Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K133017

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2013-10-07

    (11 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Estech COBRA® Genesis Bipolar Clamp is intended for the coagulation of soft tissue during general surgery. The system may also be used to coagulate blood and soft tissue to produce hemostasis.

    Device Description

    The Estech COBRA® Genesis Bipolar Clamp is a sterile, single-use device designed to guide RF energy to the target tissue via electrodes in the jaws of a surgical clamp style device. RF energy is delivered in a bipolar mode, with one jaw of the device as the ablative electrode and the other jaw as the indifferent electrode.

    AI/ML Overview

    Here's an analysis of the provided text regarding the acceptance criteria and supporting studies for the Estech COBRA® Genesis Bipolar Clamp:

    Please note: The provided document is a 510(k) summary, which is typically a high-level overview. It states that detailed testing information is available in "Section 18" and "Sections 15 and 17," but these sections are not included in the provided text. Therefore, several of the requested details about specific study methodologies (like sample sizes for specific tests, expert qualifications, adjudication methods, or MRMC studies) are not present in this summary.


    Acceptance Criteria and Reported Device Performance

    The provided 510(k) summary does not explicitly list performance acceptance criteria in a quantifiable table format. Instead, it asserts that the device is "substantially equivalent" to a predicate device (Estech COBRA Bipolar System II, K100224) based on various assessments. The "acceptance criteria" are implicitly met by demonstrating this substantial equivalence.

    Here's a table summarizing the areas of comparison:

    Acceptance Criterion (Implicit)Reported Device Performance (Estech COBRA® Genesis)
    Indications for Use EquivalenceIdentical to predicate device (coagulation of soft tissue during general surgery; coagulation of blood and soft tissue for hemostasis).
    Technology EquivalenceIdentical to predicate device (RF energy delivered in bipolar mode, one jaw ablative, other indifferent). Minor design differences (non-rotatable jaws, single active electrode) do not alter performance.
    Manufacturing Process EquivalenceIdentical to predicate device.
    Packaging EquivalenceIdentical to predicate device.
    Sterilization Method EquivalenceIdentical to predicate device.
    Mechanical IntegrityJoints and other mechanical aspects are "adequate for use" (demonstrated in Section 18, not provided).
    Ablation Performance Equivalence"Equivalent to that of the predicate" (demonstrated in Section 18, not provided).
    BiocompatibilityPatient contacting materials "demonstrated to be biocompatible" in accordance with ISO 10993 (testing in Sections 15 and 17, not provided).
    Electrical Safety (High Frequency Surgical Equipment)Conforms to pertinent sections of IEC 60601-2-2 (testing in Sections 15 and 17, not provided).

    Details of the Study/Testing That Proves the Device Meets Acceptance Criteria:

    The provided document describes the types of studies performed but lacks specific quantitative details about their methodology, sample sizes, or outcomes.

    1. Sample sizes used for the test set and the data provenance:

      • Mechanical Integrity & Ablation Performance: The document states "The testing supplied in Section 18 demonstrates that the joints and other mechanical aspects... are adequate for use. Additionally, the data demonstrates that the ablation performance... is equivalent to that of the predicate."
        • Sample Size: Not specified in this summary. The detailed report in "Section 18" would contain this information.
        • Data Provenance: Not specified (e.g., in-vitro, ex-vivo, animal model, specific country, etc.). This would likely be found in "Section 18."
        • Retrospective/Prospective: Not specified.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

      • This information is not provided in the summary. For mechanical and ablation performance, "ground truth" would likely be established through objective measurements against predefined standards or comparative measurements against the predicate device, rather than expert consensus on a test set.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

      • Not applicable/not specified for mechanical and ablation performance testing described. These are typically objective measurements.
    4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

      • No, an MRMC study is not mentioned or applicable. This device is an electrosurgical clamp, not an AI-powered diagnostic or assistive tool for human readers.
    5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

      • Not applicable. This is a physical medical device. The "standalone" performance would be its direct mechanical and electrical function.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

      • For Mechanical Integrity and Ablation Performance: Ground truth would likely be established through objective physical and functional measurements and comparisons to the predicate device's established performance specifications, rather than clinical outcomes or pathology.
      • For Biocompatibility: Ground truth is established by adherence to ISO 10993 standards and specific biological testing (e.g., cytotoxicity, sensitization, irritation).
      • For Electrical Safety: Ground truth is established by adherence to IEC 60601-2-2 standards and electrical testing.
    7. The sample size for the training set:

      • Not applicable. This device does not involve a "training set" in the context of machine learning. The design and manufacturing processes are established based on engineering principles and regulatory standards.
    8. How the ground truth for the training set was established:

      • Not applicable for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K053021

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2006-01-26

    (92 days)

    Product Code
    Regulation Number
    870.4420
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ESTECH Clearview™ MV Atrial Depressor is intended to remove excess fluids from the surgical field for filtering and return to the patient during cardiopulmonary bypass of up to 6 hours. It can also be used to retract tissue to increase visibility.

    Device Description

    The Estech Clearview™ MV Atrial Depressor is a sterile, single use, surgical instruments intended for the collection of blood and other fluids from the surgical field for return through the cardiotomy system during open heart surgery. The a molded polymer proximal connector is intended to connect into cardiopulmonary bypass system to return blood and fluids into the extracorporeal circulation circuit, most typically during stoppedheart surgical procedures. The Estech Clearview™ MV Atrial Depressor has an integrated shaft designed to connect to the Estech Small Incision Retractor. The retractor creates access to the surgical field via thoracotomy or mini-sternotomy. When mounted to the retractor, the Estech Clearview™ MV Atrial Depressor can also be used to retract tissues while collecting fluids.

    AI/ML Overview

    The provided text describes a medical device, the ESTECH Clearview™ MV Atrial Depressor, and its regulatory submission. It does not contain information about acceptance criteria for device performance, nor details of a study (clinical or otherwise) proving it meets such criteria. The document is primarily a 510(k) summary for regulatory approval, focusing on substantial equivalence to predicate devices, biocompatibility, and preclinical performance data.

    Therefore, I cannot provide the requested information. The text explicitly states "Preclinical performance data was supplied to demonstrate that the Estech Clearview™M MV Atrial Depressor can meet its labeled performance claims, and to demonstrate substantial equivalence with the predicate device," but it does not provide the details of this data or the specific acceptance criteria.

    The questions you've asked (about sample sizes, ground truth, expert qualifications, MRMC studies, standalone performance, etc.) are typical for studies assessing the diagnostic accuracy or effectiveness of AI/imaging devices. The device described here is a surgical instrument for fluid collection and tissue retraction, and its performance evaluation would likely focus on different metrics (e.g., fluid collection rate, suction integrity, material strength, biocompatibility, ease of use in a surgical setting) rather than accuracy against a ground truth in an AI context.

    Ask a Question

    Ask a specific question about this device

    K Number
    K052081

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-12-16

    (136 days)

    Product Code
    Regulation Number
    870.4210
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ESTECH Remote Access Perfusion Femoral Venous Cannula is intended for use as a venous drainage cannula during cardiopulmonary bypass up to 6 hours.

    Device Description

    The Estech Remote Access Perfusion (RAP™) Femoral Venous Cannula is a sterile, single-use, open-ended, hollow polymer tube with multiple perforations at the distal end. The barbed proximal end is intended to connect into cardiopulmonary bypass tubing to provide extracorporeal circulation of the blood, most typically during stopped-heart surgical procedures. The RAP FV cannula is provided with a flexible obturator to assist with the placement. The RAP FV cannula can be guided over a 0.035" guidewire. The RAP FV cannula has depth markers to assist the physician with placement. The RAP FV cannula ranges in outside diameter from 18 - 26 Fr, and in length from 60 -76 cm.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for a medical device (ESTECH Remote Access Perfusion FV Catheter). The information requested pertains to the acceptance criteria and the study proving the device meets those criteria, which is typical for AI/ML device submissions. However, this document does not contain any information about an AI/ML device.

    The document describes a physical medical device (a cardiopulmonary bypass cannula) and its substantial equivalence to predicate devices. It discusses:

    • Device Description: A sterile, single-use, open-ended, hollow polymer tube with multiple perforations.
    • Intended Use: As a venous drainage cannula during cardiopulmonary bypass.
    • Comparison to Predicate Device: Substantially equivalent in terms of sterile, single-use, open-ended, hollow polymer tubes inserted into the vena cava via the femoral vein.
    • Supporting Data: Biocompatibility testing consistent with ISO 11193 and preclinical performance data to demonstrate labeled performance claims and substantial equivalence.

    Therefore, I cannot provide the requested information (acceptance criteria, study details, sample sizes, ground truth establishment, MRMC studies, standalone performance) because these concepts are not applicable to the non-AI/ML device described in the provided text.

    Ask a Question

    Ask a specific question about this device

    K Number
    K051749

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2005-09-13

    (76 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Estech Cobra Surgical System is intended for the coagulation of cardiac tissue using radiofrequency (RF) energy during cardiac surgery. The System can be used during general surgery to coagulate soft tissues. The System may also be used to coagulate blood and soft tissue to produce hemostasis.

    Device Description

    The Estech Cobra Surgical System is a family of devices including RF Ablation Probes, with either a malleable or flexible shafts, used in conjunction with the Cobra Electrosurgical Unit (ESU) and other accessories, including the Estech Cobra Bipolar Recording Electrode. The System is intended for use by surgeons for the coagulation of cardiac and soft tissues during open surgical procedures.

    AI/ML Overview

    This device (Estech Cobra Surgical System) is an electrosurgical device, and the submission is a 510(k) premarket notification for an amendment to the existing device. The information provided heavily focuses on regulatory comparisons to predicate devices rather than a detailed clinical study for acceptance criteria.

    Based on the provided text, the "acceptance criteria" for this amendment are primarily based on substantial equivalence to existing predicate devices.

    Here's an analysis of the provided text in relation to your questions, noting where information is not available:

    1. Table of Acceptance Criteria and Reported Device Performance

    The acceptance criteria are implied to be substantial equivalence to predicate devices in terms of intended use, technology, design, and materials, and satisfactory preclinical performance data for any modifications. Specific numerical performance metrics (e.g., coagulation depth, lesion size) are not explicitly stated as "acceptance criteria" in a table format with corresponding reported performance, as would be typical for a new device's clinical study.

    Acceptance Criteria CategoryReported Device Performance Assessment in Submission
    Substantial Equivalence (General)The XL version of the Estech Cobra Surgical System is substantially equivalent in intended use, technology, design and materials to the predicate device.
    The Estech Cobra Surgical Probe is substantially equivalent in intended use, technology, design and materials to the predicate system.
    The Estech Cobra Bipolar Recording Electrode is substantially equivalent to the surgical console.
    BiocompatibilityNo biocompatibility data is necessary since the materials in the modified ablation probes (XL and Surgical Probe) are identical to those in the ablation probes described in the predicate 510(k)s.
    Preclinical Performance - XL VersionPreclinical performance data is supplied to demonstrate that the XL version of the cooled probe meets pertinent specifications as the predicate version.
    Preclinical Performance - Surgical ProbeNo performance data is necessary for the Estech Cobra Surgical Probe since it is identical in design and manufacture to the predicate device.
    Preclinical Performance - Bipolar Recording ElectrodePreclinical performance data is supplied to demonstrate that the Estech Cobra Bipolar Recording Electrode operates as intended.
    Compliance with StandardsAdditionally, Estech certifies that the accessories are in compliance with the pertinent standard for electrosurgical devices, AAMI HF-18.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not describe a clinical "test set" with human subjects or a specific sample size for a study proving acceptance criteria. The performance data mentioned is "preclinical performance data." This usually refers to bench testing or animal studies.

    • Sample Size: Not specified for preclinical performance tests.
    • Data Provenance: "Preclinical performance data" implies laboratory or animal testing, not human clinical trials. No country of origin for such data is specified, but it would typically be conducted by the manufacturer or a contract research organization. The context suggests that the submission is from the US, implying data generated or submitted to the US FDA.
    • Retrospective or Prospective: Not applicable as it's not a clinical study involving patients. Preclinical data would be generated prospectively as part of the validation process.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    This information is not provided. Since the described data is "preclinical performance data" and not a clinical study with a human "test set" requiring ground truth established by experts, this section is not applicable.

    4. Adjudication Method for the Test Set

    Not applicable, as there is no described clinical "test set" requiring expert adjudication.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    No MRMC study is mentioned. The device is a surgical tool, not an imaging or diagnostic device that typically undergoes MRMC studies.

    6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

    Not applicable. This is a physical surgical device, not an algorithm.

    7. Type of Ground Truth Used

    For the preclinical data provided, the "ground truth" would be established by:

    • Engineering specifications: Performance of the device (e.g., power output, temperature control, lesion size) compared against predefined engineering targets derived from the predicate device's performance.
    • Compliance with standards: Meeting the requirements of AAMI HF-18.
    • Predicate device performance: The "ground truth" for the new XL version is the established performance characteristics and safety profile of the predicate device.

    8. Sample Size for the Training Set

    Not applicable. This is a physical medical device, not an AI/ML algorithm. There's no concept of a "training set" in this context.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for a physical electrosurgical device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K041599

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2004-07-27

    (43 days)

    Product Code
    Regulation Number
    878.4400
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Estech Cobra Adhere Surgical System is intended for the coagulation of cardiac tissue using radiofrequency (RF) energy during cardiac surgery. The System can be used during general surgery to coagulate soft tissues. The System may also be used to coagulate blood and soft tissue to produce hemostasis.

    Device Description

    The Cobra Adhere Surgical System is an Electrosurgical Probe, with either a malleable or flexible shaft, used in conjunction with the Cobra Electrosurgical Unit (ESU). The System is intended for use by surgeons for the coagulation of cardiac and soft tissues during open surgical procedures.

    AI/ML Overview

    This 510(k) submission does not contain a study to prove acceptance criteria for device performance. Instead, it relies on a "substantial equivalence" argument by asserting that the device is identical to an already legally marketed predicate device.

    Here's why and what's present in the provided text:

    • No Acceptance Criteria or Performance Data: The submission explicitly states: "No data supporting equivalence between the Estech Cobra Adhere Surgical System and the Boston Scientific Cobra Adhere Surgical System is necessary because both are the same device." This indicates that no new performance studies (and therefore no new acceptance criteria) were established for this specific device because it is considered a re-branding of an existing one.

    • Reliance on Predicate Device: The core of the submission is the claim that the "Estech Cobra Adhere Surgical System is identical in nearly every way to the Boston Scientific Cobra Adhere Surgical System." The only difference cited is "labeling changes necessary to list a different final manufacturer." This implies that the Estech device inherits the performance and safety profile of the predicate device.

    Given this, I cannot fill out the table and answer the study-specific questions as they pertain to a new performance study. However, I can summarize the information provided in the context of this "substantial equivalence" claim:

    Summary of Device Equivalence Claim

    The Estech Cobra Adhere Surgical System claims substantial equivalence to the Boston Scientific Cobra Adhere Surgical System. The basis for this claim is outlined below:

    Feature/AspectEstech Cobra Adhere Surgical SystemBoston Scientific Cobra Adhere Surgical System (Predicate)Comparison/Comment
    Acceptance CriteriaNot explicitly defined (A new study was not performed).Assumed to be met by the predicate device's prior marketing clearance.The acceptance criteria for the Estech device are implicitly met by its claim of being identical to the predicate device, which had already established its safety and effectiveness. No new performance endpoints or targets were set for this 510(k).
    Reported Device PerformanceNot reported directly for this device.Assumed to be the same as the predicate device.Since the devices are asserted to be identical, the performance of the Estech device is considered to be the same as the Boston Scientific device. No new performance data is presented.

    Study-Specific Information (Based on the "Identical Device" Claim)

    1. Sample sized used for the test set and the data provenance: Not applicable. No new test set or data provenance is mentioned as this device is stated to be identical to the predicate.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. No new ground truth establishment for a test set was performed.
    3. Adjudication method (e.g. 2+1, 3+1, none) for the test set: Not applicable. No new test set for performance comparison was conducted.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an electrosurgical probe, not an AI-assisted diagnostic tool.
    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable. This device is an electrosurgical probe, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc): Not applicable. No primary ground truth was established for this submission, as it relies on the predicate device's established safety and effectiveness.
    7. The sample size for the training set: Not applicable. This device is an electrosurgical probe; there is no mention of a "training set" in the context of machine learning or algorithms.
    8. How the ground truth for the training set was established: Not applicable.

    Explanation of the "Identical Device" Argument:

    • Manufacturing: "Both devices are manufactured by Boston Scientific on the same assembly line in San Jose of the exact same materials, processes and specifications."
    • Sterilization: "Both are EtO sterilized in processes validated to ISO 11135."
    • Indications for Use: "Both devices have identical indications for use."

    Therefore, the review process for this device did not involve new performance studies against acceptance criteria but rather a verification of its sameness to the predicate device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K032632

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2003-10-22

    (57 days)

    Product Code
    Regulation Number
    870.4210
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ESTECH Remote Access Perfusion Cannula Left Axillary 21 French is intended for use in arterial perfusion of the aorta, via axillary artery, in cardiovascular surgery procedures requiring extracorporeal cardiopulmonary bypass (CPB). This device is indicated for short term cardiopulmonary bypass (< 6 hours). The device may also be used to occlude the ascending aorta, deliver cardioplegia solution and vent the aortic root.

    Device Description

    The ESTECH Remote Access Perfusion Cannula Left Axillary 21 French is a disposable 31 cm long flexible polyurethane tube with three (3) lumens with an inflatable polyurethane balloon at the distal end of the cannula. The inflatable balloon has elastomeric properties designed to provide aortic occlusion in a range of aorta sizes with internal diameters from 22 mm to 34 mm. The outside diameter of the cannula is 21 Fr. (7 mm). The cannula has a central lumen for the delivery of arterial blood through multiple distal outlets at flow rates of 1.0 to 5.0L/min, a lumen that communicates with the aorta in the area of the aortic root for delivery of cardioplegia solution and left ventricle venting, and a small lumen for control of the distal balloon. Radio-opaque arch segment of cannula and insertion depth marks aid in positioning the device. The Cannula are provided sterile in individual packages for single use.

    AI/ML Overview

    The provided 510(k) summary for the ESTECH Remote Access Perfusion Cannula Left Axillary 21 French describes non-clinical performance data and a comparison to predicate devices to establish substantial equivalence. It does not contain information about acceptance criteria or a study proving the device meets specific performance thresholds in a clinical or analytical efficacy study with human or expert-based ground truth.

    Here's an analysis based on the available text:

    1. Table of Acceptance Criteria and Reported Device Performance:

    The document does not explicitly state acceptance criteria in a quantitative, pass/fail format typical for clinical or standalone efficacy studies. Instead, it describes various performance tests conducted and generally states that the device "meets performance and safety specifications" or that results were "within specification." The comparison with predicate devices serves as the primary means of demonstrating equivalence rather than meeting independent quantitative acceptance criteria.

    Feature Tested/EvaluatedAcceptance Criteria (Implied)Reported Device Performance
    Occlusion Balloon Restricted Burst TestDemonstrate inflation characteristics up to burst in a restricted environment.Pressure data as a function of inflation volume were collected. (Implied: Data were acceptable and did not indicate premature or uncontrolled burst within intended use pressure range.)
    Occlusion Balloon Volume/Pressure/DiameterCharacterize inflation characteristics up to burst.Pressure and diameter data as a function of inflation volume were collected. (Implied: Data were acceptable and demonstrated consistent, predictable inflation across the specified ranges, including aortic occlusion in 22mm to 34mm aorta sizes).
    Occlusion Balloon Repeat Inflation TestDemonstrate balloon can be inflated repeatedly.Performed to demonstrate the balloon can be inflated repeatedly if necessary. (Implied: Test confirmed successful repeated inflation without failure or degradation.)
    Occlusion Balloon Sustained Inflation TestWithstand at least 6 hours of inflation.Performed to demonstrate that the balloon can withstand at least 6 hours of inflation. (Implied: Test confirmed successful sustained inflation for at least 6 hours.)
    Hemolysis TestingEvaluate cellular damage produced by the system compared to predicate.In-vitro test used to evaluate the relative effect of the device in the circuit as compared to the predicate devices. (Implied: Hemolysis levels were comparable to or better than predicate devices, or within acceptable limits). Note: "all cardiac bypass procedures produce some degree of hemolysis which may be at least partially compensated for in human and animal testing" suggests a comparative standard was used.
    Flow Testing - Arterial PerfusionPressure drop across cannula for 0-5 L/min flow rates to be within specification.Pressure drop across the cannula was measured for a range of arterial perfusion flow rates from 0-5 L/min and was within specification.
    Flow Testing - Cardioplegia DeliveryMinimum cardioplegia delivery flow rates for a given pressure to be within specification.Minimum cardioplegia delivery flow rates for a given pressure were also within specification.
    Biocompatibility TestingMaterials to be biocompatible.Completed for the currently marketed predicate device, and all materials used in the current device are identical. (Implied: The device shares the same biocompatibility profile as the predicate.)
    Animal TestingSuitable for intended use (insertion via axillary artery, placement, performance).Performed to demonstrate that the device was suitable for its intended use in terms of insertion via axillary artery and placement. Performance of the device, including the occlusion balloon and perfusion/cardioplegia flow, are known to be safe and effective for its intended use. (Implied: Successful insertion, placement, and functional performance were observed in animal models.)

    The study that proves the device meets the (implied) acceptance criteria:

    The "study" consists of a series of non-clinical laboratory and animal tests, designed to demonstrate the safety and effectiveness of the device and its substantial equivalence to predicate devices. These are described in Section G: "Summary of Non-Clinical Performance Data" and Section III: "Device Specifications" (though Section III is not provided in the input).

    2. Sample size used for the test set and the data provenance:

    • Sample Size: The document does not specify exact sample sizes for each test (e.g., number of balloons tested, number of flow tests performed, number of animals used). It reports results generally.
    • Data Provenance: All data referenced are from "laboratory testing" and "animal testing." This indicates in-vitro (bench testing) and in-vivo (animal) studies. No human data or external (country of origin) data provenance is explicitly stated, which is typical for 510(k) submissions that rely on predicate device equivalence. It explicitly states "The laboratory testing adhered to Good Laboratory Practices guidelines."

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    This information is not provided in the document. The tests performed are primarily engineering and simulated physiological tests, not requiring clinical expert ground truth in the way an AI diagnostic device would. For animal testing, veterinary experts would likely be involved in assessing outcomes, but their number and specific qualifications are not detailed.

    4. Adjudication method for the test set:

    This information is not applicable and therefore not provided. The testing described is objective, physical/physiological performance measurement, not subjective expert assessment requiring adjudication.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    This information is not applicable and therefore not provided. The device is a medical instrument (cannula) and not an AI-powered diagnostic or assistive tool.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    This information is not applicable and therefore not provided. The device is a physical medical instrument, not an algorithm.

    7. The type of ground truth used:

    • Bench Test Performance: For tests like burst pressure, flow rates, and sustained inflation, the "ground truth" is established by the specified engineering and material performance standards relevant to the device's function and safety. These would refer to predefined acceptable ranges or limits established by engineering design, regulatory standards, or comparison to predicate devices.
    • Biocompatibility: Established by adherence to material standards and previous testing on identical materials used in predicate devices.
    • Animal Testing: The "ground truth" is the observed physiological and mechanical performance in an animal model, validated against expected outcomes for successful device operation (e.g., proper insertion, placement, perfusion, and balloon occlusion without adverse events).

    8. The sample size for the training set:

    This information is not applicable and therefore not provided. This device is not an AI algorithm and therefore does not have a "training set" in the machine learning sense. The "training" in the context of this device refers to "On-site customized training" for users, which is not a data set.

    9. How the ground truth for the training set was established:

    This information is not applicable and therefore not provided. As mentioned above, there is no "training set" in the context of an AI algorithm.

    Ask a Question

    Ask a specific question about this device

    K Number
    K990573

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    1999-03-11

    (16 days)

    Product Code
    Regulation Number
    870.4450
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ESTECH Arterial Remote Access Perfusion Cannula is intended for use in arterial perfusion of the aorta, via a femoral artery, in cardiovascular surgery procedures requiring extracorporeal cardiopulmonary bypass (CPB). The device may also be used to occlude the ascending aorta, deliver cardioplegia solution and vent the aortic root.

    Device Description

    The ESTECH Arterial Remote Access Perfusion Cannula is a sterile, disposable, flexible polyurethane tube with three integrated lumens and an inflatable polyurethane balloon at the distal end of the cannula. The outside diameter of the cannula is 21 French. The cannula has a central lumen for delivery of arterial blood through multiple distal outlets at a flow rate up to 5 liters per minute, a small lumen for delivery of cardioplegia or venting at the aortic root, and an additional smaller lumen for control of the distal balloon. The blood outlet ports are multiple elongated openings along the cannula proximal to the occlusion balloon. Radiopaque balloon markers and insertion depth marks aid in positioning the device. The cannulas are provided sterile in individual packages for single use.

    AI/ML Overview

    The provided S510(k) summary for the ESTECH Arterial Remote Access Perfusion Cannula focuses on demonstrating substantial equivalence to a predicate device rather than presenting a detailed study of its performance against explicit acceptance criteria. The document describes various tests performed but does not explicitly state quantitative acceptance criteria or report specific performance values from these tests that can be directly compared in a table.

    Therefore, many of the requested sections below cannot be fully answered due to the nature of the provided document, which is a regulatory submission for substantial equivalence based on a predicate device, rather than a standalone performance study with defined criteria.

    Here's an analysis of the provided information, addressing what can be inferred and highlighting what is missing:


    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative acceptance criteria for the tests performed, nor does it provide specific numerical performance results for the ESTECH Arterial Remote Access Perfusion Cannula that would allow for a direct comparison in a table format.

    The document implicitly suggests that the acceptance criterion for most tests was to demonstrate performance comparable or superior to the predicate device or within "ESTECH specifications." However, these specifications are not detailed or quantified.

    Test PerformedAcceptance Criteria (Explicitly Stated in Document)Reported Device Performance (Quantified in Document)
    Occlusion Balloon Volume/Pressure/DiameterNot explicitly stated; "document the inflation characteristics up to burst"Not quantified; states data was "collected"
    Occlusion Balloon Repeat InflationNot explicitly stated; "demonstrate that the balloon can be inflated repeatedly"Not quantified
    Occlusion Balloon Inflation HoldNot explicitly stated; "demonstrate that the balloon can withstand at least 6 hours of inflation"Not quantified
    Hemolysis TestingNot explicitly stated; "not induce a higher degree of hemolysis...than expected"Not quantified; "in vitro test was used to evaluate the relative effect"
    Flow Testing (Pressure Drop)"within ESTECH specifications" (specifications not provided)Not quantified for ESTECH; predicate's "outlet blood velocity...unacceptable"
    Biocompatibility TestingNot explicitly stated; "verify biocompatibility"States "performed on all critical components"
    Animal TestingNot explicitly stated; "evaluate handling characteristics and verify performance of all intended functions""complete function...and ease of use were demonstrated"
    Clinical InvestigationNot explicitly stated; "verify the safety, efficacy, and performance of the device"States "conducted to verify the safety, efficacy, and performance"

    2. Sample Size Used for the Test Set and the Data Provenance

    The document does not provide specific sample sizes for the test sets used in any of the described tests (Occlusion Balloon, Hemolysis, Flow, Biocompatibility, Animal Testing, Clinical Investigation).

    • Animal Testing: Mentioned as "Animal studies." No number of animals or specifics provided.
    • Clinical Investigation: Mentioned as "A Clinical Investigation." No number of participants or specifics provided.

    The data provenance is generally implied as laboratory testing conducted by ESTECH. The country of origin for the studies is not specified but the company location is in Danville, CA, USA, and the submission is to the FDA. The studies are prospective in nature, as they were conducted to test the performance of the new device.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

    This information is not provided in the document. The tests described (e.g., balloon inflation, hemolysis, flow) appear to be objective engineering and biological tests that do not typically rely on expert consensus for ground truth but rather on measured physical or chemical properties. For animal and clinical studies, the evaluation would be performed by veterinarians and clinicians, respectively, but their numbers and qualifications are not disclosed.

    4. Adjudication Method for the Test Set

    This information is not applicable and not provided. The tests described are not of a nature that would typically involve an adjudication method like 2+1 or 3+1 (which are common in image interpretation studies).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    This question is not applicable to the provided document. The ESTECH Arterial Remote Access Perfusion Cannula is a medical device (a cannula), not an AI algorithm or a diagnostic tool that involves human readers interpreting cases. Therefore, an MRMC study or the effect of AI assistance on human readers is irrelevant to this submission.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This question is not applicable. The device is a physical medical instrument, not an algorithm.

    7. The Type of Ground Truth Used

    The ground truth for the various tests would be based on:

    • Engineering Measurements: For occlusion balloon characteristics and flow testing (e.g., pressure, diameter, volume, burst pressure, flow rate).
    • Laboratory Assays: For hemolysis testing (e.g., free hemoglobin levels).
    • Standardized Biocompatibility Tests: For biocompatibility (e.g., cytotoxicity, irritation, sensitization).
    • Clinical Observations/Outcomes: For animal and clinical testing (e.g., successful cannulation, adequate perfusion, absence of adverse events, physiological parameters).

    8. The Sample Size for the Training Set

    This information is not applicable. This is a physical medical device, not an AI model that requires a training set.

    9. How the Ground Truth for the Training Set Was Established

    This information is not applicable, as there is no training set for a physical medical device.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1