Search Results
Found 52 results
510(k) Data Aggregation
(269 days)
Ask a specific question about this device
(273 days)
OTTO is intended for use on swabbable needlefree connectors as a cover to protect the needlefree connectors from potential contamination. OTTO acts as a physical barrier to contamination between line accesses and also serves as a disinfecting cleaner for use prior to line access.
OTTO will disinfect the connector thirty (30) seconds after application and maintains a disinfected connector surface for up to seven (7) days if not removed.
The OTTO™ Disinfecting Cap is a sterile disposable cap designed to disinfect swabbable needlefree female connectors with 70% isopropyl alcohol (IPA) when attached to an intravenous (IV) catheter. The OTTO™ product line complements the ICU Medical Clave family of needlefree connectors; however, it is designed to work with other on-market swabbable needlefree female connectors used in peripheral intravenous catheters, indwelling central venous or arterial access catheters.
N/A
Ask a specific question about this device
(266 days)
The ClaveQS™ Bag is intended for compounding and administration of parenteral solutions. Medication transfer in and out of the ClaveQS™ Bag is done using aseptic technique.
Ask a specific question about this device
(97 days)
Clave Neutral-Displacement Connectors (MicroClave™/ NanoClave™/ Clave™Neutron) are intended for the aspiration, injection, or gravity/pump flow of IV fluids and blood upon insertion of a male luer connector. Clave Connectors may be used with power injectors at a maximum pressure of 400 psi and a maximum flow rate of 10ml/sec. Clave Connectors will prevent microbial ingress for seven (7) days.
Clave Neutral-Displacement Connectors are needlefree, bi-directional connectors that utilize a pre-slit septum which prevents microbial ingress when in the un-activated state and allows access to the fluid path when activated with an ISO 80369-7 compliant male luer. The septum offers neutral displacement of fluid during connection or disconnection of a male luer and self-seals upon disconnection to prevent fluid loss or air ingress. The Neutron incorporates additional technology that will prevent fluid displacement resulting from syringe plunger compression; patient vascular pressure changes, such as coughing or sneezing; and IV solution container run-dry. The Clave Family of Connectors do not require a cap but may be used with disinfecting caps containing 70% isopropyl alcohol.
The provided text is a 510(k) clearance letter and summary for the Clave™ Neutral-Displacement Needlefree Connectors. This document describes a medical device, not an AI/ML algorithm. Therefore, many of the requested fields related to AI/ML device performance (e.g., effect size of human readers with AI vs. without AI, ground truth for training set, number of experts for ground truth) are not applicable.
However, the document does describe the device's acceptance criteria and the studies performed to demonstrate equivalence and support new claims.
Here's the information extracted from the document:
1. A table of acceptance criteria and the reported device performance
The document primarily focuses on demonstrating substantial equivalence to a predicate device and supporting new claims through non-clinical performance data and a retrospective clinical study.
Acceptance Criteria and Device Performance (Based on "Summary of Non-Clinical Testing" and "Performance Data: Non-Clinical Testing Summary")
| Feature/Standard | Acceptance Criteria (Implied by standard conformance) | Reported Device Performance |
|---|---|---|
| Microbial Ingress (New claim) | Prevention of microbial ingress for seven (7) days | Achieved "prevention of microbial ingress for a worst-case simulated use protocol of seven (7) days, using Guidance for Industry and Staff: Intravascular Administration Sets Premarket Notification Submissions [510(k)], Section 8; Microbial Ingress Testing." |
| Power Injection Pressure (Increased claim) | Maximum pressure of 400 psi | "May be used with power injectors at a maximum pressure of 400 psi" / "The difference in pressure infusion rating... is supported by testing in compliance with ANSI/AAMI CN 27." |
| Particulate Contamination | Meets USP <788> requirements | "Particulate contamination testing was performed by following USP <788> to demonstrate particulate levels on the subject devices meets USP <788> requirements." |
| Pressure Infusion Flow Rate | Maximum flow rate of 10ml/sec | "May be used with power injectors at a maximum flow rate of 10ml/sec." |
| Biocompatibility | Conforms to ISO 10993-1, "Externally Communicating Device with Blood Path, Direct Contact, for a Prolonged Duration (>24hours to 30 days)." | "Conforms to ISO 10993-1" as per FDA guidance. |
| Sterility | Sterilized by irradiation, validated per ISO 11137-1/A1 & ISO 11137-2 | "Irradiation" / "Radiation Sterilization Validation" |
| Sterile Barrier System | Validation per ISO 11607-1, ISO 11607-2 | "Sterile Barrier System Validation" |
| Pyrogenicity | Non-pyrogenic (meets USP <85>) | "Non-pyrogenic" / "Bacterial endotoxins" testing per ANSI/AAMI ST72 & USP <85>. |
| ISO 8536-4 (Infusion Equipment) | Meet requirements for particulate contamination, leakage, tensile strength, flow rate, protective caps, chemical requirements, pyrogenicity, seal and snap tests, seal test, seal tests – post durability. | All tests passed, supporting substantial equivalence. |
| ISO 8536-8 (Sterile Infusion Sets for single use, general purpose) | Meet requirements for leakage. | All tests passed, supporting substantial equivalence. |
| ISO 8536-10 (Infusion sets for use with pressure infusion apparatus) | Meet requirements for avoidance of air bubbles. | All tests passed, supporting substantial equivalence. |
| ISO 80369-7 (Luer Connectors) | Meet dimensional requirements, fluid leakage, sub-atmospheric pressure air leakage, stress cracking, resistance to separation from axial load, resistance to separation from unscrewing, resistance to overriding. | All tests passed, supporting substantial equivalence. |
| ANSI/AAMI CN27 (Luer-activated, Needleless Connectors) | Meet requirements for non-interconnectability, flow rate, exposure to IPA, infusate compatibility, resistance to separation (axial load, unscrewing, overriding), backpressure, positive pressure fluid leakage, subatmospheric pressure air leakage (unactivated/activated), duration of activation, number of activations, priming volume, residual volume, hemolysis, power injection, microbial ingress, displacement volume. | All tests passed, supporting substantial equivalence and new claims (power injection, microbial ingress). |
| Clinical Efficacy (Bloodstream Infection) | Statistically significant lower relative risk of bloodstream infection and bloodstream infection-associated mortality when compared to non-Clave users. | RR of CLABSI was 0.93 (7% decreased risk, p=0.04) for Clave Family Connectors. RR of CLABSI was 0.81 (19% decrease, p=0.04) for high-volume users. |
2. Sample size used for the test set and the data provenance
- Non-Clinical Testing: The document does not specify the exact sample sizes for each non-clinical test (e.g., ISO 8536-4, ANSI/AAMI CN27). It states that "The subject device has been evaluated for the prevention of microbial ingress for a worst-case simulated use protocol of seven (7) days" and mentions conformance to various standards, which implicitly include testing of multiple units.
- Clinical Testing (CMS study):
- Sample Size: Not explicitly stated as a "sample size" in the context of device testing. This was a retrospective study that analyzed 2019 CMS data. The study "analyzed 2019 CMS data" to compare hospitals using Clave connectors to those not using them.
- Data Provenance:
- Country of Origin: United States (CMS data).
- Retrospective or Prospective: Retrospective.
- Specifics: "analyzed 2019 CMS data," "adjusting for Hospital characteristics," "acute-care Hospitals which utilized the Clave Family of Connectors... had a statistically significant lower relative risk... when compared to acute-care Hospitals that did not utilize Clave Connectors."
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- Not Applicable. This information is for AI/ML devices. The 510(k) is for a physical medical device (needlefree connector). Ground truth for the clinical study (bloodstream infection rates) would have been established through hospital records and CMS reporting, not by human experts adjudicating images or cases for AI training/testing.
4. Adjudication method for the test set
- Not Applicable. This information is typically for AI/ML devices involving human reviewer consensus. The clinical study used pre-existing CMS data on bloodstream infections.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not Applicable. This is not an AI/ML device. The clinical study was a comparison of hospital outcomes based on device usage, not human reader performance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not Applicable. This is not an AI/ML device.
7. The type of ground truth used
- Non-Clinical Testing: Ground truth is established by the specifications and measurement techniques defined in the referenced industry standards (e.g., ISO, ANSI/AAMI, USP). Performance is measured against these established parameters.
- Clinical Testing: The ground truth for the clinical claim (bloodstream infection reduction) was based on outcomes data from official government reporting (CMS data) regarding bloodstream infection rates (CLABSI) and bloodstream infection-associated mortality in acute-care hospitals.
8. The sample size for the training set
- Not Applicable. This is not an AI/ML device. There is no "training set."
9. How the ground truth for the training set was established
- Not Applicable. This is not an AI/ML device.
Ask a specific question about this device
(257 days)
The Plum Duo™ Precision IV Pump is intended for administration of parenteral fluids, medications, and whole blood and blood products through the following routes of administration: intravenous, intra-arterial, and subcutaneous.
The Plum Duo™ Precision IV Pump is intended for use in clinical environments in the hospital and other outpatient healthcare facilities by licensed healthcare professionals. These healthcare professionals are trained in the use of the infusion pump and the administration of therapies consistent with the intended use.
The Plum Duo™ Precision IV Pump is intended for adults, pediatric (including infants and children), and neonatal patient populations.
The Plum Duo™ Precision IV Pump is a large volume pump (LVP) with two independent pump channels that can deliver fluid to a patient from 1 to 4 lines independently. In addition, although the channels can operate independently, patient parameters can be shared across the channels to aid in ease of programming. The Plum Duo™ Precision IV Pump can only be used with dedicated PlumSet™ administration sets (not subject of this filing). Each pump channel accepts a cassette that is part of a PlumSet™ administration set and can connect to a primary and/or secondary container. The fluid is delivered from the upstream lines either serially (piggyback) or concurrently through the cassette to the downstream line. Each pump channel cassette has an independent downstream line (patient line), so the clinician can connect each downstream line to a single infusion site or two separate infusion sites. The flow rate accuracy precision has been improved (lower allowed variance) by implementing the new motor mechanism, as well as increased precision for programming concentration, flow rate and VTBI entries. The overall delivery accuracy of the system has improved to +/- 3% per TIR 101 standard condition testing and +/-5% for non-standard conditions.
The provided FDA 510(k) clearance letter and summary for the Plum Duo™ Precision IV Pump discuss the device's technical specifications and how it meets regulatory requirements for significant equivalence to a predicate device. However, this document does not describe the acceptance criteria and a study proving the device meets those criteria in the context of an AI/ML-driven medical device, as implied by the user's detailed request.
The Plum Duo™ Precision IV Pump is an infusion pump, a hardware medical device with embedded software, not an AI/ML diagnostic or prognostic tool. Therefore, the requested information (such as sample size for test sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details, etc.) does not apply to this specific device or the information provided in the 510(k) summary.
The 510(k) summary focuses on "Non-Clinical Testing" which includes:
- Verification testing of product requirements
- Human factors validation testing
- Reliability goals testing
- Safety assurance case following FDA Guidance for Infusion Pumps
- Flow rate and bolus accuracy testing per AAMI TIR101
- Software verification and validation per relevant FDA guidance documents (2005 and 2021 draft for software functions), which are focused on traditional software engineering principles, not AI/ML model performance.
- Human factors evaluations per FDA guidance and IEC 62366-1
- Electrical and Electromagnetic Compatibility testing per IEC 60601-1 and IEC 60601-1-2
- Cybersecurity testing per FDA guidance documents (2014 and 2016)
- Risk management activities per ISO 14971:2019 +A11 2021
Conclusion based on the provided document:
The provided document does not contain the information required to answer the specific questions about acceptance criteria and study proving an AI/ML device meets them. The device described, the Plum Duo™ Precision IV Pump, is an infusion pump, and its clearance relies on non-clinical performance and safety data relevant to its mechanical and software functions, not AI/ML-driven insights or diagnostics.
Therefore, I cannot populate the table or answer the specific questions regarding AI/ML study design and ground truth establishment for this device based on the given text. The text explicitly states: "Clinical evaluation is not required for this submission to support substantial equivalence." This further indicates that the type of studies and data provenance you are asking about (which are typical for AI/ML diagnostic devices) were not part of this 510(k) submission.
Ask a specific question about this device
(257 days)
The Plum Solo™ Precision IV Pump is intended for administration of parenteral fluids, medications, and whole blood and blood products through the following routes of administration: intravenous, intra-arterial, and subcutaneous.
The Plum Solo™ Precision IV Pump is intended for use in clinical environments in the hospital and other outpatient healthcare facilities by licensed healthcare professionals. These healthcare professionals are trained in the use of the infusion pump and the administration of therapies consistent with the intended use.
The Plum Solo™ Precision IV Pump is intended for adults, pediatric (including infants and children), and neonatal patient populations.
The Plum Solo™ Precision IV Pump is a large volume pump (LVP) with one pump channel that can deliver fluid to a patient from 1 to 2 lines independently. The Plum Solo™ Precision IV Pump can only be used with dedicated PlumSet™ administration sets (not subject of this filing). The pump channel accepts a cassette that is part of a PlumSet™ administration set and can connect to a primary and/or secondary container. The fluid is delivered from the upstream lines either serially (piggyback) or concurrently through the cassette to the downstream line. The flow rate accuracy precision has been improved (lower allowed variance) by implementing the new motor mechanism, as well as increased precision for programming concentration, flow rate and VTBI entries. The overall delivery accuracy of the system has improved to +/- 3% per TIR 101 standard condition testing and +/-5% for non-standard conditions.
The provided FDA 510(k) clearance letter pertains to the Plum Solo™ Precision IV Pump, which is a large volume infusion pump. The document focuses on demonstrating substantial equivalence to a predicate device, the Plum Duo™ Infusion System (K223607).
It's important to note that the detailed information typically associated with acceptance criteria and study results for AI/ML-enabled medical devices (like the number of experts, adjudication methods, MRMC studies, or specific ground truth methodologies for training data) is not present in this clearance letter. This is because an IV pump, while an advanced medical device, is not an AI/ML diagnostic or prognostic system that relies on interpreting complex data like medical images, waveforms, or patient parameters to derive a diagnosis or predict an outcome.
Instead, the acceptance criteria and study results described for the Plum Solo™ Precision IV Pump focus on engineering performance specifications, reliability, and human factors validation, which are standard for such devices.
Therefore, the following information is extracted and presented based on the provided text, and where information is not available (as it is not typically part of the clearance for this class/type of device), it will be explicitly stated.
Overview of Device and Study Focus
The Plum Solo™ Precision IV Pump is a large volume pump (LVP) designed for administering parenteral fluids, medications, and blood products. The primary goal of the 510(k) submission was to demonstrate substantial equivalence to its predecessor, the Plum Duo™ Infusion System. The key improvements highlighted are enhanced flow rate accuracy and increased precision for programming.
1. Acceptance Criteria and Reported Device Performance
The acceptance criteria are primarily defined by the performance targets for flow rate accuracy and bolus delivery accuracy, compared to the predicate device.
| Characteristic | Acceptance Criteria (Implied by Predicate Performance or New Target) | Reported Device Performance (Plum Solo™) |
|---|---|---|
| Flow Rate Delivery Accuracy | +/- 5% for 0.1 - 999 mL/hr (Predicate performance) | +/- 3% for 0.1 - 999 mL/hr |
| Cassette Service Duration for Flow Accuracy | 0 to 96 hours | 0 to 96 hours |
| Bolus Delivery Accuracy | +/- 5% for delivery volumes as low as 0.1 ml at 999 ml/hr | +/- 5% for delivery volumes as low as 0.1 ml at 999 ml/hr |
Note on Acceptance Criteria: For a device like an infusion pump, "acceptance criteria" are implicitly met when the device demonstrates performance that is either equivalent to or improved upon a legally marketed predicate, in addition to meeting relevant safety standards (e.g., IEC standards, risk management requirements, human factors). In this case, the improved flow rate accuracy (from +/-5% to +/-3%) is a key performance metric that exceeds the predicate, serving as a strong point of "acceptance" in terms of performance.
2. Sample Size Used for the Test Set and Data Provenance
The document does not specify a sample size for the test set in terms of "patient data" as this is not an AI/ML diagnostic device analyzing patient-derived data. Instead, testing involves engineering verification and validation runs on the physical device and its software. The provenance of such data would be controlled laboratory settings.
- Sample Size: Not applicable in the context of "patient data" or "images." Testing involved verification and validation of product requirements, reliability goals, and human factors. The exact number of test units or test repetitions for each performance metric (e.g., flow rate accuracy tests per AAMI TIR101) is not specified in the summary but would be detailed in internal test reports.
- Data Provenance: Not applicable in the sense of country of origin for "data" like patient scans. This is laboratory-generated performance data from device testing. The testing was conducted as part of the device's design verification and validation processes. It is implicitly retrospective in the sense that the testing was performed before the 510(k) submission.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications
This concept is not applicable to the type of device discussed. Ground truth for an infusion pump's performance (e.g., flow rate, bolus accuracy) is established through precise physical measurements using calibrated equipment and reference standards (e.g., gravimetric methods for flow rate), not through expert interpretation of medical data.
- Number of Experts: Not applicable.
- Qualifications of Experts: Not applicable.
4. Adjudication Method for the Test Set
This concept is not applicable to the type of device discussed. Adjudication (e.g., 2+1, 3+1 for imaging consensus) is used for subjective evaluations where multiple human readers assess data. Performance metrics for an infusion pump are objective and measured.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
No, a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This type of study is specifically designed for assessing the impact of AI on human reader performance, typically in diagnostic imaging or similar fields. An infusion pump's primary function is fluid delivery, not diagnostic interpretation.
6. If a Standalone (Algorithm-Only Without Human-in-the-Loop) Performance Study was done
The term "standalone performance" in the context of an infusion pump refers to its ability to accurately deliver fluids based on its internal mechanics and software programming, independent of immediate human intervention during the delivery phase. The document details that:
- Flow rate and bolus accuracy testing were conducted by following AAMI TIR101. This is a standard for evaluating the performance of infusion pumps without a human directly influencing each delivery cycle. This essentially serves as the "standalone" performance assessment for this device.
- The reported performance of +/- 3% for flow rate accuracy directly reflects this standalone capability.
7. The Type of Ground Truth Used
The ground truth for an infusion pump's performance is established through physical measurements against established standards, not through expert consensus, pathology, or outcomes data in the medical sense.
- Ground Truth: Utilized reference standards and precise measurement techniques (e.g., gravimetric measurement for flow rate) as per AAMI TIR101. The "truth" is the actual volume delivered compared to the commanded volume, measured by calibrated instruments.
- Human Factors Validation: While not "ground truth" per se, human factors evaluations were conducted to validate the effectiveness of safety-critical use-related features/functionality and use error-related mitigations, in accordance with FDA guidance and IEC 62366-1. This ensures the user interface and interactions are safe and effective.
8. The Sample Size for the Training Set
This concept is not applicable to the type of device discussed. Infusion pumps are not "trained" on large datasets in the way AI/ML algorithms are. Their operational parameters are set by design and engineering specifications. While the device contains software, it's not a machine learning model that learns from a "training set."
9. How the Ground Truth for the Training Set Was Established
This concept is not applicable because there is no "training set" for an infusion pump in the AI/ML sense. The device's functionality is a result of engineering design, mathematical algorithms for pump control, and calibration, not machine learning from data.
Ask a specific question about this device
(257 days)
The LifeShield™ Infusion Safety Software Suite is a collection of software products that facilitates networked communication between compatible systems. The Infusion Safety Software Suite provides trained healthcare professionals the ability to manage data for compatible infusion pumps. All data entry and validation of infusion parameters on compatible infusion pumps is performed by a trained healthcare professional. LifeShield™ Infusion Safety Software Suite is indicated for use in patients including adult, pediatric and neonate undergoing infusion therapy with connected compatible infusion pumps (as per the indications for use specified for the compatible infusion pump).
The LifeShield™ Drug Library Management (DLM) software product is intended to be used by pharmacists to create, configure, edit, and manage drug library data, including infusion pump settings, for use with compatible infusion pumps. Drug library contents are constructed based on the healthcare provider's defined best practices.
The LifeShield™ Clinical Dashboards & Reports (CDR) software product provides trained healthcare professionals with the capability to view and manage infusion information collected from compatible infusion pumps. Healthcare professionals may choose to use the collected infusion information to support continuous quality improvement programs, or to analyze and trend various aspects of the infusion pumps and therapies used. It is not intended to be a substitute for good clinical management practices, nor does its operation create decisions or treatment pathways.
The LifeShield™ Data Flow Management (DFM) software product is intended to facilitate bi-directional communication with compatible infusion pumps, information technology systems, and other LifeShield™ Infusion Safety Software Suite products. LifeShield™ DFM provides a way to automate the programming of infusion parameters, thereby decreasing the number of manual steps necessary to enter infusion data. LifeShield™ DFM forwards infusion-related information received from the infusion pump to compatible information technology systems.
The LifeShield™ Infusion Safety Software Suite is a cloud-based platform provided as a software-as-a-service (SaaS) designed to be compatible with the Plum Duo™ Precision IV infusion pump and Plum Solo™ Precision IV infusion pump. The LifeShield Infusion Safety Software Suite is hosted by Amazon Web Services (AWS) as its cloud provider.
LifeShield™ Infusion Safety Software Suite consists of a collection of software services which, when used together, provide a comprehensive set of data management capabilities for trained healthcare professionals when working with infusion pumps. LifeShield™ Infusion Safety Software Suite does not remotely control or program the infusion pump or provide the ability to remotely manage pump alarms such as real-time monitoring, clearing and silencing alarms.
The LifeShield™ Drug Library Management (DLM) software is used by pharmacists to create approved drug libraries that can be downloaded by the infusion pumps. The latest software version introduces enhancements to its user interface and additional drug library settings for support of Plum Duo™ and Plum Solo™ pumps.
The LifeShield™ Clinical Dashboards & Reports (CDR) software is used by clinical administrators to view infusion or device-related information received from the infusion pumps via the LifeShield™ DFM. The information presented by the software does not create decisions or treatment pathways for patients. The latest version of the software improves the data presented for ongoing infusions and dashboards.
LifeShield™ Data Flow Management (DFM) software facilitates bidirectional communications between the infusion pump and hospital information systems (HIS); it routes pharmacy-validated orders to the connected pumps and infusion-related information to the HIS. The latest software version adds the ability to forward alarms to the HIS.
The provided FDA 510(k) clearance letter for the LifeShield™ Infusion Safety Software Suite (K242117) does not contain the specific details required to construct a table of acceptance criteria with reported device performance, nor does it detail a study that proves the device meets specific performance criteria in terms of accuracy or clinical outcomes.
This document primarily focuses on demonstrating substantial equivalence to a predicate device (K223606) based on non-clinical testing (design verification, design validation, cybersecurity, risk management, and human factors) and software modifications, rather than performance metrics related to clinical accuracy or diagnostic capabilities (which would be more common for AI/ML devices in image analysis, for example).
The submission explicitly states:
- "Summary of Clinical Testing: Not applicable. A clinical study is not required for this submission to support substantial equivalence."
Therefore, I cannot fulfill all parts of your request as the information is not present in the provided text. The device is a software suite for managing infusion pump data, not an AI/ML diagnostic or predictive tool that would have performance metrics like sensitivity, specificity, or AUC against a ground truth of disease.
However, I can extract information related to the types of testing performed and what they aimed to prove, which indirectly serve as acceptance criteria for this type of software device.
Analysis of Acceptance Criteria and Study Details based on the Provided Document
Given the nature of the device (infusion safety software suite) and the information provided in the 510(k) summary, the "acceptance criteria" here are aligned with software quality, safety, and functionality, rather than clinical performance metrics typically associated with AI/ML diagnostic tools.
The study proving the device meets acceptance criteria is a comprehensive set of non-clinical tests described below, rather than a single clinical trial.
1. Table of Acceptance Criteria and the Reported Device Performance
Since this is not an AI/ML diagnostic device with performance metrics like accuracy, sensitivity, or specificity against disease presence, the "acceptance criteria" relate to software design, functionality, safety, and usability. The document reports that all these tests "pass established acceptance criteria."
| Acceptance Criteria Category | Specific Criteria (Inferred from testing) | Reported Device Performance |
|---|---|---|
| Design Requirements | Adherence to design specifications; proper function of new features (e.g., UI enhancements, new drug library settings, alarm forwarding). | Passed: "Design verification tests pass established acceptance criteria, confirming the subject device meets design requirements." |
| Software Quality | Conformance to IEC 62304:2015 software development process; reliability. | Passed: "Software verification follows the software development process outlined in IEC 62304:2015." "Verification activities also include software verification, performance, reliability, compatibility, and interoperability tests." |
| Intended User Needs/Use Environment | Device meets all intended users' needs for its intended use and environment. | Passed: "Design validation tests pass established acceptance criteria, confirming the subject device meets all intended users' needs for the device's intended use and intended use environment." |
| Cybersecurity | Reasonable security against threats; protection of data. | Passed: "Cybersecurity evaluation and testing demonstrate that the software is reasonably secure." |
| Risk Management | Identification and mitigation of risks (ISO 14971:2019); effectiveness of mitigations. | Passed: "Risk management activities...concludes that the subject device is reasonably safe." "Mitigations...tested for correct implementation and effectiveness." |
| Human Factors/Usability | Effectiveness of user interface design for new features and critical tasks, per FDA guidance and standards (IEC 62366-1:2020, ANSI/AAMI HE75:2009/(R)2018). | Passed: "Human Factors study demonstrates the effectiveness of the user interface design for additional features and their associated critical tasks." |
2. Sample Size Used for the Test Set and Data Provenance
- Test Set Sample Size: The document does not specify a "sample size" in terms of patient data or clinical cases. For software verification and validation, the "test set" would consist of numerous test cases, simulated use scenarios, and functional tests. These numbers are not detailed in the summary.
- Data Provenance: Not applicable in the traditional sense of patient data. The testing is focused on software functionality, safety, and usability. There is no mention of retrospective or prospective data collection from patients or clinical settings for performance evaluation.
3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications
- Ground Truth Establishment: For this type of software, "ground truth" isn't established by clinical experts in the same way it would be for a diagnostic image. Instead, it's defined by design specifications, regulatory requirements, industry standards (e.g., IEC 62304, ISO 14971, IEC 62366-1), and best practices in software engineering and cybersecurity.
- Experts: The development team, regulatory affairs, quality assurance, and potentially third-party cybersecurity and human factors experts would define and verify these "ground truths" (i.e., whether the software behaves as intended and safely). The specific number or qualifications of these internal or external experts is not detailed in the provided K summary. The human factors study implies the use of representative users (trained healthcare professionals), but not necessarily "experts" defining a ground truth about a medical condition.
4. Adjudication Method for the Test Set
- Not applicable in the context of clinical expert adjudication of medical cases. Adjudication in software testing would involve bug reporting, resolution, and re-testing processes managed by the development and QA teams, not described here.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- No, an MRMC study was not done. The document explicitly states: "Summary of Clinical Testing: Not applicable. A clinical study is not required for this submission to support substantial equivalence." An MRMC study is a type of clinical study often done for diagnostic AI to compare human performance with and without AI assistance. This device is not a diagnostic AI.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- This concept is not directly applicable. The "LifeShield™ Infusion Safety Software Suite" is inherently a human-in-the-loop system designed to assist trained healthcare professionals. It manages data, facilitates communication, and supports drug library management but does not "decide" or "treat" on its own. Its "standalone" performance would pertain to its functional correctness in processing data and facilitating communication, which is covered by the general "design verification" and "performance" testing.
7. The Type of Ground Truth Used
- The "ground truth" for this software pertains to its functional correctness, adherence to specifications, safety requirements, cybersecurity posture, and usability standards. It is not based on clinical "outcomes data" or "pathology" in the medical sense, but rather on:
- Design Specifications: Whether the software performs its programmed functions as intended.
- Regulatory Standards: Compliance with relevant medical device software (IEC 62304), risk management (ISO 14971), and human factors (IEC 62366-1) standards.
- User Needs/Requirements: Whether the software meets the needs of its intended users in its intended environment.
8. The Sample Size for the Training Set
- Not applicable. This device is a software suite, not an AI/ML model that undergoes a "training phase" with a distinct "training set" of data to learn patterns. The software's capabilities are based on explicit programming and configuration, not machine learning from a dataset.
9. How the Ground Truth for the Training Set Was Established
- Not applicable, as there is no "training set" for this type of software.
In summary, the FDA 510(k) clearance for the LifeShield™ Infusion Safety Software Suite is based on a demonstration of substantial equivalence to a predicate device through extensive non-clinical software verification and validation, risk management, cybersecurity testing, and human factors analysis. The "acceptance criteria" are related to software quality, safety, and functionality, rather than clinical performance metrics against a medical ground truth or AI model training data.
Ask a specific question about this device
(30 days)
Rio™ Drug Reconstitution Transfer Device is indicated for single-use reconstituting or mixing of liquid or lyophilized drug in a vial, and the aseptic transfer of the reconstituted drug into the multi-port LifeCare IV container system (IV bag) for patient infusion administration.
The Rio™ Drug Reconstitution Transfer Device (Rio) is a single use, sterile, two-way, drug transfer device that is designed to connect an ICU Medical LifeCare IV container system (IV bag)(up to 500 mL) via the drug additive port, to a drug vial having either a 13mm or 20mm stopper closure for reconstituting or mixing and aseptic transfer of the drug from the vial into the solution of the IV bag. Once connected, Rio is not separated from the IV bag or vial and should be disposed of with the IV bag when administration is complete. Rio is intended to be used in a pharmacy setting or patient care area, by trained clinicians.
The Rio design consists of a needle-free port spike that connects to the compatible IV bag on one end, and a vial spike on the other end to connect a standard liquid or lyophilized/powdered drug vial. The bag spike and vial spike contain protective caps that maintain the sterility of the device until the caps are removed prior to use. Rio also includes a flow director (rotating handle) that will isolate the fluid between the vial and bag until manipulated by the pharmacist or clinician to allow two-way fluid transfer between the vial and bag.
The provided text is a 510(k) summary for the Rio™ Drug Reconstitution Transfer Device. This document focuses on demonstrating substantial equivalence to a predicate device (K192154), rather than providing detailed acceptance criteria and study results for de novo device performance validation.
Therefore, the document does not contain the specific information required to answer most of your questions about:
- A table of acceptance criteria and reported device performance.
- Sample sizes for test sets or their provenance.
- Number of experts, their qualifications, or adjudication methods for ground truth.
- MRMC or standalone comparative effectiveness studies.
- Types of ground truth used.
- Sample size or methods for establishing ground truth for training sets.
The document indicates that clinical data was not needed to support the substantial equivalence determination, which means there are no clinical studies of the type you're asking about (e.g., MRMC studies).
However, I can extract the information that is present regarding non-clinical testing and how it supports the device's conformance:
Summary of Non-Clinical Testing and Conformance:
The manufacturer states that "Non-clinical verification has been conducted to evaluate the safety, performance and functionality. The results of these test have demonstrated the overall safety of the subject device and ultimately supports a substantial equivalence device."
The document generally states that "Test results from the performance testing conducted demonstrate the subject device met all acceptance criteria requirements." However, it does not explicitly list the quantitative acceptance criteria or the specific numerical results obtained for each test.
Here's what can be gleaned about the non-clinical testing performed:
1. A table of acceptance criteria and the reported device performance:
- Acceptance Criteria: While specific numerical criteria are not provided, the testing aimed to meet various ISO and USP standards. For example, for particulates, the device had to "meet USP <788> requirements." For sterility, it had to meet a "SAL 1x 10-6."
- Reported Device Performance: The summary states that the device "met all acceptance criteria requirements" for the listed tests. No specific numerical performance values are given.
| Test Type | Relevant Standard (if mentioned) | General Performance Description (Acceptance Criteria Implicit) | Reported Device Performance |
|---|---|---|---|
| Functional Performance | Met all requirements | ||
| Positive pressure leak | ISO 22413 | Device maintains seal under positive air pressure | Met requirements |
| Negative pressure leak | ISO 8536-4 | Device maintains seal under negative pressure | Met requirements |
| Fluid flow | ISO 22413 | Fluid flows as intended through the device | Met requirements |
| Retention Testing | ISO 8536-4 | Device retains components as designed | Met requirements |
| Fragmentation/Coring | ISO 22413 | Device does not shed particulates or core | Met requirements |
| Vapor Barrier Test | Not Specified | Device maintains vapor barrier | Met requirements |
| Dye Leak Test | ISO 8871-5 | No dye leakage detected | Met requirements |
| Bag Insertion force | Not Specified | Insertion into IV bag requires acceptable force | Met requirements |
| Tamper Clip Performance | Met all requirements | ||
| Engagement force | Not Specified | Tamper clip engages with appropriate force | Met requirements |
| Removal force | Not Specified | Tamper clip removes with appropriate force | Met requirements |
| Particulate Testing | USP <788> | Particulate levels must be below specified USP limits | Meets USP <788> requirements |
| Microbial Ingress | Not Specified | Device prevents microbial entry | Met requirements |
| Biocompatibility | ISO 10993-1, FDA Guidance (Sept. 2023) | Device material is biologically compatible with human contact | Met requirements |
| Sterilization Validation | ISO 11137 | Device is effectively sterilized to SAL 1x10^-6 | Met requirements (SAL 1x10^-6) |
| Packaging | ISO 11607 | Packaging maintains sterility and integrity | Met requirements |
| Shelf life/Aging | Not Specified | Device maintains performance over its 5-year shelf life | Met requirements (5 years) |
2. Sample sized used for the test set and the data provenance:
- Not specified in the provided text.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable/Not specified. This is not a study involving expert readers or ground truth establishment in the context of diagnostic AI.
4. Adjudication method for the test set:
- Not applicable/Not specified.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No. The document explicitly states: "Clinical data was not needed to support a substantial equivalence determination." This is not an AI device or a diagnostic device.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- No. This is a physical, mechanical device, not an algorithm.
7. The type of ground truth used:
- For the non-clinical tests, the "ground truth" is established by adherence to recognized standards (ISO, USP) and engineering specifications for mechanical and material performance. For example, for particulate testing, the USP <788> limits define the "ground truth" for acceptable particulate levels. For sterility, the
SAL 1x10^-6is the "ground truth" for validated sterility.
8. The sample size for the training set:
- Not applicable. This is not an AI/machine learning device.
9. How the ground truth for the training set was established:
- Not applicable.
Ask a specific question about this device
(163 days)
The Cogent HMS is intended for patients for whom the monitoring of CCO and calculated hemodynamic parameters is indicated for diagnostic and prognostic evaluation by a clinician. The Cogent HMS is intended for use with ICU Medical pulmonary artery catheters and central venous oximetry catheters and with ICU Medical Cogent sensors. The Cogent HMS is intended to measure and calculate venous oxygen saturation in patients. PulseCO functionality is limited to adult patients.
The Cogent HMS is designed to compute and display cardiac and oximetry parameters relevant to patient care in the hospital acute care areas including Intensive Care Units and the Operating Room. Monitoring parameters include cardiac output and blood oxygen saturation levels, as well as other derived hemodynamic parameters. Measurements are obtained through the compatible ICU Medical pulmonary artery and central venous oximetry catheters, and ICU Medical CardioFlo™ sensors.
Input data for derived parameters may be keyed in by a clinician or may be obtained from a bedside monitor.
The Cogent HMS provides the following functions:
- monitors patient cardiac output continuously (CCO) using continuous thermodilution -(TdCO), and intermittently, using bolus thermodilution (Bolus CO);
- monitors continuous cardiac output (CCO) using pulse power analysis on an arterial pressure waveform;
- monitors venous oxygen saturation (SvO2) by measuring the reflectance spectrum of the blood; and
- provides a general-purpose interface to the analog input/output channels of other monitoring devices.
The Cogent HMS consists of:
- a base unit (patient interface module or PIM);
- a dedicated touch-screen display unit (user interface module or UIM) which allows for patient monitoring remotely (up to 50 feet); and
- associated cables
The PIM and UIM modules communicate with each other in docked, tethered (wired) or wireless mode.
The provided text describes a 510(k) premarket notification for the ICU Medical Cogent™ Hemodynamic Monitoring System (HMS). However, the document focuses on demonstrating substantial equivalence to a predicate device (K152006) primarily due to updates to the operating system (Windows 7 to Windows 10), software (version 1.1.8 to 1.4.0), and minor hardware changes. The submission primarily relies on non-clinical testing and verification, rather than a clinical study with detailed acceptance criteria and performance metrics for a novel algorithm.
Based on the provided text, the device is an updated version of an already cleared hemodynamic monitoring system. Therefore, the "acceptance criteria" discussed are largely related to ensuring the updated device performs equivalently to its predicate and meets relevant safety and performance standards. No specific "acceptance criteria" in terms of clinical performance metrics of an AI algorithm are explicitly stated, as the device is not presented as an AI-powered diagnostic algorithm with a performance threshold to meet.
Here's a breakdown of the information based on your request, as much as can be extracted from the provided text:
1. Table of Acceptance Criteria and Reported Device Performance
As the submission is for an updated version of an existing device, the "acceptance criteria" largely revolve around demonstrating equivalent performance to the predicate device and compliance with relevant standards. The document doesn't provide a table of precise quantitative acceptance criteria for clinical performance (e.g., sensitivity, specificity for a diagnostic algorithm) and corresponding reported performance of a novel AI component. Instead, it states that "the measurement performance of the subject device is equivalent to that of the predicate device."
| Acceptance Criteria Category (Implied from text) | Reported Device Performance (Implied from text) |
|---|---|
| Software Performance | Verified and validated successfully (per IEC 62304 and FDA guidance). Software considered equivalent to predicate. |
| System Bench Testing (Simulated Use) | Measurement performance of the subject device is equivalent to that of the predicate device. |
| Electrical Safety | Complies with requirements per IEC 60601-1. |
| Electromagnetic Compatibility (EMC) | Complies with requirements per IEC 60601-1-2. |
| Cybersecurity | System is effective in addressing cybersecurity threats. |
| Risk Management | Risk management activities incorporated in accordance with ISO 14971:2019 and tested for correct implementation and effectiveness. |
| Functional Performance & Intended Use | Meets functional performance and intended use claims as described in device labeling. No different questions of safety and effectiveness introduced. |
| Biocompatibility | Not applicable, as the device itself does not have direct patient contact. (Patient-contacting accessories are cleared separately). |
2. Sample Size Used for the Test Set and Data Provenance
The document explicitly states: "No new human factors, animal, and/or clinical studies were conducted as was determined not required to demonstrate device safety and effectiveness for the subject device."
Therefore, there is no "test set" in the context of clinical data with a sample size or provenance for this specific 510(k) submission. The testing performed was primarily non-clinical (bench testing, software V&V).
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications
Since no new clinical studies were conducted for this submission, there is no mention of experts establishing ground truth for a test set.
4. Adjudication Method for the Test Set
Not applicable, as no new clinical test set was used.
5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
Not applicable. The device is a hemodynamic monitoring system, not an AI-assisted diagnostic imaging or interpretation tool. The submission focuses on software and hardware updates to an existing monitoring device, not the evaluation of an AI algorithm's impact on human reader performance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The "Cogent System Software Validation" and various algorithm validations (PulseCO, Bolus CO, SO2, CCO) were performed, which could be considered standalone performance evaluations of the specific algorithms within the device. However, these are validations of established medical algorithms for physiological measurements, not novel AI algorithms in the common sense. The text implies these were bench validations, not clinical standalone performance studies.
7. The Type of Ground Truth Used
For the algorithm validations (e.g., PulseCO, Bolus CO, SO2, CCO), the ground truth was established through "in vitro validation" using flow simulators or electronically generated data. This suggests a controlled laboratory environment where the "true" physiological values could be precisely set or simulated.
8. The Sample Size for the Training Set
Not applicable. The document discusses updates to an existing device and its algorithms, not the training of a new AI algorithm that would require a distinct training set. The algorithms mentioned (e.g., thermodilution, pulse power analysis) are based on established physiological principles and are not typically "trained" in the machine learning sense with large datasets.
9. How the Ground Truth for the Training Set was Established
Not applicable, as there is no mention of a training set for a novel AI algorithm.
Ask a specific question about this device
(265 days)
The Plum Duo Infusion System is intended for parenteral fluids and medications through clinically acceptable routes (limited to intravenous, intra-arterial, and subcutaneous therapies).
The Plum Duo Infusion System is intended for use in clinical environments in the hospital environment and other outpatient healthcare facilities by licensed healthcare professionals are trained in the use of the infusion pump and the administration of therapies consistent with the intended use.
The Plum Duo Infusion System is intended for adult, pediatric (including infants and children), and neonatal patient populations.
The Plum Duo™ Infusion System is the next generation of the Plum™ family of devices that is based on the fundamental technology of the Plum 360™ Infusion System cleared under K161469. The pump design incorporates state-of-the-art features such as dual channel functionality, touch screen display, and lithium iron phosphate battery technology. The pump uses the same volumetric piston type technology with a plunger stepper motor to deliver fluids to a patient as the predicate Plum 360™ Infusion System.
Plum Duo™ Infusion System is a large volume pump (LVP) with two independent pump channels that can deliver fluid to a patient on up to 4 lines and is designed so that it is possible to use one channel only. In addition, although the channels can operate independently, patient parameters can be shared across the channels to aid in the speed of programming. Each channel accepts a cassette that is part of a PlumSet™ administration set and can connect to a primary and/or secondary container.
The provided document is a 510(k) Summary for the Plum Duo™ Infusion System, which is an infusion pump. It demonstrates substantial equivalence to a predicate device. This type of regulatory submission focuses on engineering and performance validation rather than clinical studies involving patient data or human interpretation of medical images.
Therefore, the document does not contain the information requested regarding acceptance criteria and study details for an AI/ML-enabled medical device that typically involves:
- A table of acceptance criteria and reported device performance (in terms of sensitivity, specificity, AUC, etc., which are common for diagnostic or AI algorithms).
- Sample size for a test set based on patient data provenance.
- Number of experts and their qualifications for establishing ground truth.
- Adjudication methods for test sets.
- Multi-Reader Multi-Case (MRMC) comparative effectiveness studies.
- Standalone (algorithm-only) performance.
- Type of ground truth (pathology, outcomes data, expert consensus).
- Sample size for the training set and how its ground truth was established.
This document describes a more traditional medical device submission, focusing on the mechanical, electrical, and software safety and performance of an infusion pump. The "Summary of Non-Clinical Testing" section lists various engineering tests (e.g., flow rate accuracy, air in line performance, alarm testing, electrical safety, reliability testing, human factors validation), which are typical for infusion pumps to ensure their safe and effective operation, but these do not align with the criteria for evaluating an AI/ML model for diagnostic or predictive purposes.
The statement "Clinical evaluation is not required for this submission to support substantial equivalence" further confirms that the type of studies and data provenance you are asking about (which often involve clinical data and human interpretation) were not part of this specific submission.
Ask a specific question about this device
Page 1 of 6