Search Filters

Search Results

Found 7 results

510(k) Data Aggregation

    K Number
    K140846

    Validate with FDA (Live)

    Device Name
    EDS COMBO-RINSE
    Date Cleared
    2014-07-25

    (113 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    12 - 120
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EDS Combo-Rinse is intended to cleanse root canal systems by irrigating root canals.

    Device Description

    EDS Combo-Rinse is an aqueous solution for use during the irrigation phase of endodontic root canal preparation procedures. The device has a dual purpose of flushing out any loosened debris and removing the smear layer produced from root canal instrumentation.

    AI/ML Overview

    The provided text describes a 510(k) summary for the EDS Combo-Rinse device, which is an aqueous solution for use during root canal preparation. The submission focuses on demonstrating substantial equivalence to a predicate device (QMix™ 2in1 Endodontic Irrigating Solution). Therefore, the "acceptance criteria" and "study that proves the device meets the acceptance criteria" are framed in terms of equivalence to the predicate device, rather than independent performance metrics defined specifically for EDS Combo-Rinse.

    Here's the breakdown of the information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (based on predicate device equivalence)Reported Device Performance (for EDS Combo-Rinse)
    Biocompatibility equivalent to predicate device (QMix™ 2in1 Endodontic Irrigating Solution)Demonstrated equivalence to the predicate device in five biocompatibility tests according to ISO 10993.
    Smear layer removal performance equivalent to predicate device (QMix™ 2in1 Endodontic Irrigating Solution)Demonstrated equivalence to the predicate device in smear layer removal performance.
    Safety equivalent to legally marketed devices containing the same componentsPrior use of components in legally marketed devices (including predicate devices with common components) supports safety.
    Effectiveness equivalent to legally marketed devices containing the same componentsPrior use of components in legally marketed devices (including predicate devices with common components) supports effectiveness.

    2. Sample Size Used for the Test Set and Data Provenance

    The document does not explicitly state sample sizes for the biocompatibility or smear layer removal testing.
    Regarding data provenance:

    • The tests were conducted to determine equivalence to the predicate device.
    • The document implies these were non-clinical tests.
    • Country of origin is not specified but is assumed to be part of the manufacturer's testing process.
    • The studies were retrospective in the sense that they compared the new device to existing predicate devices, but the testing itself would have been prospective for the EDS Combo-Rinse samples.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. The "ground truth" here is the established safety and performance profile of the predicate device and the accepted standards (ISO 10993). There is no mention of human experts establishing ground truth for a test set in the context of device performance. The assessment is based on laboratory test results against established benchmarks.

    4. Adjudication Method for the Test Set

    Not applicable. There was no explicit adjudication method for a test set described, as the evaluation was based on objective, standardized tests (biocompatibility according to ISO 10993, and smear layer removal testing).

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No. This type of study is not relevant for a root canal irrigating solution and was not performed. The device is a physical/chemical irrigant, not an imaging or diagnostic AI tool.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Yes, in a sense. The biocompatibility and smear layer removal tests are standalone evaluations of the device's physical and chemical properties and performance, without human intervention in the device's function during testing. This is a "device only" performance assessment.

    7. The Type of Ground Truth Used

    The ground truth used for comparison was:

    • Established performance of the predicate device: QMix™ 2in1 Endodontic Irrigating Solution, particularly its biocompatibility and smear layer removal capabilities.
    • International standards: ISO 10993 for biocompatibility testing.
    • Regulatory precedence: The fact that components have been "used in legally marketed devices and were found safe for dental use."

    8. The Sample Size for the Training Set

    Not applicable. This device is not an AI/ML algorithm that requires a training set. The term "training set" is not relevant to this type of device submission.

    9. How the Ground Truth for the Training Set Was Established

    Not applicable, as there is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K132609

    Validate with FDA (Live)

    Date Cleared
    2014-01-22

    (155 days)

    Product Code
    Regulation Number
    872.4850
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EDS Ultrasonic Tips are intended for use by dental professionals for the removal of soft and hard tissue during endodontic root canal preparation procedures. They can also aid in the removal of endodontic posts and other intra-canal blockages.

    Device Description

    EDS Ultrasonic Tips are an accessory to a piezoelectric ultrasonic handpiece and scaler unit. These powered components are not included as part of the device submitted for application with the 510(k) submission. EDS Ultrasonic Tips are stainless steel and will be available in M3x0.5 and #5-40 thread with 6 different tip designs. 1 tip design is devised for post removal; the remaining 5 tips are designed for negotiating the various angles and directions of root canals.

    AI/ML Overview

    The provided text describes the acceptance criteria and the study conducted for the EDS Ultrasonic Tips.

    Here's the breakdown of the information requested:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance CriteriaReported Device Performance
    Breakage resistance testing"Mechanical testing was conducted in the form of breakage resistance testing to determine the appropriate shelf life for EDS Ultrasonic Tips." (Specific results or metrics are not detailed in the provided text, only that the test was conducted to establish shelf life.)

    2. Sample size used for the test set and the data provenance:

    • Sample size: Not explicitly stated in the provided text. The document mentions "Mechanical testing was conducted in the form of breakage resistance testing," but does not specify the number of tips tested.
    • Data provenance: Not explicitly stated. Given that the testing was conducted to support a 510(k) submission, it would typically be prospective testing conducted by the manufacturer, Essential Dental Systems.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. The study involved mechanical testing, not expert-based ground truth establishment for a diagnostic or similar device.

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • Not applicable. The study involved mechanical testing, not a method requiring adjudication of expert opinions.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, a multi-reader multi-case (MRMC) comparative effectiveness study was not done. The device is an ultrasonic tip used in endodontic procedures, not an AI-powered diagnostic tool requiring human reader studies.
    • The document explicitly states: "Discussion of Clinical Tests Performed: Not Applicable."

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not applicable. The device is a physical instrument, not an algorithm.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • For the breakage resistance testing, the "ground truth" would be established by engineering standards or specifications for material integrity and performance, likely measured through mechanical stress tests to determine the point of failure or acceptable loads. This is not categorized as expert consensus, pathology, or outcomes data.

    8. The sample size for the training set:

    • Not applicable. The device is a physical instrument, not an algorithm that requires a training set.

    9. How the ground truth for the training set was established:

    • Not applicable. As above, there is no training set for this type of device.
    Ask a Question

    Ask a specific question about this device

    K Number
    K111541

    Validate with FDA (Live)

    Date Cleared
    2012-03-16

    (288 days)

    Product Code
    Regulation Number
    872.4200
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Endo-Express® Handpiece is indicated for the removal of dentin, shaping and obturation of the root canal. This application area extends to endodontic procedures using NiTi or stainless steel endodontic instruments intended by the manufacturer for use in the mechanical preparation of root canals.

    Device Description

    Endo-Express® Handpiece is a reciprocating handpiece for use with NiTi or stainless steel instruments during endodontic procedures. The handpiece is driven by an air motor. The steps required to properly operate the handpiece are: push button to open chuck, insert instrument handle into chuck, release button to close chuck, run handpiece. After use, the handpiece should be lubricated and autoclaved. The contents of the kit are 1 Endo-Express® Handpiece, 1 Lubricant bottle, 1 nozzle attachment.

    AI/ML Overview

    The provided document is a 510(k) summary for the Endo-Express® Handpiece, stating its substantial equivalence to a predicate device. It does not contain information about acceptance criteria or a dedicated study with the specific components you requested (such as sample sizes, expert qualifications, and MRMC studies).

    Here's a breakdown of why this information isn't present in the provided text:

    • Nature of the Document: This is a premarket notification (510(k)) seeking clearance based on substantial equivalence. For Class I and many Class II devices, clinical trials with formal acceptance criteria, statistical analyses, and ground truth establishment are often not required if substantial equivalence to a legally marketed predicate device can be demonstrated through performance testing, comparison of technological characteristics, and intended use.
    • "Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as Follows: Not Applicable" (Section 7): This explicitly states that specific non-clinical tests were not performed in a way that would require detailed reporting of acceptance criteria and results in this summary. It implies that the demonstration of substantial equivalence relied on other factors, primarily comparison of intended use and technological characteristics to the predicate.
    • "Discussion of Clinical Tests Performed: Not Applicable" (Section 8): This confirms that no clinical studies were performed for the purpose of this 510(k) submission. Therefore, there would be no data on sample sizes, expert-established ground truth, MRMC studies, or standalone algorithm performance.

    Based on the provided document, the following information is available for your request:

    1. A table of acceptance criteria and the reported device performance:

    Acceptance CriteriaReported Device Performance
    Not explicitly stated in the document as formal acceptance criteria for a new study.The device is considered substantially equivalent to the predicate device (TEP E4R, K#962540) based on similar intended use and technological characteristics.

    (Note: The document implies the "acceptance" is based on the device having similar characteristics and intended use to the predicate. There are no numerical performance metrics or thresholds explicitly defined as acceptance criteria for this specific submission.)

    2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):

    • Sample Size for Test Set: Not applicable. No test set for performance comparison was described as part of a new study in this submission.
    • Data Provenance: Not applicable.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not applicable. No ground truth establishment by experts for a test set was described in this submission.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • Not applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not applicable. No MRMC study was performed or described. This device is a mechanical dental handpiece, not an AI-assisted diagnostic tool.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Not applicable. This device is a mechanical dental handpiece, not an algorithm.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

    • Not applicable.

    8. The sample size for the training set:

    • Not applicable.

    9. How the ground truth for the training set was established:

    • Not applicable.

    Summary of the study that proves the device meets the acceptance criteria:

    The document states, "Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as Follows: Not Applicable" and "Discussion of Clinical Tests Performed: Not Applicable."

    Instead of a specific performance study, the substantial equivalence was demonstrated by:

    • Comparison of Intended Use: The Endo-Express® Handpiece has the same intended use as the predicate device (TEP E4R).
    • Comparison of Technological Characteristics: The device has similar technological characteristics to the predicate, with the main differences being cosmetic appearance and maximum allowable motor speed. The key functional characteristics (latch type, file type, motor connection, reduction ratio) are identical to the predicate.

    The "study" proving the device meets the implicit acceptance criteria (i.e., being substantially equivalent to a legally marketed predicate) is primarily a direct comparison of specifications and intended use outlined in Section 6, rather than a clinical or detailed non-clinical performance study with explicit acceptance criteria and results.

    Ask a Question

    Ask a specific question about this device

    K Number
    K101504

    Validate with FDA (Live)

    Device Name
    TI-CORE FLOW
    Date Cleared
    2010-10-29

    (150 days)

    Product Code
    Regulation Number
    872.3275
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Ti-Core® Flow+ is indicated for vital or non-vital tooth core build-up (replacement of existing restorations and/or lost tooth structure), as a base prior to fabricating an indirect restoration, and as a post cement. Ti-Core® Flow+ is also indicated for cementing of endodontic posts, crowns, bridges, inlays, onlays, and for core build-ups.

    Device Description

    Ti-Core® Flow+ is a dual-cure resin cement that can be used as a luting agent and core material. It is packaged in the familiar double-barrel syringe and our materials with the puttless other dental materials. The steps required to oroperly express the cement are: remove cap, attach mixer, attach dispensing tip, press plunger. After use, the mixer/dispensing tip should be discarded and the cap replaced.

    The contents of the kit are
    2 - Dual chambered syringes each containing 9 am resin cement

    • 20 Mixers
      10 - Thin dispensing tips
      10 - Thick dispensing tips
    AI/ML Overview

    The provided document describes the acceptance criteria and the study conducted for the Ti-Core® Flow+ dental resin cement.

    Acceptance Criteria and Device Performance

    Acceptance Criteria (ISO Standard 4049)Reported Device Performance (Ti-Core® Flow+)Details
    Film thicknessMeets or exceeds ISO specificationNon-clinical performance testing conducted.
    Working and setting timeMeets or exceeds ISO specificationNon-clinical performance testing conducted.
    Sensitivity to ambient lightMeets or exceeds ISO specificationNon-clinical performance testing conducted.
    Depth of cureMeets or exceeds ISO specificationNon-clinical performance testing conducted.
    Flexural strengthMeets or exceeds ISO specificationNon-clinical performance testing conducted.
    Water solubility and sorptionMeets or exceeds ISO specificationNon-clinical performance testing conducted.
    RadiopacityDoes not meet due to "not claimed"The device's radiopacity is not claimed and thus does not meet the specification, but this is presented as acceptable.

    Study Information

    1. Sample size used for the test set and the data provenance:
      The document does not specify a distinct "test set" in the traditional sense for a clinical study. The performance evaluation was based on non-clinical tests conducted according to ISO standard 4049. The sample size for these non-clinical tests (e.g., number of specimens for film thickness, flexural strength) is not provided. The provenance of these non-clinical test materials (e.g., origin of the raw materials, specific lab where tests were conducted) is also not explicitly stated, though it's implied the testing was done by Essential Dental Systems or a contracted lab.

    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
      Not applicable. This was a non-clinical, in-vitro study based on ISO standard specifications, not a study requiring expert clinical ground truth assessment on human data.

    3. Adjudication method for the test set:
      Not applicable. This was a non-clinical, in-vitro study.

    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
      Not applicable. This document pertains to a dental resin cement, not an AI-powered diagnostic device, and no human reader studies are described.

    5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
      Not applicable. This is not an algorithm or AI device.

    6. The type of ground truth used:
      The "ground truth" for evaluating the device's performance against its acceptance criteria was the ISO standard 4049 specification limits for various physical and chemical properties of dental resin cements.

    7. The sample size for the training set:
      Not applicable. This is not a machine learning or AI device that requires a training set.

    8. How the ground truth for the training set was established:
      Not applicable. There is no training set for this device.

    Ask a Question

    Ask a specific question about this device

    K Number
    K072555

    Validate with FDA (Live)

    Device Name
    EDTA PLUS
    Date Cleared
    2007-12-13

    (94 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    EDTA Plus is a root canal cleanser designed to facilitate removal of dentinal debris from the walls of root canals prior to obturation

    Device Description

    EDTA Plus is a root canal cleanser for use in endodontic procedures. During/After endodontic instrumentation, the product should be used to clean the canal space and remove the smear layer before placement of the endodontic filling. The material can be delivered into the canal using a side-vented, closed tip irrigation needle.

    The device consists of a 16 oz. bottie of EDTA Plus, a peach-colored aqueous solution.

    AI/ML Overview

    The provided document is a 510(k) summary for a medical device called "EDTA Plus," a root canal cleanser. It does not contain any information regarding acceptance criteria, device performance studies, sample sizes, expert ground truth, adjudication methods, MRMC studies, standalone algorithm performance, or training set details.

    The document primarily focuses on:

    • Identification of the submitter and device.
    • Comparison to predicate devices (SmearClear® and Moyco-EDTA) for the purpose of demonstrating substantial equivalence.
    • Device description (16 oz. bottle of peach-colored aqueous solution).
    • Intended Use (root canal cleanser to facilitate removal of dentinal debris before obturation).
    • The FDA's decision letter granting substantial equivalence (K072555).

    In summary, the provided text does not contain the information requested for a detailed description of acceptance criteria and the study proving device performance as it pertains to AI or diagnostic device performance metrics. This type of submission (510(k) for a cleanser) typically relies on demonstrating equivalence to existing products based on composition and known mechanisms of action, rather than clinical performance studies with specific statistical endpoints and ground truth comparisons.

    Ask a Question

    Ask a specific question about this device

    K Number
    K070401

    Validate with FDA (Live)

    Device Name
    ENDO-CHX
    Date Cleared
    2007-07-11

    (149 days)

    Product Code
    Regulation Number
    N/A
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    Endo-CHXTM is intended to irrigate and cleanse root canal systems.
    ENDO-CHX™ is intended to irrigate and cleanse root canal systems.

    Device Description

    Endo-CHX™ is a root canal cleanser for use in endodontic procedures. After endodontic instrumentation, the product should be used to cleanse the canal space before placement of the endodontic filling. The material should be delivered into the canal using an irrigating neede. A side-vented, closed tip irrigation needle is most preferred.

    AI/ML Overview

    The provided document is a 510(k) summary for a root canal cleanser called Endo-CHX™. As such, it primarily focuses on establishing substantial equivalence to predicate devices rather than providing detailed acceptance criteria and a study demonstrating performance against those criteria in the way that might be expected for an AI-powered diagnostic device.

    Here's an analysis based on the information provided, while noting the limitations due to the nature of the device:

    1. Table of Acceptance Criteria and Reported Device Performance

    This information is not applicable in the traditional sense for this type of device and submission. For a root canal cleanser, "acceptance criteria" would typically relate to its chemical properties, biocompatibility, and efficacy in cleaning the root canal system, often demonstrated through in vitro or ex vivo studies. The 510(k) summary focuses on substantial equivalence to predicate devices rather than setting and meeting specific quantitative performance metrics like sensitivity or specificity.

    The "device performance" is implicitly established by demonstrating that it functions similarly to legally marketed predicate devices and is considered safe and effective.

    2. Sample Size Used for the Test Set and Data Provenance

    There is no test set in the context of an AI/diagnostic device performance study. The 510(k) relates to a chemical solution.

    3. Number of Experts Used to Establish Ground Truth and Qualifications

    This information is not applicable. Ground truth, expert consensus, or clinical adjudication as would be relevant for an AI/diagnostic device is not part of this submission type.

    4. Adjudication Method

    This information is not applicable.

    5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study

    This information is not applicable. This is not an AI-powered diagnostic device.

    6. Standalone Performance Study

    The submission mentions "Non-Clinical Testing: The performance and biocompatibility documentation provided the submission support the safety and effectiveness of the Endo-CHX root canal cleanser for the indicated uses." This suggests standalone testing was performed, but the details (e.g., specific tests, outcomes, quantitative measures) are not provided in this summary exhibit. It just states that such documentation exists and supports safety and effectiveness.

    7. Type of Ground Truth Used

    For non-clinical testing of a root canal cleanser, the "ground truth" would likely involve standardized in vitro models or ex vivo tooth specimens. The effectiveness would be measured by properties such as debris removal, antibacterial activity, or smear layer removal, often compared to controls or predicate devices. However, direct "ground truth" establishment in a clinical sense (like pathology for cancer) is not specified or generally applicable here.

    8. Sample Size for the Training Set

    There is no training set for this device as it is not an AI/machine learning product.

    9. How the Ground Truth for the Training Set was Established

    There is no training set for this device.


    Summary of Relevant Information from the Provided Text:

    • Acceptance Criteria & Performance: Not explicitly stated as quantitative metrics. Performance is inferred through "substantial equivalence" to predicate devices and general statements about non-clinical testing supporting safety and effectiveness.
    • Study Type: Non-clinical (e.g., in vitro or ex vivo) testing was performed for performance and biocompatibility. No clinical trials were deemed necessary ("Clinical Testing: Not applicable").
    • Predicate Devices: The device was compared to K061689 (Aquatine™ EC Endodontic Cleanser) and K053167 (BioPure™ MTAD™ Root Canal Cleanser). The comparison focuses on mechanism of action and composition.
    • Conclusion: The device is considered "substantially equivalent to the predicates and raises no issues of safety and effectiveness."
    Ask a Question

    Ask a specific question about this device

    K Number
    K063856

    Validate with FDA (Live)

    Date Cleared
    2007-02-06

    (40 days)

    Product Code
    Regulation Number
    872.3820
    Panel
    Dental
    Age Range
    All
    Reference & Predicate Devices
    N/A
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The EZ-Fill® Xpress Epoxy Root Canal Cement System is indicated for permanent sealing of root canals following established endodontic procedures.

    Device Description

    The EZ-Fill® Xpress Epoxy Root Canal Cement System is an obturation system for filling straight and minimally curved canals. The bi-directional spiral (cement carrier that fully coats the canal, but prevents excess cement form exiting apically) and epoxy root canal cement, combined with a single point technique, creates a seal equivalent to lateral condensation and thermoplastic gutta percha. The contents of the kit are identical except that '2 - Dual Chambered Syringes containing 9.5 gm of root canal sealer' and '20 - Mixing Tips' replace the '7.5 gm - Epoxy Root Canal Cement Gel, 8.0 gm - Powder (Epoxy Root Canal Cement), and 1 Measuring Scoop.' The full list of contents is as follows: 4 - Color Coded to ISO Size 25 Bi-Directional Spirals (1 - 21mm length and 3 - 25 mm length), 2 - Dual Chambered Syringes containing 9.5 gm root canal sealer, 20 - Mixing Tips.

    AI/ML Overview

    The provided document describes the EZ-Fill® Xpress Epoxy Root Canal Cement System and its substantial equivalence to predicate devices, but it does not contain a study that proves the device meets specific acceptance criteria in the format requested.

    The document focuses on demonstrating substantial equivalence through:

    • Comparison of device description, intended use, and technological characteristics to predicate devices.
    • Bench testing to show compliance with international standards (ADA Specification No.57 and ISO 6876) for physical properties.
    • Biocompatibility literature to show no new safety concerns compared to predicate devices.

    Therefore, many of the requested details about acceptance criteria, study design, and ground truth are Not Applicable (N/A) based on the provided text, as a direct study demonstrating achievement of criteria is not presented.

    Here's a breakdown of the available information:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance Criteria (Standard)Reported Device Performance
    ADA Specification No.57 (Dental Root Canal Filling Material)Meet/exceed ADA Specification No.57
    ISO 6876 (Endodontic Filling Materials)Meet/exceed ISO 6876
    Specific Physical Properties (Flow, Film Thickness, Dimensional Stability, Solubility, Disintegration)Meet/exceed these properties as per the above standards
    Bi-Directional Spiral Dimensional Specifications (implied)Checked to meet all required specifications
    Biocompatibility (implied: no new safety concerns)Literature shows no new safety/biocompatibility concerns compared to predicate devices

    2. Sample size used for the test set and the data provenance

    • Sample Size: Not specified. The document only mentions "Bench Testing performed" without detailing the number of samples or tests conducted.
    • Data Provenance (country of origin, retrospective/prospective): Not specified.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

    • Not Applicable. Bench testing against pre-defined physical property standards does not typically involve expert ground truth establishment in the same way clinical or diagnostic studies do. The "ground truth" here is the standard itself.

    4. Adjudication method for the test set

    • Not Applicable. See point 3.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • No. The document explicitly states: "Discussion of Clinical Tests Performed: Not Applicable." This is a material testing for a root canal cement, not a diagnostic AI device.

    6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • Not Applicable. This is not an AI/algorithm-based device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

    • For physical properties, the ground truth is based on the requirements and methodologies outlined in ADA Specification No.57 and ISO 6876.
    • For biocompatibility, the ground truth is established by biocompatibility literature and Material Safety Data Sheets, indicating that the materials do not raise new safety concerns compared to legally marketed predicate devices.

    8. The sample size for the training set

    • Not Applicable. This is not an AI/machine learning device that requires a training set.

    9. How the ground truth for the training set was established

    • Not Applicable. See point 8.
    Ask a Question

    Ask a specific question about this device

    Page 1 of 1