Found 8 results

510(k) Data Aggregation

    K Number
    K080091

    Validate with FDA (Live)

    Date Cleared
    2008-02-05

    (22 days)

    Product Code
    Regulation Number
    870.2910
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Health Buddy® Appliance is indicated for use in non-clinical settings to collect and transmit historical medical information to healthcare professionals to help support effective management of their patients.

    Device Description

    The Health Buddy Appliance is a communications product that connects to a telephone line or Ethernet connection. It is used by patients in conjunction with the Health Hero Online service to answer questions and furnish information to their healthcare professional(s) between office visits. The Health Buddy Appliance contains software that can be activated to function with specific medical devices (including blood glucose meters, non-invasive blood pressure cuffs, patient weight scales, peak flow meters and pulse oximeters). The Health Buddy Appliance retrieves data from a specific medical device and stores it for transmission to a healthcare provider.

    The physiologic patient parameters available for retrospective display and evaluation include pulse rate, blood pressure, blood glucose level, weight, and Peak Expiratory Flow (PEF) and FEV1 (forced expiratory volume) measurements, and blood oxygen saturation (%SpO2). The Health Buddy receives connections from these medical devices, and through the data port, wireless hub, or infrared device, downloads readings from the identified attached device and transmits the responses over the telephone or Ethernet connection at predetermined times to the patient's health care professional.

    The Health Buddy Appliance is a simple, user-friendly device that connects to the patient's standard home telephone line. The device connects to a Data Center via a toll-free number or an Ethernet connection to send patient responses since the previous data transfer and to retrieve the new dialogue.

    The screen displays information sent by the patient's healthcare provider and asks questions about vital signs, symptoms and behaviors, and allows the patient to respond via four large buttons. The Health Buddy will respond to the patient's answers with education, reinforcement and messages that prompt patient action. The patient responses are sent to the patient's health care provider.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Health Buddy® Appliance, which is a communication product designed to collect and transmit patient data from various medical devices to healthcare professionals.

    Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided document:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document doesn't provide specific numerical acceptance criteria for performance metrics (e.g., sensitivity, specificity, accuracy). Instead, it states that "Verification and validation testing activities were conducted to establish the performance, functionality and reliability characteristics of the Health Buddy Appliance." And concludes that "The Health Buddy Appliance performed as intended."

    The core "performance" of this device is its ability to communicate and transmit data correctly. The acceptance criteria, therefore, would be that these functions work as expected.

    Acceptance Criteria CategorySpecific Criteria (Inferred)Reported Device Performance
    Functional Performance- Ability to correctly connect to medical devices (blood glucose meters, BP cuffs, weight scales, peak flow meters, pulse oximeters)"The Health Buddy Appliance performed as intended."
    - Ability to retrieve data from connected medical devices"The Health Buddy Appliance performed as intended."
    - Ability to store retrieved data"The Health Buddy Appliance performed as intended."
    - Ability to transmit data to healthcare professionals via telephone or Ethernet connection at predetermined times"The Health Buddy Appliance performed as intended."
    - Ability to display information to the patient and receive responses via buttons"The Health Buddy Appliance performed as intended."
    - Ability to respond to patient answers with education, reinforcement, and messages"The Health Buddy Appliance performed as intended."
    Reliability- Consistent operation over time"Verification and validation testing activities were conducted to establish the... reliability characteristics of the Health Buddy Appliance."
    Safety- Electrical safety compliance"Testing included electrical safety..."
    - EMC environmental testing compliance"...EMC environmental testing..."
    Integration- Proper integration with infrared, Ethernet network connectivity, and new LCD technologies"Integration testing..."
    Signal Integrity- Accurate and unaltered transmission of data from medical devices"Signal integrity testing..."
    System Functionality- Overall system operation as a complete unit"System testing..."
    Physical Durability- Ability to withstand shipping and distribution conditions"Shipping/distribution..."
    Risk Mitigation- Identification and mitigation of hazards through design or labeling"A hazard risk analysis was also performed and associated risks mitigated through design or labeling."

    2. Sample Size Used for the Test Set and Data Provenance:

    The document does not specify the exact sample size used for the test set or the data provenance (e.g., country of origin, retrospective/prospective). It generally refers to "Verification and validation testing activities."

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

    The document does not mention using experts or establishing ground truth in the context of clinical performance for the test set. This device is an accessory for data transmission, not a diagnostic tool requiring expert interpretation of results. Its performance is evaluated based on its technical functionality (e.g., proper data relay), rather than clinical accuracy adjudicated by experts.

    4. Adjudication Method for the Test Set:

    Given that the device's function is data transmission and not diagnostic interpretation, an adjudication method for a test set (like 2+1 or 3+1) is not applicable and not mentioned in the document.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

    A MRMC comparative effectiveness study is not applicable and not mentioned. The Health Buddy Appliance is not an AI-powered diagnostic tool that assists human readers in interpreting medical images or data. Therefore, there is no "human readers improve with AI vs without AI assistance" effect size to report.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study:

    A standalone performance study focused on the algorithm itself was implicitly done through the "Verification and validation testing activities." The software within the Health Buddy Appliance was tested to ensure it correctly handled data retrieval, storage, and transmission. However, the document doesn't quantify this performance with typical standalone metrics like sensitivity or specificity because the device's primary function is communication, not diagnostic interpretation. Its "standalone" performance is its ability to perform its communication tasks correctly.

    7. Type of Ground Truth Used:

    For the Health Buddy Appliance, the "ground truth" would be related to the correctness of data transmission and device functionality. This would involve:

    • Known input data: Inputting specific physiological data from medical devices and verifying that the exact same data is received by the healthcare professional's system.
    • Functional specifications: Verifying that all programmed features (e.g., displaying questions, receiving button presses, responding with messages, connecting via Ethernet/phone) work according to their design specifications.
    • Electrical and environmental standards: Compliance with relevant industry standards for electrical safety and electromagnetic compatibility.

    This is not "expert consensus," "pathology," or "outcomes data" in the traditional clinical sense, but rather a verification against defined technical and functional requirements.

    8. Sample Size for the Training Set:

    The Health Buddy Appliance appears to be a rule-based or firmware-driven device for data transmission, not an AI/machine learning product that requires a "training set" in the conventional sense. Therefore, the concept of a training set sample size is not applicable and not mentioned in the document.

    9. How the Ground Truth for the Training Set Was Established:

    As there is no mention of a training set for an AI/ML algorithm, the method for establishing its ground truth is not applicable and not mentioned.

    Ask a Question

    Ask a specific question about this device

    K Number
    K070543

    Validate with FDA (Live)

    Date Cleared
    2007-03-28

    (30 days)

    Product Code
    Regulation Number
    870.2910
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Health Buddy® Appliance is indicated for use in non-clinical settings to collect and transmit historical medical information to healthcare professionals to help support effective management of their patients.

    Device Description

    The Health Buddy appliance is a communications product that connects to a telephone line. It is used by patients in conjunction with the Health Hero Online service to answer questions and furnish information to their healthcare professional(s) between office visits. The Health Buddy appliance contains software that can be activated to function with specific medical devices (including blood glucose meters, non-invasive blood pressure cuffs, patient weight scales, peak flow meters and pulse oximeters). The Health Buddy appliance retrieves data from a specific medical device and stores it for transmission to a healthcare provider. The physiologic patient parameters available for retrospective display and evaluation include blood pressure, blood glucose level, weight, and Peak Expiratory Flow (PEF) and FEV1 (forced expiratory volume) measurements, and blood oxygen saturation (%SpO₂) and pulse rate using a digital pulse oximeter. The Health Buddy receives connections from these medical devices, and through the data port, wireless hub, or infrared device, downloads readings from the identified attached device and transmits the responses over the phone lines at predetermined times to the patient's health care professional.

    The Health Buddy appliance is a simple, user-friendly device that connects to the patient's standard home telephone line. The device connects to a Data Center via a toll-free number to send patient responses since the previous data transfer and to retrieve the new dialogue.

    The screen displays information and asks questions about vital signs, symptoms and behaviors sent by the patient's healthcare provider, and allows the patient to respond via four large buttons. The Health Buddy will respond to the patient's answers with education, reinforcement and messages that prompt patient action. The patient responses are sent to the patient's health care provider.

    AI/ML Overview

    Here's an analysis of the provided text regarding the Health Buddy® Appliance, addressing your requested information:

    Key Takeaway: The provided 510(k) summary for the Health Buddy® Appliance is for a special 510(k) of device modifications. This means it is based on modifications to an already cleared device and demonstrates substantial equivalence to predicate devices. As such, it does not contain a new study with acceptance criteria and device performance results in the way you might expect for an initial clearance or a comprehensive clinical trial. The focus is on demonstrating that the modifications do not introduce new questions of safety or effectiveness.

    Therefore, many of your requested items (like a table of acceptance criteria, sample sizes, expert ground truth, MRMC studies, standalone performance studies) are not present or applicable to this specific type of submission as described.


    Summary of Requested Information from the Provided Text:

    1. A table of acceptance criteria and the reported device performance

    • Not applicable/Not provided in this document. This special 510(k
      ) submission focuses on demonstrating substantial equivalence of modifications to an existing device, rather than presenting a de novo performance study against new acceptance criteria. The document asserts that the safety and effectiveness are "reasonably assured" based on equivalence to predicate devices.

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    • Not applicable/Not provided. No specific test set or data provenance is mentioned as this is a modification submission based on substantial equivalence.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    • Not applicable/Not provided. No ground truth establishment for a test set is discussed in this context.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    • Not applicable/Not provided. No test set or adjudication method is discussed.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable/Not provided. This device is a data management system/accessory, not an AI-powered diagnostic tool, and no MRMC study is mentioned.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable/Not provided. The device's function is to collect and transmit historical data from other medical devices to healthcare professionals. It does not perform an "algorithm only" diagnostic function that would typically be evaluated in a standalone performance study. Its role is as an accessory and communication tool. The indication explicitly states it is "not intended to provide automated treatment decisions, nor is it to be used as a substitute for a professional healthcare judgment."

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • Not applicable/Not provided. As there is no new performance study described, there is no explicit mention of ground truth types. The device transmits historical data from other medical devices, so the "ground truth" for the data itself would derive from the accuracy of those connected medical devices (e.g., blood glucose meter, blood pressure cuff).

    8. The sample size for the training set

    • Not applicable/Not provided. The document describes a medical device for data transmission, not an AI/ML algorithm that would typically require a training set.

    9. How the ground truth for the training set was established

    • Not applicable/Not provided. See the explanation for point 8.

    Detailed Information from the Provided Text (Relevant to Device Description and Predicate):

    Acceptance Criteria & Device Performance (Indirect):

    Since this is a Special 510(k) for modifications, the "acceptance criteria" are generally understood to be that the modifications do not alter the fundamental operation or intended use such that the device is no longer substantially equivalent to its predicates, and that the modifications do not raise new questions of safety or effectiveness. The "performance" is implicitly deemed acceptable because it maintains substantial equivalence to already cleared devices.

    Acceptance Criteria (Implied for Special 510(k) Modification)Reported Device Performance (Implied from Substantial Equivalence)
    Modifications do not change the indications for use.The device maintains its indicated use in non-clinical settings to collect and transmit historical medical information.
    Modifications do not change the fundamental technology.The core technology of connecting to a telephone line and transmitting data to a Data Center remains. Modifications relate to supporting additional medical devices via data port, wireless hub, or infrared.
    Modifications do not raise new questions of safety/effectiveness.Safety and effectiveness are "reasonably assured" as per the substantial equivalence conclusion.
    Device maintains ability to retrieve and store data from specific medical devices.The Health Buddy appliance continues to retrieve and store data from connected medical devices (blood glucose meters, non-invasive blood pressure cuffs, patient weight scales, peak flow meters, pulse oximeters) for transmission.

    Predicate Devices (Form the Basis of Equivalence, not a "study"):

    Indication for Use:

    "Health Buddy® Appliance is indicated for use in non-clinical settings to collect and transmit historical medical information to healthcare professionals to help support effective management of their patients."

    Conclusion from the 510(k) Summary:

    "The Health Buddy Appliance is substantially equivalent in technology, features, and indications for use to devices cleared under the Federal Food. Drug and Cosmetic Act. Therefore, safety and effectiveness are reasonably assured, justifying 510(k) clearance of this modified device."

    Ask a Question

    Ask a specific question about this device

    K Number
    K063612

    Validate with FDA (Live)

    Date Cleared
    2006-12-29

    (25 days)

    Product Code
    Regulation Number
    870.2910
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Health Buddy® Appliance is indicated for use in non-clinical settings to collect and transmit historical medical information to healthcare professionals to help support effective management of their patients.

    Device Description

    The Health Buddy appliance is a communications product that connects to a telephone line. It is used by patients in conjunction with the Health Hero Online service to answer questions and furnish information to their healthcare professional(s) between office visits. The Health Buddy appliance contains software that can be activated to function with specific medical devices (including blood glucose meters, non-invasive blood pressure cuffs, patient weight scales, peak flow meters and pulse oximeters). The Health Buddy appliance retrieves data from a specific medical device and stores it for transmission to a healthcare provider. The physiologic patient parameters available for retrospective display and evaluation include blood pressure, blood glucose level, weight, and Peak Expiratory Flow (PEF) and FEV1 (forced expiratory volume) measurements, and blood oxygen saturation (%SpO₂) and pulse rate using a digital pulse oximeter. The Health Buddy receives connections from these medical devices, and through the data port, wireless hub, or infrared device, downloads readings from the identified attached device and transmits the responses over the phone lines at predetermined times to the patient's health care professional.

    The Health Buddy appliance is a simple, user-friendly device that connects to the patient's standard home telephone line. The device connects to a Data Center via a toll-free number to send patient responses since the previous data transfer and to retrieve the new dialoque.

    The screen displays information and asks questions about vital signs, symptoms and behaviors sent by the patient's healthcare provider, and allows the patient to respond via four large buttons. The Health Buddy will respond to the patient's answers with education, reinforcement and messages that prompt patient action. The patient responses are sent to the patient's health care provider.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for a modification to the Health Buddy® Appliance. However, it does NOT contain detailed information about specific acceptance criteria or a study that rigorously proves the device meets those criteria in terms of performance metrics like accuracy, sensitivity, or specificity.

    The submission focuses on establishing substantial equivalence to previously cleared devices for a modification to the software to support the connection of additional medical devices. This is a regulatory pathway that relies more on demonstrating that the changes do not introduce new questions of safety or effectiveness, rather than providing new performance data for the device as a whole.

    Therefore, many of the requested details are not present in the provided document.

    Here's a breakdown based on the available information:


    1. Table of Acceptance Criteria and Reported Device Performance

    This information is not provided in the document. The 510(k) submission for this device modification focuses on establishing substantial equivalence rather than presenting new performance data against specific acceptance criteria. The document explicitly states: "This submission represents a modification to the software in the Health Buddy® Appliance to support the connection of additional medical devices."


    2. Sample Size Used for the Test Set and Data Provenance

    This information is not provided in the document. No specific test set or study details with sample sizes are mentioned.


    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    This information is not provided in the document. There is no mention of experts or ground truth establishment for a test set.


    4. Adjudication Method for the Test Set

    This information is not provided in the document.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of Human Readers Improve with AI vs. Without AI Assistance

    This information is not applicable/provided. The Health Buddy Appliance is a data management and communication system for physiological parameters, not an AI-assisted diagnostic tool that would typically involve an MRMC study comparing human readers with and without AI. It collects and transmits historical data for healthcare professionals to support patient management.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Study Was Done

    This information is not provided in the document. The device primarily functions as a data conduit and communication tool, not an autonomous algorithm requiring standalone performance evaluation in the typical sense of a diagnostic or therapeutic algorithm.


    7. The Type of Ground Truth Used

    This information is not provided in the document. Given the nature of the device as a data transmission system, the "ground truth" would likely relate to the accuracy of data transmission and correlation with the attached medical devices, but specific details are not outlined.


    8. The Sample Size for the Training Set

    This information is not provided in the document. There is no mention of a training set for an algorithm.


    9. How the Ground Truth for the Training Set Was Established

    This information is not provided in the document.


    Summary of the Device and its Regulatory Review:

    The Health Buddy® Appliance is a communication product that connects to a telephone line. It functions by retrieving data from various medical devices (blood glucose meters, blood pressure cuffs, scales, peak flow meters, pulse oximeters) and transmitting this historical patient information to healthcare professionals. It also allows healthcare providers to send questions and educational messages to patients.

    This specific 510(k) submission (K063612) is a modification to the software of an already cleared Health Buddy® Appliance to support the connection of additional medical devices. The regulatory review relies on demonstrating substantial equivalence to previously cleared Health Buddy versions and other similar telemonitoring systems.

    The core assertion for clearance is that the modified device is "substantially equivalent in technology, features, and indications for use to devices cleared under the Federal Food, Drug and Cosmetic Act," and therefore its "safety and effectiveness are reasonably assured." This implies that the performance characteristics (e.g., accuracy of data transmission, reliability of communication) of the modified device are considered to be comparable to those of the predicate devices without requiring new, extensive performance studies to define new acceptance criteria and prove them.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060843

    Validate with FDA (Live)

    Date Cleared
    2006-04-20

    (23 days)

    Product Code
    Regulation Number
    870.2700
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Health Buddy® Appliance is indicated for use in non-clinical settings to collect and transmit historical medical information to healthcare professionals to help support effective management of their patients.

    Device Description

    The Health Buddy appliance is a communications product that connects to a telephone line. It is used by patients in conjunction with the Health Hero Online service to answer questions and furnish information to their healthcare professional(s) between office visits. The Health Buddy appliance contains software that can be activated to function with specific medical devices (including blood glucose meters, non-invasive blood pressure cuffs, patient weight scales, peak flow meters and pulse oximeters). The Health Buddy appliance retrieves data from a specific medical device and stores it for transmission to a healthcare provider. The physiologic patient parameters available for retrospective display and evaluation include blood pressure, blood glucose level, weight, and Peak Expiratory Flow (PEF) and FEV1 (forced expiratory volume) measurements, and blood oxygen saturation (%SpO₂) and pulse rate using a digital pulse oximeter. The Health Buddy receives connections from these medical devices, and through the data port, downloads readings from the identified attached device and, at predetermined times, transmits the responses over the phone lines to the patient's health care professional.

    The Health Buddy appliance is a simple, user-friendly device that connects to the patient's standard home telephone line. The device connects to a Data Center via a toll-free number to send patient responses since the previous data transfer and to retrieve the new dialoque.

    The screen displays information and asks questions about vital signs, symptoms and behaviors sent by the patient's healthcare provider, and allows the patient to respond via four large buttons. The Health Buddy will respond to the patient's answers with education, reinforcement and messages that prompt patient action. The patient responses are sent to the patient's health care provider.

    AI/ML Overview

    The provided text describes a 510(k) submission for modifications to the Health Buddy® Appliance. However, it does not contain any information about acceptance criteria or a study designed to prove the device meets those criteria.

    The document focuses on:

    • Regulatory Authority and Company Information
    • Device Name and Classification
    • Predicate Devices: Identifying previously cleared devices that the modified device is substantially equivalent to.
    • Device Description: Explaining what the Health Buddy appliance is and how it functions as a communication tool for collecting and transmitting physiological data from various medical devices.
    • Indication for Use Statement: Defining the intended purpose of the device.
    • Substantial Equivalence Comparison: Arguing that the modified device is substantially equivalent to existing cleared devices.
    • Conclusion: Stating that safety and effectiveness are reasonably assured due to substantial equivalence.
    • FDA Correspondence: The FDA's letter granting 510(k) clearance based on substantial equivalence.

    Therefore, I cannot provide the requested information about acceptance criteria or a study that proves the device meets them because this information is not present in the provided text.

    In a real-world scenario for a medical device requiring performance testing, the requested information would typically be found in a separate section detailing verification and validation activities, including test protocols, acceptance criteria, and specific study results. This document bypasses such a detailed study by asserting substantial equivalence to previously cleared devices for modifications that are primarily software-related to support additional medical device connections.

    Ask a Question

    Ask a specific question about this device

    K Number
    K050567

    Validate with FDA (Live)

    Date Cleared
    2005-04-05

    (32 days)

    Product Code
    Regulation Number
    870.2910
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Health Buddy® Appliance is indicated for use in non-clinical settings to collect and transmit historical medical information to healthcare professionals to help support effective management of their patients.

    Device Description

    The Health Buddy appliance is a communications product that connects to a telephone line. It is used by patients in conjunction with the Health Hero Online service to answer questions and furnish information to their healthcare professional(s) between office visits. The Health Buddy appliance contains software that can be activated to function with specific medical devices (including blood glucose meters, non-invasive blood pressure cuffs, patient weight scales and peak flow meters). The Health Buddy appliance retrieves data from a specific medical device and stores it for transmission to a healthcare provider. The physiologic patient parameters available for retrospective display and evaluation include blood pressure, blood glucose level, weight, and Peak Expiratory Flow (PEF) and FEV1 (forced expiratory volume) measurements. In this submission, software changes are made to support hardware replacement to components no longer available in the Health Buddy. In addition, the operating system has changed. The Health Buddy device has data ports that allow it to download readings from identified attached devices, activated only upon appropriate patient enrollment.

    The Health Buddy appliance is a simple, user-friendly device that connects to the patient's standard home telephone line. The device connects to a Data Center via a toll-free number to send patient responses from the previous day and to retrieve the current day's dialoque.

    The screen displays the information or questions about vital signs, symptoms and behaviors sent by the patient's healthcare provider, and allows the patient to respond via four large buttons. The Health Buddy will respond to the patient's answers with education, reinforcement and messages that prompt patient action. The patient information is sent to the patient's health care provider.

    The Health Buddy® is an accessory device, intended to be a communication tool to enable healthcare providers to receive historical patient information. The product is used in conjunction with Health Hero Network's Online Service, a communication tool to enable health care providers to educate, motivate, and receive patient information. Health Buddy Appliance is not intended to provide automated treatment decisions, nor is it to be used as a substitute for a professional healthcare judgment. All patient medical diagnosis and treatment are to be performed under the supervision and oversight of an appropriate healthcare professional.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a modification to the Health Buddy® Appliance. This type of submission focuses on demonstrating "substantial equivalence" to a predicate device, rather than proving safety and efficacy through new, independent studies with acceptance criteria in the traditional sense.

    Therefore, many of the requested categories (acceptance criteria, specific study design, sample sizes for test and training sets, ground truth establishment for novel algorithms, MRMC studies, and standalone performance) are not applicable to this type of submission.

    Here's an breakdown based on the provided text, explaining why certain information is absent:


    1. Table of Acceptance Criteria and Reported Device Performance

    • Not Applicable. This 510(k) is for modifications to an already cleared device (Health Buddy® Appliance) and focuses on demonstrating substantial equivalence to its predicates. It does not present new acceptance criteria or new performance metrics with specific thresholds that a study would aim to meet, as would be the case for a novel device or a device with new clinical indications. The "performance" here is primarily about maintaining the functionality and safety of the original device despite hardware and OS changes.

    2. Sample Size Used for the Test Set and Data Provenance

    • Not Applicable. No explicit "test set" in the context of an algorithm's performance on a specific dataset is mentioned. The changes are to hardware components and the operating system of an existing communication device. The "testing" would have focused on ensuring the modified device continues to perform its intended functions (collecting and transmitting data from attached medical devices) as reliably as the original, in line with its cleared indications. This would likely involve engineering verification and validation, not a clinical study on a patient data set.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    • Not Applicable. As there's no explicit "test set" involving diagnostic or interpretative tasks, there's no mention of experts establishing ground truth for such a set.

    4. Adjudication Method for the Test Set

    • Not Applicable. No test set requiring expert adjudication is described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Not Applicable. An MRMC study is typically conducted for diagnostic devices where human readers interpret medical images or data. The Health Buddy® Appliance is a data collection and transmission device, not an interpretive diagnostic tool. There is no AI component that would "assist" human readers in the way an MRMC study evaluates.

    6. Standalone (Algorithm Only) Performance Study

    • Not Applicable. The Health Buddy® Appliance acts as a communication tool and an accessory to other medical devices (blood glucose meters, blood pressure cuffs, etc.). It does not contain a standalone "algorithm" for making medical decisions or interpretations, nor does it perform any diagnostic functions independently. Its software updates are related to hardware compatibility and the operating system, not to a novel clinical algorithm.

    7. Type of Ground Truth Used

    • Not Applicable. Given the nature of the device (a data communication system), the concept of "ground truth" as it applies to diagnostic accuracy (e.g., pathology, clinical outcome) is not relevant. The device's function is to accurately transmit data, not to interpret it.

    8. Sample Size for the Training Set

    • Not Applicable. This device does not use machine learning algorithms that require a "training set" of data for learning or model development. The software changes are related to hardware drivers and the operating system.

    9. How the Ground Truth for the Training Set Was Established

    • Not Applicable. As no training set is described (see point 8), there is no ground truth establishment for a training set.

    Summary of the 510(k) Submission:

    The provided document describes a Special 510(k) submission for modifications to the "Health Buddy® Appliance." The key points are:

    • Device Type: A communication product that connects to a telephone line, used by patients to answer questions and transmit data from specific medical devices (blood glucose meters, blood pressure cuffs, scales, peak flow meters) to healthcare professionals.
    • Modifications: Software changes to support hardware replacement due to components no longer being available, and an update to the operating system.
    • Indication for Use: "Health Buddy® Appliance is indicated for use in non-clinical settings to collect and transmit historical medical information to healthcare professionals to help support effective management of their patients." It explicitly states it is not intended for automated treatment decisions or as a substitute for professional healthcare judgment.
    • Predicate Devices: The submission claims substantial equivalence to previously cleared versions of the Health Buddy® (K042273, K040086, K993128) and other similar telehealth monitors/systems (CareMatix Wellness System K040966, Viterion 100 TeleHealth Monitor K030419).
    • Conclusion: The changes maintain the technological features and indications for use, thus demonstrating substantial equivalence to cleared devices, assuring safety and effectiveness.

    This 510(k) is a regulatory filing for a device modification, not a report on a clinical study establishing new performance metrics for a novel algorithm or diagnostic capability. The "study" here is essentially the demonstration that the modified device maintains its previously cleared functionality without raising new questions of safety or effectiveness.

    Ask a Question

    Ask a specific question about this device

    K Number
    K042273

    Validate with FDA (Live)

    Date Cleared
    2004-08-31

    (8 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Health Buddy® with Device Connectivity is indicated for use in non-clinical settings to collect and transmit historical medical information to healthcare professionals to help support effective management of their patients.

    Device Description

    The Health Buddy appliance is a communications product that connects to a telephone line. It is used by patients in conjunction with the Health Hero Online service to answer questions and furnish information to their healthcare professional(s) between office visits. The Health Buddy appliance contains software that can be activated to function with specific medical devices (including blood glucose meters, non-invasive blood pressure cuff, patient weight scale and peak flow meter). The Health Buddy with Device Connectivity retrieves data from a specific medical device and stores it for later transmission to a healthcare provider. The physiologic patient parameters available for retrospective display and evaluation include blood pressure, blood glucose level, and weight. In this submission, minor software changes are made in the Health Buddy ROM to allow it to connect to a Peak Flow Meter and monitor Peak Expiratory Flow (PEF) and FEV1 measurements. The Health Buddy has the ability to receive connections from optional, compatible devices, including those identified above, and others, such as digital thermometers, pulse oximeters and prothrombin/INR meters. The Health Buddy device has a data port that allows it to download readings from an identified attached device, activated only upon appropriate patient enrollment.

    The Health Buddy with Device Connectivity is a simple, user-friendly device that connects to the patient's standard home telephone line. The device connects to the Data Center via a toll-free number to send patient responses from the previous day and to retrieve the current day's dialogue.

    The screen displays the information or questions about vital signs, symptoms and behaviors sent by the patient's healthcare provider, and allows the patient to respond via four large buttons. The Health Buddy will respond to the patient's answers with education, reinforcement and messages that prompt patient action. The patient information is sent to the patient's health care provider.

    AI/ML Overview

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The provided document is a 510(k) premarket notification for a minor software modification, asserting substantial equivalence to previously cleared devices. It does not contain explicit performance acceptance criteria or a study with reported device performance metrics against such criteria.

    The core of the submission is to demonstrate that the minor software changes to allow connection to a Peak Flow Meter do not alter the fundamental safety and effectiveness of the existing device. It relies on the concept of "substantial equivalence" rather than a de novo performance study.

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not applicable (N/A) – No specific test set to evaluate performance is described in this substantial equivalence submission.
    • Data Provenance: Not applicable (N/A) – No new clinical data or data from a test set is presented. The submission relies on establishing equivalence to previously cleared devices.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Not applicable (N/A) – No new ground truth establishment for a test set is described.
    • Qualifications of Experts: Not applicable (N/A)

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable (N/A) – No test set or ground truth adjudication is described.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • MRMC Study Done: No. This type of study is not mentioned or implied in the provided document. The device is a data collection and transmission accessory, not an AI diagnostic tool intended to assist human readers in image interpretation or similar tasks.
    • Effect Size of Human Readers Improve with AI vs. Without AI Assistance: Not applicable (N/A)

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Standalone Study Done: No. The Health Buddy is described as a "communication tool" and an "accessory device" that collects and transmits historical data. It is not an algorithm that performs standalone diagnostic or interpretive functions.

    7. Type of Ground Truth Used

    • Type of Ground Truth: Not applicable (N/A) – No ground truth is established or used in this submission. The device's function is to transmit existing physiological data, not to make a diagnosis or interpretation itself.

    8. Sample Size for the Training Set

    • Sample Size for Training Set: Not applicable (N/A) – The document does not describe a machine learning algorithm or a training set. The "software changes" mentioned are related to device connectivity for specific medical devices, not an AI/ML model.

    9. How the Ground Truth for the Training Set Was Established

    • How Ground Truth Was Established: Not applicable (N/A) – No training set or ground truth establishment for such a set is described.

    Summary of Device's Approach to Meeting Requirements:

    The document asserts that the modified Health Buddy® with Device Connectivity meets safety and effectiveness requirements by demonstrating substantial equivalence to predicate devices that have already received 510(k) clearance. The changes are described as "minor software changes" to allow connection to a Peak Flow Meter. The justification relies on the fact that the fundamental function of collecting and transmitting historical data for healthcare professionals (without providing automated treatment decisions) remains unchanged and aligns with the cleared functions of the predicate devices. Therefore, a new performance study with acceptance criteria and ground truth establishment was not deemed necessary for this specific submission.

    Ask a Question

    Ask a specific question about this device

    K Number
    K040086

    Validate with FDA (Live)

    Date Cleared
    2004-03-10

    (54 days)

    Product Code
    Regulation Number
    870.1130
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Health Buddy® with Buddylink is indicated for use in non-clinical settings to collect and transmit historical medical information to healthcare professionals to help support effective management of their patients.

    Device Description

    The Health Buddy appliance is a communications product that connects to a telephone line. It is used by patients in conjunction with the Health Hero Online service to answer questions and furnish information to their healthcare professional(s) between office visits. The Health Buddy appliance contains software that can be activated to function with specific medical devices (such as blood glucose meter, non-invasive blood pressure cuff and patient weight scale). The Health Buddy with Buddylink retrieves data from a medical device (blood glucose meter, non-invasive blood pressure cuff, or patient weight scale) and stores it for later transmission to a healthcare provider. The device has a data port that allows it to download readings from the attached device, only activated upon appropriate patient enrollment.

    The Health Buddy with Buddylink is a simple; user-friendly device that connects to the patient's standard home telephone line. The device connects to the Data Center via a toll-free number to send responses from the previous day and to retrieve the current day's dialogue.

    The screen displays the information or questions about vital signs, symptoms and behaviors sent by the patient's healthcare provider, and allows the patient to respond via four large buttons. The Health Buddy will respond to the patient's answers with education, reinforcement and messages that prompt patient action. The patient information is sent to the patient's health care provider.

    AI/ML Overview

    This device, the Health Buddy® with Buddylink, is a communication product that connects to a telephone line. It is used by patients to answer questions and furnish information to their healthcare professional(s) between office visits. It can retrieve data from an attached medical device (blood glucose meter, non-invasive blood pressure cuff, or patient weight scale) and stores it for later transmission to a healthcare provider.

    Based on the provided information, the submission focuses on a modification to the software in the Health Buddy® with Buddylink, not a new stand-alone study to prove effectiveness against specific acceptance criteria. The device is cleared based on substantial equivalence to predicate devices. Therefore, the traditional acceptance criteria and detailed study results, as one might expect for a de novo device, are not explicitly provided in this document.

    Here's an attempt to answer your questions based on the available information:


    1. Table of Acceptance Criteria and Reported Device Performance

    Given that this submission is for a modification to software and relies on substantial equivalence, explicit "acceptance criteria" and "reported device performance" in the quantifiable sense of a clinical trial are not presented. The primary "acceptance criterion" met here is demonstrating substantial equivalence to legally marketed predicate devices.

    Acceptance Criteria CategorySpecific Criteria (Implicitly Met by Substantial Equivalence)Reported Device Performance
    SafetyDevice is safe for its intended use.Implied by substantial equivalence to cleared devices. No new safety concerns identified for software modification.
    EffectivenessDevice effectively collects and transmits historical medical information to healthcare professionals and supports effective patient management.Implied by substantial equivalence to cleared devices (K993128, K984527, K020184). The software modification is deemed to maintain or improve effectiveness in data collection and transmission.
    Technical PerformanceCommunication functionality (connecting via telephone line, sending and retrieving data); data download from medical devices; user interface for patient interaction.Functions as described for data collection and transmission, similar to predicate devices. Software modification did not degrade these functions.
    Indications for UseAligns with the stated indications: "collect and transmit historical medical information to healthcare professionals to help support effective management of their patients in non-clinical settings."The device, with the software modification, continues to meet these indications.

    2. Sample Size Used for the Test Set and Data Provenance

    No specific test set or sample size for a performance study is described as this is a substantial equivalence submission for a software modification. The comparison is against predicate devices.

    • Sample Size for Test Set: Not applicable / Not provided.
    • Data Provenance: Not applicable / Not provided for an independent performance study. The comparison is based on the characteristics and cleared status of the predicate devices.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. As no independent test set performance data is described, there's no mention of experts establishing ground truth for such a set.


    4. Adjudication Method for the Test Set

    Not applicable. No discrete test set requiring adjudication is described.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

    No, an MRMC comparative effectiveness study was not done or described in this document. The submission is based on substantial equivalence to predicate devices, not improvement over human readers or AI assistance.


    6. If a Standalone (algorithm only without human-in-the-loop performance) was done

    No, a standalone algorithm performance study is not described. The device is a communication tool, and its function involves human input (data from medical devices, patient responses) and human interaction (healthcare providers receiving data). The evaluation focused on its equivalence as a communication tool.


    7. The Type of Ground Truth Used

    Not applicable in the context of diagnostic or interpretive performance. The "ground truth" for the device's function is its ability to reliably collect and transmit data, which is implicitly accepted through the substantial equivalence argument to already cleared devices.


    8. The Sample Size for the Training Set

    Not applicable. This document describes a medical device (hardware with software), not a machine learning algorithm requiring a training set for model development in the sense of AI performance evaluation. The software modification is likely related to connectivity or user interface enhancements rather than a learning algorithm.


    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as no training set for a machine learning model is mentioned.

    Ask a Question

    Ask a specific question about this device

    K Number
    K993128

    Validate with FDA (Live)

    Date Cleared
    1999-11-23

    (64 days)

    Product Code
    Regulation Number
    862.1345
    Reference & Predicate Devices
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Health Buddy with Device Connect [Glucose Meter] is indicated for use in non-clinical settings to collect and transmit historical blood glucose meter data to healthcare professionals to help support effective management of diabetes patients.

    Health Buddy with Device Connect is intended to be a communication tool to enable healthcare providers to receive patient blood glucose information. The product is intended to be used in conjunction with Health Hero Network Online Service, a communication tool to enable healthcare providers to educate, motivate, and receive patient information. Health Buddy with Device Connect and the Health Hero Network Online Service are not intended to provide automated treatment decisions, nor to be used as a substitute for a healthcare professional's judgment. All patient medical diagnosis and treatment are to be performed under the supervision and oversight of an appropriate healthcare professional.

    Device Description

    Health Buddy with Device Connect [Glucose Meter] is indicated for use in non-clinical settings to collect and transmit historical blood glucose meter data to healthcare professionals to help support effective management of diabetes patients.

    AI/ML Overview

    The provided text describes a 510(k) submission for the Health Buddy® with Device Connect [Glucose Meter], a device intended to collect and transmit historical blood glucose meter data to healthcare professionals. However, the document does not contain any information regarding specific acceptance criteria, a study proving the device meets acceptance criteria, or details regarding device performance metrics.

    The document is a US FDA 510(k) summary and the accompanying FDA clearance letter, which focuses on:

    • Device Description and Intended Use: Explains what the device does (collects and transmits blood glucose data) and its purpose (to help healthcare professionals manage diabetes patients).
    • Substantial Equivalence: States that the device is substantially equivalent to legally marketed predicate devices (LifeScan ONE TOUCH Profile Diabetes Tracking System and LifeScan IN TOUCH Diabetes Management Software). This is the primary basis for 510(k) clearance, rather than a de novo performance study against pre-defined acceptance criteria.
    • Regulatory Classification: Identifies it as a Class II device under 21 CFR §862.1345 (Glucose Test System).
    • Administrative Details: Contact information, dates, and FDA correspondence.

    Therefore, I cannot fulfill your request for:

    1. A table of acceptance criteria and the reported device performance: This information is not present in the provided text. The 510(k) clearance is based on substantial equivalence, not a direct comparison against explicit performance acceptance criteria in the provided document.
    2. Sample size used for the test set and the data provenance: Not available. No "test set" in the context of a performance study is described.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No ground truth establishment is described.
    4. Adjudication method: Not applicable.
    5. Multi-reader multi-case (MRMC) comparative effectiveness study: Not mentioned.
    6. Standalone performance study: Not mentioned.
    7. Type of ground truth used: Not applicable.
    8. Sample size for the training set: Not applicable.
    9. How the ground truth for the training set was established: Not applicable.

    The document is a regulatory submission for substantial equivalence clearance, not a clinical or performance study report that would contain the requested details on acceptance criteria and performance data.

    Ask a Question

    Ask a specific question about this device