510(k) Data Aggregation

    K Number
    K243886

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-04-29

    (132 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ExpressCore Full Core Biopsy Device is intended for use in obtaining biopsies from soft tissues.

    Device Description

    The ExpressCore Device is a sterile, single patient use, hand held full core biopsy device comprised of a plastic handle and three stainless steel needle components that work in sequence to capture a full core biopsy sample. The device is capable of infinitely variable throw distance between 10 and 30mm and is equipped with multiple firing modes to accommodate procedural requirements and physician preference. Multiple device gauge size and needle option combinations have been developed from 12ga x 10cm to 18ga x 25cm.

    AI/ML Overview

    This FDA 510(k) clearance letter pertains to a device, the ExpressCore Biopsy Device, not an AI/ML software. The provided text details the device's characteristics, intended use, and comparison to predicate devices, but it does not include information about a study that proves the device meets specific acceptance criteria in the context of an AI/ML algorithm.

    Therefore, I cannot fulfill your request for:

    • A table of acceptance criteria and reported device performance (for an AI/ML algorithm)
    • Sample sized used for the test set and data provenance (for an AI/ML algorithm's test set)
    • Number of experts used to establish ground truth and qualifications (for an AI/ML algorithm's ground truth)
    • Adjudication method (for an AI/ML algorithm's test set)
    • MRMC comparative effectiveness study (as this is a physical device, not an AI system assisting human readers)
    • Standalone performance (as this is a physical device, not an AI algorithm)
    • Type of ground truth used (for an AI/ML algorithm)
    • Sample size for the training set (as this is a physical device, not an AI algorithm)
    • How the ground truth for the training set was established (as this is a physical device, not an AI algorithm)

    The document focuses on the physical and functional characteristics of the biopsy device and demonstrates substantial equivalence to existing predicate devices through bench performance testing, biocompatibility, sterilization, packaging, and shelf life data.

    Key information from the provided text relevant to device testing (not AI/ML):

    • Type of Study: Non-Clinical Bench Performance Testing, Biocompatibility, Sterilization, Packaging, Shelf Life.
    • Purpose of Study: To confirm the device meets all system requirements and is substantially equivalent to the predicate device.
    • Specific Tests Mentioned:
      • Usability of the Device
      • Biopsy Sample Collection
      • Finger Cannula Tensile Test
      • Spoon Cannula Tensile Test
      • Stylet Tensile Test
      • Representative Tissue Sampling

    This document is a 510(k) clearance for a physical medical device, not a review of an AI/ML-based medical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K240429

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-11-07

    (268 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    2 - 65
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Trilogy Tissue Marker is intended for use to attach to breast tissue at the surgical breast biopsy or percutaneous breast biopsy to radiographically mark the biopsy procedure, and be permanently visible under MRI, x-ray and ultrasound.

    Device Description

    The Trilogy Tissue Marker is a sterile, single use device comprised of a disposable delivery device preloaded with a tissue marker. The disposable delivery device includes an introducer needle comprised of a plastic molded deployment handle, a thumb slide, a 14 ga. cannula with 1 cm depth marks and a push rod. The tissue marker is preloaded in the distal end of the cannula. Trilogy tissue markers are made of a non-resorbable polymer embedded with a nitinol shape, allowing for permanent visibility under ultrasound, x-ray, and MRI. This device offers the choice of three unique tissue marker shapes (i) Ring (ii) Cross (iii) Ribbon.

    AI/ML Overview

    The provided document is a 510(k) summary for the Trilogy Tissue Marker. It details a comparison between the subject device and a predicate device (EasyMark™ Tissue Marker) to demonstrate substantial equivalence, rather than a study about acceptance criteria for an AI device. Therefore, much of the requested information regarding AI device acceptance criteria and study design is not available in this document.

    However, I can extract the information related to the device performance and the testing conducted to support its substantial equivalence with the predicate device.

    Here's a breakdown of the available information:

    1. A table of acceptance criteria and the reported device performance

    The document does not detail specific quantitative acceptance criteria or performance metrics in a readily extractable table form. Instead, it lists the types of performance tests conducted and states that the device meets "all system requirements" and is "substantially equivalent" to the predicate device.

    Performance Test CategoryDevice Performance (as stated in the document)
    Performance Testing – BenchConfirmed to meet all system requirements and is substantially equivalent to the predicate device.
    Accuracy of Marker Deployment(Implicitly meets requirements for substantial equivalence)
    Marker Deployment Force(Implicitly meets requirements for substantial equivalence)
    Usability of the Device(Implicitly meets requirements for substantial equivalence)
    Imaging Assessment(Implicitly meets requirements for substantial equivalence)
    Safety and Compatibility in Magnetic Resonance (MR) Environment(Implicitly meets requirements for substantial equivalence)
    Tissue Marker Migration Potential(Implicitly meets requirements for substantial equivalence)
    Delivery Device Cannula Tensile Test(Implicitly meets requirements for substantial equivalence)
    Delivery Device Push Rod Tensile Test(Implicitly meets requirements for substantial equivalence)
    Biocompatibility(Meets requirements for substantial equivalence)
    Sterilization(Meets requirements for substantial equivalence)
    Packaging(Meets requirements for substantial equivalence)
    Shelf Life(Meets requirements for substantial equivalence)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not provide details on the sample sizes used for the "Non-Clinical Bench Performance Testing." It also does not specify data provenance (e.g., country of origin or retrospective/prospective nature), as the testing appears to be primarily bench-based (laboratory testing).

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This information is not applicable and not provided. The study described is not a clinical study involving expert interpretation of medical images or data from human subjects. It focuses on the physical and functional aspects of a medical device (a tissue marker).

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    This information is not applicable and not provided, as the study is not an AI performance evaluation involving multiple readers or complex ground truth adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A multi-reader multi-case (MRMC) comparative effectiveness study was not done. This type of study relates to the performance of AI systems in assisting human readers, which is not the subject of this 510(k) submission.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    A standalone performance study of an algorithm was not done. This submission is for a physical medical device (tissue marker), not an AI algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    The concept of "ground truth" as typically used in AI studies (e.g., expert consensus, pathology, outcomes data) is not directly applicable here. The "ground truth" for the bench performance testing of this physical device would stem from objective measurement standards and engineering specifications. For instance, "Accuracy of Marker Deployment" would be evaluated against designed specifications for marker placement, and "Tissue Marker Migration Potential" would be assessed against defined metrics for stability.

    8. The sample size for the training set

    This information is not applicable and not provided, as this is not an AI device that requires a training set.

    9. How the ground truth for the training set was established

    This information is not applicable and not provided, as this is not an AI device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K183503

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2019-10-17

    (304 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EasyMark Tissue Marker is intended for use to attach to soft breast tissue at the surgical site during an open surgical breast biopsy or a percutaneous breast biopsy to radiographically mark the biopsy procedure.

    Device Description

    The EasyMark Tissue Marker is a sterile, single use device comprised of a disposable delivery device preloaded with a tissue marker. The disposable delivery device includes an introducer needle comprised of a plastic molded deployment handle, a thumb slide, a 17 ga. cannula with 1 cm depth marks and a push rod. The tissue marker is preloaded in the distal end of the cannula. This device offers the user the choice of two unique tissue markers (i) 316 LVM Stainless Steel Anchor shaped Marker (ii) Titanium 6AI-4V ELI Ribbon shaped Marker.

    AI/ML Overview

    The provided text describes the 510(k) premarket notification for the EasyMark Tissue Marker. It details the device, its indications for use, and a comparison to predicate and reference devices. However, it does not contain specific acceptance criteria or the study results proving the device meets those criteria, nor does it include information about a standalone study, MRMC study, sample sizes, ground truth establishment, or expert details for such studies.

    Based on the information provided, here's what can be inferred and what is missing:

    1. A table of acceptance criteria and the reported device performance
    This information is not provided in the document. The document lists the types of non-clinical tests performed but does not state the specific acceptance criteria for each test or the quantitative results from those tests.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    This information is not provided as the document only states "non-clinical testing and evaluation was conducted." It doesn't detail specific sample sizes for any of the performance tests.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    This information is not provided as the document does not describe any studies involving human experts or ground truth establishment in this manner.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    Not applicable, as no studies involving human interpretation or adjudication are described.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    This information is not provided. The EasyMark Tissue Marker is a physical implantable device, not an AI software. Therefore, an MRMC study with human readers improving with AI assistance would not be relevant to this device.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    This information is not provided. The device is a physical tissue marker, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    This information is not provided. The document describes performance testing, but not the specific methods used to establish ground truth for metrics like "Accuracy of Marker Deployment." It likely refers to objective measurements against predefined specifications.

    8. The sample size for the training set
    Not applicable. The EasyMark Tissue Marker is a physical device, not an AI model that requires a training set.

    9. How the ground truth for the training set was established
    Not applicable. The EasyMark Tissue Marker is a physical device, not an AI model that requires a training set.


    Summary of Provided Information (within the scope of the document):

    Device Name: EasyMark Tissue Marker
    Intended Use: To attach to soft breast tissue at the surgical site during an open surgical breast biopsy or a percutaneous breast biopsy to radiographically mark the biopsy procedure.

    Non-clinical Testing and Evaluation Performed to demonstrate substantial equivalence to the predicate device:

    • Performance Testing Bench:
      • Accuracy of Marker Deployment
      • Marker Deployment Force
      • Usability of the Device
      • Imaging Assessment (X-ray and MRI)
      • Safety and Compatibility in Magnetic Resonance (MR) Environment
      • Tissue Marker Migration Potential
      • Delivery Device Cannula Tensile Test
      • Delivery Device Push Rod Tensile Test
    • Biocompatibility
    • Sterilization
    • Packaging
    • Shelf Life

    Predicate Device: UltraClip® Tissue Marker, K993785
    Reference Device (for biocompatibility of delivery device): SelectCore, K093256

    Conclusion: The non-clinical testing and evaluation demonstrated that the EasyMark Tissue Marker "does not raise new concerns of safety or effectiveness and is substantially equivalent to the predicate device."

    Disclaimer: The provided document is a 510(k) summary for a medical device. It focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing. It does not typically include detailed acceptance criteria or study results in the format requested, especially since the device is a physical marker and not an AI or diagnostic software tool.

    Ask a Question

    Ask a specific question about this device

    K Number
    K112945

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2011-11-03

    (30 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lymph nodes and various soft tissue tumors.

    Device Description

    The EliteCore™ Full Core Biopsy Device is a sterile, disposable device which features a stainless steel cutting cannula with spoon, rotating coring cannula and stylet. The device is comprised of a plastic housing that contains the mechanically actuated components. The key performance attribute of the EliteCore™ Full Core Biopsy Device is the capability of obtaining a full core specimen. A variable throw feature allows the user to choose a throw setting ranging from 13 to 25 mm.

    AI/ML Overview

    The EliteCore™ Full Core Biopsy Device is a sterile, disposable device for obtaining biopsies from soft tissues. The performance of this device was assessed by comparing it to a predicate device, the Revolution® Full Core Biopsy Device (K090611). The tests conducted aimed to confirm that the quality of samples obtained with the EliteCore™ device is equivalent to that of the predicate device.

    1. Table of Acceptance Criteria and Reported Device Performance:

    TestAcceptance Criteria (How Equivalent)Reported Device Performance
    Package InspectionIdentical testing performed on predicate device with equivalent results.Visually inspected packaging for damage, loose components, secure tip protector, and variable throw set at 22mm. Results were equivalent to predicate device.
    Burst TestIdentical testing performed on predicate device with equivalent results.Measured pressure at which sealed tray fails. Results were equivalent to predicate device.
    Device InspectionIdentical testing performed on predicate device with equivalent results.Visually inspected device for damage. Results were equivalent to predicate device.
    Dimensional (Variable Throw)Identical testing performed on predicate device with equivalent results.Measured variable throw desired vs. actual. Results were equivalent to predicate device.
    Dimensional (Variable Throw after Dry Fires)Identical testing performed on predicate device with equivalent results.Measured variable throw after dry fires, desired vs. actual. Results were equivalent to predicate device.
    Force to Fire Front Triggers (RH / LH)Identical testing performed on predicate device with equivalent results.Measured force to fire front triggers. Results were equivalent to predicate device.
    Force to Fire Rear TriggerIdentical testing performed on predicate device with equivalent results.Measured force to fire rear trigger. Results were equivalent to predicate device.
    Tissue Test (Chicken and Liver)Identical testing performed on predicate device with equivalent results.Able to take tissue samples in chicken and liver. Results were equivalent to predicate device.
    Tissue Test (Various Throw Distances)Additional testing that was not performed on predicate.Able to take tissue samples at various throw distances. (No direct comparison to predicate stated for this specific aspect, but overall tissue test performance confirmed equivalence). The document states identical testing performed on predicate device with equivalent results.
    Tissue Test (1st vs. Last Sample Weight)Identical testing performed on predicate device with equivalent results.Measured 1st sample as compared to last sample for sample weight and verified no loss of effectiveness. Results were equivalent to predicate device.
    Mates with Adjustable Coaxial and Able to Take SampleIdentical testing performed on predicate device with equivalent results.The device successfully mated with an adjustable coaxial and was able to take a sample. The results were equivalent to the predicate device.
    Verify Weld Integrity (Visual Inspection after Functional Testing)Identical testing performed on predicate device with equivalent results.Visually inspected device for damage along weld joints after functional testing. Results were equivalent to predicate device.
    Dimensional (Cannula, Spoon, and Stylet Migration)Identical testing performed on predicate device with equivalent results.Measured cannula, spoon, and stylet for migration. Results were equivalent to predicate device.
    Dimensional (Joint Strength for Cannula, Spoon, and Stylet)Identical testing performed on predicate device with equivalent results.Measured the joint strength for the cannula, spoon, and stylet. Results were equivalent to predicate device.

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not explicitly state the specific number of devices or biopsy attempts used for each test. Instead, it broadly states that "Identical testing performed on predicate device w/ equivalent results" for most tests. The tissue tests mention "chicken and liver" and "various throw distances," indicating a lab-based, prospective evaluation with animal tissue. No country of origin for the data is specified, but given the submission to the FDA, it is presumed to be conducted by the manufacturer, Inrad, Inc., located in Kentwood, MI, USA.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

    This information is not provided in the document. The performance testing appears to be based on objective physical measurements and visual inspections rather than expert interpretation of biopsy samples.

    4. Adjudication Method for the Test Set:

    Not applicable. The tests are focused on device performance metrics (e.g., dimensions, forces, ability to take tissue samples) rather than diagnostic accuracy that would require an adjudication process.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    Not applicable. This device is a biopsy instrument, not an AI-powered diagnostic tool. The performance evaluation is based on engineering and basic tissue procurement capabilities.

    6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This is a physical medical device, not an algorithm.

    7. The Type of Ground Truth Used:

    The ground truth for most tests was based on objective physical measurements (e.g., pressure, force, dimensions, weight) and visual inspections for damage or proper function. For tissue tests, the ground truth was the ability to successfully obtain a tissue sample and verify its properties (e.g., weight, no loss of effectiveness). The comparison was made against the performance of a legally marketed predicate device.

    8. The Sample Size for the Training Set:

    Not applicable. This device is not an AI/ML device requiring a training set. The performance testing is for a physical medical instrument.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable. (See #8)

    Ask a Question

    Ask a specific question about this device

    K Number
    K093399

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-11-19

    (17 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for use in obtaining biopsies from soft tissues such as liver, breast, kidney, prostate, spleen, lymph nodes and various soft tissue tumors. For breast biopsy this product is for diagnosis only - not for therapeutic use.

    Device Description

    The PreciseCore™ Biopsy Device is a sterile, disposable device which features a stainless steel cutting cannula and stylet. The device is comprised of a plastic housing that contains the mechanically actuated components. The stylet of the device is stationary, therefore acquiring tissue without an advancement of the needle set beyond the visualized tip.

    AI/ML Overview

    This 510(k) submission (K093399) for the PreciseCore™ Biopsy Device primarily focuses on demonstrating substantial equivalence to a predicate device through performance testing. However, it does not provide detailed acceptance criteria, specific performance metrics in a tabular format, or a comprehensive study report with the kind of information typically requested (e.g., sample sizes for different sets, ground truth establishment details, expert qualifications, MRMC studies).

    Here's an analysis based on the provided text, highlighting what is and is not present:


    Acceptance Criteria and Device Performance (Based on available information)

    The document states: "Performance testing confirms that the quality of samples obtained with the PreciseCore™ Biopsy Device is equivalent to that of the predicate device." This implies the acceptance criterion was "equivalence in sample quality" to the predicate, but no specific metrics or thresholds are provided.

    Acceptance Criteria (Inferred)Reported Device Performance
    Equivalent sample quality to predicate device (K994272 – Promex Automated Core Biopsy Device, also known as Inrad® AccuCore Single Action Core Biopsy Device and US Biopsy SABD)Confirmed to be equivalent to the predicate device in sample quality.

    Missing Information (Based on the provided text):

    The provided 510(k) summary does not contain the following details:

    1. Sample sizes used for the test set and the data provenance: There is no mention of a specific clinical "test set" in terms of patient or tissue samples. The performance testing appears to be functional or mechanical, comparing the device's output (sample quality) to a predicate.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: This is not applicable as there isn't a described "test set" requiring expert ground truth establishment in the context of clinical images or diagnostic outcomes. The "quality of samples" likely refers to characteristics like length, integrity, or cellularity, evaluated by laboratory personnel or pathologists, but no specifics are given.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable for the reasons above.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance: This is not applicable. The device is a physical biopsy instrument, not an AI-powered diagnostic or imaging tool.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: This is not applicable, as it is a physical biopsy device, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): While the "quality of samples" would likely refer to histopathological evaluation, the specific method of establishing this ground truth for the comparison is not detailed.
    7. The sample size for the training set: Not applicable, as this is not an AI/ML device requiring a training set.
    8. How the ground truth for the training set was established: Not applicable.

    Summary of Study (Based on provided text):

    The submission relies on a "Performance Testing Summary" to demonstrate substantial equivalence.

    • Study Design/Purpose: To confirm that the quality of samples obtained with the PreciseCore™ Biopsy Device is equivalent to that of the predicate device (K994272 – Promex Automated Core Biopsy Device, Inrad® AccuCore Single Action Core Biopsy Device, and US Biopsy SABD).
    • Methodology (General): "Performance testing" was conducted. While no specifics are given, for a biopsy device, this typically involves mechanical testing (e.g., penetration force, firing reliability) and potentially benchtop testing on tissue phantoms or ex vivo tissue to assess core length, integrity, and cellular yield when compared to the predicate device.
    • Data Provenance: Not specified, but likely laboratory/benchtop testing rather than clinical data from a specific country.
    • Result: The testing "confirms that the quality of samples obtained with the PreciseCore™ Biopsy Device is equivalent to that of the predicate device."

    Conclusion:

    The K093399 submission successfully demonstrated substantial equivalence for the PreciseCore™ Biopsy Device by asserting that its performance testing confirmed sample quality equivalent to its predicate. However, it lacks the detailed breakdown of acceptance criteria, specific performance metrics, and clinical study information (like sample sizes, ground truth establishment, expert qualifications, or MRMC studies) that would be expected for an AI/ML device or a device requiring detailed clinical validation data. This is typical for 510(k)s for predicate-based medical devices where the equivalence is often shown through functional, mechanical, and benchtop testing rather than extensive clinical efficacy trials.

    Ask a Question

    Ask a specific question about this device

    K Number
    K093256

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-11-17

    (29 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for use in obtaining biopsies from soft tissues such as liver, breast, kidney, prostate, spleen, lymph nodes and various soft tissue tumors. For breast biopsy this product is for diagnosis only - not for therapeutic use.

    Device Description

    The SelectCore™ Variable Throw Biopsy Device is a sterile, disposable device which features a stainless steel cutting cannula and stylet. The device is comprised of a plastic housing that contains the mechanically actuated components. The device offers the user the choice of either single fire or fully automatic modes. A variable throw feature allows the user to choose a throw setting ranging from 10 to 25 mm.

    AI/ML Overview

    The provided text does NOT contain information about acceptance criteria or a study that proves the device meets specific performance criteria. The document is a 510(k) summary for the SelectCore™ Variable Throw Biopsy Device, which focuses on establishing substantial equivalence to a predicate device.

    Specifically, the "Performance Testing Summary" section (in {1}) states: "Performance testing confirms that the quality of samples obtained with the SelectCore™ Variable Throw Biopsy Device is equivalent to that of the predicate device." However, it does not provide details on:

    • Specific acceptance criteria: There are no defined numerical or qualitative targets for device performance.
    • The study design: No information is given about how the "quality of samples" was assessed, what metrics were used, or what kind of study was conducted.
    • Any of the requested details about sample size, ground truth, expert involvement, or comparative effectiveness.

    Therefore, I cannot fulfill your request for a table of acceptance criteria and reported performance, nor can I provide the detailed study information. The provided text does not contain this level of detail.

    Ask a Question

    Ask a specific question about this device

    K Number
    K090611

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2009-06-04

    (90 days)

    Product Code
    Regulation Number
    876.1075
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for use in obtaining biopsies from soft tissues such as liver, breast, kidney, prostate, spleen, lymph nodes and various soft tissue tumors.

    Device Description

    The Revolution™ Full Core Blopsy Device is a sterile, disposable device which features a stainless steel cutting cannula with spoon, rotating coring cannula and stylet. The device is comprised of a plastic housing that contains the mechanically actuated components. The key performance attribute of the Revolution™ Full Core Biopsy Device is the capability of obtaining a full core specimen. A variable throw feature allows the user to choose a throw setting ranging from 10 to 25 mm.

    AI/ML Overview

    The provided text describes a medical device, the Revolution™ Full Core Biopsy Device, and its 510(k) summary for FDA clearance. However, it does not contain specific acceptance criteria, detailed study designs, or performance metrics beyond a general statement of equivalency.

    Therefore, I cannot fulfill your request for a table of acceptance criteria and reported device performance, nor can I provide information on sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, or training set details as this information is not present in the provided document.

    The document states:

    • "Performance testing confirms that the quality of samples obtained with the Revolution™ Full Core Biopsy Device is equivalent to that of the predicate device." This is a high-level statement indicating a study occurred, but without details on how "quality" was measured or the specific criteria for "equivalency."

    To answer your questions, the following information would be needed:

    • A section detailing the specific performance tests conducted.
    • The acceptance criteria for each test (e.g., in terms of specimen length, integrity, diagnostic yield, etc.).
    • The results of those tests compared to the predicate device or a defined threshold.
    • Details about the methodology of sample quality assessment (e.g., histological review, pathological assessment).
    Ask a Question

    Ask a specific question about this device

    K Number
    K063238

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2007-01-30

    (96 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The intended use of the Ultraclip II WING and Ultraclip II COIL is to attach to soft breast tissue at the surgical site during an open surgical breast biopsy or a percutaneous breast biopsy to radiographically mark the location of the biopsy procedure.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but based on the provided document, there is no information available regarding the acceptance criteria, study details, or performance metrics of the device (Ultraclip® II WING, Ultraclip® II COIL).

    The document is a 510(k) clearance letter from the FDA, which determines substantial equivalence to a predicate device. It confirms that the device can be marketed but does not include the detailed study results or acceptance criteria that would typically be found in a clinical study report or a more comprehensive submission document.

    Therefore, I cannot provide the requested information, including:

    1. A table of acceptance criteria and the reported device performance
    2. Sample sized used for the test set and the data provenance
    3. Number of experts used to establish the ground truth for the test set and their qualifications
    4. Adjudication method
    5. If a multi reader multi case (MRMC) comparative effectiveness study was done and its effect size
    6. If a standalone performance study was done
    7. The type of ground truth used
    8. The sample size for the training set
    9. How the ground truth for the training set was established
    Ask a Question

    Ask a specific question about this device

    K Number
    K042341

    Validate with FDA (Live)

    Device Name
    ULTRACLIP II US
    Manufacturer
    Date Cleared
    2004-09-20

    (21 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ultraclip II US intended use is to attach to soft breast tissue at the surgical site during an open surgical breast biopsy or a percutaneous breast biopsy procedure.

    Device Description

    Not Found

    AI/ML Overview

    I am sorry, but the provided text from the FDA 510(k) K042341 for the Inrad, Inc. Ultraclip® II US does not contain information about acceptance criteria, device performance studies, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, or training set details.

    This document is a letter from the FDA granting substantial equivalence to the Ultraclip® II US to a legally marketed predicate device. It identifies the device, its regulation number, and states that it can be marketed. It also includes an "INDICATIONS FOR USE STATEMENT" which describes the intended use of the device.

    Therefore, I cannot provide the requested table and study details based on the given input.

    Ask a Question

    Ask a specific question about this device

    K Number
    K041201

    Validate with FDA (Live)

    Device Name
    ULTRACLIP II MR
    Manufacturer
    Date Cleared
    2004-06-03

    (27 days)

    Product Code
    Regulation Number
    878.4750
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Ultraclip II MR intended use is to attach to soft breast tissue at the time of an open surgical breast biopsy or a percutaneous breast biopsy to radiographically mark the location of the biopsy procedure.

    Device Description

    Not Found

    AI/ML Overview

    This document is a 510(k) clearance letter from the FDA for a medical device called Ultraclip® II MR. It does not contain the detailed study information, acceptance criteria, or performance data of the device.

    Therefore, I cannot provide the requested information, which typically would be found in the manufacturer's submission or a separate scientific publication. This letter only confirms that the device has received market clearance based on substantial equivalence to a predicate device.

    Ask a Question