Search Filters

Search Results

Found 3 results

510(k) Data Aggregation

    K Number
    K223523

    Validate with FDA (Live)

    Device Name
    Sonic DL
    Date Cleared
    2023-05-30

    (188 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Sonic DL is a Deep Learning based image reconstruction technique that is available for use on GE Healthcare 1.5T and 3.0T MR systems. Sonic DL reconstructs MR images from highly under-sampled data, and thereby enables highly accelerated acquisitions. Sonic DL is intended for cardiac imaging, and for patients of all ages.

    Device Description

    Sonic DL is a new software feature intended for use with GE Healthcare MR systems. It consists of a deep learning based reconstruction algorithm that is applied to data from MR cardiac cine exams obtained using a highly accelerated acquisition technique.

    Sonic DL is an optional feature that is integrated into the MR system software and activated through a purchasable software option key.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study details for the Sonic DL device, based on the provided document:

    Sonic DL Acceptance Criteria and Study Details

    1. Table of Acceptance Criteria and Reported Device Performance

    The document describes the performance of Sonic DL in comparison to conventional ASSET Cine images. While explicit numerical acceptance criteria for regulatory clearance are not stated, the studies aim to demonstrate non-inferiority or superiority in certain aspects. The implicit acceptance criteria are:

    • Diagnostic Quality: Sonic DL images must be rated as being of diagnostic quality.
    • Functional Measurement Agreement: Functional cardiac measurements (e.g., LV volumes, EF, CO) from Sonic DL images must agree closely with those from conventional ASSET Cine images, ideally within typical inter-reader variability.
    • Reduced Scan Time: Sonic DL must provide significantly shorter scan times.
    • Preserved Image Quality: Image quality must be preserved despite higher acceleration.
    • Single Heartbeat Imaging (Functional): Enable functional imaging in a single heartbeat.
    • Rapid Free-Breathing Functional Imaging: Enable rapid functional imaging without breath-holds.
    Implicit Acceptance CriterionReported Device Performance
    Diagnostic Quality"on average the Sonic DL images were rated as being of diagnostic quality" (second reader study).
    Functional Measurement Agreement"the inter-method variability (coefficient of variability comparing functional measurements taken with Sonic DL images versus measurements using the conventional ASSET Cine images) was smaller than the inter-observer intra-method variability for the conventional ASSET Cine images for all parameters, indicating that Sonic DL is suitable for performing functional cardiac measurements" (first reader study). "Functional measurements using Sonic DL 1 R-R free breathing images from 10 subjects were compared to functional measurements using the conventional ASSET Cine breath hold images, and showed close agreement" (additional clinical testing for 1 R-R free breathing).
    Reduced Scan Time"providing a significant reduction in scan time compared to the conventional ASSET Cine images" (second reader study). "the Sonic DL feature provided significantly shorter scan times than the conventional Cine imaging" (overall conclusion).
    Preserved Image Quality"capable of reconstructing Cine images from highly under sampled data that are similar to the fully sampled Cine images in terms of image quality and temporal sharpness" (nonclinical testing). "the image quality of 13 Sonic DL 1 R-R free breathing cases was evaluated by a U.S. board certified radiologist, and scored higher than the corresponding conventional free breathing Cine images from the same subjects" (additional clinical testing for 1 R-R free breathing).
    Single Heartbeat Functional Imaging"Sonic DL is capable of achieving a 12 times acceleration factor and obtaining free-breathing images in a single heartbeat (1 R-R)" (additional clinical testing).
    Rapid Free-Breathing Functional Imaging"Sonic DL is capable of... obtaining free-breathing images in a single heartbeat (1 R-R)" (additional clinical testing).

    2. Sample Size Used for the Test Set and Data Provenance

    The document describes two primary reader evaluation studies and additional clinical testing.

    • First Reader Study (Functional Measurements):
      • Sample Size: 107 image series from 57 unique subjects (46 patients, 11 healthy volunteers).
      • Data Provenance: Data from 7 sites: 2 GE Healthcare facilities and 5 external clinical collaborators. This indicates data from multiple sources, likely a mix of prospective and retrospective collection. The geographic origin of these sites is not explicitly stated but implies a multi-center study potentially from different countries where GE Healthcare operates or collaborates.
    • Second Reader Study (Image Quality Assessment):
      • Sample Size: 127 image sets, which included a subset of the subjects from the first study.
      • Data Provenance: Same as the first reader study (clinical sites and healthy volunteers at GE Healthcare facilities).
    • Additional Clinical Testing (1 R-R Free Breathing):
      • Functional Measurements: 10 subjects.
      • Image Quality Evaluation: 13 subjects.
      • Data Provenance: In vivo cardiac cine images from 19 healthy volunteers. This implies prospective collection or a subset of prospectively collected healthy volunteer data.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

    • First Reader Study (Functional Measurements): Three radiologists. Qualifications are not explicitly stated, but their role in making quantitative measurements implies expertise in cardiac MRI.
    • Second Reader Study (Image Quality Assessment): Three radiologists. Qualifications are not explicitly stated, but their role in blinded image quality assessments implies expertise in cardiac MRI interpretation.
    • Additional Clinical Testing (1 R-R Free Breathing Image Quality): One U.S. board certified radiologist.

    4. Adjudication Method for the Test Set

    The document does not explicitly state an adjudication method (like 2+1, 3+1, or none) for either the functional measurements or the image quality assessments. For the first study, it mentions "inter-method variability" and "inter-observer intra-method variability," suggesting that the readings from the three radiologists were compared against each other and against the conventional method, but not necessarily adjudicated to establish a single "ground truth" per case. For the second study, "blinded image quality assessments" were performed, and ratings were averaged, but no adjudication process is described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and the effect size

    A clear MRMC comparative effectiveness study, in the sense of measuring human reader improvement with AI vs. without AI assistance, is not explicitly described.

    The studies compare the performance of Sonic DL images (algorithm output) against conventional images, with human readers evaluating both.

    • The first reader study compares quantitative measurements from Sonic DL images to conventional images, indicating suitability for performing functional cardiac measurements by showing smaller inter-method variability than inter-observer intra-method variability for conventional images. This suggests Sonic DL is at least as reliable as the variability between conventional human measurements.
    • The second reader study involves blinded image quality assessments of both conventional and Sonic DL images, confirming that Sonic DL images were rated as diagnostic quality.
    • The additional clinical testing for 1 R-R free breathing shows that Sonic DL images were "scored higher than the corresponding conventional free breathing Cine images" by a U.S. board-certified radiologist.

    These are comparisons of the image quality and output from the AI system versus conventional imaging, interpreted by readers, rather than measuring human reader performance assisted by the AI system.

    Therefore, the effect size of how much human readers improve with AI vs. without AI assistance is not provided because the studies were designed to evaluate the image output quality and measurement agreement of the AI-reconstructed images themselves, not to assess an AI-assisted workflow for human readers.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Yes, standalone performance was assessed for image quality metrics.

    • Nonclinical Testing: "Model accuracy metrics such as Peak-Signal-to-Noise (PSNR), Root-Mean-Square Error (RMSE), Structural Similarity Index Measure (SSIM), and Mean Absolute Error (MAE) were used to compare simulated Sonic DL images with different levels of acceleration and numbers of phases to the fully sampled images." This is a standalone evaluation of the algorithm's output quality against a reference.
    • In Vivo Testing: "model accuracy and temporal sharpness evaluations were conducted using in vivo cardiac cine images obtained from 19 health volunteers." This is also a standalone technical evaluation of the algorithm's output on real data.

    7. The Type of Ground Truth Used

    • Nonclinical Testing (Simulated Data): The ground truth was the "fully sampled images" generated from an MRXCAT phantom and a digital phantom.
    • Clinical Testing (Reader Studies):
      • Functional Measurements: The "ground truth" for comparison was the measurements taken from the "conventional ASSET Cine images." The variability of these conventional measurements across readers also served as a baseline for comparison. This is a form of clinical surrogate ground truth (comparing to an established accepted method).
      • Image Quality Assessments: The "ground truth" was the expert consensus/opinion of the radiologists during their blinded assessments of diagnostic quality.
      • Additional Clinical Testing (1 R-R Free Breathing): Functional measurements were compared to "conventional ASSET Cine breath hold images" (clinical surrogate ground truth). Image quality was based on the scoring by a "U.S. board certified radiologist" (expert opinion).

    No pathology or outcomes data were used as ground truth. The ground truth in the clinical setting was primarily based on established imaging techniques (conventional MR) and expert radiologist assessments.

    8. The Sample Size for the Training Set

    The document does not explicitly state the sample size for the training set used for the deep learning model. It only describes the data used for testing the device.

    9. How the Ground Truth for the Training Set Was Established

    Since the training set size is not provided, the method for establishing its ground truth is also not described in the provided text. Typically, for deep learning reconstruction, the "ground truth" for training often involves fully sampled or high-quality reference images corresponding to the undersampled input data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K213717

    Validate with FDA (Live)

    Device Name
    AIR Recon DL
    Date Cleared
    2022-06-08

    (196 days)

    Product Code
    Regulation Number
    892.1000
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    AIR Recon DL is a deep learning based reconstruction technique that is available for use on GE Healthcare 1.5T, 3.0T, and 7.0T MR systems. AIR Recon DL reduces noise and ringing (truncation artifacts) in MR images, which can be used to reduce scan time and improve image quality. AIR Recon DL is intended for use with all anatomies, and for patients of all ages. Depending on the anatomy of interest being imaged, contrast agents may be used.

    Device Description

    AIR Recon DL is a software feature intended for use with GE Healthcare MR systems. It is a deep learning based reconstruction technique that removes noise and ringing (truncation) artifacts from MR images. AIR Recon DL is an optional feature that is integrated into the MR system software and activated through a purchasable software option key.

    AI/ML Overview

    Here's a breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided FDA 510(k) summary for AIR Recon DL:

    1. Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implicitly defined by the desired improvements and non-inferiority claims of the AIR Recon DL feature. The reported device performance demonstrates how these criteria were met.

    Acceptance Criteria CategorySpecific Criteria (Implicit from Claims)Reported Device Performance (as stated in the document)
    Image Quality - Noise ReductionEquivalent or better Apparent Signal to Noise Ratio (SNR)133 out of 133 cases showed equivalent or better apparent SNR.
    Image Quality - SharpnessEquivalent or better sharpness133 out of 133 cases showed equivalent or better sharpness.
    Image Quality - Lesion ConspicuityEquivalent or better lesion conspicuity for pathological cases123 out of 124 cases with pathology showed equivalent or better lesion conspicuity.
    Impact on Quantitative MeasurementsDoes not adversely affect accuracy of quantitative measurements (e.g., contrast pharmacokinetics, lesion sizes, brain volumetry).Strong agreement between measurements made using conventional and AIR Recon DL images.
    Scan Time ReductionImage quality maintained or improved even with reduced scan time.For 22 image pairs with shorter scan times (AIR Recon DL) vs. longer scan times (conventional), AIR Recon DL images were rated as better or equivalent image quality in all cases.
    ArtifactsDoes not significantly change the appearance of motion artifacts.Sample images show AIR Recon DL does not significantly change the appearance of motion artifacts.
    Overall Radiologist PreferenceRadiologists prefer AIR Recon DL images over conventional images.Radiologists preferred AIR Recon DL images over conventional images in 99% of evaluations.
    Non-clinical Performance (Phantoms)Improved SNR, sharpness; maintained low contrast detectability; ADC maps not adversely impacted.Nonclinical testing passed defined acceptance criteria; demonstrated improved SNR and sharpness, maintained low contrast detectability, and no adverse impact on ADC maps.

    2. Sample Size and Data Provenance

    • Test Set Sample Size: 133 cases.
      • 129 patient cases
      • 4 healthy subjects
    • Data Provenance:
      • Country of Origin: Not explicitly stated, but "10 different clinical sites" suggests a multi-center study, and "a GE Healthcare facility" could indicate a US or international site. Given the FDA submission, it's highly likely to include US data.
      • Retrospective or Prospective: Not explicitly stated, but the description "images acquired across a variety of pulse sequences and anatomies" and involvement of "10 different clinical sites" could imply a retrospective collection for the reader study, where images were pre-collected. However, without explicit mention, it's not definitive. The phrasing "acquired from the same acquired raw data" suggests a paired comparison based on existing data.

    3. Number of Experts and Qualifications

    • Number of Experts: Three radiologists.
    • Qualifications: "Radiologists" implies medical doctors specialized in radiology. No further specifics on years of experience or subspecialty were provided.

    4. Adjudication Method for the Test Set

    The adjudication method appears to be 2+1 (or 3/3 agreement is ideal, but 2 out of 3 for consensus is common).
    "Each image pair was evaluated independently by three radiologists."
    "The results confirmed that the AIR Recon DL feature provides images with equivalent or better image quality in terms of apparent signal to noise ratio (133 out of 133 cases), sharpness (133 out of 133 cases), and lesion conspicuity (123 out of 124 cases with pathology)."
    "The radiologists reading the images also indicated a preference for the AIR Recon DL images over conventional images in 99% of the evaluations."
    This indicates that the claims are based on the collective agreement or majority opinion of the three readers for each case. The exact decision rule (e.g., simple majority, unanimous) is not stated, but the high consistency (e.g., 133/133, 123/124) implies strong agreement or effective resolution.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    Yes, a multi-reader multi-case (MRMC) comparative effectiveness study was done.

    • Comparison: Radiologists compared AIR Recon DL images to conventional images (without AIR Recon DL) reconstructed from the same acquired raw data.
    • Effect Size of Human Reader Improvement: The study demonstrates a significant preference and improvement in perceived image quality attributes by human readers when using AIR Recon DL assisted images compared to conventional images.
      • Preference: Radiologists preferred AIR Recon DL images over conventional images in 99% of evaluations. This indicates a strong positive effect on reader perception and diagnostic confidence.
      • Qualitative Improvement:
        • Equivalent or better SNR in 100% of cases (133/133).
        • Equivalent or better sharpness in 100% of cases (133/133).
        • Equivalent or better lesion conspicuity in ~99.2% of pathological cases (123/124).
      • Enablement of Shorter Scans: For shorter scan time acquisitions, AIR Recon DL images were rated as better or equivalent image quality in 100% of 22 image pairs, which suggests human readers are able to maintain or even improve their assessment quality despite reduced acquisition time. Overall, the effect size is very large and consistently positive across all measured subjective criteria.

    6. Standalone (Algorithm Only) Performance

    The document describes "nonclinical testing" on phantoms, which represents a form of standalone testing where the algorithm's output is directly measured against predefined physical criteria:

    • "AIR Recon DL has undergone phantom testing to evaluate the feature and its impact on image quality, including SNR, sharpness, and low contrast detectability."
    • "The nonclinical testing demonstrated that AIR Recon DL does improve SNR and image sharpness while maintaining low contrast detectability."
    • "ADC maps were not adversely impacted by the use of AIR Recon DL."
      This evaluates the algorithm's effect on image characteristics absent human interpretation of clinical cases.

    7. Type of Ground Truth Used

    The study primarily used expert consensus (radiologist agreement) as the ground truth for evaluating image quality attributes (SNR, sharpness, lesion conspicuity, overall preference) and the impact of the algorithm.
    For the "presence of pathology" in the test set, it's assumed that this was either identified beforehand from clinical reports or pathology (e.g., biopsy results) or by consensus among the evaluating radiologists prior to their evaluation of the AI-enhanced images. However, the exact source of ground truth for pathology presence/absence isn't detailed, only that 124 cases "with pathology" and 9 cases "without pathology" were included.

    8. Sample Size for the Training Set

    The document states: "Both the proposed AIR Recon DL and the predicate device use neural networks that have similar architecture, and were trained using similar methods and data."
    However, the specific sample size for the training set is NOT provided in this summary.

    9. How Ground Truth for Training Set was Established

    The document states: "Both the proposed AIR Recon DL and the predicate device use neural networks that have similar architecture, and were trained using similar methods and data."
    The method for establishing ground truth for the training set is NOT explicitly detailed. Typically, for deep learning reconstruction, the "ground truth" during training often refers to high-quality, fully sampled MR images (or images from a prior, high-quality reconstruction method) that the AI attempts to match or improve upon, rather than a clinical diagnosis per se. The goal during training would be to generate images that are less noisy and sharper while preserving underlying anatomical and pathological information, learned by comparing "corrupted" (e.g., undersampled, noisy) inputs to "clean" reference images.

    Ask a Question

    Ask a specific question about this device

    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The system is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission projection data from the same axial plane taken at different angles. The systems with 160 mm detector coverage have the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, liver, kidney, pancreas, etc. The system may acquire data using Axial, Cardiac, and Gated CT scan techniques from patients of all ages. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.

    This device may include data and image processing to produce images in a variety of trans-axial and reformatted planes. Further, the images can be post processed to produce additional imaging planes or analysis results.

    The system is indicated for head, whole body, cardiac, and vascular X-ray Computed Tomography applications.

    The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.

    If the spectral imaging option is included on the system can acquire CT images using different kV levels of the same anatomical region of a patient in a single source. The differences in the energy dependence of the attenuation coefficient of the different materials provide information about the chemical composition of body materials. This approach enables images to be generated at energies selected from the available spectrum to visualize and analyze information about anatomical and pathological structures.

    GSI provides information of the chemical composition of renal calculi by calculation and graphical display of the spectrum of effective atomic number. GSI Kidney stone characterization provides additional information to aid in the characterization of uric acid versus nonuric acid stones. It is intended to be used as an adjunct to current standard methods for evaluating stone etiology and composition.

    Device Description

    The Revolution CT family of products, including Revolution CT, Revolution CT ES, Revolution CT with Apex edition, Revolution CT ES with Apex edition Apex, Revolution Apex Elite, Revolution Apex Plus, and Revolution Apex Select, Revolution CT Power, Revolution Apex Pro are multi-slice CT scanners consisting of a gantry, patient table, scanner desktop (operator console), system cabinet, power distribution unit (PDU), and interconnecting cables. The system includes image acquisition hardware, image acquisition and reconstruction software, and associated accessories.

    GE has modified the cleared Revolution CT (K19177) within our design controls to include the 0.23 s/rot option. The 0.23s/rot option can be used with axial scan acquisitions and is especially beneficial during certain cardiac scan acquisitions. The scan workflow and user interface remain identical to the of the predicate device, with the exception that the user now has the option to select 0.23 s/rot in addition to other gantry rotation speeds.

    The addition of a new maximum gantry rotation speed leads to updates to system performance claims about maximum temporal resolution when combined with the optional Snapshot Freeze 2 (K183161) feature.

    AI/ML Overview

    The provided text does not contain detailed acceptance criteria for the device or a study explicitly proving the device meets said criteria. The document is a 510(k) premarket notification for a Computed Tomography (CT) system (Revolution CT family) that includes the addition of a 0.23 s/rot option. The focus of the submission is to demonstrate substantial equivalence to a predicate device (Revolution CT, K191777), not to present an independent study with explicit acceptance criteria for a novel device performance claim.

    However, based on the information provided, we can infer some aspects related to evaluating the new 0.23 s/rot option:

    Inferred Acceptance Criteria and Reported Device Performance (Focusing on the 0.23 s/rot option)

    Acceptance Criteria (Inferred)Reported Device Performance (Inferred from context)
    Maintain safety and effectiveness of the predicate device (Revolution CT, K191777)The device is deemed "as safe and effective" and performs in a "substantially equivalent manner" to the predicate device.
    Compliance with relevant standards (IEC 60601-1 Ed. 3.1, 21CFR Subchapter J, NEMA XR-25, XR-26, and XR-28)"Completed testing and is in compliance with IEC 60601-1 Ed. 3.1 and its associated collateral and particular standards, 21CFR Subchapter J, and NEMA XR-25, XR-26, and XR-28."
    Successful verification through quality system and engineering testing"Successfully completed all testing per our quality system. The risk management and design verification activities did not raise any new questions about safety and effectiveness."
    Technical substantiation of 0.23 s/rot performance claims (especially for maximum temporal resolution combined with Snapshot Freeze 2)"Performance evaluation testing included use of a cardiac phantom and mathematical modeling to provide technical substantiation of the 0.23 s/rot performance claims... Various mathematical and statistical analyses were performed to demonstrate that each performance item was successfully verified and substantiated."

    Detailed Breakdown of Study Information (Focusing on the 0.23 s/rot option):

    1. Sample size used for the test set and the data provenance:

      • Test Set Sample Size: Not explicitly stated by number of cases or patients. The non-clinical testing involved "a cardiac phantom" and "mathematical modeling." This implies a phantom-based study for physical performance and computational analysis.
      • Data Provenance: The study was non-clinical and conducted by the manufacturer, GE Medical Systems, LLC. It's an internal validation of the new feature.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in a way that aligns with "ground truth" for a clinical test set, as no clinical study was performed. The evaluation relied on technical performance metrics validated against engineering standards and a cardiac phantom, rather than expert interpretation of medical images.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable as there was no clinical test set requiring expert adjudication. The substantiation was based on physical measurements from a phantom and mathematical modeling.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No MRMC comparative effectiveness study was done. This submission is for a hardware/software update to a CT scanner, not an AI-assisted diagnostic tool.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • The evaluation focused on the standalone performance of the CT system with the new gantry rotation speed. The "algorithm" here refers to the system's operational parameters and image reconstruction, rather than a diagnostic AI algorithm. The performance items were verified and substantiated through phantom studies and mathematical analysis.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For the non-clinical testing, the "ground truth" was likely established by:
        • Physical measurements/known properties of the cardiac phantom: For evaluating temporal resolution and other imaging characteristics.
        • Engineering specifications and mathematical models: To predict and verify performance parameters.
    7. The sample size for the training set:

      • Not applicable in the context of this submission. The device is a CT scanner with a hardware and software update, not a machine learning algorithm that requires a specific "training set" of patient data in the typical sense. The development involved general engineering design, development, and verification processes.
    8. How the ground truth for the training set was established:

      • Not applicable as there was no distinct "training set" for a machine learning model. The system's design and engineering would be based on established physics, engineering principles, and prior knowledge from existing CT technology, rather than a labeled training dataset.

    Summary of the Study:

    The study was a non-clinical performance evaluation conducted by GE Medical Systems, LLC. It involved:

    • Engineering testing: To ensure compliance with electrical, mechanical, and safety standards (IEC 60601-1 Ed. 3.1, 21CFR Subchapter J, NEMA XR-25, XR-26, and XR-28).
    • Performance evaluation testing: Utilized a cardiac phantom and mathematical modeling to technically substantiate claims related to the 0.23 s/rot option, particularly in conjunction with the Snapshot Freeze 2 feature for temporal resolution.
    • Quality System Compliance: The development followed the Quality System Regulations of 21CFR 820 and ISO 13485, including risk analysis, technical reviews, design reviews, code inspections, and various levels of testing (unit, integration, performance, safety, simulated use).

    The conclusion was that the device, with the new 0.23 s/rot option, is "as safe and effective, and performs in a substantially equivalent manner to the predicate device Revolution CT (K191777)." No clinical testing was deemed necessary because the modifications were fully testable on an engineering bench.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1