Search Filters

Search Results

Found 8 results

510(k) Data Aggregation

    K Number
    K113475

    Validate with FDA (Live)

    Date Cleared
    2012-03-20

    (119 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ESTECH Cobra Adhere XL2 is intended to ablate cardiac tissue during cardiac surgery using radiofrequency (RF) energy when connected directly to the Estech Electrosurgical Unit (ESU).

    The ESTECH Cobra Adhere XL 2 may be used for temporary cardiac pacing, sensing, recording, and stimulation during the evaluation of cardiac arrhythmias during surgery when connected to a temporary external pacemaker or recording device.

    Device Description

    The Estech COBRA Adhere XL 2 is a sterile, single use device intended for the ablation of The Esteon CODIA Cranel to deliver RF energy to the target tissue via electrodes in the device when connected to the COBRA Electrosurgical Unit (ESU).

    When the COBRA Adhere XL 2 is connected to an auxiliary temporary external pacemaker or recording device it can be used to provide temporary cardiac pacing, sensing, recording, and stimulation during the evaluation of cardiac arrhythmias during surgery. The COBRA Adhere XL 2 must be disconnected from the ESU and connected to a temporary external pacemaker or recording device using the accessory cable provided.

    AI/ML Overview

    The provided text is a 510(k) summary for the Estech COBRA Adhere XL 2, a surgical device for ablation of cardiac tissue. It describes the device, its intended use, and compares it to predicate devices to establish substantial equivalence.

    However, the document does NOT contain information about specific acceptance criteria or a detailed study proving the device meets acceptance criteria, as one would find in a clinical trial report or a performance study summary with quantified results.

    The text focuses on demonstrating substantial equivalence to predicate devices through technical comparisons and general non-clinical testing. It states that "Appropriate testing, included in this application, has been performed to ensure that the ESTECH COBRA Adhere XL 2 meets its product specifications" and that "In-vitro and in-vivo testing demonstrate that the COBRA Adhere XL2 is able to pace, sense, and stimulate and ablate cardiac tissue as safely and effectively as the predicate devices."

    Without quantitative performance metrics, sample sizes for test sets, ground truth details, expert qualifications, or comparative effectiveness studies, it is impossible to complete a table of acceptance criteria versus reported device performance or address most of the requested points.

    Therefore, the following points cannot be extracted from the provided text:

    • A table of acceptance criteria and the reported device performance
    • Sample sizes used for the test set
    • Data provenance for the test set
    • Number of experts used to establish the ground truth for the test set and their qualifications
    • Adjudication method for the test set
    • Information on a multi-reader multi-case (MRMC) comparative effectiveness study
    • Information on a standalone (algorithm only) performance study
    • Type of ground truth used for the test set
    • Sample size for the training set
    • How the ground truth for the training set was established

    The document focuses on non-clinical testing for safety and effectiveness in establishing substantial equivalence to prior devices, rather than a detailed clinical performance study with specific acceptance criteria as you've requested.

    Ask a Question

    Ask a specific question about this device

    K Number
    K110913

    Validate with FDA (Live)

    Date Cleared
    2011-09-23

    (175 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Estech Cobra Surgical System is intended for the coagulation of cardiac tissue using radiofrequency (RF) energy during cardiac surgery. The System can be used during general surgery to coagulate soft tissues. The System may also be used to coagulate blood and soft tissue to produce hemostasis.

    Device Description

    The ThermaShield is a flat, bell-shaped silicone pad that is designed to reliably fill the anatomic space between the target tissue of an RF ablation and adjacent non-target tissue. Its thickness and rounded edges provide tactile feedback to the surgeon and facilitates confidence in proper placement. The ThermaShield is primarily designed for use in open chest cardiac surgery. The addition of the ThermaShield to the ESTECH Cobra Surgical System provides a more convenient option for the physician to thermally insulate surrounding non-target tissue during RF ablations. The ThermaShield is intended to improve upon circumstances where no insulation is used or to replace the current use of a surgical glove as a thermal insulator between target and surrounding tissues.

    AI/ML Overview

    The provided text describes an update to an existing medical device, the ESTECH Cobra Surgical System, with the addition of a ThermaShield accessory. This is a special 510(k) submission, meaning the changes are not significant enough to warrant a new full review but rather an amendment to an existing clearance. As such, the study described is a bench test to demonstrate substantial equivalence, not a clinical trial with human subjects.

    Here's a breakdown of the requested information based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    The document does not explicitly state quantitative acceptance criteria in a table format. However, it indicates the criteria were met by demonstrating the ThermaShield's performance.

    Acceptance Criteria (Implied)Reported Device Performance
    Thermally insulate adjacent tissues during RF ablation"Bench data provided in Section 18 demonstrates that the ThermaShield performs as intended" and "demonstrates that the ThermaShield technology is equivalent or better than surgical gloves, commonly used clinically, in thermally insulating adjacent tissues and is an improvement over having no insulation."
    Not raise new types of safety or efficacy questions"Bench data provided in Section 18 demonstrates that the ThermaShield performs as intended and does not raise new types of safety or efficacy questions."
    Biocompatibility of patient-contacting materials"The patient contacting materials of the ESTECH ThermaShield technology have been demonstrated to be biocompatible, in accordance with ISO 10993: Biological Evaluation of Medical Devices."
    Safety of electrical design (Electromagnetic Compatibility)"the safety of the electrical design is in conformance with the pertinent sections of IEC 60601-1-2, (Second Edition, 2001)... and IEC 60601-2-2."
    Substantial Equivalence to Predicate Device"ESTECH concludes that the ESTECH Cobra Surgical System with ThermaShield accessory is substantially equivalent to the predicate ESTECH Cobra Surgical System."

    2. Sample Size Used for the Test Set and Data Provenance

    • Sample Size for Test Set: Not explicitly stated as this was a bench test, not a clinical study involving human or animal subjects. The "bench data" would refer to a series of experimental setups to measure thermal insulation properties.
    • Data Provenance: Bench test data, not explicitly stated origin (likely internal company testing). It is retrospective in the sense that it evaluates the performance of the newly designed accessory.

    3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

    Not applicable. This device is an accessory to an electrosurgical system, and the evaluation was based on objective thermal insulation measurements, not expert interpretation of images or symptoms.

    4. Adjudication Method for the Test Set

    Not applicable. The "test set" here refers to bench testing, and data would be gathered and analyzed by technical personnel, not adjudicated by experts in the context of clinical interpretation.

    5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

    Not applicable. This document describes a physical medical device accessory, not an AI-based diagnostic or assistive technology.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    Not applicable, as this is not an algorithm-based device. However, "standalone" in a different context could refer to the ThermaShield's performance when incorporated into the Cobra Surgical System, which was indeed evaluated in the bench tests.

    7. The Type of Ground Truth Used

    The ground truth used was objective physical measurements of thermal insulation properties during bench testing, as well as compliance with established biocompatibility and electrical safety standards. This is not ground truth in the pathology/outcomes sense.

    8. The Sample Size for the Training Set

    Not applicable. This is not a machine learning or AI-based device, so there is no concept of a "training set."

    9. How the Ground Truth for the Training Set Was Established

    Not applicable. As there is no training set for an AI/ML model, no ground truth was established for it.

    Ask a Question

    Ask a specific question about this device

    K Number
    K100224

    Validate with FDA (Live)

    Date Cleared
    2010-04-07

    (71 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ESTECH Cobra Bipolar II System is intended for the coagulation of soft tissue during general surgery. The system may also be used to coagulate blood and soft tissue to produce hemostasis.

    Device Description

    The ESTECH Cobra Bipolar System is designed to guide RF energy to the target tissue via electrodes in the jaws of a surgical clamp style device. RF energy is delivered in a bipolar mode, with one jaw of the device as the ablative electrode and the other jaw as the indifferent electrode.

    AI/ML Overview

    The provided text does not describe a study involving a medical device with acceptance criteria and reported performance in the context of an AI/algorithm-based system. Instead, it is a 510(k) premarket notification for a traditional medical device, the ESTECH Cobra Bipolar II System.

    The document states that the new device is "identical in indications for use, technology, manufacture, packaging and sterilization" to a predicate device (ESTECH Cobra Bipolar System, K053100). The "acceptance criteria" and "study" described are primarily related to demonstrating equivalence to the predicate device and ensuring the safety and basic functionality of the new design.

    Therefore, many of the requested fields are not applicable to this type of submission. Below is a breakdown based on the information available:

    1. Table of Acceptance Criteria and Reported Device Performance

    Category/CriteriaAcceptance CriteriaReported Device Performance
    Mechanical IntegrityAdequacy for use of joints and other mechanical aspects."The bench testing supplied in Section 18 demonstrates that the joints and other mechanical aspects of the design of the ESTECH Cobra Bipolar II System are adequate for use."
    Electrode PerformancePerformance equivalent to the predicate device (ESTECH Cobra Bipolar System)."Additionally, the data demonstrates that the performance of the ESTECH Cobra Bipolar II System electrodes is equivalent to that of the predicate."
    BiocompatibilityConformance with ISO 10993: Biological Evaluation of Medical Devices."The patient contacting materials of the ESTECH Cobra Bipolar II System have been demonstrated to be biocompatible, in accordance with ISO 10993: Biological Evaluation of Medical Devices..."
    Electrical SafetyConformance with pertinent sections of IEC 60601-2-2: Medical electrical equipment - Part 2-2: Particular requirements for the safety of high frequency surgical equipment."...and the safety of the electrical design is in conformance with the pertinent sections of IEC 60601-2-2: Medical electrical equipment - Part 2-2: Particular requirements for the safety of high frequency surgical equipment."
    Substantial Equivalence (Overall)The device is "identical in indications for use, technology, manufacture, packaging and sterilization" to the predicate, with design differences not altering performance."ESTECH concludes that the ESTECH Cobra Bipolar II System is substantially equivalent to the predicate ESTECH Cobra Bipolar System (K053100)." (FDA concurred with this conclusion).

    2. Sample size used for the test set and the data provenance: Not applicable. The "study" here involves bench testing of device components and performance characteristics, not a clinical study with a "test set" of patient data or images.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a clinical "test set" is not relevant for this type of device submission. The "ground truth" for the device's performance is established by established engineering and international standards (e.g., ISO, IEC).

    4. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable.

    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is an electrosurgical tool, not an AI-assisted diagnostic or interpretative system that would involve human readers.

    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This is not an algorithm or AI device.

    7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" is based on established engineering principles, international standards (ISO 10993 for biocompatibility, IEC 60601-2-2 for electrical safety), and comparison to the well-understood performance of the predicate device.

    8. The sample size for the training set: Not applicable. There is no AI model or algorithm being trained.

    9. How the ground truth for the training set was established: Not applicable.

    Ask a Question

    Ask a specific question about this device

    K Number
    K091542

    Validate with FDA (Live)

    Date Cleared
    2009-10-20

    (147 days)

    Product Code
    Regulation Number
    870.4210
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ESTECH EasyFlow Cannula with Guidewire is intended for use with cardiopulmonary bypass as an arterial return or venous drainage cannula for up to 6 hours.

    Device Description

    The ESTECH EasyFlow Cannula with Guidewire is a sterile, single-use device. The single-lumen wire reinforced polymer tube incorporates multiple perforations at the distal end to provide a diffuse flow pattern. The barbed proximal end is intended to connect into cardiopulmonary bypass tubing to provide extracorporeal circulation of the blood, most typically during stopped-heart surgical procedures. Each cannula is provided with a flexible obturator to assist with the placement and positioning of the cannula and an optional stylet which inserts into the obturator to provide additional column strength for insertion. The product also comes with a 0.035" diameter guidewire, up to 3 vessel dilators, an 18 gauge. needle, and scalpel for introduction

    AI/ML Overview

    The provided text describes the ESTECH EasyFlow Cannula with Guidewire and its premarket notification for substantial equivalence, but it does not contain information related to acceptance criteria, specific device performance numbers, or a study design with sample sizes, ground truth establishment, or expert involvement.

    The document states:

    • "Bench testing has demonstrated that the device meets the proposed product specifications."
    • "Biocompatibility analysis demonstrates that the device is in compliance with ISO 10993"

    However, it does not provide details on:

    1. A table of acceptance criteria and the reported device performance: The document mentions "proposed product specifications" and "compliance with ISO 10993" but does not detail what these specifications or the outcome of the compliance testing were.
    2. Sample size used for the test set and the data provenance: No sample sizes or data provenance are mentioned for any testing.
    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable as no ground truth creation process is described.
    4. Adjudication method for the test set: Not applicable as no ground truth creation process is described.
    5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a medical cannula, not an AI-assisted diagnostic tool.
    6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done: Not applicable. This device is a medical cannula, not an algorithm.
    7. The type of ground truth used: Not applicable. The "ground truth" for a medical device like this would typically involve engineering specifications, material properties, and functional performance metrics, not diagnostic "ground truth."
    8. The sample size for the training set: Not applicable as there is no mention of a "training set" for this type of device.
    9. How the ground truth for the training set was established: Not applicable.

    In summary, the provided document focuses on establishing substantial equivalence to predicate devices based on design, materials, manufacturing, packaging, and intended use, along with general statements about bench testing and biocompatibility. It does not present a detailed study with specific acceptance criteria and performance data in the format requested.

    Ask a Question

    Ask a specific question about this device

    K Number
    K091350

    Validate with FDA (Live)

    Date Cleared
    2009-09-15

    (131 days)

    Product Code
    Regulation Number
    876.1500
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    N/A
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Hawkeye™ Introducer Systems, and accessories, are intended for use in minimally invasive surgery allowing access for delivery and placement of surgical instruments (e.g., ESTECH COBRA Adhere XL, AFfirm Pacing Probe). They are indicated for patients requiring blunt dissection of tissue including structures in the thoracic space.

    Device Description

    The Hawkeye family consists of two products; the ESTECH Hawkeye Magnetic Introducer System and the ESTECH Hawkeye Scope System.

    The Hawkeye Magnetic Introducer System is comprised of three main components:

      1. Hawkeye Magnetic Cannula Assembly
      1. Magnetic Introducer
      1. Malleable Stylet

    Both Systems allow for endoscopic visualization within the thoracic cavity, blunt dissection, and facilitate the placement of surgical equipment. These pieces may be packaged, sterilized and sold separately. They are single-use only.

    AI/ML Overview

    Please provide the acceptance criteria and the study that proves the device meets the acceptance criteria. The requested information is specific to clinical performance, interpretability, or diagnostic accuracy studies.

    The provided document describes the Hawkeye™ Introducer Systems, a device intended for minimally invasive surgery to allow access for delivery and placement of surgical instruments and for blunt dissection of tissue.

    However, the document does not contain information regarding:

    • Any specific acceptance criteria related to the device's clinical performance, interpretability, or diagnostic accuracy (e.g., sensitivity, specificity, accuracy thresholds).
    • Details of a study that tested the device against such criteria. The "Summary of Supporting Data" only mentions:
      • Biocompatibility analysis (compliance with ISO 10993)
      • Bench testing (met proposed product specifications)
      • Animal testing
      • Cadaver testing

    Therefore, I cannot provide the requested table or answer the specific questions about sample sizes, ground truth establishment, expert qualifications, adjudication methods, MRMC studies, or standalone performance for a clinical or diagnostic accuracy study. The document focuses on establishing substantial equivalence based on material properties, design, and intended use compared to a predicate device, rather than providing clinical performance data.

    Ask a Question

    Ask a specific question about this device

    K Number
    K060101

    Validate with FDA (Live)

    Date Cleared
    2006-05-10

    (118 days)

    Product Code
    Regulation Number
    870.4210
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ESTECH Aortic Easy Flow Cannula is intended for use in the perfusion of the aorta during cardiovascular surgery procedures reguiring extracorporeal cardiopulmonary bypass (CPB) up to 6 hours.

    Device Description

    The Estech Aortic Easy Flow Cannula is a sterile, single-use, device. The single-lumen polymer tube incorporates wire-wrapping with multiple perforations at the distal end to provide a diffuse flow pattern. The barbed proximal end is intended to connect into cardiopulmonary bypass tubing to provide extracorporeal circulation of the blood, most typically during stopped-heart surgical procedures. Each cannula is provided with a flexible obturator to assist with the placement and positioning of the cannula.

    AI/ML Overview

    The Estech Aortic Easy Flow Cannula is a medical device intended for use in cardiovascular surgery procedures requiring extracorporeal cardiopulmonary bypass (CPB) up to 6 hours. The documentation provided focuses on demonstrating substantial equivalence to predicate devices rather than presenting a standalone study with specific acceptance criteria and performance metrics for the new device based on clinical outcomes.

    Here's an analysis of the provided information:

    Acceptance Criteria and Reported Device Performance

    The provided text does not explicitly state specific quantitative acceptance criteria or a table of performance metrics for the ESTECH Aortic Easy Flow Cannula. Instead, the core of the submission relies on demonstrating substantial equivalence to predicate devices.

    The acceptance criteria implied are:

    • Biocompatibility: The device must meet the criteria for biocompatibility consistent with ISO 11193.
    • Labeled Performance Claims: The device must be able to meet its labeled performance claims.
    • Substantial Equivalence: The device must be demonstrated to be substantially equivalent to the predicate devices in terms of intended use, technology, design, materials, manufacture, and packaging.

    The reported device performance is:

    Acceptance CriteriaReported Device Performance
    Biocompatibility"The ESTECH Aortic Easy Flow Cannula met the criteria for biocompatibility."
    Labeled Performance Claims"Preclinical performance data was supplied to demonstrate that the ESTECH Aortic Easy Flow Cannula can meet its labeled performance claims."
    Substantial Equivalence"Estech concludes that the ESTECH Aortic Easy Flow Cannula is substantially equivalent to the predicate devices."

    Study Information

    Based on the provided text, the "study" conducted is a preclinical performance data assessment and biocompatibility testing to support the claim of substantial equivalence. This is typical for a 510(k) submission where a new device is compared to existing legally marketed devices. There is no mention of a human clinical trial or a study designed to quantify device performance in terms of clinical outcomes, sensitivity, specificity, accuracy, or similar metrics typically associated with AI/diagnostic device evaluations.

    Given this context, most of the requested information regarding sample size, expert involvement, and ground truth establishment (which are common for AI/diagnostic device studies) is not directly applicable or explicitly detailed in this 510(k) summary.

    1. Sample size used for the test set and the data provenance: Not explicitly stated as a separate "test set" in the context of a clinical trial. The "preclinical performance data" would involve bench testing or possibly animal studies, but no specific sample sizes or data provenance (country of origin, retrospective/prospective) are detailed in this summary.
    2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. The "ground truth" for this type of submission is typically based on established engineering principles, regulatory standards (like ISO 11193 for biocompatibility), and comparison to the known performance characteristics of the predicate devices. Expert consensus in the context of human data interpretation (e.g., radiology reads) is not relevant here.
    3. Adjudication method (e.g., 2+1, 3+1, none) for the test set: Not applicable. Adjudication methods are typically used for establishing ground truth in human subject studies, which are not detailed here.
    4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable. This device is a cardiopulmonary bypass vascular cannula, not an AI-powered diagnostic tool.
    5. If a standalone (i.e., algorithm only without human-in-the loop performance) was done: Not applicable. This is a physical medical device, not an algorithm.
    6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.): The "ground truth" for the preclinical performance data would be derived from:
      • Regulatory Standards: ISO 11193 for biocompatibility.
      • Engineering Specifications: Performance characteristics (e.g., flow rates, pressure drops, durability) measured against design specifications and predicate device performance.
      • Predicate Device Data: Performance characteristics of the legally marketed predicate devices serving as benchmarks.
    7. The sample size for the training set: Not applicable. This is not an AI/machine learning device that would require a "training set."
    8. How the ground truth for the training set was established: Not applicable.

    Conclusion

    The provided document details a 510(k) summary for a cardiopulmonary bypass vascular cannula, focusing on demonstrating substantial equivalence to predicate devices through biocompatibility testing and preclinical performance data. It does not contain the kind of detailed information about acceptance criteria, test sets, expert involvement, or ground truth establishment that would be relevant for an AI/diagnostic device performance study.

    Ask a Question

    Ask a specific question about this device

    K Number
    K053326

    Validate with FDA (Live)

    Date Cleared
    2006-01-06

    (36 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The Estech Cobra Cardiac Electrosurgical System is intended for the coaqulation of cardiac tissue using radiofrequency (RF) energy during cardiac surgery. The System can be used during general surgery to coagulate soft tissue. The System may also be used to coagulate blood and soft tissue to produce hemostasis.

    Device Description

    The Estech Cobra Cardiac Electrosurgical Unit and Estech Cable comprise a system that is identical to the Boston Scientific Cobra Cardiae Electrosurgical System. Both Systems are comprised of three components: the (radiofrequency) RF Probe, Flectrosurgical Unit (ESU) and Instrument (Cobra) Cable is an accessory to the ESC. The Estech RF Probes have been the subject of previous premarket notifications. The ESU is a software controlled high frequency electronic instrument, provided with controls for set temperature, power limit, and number of active electrodes. The ESU delivers 460 kHz of RF energy to selected Probe electrodes. The ESU measures temperatures from the Probe thermocouples and modulates the RF encrgy to keep all sclected electrodes' temperatures essentially the same; it adjusts the power output to maintain the maximum temperature of all selected electrodes close to the set point. The ESU has readouts for temperature, time of energy delivery, and delivered power. Front panel connectors include connections for the Instrument Cable, and third party dispersive or indifferent (DIP) electrodes. The Instrument Cable (Cobra Cable) connects the ESU to the RF Probe. It is supplied sterile to the user. The User can resterilize the Cable.

    AI/ML Overview

    This document describes a 510(k) submission for the Estech Cobra Cardiac Electrosurgical Unit and Cable. The submission argues for substantial equivalence to a predicate device, the Boston Scientific Cobra Cardiac Electrosurgical System (K013873), rather than presenting a study with specific acceptance criteria and performance data for the Estech device itself. Therefore, many of the requested categories are not applicable to this type of submission.

    Here's a breakdown based on the provided text:

    1. Table of Acceptance Criteria and Reported Device Performance

    Not applicable. The submission asserts identity to a predicate device, meaning it is considered to have the same performance characteristics as the predicate. No new performance data or acceptance criteria are provided or required for this type of 510(k).

    2. Sample Size Used for the Test Set and Data Provenance

    Not applicable. No new test set data was generated for the Estech device as it is asserted to be identical to the predicate.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Not applicable. See point 2.

    4. Adjudication Method for the Test Set

    Not applicable. See point 2.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and its effect size

    Not applicable. See point 2.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

    Not applicable. This is a hardware device (electrosurgical unit and cable), not an algorithm or AI system.

    7. The Type of Ground Truth Used

    Not applicable. See point 2. The "ground truth" for this submission is the established performance and safety profile of the predicate device.

    8. The Sample Size for the Training Set

    Not applicable. This is a hardware device, not an AI or machine learning system that would require a training set.

    9. How the Ground Truth for the Training Set was Established

    Not applicable. See point 8.


    Summary of the Acceptance Criteria and Study (as presented in the document):

    The Estech Cobra Cardiac Electrosurgical Unit and Cable passed acceptance criteria by demonstrating substantial equivalence to a previously cleared predicate device, the Boston Scientific Cobra Cardiac Electrosurgical System (K013873).

    The "study" in this context is the comparison of the new device to the predicate device, which concluded they are identical.

    Key points from the document regarding this "study":

    • Acceptance Criteria (Implicit): The primary acceptance criterion was that the Estech device be identical in intended use, technology, design, materials, manufacture, and packaging to the Boston Scientific Cobra Cardiac Electrosurgical System.
    • Reported Device Performance (Implicit): Since the device is stated to be identical to the predicate, its performance is assumed to be identical to the predicate device, which had already met FDA requirements.
    • Supporting Data: The submission explicitly states, "Supporting data is not necessary to support this submission since the Estech Cobra Cardiac FSU is identical to the predicate device..." This means no new tests, clinical trials, or performance studies were conducted for the Estech device. The FDA's 510(k) clearance confirms their agreement with this assessment of substantial equivalence.
    Ask a Question

    Ask a specific question about this device

    K Number
    K053100

    Validate with FDA (Live)

    Date Cleared
    2005-12-19

    (46 days)

    Product Code
    Regulation Number
    878.4400
    Age Range
    All
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The ESTECH Cobra Bipolar System is intended for the coagulation of soft tissue during general surgery. The System may also be used to coagulate blood and soft tissue to produce hemostasis.

    Device Description

    The Estech Cobra Bipolar System is comprised of two components: a reusable surgical clamp an associated coagulating device known as the Cobra Bipolar Insert. The Bipolar Insert is a sterile, single use device that transmits radiofrequency (RF) energy to coagulate soft tissues and fits in a jaw of the clamp. The Bipolar System will be used with the Boston Scientific Cobra Electrosurgical Unit (ESU). The ESU delivers 460 kHz of RF energy to selected Bipolar Insert electrodes, modulates the RF energy to keep all selected Bipolar Insert electrodes' temperatures essentially the same, and adjusts the power output to maintain the maximum temperature of all selected electrodes close to the set point.

    AI/ML Overview

    This document is a 510(k) Premarket Notification for the ESTECH Cobra Bipolar System, which is essentially a re-filing of an existing device (Boston Scientific Cobra Bipolar System, K023288) due to a transfer of sales and responsibility. Therefore, the information provided focuses on demonstrating substantial equivalence to the predicate device rather than presenting new performance data from a dedicated study.

    Based on the provided text, the acceptance criteria and study information are as follows:

    1. Table of Acceptance Criteria and Reported Device Performance

    Acceptance CriteriaReported Device Performance
    Intended UseIdentical to predicate device (Boston Scientific Cobra Bipolar System)
    TechnologyIdentical to predicate device
    DesignIdentical to predicate device
    MaterialsIdentical to predicate device
    ManufactureIdentical to predicate device
    PackagingIdentical to predicate device
    SterilizationIdentical to predicate device
    Safety & EffectivenessSubstantially equivalent to predicate device

    2. Sample Size Used for the Test Set and Data Provenance

    The document explicitly states that "Supporting data is not necessary to support this submission since the Estech Cobra Bipolar System is identical to the predicate device, the Boston Scientific Cobra Bipolar System (K023288)."

    Therefore, there was no dedicated test set or new performance data generated for this specific 510(k) submission. The data provenance relies on the prior approval of the predicate device.

    3. Number of Experts Used to Establish Ground Truth for the Test Set and Their Qualifications

    Since no new test set was used, no experts were specifically used to establish ground truth for this submission. The assumption is that the predicate device's original approval relied on appropriate expert evaluations as part of its regulatory process.

    4. Adjudication Method for the Test Set

    As there was no new test set, there was no adjudication method employed for this submission.

    5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

    No MRMC comparative effectiveness study was done for this submission. The device is claimed to be identical to the predicate, implying similar performance.

    6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

    This device is an electrosurgical probe, not a software algorithm. Therefore, a standalone algorithm performance study is not applicable and was not performed.

    7. The Type of Ground Truth Used

    The ground truth for this submission is based on the established performance and safety of the predicate device (Boston Scientific Cobra Bipolar System, K023288), which was previously cleared by the FDA. The submission asserts that the new device is identical in all relevant aspects, hence its functionality and safety are equivalent to the pre-existing device.

    8. The Sample Size for the Training Set

    Since this submission relies on demonstrating substantial equivalence to an identical predicate device and no new performance studies were conducted, there was no new training set used or described for this particular submission.

    9. How the Ground Truth for the Training Set Was Established

    As no training set was used for this submission, there is no description of how its ground truth was established. The regulatory burden was met by declaring the new device identical to the predicate.

    Ask a Question

    Ask a specific question about this device

    Page 1 of 1