Search Filters

Search Results

Found 4 results

510(k) Data Aggregation

    K Number
    K193189

    Validate with FDA (Live)

    Date Cleared
    2020-05-06

    (169 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tag of the RFLS is intended for percutaneous placement in the breast to mark (>30 days) a lesion intended for surgical removal. Using image guidance (such as ultrasound or radiography) or aided by non-imaging guidance (RFLS), the RFID Tag is located and surgically removed with the target tissue. The RFLS is intended only for the non-imaging detection and localization of the Tag that has been implanted in a lesion intended for surgical removal.

    Device Description

    The proposed RFID Localization System is a marker-with-detector localization device that employs miniature RFID tags as markers and a hand-held reader that can measure distance to the tag. The RFLS is comprised of a Tag, Tag Applicator/Tag Applicator S, LOCalizer Reader) and LOCalizer Surgical Probe (Surgical Probe). The Tag, when used in conjunction with the Reader and Surgical Probe, can be used as a guide for the surgeon to refer to in the excision of tissue. The RFLS is a prescription device meant only for use by trained professionals.

    AI/ML Overview

    Since the provided text is a 510(k) summary for a medical device (RFID Localization System), and not a study report for a diagnostic AI/ML device, much of the requested information regarding acceptance criteria for AI/ML performance, sample sizes for AI test/training sets, expert ground truth establishment, and MRMC studies is not directly applicable or present in this document.

    The document discusses performance data related to the physical device's function, safety, and compatibility, rather than the performance of an AI model in a diagnostic context. The "RFID Localization System" is a marker-with-detector device, not an AI-powered diagnostic system.

    However, I can extract the acceptance criteria and performance data as described for this physical device, along with other relevant information that is present.

    Here's the closest possible interpretation of your request based on the provided text, focusing on the device validation rather than AI performance:


    Device: RFID Localization System (RFLS)
    Intended Use: Percutaneous placement in the breast to mark a lesion intended for surgical removal. Aids in non-imaging detection and localization of the implanted Tag for surgical removal.

    1. Table of Acceptance Criteria and Reported Device Performance

    For a physical device like the RFLS, "acceptance criteria" relate more to meeting design specifications and safety/performance standards. The document states that the RFLS met the applicable design and performance requirements. The "performance" is demonstrated by passing the various tests listed, affirming its functionality and safety.

    Feature/Test TypeAcceptance Criteria (Implied: Met applicable design/performance requirements & standards)Reported Device Performance
    RFLS System Design VerificationDevice meets product specifications defined in design requirement documentation."RFID Localization System verification testing was conducted to demonstrate that the RFID Localization System (RFLS) meets product specifications as defined in the design requirement documentation." "Testing was conducted to measure the effect of multiple tags on tag-to-probe distance readings." (Implied successful demonstration, as it supports substantial equivalence).
    Tissue Marker Migration EvalRFID Tag does not migrate due to MRI-induced forces or simple body movement beyond acceptable limits."A migration evaluation was performed to assess migration of the RFID Tag due to MRI-induced forces as well as simple migration resultant from body movement." "Migration has previously been established through evaluation for the RFID Localization System (K163667 and K181692)." (Implied successful demonstration, relying on previous clearances).
    Usability Verification & ValidationDevice design meets usability requirements."Usability testing was conducted to demonstrate that the RFID Localization System (RFLS) design meets Usability requirements." "Usability has previously been established through evaluation for the RFID Localization System (K181692 and K190932)." (Implied successful demonstration, relying on previous clearances).
    Electrosurgery & MRI CompatibilityRFID Tag function is not adversely affected by electrosurgery. Device is safe and compatible in the MR environment."Testing was conducted to assess electromagnetic interference and RFID Tag functionality after direct exposure to electrical current from an electrosurgery instrument." "Electrosurgical compatibility has previously been established through testing for the RFID Localization System (K163667)." "Testing was conducted to evaluate the interaction (safety and compatibility) of the RFID Tag in the magnetic resonance (MR) environment." "MRI compatibility has previously been established through testing for the RFID Localization System (K163667 and K190932)." (Implied successful demonstration, relying on previous clearances).
    Packaging ValidationSterile components meet standards for packaging integrity and sterility."Package Qualification Testing for sterile components was completed in accordance with the following standards: ISO 11607-1, ASTM D4169-09, ASTM F1886/F1886M-16, ASTM F2096-11, ASTM F88/F88M-15, ASTM F1929-98 (2004)." "Ship testing was completed in accordance with ASTM D4149 and ISTA 2A." "Packaging qualification has previously been established through testing for the RFID Localization System (K163667)." "Shipping evaluations has previously been established through testing for the RFID Localization System (K190932)." (Implied successful demonstration, relying on previous clearances).
    Sterilization ValidationSterile components meet sterilization requirements."Sterilization Validation for the sterile components was completed in accordance with the following standards, as appropriate: AAMI ANSI ISO 11137-1, AAMI ANSI ISO 11137-2, AAMI ANSI ISO 10993-7." "Sterilization parameters have previously been validated through testing for the RFID Localization System (K163667, K181692 and K190932)." (Implied successful demonstration, relying on previous clearances).
    Stability TestingDevice maintains performance and safety over its intended shelf-life."Shelf-life evaluations were conducted in accordance the Food & Drug Administrations; "Shelf Life of Medical Devices" and the ASTM Standard "F1980, Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices"." "Device shelf-life has previously been established through testing for the RFID Localization System (K163667, K181692 and K190932)." (Implied successful demonstration, relying on previous clearances).
    BiocompatibilityPatient-contacting materials are biocompatible per ISO 10993-1 for the expected contact category, type, and duration."Testing completed met the requirements of the respective test methods per the International Standard ISO 10993-1 'Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing Within a Risk Management Process' as appropriate for the expected contact category, type and duration." "Biocompatibility has previously been established through biocompatibility testing for the RFID Localization System patient contacting materials (K190932 and DEN040007/K033440)." (Implied successful demonstration, relying on previous clearances).
    Electrical Safety & EMCDevice meets electrical safety and electromagnetic compatibility standards (IEC 60601-1, IEC 60601-1-2)."Electrical safety and EMC safety testing was performed to, and passed, the following standards: IEC 60601-1, IEC 60601-1-2." (Explicitly states "passed the following standards").
    Software V&V TestingSoftware meets FDA guidance for medical device software and cybersecurity."Software verification and validation testing for the proposed RFID Localization System (RFLS) was conducted and documentation provided as recommended by FDA's Guidance for Industry and FDA Staff, 'Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices' and 'Guidance for the Content of Premarket Submissions for Management of Cybersecurity in Medical Devices'." (Implies successful completion and adherence to guidance, relying on previous clearances K181692 and K190922).

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify a "test set sample size" in the context of an AI/ML model. Instead, it refers to various engineering and biological validation tests. For many tests, it explicitly states that the performance was "previously established" through prior 510(k) clearances (e.g., K190932, K181692, K163667, DEN040007/K033440). This indicates that the current submission relies on historical test data and validations, with some additional testing (e.g., multiple tags, updated packaging/shipping in K190932, and the explicitly mentioned electrosurgery and MRI compatibility for K163667, and the current submission's specific software and electrical safety/EMC testing).

    • Data Provenance: The document doesn't specify the country of origin for the underlying test data, but it refers to FDA-cleared predicate and reference devices, implying compliance with US regulatory standards. All validations appear to be retrospective or performed as part of device development and previous submissions.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications:

    This information is not applicable. This is a physical medical device, not an AI/ML diagnostic algorithm requiring expert "ground truth" for interpretations of medical images or data. The "ground truth" for the device's performance would be established by objective measurements against engineering specifications and validated medical standards.

    4. Adjudication Method for the Test Set:

    Not applicable. This concept applies to human reader studies often used for AI/ML validation, not the direct performance testing of a physical medical device.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    Not applicable. This type of study is for evaluating observer performance with or without AI assistance, which is not the subject of this 510(k). The device is a physical marker and localization system, not a diagnostic AI.

    6. Standalone (Algorithm Only Without Human-in-the-Loop) Performance:

    Not applicable. This device is not an algorithm; it's a marker-with-detector system that a human surgeon uses. Its performance is inherent in its physical properties and functionality.

    7. Type of Ground Truth Used:

    The "ground truth" for this device's validation is based on:

    • Engineering specifications and design requirements.
    • Objective measurements (e.g., distance readings, physical stability, electrical safety parameters).
    • Compliance with established international and ASTM standards (e.g., ISO 11607-1, ASTM D4169-09, ISO 10993-1, IEC 60601-1).
    • Biocompatibility assessments of materials.
    • Validation of sterilization processes.

    8. Sample Size for the Training Set:

    Not applicable. This device does not use an AI/ML model that requires training data in the conventional sense.

    9. How the Ground Truth for the Training Set was Established:

    Not applicable, as there is no AI/ML training set. The "ground truth" for the device's design and manufacturing is established through scientific principles, engineering validated methods, and adherence to medical device regulatory standards and best practices.

    Ask a Question

    Ask a specific question about this device

    K Number
    K190932

    Validate with FDA (Live)

    Date Cleared
    2019-09-13

    (156 days)

    Product Code
    Regulation Number
    878.4300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tag of the RFLS is intended for percutaneous placement in the breast to mark (>30 days) a lesion intended for surgical removal. Using image guidance (such as ultrasound or radiography) or aided by non-imaging guidance (RFLS), the RFID Tag is located and surgically removed with the target tissue. The RFLS is intended only for the non-imaging detection and localization of the Tag that has been implanted in a lesion intended for surgical removal.

    Device Description

    The Tag, Tag Applicator, Tag Applicator S, LOCalizer Reader, and LOCalizer Surgical Probe are components of the Health Beacons RFID Localization System (RFLS). The proposed device is a marker-with-detector localization device that employs miniature RFID tags as markers and a handheld reader that can measure distance to the Tag. The Tag, when used in conjunction with the Reader and Surgical Probe, can be used as a guide for the surgeon during the excision of tissue. The RFLS is a prescription device meant only for use by trained professionals, specifically breast surgeons and diagnostic radiologists.

    AI/ML Overview

    This document describes the 510(k) summary for the Health Beacons, Inc. RFID Localization System (RFLS). However, based on the provided text, there is no detailed information about acceptance criteria and a specific study proving the device meets those criteria, especially in the context of AI assistance or human reader performance improvement.

    The document is a 510(k) summary for a medical device (RFID Localization System) intended to mark and locate lesions in the breast for surgical removal. While it mentions performance testing was conducted to support substantial equivalence, it does not provide the specific acceptance criteria or the results from those tests in a format that would allow filling out the requested table or answering many of the follow-up questions.

    The device itself is an RFID localization system, which uses miniature RFID tags as markers and a handheld reader to measure distance to the tag. This is a physical localization device, not an AI/software-based diagnostic tool that would typically involve acceptance criteria related to sensitivity, specificity, or human reader improvement with AI assistance.

    Therefore, many of the requested details, particularly those pertaining to AI/ML algorithms (e.g., acceptance criteria for diagnostic accuracy, standalone algorithm performance, MRMC studies, training/test set details, ground truth for AI models), are not applicable or not present in this specific 510(k) summary.

    Here's what can be extracted and what cannot, based on the provided text:

    1. A table of acceptance criteria and the reported device performance:

    • Not explicitly provided in the document. The document states that "All verification and validation activities identified as necessary were performed... and results demonstrate that predetermined acceptance criteria were met." However, the specific criteria and the numerical performance results are not tabulated or detailed.
    • The "performance testing" subsections (VIII) list categories of tests but not acceptance criteria or outcomes. These tests include:
      • Magnetic field emission testing per IEC 60601-1-2:2014
      • Delivery testing
      • Deployment Force testing
      • Needle Penetration Force testing
      • Usability testing

    2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective):

    • Not provided. The document states "performance testing was provided," but offers no details on sample size, study design (retrospective/prospective), or data provenance.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable/Not provided. This is relevant for diagnostic AI/ML devices where human expert consensus often establishes ground truth. This device is a physical localization system, not a diagnostic imaging AI.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable/Not provided. Same reasoning as point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No. This device is a physical localization system; it's not an AI assisting human readers with diagnostic tasks.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • Not applicable/Not provided. The device is a "marker-with-detector localization device" that involves a Tag and a Reader/Probe used by a surgeon. There isn't an "algorithm only" component in the sense of a standalone diagnostic AI.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • Not explicitly stated for the listed performance tests. For a physical device, ground truth for performance tests would likely involve physical measurements (e.g., actual vs. measured distance, force applied, magnetic field readings).

    8. The sample size for the training set:

    • Not applicable/Not provided. This device is not an AI/ML algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable/Not provided. Same reasoning as point 8.

    In summary of the provided text:

    This FDA 510(k) summary describes a physical medical device (RFID Localization System), not an AI/ML diagnostic software. The document asserts that performance testing was conducted and met predetermined acceptance criteria, which supported the substantial equivalence determination to a predicate device. However, it does not provide the specific numerical acceptance criteria or the detailed results of those performance tests. Information regarding sample sizes, data provenance, expert involvement for ground truth, or MRMC studies (which are highly relevant for AI/ML diagnostic devices) is absent as these concepts are not directly applicable to the type of device described.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tag of the RFLS is intended for percutaneous placement in the breast to mark (>30 days) a lesion intended for surgical removal. Using image guidance (such as ultrasound or radiography) or aided by non-imaging guidance (RFLS), the RFID Tag is located and surgically removed with the target tissue. The RFLS is intended only for the non-imaging detection and localization of the Tag that has been implanted in a lesion intended for surgical removal.

    Device Description

    The Tag, Tag Applicator, LOCalizer Reader, and LOCalizer Surgical Probe are components of the Health Beacons RFID Localization System (RFLS). The proposed device is a marker-with-detector localization device that employs miniature RFID tags as markers and a hand-held reader that can measure distance to the Tag, when used in conjunction with the Reader and Surgical Probe, can be used as a guide for the surgeon during the excision of tissue. The RFLS is a prescription device meant only for use by trained professionals, specifically breast surgeons and diagnostic radiologists.

    AI/ML Overview

    This document is a 510(k) Summary for the RFID Localization System (RFLS). It describes a medical device and provides information on its performance data, but it does not contain the results of a clinical study with acceptance criteria and a detailed comparison against it. The document primarily focuses on technical performance testing and equivalence to predicate devices, rather than a clinical effectiveness study measuring human reader improvement or standalone algorithm performance against a clinical ground truth.

    Therefore, many of the requested fields cannot be filled from the provided text.

    Here's a breakdown of what can be extracted and what is missing:

    1. A table of acceptance criteria and the reported device performance:

    The document lists "Performance Testing" categories, but it does not provide specific acceptance criteria or quantitative performance results for clinical endpoints. Instead, it states that tests were performed to demonstrate that the RFLS "met the applicable design and performance requirements."

    Acceptance CriteriaReported Device Performance
    (Not explicitly stated in terms of clinical performance or specific thresholds for accuracy, sensitivity, specificity, etc.)- RFLS System Verification performed.- Reader/Probe Design Verification performed.- Localizer System and Component Accuracy, Repeatability Evaluation performed.- Test Media Evaluation performed.- Reader and Surgical Probe Operating Environment Evaluation performed.- Tissue Marker Migration Evaluation performed.- Usability Verification and Validation performed.- Electrosurgical tool compatibility performed.- Packaging Validation performed.- Sterilization Validation performed.
    Biocompatibility requirements per ISO 10993-1Met requirements; previously established for reference device (K163667) using identical patient-contacting materials.
    Electrical Safety and EMC per IEC 60601-1 and IEC 60601-1-2Passed testing for both standards.
    Software Verification and Validation testing per FDA GuidancesConducted and documentation provided.

    2. Sample size used for the test set and the data provenance: Not mentioned. The performance data refers to device-level technical testing rather than a clinical test set.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable, as this document does not describe a study involving expert-established ground truth for a clinical test set.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This document does not describe an MRMC study or AI assistance. The device is a localization system, not an AI diagnostic tool for image interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable in the context of an "algorithm only" performance. The device is a "marker-with-detector localization device" used by trained professionals. The performance tests mentioned are for the physical device system.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable, as this document does not describe a study involving clinical ground truth. The "ground truth" for the device's technical performance would be engineering specifications and measurements.

    8. The sample size for the training set: Not applicable. This is not an AI/ML device that requires a training set in the conventional sense.

    9. How the ground truth for the training set was established: Not applicable.

    Summary of what the document does provide:

    • Device Name: RFID Localization System (RFLS)
    • Intended Use: Percutaneous placement of a Tag in the breast to mark a lesion for surgical removal. The RFLS (hand-held reader and surgical probe) assists in non-imaging detection and localization of the Tag.
    • Predicate Device: Cianna Medical SAVI Scout Reflector and SAVI Scout System (K171767)
    • Comparison: The RFLS has the "same indications for use, principles of operation and similar fundamental scientific technology and materials" as the predicate.
    • Performance Data Types: System verification, design verification, accuracy/repeatability, test media evaluation, operating environment, tissue marker migration, usability, electrosurgical tool compatibility, packaging, sterilization, biocompatibility, electrical safety, and software V&V.
    • Biocompatibility: Established based on an identical reference device (K163667) and adherence to ISO 10993-1.
    • Electrical Safety & EMC: Passed IEC 60601-1 and IEC 60601-1-2 standards.
    • Software V&V: Conducted per FDA guidances.

    The document concludes that the RFLS is substantially equivalent to its predicate device based on these performance and technical characteristic comparisons.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Tag of the RFLS is intended for percutaneous placement in the breast to temporarily (<30 days) mark a lesion intended for surgical removal. Using image guidance (such as ultrasound or radiography) or aided by non-imaging guidance (RFLS), the RFID Tag is located and surgically removed with the target tissue.

    The RFLS is intended only for the non-imaging detection and localization of the Tag that has been implanted in a lesion intended for surgical removal.

    Device Description

    The RFID Localization System (RFLS) is a marker-with-detector localization device that employs miniature RFID tags as markers and a hand-held reader that can measure distance to the tag. The RFLS is comprised of a Tag, Tag Applicator, Reader, and Surgical Probe. The Tag, when used in conjunction with the Reader and Surgical Probe, can be used as a guide for the surgeon during the excision of tissue. The RFLS is a prescription device meant only for use by trained professionals, specifically breast surgeons and diagnostic radiologists.

    AI/ML Overview

    The provided text details the 510(k) summary for the RFID Localization System (RFLS). While it outlines various performance tests conducted, it does not provide a specific table of acceptance criteria with corresponding reported device performance, nor does it detail a comparative effectiveness study involving human readers (MRMC). The information focuses on demonstrating substantial equivalence to a predicate device through non-clinical testing.

    Based on the provided text, here's an attempt to answer the questions, highlighting what information is available and what is not:


    Acceptance Criteria and Study for RFID Localization System (RFLS)

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not provide a formal table with specific numerical acceptance criteria and corresponding reported performance for each criterion. It broadly states that "performance testing was conducted to evaluate and characterize the performance of the RFLS."

    However, one specific performance metric is mentioned under "Localization Accuracy" in the comparison table:

    Acceptance Criterion (Implicit)Reported Device Performance
    Localization Accuracy±7 mm

    Other performance aspects mentioned (without specific numerical criteria or reported values):

    • RFLS System Verification
    • Reader/Probe Design Verification
    • Localizer System and Component Accuracy, Repeatability Evaluation
    • Test Media Evaluation
    • Reader and Surgical Probe Operating Environment Evaluation
    • Tissue Marker Migration Evaluation
    • Usability Verification and Validation
    • Electrosurgical tool compatibility
    • EMC/Safety Testing
    • Biocompatibility
    • Magnetic Resonance Testing
    • Packaging Validation
    • Sterilization Validation

    The document states that these tests "demonstrates the safety of the system for its intended use and confirms the performance of the system for use in intra-operative surgical guidance. Specifically, the performance testing demonstrates the safe use of the system to facilitate intra-operative guided surgical procedures by detecting the presence of the Tag by providing an audible tone which increases pitch and volume with decreased distance, a visible bar indicator, and a measurement readout."

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify the sample sizes used for the various performance tests (e.g., number of RFLS tags tested, number of measurement repetitions for accuracy, etc.).

    It also does not detail the data provenance in terms of country of origin or whether the studies were retrospective or prospective. Given that this is a 510(k) summary for a medical device cleared by the FDA (USA), it's highly probable the testing was conducted in the USA, but this is not explicitly stated. The tests are described as "non-clinical testing," implying they were not conducted on human patients.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    The document does not mention the use of experts or the establishment of ground truth by experts for the described performance tests. The tests appear to be primarily engineering and bench-top evaluations (e.g., accuracy, repeatability, EMC, sterilization validation) rather than studies requiring expert clinical judgment for ground truth.

    4. Adjudication Method for the Test Set

    Not applicable, as expert involvement and clinical ground truth establishment are not described for the non-clinical performance tests.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done

    No, the document does not indicate that a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was performed to assess how human readers improve with AI vs. without AI assistance. The RFLS is a localization device, not an AI-assisted diagnostic tool that would typically involve comparative reader studies for diagnostic accuracy.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was Done

    The performance testing appears to be primarily focused on the standalone performance of the device's components and system operation (e.g., localization accuracy of ±7mm, detection of the tag, providing audible tone and visible indicators). The device's function is explicit: "The RFLS is intended only for the non-imaging detection and localization of the Tag that has been implanted in a lesion intended for surgical removal." This implies an "algorithm only" type of measurement in terms of its core function, but it's part of a human-guided surgical procedure.

    7. The Type of Ground Truth Used

    For "Localization Accuracy (±7 mm)," the ground truth would likely be established by a highly precise physical measurement system (e.g., an optical tracking system, a precisely calibrated measurement jig) during bench testing, rather than expert consensus, pathology, or outcomes data, as it's a direct physical measurement. For other tests like "Sterilization Validation" or "Biocompatibility," the ground truth would be adherence to established standards and test protocols.

    8. The Sample Size for the Training Set

    The document does not mention a training set. The RFLS is described as a marker-with-detector localization device, relying on RFID technology for detection and distance measurement. This does not suggest the use of machine learning or algorithms that require a distinct "training set" in the conventional sense (e.g., for image classification or prediction). The device's function appears to be based on direct physical principles of radiofrequency wave detection rather than learned patterns from a training dataset.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as a training set is not mentioned.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1