Search Filters

Search Results

Found 6 results

510(k) Data Aggregation

    K Number
    K161502

    Validate with FDA (Live)

    Device Name
    MC1 Plus
    Date Cleared
    2016-08-16

    (76 days)

    Product Code
    Regulation Number
    890.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MC1 Plus is indicated for:
    a) Therapeutic Massager:

    1. Provides temporary relief of minor muscle aches and pains;
    2. Relieves muscle spasms;
    3. Temporarily improves local blood circulation;
    4. Temporarily reduces the appearance of cellulite.
      b) Ultrasonic Diathermy:
    5. Relief of pain;
    6. Muscle spasms;
    7. Joint contractures;
    8. NOT for the treatment of malignancies.
    Device Description

    MC1 Plus is a computerized body massager and ultrasound diathermy system. The MC1 Plus is supplied with two handpieces: one ultrasonic handpiece with two 1 MHz transducers mounted on it and one handpiece for computerized body massager.

    AI/ML Overview

    The provided text is a 510(k) Summary for the General Project MC1 Plus device, a computerized body massager and ultrasound diathermy system. The document focuses on demonstrating substantial equivalence to a legally marketed predicate device (MC1, K091615), rather than describing a study with acceptance criteria for a novel device's performance.

    Therefore, the requested information categories (1-7, 9) cannot be fully populated as they pertain to clinical studies designed to establish specific performance metrics against pre-defined acceptance criteria for a new device. This document primarily relies on engineering and electrical safety testing, and performance testing for equivalence to a predicate, not clinical trials to prove device efficacy against acceptance criteria.

    However, based on the provided text, here's what can be extracted and inferred:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly state "acceptance criteria" in the context of a clinical study for intended use. Instead, it demonstrates compliance with recognized electrical safety and performance standards, and comparison to a predicate device.

    Acceptance Criteria CategorySpecific Criteria (from standards or predicate comparison)Reported Device Performance (MC1 Plus)
    Electrical SafetyConformance to AAMI/ANSI ES60601-1Conforms
    Electromagnetic Compatibility (EMC)Conformance to IEC 60601-1-2Conforms
    Ultrasonic Physiotherapy Equipment Electrical SafetyConformance to IEC 60601-2-5 (3rd ed.)Conforms
    Tissue Heating PerformanceEquivalence to MC1 (K091615) in terms of increasing tissue temperature to at least 40°C"Testing showed that the MC1 Plus increases tissue temperature as required for ultrasonic diathermy.""The two 1 MHz ultrasound transducers can be considered equivalent to the MC1 in terms of heating tissue temperature to at least 40° C."
    Leakage Current (Ground NC)<0.500 mA (from IEC 60601-1)0.085 mA
    Leakage Current (Ground SFC)<1.0 mA (from IEC 60601-1)0.107 mA
    Leakage Current (Patient NC DC)<0.01 mA (from IEC 60601-1)<0.001 mA
    Leakage Current (Patient NC AC)<0.10 mA (from IEC 60601-1)<0.01 mA
    Leakage Current (Patient SFC DC)<0.05 mA (from IEC 60601-1)<0.001 mA
    Leakage Current (Patient SFC AC)<0.50 mA (from IEC 60601-1)<0.01 mA
    Maximum Ultrasonic Power OutputComplies with IEC 60601-2-5 3rd ed. (reduced from prior device)15 W (reduced from 21 W)
    Maximum Ultrasonic IntensityComplies with IEC 60601-2-5 3rd ed. (reduced from prior device)3 W/cm² (reduced from 7 W/cm²)
    ERA (Effective Radiating Area)Complies with IEC 60601-2-5 (modified from prior device)4.9 cm² (modified from 3 cm²)

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not describe a "test set" in the context of human subjects or patient data. The "performance data" refers to engineering and electrical safety testing of the device itself. There is no information about data provenance (country of origin, retrospective/prospective) because it's not a clinical data study.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Not applicable. Ground truth establishment by clinical experts is not mentioned as this is not a study involving diagnostic interpretation or clinical outcomes adjudicated by experts.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable. There is no "test set" requiring adjudication in the context of clinical or diagnostic performance.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. This device is a therapeutic massager and ultrasonic diathermy system; it does not involve "human readers" or "AI assistance" for diagnostic interpretation.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    The device is a physical therapeutic device, not an algorithm. Therefore, "standalone" performance in the context of AI algorithms is not applicable. The device's performance stands alone in its function without a human interpretation loop.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    The "ground truth" for the performance claims appears to be:

    • Engineering measurements and standards: Conformance to internationally recognized electrical safety and EMC standards (AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-2-5) through direct testing of the device's electrical parameters and output.
    • Comparison to predicate device: Demonstration that the MC1 Plus performs "similarly" to the predicate MC1 in terms of its ability to increase tissue temperature to at least 40°C. This implies that the predicate's established performance serves as the benchmark for this equivalence.

    8. The sample size for the training set

    Not applicable. This is not an AI/machine learning device that would require a training set.

    9. How the ground truth for the training set was established

    Not applicable. As there is no training set for an AI model, there is no ground truth establishment for a training set.

    Ask a Question

    Ask a specific question about this device

    K Number
    K091615

    Validate with FDA (Live)

    Device Name
    MED CONTOUR
    Date Cleared
    2010-05-06

    (337 days)

    Product Code
    Regulation Number
    890.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MC1 is indicated for: a) Therapeutic Massager: 1. Provides temporary relief of minor muscle aches and pains; 2. Relieves muscle spasms, 3. Temporarily improves local blood circulation; 4. Temporarily reduces the appearance of cellulite. b) Ultrasonic Diathermy: 1. Relief of pain; 2. Muscle spasms; 3. Joint contractures; 4. NOT for the treatment of malignancies.

    Device Description

    The MC1 is a computerized body massager and ultrasound diathermy system. The MC1 is supplied with two handpieces: one ultrasonic handpiece with two 1 MHz transducers mounted on it and one handpiece for computerized body massage.

    AI/ML Overview

    Here's a summary of the acceptance criteria and study information for the General Project MC1, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Electrical SafetyConforms with IEC 60601-1
    Electromagnetic Compatibility (EMC)Conforms with IEC 60601-1-2
    Electrical Safety in Ultrasonic Physiotherapy EquipmentConforms with IEC 60601-2-5
    Tissue Heating for Ultrasonic Diathermy (Required for therapeutic effect, specifically for tissue temperature to reach at least 40° C)The MC1 increases tissue temperature as required for ultrasonic diathermy; "The two 1 MHz ultrasound transducers can be considered equivalent to the MED Sculpt in terms of heating tissue temperature to at least 40° C." Results were similar to the predicate device (MED Sculpt).

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state a "test set" in the context of patient or image data for evaluating clinical performance. The performance testing described focuses on engineering and physical performance (electrical safety, EMC, tissue heating).

    • Sample Size: Not applicable in the traditional sense of a clinical test set. The evaluation compared the MC1's physical performance to a predicate device, the MED Sculpt.
    • Data Provenance: Not applicable. The "data" refers to measurements from the device itself during controlled testing, not data derived from human subjects or retrospective/prospective studies on patient outcomes.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

    • Number of Experts: Not applicable. Ground truth for clinical performance was not established by experts in this context.
    • Qualifications of Experts: Not applicable.

    4. Adjudication Method for the Test Set

    • Adjudication Method: Not applicable. No human readers or adjudicators were involved in evaluating the device's technical performance against a ground truth.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    • Was an MRMC study done? No. The submission describes technical and performance testing of the device itself, not a study evaluating human reader performance with or without AI assistance.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    • Was a standalone study done? This question is generally relevant for AI/software devices. For this physical medical device (ultrasound and massager), a "standalone" study in the AI sense is not applicable. The performance testing was of the device operating on its own.

    7. Type of Ground Truth Used

    • Type of Ground Truth: The ground truth for the device's physical performance was based on established engineering standards (IEC) for safety and electromagnetic compatibility, and a specific temperature threshold (at least 40° C) for tissue heating as a measure of therapeutic efficacy for ultrasonic diathermy. This could be considered measured physical properties and engineering standards. It was not pathology, expert consensus on clinical cases, or outcomes data.

    8. Sample Size for the Training Set

    • Sample Size: Not applicable. This device does not use an "AI" or "machine learning" algorithm that requires a training set. The device's function is mechanical and ultrasonic.

    9. How the Ground Truth for the Training Set Was Established

    • How Ground Truth Established: Not applicable, as there is no training set for an AI algorithm.
    Ask a Question

    Ask a specific question about this device

    K Number
    K062312

    Validate with FDA (Live)

    Date Cleared
    2006-10-16

    (69 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Axiom Intense Pulsed Light System is a medical device emitting light radiation in the range from 590 nm to 1200 nm. It is intended for use in:

    • Removal of unwanted hair from all skin types and to effect stable long-term or permanent, hair reduction
    • Treatment of mild to moderate inflammatory acne (acne vulgaris) .
    • Treatment of benign pigmented epidermal and cutaneous lesions including warts, scars, striae, lentigines, nevi, melasma and cafè-au-lait
    Device Description

    The Axiom Intense Pulsed Light System is a medical device emitting light radiation in the range from 590 mm to 1200 mm. The Axion Intense Fullood Eight Oyel Cyell is based on a quick power discharge of Capacitors in a Xenon ramp, mountou on a hanspires include the charge intense Fulsed Eight. The four philopal pairol panel, and a handpiece with exchangeable lamp box

    AI/ML Overview

    The provided text is a 510(k) summary for the AXIOM Intense Pulsed Light System. It states explicitly that no performance data was required or submitted for this Class II device by the Food and Drug Administration (Office of Device Evaluation).

    Therefore, the following information cannot be extracted from the document:

    • A table of acceptance criteria and the reported device performance: No acceptance criteria or performance data is presented.
    • Sample sized used for the test set and the data provenance: No test set or data provenance is mentioned.
    • Number of experts used to establish the ground truth for the test set and the qualifications of those experts: No ground truth establishment is described.
    • Adjudication method for the test set: No adjudication method is mentioned.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: No MRMC study was conducted.
    • If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: The device is an Intense Pulsed Light System, not an algorithm, so this is not applicable.
    • The type of ground truth used (expert consensus, pathology, outcomes data, etc): No ground truth is mentioned.
    • The sample size for the training set: No training set is mentioned.
    • How the ground truth for the training set was established: No ground truth for a training set is mentioned.

    The document focuses on demonstrating substantial equivalence to predicate devices (Flash1, Med FlashII, SpaTouch, IPL Quantum SR, Clareon Pulsed Light System, Photosilk Plus) based on shared indications for use and similar technological characteristics, controls, displays, and light sources. The FDA concluded that the device is substantially equivalent, allowing it to be marketed.

    Ask a Question

    Ask a specific question about this device

    K Number
    K053041

    Validate with FDA (Live)

    Device Name
    MED SCULPT
    Date Cleared
    2006-08-02

    (279 days)

    Product Code
    Regulation Number
    890.5300
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MED Sculpt Computerized Body Massager and M-Sonic Ultrasound Diathermy is indicated for:
    a. Therapeutic Massager:
    i. Provides temporary relief of minor muscle aches and pains . -
    ii . Relieves muscle spasms
    iii. Temporarily improves local blood circulation
    iv. Temporarily reduces the appearance of cellulite
    b. Ultrasonic Diathermy:
    i. Relief of pain
    ii. Muscle spasms
    III. Joint contractures
    iv. NOT for the treatment of malignancies

    Device Description

    The MED SCULPT Computerized Body Massager and M-Sonic Ultrasound Diathermy system has massage and ultrasound diathermy components. The massage portion of the device uses a specially designed membrane, made up of elastomeric material. This membrane, fastened onto the handbiece, moves in an alternate way due to changeable depression created by a vacuum pump; such movement is due to the opening and closing of two electronically controlled electrovalves. The unit also includes an ultrasonic handpiece that can be used to perform ultrasound diathermy. Ultrasound is a particular form of mechanical energy which leads to an increase in the molecular vibration of the tissue resulting in heat generation and according to this principle the ultrasonic handpiece of the unit, using an ultrasound wave at the fixed frequency of 3 MHz, produces thermal changes in the areas of the body where it is applied.

    AI/ML Overview

    Here's an analysis of the provided 510(k) summary regarding the MED Sculpt Computerized Body Massager and M-Sonic Ultrasound Diathermy, focusing on acceptance criteria and study information:

    Based on the provided document, there is no performance data or clinical study detailed. The 510(k) summary explicitly states under "6. Performance Data": "None presented."

    This type of 510(k) submission typically relies on substantial equivalence to a predicate device without requiring new clinical performance data if the new device's technology and intended use are similar enough to an already legally marketed device. In this instance, the MED Sculpt device is deemed substantially equivalent to the Dermosonic manufactured by Sybaritic, Inc. (K024307).

    Therefore, most of the questions regarding acceptance criteria, study design, and performance metrics cannot be answered from the provided text.


    1. Table of Acceptance Criteria and Reported Device Performance

    Not applicable. The document explicitly states "None presented" for performance data. There are no acceptance criteria or reported device performance metrics provided in this 510(k) summary. The submission relies on substantial equivalence to a predicate device.


    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable. No test set was used to generate performance data for this submission.


    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    Not applicable. No test set requiring ground truth establishment was conducted or reported.


    4. Adjudication Method for the Test Set

    Not applicable. No test set requiring adjudication was conducted or reported.


    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was Done, and Effect Size

    No. A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done or reported in this 510(k) summary.


    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This device is a physical medical device (massager and ultrasound diathermy), not an algorithm or AI system, so the concept of standalone algorithm performance is not relevant.


    7. The Type of Ground Truth Used

    Not applicable. No ground truth data was used, as no performance study was conducted or reported.


    8. The Sample Size for the Training Set

    Not applicable. This device is a physical medical device, not an AI or algorithm-based system that uses a training set.


    9. How the Ground Truth for the Training Set was Established

    Not applicable. As a physical device, there is no training set or ground truth establishment relevant to an algorithm's training.


    In summary: The provided 510(k) submission for the MED Sculpt Computerized Body Massager and M-Sonic Ultrasound Diathermy relies on the concept of substantial equivalence to a predicate device (Dermosonic, K024307) rather than presenting new performance data or clinical studies for the device itself. Therefore, the document does not contain information about acceptance criteria, study design, sample sizes, expert involvement, or ground truth establishment for performance data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K051508

    Validate with FDA (Live)

    Date Cleared
    2005-10-05

    (120 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    All
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The MED flash II Intense Pulsed Light System is indicated for use in:

    • Removal of unwanted hair from all skin types and to effect stable long-term or permanent, hair reduction
    • Treatment of inflammatory acne (acne vulgaris)
    • Treatment of benign pigmented epidermal and cutaneous lesions including warts, scars, striae, lentigines, nevi, melasma and café-au-lait
    • Treatment of vascular lesions, including port wine stains, hemangiomas, angiomas, telangiectasias, rosacea, facial and leg veins
    • The integrated cooling handpiece is intended to provide pre-cooling of the epidermis, to reduce thermal injury to the epidermis, and to reduce patient pain, & discomfort associated with light applications.
    Device Description

    The MED flash II Intense Pulsed Light System is a medical device emitting light radiation in the range from 590 nm to 1200 nm. The system is based on a quick power discharge of capacitors in a Xenon lamp, mounted on a handpiece. This generates a rapid and Intense Pulsed Light.
    The Med flash II system contains five principal components:

    • Charge system of the capacitors .
    • Electronic control system .
    • Control panel .
    • Handpiece with exchangeable lamp box ●
    • Handpiece cooling system .
    AI/ML Overview

    Here's an analysis of the provided 510(k) summary regarding the acceptance criteria and supporting study for the MED flash II Intense Pulsed Light System:

    1. Table of Acceptance Criteria and Reported Device Performance

    Performance Measure/Acceptance CriteriaReported Device PerformanceComments
    Specific Performance DataNone PresentedThe 510(k) summary explicitly states "None presented" under the "Performance Data" section.
    Premarket Equivalence to Predicate DevicesSubstantially equivalent to (K041086, K033549), Palomar's EsteLux V (K040081), Palomar's EsteLux G (K020941), Lumenis' IPL Quantum SR and the Real Time Chiller (K020839), Novalis Medical's Clareon Pulsed Light System (K043319), Cutera's Optional Pulsed Light Hand Piece Family (K050047), Cynosure's Photosilk and Photosilk Plus (K041095) and General Project's Flash 1 (K022583).The device's "acceptance criteria" are implicitly met by demonstrating substantial equivalence to a combination of several predicate devices. This means that if the predicate devices were deemed safe and effective for their indications, and the subject device is similar enough, then the subject device is also considered safe and effective.
    Indications for Use (Implicit Acceptance)- Removal of unwanted hair from all skin types and to effect stable long-term or permanent, hair reduction - Treatment of inflammatory acne (acne vulgaris) - Treatment of benign pigmented epidermal and cutaneous lesions including warts, scars, striae, lentigines, nevi, melasma and café-au-lait - Treatment of vascular lesions, including port wine stains, hemangiomas, angiomas, telangiectasias, rosacea, facial and leg veins - Integrated cooling handpiece intended to provide pre-cooling of the epidermis, to reduce thermal injury to the epidermis, and to reduce patient pain, & discomfort associated with light applications.The device is deemed acceptable for these indications based on its substantial equivalence to predicate devices that have similar indications and technological characteristics.
    Technical Specifications (Implicit Acceptance for Equivalence)- Emitting light radiation in the range from 590 nm to 1200 nm. - Based on quick power discharge of capacitors in a Xenon lamp. - Handpiece with exchangeable lamp box. - Handpiece cooling system.These technical specifications are presented in the device description, implying that they align with or are sufficiently similar to the predicate devices to support substantial equivalence.

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states "None presented" under the "Performance Data" section. This indicates that no specific test set data, sample size, or information on data provenance (country of origin, retrospective/prospective) was provided or required for this 510(k) submission. The approval was based on demonstrating substantial equivalence to predicate devices, not on new clinical performance data from a specific test set.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

    Not applicable. Since no specific performance data or test set was presented, there was no ground truth to be established by experts in the context of this 510(k) summary.

    4. Adjudication Method for the Test Set

    Not applicable. As no test set data was presented, no adjudication method was mentioned or required.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. The document explicitly states "None presented" for performance data. There is no mention of an MRMC study, or any comparative effectiveness study with or without AI assistance. This type of study would typically be part of a robust performance data section, which is absent here.

    6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) Was Done

    Not applicable/No. This device is an Intense Pulsed Light System, a physical medical device, not a software algorithm. Therefore, "standalone" algorithm performance without human interaction is not relevant to its type. The 510(k) summary does not mention any software or AI components that would require such a study.

    7. The Type of Ground Truth Used

    No specific ground truth type was used or reported in the context of a new performance study. The basis for clearance is substantial equivalence to predicate devices. This implies that the safety and effectiveness of the device are inferred from the established safety and effectiveness of the predicate devices.

    8. The Sample Size for the Training Set

    Not applicable. As the approval is based on substantial equivalence to existing predicate devices and no new performance study data was presented, there are no references to a training set for an algorithm.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. Since no training set was mentioned in this 510(k) summary, there is no information on how a ground truth for a training set would have been established.

    Ask a Question

    Ask a specific question about this device

    K Number
    K004014

    Validate with FDA (Live)

    Device Name
    MD 30 COMPACT
    Manufacturer
    Date Cleared
    2001-03-16

    (84 days)

    Product Code
    Regulation Number
    878.4810
    Age Range
    N/A
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use
    Device Description
    AI/ML Overview
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1