510(k) Data Aggregation

    K Number
    K253355

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-10-29

    (29 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Difiney Advanced Anti Snoring Device 4.0 is intended to help reduce snoring in people 18 years of age or older.

    Device Description

    The Difiney Advanced Anti Snoring Device 4.0 is designed to reduce snoring by gently repositioning the lower jaw forward, which opens the airway and allows for smoother airflow during sleep. By adjusting the position of the lower jaw, the device helps prevent the throat tissues from collapsing and blocking airflow, which is a common cause of snoring.

    The 4.0 version is customizable, allowing users to set the lower jaw advancement up to 6mm, providing a personalized fit for maximum comfort and effectiveness. This forward positioning also reduces vibrations in the airway, which are often responsible for the snoring sound.

    AI/ML Overview

    N/A

    Ask a Question

    Ask a specific question about this device

    K Number
    K250482

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2025-05-08

    (78 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Advanced Anti Snoring Device 4.0 is intended to help reduce snoring in people 18 years of age or older.

    Device Description

    The Advanced Anti Snoring Device 4.0 is designed to reduce snoring by gently repositioning the lower jaw forward, which opens the airway and allows for smoother airflow during sleep. By adjusting the position of the lower jaw, the device helps prevent the throat tissues from collapsing and blocking airflow, which is a common cause of snoring. The 4.0 version is customizable, allowing users to set the lower jaw advancement up to 6mm, providing a personalized fit for maximum comfort and effectiveness. This forward positioning also reduces vibrations in the airway, which are often responsible for the snoring sound.

    AI/ML Overview

    The provided FDA 510(k) clearance letter and summary for the Advanced Anti Snoring Device 4.0 (K250482) explicitly state "Performance Testing - Clinical: Not Applicable." This indicates that no human clinical studies were performed to demonstrate the device's performance in reducing snoring.

    Therefore, based solely on the provided document, I cannot fulfill most of your request as the information regarding clinical performance, expert ground truth, sample sizes for test sets, MRMC studies, or training sets is not present. The device's clearance is based on substantial equivalence to a predicate device (K233850) through non-clinical testing.

    However, I can extract the information related to acceptance criteria (implied by the comparison to the predicate) and the non-clinical studies performed.

    Here's a breakdown of what can be inferred and what cannot be provided from the given text:

    What Can Be Inferred/Extracted from the Document:

    • Acceptance Criteria (Implied by Predicate Equivalence): The primary "acceptance criteria" here is the demonstration of substantial equivalence to the predicate device (Difiney Anti Snoring Device, K233850) based on shared technological characteristics and non-clinical performance. The device must meet the regulatory requirements for an intraoral device for snoring.
    • Reported Device Performance: The device's performance is not reported in terms of clinical efficacy in reducing snoring, but rather in meeting material and mechanical specifications.
    • Non-Clinical Studies: Biocompatibility and physical properties testing.

    What Cannot Be Provided from the Document (as "Not Applicable" or Not Mentioned):

    • A table of acceptance criteria and reported device performance related to snoring reduction (as no clinical performance data is provided).
    • Sample sizes used for a test set (no clinical test set described).
    • Data provenance for a test set.
    • Number of experts used to establish ground truth for a test set.
    • Qualifications of those experts.
    • Adjudication method for a test set.
    • If a multi-reader multi-case (MRMC) comparative effectiveness study was done.
    • Effect size of human readers improving with AI vs. without AI assistance.
    • If a standalone (algorithm only) performance was done.
    • The type of ground truth used (clinical outcomes).
    • The sample size for the training set (no AI/ML algorithm mentioned requiring a training set).
    • How the ground truth for the training set was established.

    Summary Based on Provided Document:

    1. Table of Acceptance Criteria and Reported Device Performance

    Feature/TestAcceptance Criteria (Implied by Predicate Equivalence)Reported Device Performance
    Intended UseIntended to help reduce snoring in people 18 years of age or older."The Advanced Anti Snoring Device 4.0 is intended to help reduce snoring in people 18 years of age or older." (Stated in Indications for Use, demonstrating conformance to the predicate's intended use.)
    Mechanism of ActionReduces snoring by gently repositioning the lower jaw forward, opening the airway, and allowing smoother airflow during sleep, preventing throat tissues from collapsing and blocking airflow."helps reduce snoring by moving the lower jaw forward to open your airway to improve breathing allowing air to flow freely without the unhealthy and annoying sound of snoring." (Matches predicate's mechanism)
    MaterialsMust use materials comparable to the predicate (Polycarbonate resin, Ethylene vinyl acetate copolymer).Uses Polycarbonate resin and Ethylene vinyl acetate copolymer.
    Desired CharacteristicsHome use, heat sensitive/moldable, adjustable jaw advancement position.Home use, heat sensitive/moldable, adjustable jaw advancement position.
    Physical/MechanicalCustom-fitted intraoral device, repositions mandible anteriorly, reusable.Custom-fitted intraoral device, repositions mandible anteriorly, reusable. (Similar to predicate)
    Mandibular AdvancementPredicate offers up to 4mm advancement in 1mm increments. Subject device offers up to 6mm advancement in 1mm increments. (Difference assessed not to impact safety/effectiveness).Up to 6mm at 1mm increment.
    DesignPredicate has a small anterior protrusion. Subject device omits this for comfort. (Difference assessed not to affect safety/effectiveness).Small anterior protrusion present (referring to the subject device having one, but the text "Different 2" says "No anterior protrusion" for subject device and "Small anterior protrusion present" for predicate. This seems like a typo in the table itself. Assuming the text description of the difference is correct: Small anterior protrusion absent).
    Manufacturing ProcessInjection molding.Injection molding.
    SterilityNon-sterile.Non-sterile.
    BiocompatibilityMust meet ISO 10993 standards."Bio-compatibility assessment was conducted in accordance with ISO 10993-1..." and passed.
    Physical Properties of MaterialsMaterials must meet performance tests for flexural strength, flexural modulus, tensile strength, elongation at break, water absorption, melting temperature, and density.Materials passed these performance tests.

    2. Sample Size Used for the Test Set and Data Provenance:
    * N/A. The document explicitly states "Performance Testing - Clinical: Not Applicable." The clearance is based on non-clinical testing and substantial equivalence to a predicate device.
    * The non-clinical tests (biocompatibility, physical properties) are typically conducted on material samples or device prototypes, not human subjects, and thus "sample size" in the context of a clinical test set is not relevant here.
    * Data Provenance: Not applicable for clinical data. For non-clinical tests, the tests were conducted by the manufacturer, Fissiontech LLC. The country of origin for the device manufacturing is likely China, given the correspondent's location, but this is not explicitly stated for the testing location itself.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of those Experts:
    * N/A. As no clinical performance testing or "test set" for efficacy was conducted or reported, no experts were used to establish ground truth for clinical outcomes in this context.

    4. Adjudication Method for the Test Set:
    * N/A. No clinical test set or adjudication process for human data was described.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done:
    * N/A. No clinical studies were reported, and the device is not an imaging AI product that would typically undergo MRMC studies.

    6. If a Standalone Performance Study was done (e.g., algorithm only without human-in-the-loop performance):
    * N/A. The device is a physical intraoral device, not an algorithm. No standalone performance study related to AI/ML was relevant or conducted.

    7. The Type of Ground Truth Used:
    * For the non-clinical testing, the "ground truth" was established by standardized test methods and specifications (e.g., ISO 10993 for biocompatibility, and various physical property tests for material characteristics). The device's conformance to these predefined standards served as the "ground truth" for material and manufacturing quality.
    * For clinical efficacy in snoring reduction, no ground truth was established or presented in this document.

    8. The Sample Size for the Training Set:
    * N/A. This device does not incorporate AI/ML, so there is no training set.

    9. How the Ground Truth for the Training Set Was Established:
    * N/A. No training set exists.

    In conclusion, the FDA 510(k) clearance for the Advanced Anti Snoring Device 4.0 was granted based on demonstrably meeting all non-clinical design specifications and showing substantial equivalence to a legally marketed predicate device, rather than through clinical performance testing for snoring reduction.

    Ask a Question

    Ask a specific question about this device

    K Number
    K233850

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2024-04-18

    (135 days)

    Product Code
    Regulation Number
    872.5570
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Difiney Anti Snoring Device is intended to help reduce snoring in adults 18 years of age or older.

    Device Description

    The Difiney Anti Snoring device is an intraoral device that is composed of an upper and lower intraoral trays with a handle that is used during the process of customizing the trays. The Difiney Anti Shoring device repositions the jaw anteriorly relative to the maxilla in order to increase the patient's pharyngeal space to improve the ability to exchange air and decrease air turbulence, a causative factor in snoring.

    AI/ML Overview

    The provided text is a 510(k) summary for the Difiney Anti Snoring Device. It outlines the regulatory classification, intended use, and a comparison to a predicate device (SnoreRx). Crucially, the document states "Performance Testing - Clinical: Not Applicable" and "Performance Testing - Animal: Not Applicable."

    This means the submission does not include data from a clinical study or animal study to demonstrate the device's performance based on specific acceptance criteria. Instead, the substantial equivalence determination is based on technological characteristics and non-clinical testing (biocompatibility and physical properties).

    Therefore, I cannot provide the information requested in your prompt regarding acceptance criteria, device performance from a study, sample sizes, expert involvement, or MRMC studies, because this information is explicitly stated as "Not Applicable" in the provided document.

    The document's conclusion is that the device is substantially equivalent to the predicate based on technological characteristics and non-clinical testing, not on clinical performance metrics.