This information is not applicable and therefore not provided. This is not a machine learning or AI-based device, so there is no concept of a "training set."

    9. How the ground truth for the training set was established:

    This information is not applicable and therefore not provided for the reason stated in point 8.

    Ask a Question

    Ask a specific question about this device

    K Number
    K181469

    Validate with FDA (Live)

    Date Cleared
    2018-11-09

    (158 days)

    Product Code
    Regulation Number
    884.6160
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cryotop®US-flash and Cryotop®US-scoop are cryopreservation storage devices that are intended for use in vitrification procedures to contain and maintain human oocytes (MII) and embryos.

    Device Description

    This 510(k) covers two subject devices, the Cryotop®US-flash and Cryotop®US-scoop that are modified versions of the predicate device (Cryotop®US, K153027). Both devices are cryopreservation storage devices intended for embryo/oocyte vitrification. Cryotop®US-flash and Cryotop®US-scoop are provided sterile and are for single-use only.

    The Cryotop®US-flash is composed of a polypropylene straw and an acrylonitrile butadiene styrene handle with a polyethylene terephthalate fine tip where oocytes or embryos are loaded. The fine device is flat with a triangular-shaped area for oocyte/embryo loading. The Cryotop®US-scoop is composed of a polypropylene straw and a polystyrene handle and fine tip where oocytes or embryos are loaded. The fine tip of this device is U-shaped. After oocyte or embryo loading, the handle of the subject devices is inserted into the straw enclosure. A hermetic seal is created via a tapered handle with a stop location integrated into the handle. As the handle is placed into the straw enclosure it creates a closed system keeping the fine tip isolated from the liquid nitrogen. The straw component of both devices also includes a weight at its distal end to aid in maintaining correct device position during storage in liquid nitrogen.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for two new cryopreservation storage devices: Cryotop®US-flash and Cryotop®US-scoop. The submission aims to demonstrate substantial equivalence to a predicate device, Cryotop®US (K153027). The document focuses on non-clinical performance data to support this claim, rather than a clinical study comparing the devices.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based only on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    Test CategorySpecific Test / CriterionAcceptance CriteriaReported Device Performance (Cryotop®US-flash & Cryotop®US-scoop)
    PerformanceCooling Rate3,000°C/minPassed: 3,000°C/min
    Warming Rate44,000°C/minPassed: 44,000°C/min
    Physical/MechanicalDimensional TestingOuter diameter and length per specifications (specific values not provided in this extract)Passed (implied)
    Durability TestingNo burst or liquid nitrogen ingress into straw after 30-second immersionPassed
    Mechanical Tensile TestingTensile strength to withstand 5NPassed
    BiocompatibilityBacterial Endotoxins<0.5 EU/device per USP <85> and ANSI/AAMI ST72:2002/(R)2010Passed: <0.5 EU/device
    Mouse Embryo Assay (MEA)≥80% of 1-cell embryos developed to blastocysts at 96 hoursPassed: ≥80%
    EnvironmentalTransportation Simulation TestingPer ASTM D4169-09Passed (implied)
    Shelf-LifePackage Integrity Testing: Dye penetrationPer ASTM F1929-15Passed (implied)
    Package Integrity Testing: Seal strengthPer ASTM F88/F88M-15Passed (implied)
    Shelf-Life (Dimensional, Mechanical, Durability, Endotoxin, MEA)In accordance with methods & acceptance criteria mentioned above for each testPassed (implied)

    2. Sample size used for the test set and the data provenance:

    • Sample Size: Not explicitly stated for each individual test. The document mentions "non-clinical performance testing" was conducted, implying multiple tests on multiple units, but no specific numbers of devices tested are provided.
    • Data Provenance: The tests were conducted by the manufacturer, Kitazato Corporation, as "internal requirements." The text does not specify the country of origin of the data beyond "Kitazato Corporation 81 Nakajima, Fuji-city Shizuoka 416-0907 JAPAN." These are retrospective tests performed for the 510(k) submission.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided as the tests are non-clinical, laboratory-based performance tests, not clinical studies requiring expert interpretation of outcomes.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This information is not applicable as the tests are non-clinical and do not involve human interpretation or adjudication in the context of clinical outcomes. The results are objective measurements against defined criteria.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done. This is a 510(k) submission for a medical device (cryopreservation storage device), not an AI-powered diagnostic tool, and therefore, an MRMC study is not applicable.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This is not applicable as the submission is for a medical device (cryopreservation storage device), not an algorithm or AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the non-clinical performance tests, the "ground truth" is established by objective physical, chemical, and biological measurements against pre-defined, quantitative standards and specifications (e.g., specific temperatures, mechanical forces, endotoxin levels, embryo development rates).

    8. The sample size for the training set:

    • This concept is not applicable to this submission, as it's for physical medical devices and the non-clinical tests described do not involve machine learning or a training set.

    9. How the ground truth for the training set was established:

    • This question is not applicable as there is no training set for the described non-clinical performance tests for this medical device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K182002

    Validate with FDA (Live)

    Device Name
    Cumulus Remover
    Date Cleared
    2018-10-23

    (89 days)

    Product Code
    Regulation Number
    884.6180
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Cumulus Remover is for the removal of the cumulus complex and corona radiata surrounding the oocyte in preparation for ICSI.

    Device Description

    Cumulus Remover is an enzyme solution containing recombinant human hyaluronidase that digests hyaluronic acid that binds the cumulus and corona cells surrounding oocytes together. This function of hyaluronidase can be used for denuding cumulus and coronal cells from oocytes prior to performing Intracytoplasmic Sperm Injection (ICSI) fertilization procedures. Cumulus Remover is provided in a polypropylene vial (package size 0.5 mL), and five vials are packaged together in a box. This product is aseptically processed and has a shelf-life of six months when stored at 2-8°C. Cumulus Remover is tested for pH, osmolality, endotoxin, sterility, embryotoxicity, and hyaluronidase activity before lot release.

    AI/ML Overview

    The provided text describes the "Cumulus Remover," a hyaluronidase solution used for removing cumulus complex and corona radiata from oocytes in preparation for ICSI. Here's a breakdown of the acceptance criteria and the study details:

    1. Table of Acceptance Criteria and Reported Device Performance

    TestAcceptance CriteriaReported Device Performance
    pH7.2-7.67.2-7.6 (specified as falling within this range)
    Osmolality270-295 mOsm270-295 mOsm (specified as falling within this range)
    SterilityNo microbial growth (per USP <71>)No microbial growth
    Endotoxin<0.25 EU/ml (per USP <85>)<0.25 EU/ml
    Mouse Embryo Assay (MEA)≥80% of embryos expand to blastocyst stage by 96hMet (implied by statement "in comparison with the control group" and overall conclusion of substantial equivalence)
    Hyaluronidase Activity70-90 units/ml (per Japanese Pharmacopoeia, JP17)70-90 units/ml (average 80 units/ml)
    Shelf-lifeAll above specifications met at time zero and 6 monthsMet (implied by statement that testing was conducted)

    2. Sample Size Used for the Test Set and Data Provenance

    • pH, Osmolality, Sterility, Endotoxin, Hyaluronidase activity: The sample sizes for these tests are not explicitly stated in terms of number of batches or individual samples. However, the tests are conducted as part of lot release, suggesting a representative sample from each manufactured lot.
    • Mouse Embryo Assay (MEA): The sample size for embryos used in the MEA is not specified.
    • Data Provenance: The document does not explicitly state the country of origin for the data or whether it was retrospective or prospective. Given that the manufacturer is Kitazato Corporation in Japan, it's likely the studies were conducted there. The studies appear to be prospective, laboratory-based tests conducted on the manufactured product.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This type of information is generally not applicable to the non-clinical performance tests described for this device (pH, sterility, MEA, etc.). These are objective, laboratory-measured parameters, not subject to expert interpretation for ground truth establishment.

    4. Adjudication Method for the Test Set

    Not applicable. The tests are objective laboratory measurements, not assessments requiring adjudication.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done

    No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not performed. This type of study is typically relevant for interpretative diagnostic devices where human reader performance is a key metric.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, the studies described are standalone performance evaluations of the device itself (the Cumulus Remover solution) in controlled laboratory settings, independent of human interaction within a clinical workflow.

    7. The Type of Ground Truth Used

    The "ground truth" for each specific test is defined by established scientific and regulatory standards:

    • pH: Measured directly against a standard scale.
    • Osmolality: Measured directly against a standard scale.
    • Sterility: Absence of microbial growth, determined by standard microbiological culture methods.
    • Endotoxin: Concentration of endotoxins, determined by standard LIMULUS Amebocyte Lysate (LAL) assay.
    • Mouse Embryo Assay (MEA): The ability of mouse embryos to develop to the expanded blastocyst stage, which is a biological outcome observed and assessed without subjective expert consensus regarding "ground truth" per se, but rather against a predefined biological benchmark.
    • Hyaluronidase Activity: Measured directly against a standard activity unit, per Japanese Pharmacopoeia, JP17.

    8. The Sample Size for the Training Set

    Not applicable. This device is a chemical solution assessed through laboratory performance tests, not an AI/ML algorithm that requires a training set of data.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K171748

    Validate with FDA (Live)

    Date Cleared
    2017-12-14

    (184 days)

    Product Code
    Regulation Number
    884.6180
    Age Range
    All
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Vitrification Kit is indicated for use in the preparation, vitrification and storage of oocytes (MI), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos.

    The Thawing Kit is indicated for use in the preparation and thawing of vitrified oocytes (MII), pronuclear (PN) zygotes through day 3 cleavage stage embryos, and blastocyst stage embryos.

    Device Description

    The Vitrification and Thawing Kits are composed of a set of six media to vitrify and warm MII oocytes, and pronuclear (PN) zygotes through blastocyst stage embryos for Assisted Reproductive Technology (ART) procedures.

    The Vitrification Kit includes three media components, Basic Solution (BS), Equilibration Solution (ES) and Vitrification Solution (VS), containing the cryoprotectants ethylene glycol, trehalose, and dimethyl sulfoxide. During the vitrification process, embryos are first exposed to ES and then to VS. In the case of the oocytes, use BS and ES. Using this methodology, the permeating cryoprotectants can replace water in the occyte, PN through blastocyst stage embryos prior to vitrification and storage in liquid nitrogen. The Vitrification Kit comes prepackaged with one 1.5 ml vial of BS and ES, two 1.5 ml vials of VS, 4 Cryotop devices (Cryotop CL, Cryotop SC, or Cryotop US), and 2 Repro Plates.

    The Thawing Kit is composed of three media used stepwise for thawing cryoprotectants from vitrified oocytes, and PN through blastocyst stage embryos. The Thawing Kit is composed of TS (Thawing Solution), DS (Dilution Solution) and WS (Wash Solution). The Thawing Kit comes pre-packaged with two 4.0 ml vials of thawing solution, one 4.0 ml vial of dilution solution, one 4.0 ml vial of washing solution, one Repro Plate, and two 35 mm dishes.

    All the media in the Vitrification Kit contain Gentamicin. The media in these kits undergo aseptic filtration, while storage devices and plates are sterilized by radiation.

    AI/ML Overview

    The provided document describes the Vitrification Kit and Thawing Kit (K171748) and its substantial equivalence to a predicate device. Below is an attempt to extract the requested information, though it's important to note that this document is a 510(k) Summary, which focuses on demonstrating substantial equivalence rather than a full study report with detailed acceptance criteria and performance data in the format often associated with AI/software performance studies. The device is a "Reproductive Media and Supplements," which are chemical reagents, not an AI/software device, so many of the requested fields (like AI-specific performance metrics, reader studies, etc.) are not directly applicable.

    Here's the information based on the provided text:

    Acceptance Criteria and Device Performance

    Since this is a submission for a "Reproductive Media and Supplements" kit, the acceptance criteria are related to the biological outcome (survival, development, etc.) of oocytes and embryos rather than typical device performance metrics like accuracy, sensitivity, or specificity of an AI algorithm. The performance is compared to similar existing products (predicate device or other vitrification media).

    Acceptance Criteria (Bench/Literature Study)Reported Device Performance (as demonstrated by literature or similar device)
    Oocyte Survival Rate (compared to surrogate device/vitrification media with serum substitute)Comparable oocyte survival rate between a surrogate device (with similar formulation and cryoprotectants to the subject device) and vitrification media containing serum substitute supplement. Also, comparable oocyte survival rate to other methods of vitrification.
    Implantation Rate (following vitrification using a surrogate device)Comparable implantation rate between a surrogate device (with similar formulation and cryoprotectants to the subject device) and vitrification media containing serum substitute supplement.
    Clinical Pregnancy Rate (following vitrification using a surrogate device)Comparable clinical pregnancy rate between a surrogate device (with similar formulation and cryoprotectants to the subject device) and vitrification media containing serum substitute supplement.
    Live Birth Rate (following vitrification using a surrogate device)Comparable live birth rate between a surrogate device (with similar formulation and cryoprotectants to the subject device) and vitrification media containing serum substitute supplement. Birth rates following use of vitrified oocytes were shown to be comparable to the methods used in the predicate device.
    Human Blastocyst Survival Rate (compared to surrogate device/vitrification media with serum substitute)Comparable human blastocyst survival rate following vitrification between a surrogate device (with similar formulation to the predicate device) and vitrification media containing serum substitute supplement.
    Fertilization Rate of Oocytes (following vitrification using methods similar to the subject device)Fertilization rate comparable to fresh oocytes.
    Quality Blastocyst Rate (following vitrification using methods similar to the subject device)Quality blastocyst rate comparable to fresh oocytes.
    Survival rates of oocytes and embryos (general consistency with normal ART procedures)Consistent with normal ART procedures using similar IVF treatments and cryopreservation techniques.
    Endotoxin (LAL methodology for Media)< 0.25 EU/mL
    Mouse Embryo Assay (development to blastocyst at 96 hours)> 80% development to blastocyst at 96 hours
    Sterility TestingPasses
    pH Test7.20 - 7.60
    BiocompatibilityPasses
    Sterilization Validation, Packaging Validation, Performance (bench) testing (for identical device cleared under K160864, leveraged in this submission)Passed all testing.

    Study Details

    1. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

      • Sample Size: Not explicitly stated as a single "test set" number. The evidence comes from three published scientific papers.
        • Literature 1 (Coello et al, 2016): Retrospective cohort study.
        • Literature 2 (Mori et al, 2015): Not specified in the summary, but likely a study comparing different methods.
        • Literature 3 (Inoue et al, 2014): Not specified.
      • Data Provenance: Not explicitly stated for all, but Coello et al. is published in Journal of Assisted Reproduction Genetics, typically international. Mori et al. published in Reproductive BioMedicine Online. Inoue et al. in Low Temp Med. Specific countries of origin for the patient data are not detailed in this summary. The studies appear to be retrospective and prospective clinical/bench studies, but specific details for each are not fully provided in this summary.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

      • Not applicable in the context of this device (Reproductive Media Kit). The "ground truth" for the performance claims would be the observed biological outcomes (survival, fertilization, development to blastocyst, pregnancy, live birth rates) reported in the referenced scientific literature, likely assessed by trained embryologists and clinicians.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set

      • Not applicable. This is not an AI/image analysis device requiring expert adjudication of outputs.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

      • Not applicable. This is not an AI device. The comparison is between different media formulations and vitrification methods.
    5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

      • Not applicable. This is not an AI/algorithm device.
    6. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

      • The "ground truth" here is outcomes data and biological observations from clinical and laboratory studies reported in published literature, such as oocyte/embryo survival rates, fertilization rates, blastocyst development, implantation rates, clinical pregnancy rates, and live birth rates.
    7. The sample size for the training set

      • Not applicable. This is a medical device (chemical media), not an algorithm or AI model that requires a training set.
    8. How the ground truth for the training set was established

      • Not applicable. (See #7).
    Ask a Question

    Ask a specific question about this device

    K Number
    K162878

    Validate with FDA (Live)

    Date Cleared
    2017-07-13

    (272 days)

    Product Code
    Regulation Number
    884.6110
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Kitazato ET Catheters are intended for ultrasound guided introduction of embryos into the uterine in vitro fertilisation.

    Device Description

    Kitazato ET Catheters are sterile, single-use catheters for embryo transfer. The ET Catheters consist of a transfer catheter, guide catheter, and a stylet sheath. The Transfer catheter (hereafter referred to as 'catheter' or 'ET Catheter') is composed of a catheter shaft, a connector and a shaft protective sleeve (protector). The catheter types include a normal catheter (Type1, EC-PRO Normal), a catheter with a stainless outer stiffener (Type2, EC-PRO Supported), a catheter with Echo Line (Type3, EC-PRO Master), and a catheter that has both the stainless outer stiffener and the echo line (Type4, EC-PRO Master Supported), all of which are offered in lengths of 18cm and 23cm, and various versions (v1-v8). A corresponding Trial catheter of the same length is provided for all four Types of the Kitazato ET Catheters. The Guide catheter is offered both with a straight shaft and a 30° pre-curved distal shaft. The Stylet Cannula (Type5) is an optional accessory used when additional rigidity is needed during insertion of the cannula through the cervix, and is offered in two lengths, 13.5 cm and 18.5 cm.

    During the implantation procedure of the embryo into the uterus, the catheter is introduced into the uterine cavity through the cervix, then the embryo is delivered into the uterine cavity by plunging the syringe that is coupled at the connector end of the catheter via a 6% taper luer lock. A syringe is not included with the Kitazato ET Catheters.

    AI/ML Overview

    The provided text describes the Kitazato ET Catheters, an embryo transfer catheter, and its substantial equivalence to a predicate device. It includes non-clinical performance data to support its safety and effectiveness.

    Here's the information requested, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Test TypeAcceptance CriteriaReported Device Performance
    Biocompatibility
    CytotoxicityPass (ISO 10993-5:2009)Passed
    SensitizationPass (ISO 10993-10:2010)Passed
    IrritationPass (ISO 10993-10:2010)Passed
    Mouse Embryo Assay (MEA)≥ 80% expanded to blastocyst at 96 hours (1-Cell MEA)Passed (1-cell mouse embryos, ≥ 80% expanded blastocysts at 96 hrs)
    Endotoxin< 20 EU/device< 20 EU/device (conforms to specification)
    Sterilization ValidationConformance to ISO 11135:2014 & ISO 10993-7:2008Passed
    Dimensional VerificationFinal sterile devices conform to dimensional specifications; luer connectors conform to ISO 594-1 and ISO 594-2Passed
    Appearance TestingPass visual inspection for scratches, burrs, and foreign objectsPassed
    Mechanical Tensile TestingMet established specificationsTensile strength met the established specifications
    Depth Mark Color-fastnessClearly legible after soaking in saline for 24 hours and 100 cycles of rubbingRemained clearly legible
    Package IntegrityPass dye penetration, seal strength, and visual inspection per ASTM 1929-15, ASTM 1866/F 1886M-09, and ASTM F88/F88M-15Passed
    Transport TestingPackaging maintained device integrityPackaging maintained device integrity
    Shelf Life TestingMaintained performance for 3 years (tensile strength, dimension, appearance, sterility, endotoxin, MEA, and color fastness)Established three (3) year shelf life following real time aging; parameters assessed (tensile strength, dimension, appearance, sterility, endotoxin, MEA, and color fastness)

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes non-clinical performance testing of the device itself and does not involve human subjects or clinical data for a "test set" in the context of AI/diagnostic algorithm evaluation. Therefore, details regarding sample size for a test set and data provenance (country of origin, retrospective/prospective) are not applicable here. The testing was performed on the device prototypes and manufactured units.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    This information is not applicable as the study is a non-clinical device performance test and does not involve establishing ground truth from human expert interpretations of data (e.g., medical images).

    4. Adjudication Method for the Test Set

    This information is not applicable as the study is a non-clinical device performance test and does not involve adjudication of expert interpretations.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The document describes non-clinical performance testing of a medical device (catheter), not an AI algorithm intended to assist human readers. Therefore, an effect size of human readers' improvement with vs. without AI assistance is not applicable.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    No, a standalone performance study in the context of an AI algorithm was not done. This document pertains to the physical performance characteristics of a medical catheter, not an AI or diagnostic algorithm.

    7. Type of Ground Truth Used

    For the non-clinical performance tests, the "ground truth" used are the defined specifications and standards for the device's physical, chemical, and biological properties. This includes:

    • ISO standards (e.g., ISO 10993 for biocompatibility, ISO 11135 for sterilization, ISO 11607 for packaging)
    • ANSI/AAMI standards (e.g., ANSI/AAMI ST72 for endotoxin)
    • ASTM standards (e.g., ASTM D4169 for transport, ASTM F88/F88M-15 for seal strength)
    • JIS standards (e.g., JIS L 0849 / JIS L 0803 for color-fastness)
    • Internal established specifications for tensile strength, dimensions, and appearance.
    • Biological assays (Mouse Embryo Assay) with specific biological endpoints (blastocyst development rate).

    8. Sample Size for the Training Set

    This information is not applicable. The document does not describe the development or testing of an AI algorithm, so there is no training set in that context. The "training" for the device would be its design and manufacturing process, and performance is tested against established specifications.