Search Results
Found 14 results
510(k) Data Aggregation
(237 days)
The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon, electron, and proton treatment plans and displays, on-screen and in hard-copy, two- or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups. The Monaco product line is intended for use in radiation treatment planning. It uses generally accepted methods for:
- contouring
- image manipulation
- simulation
- image fusion
- plan optimization
- QA and plan review
The Monaco RTP System accepts patient diagnostic imaging data from CT and MR scans, and source dosimetry data, typically from a linear accelerator. The system then permits the user to display and define (contour) the target volume to be treated and critical structures which must not receive above a certain level of radiation, on these diagnostic images. Based on the prescribed dose, the user, a Dosimetrist or Medical Physicist, can then create multiple treatment scenarios involving the number, position(s) and energy of radiation beams and the use of a beam modifier (MLC, block, etc.) between the source of radiation and the patient to shape the beam. Monaco RTP system then produces a display of radiation dose distribution within the patient, indicating not only doses to the target volume but to surrounding tissue and structures. The optimal plan satisfying the prescription is then selected, one that maximizes dose to the target volume while minimizing dose to surrounding healthy volumes.
The parameters of the plan are output for later reference and for inclusion in the patient file. Monaco planning methods and modalities:
- Intensity Modulated Radiation Treatment (IMRT) planning
- Electron, photon and proton treatment planning
- Planning for dynamic delivery methods (e.g., dMLC, dynamic conformal)
- Volumetric Modulated Arc Therapy (VMAT)
- Stereotactic planning and support of cone-based stereotactic
- 3D conformal planning
- Distributed planning configurations (e.g., for conventional linac)
- Adaptive planning capabilities (e.g., for MR-Linac & conventional linac)
- Auto planning features (e.g., for conventional linac)
Monaco basic systems tools, characteristics, and functions:
- Plan review tools
- Manual and automated contouring tools (Segmentation component for MR images)
- DICOM connectivity
- Windows operating system
- Simulation
- Support for a variety of beam modifiers (e.g. MLCs, blocks, etc.)
- Standardized uptake value (SUV)
- Specialty Image Creation (MIP, MinIP, and Avg)
- Monaco dose and Monitor Unit (MU) calculation
- Dose calculation algorithms for electron, photon, proton planning
Monaco is programmed using C, C++ and C# computer programming languages. Monaco runs on Windows operating system and off-the-shelf computer server/hardware.
The provided FDA 510(k) clearance letter and summary for the Monaco RTP System (6.3) outlines the acceptance criteria and a study supporting the substantial equivalence of the new features. Here's a breakdown of the requested information:
1. A table of acceptance criteria and the reported device performance
| Changed Feature | Acceptance Criteria | Reported Device Performance |
|---|---|---|
| Segmentation component for invoking MR auto-segmentation algorithms (AI-based) | Primary metric: Average Hausdorff Distance (AVD) ≤ 3 mm. Secondary metric (value of interest): DICE or AUC ≥ 0.7 for specific structures. Additionally, qualitative analysis based on a 5-point Likert scale to determine if automatically generated structures provide a valuable starting point for clinical delineation. Investigation of any failures to meet the DICE confidence value of 0.7, with findings included as "Limitations." Sub-group analysis based on patient size, pixel size, slice spacing, and number of slices. | For all evaluated structures across all models (Female Pelvis Intact & Hysterectomy, Male Pelvis, and Head & Neck), the mean Absolute Volume Difference (AVD) was less than 3 mm. Structure-specific statistical analyses supported this conclusion. Patterns of failure for any structure failing the DICE confidence value of 0.7 were investigated and included as "Limitations." Qualitative analysis concluded that automatically generated structures provided a valuable starting point for clinical delineation. |
| Auto-planning | All pre-defined acceptance criteria related to workflow performance, protocol management, plan creation, interoperability, and error handling must be met. Plans generated must be clinically acceptable for the intended use and not introduce new safety or effectiveness concerns. | All testing met pre-defined acceptance criteria. Treatment plans generated were reviewed within the clinical workflow and determined to be suitable for clinical use, without introducing new safety or effectiveness concerns. |
| Extending adaptive planning capabilities to EMLA for offline adaptive planning | Correct system behavior during image registration, structure propagation, dose recalculation/re-optimization, offline adaptive plan generation, and workflow execution under representative clinical scenarios. No defects, unexpected behavior, or data integrity issues. Plans generated must be clinically acceptable for the intended use. All predefined acceptance criteria for verification and validation must be met. | No defects, unexpected behavior, or data integrity issues were identified during testing. Validation demonstrated that offline adaptive planning using CT‑to‑CBCT supports the creation of clinically acceptable treatment plans for the intended use. Offline adaptive plans were reviewed within the clinical workflow and determined to be suitable for use. Verification and validation testing met pre-defined acceptance criteria. |
| Interoperability with 3rd party software for image management and contouring | Correct DICOM export functionality, preservation of data integrity, and successful creation of an offline adaptive plan using third-party contouring. Third-party contouring outputs must be clinically acceptable and comparable to reference contours produced by qualified users. All planned Solution Interoperability test cases successfully executed and passed. All verification and validation testing met predefined acceptance criteria. | All verification and validation testing met the predefined acceptance criteria. All planned Solution Interoperability test cases have been successfully executed and passed. |
2. Sample size used for the test set and the data provenance
- AI-based segmentation component:
- Female Pelvis Intact & Hysterectomy models: 529 images (joint image set).
- Male Pelvis model: 250 images.
- Head & Neck model: 1862 images.
- Data Provenance: Not explicitly stated regarding country of origin or whether the data was retrospective or prospective.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
- AI-based segmentation component: "reference contours produced by qualified users" and "clinical delineation." The exact number or specific qualifications (e.g., radiologist with X years of experience) of these experts are not specified in the provided document.
4. Adjudication method for the test set
- AI-based segmentation component: For the qualitative analysis, it states "a conclusion that the automatically generated structures provided a valuable starting point for clinical delineation." This implies human review and evaluation. However, a formal adjudication method like "2+1" or "3+1" is not explicitly mentioned.
- For other features (Auto-planning, Adaptive planning, Interoperability), reviews mention evaluation within the "clinical workflow" and determination of "suitability for clinical use," but a specific adjudication method beyond internal reviews is not detailed.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- A formal MRMC comparative effectiveness study to quantify human reader improvement with AI assistance is not mentioned in the provided text. The AI component was evaluated in a standalone manner for its segmentation accuracy, and qualitatively for its utility as a "starting point for clinical delineation."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Yes, for the AI-based segmentation component, a standalone algorithm-only performance evaluation was done using metrics like Average Hausdorff Distance (AVD), DICE, and AUC.
7. The type of ground truth used
- For the AI-based segmentation component, the ground truth for the test set involved "reference contours produced by qualified users" and "clinical delineation." This implies expert consensus/delineation rather than pathology or outcomes data.
- For other features, "clinically acceptable treatment plans" and "suitable for clinical use" imply evaluation against accepted clinical standards, likely by qualified personnel.
8. The sample size for the training set
- The training set sample sizes are indicated for the AI-based segmentation models:
- Female Pelvis Intact & Hysterectomy: 529 images (joint image set used for training).
- Male Pelvis: 250 images (used for training).
- Head & Neck: 1862 images (used for training).
9. How the ground truth for the training set was established
- The document implies that the training data for the AI-based segmentation models would have been expertly annotated to establish the ground truth, given the mention of "reference contours produced by qualified users" for evaluation. However, the exact method for establishing ground truth for the training set is not explicitly detailed beyond this inference.
Ask a specific question about this device
(185 days)
The Elekta Medical Linear Accelerator (EMLA) is intended to be used for external beam radiation therapy (EBRT) treatments as determined by a licensed medical practitioner.
It is intended to assist a licensed medical practitioner in the delivery of EBRT to defined target volumes, while sparing surrounding normal tissue and critical organs from excess radiation.
Elekta Synergy and Elekta Harmony are the default entry-level configurations. It is intended to be used for single or multiple fractions using standard dose fractionation, hyperfractionation, and hypofractionation in all areas of the body where such treatment is indicated.
Elekta Infinity and Elekta Harmony Pro are the default mid-level configuration. It is intended to be used for single or multiple fractions using standard dose fractionation, hyperfractionation, hypofractionation and stereotactic delivery (stereotactic body radiation therapy – SBRT; stereotactic ablative radiotherapy – SABR) in all areas of the body where such treatment is indicated.
Versa HD and Elekta Evo are the default high-level configuration. It is intended to be used for single or multiple fractions using standard fractionation, hyperfractionation, hypofractionation and stereotactic delivery (stereotactic body radiation therapy – SBRT; stereotactic ablative radiotherapy – SABR; stereotactic radio surgery - SRS) in all areas of the body where such treatment is indicated and for the treatment of functional disorders, such as trigeminal neuralgia.
The EMLA is indicated for the delivery of curative and palliative intent EBRT to Adult and Pediatric patients with primary benign and malignant tumor and metastasis (or secondaries) anywhere in the body.
The Elekta Medical Linear Accelerator (EMLA) is an external beam, image guided Radiation Therapy device to assist a licensed practitioner in the delivery of ionizing radiation to a defined target volume. The system is located in a radiation-shielded treatment room and consists of several sub-systems, such as, the electron accelerator, beam shaping, imaging, computerized control systems and a treatment table to support the patient with accessories for patient positioning and set-up to deliver therapeutic treatments.
The EMLA is equipped with a MV portal imaging sub-system, i.e. iViewGT, and an optional kV imaging sub-system, i.e. XVI. The table is capable of linear and rotational movements.
The user interface controlling devices are located partly in the treatment room and partly in the control room.
The EMLA is made available in the following models: Elekta Synergy, Elekta Harmony, Elekta Infinity, Elekta Harmony Pro, Versa HD, Elekta Evo. The major differences are described in section VII.
The provided FDA 510(k) clearance letter and summary for the Elekta Medical Linear Accelerator (EMLA) describes performance testing for differences between the subject devices (new EMLA models) and the predicate devices (older EMLA models, K210500). The primary focus of the performance testing detailed in the summary is related to improvements in CBCT image quality and reconstruction.
Here’s a breakdown of the requested information based on the provided text:
1. Table of acceptance criteria and the reported device performance
The document does not explicitly present a table of quantitative acceptance criteria alongside corresponding reported device performance values. Instead, it describes general improvements and conformance to standards. The acceptance is implicitly based on meeting or exceeding the predicate device's performance, especially for the high-definition (HD) reconstruction.
| Acceptance Criterion (Implicit) | Reported Device Performance (Summary of Test Results) |
|---|---|
| Conformance to applicable consensus standards (e.g., IEC 60601-2-68 for image quality, IEC 60601-2-1 Ed. 4 for Linac control) | Test results showed conformance of the subject devices to the applicable consensus standards, Elekta defined performance specifications, and associated risk management requirements. |
| FDK based reconstruction function: image quality for IGRT (uniformity, spatial resolution, low contrast visibility, geometric accuracy) | Improved image quality in uniformity, volume outline, and spatial resolution compared to the predicate device, with no adverse impact on registration accuracy. |
| FDK based reconstruction function: image registration accuracy | Accurate registration. The conclusion mentions "no adverse impact to the accuracy of registration" for the FDK based an improved FDK based reconstruction function. |
| HD Reconstruction function (pelvic anatomies): improved image quality for IGRT (uniformity, HU consistency, SNR, CNR, contrast consistency) compared to FDK based reconstruction | Image quality improved in terms of better uniformity and HU accuracy. Improved image quality results in better performance of the automatic registration function, often not requiring manual adjustment. Clinical survey showed a preference for HD Reconstruction over the predicate. |
| HD Reconstruction function (pelvic anatomies): image registration accuracy compared to FDK based reconstruction | Improved image quality often leads to better performance of the automatic registration function, not requiring any manual adjustment post registration. |
| HD Reconstruction function: Clinical image quality (qualitative comparison) | Users qualitatively compared image quality between the predicate device and the subject device, reporting improved image quality. Clinical survey showed a preference towards the HD Reconstruction of the subject device over the predicate. |
| Cybersecurity improvements for linac and imaging system control | The control system for the subject device has improvements to cybersecurity; enables compliance with IEC 60601-2-1 Ed. 4; supports an integrated beam gating interface in compliance with IEC 60601-2-1 Ed. 4 and based on NEMA RT 1-2014 standard. |
| Functional performance characteristics (e.g., photon and electron energy/dose rates) | Most characteristics are "Same" as predicate (e.g., dose rates). Harmony Pro supports more photon and electron energies than predicate Harmony. Subject Synergy uses Agility BLD (which covers MLCi2 performance) whereas predicate Synergy supports both. Elekta Evo supports HexaPOD evo RT System (highest performance). All subject devices conform to the same patient-contact materials and rely on predicate device test data where technological characteristics are the same. |
2. Sample size used for the test set and the data provenance
- FDK based reconstruction function (Image Quality Evaluation): "acquired image quality phantom data" - Specific sample size not provided, likely laboratory phantom data.
- FDK based reconstruction function (Image Registration Accuracy): "phantom data and CT reference data" - Specific sample size not provided, likely laboratory phantom data.
- HD Reconstruction vs. FDK (Image quality comparison - clinical data sets): A total of 124 different clinical data sets.
- HD Reconstruction vs. FDK (Image registration accuracy - clinical patient CBCT projection data sets): A total of 13 different clinical patient CBCT projection data sets.
- HD Reconstruction vs. FDK (Clinical image quality evaluation - clinical patient data): Specific number of patients/data sets not explicitly stated, but clinical patient data was used for qualitative comparison.
Data Provenance: The document does not explicitly state the country of origin for the clinical data or explicitly state whether it was retrospective or prospective. Given it's clinical data sets for evaluating reconstruction, it's highly likely to be retrospective data collected from clinical operations.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts
The document states:
- For the FDK based reconstruction: "reconstruct a CBCT volume image which is suitable for visualizing anatomies to enable certain clinical judgment" - this implies expert judgment in assessment, but does not specify the number or qualifications.
- For the HD Reconstruction: "suitable for visualizing the pelvic anatomies to enable certain clinical judgment".
- "A clinical image quality evaluation was performed between HD Reconstruction function and FDK based reconstruction function, using clinical patient data, where user qualitatively compared image quality between the predicate device and the subject device and reported improved image quality." - This indicates qualitative evaluation by "user", but the number and specific qualifications (e.g., radiologist with X years of experience) are not defined. The document also mentions "Formal validation of the clinical workflows has been performed by competent and professionally qualified personnel" but does not specify them in relation to ground truth establishment for the test set.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
The document does not describe a formal adjudication method (like 2+1 or 3+1 consensus) for establishing ground truth or evaluating the clinical image quality. It generally refers to qualitative comparison by "user" or "competent and professionally qualified personnel".
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
The document mentions "A clinical image quality evaluation was performed between HD Reconstruction function and FDK based reconstruction function, using clinical patient data, where user qualitatively compared image quality between the predicate device and the subject device and reported improved image quality." It also notes "A clinical survey shows a preference towards the HD Reconstruction of the subject device over the predicate."
This suggests a form of reader study or survey was conducted, where users (human readers) compare images from the legacy FDK method (without the new AI-ML component) to the HD Reconstruction (which "includes an AI-ML based component to estimate the scatter"). However, it's not explicitly labeled as a "multi reader multi case (MRMC) comparative effectiveness study" in the formal sense, and no quantitative effect size of improvement for human readers is provided. The improvement is described qualitatively (e.g., "improved image quality," "better uniformity and HU accuracy," "preference").
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, standalone algorithm performance evaluation was done for the FDK and HD reconstruction functions. This is evident from:
- "Image quality evaluation was performed for the FDK based reconstruction function, using acquired image quality phantom data, to evaluate uniformity, spatial resolution, low contrast visibility, and geometric accuracy in accordance with IEC 60601-2-68;"
- "Image quality comparison was performed between HD Reconstruction function and FDK based reconstruction function, using data acquired on phantom data, to evaluate uniformity, spatial resolution, low contrast visibility, and geometric accuracy in accordance with IEC 60601-2-68."
- "Image quality comparison was performed between HD Reconstruction function and FDK based reconstruction function, using a total of 124 different clinical data sets, to evaluate uniformity, Hounsfield Unit consistency, signal-to-noise ratio, contrast-to-noise ratio, and contrast consistency."
These evaluations measure the intrinsic performance of the reconstruction algorithms without explicit human interaction beyond setting up the evaluation.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The types of ground truth used include:
- Phantom Data: For evaluating image quality metrics (uniformity, spatial resolution, low contrast visibility, geometric accuracy) and image registration accuracy (against CT reference data). Phantoms often have known properties or are designed to allow for quantitative measurement of image accuracy.
- CT Reference Data: Used for comparing image registration accuracy. CT imaging is considered a high-fidelity reference.
- Clinical Patient Data: Used for comparing various image quality metrics (Hounsfield Unit consistency, signal-to-noise ratio, contrast-to-noise ratio, contrast consistency) and for qualitative "user" comparison of image quality. The "ground truth" for clinical image quality comparison would effectively be the subjective assessment of the users or experts performing the comparison against each other, and against the clinical utility standards.
- Simulated Monte Carlo data: Used to train the neural network component of the HD reconstruction that estimates scatter. This implies a simulated ground truth for scatter estimation.
8. The sample size for the training set
The document states: "It includes an AI-ML based component to estimate the scatter to enable its automatic removal from the projection images acquired by the imager ahead of the volume reconstruction. Simulated Monte Carlo data is used to train the network."
The specific sample size for the training set (i.e., the amount of simulated Monte Carlo data) for the AI-ML component is not provided.
9. How the ground truth for the training set was established
The ground truth for the training set of the AI-ML component was established through Simulated Monte Carlo data. Monte Carlo simulations are a computational method that can model the physical interactions of radiation with matter, providing a "ground truth" for scatter estimation in this context.
Ask a specific question about this device
(160 days)
ProKnow DS is intended to be a data archive and processing software which supports review, analytics, and clinical decision-making with a focus on the data specific to radiation oncology patients.
ProKnow DS should be used in healthcare facilities by trained medical professionals including, but not limited to physicians, medical technologists, physicists and dosimetrists.
ProKnow DS provides a high-performance, scalable medical image management and processing system. It allows users to archive, inspect, analyze, and interact with radiation therapy patient data for both retrospective and prospective studies. Its features are centered on two primary areas of interaction: (1) single patient datasets and (2) collections of patient datasets, i.e., patient cohorts. Users are able to import patient data from existing imaging, contouring, and treatment planning systems (via DICOM formats). Once imported, patient data is archived for long-term storage and is also available for inspection and analysis (examples of supported inspection and analysis tasks include: visualizing images, structures, and dose distributions; inspecting plan information; inspecting dose volume histograms; and extracting metrics).
ProKnow DS also allows users to edit physical models associated with a patient for use in retrospective studies or to be exported to commercially available radiation treatment planning systems. Once a set of patients have been established in ProKnow DS, users may create collections of related patients allowing them to analyze and correlate dosimetric values of interest and clinical endpoints across large treatment populations.
N/A
Ask a specific question about this device
(140 days)
Leksell Gamma Knife® is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters e.g.
- Metastatic tumors
- Recurrent glioblastomas
- Trigeminal neuralgia
- Medically refractory essential tremor
- Orbital tumors
- Ocular tumors
- Optic nerve tumors
- Benign diseases (such as meningiomas, vestibular schwannomas, post-surgical pituitary adenomas, craniopharyngioma, hemangioblastomas, schwannomas, arteriovenous malformations, cavernous malformations, chordomas, glomus tumors, hemangiomas)
- Skull base tumors
- Head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, oropharynx, nasopharynx, sinonasal, salivary gland)
- Pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, craniopharyngioma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformations, cavernous malformations, skull base tumors)
- Refractory mesial temporal lobe epilepsy in adults
- Refractory epilepsy associated with structural changes such as hamartomas, cerebral cavernous malformations, and arteriovenous malformations (adult and pediatric)
Leksell Gamma Knife® (available models Elekta Esprit, Icon™, and Perfexion™) is a radiosurgery system for use in the stereotactic irradiation of head structures. Surgery is achieved by delivering a prescribed dose as one or more shots of ionizing radiation to the exact site of the target.
The provided FDA 510(k) clearance letter pertains to the Leksell Gamma Knife® (Elekta Esprit, Icon™, Perfexion™) and describes the basis for its substantial equivalence to a predicate device (K222047).
It's important to note that the document outlines the non-clinical performance testing and explicitly states that no animal or clinical tests were performed to establish substantial equivalence for this specific 510(k) submission. This is because the submission primarily focuses on the replacement of obsolete hardware components and a new control system software version to support these components, along with the addition of new indications (Refractory mesial temporal lobe epilepsy in adults and Refractory epilepsy associated with structural changes) that are deemed to fall within the existing intended use and have a favorable benefit-risk profile based on established knowledge of Gamma Knife treatment for certain neurological conditions.
Therefore, the typical structure for describing acceptance criteria and a study proving device performance (especially for AI/ML-driven devices with specific performance metrics like sensitivity, specificity, etc.) is not fully applicable in this context. The acceptance criteria here are more related to functional performance, safety, and maintaining equivalence with the predicate device after hardware and software updates.
However, I can extract the information relevant to what testing was performed and how the device's continued safety and effectiveness were assured.
Acceptance Criteria and Study to Prove Device Meets Acceptance Criteria
Given that this 510(k) focuses on hardware replacement and a software update to support those replacements, rather than the introduction of a novel AI/ML algorithm with specific diagnostic performance metrics, the acceptance criteria are not in the form of typical clinical performance measures (e.g., sensitivity, specificity). Instead, they revolve around functional validation, safety, and maintenance of equivalence to the predicate device.
Key Acceptance Criteria (Inferred from the document):
- Functional Equivalence: The new hardware and software must maintain the same fundamental scientific technology, clinical workflow, and functionality as the predicate device.
- Safety: The device must continue to meet established safety standards and not introduce new or increased risks.
- Performance: The updated system must perform as intended, ensuring accurate and precise stereotactic irradiation.
- Usability: User interface and labeling updates must support safe and effective use.
- Biocompatibility: All new materials in contact with patients must be biocompatible.
Study that Proves the Device Meets the Acceptance Criteria:
The "study" described is a comprehensive set of non-clinical verification and validation activities.
1. Table of Acceptance Criteria and Reported Device Performance
| Acceptance Criterion (Inferred) | Reported Device Performance |
|---|---|
| Functional Equivalence | - "New hardware has been introduced to replace old obsolete components. A new control system software version has been introduced to support and enable the new components."- "The subject device does not introduce any new clinical requirements or functions."- "The fundamental technical characteristics... have not changed and are substantially equivalent."- "The clinical workflow, safety, functionality, and performance is unchanged thus ensuring substantial equivalence." |
| Safety (Risk Control & Compliance with Standards) | - "The change in design required a new risk control to be added to ensure the same level of safety as predicate device and thus ensuring substantial equivalence."- Verification testing performed to IEC 60601-1 series, IEC 62304, ISO 10993 series. |
| Performance (Accuracy & Consistency) | - "Testing in the form of module, integration and system level verification was performed to evaluate the performance and functionality of the new control system and the new hardware against requirement specification."- "Regression test of unchanged functionalities... to ensure that new and updated functionalities did not introduce any undesirable effects." |
| Usability | - "Performance testing, including design and usability validation of the system, has been performed by competent and professionally qualified personnel to ensure that the product fulfils the intended use and user needs."- "The design and usability validation was also made to ensure that the risk control measures associated with functions related to safety (FRS) for the affected functionality... were effective."- "Updated labeling was in scope of usability evaluation according to IEC 62366-1. The result demonstrate that the labeling support the safe and effective use of the device." |
| Biocompatibility | - "Biocompatibility testing has been performed to ensure that all materials are biocompatible in relation to the use of the device." |
| Substantial Equivalence to Predicate Device (Overall) | - "The performance data demonstrate that the subject device is as safe and effective and performs as well as the predicate devices (K222047)."- "The result of verification and validation as well as conformance to relevant safety standards demonstrate that the Leksell Gamma Knife® meets the established safety and performance criteria and is substantially equivalent to the predicate device." |
2. Sample Size Used for the Test Set and Data Provenance
- Sample Size: Not explicitly stated as a number of "cases" or "patients" in the context of clinical data. The testing involved system-level verification, module testing, integration testing, and regression testing. This implies testing across the functionality of the device.
- Data Provenance: The nature of the testing is non-clinical performance and functional validation. It's conducted by Elekta Solutions AB (Sweden) as part of their design control and quality system processes. It's not based on retrospective or prospective clinical patient data for efficacy or diagnostic performance.
3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts
- Experts: Not applicable in the sense of clinical experts establishing "ground truth" for disease diagnosis or outcome.
- Qualifications: For performance, design, and usability validation, the document states: "Performance testing, including design and usability validation of the system, has been performed by competent and professionally qualified personnel." No specific number or precise qualifications (e.g., "radiologist with 10 years of experience") are provided within this public summary, as this refers to internal quality and engineering validation teams.
4. Adjudication Method for the Test Set
- Adjudication Method: Not applicable. The testing is focused on system functionality, safety, and performance against defined engineering requirements and standards, not on interpreting images or making clinical judgments requiring adjudicated consensus.
5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done
- MRMC Study: No. The document explicitly states: "No animal or clinical tests were performed to establish substantial equivalence with the predicate device." Therefore, no MRMC study comparing human readers with and without AI assistance was conducted for this specific 510(k) submission.
6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done
- Standalone Performance: No. This device is a radiation therapy system, not an AI diagnostic algorithm. While it has control software, "standalone performance" in the context of AI metrics like sensitivity/specificity for disease detection is not relevant here. The software's performance is integral to the entire system's functionality and safety.
7. The Type of Ground Truth Used
- Ground Truth: For non-clinical validation, the "ground truth" is typically defined by:
- Requirement Specifications: The device's performance is measured against its designed functional and safety requirements.
- Established Standards: Adherence to recognized national and international standards (e.g., IEC 60601, IEC 62304, ISO 10993, IEC 62366-1).
- Predicate Device Equivalence: The gold standard is the documented performance and safety profile of the legally marketed predicate device (K222047).
8. The Sample Size for the Training Set
- Training Set Sample Size: Not applicable. This 510(k) does not describe the development of a novel machine learning algorithm that requires a distinct training set in the AI sense. The software updates are for system control and enabling new hardware, not for learning from vast datasets.
9. How the Ground Truth for the Training Set was Established
- Ground Truth for Training Set: Not applicable, as there is no mention of a training set for an AI/ML algorithm.
Ask a specific question about this device
(146 days)
Leksell GammaPlan® is a computer-based system designed for Leksell Gamma Knife® treatment planning.
Leksell GammaPlan® is a powerful, computer-based treatment planning system specifically designed for the simulation and planning of stereotactic Leksell Gamma Knife® radiosurgery based on tomographic and projectional images.
The basis of treatment planning is the acquisition and processing of digital images by a computer workstation running the treatment planning application software. The program is capable of handling a range of different imaging modalities. Images from tomographic sources such as Computer Tomography (CT), Magnetic Resonance (MR) and Positron Emission Tomography (PET) scanners can be used as well as projectional images from angiograms (AI). This allows the direct comparison between vascular structures in projectional images and tissue structures in CT and MR.
Digital images can be imported into the system via the computer network.
The treatment planning application has the ability to plan a patient's treatment protocol based on a single target or multiple targets.
The basic elements of treatment planning are:
- defining the cranial target or targets .
- . devising the configuration of the collimators to be used during treatment
- determining the parameters of the radiation shots to be delivered by Leksell • Gamma Knife®.
The provided FDA 510(k) summary for the Leksell GammaPlan® (LGP) (K232854) does not contain specific details regarding acceptance criteria and the comprehensive study results in the way requested for a typical AI/ML device.
This submission is for an updated version of a treatment planning system, which is generally software for medical device operation, rather than an AI/ML algorithm that performs autonomous diagnostic or prognostic tasks. As such, the performance testing focuses on software verification and validation against requirements, rather than clinical performance metrics in terms of sensitivity, specificity, etc., with a ground truth established by experts.
However, I can extract the information that is present and indicate where the requested information is not available in the document.
1. A table of acceptance criteria and the reported device performance
| Acceptance Criteria | Reported Device Performance |
|---|---|
| For software functionality and performance: | "Testing in the form of module, integration and system level verification was performed to evaluate the performance and functionality of the new and existing features against requirement specifications." |
| "Formal design and usability validation has been performed on a clinically equivalent device by competent and professionally qualified personnel to ensure that the product fulfils the intended use and user needs." | "Results from verification and validation testing demonstrate that conformance to applicable technical requirement specification and user needs have been met and that the system is confident and stable." |
| "The design and usability validation was also made to ensure that the risk control measures associated with functions related to safety for the new functionality was effective." | "The results of verification and validation as well as conformance to relevant safety standards demonstrate that LGP v11.4 meets the established safety and performance criteria..." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Sample Size for Test Set: Not specified in the document. The testing described is software verification and validation, not a clinical trial with a "test set" in the sense of patient data for AI model evaluation.
- Data Provenance: Not specified, as it's not a clinical data-driven study. The document mentions "clinically equivalent device" for validation, implying internal testing and possibly simulated or anonymized data, but details are absent.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Number of Experts: Not specified.
- Qualifications of Experts: The document mentions "competent and professionally qualified personnel" performed design and usability validation, but no specific professional qualifications (e.g., radiologist, medical physicist) or experience levels are provided. Ground truth in the context of a treatment planning system primarily relates to the accuracy of computational models and adherence to clinical guidelines, rather than expert interpretation of images for diagnosis.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Adjudication Method: Not specified. It's unlikely this type of adjudication was performed, as the evaluation focused on software functionality and adherence to specifications, not on resolving disagreements in expert clinical assessment.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- MRMC Study: No, an MRMC study was not performed. This device is a treatment planning system, not an AI diagnostic or assistance tool intended to improve human reader performance in interpreting medical images. The document states, "No animal or clinical tests were performed to establish substantial equivalence with the predicate device."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Standalone Performance: The device itself is software for treatment planning. Its "standalone" performance would be its ability to correctly calculate and display treatment plans based on inputs, which is covered by the mentioned verification and validation testing. However, this is not "standalone AI algorithm performance" in the typical sense of a diagnostic AI. The system is inherently "human-in-the-loop" as it requires a clinician to define targets and parameters.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
- Type of Ground Truth: For the type of software described (treatment planning), the "ground truth" for verification and validation typically involves:
- Specification Compliance: Verification that the software performs according to its written requirements and design specifications.
- Industry Standards: Conformance to relevant medical device software (e.g., IEC 62304) and medical physics standards.
- Phantom/Physical Measurement Comparisons: For dose calculations, comparison with physical measurements in phantoms or known analytical solutions (though not explicitly detailed for this specific 510(k)).
- Clinical Equivalence/Usability: Validation that the software supports intended clinical use, as performed by "competent and professionally qualified personnel."
The document does not specify "expert consensus," "pathology," or "outcomes data" as ground truth for this submission, as these are more relevant for diagnostic or prognostic AI.
8. The sample size for the training set
- Sample Size for Training Set: Not applicable and not mentioned. This device is a treatment planning system, not a machine learning model that requires a training set in the conventional sense.
9. How the ground truth for the training set was established
- Ground Truth for Training Set Establishment: Not applicable and not mentioned, as there is no machine learning "training set" for this device.
Ask a specific question about this device
(129 days)
Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan.
Elekta Unity is intended for use with compatible Treatment Planning and Oncology Information Systems. The Elekta Unity 1.5T MR scanner is a magnetic resonance imaging system that produces cross-sectional images in any orientation of the internal structure of the whole body before, during, and after the radiotherapy treatment.
When interpreted by a trained physician magnetic resonance images acquired before, during, and after the radiotherapy treatment yield information that can be useful in diagnosis and may assist therapy planning, patient positioning and treatment delivery related to radiation oncology.
Elekta Unity is a multifunctional digital linear accelerator designed to assist licensed medical practitioners in the delivery of ionizing radiation to defined volumes (e.g. malignant and benign tumors). Elekta Unity is capable of both intensity modulated radiation therapy (IMRT) and image guided radiation therapy (IGRT).
The Elekta Unity 1.5T MRI scanner is a magnetic resonance imaging sub-system that produces cross-sectional images in any orientation of the internal structure of the whole body before, during, and after the radiotherapy treatment.
When interpreted by a trained physician the images acquired before, during, and after the radiotherapy treatment vield information that can be useful in diagnosis and may assist therapy planning, patient positioning, and treatment delivery related to radiation oncology.
In addition to the motion monitoring with manual interrupt previously cleared with predicate Unity devices, the subject device Elekta Unity has added the following motion management strategies:
- · Anatomic Position Monitoring (APM) with Manual interrupt (also referred to as True Tracking as it 'tracks' in 3D and enables the system to gate using the 3D position of the anatomy's motion)
- · APM with Anatomic Tolerance Check (ATC)
- Adaptive Therapy with optional Baseline Shift (BLS) Recovery.
These modifications are not contained solely within the Elekta Unity Device, as the full clinical benefit is achieved with interoperability of Unity, Monaco RTP, and MOSAIQ OIS.
The provided text is a 510(k) summary for the Elekta Unity MR Linac and describes updates to an existing device rather than a new standalone AI/ML device with a clinical performance study. Therefore, a direct answer to all points regarding "acceptance criteria and the study that proves the device meets the acceptance criteria" in the context of an AI/ML algorithm's clinical performance cannot be fully provided from this document.
Specifically, the document states: "No animal or clinical tests were performed to establish substantial equivalence with the predicate device." This indicates that a human-in-the-loop (MRMC) or standalone (algorithm only) clinical performance study as typically understood for AI/ML diagnostic or treatment planning algorithms was not conducted for this 510(k) submission. The clearance is based on substantial equivalence by demonstrating that the new motion management strategies do not change the fundamental scientific technology or raise different questions of safety or effectiveness.
However, I can extract the information relevant to the device's technical specifications and the non-clinical testing performed to establish substantial equivalence.
Here's a breakdown of the available information based on the provided document:
1. Table of Acceptance Criteria and Reported Device Performance
The document does not present "acceptance criteria" in the traditional sense of a clinical performance study against a ground truth. Instead, it provides a comparison of "Technological Characteristics" between the subject device (Elekta Unity with new features) and its predicate device (Elekta Unity K212114). The "acceptance" here relates to demonstrating that the new features do not negatively impact the established performance and safety characteristics, or that they meet expected performance defined by engineering specifications.
| Technological Characteristic | Acceptance Criteria (Predicate Performance - Implied) | Reported Device Performance (Subject Device) |
|---|---|---|
| Radiation Source / Beam | 7MV Bremsstrahlung X-Rays | ✓ (Same) |
| Method of IMRT | MLC based cone-beam delivery | ✓ (Same) |
| Collimation | Field shaping, Multi Leaf Collimator (MLC) | ✓ (Same) |
| MLC material | Tungsten Alloy | ✓ (Same) |
| Number of leaves | 80 leaf pairs | ✓ (Same) |
| Range of MLC collimated beam size | 0.5cm x 0.5cm to 57.4cm x 22cm | ✓ (Same) |
| Gantry | Ring Gantry, collision with patient not possible | ✓ (Same) |
| Radiation Head Shielding | Lead, Tungsten Alloy, and Steel shielding | ✓ (Same) |
| Source control mechanism | Dual channel dose monitoring system | ✓ (Same) |
| Radiation Transmission through head | 0.2% of the primary beam | ✓ (Same) |
| Isocenter distance | 143.5 cm | ✓ (Same) |
| Isocenter accuracy (Radius) | 0.5 mm | ✓ (Same) |
| Max Dose Rate | Clinical use: 450 cG/min at isocentre | ✓ (Same) |
| Static Dose Accuracy | >95% points passing 3%/3mm (high dose); >95% passing 5mm/5% (low dose); 1% agreement for output factors | ✓ (Same) |
| Motion Management Strategies | Motion monitoring with manual interrupt | ✓ (Same + APM with manual interrupt, APM with ATC, Adaptive Therapy with optional BLS Recovery) |
| End-to-End Gating Latency | Not applicable to previous predicate; new feature | [-200 to +200 ms] (radiation off) and [-200 to +280 ms] (radiation on) for ATC |
| Patient table degrees of freedom | 2 (vertical & longitudinal) | ✓ (Same) |
| Integrated imaging | Magnetic resonance imaging system | ✓ (Same) |
| MR Physical Characteristics (Bore Diameter) | 700 mm | ✓ (Same) |
| MRI Frequency | 64 MHz | ✓ (Same) |
| Field Strength | 1.5T | ✓ (Same) |
| Field of View | Up to 500 mm Sequence dependent | ✓ (Same) |
| Field Homogeneity | < 2 ppm over 50 cm x 50 cm x 45 cm volume | ✓ (Same) |
| Field Stability | ≤ 0.1 ppm/hour | ✓ (Same) |
| 3D Imaging Volumes (cm) | 0.5 x 0.5 x 0.8 (min) / 56 x 56 x 40 (max) | ✓ (Same) |
| 3D Imaging Resolution (cm) | 0.01 x 0.01 x 0.1 (min) / 0.875 x 0.875 x 1 (max) | ✓ (Same) |
| 2D Imaging Planes (cm) | AP x HF: 0.5 x 0.5 (min) / 56 x 56 (max) | ✓ (Same) |
| 2D Imaging Resolution (mm) | Selectable: 0.011 x 0.011 mm (min) / 8.75 x 8.75 mm (max); Slice thickness: 0.01 mm (min) / 705 mm (max); ≤1 mm over 20 cm FOV | ✓ (Same) |
| Geometric Accuracy | ≤ 2 mm over 34 cm FOV (Guaranteed); ≤ 2 mm over 42 cm FOV (Typical) | ✓ (Same) |
| Signal to Noise | 120 | ✓ (Same) |
| Diffusion Weighted Imaging (DWI) | DWI can be imported for Off-line review | ✓ (Same) |
| Treatment Planning and Delivery System | GPU-based Monte Carlo dose calculation algorithm (GPUMCD) using Elekta Monaco | ✓ (Same) |
| Dose Output Modelling | Dose output modelled with monitor units | ✓ (Same) |
| Dose Display | Display of Linac delivery parameters | ✓ (Same) |
The specific performance of the new motion management strategies (APM with manual interrupt, APM with ATC, Adaptive Therapy with optional BLS Recovery) is demonstrated through consistency with existing features and the quantitative specification for End-to-End Gating Latency.
2. Sample size used for the test set and data provenance
Not applicable, as no clinical performance study was conducted. The assessment was based on non-clinical (engineering and software) verification and validation testing, and comparison of technological characteristics to a predicate device.
3. Number of experts used to establish the ground truth for the test set and qualifications of those experts
Not applicable, as no clinical performance study requiring expert adjudication for ground truth was conducted. Ground truth for the MR images themselves is implicitly the physical reality captured by the scanner, with interpretation by "trained physician" for diagnosis, therapy planning, patient positioning, and treatment delivery.
4. Adjudication method (e.g., 2+1, 3+1, none) for the test set
Not applicable, as no clinical performance study was conducted.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC study was done, stated as "No animal or clinical tests were performed to establish substantial equivalence with the predicate device."
6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done
No standalone clinical performance study was done for the purpose of this 510(k) clearance, as explicitly stated that no animal or clinical tests were performed. The "algorithm" here refers to the system's control and motion tracking, which is part of the integrated device functionality, not a separate AI diagnostic or treatment planning algorithm being evaluated for standalone performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the non-clinical testing, the "ground truth" would be engineering specifications, physical measurements, and adherence to recognized standards (e.g., IEC 60601 series for safety and essential performance). For the imaging component, the imaging system produces cross-sectional images for interpretation by a "trained physician," where the intended ground truth for diagnostic or guidance purposes is the patient's anatomy and pathology as determined by clinical practice.
8. The sample size for the training set
Not applicable. This document describes a medical device with updated motion management software, not a de novo AI/ML algorithm requiring a training set in the typical sense for machine learning model development. The software capabilities are validated through engineering and functional testing.
9. How the ground truth for the training set was established
Not applicable. See point 8.
Ask a specific question about this device
(128 days)
MOSAIQ® is an oncology information system used to manage workflows for treatment planning and delivery. It supports information flow among healthcare facility personnel and can be used wherever radiotherapy and/or chemotherapy are prescribed. Users can configure MOSAIQ® for Medical Oncology use, Radiation Oncology use, or the two together. Medical oncology dose calculations are designed to support both adult and pediatric patients. It lets users:
- · Assemble electronic patient charts and treatment plans, order diagnostic tests, and prescribe medications.
- · Generate and keep medication formulary lists and calculate applicable medication dosages for medical oncology.
- · Import, view, annotate, adjust, enhance, manage and archive images.
- · Compare radiation treatment plans and evaluate dose coverage.
- · Design leaf plans for operation with radiotherapy treatment machines that have multi-leaf collimators.
- · Make sure radiation treatment plans imported from treatment planning systems agree with treatment machine constraints. MOSAIQ® reads actual settings from the treatment machine communication interface. It compares these settings with predefined values. If a mismatch occurs between the planned values and the actual machine settings, the system warns the user.
- · View reference images to setup treatment. MOSAIQ® refers to predefined settings to help treatment machine setup and communicates patient and machine setup instructions.
- · Record actual delivered radiation values in an electronic chart to track treatment.
- · Use stereotactic localization to calculate set-up coordinates for treatments.
- · Monitor Intrafractional motion with real time image overlay.
MOSAIQ® is not intended for use in diagnosis.
The MOSAIQ® Oncology Information System (OIS) is an image-enabled electronic medical record system. It manages clinical and administrative workflows within oncology departments and facilitates efficient patient care. It can be configured for Medical Oncology, Radiation Oncology, or both.
The Medical Oncology (MO) configuration is a medical oncology charting solution that includes customizable regimens (Care Plans) that automate chemotherapy orders for labs, procedures, and appointments. Configurable flowsheet views are used for reviewing treatment administration, documents, assessment and lab data. Users can enter medications and screen for drug/drug and drug/allergy interactions. MOSAIQ also performs standard calculations such as Body Surface Area (BSA) and Area Under the Curve (AUC). The Medical Administration Record (MAR) supports all information related to chemotherapy and blood product administration, clinical trial study drugs, dose amounts, infusion time, multiple administration, etc.
The Radiation Oncology configuration is also a charting solution with Computerized Physician Order Entry (CPOE) capability, along with added features for image management, patient setup and positioning, verify and record, plan import, review, and approval, stereotactic localization, and pretreatment checks. MOSAIQ's Radiation Oncology functionality can be used to support a wide variety of treatment modalities including Intensity Modulated Radio Therapy (IMRT), Image Guided Radio Therapy (IGRT), particle therapy, and stereotactic radiotherapy. It can import and store treatment plans from Therapy Planning Systems (TPS) via DICOM import/DICOM RT import.
In addition to these, the current version of MOSAIQ introduces the following modifications for radiation oncology:
- · Anatomic Position Monitoring (APM) with Manual interrupt (also referred to as True Tracking as it 'tracks' in 3D and enables the system to gate using the 3D position of the anatomy's motion)
- · APM with Anatomic Tolerance Check (ATC)
- · Adaptive Therapy with optional Baseline Shift (BLS) Recoverv
- · Care Rules for motion management.
These modifications are not contained solely within MOSAIQ, as the full clinical benefit is achieved with interoperability of Unity, Monaco RTP, and MOSAIQ OIS.
The provided text is a 510(k) summary for the MODAIQ® Oncology Information System. Based on this document, the following information can be extracted regarding acceptance criteria and supporting studies:
1. A table of acceptance criteria and the reported device performance
The document does not explicitly present a table of quantitative acceptance criteria and corresponding reported device performance metrics in the format typically seen for diagnostic or AI-driven systems (e.g., sensitivity, specificity, AUC).
Instead, the "acceptance criteria" are implied to be conformance with various standards and successful completion of design verification and performance testing. The "reported device performance" is essentially a statement that the device met these criteria through testing.
Therefore, a table of stated acceptance criteria and reported performance would look like this, based on the narrative:
| Acceptance Criterion (Implicit) | Reported Device Performance |
|---|---|
| Conformance with FDA's Quality System Regulation (21 CFR §820.30) | Design verification and performance testing carried out in accordance. |
| Conformance with ISO 13485 Quality Management System Requirements | Design verification and performance testing carried out in accordance. |
| Conformance with ISO 14971 Risk Management Requirements | Design verification and performance testing carried out in accordance. |
| Conformance with IEC 62304 Software Life-Cycle Processes | Design verification and performance testing carried out in accordance. |
| Conformance with FDA guidance for major level of concern software (Class C per IEC 62304) | Software verification testing conducted and documented. |
| Conformance with ISO 14971 (Risk Management) | Satisfied. |
| Conformance with IEC 62304 (Software Life-Cycle) | Satisfied. |
| Conformance with ISO 62083 (Safety of Radiotherapy Planning Systems) | Satisfied. |
| Conformance with IEC 61217 (Radiotherapy Equipment - Coordinates, Movements and Scales) | Satisfied. |
| Conformance with IEC 62274 (Safety of Radiotherapy Record and Verify Systems) | Satisfied. |
| Conformance with AAMI RT2:2017 (Radiation Therapy Readiness Check) | Satisfied. |
| Conformance with IEC 62366-1 (Usability Engineering for Medical Devices) | Satisfied. |
| Conformance with ISO 15223 (Symbols for Medical Devices) | Satisfied. |
| Meeting established safety and performance criteria | Demonstrated through verification and validation. |
| Being substantially equivalent to the predicate device | Determined through non-clinical testing. |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document states: "No animal or clinical tests were performed to establish substantial equivalence with the predicate device." This indicates that no test set involving patient data was used for performance evaluation in the context of clinical outcomes or diagnostic accuracy. The testing was non-clinical.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
Since no clinical or animal tests were performed, there was no need for experts to establish ground truth for a test set in the conventional sense of clinical performance studies. The "ground truth" for the non-clinical testing would have been established by engineering specifications, design requirements, and standard compliance.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
As no clinical test set was used for performance evaluation that would require human expert adjudication of results, no adjudication method was employed.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No MRMC comparative effectiveness study was conducted. The device is an Oncology Information System and not an AI-assisted diagnostic tool that would typically involve human readers interpreting cases with or without AI assistance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
The document describes "Design verification and performance testing" and "Software verification testing." These are standalone evaluations of the device's functionality and adherence to technical specifications and regulatory standards. However, it's not a standalone clinical performance study as would be seen for a diagnostic algorithm. The device's modifications for "Anatomic Position Monitoring" and "Adaptive Therapy" are explicitly mentioned to involve "interoperability of Unity, Monaco RTP, and MOSAIQ OIS," implying a system-level performance rather than a single algorithm's standalone clinical performance.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
For the non-clinical testing performed, the ground truth was based on:
- Design and risk management requirements: The device's functions were tested against pre-defined specifications.
- Recognized consensus standards: Compliance with standards like ISO 14971, IEC 62304, ISO 62083, etc., served as the "ground truth" for its safety and essential performance.
- Predicate device characteristics: Substantial equivalence was established by demonstrating that the new device's characteristics and performance align with or don't adversely deviate from the predicate device (MOSAIQ OIS, K203172).
8. The sample size for the training set
The document does not mention any "training set." This type of testing is for an updated version of an existing Oncology Information System, focusing on functional verification and regulatory compliance, not machine learning model training.
9. How the ground truth for the training set was established
As no training set is mentioned or implied, this question is not applicable.
Ask a specific question about this device
(127 days)
The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon, electron, and proton treatment plans and in hard-copy, two- or three-dimensional radiation dose distributions inside patients for given treatment plan set-ups.
The Monaco product line is intended for use in radiation treatment planning. It uses generally accepted methods for:
- contouring
- image manipulation
- simulation
- image fusion
- plan optimization
- QA and plan review
The Monaco RTP System accepts patient diagnostic imaging data from CT and MR scans, and source dosimetry data, typically from a linear accelerator. The system then permits the user to display and define (contour) the target volume to be treated and critical structures which must not receive above a certain level of radiation, on these diagnostic images. Based on the prescribed dose, the user, a Dosimetrist or Medical Physicist, can then create multiple treatment scenarios involving the number, position(s) and energy of radiation beams and the use of a beam modifier (MLC, block, etc.) between the source of radiation and the patient to shape the beam. Monaco RTP system then produces a display of radiation dose distribution within the patient, indicating not only doses to the target volume but to surrounding tissue and structures. The optimal plan satisfying the prescription is then selected, one that maximizes dose to the target volume while minimizing dose to surrounding healthy volumes.
The parameters of the plan are output for later reference and for inclusion in the patient file. Monaco planning methods and modalities:
- Intensity Modulated Radiation Treatment (IMRT) planning
- Electron, photon and proton treatment planning
- Planning for dynamic delivery methods (e.g. dMLC, dynamic conformal, Volumetric Modulated Arc Therapy (VMAT))
- Stereotactic planning and support of cone-based stereotactic
- 3D conformal planning
- Adaptive planning (e.g. for the Elekta Unity MR-Linac)
- Monaco basic systems tools, characteristics, and functions:
- Plan review tools
- Manual and automated contouring tools
- DICOM connectivity
- Windows operating system
- Simulation
- Support for a variety of beam modifiers (e.g. MLCs, blocks, etc.)
- Standardized uptake value (SUV)
- Specialty Image Creation (MIP, MinIP, and Avg)
- Monaco dose and Monitor Unit (MU) calculation:
- Dose calculation algorithms for electron, photon, proton planning
Monaco is programmed using C and C++ computer programming languages. Monaco runs on Windows operating system and off-the-shelf computer server/hardware.
The provided text is a 510(k) summary for the Monaco RTP System, an updated version of a previously cleared device. It largely focuses on demonstrating substantial equivalence to the predicate device and does not contain detailed acceptance criteria or a study proving the device meets them in the format requested. The document explicitly states: "No animal or clinical tests were performed to establish substantial equivalence with the predicate device." Therefore, I cannot provide a table of acceptance criteria, reported device performance, or details about a clinical study.
However, based on the non-clinical performance testing described, here's what information can be extracted:
1. A table of acceptance criteria and the reported device performance:
Based on the provided text, specific numerical acceptance criteria and corresponding reported device performance values are not available. The document states that "Design verification and performance testing were carried out in accordance with design controls... against design and risk management requirements at sub-system, integration and system levels." It also mentions "Software verification testing was conducted and documented in accordance with FDA quidance 1 for devices that pose a major level of concern (Class C per IEC 62304)." However, the actual criteria for these verifications (e.g., specific error margins for dose calculation, response times for image manipulation) and the measured performance against those criteria are not detailed.
2. Sample size used for the test set and the data provenance:
- Sample size used for the test set: Not specified. The document mentions "sub-system, integration and system levels" testing, but no specific number of test cases or data sets are provided.
- Data provenance: Not specified. As no clinical studies were performed, the "data" would refer to test cases, models, or simulated data used in the non-clinical verification. The origin of this data is not mentioned.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable/Not specified. Since no clinical studies or expert consensus activities are described for establishing ground truth on a test set, this information is not available.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable/None specified. No adjudication method is mentioned as there were no clinical studies.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No MRMC comparative effectiveness study was conducted. The device is a Radiation Treatment Planning (RTP) system, not an AI-assisted diagnostic tool that would typically involve human readers in this context. The document explicitly states "No animal or clinical tests were performed."
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- The document describes "Design verification and performance testing" and focuses on the system's ability to calculate dose, perform image manipulation, optimization, etc., which implies a standalone performance evaluation of the software components. However, specific performance metrics for the algorithm only without any human interaction involved in setting up the plan or interpreting the output are not quantified. The mention of "Software verification testing" suggests an evaluation of the algorithm's correctness.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the non-clinical testing, the "ground truth" would likely be established through:
- Reference data/models: For dose calculations, comparison against established physics models, phantom measurements, or other validated dose calculation systems.
- Known input/output pairs: For software functionalities like image manipulation or contouring, where the expected output for a given input is pre-defined.
- Compliance with standards: The document lists several ISO and IEC standards (e.g., ISO 62083 for radiotherapy treatment planning systems), implying that meeting the requirements of these standards serves as a form of "ground truth" for safety and performance.
8. The sample size for the training set:
- Not applicable/Not specified. The document does not describe the device as employing machine learning or AI that would require a "training set" in the conventional sense for a diagnostic algorithm. It describes a physics-based dose calculation and treatment planning system.
9. How the ground truth for the training set was established:
- Not applicable. As no training set is described, there's no information on how its ground truth would be established.
In summary, the provided document is a 510(k) premarket notification for an updated Radiation Treatment Planning (RTP) system. It focuses on demonstrating substantial equivalence to a predicate device through non-clinical verification and validation testing against design requirements and recognized standards, rather than providing details of a clinical study or specific quantitative acceptance criteria and performance data.
Ask a specific question about this device
(106 days)
Leksell Gamma Knife® is a teletherapy device intended for stereotactic irradiation of head structures ranging from very small target sizes of a few millimeters to several centimeters e.g.
- · Metastatic tumors
- Recurrent glioblastomas
- · Trigeminal neuralgia
- · Medically refractory essential tremor
- · Orbital tumors
- · Ocular tumors
- · Optic nerve tumors
· Benign diseases (such as meningiomas, vestibular schwannomas, post-surgical pituitary adenomas, craniopharyngioma, hemangioblastomas, schwannomas, arteriovenous malformations, chordomas, chordomas, glomus tumors, hemangiomas) - · Skull base tumors
· Head and neck tumors (such as unknown primary of the head and neck, oral cavity, hypopharynx, - nasopharynx, sinonasal, salivary gland)
· Pediatric tumors (such as glioma, ependymoma, pituitary tumors, hemangioblastoma, meningioma, metastasis, medulloblastoma, nasopharyngeal tumors, arteriovenous malformations, skull base tumors)
Leksell Gamma Knife® (available models Icon™, Perfexion™ and Elekta Esprit) is a radiosurgery system for use in the stereotactic irradiation of head structures. Surgery is achieved by delivering a prescribed dose as one or more shots of ionizing radiation to the exact site of the target.
The provided text does not contain detailed information about specific acceptance criteria and a study that proves the device meets those criteria as it relates to a specific performance metric (like diagnostic accuracy). Instead, it focuses on the substantial equivalence of a new device model (Leksell Gamma Knife® - Elekta Esprit) to a predicate device (Leksell Gamma Knife® K173789) through non-clinical performance testing.
Here's an analysis based on the provided text, highlighting what is and isn't available:
Overall Conclusion from the Text: The document claims the new device is substantially equivalent to the predicate device because its fundamental technical characteristics are unchanged despite hardware and software updates to the user interface. Therefore, the performance is assumed to be equivalent, and no new clinical studies were conducted for this 510(k) submission.
Missing Information (per the request, but not available in the provided text):
- A table of acceptance criteria and reported device performance for a specific performance characteristic (e.g., diagnostic accuracy, sensitivity, specificity).
- Sample size used for a "test set" in the context of diagnostic performance.
- Data provenance (country of origin, retrospective/prospective).
- Number of experts, their qualifications, and adjudication methods for establishing ground truth.
- Multi-reader multi-case (MRMC) comparative effectiveness study results.
- Standalone (algorithm only) performance.
- Specific type of ground truth used (pathology, outcomes data, etc.) for performance evaluation.
- Sample size for a "training set" for an algorithm.
- How ground truth for a "training set" was established.
Information Available from the Text (interpreted in the context of "acceptance criteria" as general device functionality and safety):
The "acceptance criteria" here are implicitly related to ensuring the device performs as intended and is as safe and effective as its predicate, primarily through non-clinical testing of hardware and software updates.
1. A table of acceptance criteria and the reported device performance:
The document describes non-clinical performance testing to ensure the new model functions as intended and meets requirements.
| Acceptance Criterion (Implicit) | Reported Device Performance (Summary) |
|---|---|
| Conformance to technical requirement specification | "Results from verification and validation testing demonstrate that conformance to applicable technical requirement specification, user needs have been met..." (Page 5) |
| Functionality and Performance as intended | "Testing in the form of module, integration and system level verification was performed to evaluate the performance and functionality of the new control system and the new hardware against requirement specification." (Page 5) "Results from verification and validation testing demonstrate... that the device functions as intended." (Page 5) |
| No undesirable effects from new/updated functionalities | "Regression test of unchanged functionalities in the developed system was done to ensure that new and updated functionalities did not introduce any undesirable effects." (Page 5) |
| Usability and design validation (meets intended use/user needs) | "Design and usability validation of the system have been performed by competent and professionally qualified personnel to ensure that the product fulfils the intended use and user needs." (Page 5) |
| Effectiveness of risk control measures related to safety (FRS) | "The design and usability validation was also made to ensure that the risk control measures associated with functions related to safety (FRS) for the affected functionality, mainly the user interface of the Control Panel were effective." (Page 5) |
| Biocompatibility | "Biocompatibility testing has been performed to ensure that all materials are biocompatible in relation to the use of the device." (Page 5) |
| Safety and Effectiveness Equivalence to Predicate | "The performance data demonstrate that the Leksell Gamma Knife - Elekta Esprit is as safe and effective and performs as well as the predicate devices Leksell Gamma Knife Icon and Leksell Gamma Knife Perfexion." (Page 5) "The device safety and performance have been addressed by non-clinical testing in conformance with predetermined performance criteria, FDA guidance, and recognized consensus standards." (Page 5) "The results of verification and validation as well as conformance to relevant safety standards demonstrate that the Leksell Gamma Knife® – Elekta Esprit meets the established safety and performance criteria and is substantially equivalent to the predicate device." (Page 5) |
2. Sample size used for the test set and the data provenance:
- Sample Size: Not specified in terms of patient cases or images, as the testing was non-clinical and focused on system functionality. It refers to "module, integration and system level verification testing."
- Data Provenance: Not applicable in the context of patient data. The testing was conducted internally by Elekta Solutions AB (Sweden).
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable to this 510(k) submission, as no clinical performance study involving diagnostic accuracy with expert-established ground truth was performed for substantial equivalence. "Design and usability validation... performed by competent and professionally qualified personnel" (Page 5) speaks to the expertise in verifying engineering and human factors aspects, not medical image interpretation.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
- Not applicable.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- No, an MRMC comparative effectiveness study was not done. The submission explicitly states: "No animal or clinical tests were performed to establish substantial equivalence with the predicate device." (Page 5). This device is a radiation therapy system, not an AI-assisted diagnostic tool for human readers.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
- Not applicable in the context of an algorithm's diagnostic performance. The device itself is a treatment system, and its standalone performance refers to its ability to deliver radiation as specified, validated through non-clinical tests.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):
- For the non-clinical testing, the "ground truth" was the device's technical specifications and intended functionality. Verification and validation ensured the device met these engineering and design requirements. There was no medical "ground truth" (e.g., pathology, clinical outcomes) established for diagnostic or treatment efficacy in a clinical setting in this submission.
8. The sample size for the training set:
- Not applicable, as this device is not an AI/ML algorithm trained on a dataset in the sense of image recognition or diagnostic prediction.
9. How the ground truth for the training set was established:
- Not applicable.
Ask a specific question about this device
(162 days)
The Monaco system is used to make treatment plans for patients with prescriptions for external beam radiation therapy. The system calculates dose for photon, electron, and proton treatment plans and displays, on-screen and in hard-copy, two- or threedimensional radiation dose distributions inside patients for given treatment plan set-ups.
The Monaco product line is intended for use in radiation treatment planning. It uses generally accepted methods for:
- contouring
- image manipulation
- simulation
- image fusion
- . plan optimization
- QA and plan review
The Monaco RTP System accepts patient diagnostic imaging data from CT and MR scans, and source dosimetry data, typically from a linear accelerator. The system then permits the user to display and define (contour) the target volume to be treated and critical structures which must not receive above a certain level of radiation, on these diagnostic images. Based on the prescribed dose, the user, a Dosimetrist or Medical Physicist, can then create multiple treatment scenarios involving the number, position(s) and energy of radiation beams and the use of a beam modifier (MLC, block, etc.) between the source of radiation and the patient to shape the beam. Monaco RTP system then produces a display of radiation dose distribution within the patient, indicating not only doses to the target volume but to surrounding tissue and structures. The optimal plan satisfying the prescription is then selected, one that maximizes dose to the target volume while minimizing dose to surrounding healthy volumes.
The parameters of the plan are output for later reference and for inclusion in the patient file.
Monaco planning methods and modalities:
- Intensity Modulated Radiation Treatment (IMRT) planning .
- . Electron, photon and proton treatment planning
- . Planning for dynamic delivery methods (e.g. dMLC, dynamic conformal, Volumetric Modulated Arc Therapy (VMAT))
- . Stereotactic planning and support of cone-based stereotactic
- . 3D conformal planning
- . Adaptive planning (e.g. for the Elekta Unity MR-Linac)
Monaco basic systems tools, characteristics, and functions:
- . Plan review tools
- . Manual and automated contouring tools
- DICOM connectivity .
- . Windows operating system
- . Simulation
- . Support for a variety of beam modifiers (e.g. MLCs, blocks, etc.)
- . Standardized uptake value (SUV)
- Specialty Image Creation (MIP, MinIP, and Avq) •
- . Monaco dose and Monitor Unit (MU) calculation:
- Dose calculation algorithms for electron, photon, proton planning .
Monaco is programmed using C and C++ computer programming languages. Monaco runs on Windows operating system and off-the-shelf computer server/hardware.
This document, K213787, is an FDA 510(k) premarket notification for the Elekta Monaco RTP System, Release 6.1. It asserts substantial equivalence to a predicate device, Monaco RTP System (K202789). The document focuses on the non-clinical performance testing of the device, particularly for enhancements in proton therapy functionality.
Here's an analysis of the acceptance criteria and the study that proves the device meets them, based on the provided text:
1. A table of acceptance criteria and the reported device performance
The document does not provide a specific table with numerical acceptance criteria and corresponding reported device performance values in a format like "Target Value (X%) vs. Observed Value (Y%)". Instead, it describes a more qualitative approach to verifying the enhancements.
However, based on the "SUMMARY OF PERFORMACE TESTING (NON-CLINICAL)" section, we can infer the following:
| Acceptance Criteria Category | Reported Device Performance |
|---|---|
| General Device Performance & Functionality | "Development, verification, and validation activities for the modified system were carried out in accordance with design controls... applicable ISO 13485 Quality Management System requirements, ISO 14971 Risk Management requirements, and IEC 62304 requirements for software life-cycle processes. Non-clinical testing was performed to evaluate device performance and functionality in accordance with design and risk management requirements at subsystem, integration and system levels including interoperability." |
| LET Calculation Accuracy (Monoenergetic Spots) | "The LET to water and medium were calculated for monoenergetic spots in Monaco and Geant4 for a range of energies and materials. The results were quantitatively compared to each other as well as qualitatively compared to other published results." (Implies the results were acceptably close to reference values). |
| LET Distribution Accuracy (Complex Spot/Beam Arrangements) | "LET distributions for complex multiple spot and multiple beam arrangements as well as plan summations were calculated in Monaco and compared to expected values as obtained through manual summation of individual spots according to the design equations." (Implies that Monaco's calculations matched the expected values). |
| Clinical Workflow Validation | "Formal validation of the clinical workflows has been performed on a clinically representative production equivalent system by competent and professionally qualified personnel." (Implies successful validation). |
| Safety and Performance (Overall) | "The device safety and performance have been addressed by non-clinical testing in conformance with predetermined performance criteria, FDA guidance, and recognized consensus standards. The results of verification and validation as well as conformance to relevant safety standards demonstrate that the Monaco RTP System meets the established safety and performance criteria and is substantially equivalent to the predicate device." |
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
The document does not explicitly state a "sample size" in terms of number of patient cases for the non-clinical testing. The tests described are computational comparisons (Monaco vs. Geant4, Monaco vs. manual summation) for various energies, materials, spot arrangements, and beam arrangements. These are not patient-specific data sets but rather simulated or theoretical scenarios designed to evaluate the computational accuracy of the new features.
There is no mention of "country of origin" or whether it was "retrospective or prospective" as these terms typically apply to studies involving patient data, which this non-clinical testing does not appear to use.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
For the non-clinical tests described:
- LET calculation accuracy: The ground truth for monoenergetic spots was established by comparison to Geant4 (a Monte Carlo simulation toolkit) and "other published results". This implies a reliance on established physics models and potentially peer-reviewed literature rather than human expert interpretation of images.
- LET distribution accuracy: Ground truth for complex arrangements was established by "expected values as obtained through manual summation of individual spots according to the design equations." This suggests a mathematical derivation as the ground truth.
- Clinical workflows: "Formal validation... by competent and professionally qualified personnel." No specific number or detailed qualifications (e.g., "radiologist with 10 years of experience") are provided for these personnel.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
Given that the non-clinical tests involve comparisons to established algorithms (Geant4), published results, and mathematical derivations, there is no mention or indication of an adjudication method like "2+1" or "3+1", which are typically used for establishing consensus among human interpreters. The comparisons are to objective, established computational or theoretical benchmarks.
5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
No, a multi-reader, multi-case (MRMC) comparative effectiveness study was not done. The document explicitly states: "No animal or clinical tests were performed to establish substantial equivalence with the predicate device." Therefore, there is no information on how much human readers improve with or without AI assistance, as the changes are to the dose calculation algorithms themselves (new proton functionalities like robust optimization, robust evaluation, and LET calculation) rather than an AI-assisted diagnostic or contouring tool that directly impacts human reader performance.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
Yes, the described performance testing is primarily a standalone (algorithm only) evaluation. The comparisons are:
- Monaco vs. Geant4 (algorithm vs. algorithm/physics model)
- Monaco vs. manual summation based on design equations (algorithm vs. mathematical derivation)
The "clinical workflow validation" does involve "competent and professionally qualified personnel" interacting with the system, but the core performance evaluation of the new proton features (LET, robustness) is algorithmic.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)
The ground truth used for the non-clinical testing appears to be a combination of:
- Established physics models/simulations: Geant4 for monoenergetic LET calculations.
- Mathematical derivations/design equations: For complex LET distributions.
- "Published results": For qualitative comparison of monoenergetic LET.
There is no mention of expert consensus (for image interpretation), pathology, or outcomes data being used as ground truth for these specific non-clinical tests.
8. The sample size for the training set
The document does not mention a "training set" or "training data". This is because the Monaco RTP System (as described in this submission) uses deterministic algorithms for dose calculation (e.g., Monte Carlo algorithm) rather than machine learning or AI models that require a separate training phase. The "development" mentioned refers to software engineering and algorithm implementation, not machine learning model training.
9. How the ground truth for the training set was established
Since no training set is mentioned or implied for the deterministic algorithms described, this question is not applicable based on the provided document.
Ask a specific question about this device