510(k) Data Aggregation

    K Number
    K223370

    Validate with FDA (Live)

    Date Cleared
    2023-02-13

    (101 days)

    Product Code
    Regulation Number
    878.4850
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The NeatNick Heel Safety Lancet is a single use device with a sterile blade that is used by medical professionals to obtain a blood sample from the heel of neonates and infants for subsequent diagnostic testing.

    Device Description

    The NeatNick Heel Safety Lancet is a single use, blood sampling device with a sterile blade that are used by medical professionals to obtain a blood sample from the feet of neonates and infants for diagnostic purposes. The NeatNick Lancet incorporates an integral sharps injury prevention feature that also prevents reuse of the lancet.

    The NeatNick Lancet has been in commercial distribution since 2006.

    NeatNick Lancets are available in two configurations (Preemie and Full Term) are intended to be used by medical professionals to obtain a blood sample for diagnostic testing. The lancets consist of a stainless steel blade overmolded with low density polyethylene (LDPE) which protects the blade until use and functions as a sterile barrier. The lancet is single use, disposable and is sterilized by gamma radiation to a SAL of 10-6.

    Biocompatibility testing has been conducted per ISO 10993 at a GLP testing facility. The lancet is classified as Class II.

    The lancet is currently in commercial distribution in the United States and Canada. Since commercial distribution of the current lancets began, there have been no design changes.

    AI/ML Overview

    The provided text is a 510(k) Pre-Market Notification for the NeatNick Heel Safety Lancet. This type of FDA submission focuses on demonstrating substantial equivalence to an already legally marketed device (predicate device) rather than proving general safety and effectiveness through extensive clinical trials with specific acceptance criteria and detailed study results.

    Therefore, the document does not contain the detailed information typically associated with a study proving a device meets specific acceptance criteria in the way a clinical trial or algorithm performance study would. It explicitly states, "Clinical testing is not applicable as the risk analysis confirmed that all identified risks were addressed and mitigated appropriately."

    However, based on the non-clinical testing mentioned and the comparison table, here's what can be inferred and what information is missing:

    1. Table of acceptance criteria and the reported device performance:

    The document describes "predetermined criteria" being met for non-clinical bench testing and satisfaction of "special controls (21 CFR 878.4850)". The comparison table (Table 1) essentially serves as a performance comparison against the predicate device, where "Same" for the subject device implies it meets or performs comparably to the predicate for those characteristics.

    Acceptance Criteria (Implied from Predicate Characteristics/Special Controls)Reported Device Performance (Subject Device)
    Indication for Use: Obtain blood sample from heel of neonates and infants for diagnostic testing.Same (as predicate)
    Intended Use: Single-use, by medical professionals, for blood sample from neonates and infants for diagnostic testing, with spring-actuated sterile sweeping action blade that retracts and renders device inoperable.Same (as predicate)
    Trigger Actuation Force: 8N-36NSame (as predicate)
    Length of Cut (Full term): 2.9mm-4.1mmSame (as predicate)
    Length of Cut (Preemie): 1.8mm-3.0mmSame (as predicate)
    Depth of Cut (Full term): 1.0mm-1.5mmSame (as predicate)
    Depth of Cut (Preemie): 0.65mm-1.25mmSame (as predicate)
    Biocompatibility: Conforms to ISO10993-1Same (as predicate)
    Blade Sterility: Sterilized by Gamma Radiation, SAL = 10^-6Same (as predicate)
    Mechanical Design: Verification and validation to ensure risks managed and mechanical functions suitable for shelf life.Non-clinical bench testing performed to ensure criteria met and special controls satisfied.

    Missing Information:

    • Specific quantitative acceptance limits for the "mechanical design verification and validation testing" are not provided. The document only states these tests were performed and criteria were met.
    • Actual measured values for the subject device's Trigger actuation force, Length of cut, and Depth of cut are not provided, only that they are "Same" as the predicate, which also presents a range.

    The following questions cannot be answered from the provided document as it describes a 510(k) submission for a physical medical device (lancet), not an AI/software device or a study involving human readers or ground truth derived from expert consensus/pathology. The document explicitly states "Clinical testing is not applicable".

    2. Sample size(s) used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
    Not Applicable/Not Provided: This type of information is relevant for clinical or algorithm performance studies, which were not conducted for this 510(k) submission. Non-clinical bench testing was performed, but sample sizes for these tests are not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
    Not Applicable/Not Provided: This information pertains to studies where ground truth is established by experts (e.g., in imagery or diagnostic tasks). This submission relies on engineering specifications and bench testing.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
    Not Applicable/Not Provided: This is relevant for expert-based ground truth establishment or clinical trial adjudication, neither of which occurred.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
    Not Applicable/Not Provided: This is specific to AI-assisted diagnostic tools and human reader performance studies. The device is a physical lancet.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
    Not Applicable/Not Provided: This is specific to AI algorithm performance.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
    Not Applicable/Not Provided (in the context of clinical "ground truth"): For the non-clinical testing, the "ground truth" would be the engineering specifications and established test standards (e.g., ISO 10993-1 for biocompatibility, SAL of 10^-6 for sterility, and the specified ranges for cut length/depth and actuation force).

    8. The sample size for the training set
    Not Applicable/Not Provided: This is specific to machine learning/AI models.

    9. How the ground truth for the training set was established
    Not Applicable/Not Provided: This is specific to machine learning/AI models.

    Ask a Question

    Ask a specific question about this device

    K Number
    K223099

    Validate with FDA (Live)

    Date Cleared
    2022-11-28

    (59 days)

    Product Code
    Regulation Number
    878.4850
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Facet Manatee Reusable Lancet Base (commonly referred to as a lancing device) is a non-sterile reusable device which provides a spring-loaded mechanism to quickly eject and retract a standard or regular version lancet to effect a lancing event for the purpose of obtaining a blood sample for diagnostic testing in children, adolescents (use on children and adolescents should be by or under supervision of an adult), and adults in a home setting. The device is designed to be cleaned and disinfected between uses on a single patient.

    Device Description

    The Facet Manatee Reuseable Lancet Base is a reuseable blood sampling device used in conjunction with a standard or universal lancet blade to obtain a sample of capillary blood for diagnostic purposes, primarily for blood glucose monitoring in diabetic patients.

    AI/ML Overview

    The provided text describes the Facet Manatee Reusable Lancet Base, a reusable lancing device for obtaining blood samples for diagnostic testing. The submission is a 510(k) premarket notification, which aims to demonstrate substantial equivalence to a legally marketed predicate device, rather than proving novel effectiveness. Therefore, the information typically requested regarding AI/CAD efficacy studies, such as effects on human readers, ground truth establishment for AI models, and training set details, is not applicable or present in this document.

    Here's the breakdown of the acceptance criteria and the study that proves the device meets them, based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document describes non-clinical bench testing performed to ensure the device meets predetermined criteria and satisfies special controls. While specific numerical acceptance values are not provided, the "Importance" column outlines the intended performance. The conclusion states that the tests demonstrate the candidate device is substantially equivalent to the predicate.

    Property/CharacteristicTest MethodImportance (Acceptance Criteria)Reported Performance (Implied by Conclusion)
    Endcap Removal ForceTensile strength testerEnsure cap can be removed easily to insert lancet blade.Test was performed and criteria met (device is substantially equivalent).
    Endcap Attachment ForceTensile strength testerEnsure cap will remain on lancet base during lancing event.Test was performed and criteria met (device is substantially equivalent).
    Depth of PunctureCalibrated High Speed VideoEnsure depth of puncture is repeatable for various depth settings.Test was performed and criteria met (device is substantially equivalent).
    Over-Charging ForceTensile force gaugeEnsure that device can be charged to engage actuation spring.Test was performed and criteria met (device is substantially equivalent).
    Charging ForceTensile force gaugeEnsure device can be charged to engage actuation spring.Test was performed and criteria met (device is substantially equivalent).
    Lancet Insertion ForceTensile force gaugeEnsure lancet can be easily inserted.Test was performed and criteria met (device is substantially equivalent).
    Depth Adjust TorqueTorque gaugeEnsure that depth adjustment can be easily adjusted and that adjustment will not change during use.Test was performed and criteria met (device is substantially equivalent).
    Button Activation ForceTensile force gaugeEnsure activation force is within specification.Test was performed and criteria met (device is substantially equivalent).
    Lancet RotationTorque gaugeEnsure lancet does not rotate during device actuation.Test was performed and criteria met (device is substantially equivalent).
    Drop TestSimulated Use after dropEnsure that device can withstand a 1 meter drop to a hard surface and still function.Test was performed and criteria met (device is substantially equivalent).
    Life Cycle TestSimulated useEnsure device can withstand 3 years of expected use.Test was performed and criteria met (device is substantially equivalent).
    Chemical TestCleaning and disinfection studiesEnsure device can withstand recommended cleaning and disinfection over useful life.Test was performed and criteria met (device is substantially equivalent).
    Storage Temperature TestSimulated UseEnsure device can function after exposure to heat and cold cycles.Test was performed and criteria met (device is substantially equivalent).
    BiocompatibilityCytotoxicity, Sensitization, Irritation or Intracutaneous reactivity, Acute Systemic toxicity (materials mediated pyrogen)Ensure material of construction are biocompatible for their intended use.Conducted per ISO 10993 at a GLP testing facility; tests were performed and criteria met (device is substantially equivalent).

    2. Sample size used for the test set and the data provenance:

    The document describes non-clinical bench testing. It does not specify a "test set" in the context of patient data or clinical trials, nor does it provide details on sample sizes for each bench test beyond the phrase "non-clinical bench testing was performed." The provenance of this data would be laboratory testing.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    Not applicable. This is a non-clinical device submission not involving expert review of diagnostic images or interpretations.

    4. Adjudication method for the test set:

    Not applicable. This is a non-clinical device submission.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    No MRMC study was done. This device is a mechanical lancing device, not an AI-assisted diagnostic tool. Clinical testing was deemed "not applicable" as per the document.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    Not applicable. This device is a mechanical lancing device, not an algorithm.

    7. The type of ground truth used:

    For the non-clinical tests, the "ground truth" is defined by the engineering specifications and performance requirements for the device (e.g., specific force values for button activation, a device functioning after a 1-meter drop). For biocompatibility, the ground truth is established by the relevant ISO 10993 standards and the results from GLP testing, demonstrating the material's biological safety.

    8. The sample size for the training set:

    Not applicable. This is a non-clinical device submission and does not involve AI/ML training.

    9. How the ground truth for the training set was established:

    Not applicable. This is a non-clinical device submission and does not involve AI/ML training.

    Ask a Question

    Ask a specific question about this device

    K Number
    K222539

    Validate with FDA (Live)

    Date Cleared
    2022-11-18

    (88 days)

    Product Code
    Regulation Number
    878.4850
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Facet Lancet is a sterile, disposable single used with a compatible proprietary lancet base (lancing device) to obtain a droplet of capillary blood from the finger for subsequent diagnostic testing. The Lancet is to be properly disposed of after a single use on an individual child, adolescent, or adult patient in a home setting.

    Device Description

    The Facet Lancet is a sterile, single use, blood sampling device used to obtain a sample of capillary blood for diagnostic primarily for blood glucose monitoring in diabetic patients. The lancet is available in two needle sizes, 30 gauge and 33 gauge.

    Facet lancets are intended to be used by diabetic patients to obtain a blood sample for assistance in self-monitoring of their blood glucose levels. The lancets consist of a 30G or 33G stainless steel needle overmolded with low density polyethylene (LDPE) and integral sterility cap. The lancet is single use, disposable and is sterilized by gamma radiation to a SAL of 10-9.

    AI/ML Overview

    The provided text describes a 510(k) premarket notification for the Facet Blood Lancets, seeking to demonstrate substantial equivalence to a predicate device. The information focuses on non-clinical bench testing to ensure the device meets predetermined criteria and special controls.

    Here's an analysis of the provided information regarding acceptance criteria and the study that proves the device meets them:

    1. A table of acceptance criteria and the reported device performance

    The document provides a "Summary of Nonclinical Tests Performed" (Table 2) which lists the properties/characteristics tested and their importance. While it doesn't explicitly state quantitative acceptance criteria for each test, it implies that meeting these criteria demonstrates acceptable performance for substantial equivalence. The reported performance is generally stated as "met predetermined criteria" in the text, rather than specific numerical results.

    Property/CharacteristicTest MethodImportanceReported Device Performance
    Bond Strength of lancet blade to lancet bodyTensile strength testerEnsure needle does not detach from lancet body during useMet predetermined criteria; substantially equivalent to predicate device
    Sterility Cap TorqueTorque testerEnsure that the cap can be easily removed by user when desiredMet predetermined criteria; substantially equivalent to predicate device
    Compatibility with lancet basesSimulated useEnsure usability with commercially available lancet basesMet predetermined criteria; substantially equivalent to predicate device
    BiocompatibilityCytotoxicity, Sensitization, Irritation or Intracutaneous reactivity, Acute Systemic toxicity (materials mediated pyrogen)Ensure material of construction are biocompatible for their intended useMet predetermined criteria; substantially equivalent to predicate device

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document does not specify the sample size used for the non-clinical bench tests (test set). It also does not explicitly state the data provenance (e.g., country of origin, retrospective or prospective), but given it describes a 510(k) submission, the tests were conducted by or for the manufacturer (Facet Technologies LLC) as part of a prospective evaluation for regulatory approval.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    This type of information is not applicable to the non-clinical, bench testing described for this device. Ground truth, in the context of expert consensus, is typically relevant for studies involving human interpretation or diagnosis, which is not the nature of these mechanical and biocompatibility tests.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Adjudication methods are typically associated with studies involving human interpretation or subjective assessments. For the objective, non-clinical bench tests described here, an adjudication method is not applicable. The results would be based on direct measurements and adherence to specified test procedures.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    A Multi-Reader Multi-Case (MRMC) comparative effectiveness study is not applicable to this device. These studies are typically used for diagnostic imaging devices where human readers (e.g., radiologists) interpret images with and without AI assistance. The Facet Blood Lancet is a physical medical device for blood sampling, not an AI-powered diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    A "standalone" algorithm performance assessment is not applicable here as the Facet Blood Lancet is a physical medical device and does not involve any algorithms.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    For the non-clinical tests described:

    • Bond Strength and Cap Torque: The "ground truth" would be established by the engineer's design specifications and defined pass/fail criteria based on mechanical engineering principles and relevant standards.
    • Compatibility: The "ground truth" would be successful operation with commercially available lancing devices, likely defined by visual inspection and ease of use.
    • Biocompatibility: The "ground truth" is established by adherence to recognized international standards (ISO 10993 series) which dictate the acceptable biological response to device materials.

    8. The sample size for the training set

    This device does not involve a "training set" in the context of machine learning or AI models. The tests performed are physical and biological evaluations of the device itself.

    9. How the ground truth for the training set was established

    As there is no "training set" for this type of device and study, this question is not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K221433

    Validate with FDA (Live)

    Date Cleared
    2022-07-15

    (59 days)

    Product Code
    Regulation Number
    878.4850
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Facet 28G Universal Lancet is a sterile, disposable single used to obtain a droplet of capillary blood for subsequent diagnostic testing from the finger or an alternative site, such as the palm, upper arm, or forearm. The Lancet is to be properly disposed of after a single use on an individual child, adolescent, or adult patient in a home setting.

    Device Description

    The Facet 28G Universal Lancet) is a sterile, single use, blood sampling device used to obtain a sample of capillary blood for diagnostic purposes, primarily for blood glucose monitoring in diabetic patients. The lancets consist of a 28G stainless steel needle overmolded with low density polyethylene (LDPE) and integral sterility cap. The lancet is single use, disposable and is sterilized by gamma radiation to a SAL of 10-9.

    AI/ML Overview

    This document is a 510(k) summary for the Facet 28G Universal Lancet, a medical device. It does not describe an AI/ML-enabled device, nor does it contain information about clinical studies with human participants, expert ground truth establishment, or multi-reader multi-case (MRMC) studies. Therefore, many of the requested points regarding acceptance criteria and study details for AI/ML devices cannot be answered from this document.

    However, I can extract the acceptance criteria and performance related to the non-clinical bench testing for this physical device.


    Device: Facet 28G Universal Lancet

    Nature of the Device: This is a physical medical device (blood lancet), not an AI/ML software device. The acceptance criteria and performance data presented are for physical and material properties, and compatibility, not for diagnostic accuracy or algorithmic performance.


    1. Table of Acceptance Criteria and the Reported Device Performance

    The document describes non-clinical bench testing, which serves as the "study" to prove the device meets pre-determined criteria. The reported performance is not explicitly given in numerical terms for all tests, but rather stated that "predetermined criteria were met" and "risks were appropriately managed." However, the acceptance criteria (requirements) are provided.

    Property/CharacteristicAcceptance Criteria (Requirement)Reported Device Performance (Implied)
    Bond Strength of lancet blade (needle) to lancet body$\geq$ 13.23 NMet the criteria (stated "predetermined criteria were met")
    Sterility Cap Torque$\leq$ 4.9 N-cmMet the criteria (stated "predetermined criteria were met")
    Compatibility with lancet bases (lancing devices)Ensure usability with commercially available lancet basesMet the criteria (stated "predetermined criteria were met")
    BiocompatibilityMeet ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-11 requirements (Cytotoxicity, Sensitization, Irritation or Intracutaneous reactivity, Acute Systemic toxicity)Met the criteria (stated "predetermined criteria were met")
    SterilitySterilized by Gamma Radiation SAL = $10^{-6}$Met the criteria (Same as predicate)

    Note: The document only states that the criteria were "met" or that the performance was "same" as the predicate. Specific numerical results of the tests (e.g., the exact bond strength measured) are not provided in this summary.


    2. Sample Size Used for the Test Set and the Data Provenance

    • Sample Size: The document does not specify the numerical sample sizes for each non-clinical test (e.g., how many lancets were tested for bond strength or cap torque). It broadly states that "non-clinical bench testing was performed."
    • Data Provenance: This is not applicable in the context of clinical data for an AI/ML device. For this physical device, the data provenance is from internal non-clinical bench testing performed by Facet Technologies, LLC. The testing ensures the device meets specified engineering and biocompatibility requirements.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    • Not applicable. This device is a physical lancet, not an AI/ML system requiring expert interpretation of data. The "ground truth" here is objective measurement of physical properties (e.g., force, torque) and standard biocompatibility testing.

    4. Adjudication Method for the Test Set

    • Not applicable. There is no human interpretation or adjudication required for the physical property tests or biocompatibility assessments described.

    5. If a Multi-Reader Multi-Case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • Not applicable. This is not an AI-enabled device and no MRMC studies were performed or are relevant.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    • Not applicable. This is not an AI/ML algorithm.

    7. The type of ground truth used

    • For physical properties (Bond Strength, Cap Torque): The ground truth is established by objective physical measurements against engineering specifications.
    • For Biocompatibility: The ground truth is established by standardized international (ISO 10993 series) and FDA-recognized test methods performed in a laboratory setting.
    • For Compatibility: The ground truth is established via simulated use with commercially available lancing devices.

    8. The sample size for the training set

    • Not applicable. This is not an AI/ML device, so there is no "training set." The focus is on verifying compliance with design specifications and regulatory standards.

    9. How the ground truth for the training set was established

    • Not applicable. There is no training set as this is not an AI/ML device.