Search Filters

Search Results

Found 2 results

510(k) Data Aggregation

    K Number
    K213092

    Validate with FDA (Live)

    Date Cleared
    2021-12-15

    (82 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for the management of minor scrapes, minor bruises, minor abrasions, minor lacerations, and minor burns.

    Device Description

    Cellusheet Dressing is a soft, sterile, disposable, absorbent & conformable single use wound dressing. It is hydrophilic dressing composed of fibrous bovine collagen and a non-adherent backing layer. The dressing absorbs wound drainage to cover the topical wounds to provide moist wound environment. The Cellusheet Dressing will be available in 2"X2" and 8"X12" sizes. Cellusheet Dressing is made up of collagen which is the most abundant protein in the human body and to cover the topical wounds to provide moist wound environment.

    Cellufil Particles is a bovine (cattle) collagen wound dressing that is sterile, off-white powder that absorbs wound drainages. The dressing is an off-white, colored powder made up of collagen cover the topical wounds to provide moist wound environment. Cellufil Particles consists of spherical hydrophilic particles 0.1mm to 0.3mm in diameter.

    AI/ML Overview

    This document is a 510(k) summary for medical devices, specifically Cellusheet Dressing and Cellufil Particles, seeking Over-The-Counter (OTC) clearance. It compares these devices to their previously cleared Prescription Use (Rx) predicates, SkinTemp II Dressing and Medifil II Particles, respectively. The core of the submission argues for substantial equivalence primarily based on the devices being identical to their predicates in all aspects except the change from Rx to OTC indication for use.

    Here's a breakdown of the requested information, though much of it regarding acceptance criteria and separate performance studies is not applicable or not explicitly detailed in this type of submission, which focuses on substantial equivalence rather than new performance data for this specific submission.


    1. A table of acceptance criteria and the reported device performance

    This 510(k) submission is for a "Special 510(k)" which modifies the Indications for Use from Prescription (Rx) to Over-The-Counter (OTC) for two previously cleared devices. Therefore, the acceptance criteria and device performance reported are not against new performance standards, but rather demonstrate that the device remains substantially equivalent to its predicates, which already met their respective performance criteria. The document explicitly states: "The subject devices are the same as the predicate devices in all technological characteristics, geometric specifications, packaging, sterilization, manufacturing process and raw materials used. The subject devices do not raise any different questions of safety and effectiveness with regards to the technological characteristics."

    Essentially, the acceptance criterion for this submission is Substantial Equivalence to the predicate devices, established by demonstrating no changes in design, materials, manufacturing, or other technological characteristics that would raise new questions of safety or effectiveness for the proposed OTC use.

    Acceptance Criteria (for Substantial Equivalence)Reported Device Performance (vs. Predicate)
    Intended Use Comparison: Similar intended use.Same: "Management of wounds" for both Cellusheet Dressing/Cellufil Particles and SkinTemp II Dressing/Medifil II Particles.
    Indications for Use Comparison: Proposed indications for use do not raise new questions of safety or effectiveness.Different, but argued as equivalent: Proposed OTC indications are "Intended for management of minor cuts, minor scrapes, minor bruises, minor abrasions, minor lacerations, and minor burns." Predicate Rx indications are broader: "The device is intended for the management of burns, sores, blisters, scrapes, ulcers and other wounds." The document argues: "The difference in indications from prescription to OTC do not raise different questions of safety or effectiveness. The use of these products for minor wounds is intuitive. Similar products already exist in the OTC market (see K081724)."
    Technological Characteristics Comparison: Similar design, materials, sterilization, shelf life, configurations, etc.Same: - Sterilization: E-beam. Product supplied STERILE.- Storage Conditions: Store in a cool and dry place.- Design: Soft, sterile, disposable, absorbent & conformable single-use wound dressing with non-adherent backing (Cellusheet); Bovine collagen wound dressing that is sterile, off-white powder (Cellufil).- Primary Constituent Materials: Bovine Collagen, Demineralized Water (Cellusheet only), Buffering agent / Chemical Z.- Usage Type: Single Patient Use.- Shelf Life: 5 years.- Configurations/Particle Size: 2"X2", 3"X4", 8"X12" (Cellusheet); 0.1mm-0.3mm (Cellufil).- Packaging: 5 sheets/pouches per box.- Device Class/Product Code: Unclassified, KGN.
    Biocompatibility: No new biocompatibility concerns.Same: "All testing performed on predicate will be applicable since there is no change in the product specifications. All testing was performed as per FDA biocompatibility guidance on ISO 10993."
    Product Integrity/Stability: No new concerns over shelf-life.Same: "the product integrity testing over the shelf-life period will also be applicable to the subject device in this case."

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    This submission does not describe a new "test set" or new performance data for the devices. The basis for substantial equivalence is that the devices are identical to previously cleared predicate devices except for the change in marketing from Rx to OTC. Therefore, no new clinical or non-clinical studies specifically for a "test set" for this submission were conducted or reported here. The document states: "The subject devices are the same as the predicate devices... All the testing was submitted in K122325 for SkinTemp II Dressings i.e., the predicate for Device 1 and K910944 for Medifil II Particles i.e., the predicate for Device 2."


    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. As this is a Special 510(k) based on substantial equivalence to existing predicate devices, no new "test set" requiring expert-established ground truth for device performance is described.


    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. No new test set or clinical study requiring adjudication is described in this submission.


    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This document pertains to wound dressings and particles, not AI software or diagnostic imaging devices that would typically involve MRMC studies.


    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This does not involve an algorithm.


    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    Not applicable for this Special 510(k) submission. The "ground truth" for the predicate devices, which established their safety and effectiveness for their original Rx indications, would have relied on standard testing and potentially clinical data appropriate for wound care products at the time of their original clearance (K122325, K910944, K925545). This information is not detailed in the current submission.


    8. The sample size for the training set

    Not applicable. There is no mention of a "training set" as this is not an AI/machine learning device.


    9. How the ground truth for the training set was established

    Not applicable. There is no mention of a "training set."

    Ask a Question

    Ask a specific question about this device

    K Number
    K122325

    Validate with FDA (Live)

    Device Name
    SKINTEMP II
    Date Cleared
    2012-10-26

    (86 days)

    Product Code
    Regulation Number
    N/A
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The device is intended for the management of burns, sores, blisters, scrapes, ulcers, and other wounds.

    Device Description

    Skin Temp. H. Dressing is a sterile, disposable, single use wound dressing. It is a hydrocolloid and hydrophillic dressing composed of fibrous type I bovine collagen. The device is to be provided in hydrated as well as dry form. The device is to be used for the management of burns, blisters, sores, scrapes, ulcers and other wounds. The Skin.Temp. II. Dressing will be available in 2"X2", 3"X4" and 8"X12" sizes and additional sizes may introduced at a later time.

    AI/ML Overview

    This submission details the SkinTemp II Dressing, a collagen-based wound dressing, as substantially equivalent to the predicate device, BioCore SkinTemp sheet. The core of the submission revolves around demonstrating this substantial equivalence through a comparison of technological characteristics and non-clinical testing.

    Here's an analysis of the provided information concerning acceptance criteria and the supporting study:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly present a formal table of "acceptance criteria" with quantitative performance metrics for the SkinTemp II Dressing. Instead, it relies on demonstrating substantial equivalence to a predicate device. The acceptance is implied by meeting or being analogous to the predicate device's characteristics and performance.

    However, based on the non-clinical testing and comparison to the predicate, we can infer some "performance criteria" that the SkinTemp II Dressing met:

    Accepted Performance Criteria CategoryTarget (Analogous to Predicate)SkinTemp II Dressing Performance
    Intended Use/IndicationsManagement of burns, sores, blisters, scrapes, ulcers, and other wounds.Identical: "The HBS SkinTemp II Dressing is indicated for the management of burns, sores, blisters, scrapes, ulcers, and other wounds." (Page 2, Table 7.1)
    Principles of UseCleanse wound, apply medication if needed, apply dressing, cover with non-adherent dressing, change as required.Identical: "Cleanse the wound. Apply medication if needed. Apply dressing and cover with non-adherent dressing. Change as required." (Page 2, Table 7.1)
    Scientific Technology100% type I bovine collagen as primary constituent.Identical: "The SkinTemp II Dressing uses 100% type I bovine collagen as its primary constituent." (Page 2, Table 7.1)
    Configuration / System ComponentsSterile, disposable, single use, collagen-based wound dressing with non-adherent backing. Hydrocolloid, absorbent, gel formation.Identical: "Sterile, disposable, single use, collagen based wound dressing with non adherent backing. Further properties include: a) Hydrocolloid b) Absorbent Dressing c) Gel formation" (Page 2, Table 7.1)
    Sterility MethodElectron Beam Sterilization.Identical: "Each dressing is sterilized using Electron Beam Sterilization." (Page 2, Table 7.1) Validated per AAMI TIR-27 and ISO 11137-1:2006. (Page 4-5)
    MaterialsType I Bovine Collagen, Distilled Water, Buffering Agent (Chemical Z).Identical: "Type I Bovine Collagen, Distilled Water, Buffering agent / Chemical Z (proprietary name)" (Page 2, Table 7.1). The inclusion of Chemical Z was the specific reason for this new 510(k) submission to formally document an ingredient that was previously used but not formally documented in the predicate's 510(k). (Page 5)
    BiocompatibilityMeet all biocompatibility testing requirements.Confirmed: "Testing has confirmed that SkinTemp II Dressing meets all biocompatibility testing requirements stated by FDA regulations and ISO 10993." (Page 4) Performed by North American Science Association, Inc. (NAmSA).
    Sterilization AssuranceValidated sterilization process.Confirmed: Sterilized per AAMI TIR-27, validated, and routine control created in accordance with ISO 11137-1:2006. (Page 4-5)
    Physical RequirementsGeneral Physical Requirements, Dimensional Analysis, Visual Inspections, Moisture Content testing, Package and Labeling requirements, Operational Environment Requirements.Confirmed: Design verification testing via functional testing detailed in Section 20.2 (not provided in the excerpt). This included General Physical Requirements, Dimensional Analysis (e.g., 2" ± 0.25" x 2" ± 0.25", < 17% Moisture Content, < 10 ppm Heavy Metals, Page 1), Visual Inspections, Moisture Content testing, Package and Labeling requirements and Operational Environment Requirements. (Page 5)
    Manufacturing ProcessManufactured in a controlled environment with validated processes.Confirmed: Manufactured in an ISO 13485:2003 certified cleanroom environment with complete process validation testing. (Page 5)

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not specify a "test set" in the context of clinical or performance data from human subjects. The study presented here is a non-clinical substantial equivalence comparison.

    • Sample Size: Not applicable in the traditional sense of a clinical trial. The testing involves samples of the SkinTemp II Dressing undergoing physical, chemical, and biological evaluations. Specific sample sizes for each non-clinical test (e.g., biocompatibility samples, sterilization validation samples, dimensional analysis samples) are not provided in this summary.
    • Data Provenance: The testing was performed internally by Human BioSciences, Inc. and by a third-party laboratory, North American Science Association, Inc. (NAmSA), for biocompatibility. The summary does not provide country of origin or whether it's retrospective or prospective, but given it's non-clinical testing for a 510(k) submission, it would be prospective testing on the manufactured device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This is a non-clinical submission demonstrating substantial equivalence to a predicate device, not a study requiring expert readers or interpretation of clinical data. The "ground truth" for the non-clinical tests would be the established scientific and regulatory standards (e.g., FDA regulations, ISO standards, AAMI guidelines) and the documented characteristics of the predicate device.

    4. Adjudication Method for the Test Set

    Not applicable. There is no expert adjudication or clinical assessment involved in this type of substantial equivalence submission.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. There is no mention of an MRMC study. This submission focuses on the device itself and its equivalence to a predicate, not on the improvement of human reader performance with or without AI assistance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable. This is a medical device (wound dressing), not an algorithm or AI-based device.

    7. The Type of Ground Truth Used

    The "ground truth" used for this submission is primarily:

    • Predicate Device Characteristics: The detailed specifications and approved 510(k) information of the BioCore SkinTemp sheet (K913023 & K925545). The new device's characteristics were compared point-by-point to this established device.
    • Regulatory Standards: ISO 10993 (Biocompatibility), AAMI TIR-27 (Sterilization), ISO 11137-1:2006 (Sterilization), and FDA regulations (e.g., 21 CFR Part 807.92). Performance parameters such as sterility, biocompatibility, and manufacturing quality are measured against these established standards.
    • Internal Specifications: For physical characteristics like size, moisture content, and heavy metals, the "ground truth" is the internal design specifications of the SkinTemp II Dressing, which are designed to be consistent with or superior to the predicate.

    8. The Sample Size for the Training Set

    Not applicable. There is no machine learning "training set" involved as this is not an AI/algorithm-driven device.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for an AI/algorithm.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1